Operating Manual
`
`OPERATING MANUAL
`BLU-U® Blue Light Photodynamic Therapy Illuminator
`Model 4170
`When Printed This Document is "For Reference Use Only"
`
`Operating Manual
`
`
`
`BLU-U® Blue Light Photodynamic Therapy
`Illuminator Model 4170
`
`
`
`
` MAN-1211 Rev B
`
`

`

`OPERATING MANUAL
`BLU-U® Blue Light Photodynamic Therapy Illuminator
`Model 4170
`When Printed This Document is "For Reference Use Only"
`Table of Contents
`
`3
`SYMBOLS AND DEFINITIONS:…………………………………………………………………………
`4
`INDICATIONS FOR USE: ………………………………………………………………………………
`4
`CAUTIONS AND WARNINGS: …………………………………………………………………………
`5
`PRODUCT SPECIFICATIONS: …………………………………………………………………………
`5
`DESCRIPTION: ………………………………………………………………………………………..
`6
`CONTROLS: …………………………………………………………………………………………..
`MAIN POWER SWITCH: …………………………………………………………………………….. 6
`6
`KEY SWITCH: ………………………………………………………………………………………
`6
`TIMER: …………………………………………………………………………………………….
`INDICATOR LIGHTS: …………………………………………………………………………………. 7
`INSTRUCTIONS FOR USE WITH THE LEVULAN® KERASTICK® FOR THE TREATMENT OF ACTINIC
`9
`KERATOSES: ………………………………………………………………………………………….
`INITIAL SET-UP: ……………………………………………………………………………………. 9
`SET-UP: …………………………………………………………………………………………… 9
`9
`PROCEDURE FOR TREATING LESIONS ON THE FACE: …………………………………………....
`FOLLOWING PATIENT TREATMENT: ………………………………………………………………. 10
`PROCEDURE FOR TREATING LESIONS ON THE SCALP: ………………………………………….. 10
`FOLLOWING PATIENT TREATMENT: ……………………………………………………………….. 11
`INSTRUCTIONS FOR USE FOR THE LIGHT ONLY TREATMENT OF ACNE: …………………………….. 12
`INITIAL SET-UP: …………………………………………………………………………………… 12
`SET-UP: …………………………………………………………………………………………… 12
`PROCEDURE FOR TREATING ACNE: ………………………………………………………………. 12
`FOLLOWING PATIENT TREATMENT: ………………………………………………………………. 13
`TROUBLESHOOTING/SERVICE AND REPAIR: ………………………………………………………… 14
`TRANSPORT AND STORAGE: …………………………………………………………………………. 16
`CLEANING/DISINFECTING: ……………………………………………………………………………. 16
`CUSTOMER SERVICE: ………………………………………………………………………………… 17
`FOR SERVICE, REPAIR, OR CALIBRATION OF THIS EQUIPMENT CALL: ……………………………. 17
`WARRANTY COVERAGE AND DISCLAIMERS: ……………………………………………………… 17
`
`
`
`
`
`
`
`
`Page 2
`
` MAN-1211 Rev B
`
`

`

`OPERATING MANUAL
`BLU-U® Blue Light Photodynamic Therapy Illuminator
`Model 4170
`When Printed This Document is "For Reference Use Only"
`Symbols and Definitions:
`
`
`
`!
`
`
`Reference User Manual
`
`T
`5AL 250V
`Indicates the Electrical Specification and
`Location of the Required Fuses
`
`
`
`60 oC
`140 oF
`
`-20 oC
`-4 oF
`
`
`Indicates the Upper and Lower
`Temperatures Allowed for Storage and
`Transport
`
`
`Indicates Need to Wear Eye Protection
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Warning/Caution
`
`This Equipment Conforms to the IEC
`Publication 878-02-02 Type B
`Classification
`
`0-95%
`
`
`Indicates the % Humidity Allowed for
`Storage and Transport
`
`
`
`
`
`Page 3
`
` MAN-1211 Rev B
`
`

`

`OPERATING MANUAL
`BLU-U® Blue Light Photodynamic Therapy Illuminator
`Model 4170
`When Printed This Document is "For Reference Use Only"
`
`Indications for Use:
`DUSA Pharmaceuticals, Inc.® BLU-U® Blue Light Photodynamic Therapy Illuminator Model
`4170, in combination with the Levulan® Kerastick® (aminolevulinic acid HCl) for Topical
`Solution, 20%, is indicated for the treatment of non-hyperkeratotic actinic keratoses (AK) of
`the face or scalp.
`The BLU-U® Blue Light Photodynamic Therapy Illuminator Model 4170 is intended to provide
`phototherapeutic light to the body. The BLU-U® 4170 is generally indicated to treat
`dermatological indications. The BLU-U® 4170 is specifically indicated to treat moderate
`inflammatory acne vulgaris.
`
`Cautions and Warnings:
`
`
`
`WARNING:
`The BLU-U® Blue Light Photodynamic Therapy Illuminator, in combination with the Levulan®
`Kerastick® for Topical Solution, 20% is indicated for the treatment of non-hyperkeratotic
`actinic keratoses of the face or scalp. Do not use this device with other photosensitizing
`drugs. Refer to the Levulan® Kerastick® for Topical Solution, 20% Package Physician Insert
`for additional information.
`When using the BLU-U® for acne, do not use this device with photosensitizing drugs.
`WARNING:
`All personnel should read and understand the instructions in this manual before the system
`is used. Failure to do so may result in improper operation of the system.
`WARNING:
`Use only eyewear which blocks light with wavelengths of at least 500nm and shorter with an
`Optical Density (OD) of two or greater.
`
`WARNING:
`To avoid risk of electrical shock, connect the power cord to a properly wired grounding
`receptacle only.
`
`WARNING:
`Never attempt to service the device when it is connected to a power source. Hazardous
`voltages inside the device may cause severe electrical shock. Disconnect the power cord
`before servicing.
`
`WARNING:
`Do not allow fluids to enter the device. Damage to the device may result.
`CAUTION:
`The device should not be serviced or opened except by qualified service technicians.
`Tampering by unqualified persons may cause damage to the unit or personal injury.
`CAUTION:
`Federal law restricts this device to sale by or on the order of a physician or licensed medical
`practitioner.
`
`
`
`Page 4
`
` MAN-1211 Rev B
`
`

`

`OPERATING MANUAL
`BLU-U® Blue Light Photodynamic Therapy Illuminator
`Model 4170
`When Printed This Document is "For Reference Use Only"
`
`Product Specifications:
`The BLU-U® is an electrical Class I, Type B device designed for indoor use only. It has
`been tested to the following standards: UL2606.1, CAN/CSA C22.2 No.601.1, EN60601-1,
`and EN 60601-1-2. General specifications are listed below.
`Power cord
`3-conductor hospital grade electrical cord
`Power requirements
`117 VAC, 2.5 Amps, 60 Hz
`Footprint (light unit + stand)
`Width: 99.70 cm (39.25 inches)
`Depth: 67.31 cm (26.5 inches)
`Height Minimum: 127 cm (50 inches)
`Height Maximum: 160 cm (63 inches)
`Overall dimensions of the light unit Width: 52.71 cm (20.75 inches)
`Depth: 47.63 cm (18.75 inches)
`Height: 38.74 cm (15.25 inches)
`72.6 kg (160 lbs.)
`20 – 30o C (68 – 86o F)
`
`Weight (light unit + stand)
`Operating Temperature Range
`
`
`Description:
`The BLU-U® is a compact light source designed to provide a uniform distribution of blue
`light to areas of the patient’s face or scalp for the use stated above. It is comprised of 7
`horizontally mounted U-shaped fluorescent tubes within a plastic chassis. The tubes are
`covered by a polycarbonate shield, which directs cooling airflow within the unit and
`significantly minimizes the risk of glass-patient contact in the event of tube breakage.
`The BLU-U® is mounted on a floor-stand, which permits rapid positioning as well as
`adjustment for patient height. The control panel is also affixed to the floor stand.
`The BLU-U® has a built-in power output monitoring and diagnostic system, which
`illuminates a neon light to inform the user of the system’s status.
`The BLU-U® has a system timer used to set the light dose delivered to the patient.
`The BLU-U® is rated for short-time operation.
`
`Page 5
`
` MAN-1211 Rev B
`
`

`

`OPERATING MANUAL
`BLU-U® Blue Light Photodynamic Therapy Illuminator
`Model 4170
`When Printed This Document is "For Reference Use Only"
`
`Controls:
`Controls for the BLU-U® are located on the floor stand.
`
`Main Power Switch:
`The Main Power Switch is a two-position rocker switch, labeled “l” and “0,” located next to
`the electrical cord.
` l
`=
`ON
`0
`=
`OFF
`Push the Main Power Switch to “l” to turn on power to the system.
`Push the Main Power Switch to “0” to disconnect all electrical components within the
`BLU-U® from the AC line.
`
`
`
`Key Switch:
`The Key Switch is the “ON/OFF” switch for the BLU-U® and requires a special key to
`operate.
`(cid:206) Remove the key and store it securely whenever the unit is not in use, to prevent
`unauthorized use of the BLU-U®.
`Turn the Key Switch to “l” to turn on power to the BLU-U® control electronics. This
`activates the Timer so that the prescribed exposure time can be entered. When activated,
`the timer will remember and display the last treatment time setting.
`Note:
`When the Key Switch is turned off, it should not be turned back on again for at least
`thirty (30) seconds to ensure that the control electronics have properly powered
`down and reset.
`
`Timer:
`The system Timer is used to control the operation of the fluorescent tubes. Use the Timer
`to:
`• Set the exposure time,
`Initiate light exposure,
`•
`and after the set exposure time has elapsed,
`• Automatically turn off the tubes.
`
`Page 6
`
` MAN-1211 Rev B
`
`

`

`OPERATING MANUAL
`BLU-U® Blue Light Photodynamic Therapy Illuminator
`Model 4170
`When Printed This Document is "For Reference Use Only"
`The following buttons control operation of the Timer:
`
`FUNCTION / OPERATION
`
`BUTTON
`Time Select
`
` (cid:199) (cid:200)
`
`Start/Stop
`
`
`The Time Select buttons are used to set the exposure time.
`Depress the (cid:199) (up arrow) button to increase time.
`Depress the (cid:200) (down arrow) button to decrease time.
`When first depressed, these buttons change the displayed reading
`slowly; if they remained depressed, the display changes quickly.
`Depressing and releasing these buttons quickly makes small
`adjustments to the displayed time.
`The Start/Stop button toggles between the running and stopped
`states of the Timer and Tubes.
`After the exposure time has been set, depress this button once to
`turn on the tubes and initiate the Timer countdown sequence.
`Depress it a second time to turn off the tubes and stop the Timer.
`
`Note:
` If the light treatment for actinic keratoses of the face or scalp is interrupted, see
`the Levulan® Kerastick® for Topical Solution, 20% package insert for further
`details before proceeding.
`If the light treatment for acne is interrupted, reset and continue treatment.
`Indicator Lights:
`Indicator lights for the BLU-U® are located on the control panel on the floor stand.
`
`INDICATOR
`System Status
`Indicator
`
`
`FUNCTION / OPERATION
`The red neon light, located to the right of the Timer, indicates
`system status. At the beginning of each light treatment, the System
`Status Indicator flashes three (3) times to indicate that the system
`control electronics and the neon light are functioning normally, and
`that the BLU-U® is ready for use.
`
`If System Status Indicator light fails to flash three (3) times immediately after the initiation
`of timed light treatment, the BLU-U® should not be used until the problem has been
`identified or a qualified service technician has serviced the unit.
`(See the Troubleshooting Table)
`
`If a patient has been dosed with the Levulan® Kerastick® for Topical Solution, 20%, for the
`treatment of actinic keratoses of the face or scalp, see the Levulan® Kerastick® for Topical
`Solution, 20% Package Physician Insert for further instructions on early termination or
`cancellation of light treatment.
`If the System Status Indicator lights at any other time, refer to Table 1 on the
`following page.
`
`Page 7
`
` MAN-1211 Rev B
`
`

`

`OPERATING MANUAL
`BLU-U® Blue Light Photodynamic Therapy Illuminator
`Model 4170
`When Printed This Document is "For Reference Use Only"
`System Status Indicator Error Conditions:
`
`
`Table 1
`
`CONDITION
`Rapid Flashing
`Continuous rapid flashing of the
`System Status Indicator immediately
`after initiation of the timed light
`treatment indicates a problem with the
`electronic control system.
`If this happens, the BLU-U® will not be
`operational and will not light.
`
`Slow Flashing
`Continuous slow flashing of the
`System Status Indicator (3 flashes
`every 4 seconds) after initiation of the
`timed light treatment indicates that
`either:
`• The BLU-U® output power is too
`high, or
`• A problem exists with the BLU-U®
`electronic control system.
`
`Steady On
`The System Status Indicator lights
`steadily during the treatment (for at
`least 10 seconds at a time)
`This indicator code, which may occur
`after initiation of the timed light
`treatment, indicates that:
`• The BLU-U® output power is too
`low, or
`• The end of tube lifetime has been
`reached.
`Exposure Time Indicator
`
`
`ACTION
`
`
`Discontinue the treatment.
`Turn the Key Switch and the Main Power Switch to
`the “0” (off) position, and call customer service.
`If the patient has been dosed with the Levulan®
`Kerastick® for Topical Solution, 20%, see the
`Levulan® Kerastick® for Topical Solution, 20%
`Package Physician Insert for further instructions on
`early termination or cancellation of light treatment.
`
`
`Discontinue the treatment.
`Turn the Key Switch and the Main Power Switch to
`the “0” (off) position, and call DUSA’s customer
`service department.
`If the patient has been dosed with the Levulan®
`Kerastick® for Topical Solution, 20%, see the
`Levulan® Kerastick® for Topical Solution, 20%
`Package Physician Insert for further instructions on
`early termination or cancellation of light treatment.
`
`• Complete the treatments of any patients who have
`already been dosed with the Levulan® Kerastick®,
`and
`• Call customer service.
`
`This four-digit red LED located on the Timer unit
`displays the remaining exposure time in minutes and
`seconds. Prior to pushing the Start button to begin
`light exposure, the display indicates the amount of
`exposure time set.
`When you press the Start button, the Exposure Time
`Indicator display counts down the amount of exposure
`time remaining. The tubes turn off automatically when
`the display reaches “00:00”.
`
`Page 8
`
` MAN-1211 Rev B
`
`

`

`OPERATING MANUAL
`BLU-U® Blue Light Photodynamic Therapy Illuminator
`Model 4170
`When Printed This Document is "For Reference Use Only"
`
`Instructions for Use with the Levulan® Kerastick® for the treatment of Actinic
`Keratoses:
`Note - If a patient’s light treatment is interrupted, terminated prematurely, or cannot
`be administered, see the Levulan® Kerastick® for Topical Solution, 20% Package
`Physician Insert for important further instructions.
`
`Initial Set-Up:
`1.
`Inspect the unit power fuse holder to ensure the correct line voltage has been selected.
`2. Plug the female end of the supplied electrical cord into the mating jack on the floor
`stand base and plug the other end into a standard 120∼ outlet.
`3. Press the Main Power Switch to the “l” (on) position.
`
`Set-Up:
`1. Using the key, turn the Key Switch to the “l” (on) position. Verify that the red Timer
`display is active.
`2. Position the eye protection on the patient prior to treatment. Place the eye protection
`over the patient’s eyes and insure the eye protection is secure against the patient’s
`face. Verify that the eye protection does not cover or shadow any area intended for
`treatment.
`3. Place the patient in an upright, sitting position. The procedure for positioning the
`BLU-U® depends on the location of the lesions to be treated and is found in the
`following 2 sections. The patient’s head may be supported during treatment. Ensure
`that the method of head support does not cover or shadow any area intended for
`treatment.
`
`Procedure for Treating Lesions on the Face:
`1. Loosen the knobs on either side of the light unit and rotate it to the vertical position (the
`“U” shaped bulbs stacked vertically). Retighten the knobs to lock the light unit in place.
`2. Position the BLU-U® around the patient’s head so the entire surface area to be treated
`lies between 2” and 4” from the BLU-U® surface:
`a.) The patient’s nose should be no closer than 2” from the surface
`
`and
`b.) The patient’s forehead and cheeks should be no further than 4” from the surface
`
`and
`c.) The sides of the patient’s face and the patient’s ears should be no closer than 2”
`from the BLU-U® surface
`
`Note:
`The patient’s hair should not cover or shadow the area to be treated.
`However, the patient’s hair may be closer than 2” to the surface of the
`BLU-U® without any deleterious effects.
`
`Page 9
`
` MAN-1211 Rev B
`
`

`

`4.
`
`OPERATING MANUAL
`BLU-U® Blue Light Photodynamic Therapy Illuminator
`Model 4170
`When Printed This Document is "For Reference Use Only"
`3. Set the Timer to the prescribed treatment time of 16 minutes 40 seconds by
`depressing the Time Select buttons. Continue to depress these buttons until the correct
`time is displayed.
`Insure that all personnel are wearing appropriate eye protection and then depress the
`Start/Stop switch on the BLU-U® Timer. The System Status Indicator will flash three
`(3) times and go off. If it does not flash three times, try the remedies in the
`Troubleshooting Table. If the System Status Indicator still does not flash three
`times, do not use the BLU-U® even if the Timer works and the tubes light; system
`output may be incorrect under these circumstances. [See the Levulan® Kerastick® for
`Topical Solution, 20% Package Physician Insert for further instructions on cancellation
`of light treatment.]
`5. Verify that all seven tubes are lit. If one or more does not light, discontinue the light
`treatment and call for service. See the Levulan® Kerastick® for Topical Solution, 20%
`Package Physician Insert for further details on early termination or cancellation of light
`treatment.
`6. Periodically check the System Status Indicator. If the System Status Indicator lights
`during treatment, see “Indicator Lights” in the Controls section.
`7. Take care that the patient does not move during the time the BLU-U® is on as this may
`result in under exposure of the lesion(s).
`8. At the end of the treatment period, the Timer will automatically turn off the BLU-U®.
`
`Following Patient Treatment:
`1. Remove the patient from the BLU-U® and remove the patient’s eye protection.
`2. Turn the Key Switch on the BLU-U® to the “0” position
`3. Remove the key from the BLU-U® and store it in a secure location where unauthorized
`personnel cannot use it.
`
`Procedure for Treating Lesions on the Scalp:
`1. Loosen the knobs on either side of the light unit and rotate it to the horizontal position
`(the “U” shaped bulbs stacked horizontally). Retighten the knobs to lock the light unit in
`place.
`2. Position the BLU-U® around the patient’s head so the entire surface area to be treated
`lies between 2” and 4” from the BLU-U® surface:
`a.) The patient’s scalp should be no closer than 2” from the surface
`
`and
`b.) The patient’s scalp should be no further than 4” from the surface
`
`and
`c.) The sides of the patient’s face and the patient’s ears should be no closer than 2”
`from the BLU-U® surface
`
`Note:
`The patient’s hair should not cover or shadow the area to be treated.
`However, the patient’s hair may be closer than 2” to the surface of the
` BLU-U® without any deleterious effects.
`
`Page 10
`
` MAN-1211 Rev B
`
`

`

`4.
`
`OPERATING MANUAL
`BLU-U® Blue Light Photodynamic Therapy Illuminator
`Model 4170
`When Printed This Document is "For Reference Use Only"
`3. Set the Timer to the prescribed treatment time of 16 minutes 40 seconds by
`depressing the Time Select buttons. Continue to depress these buttons until the correct
`time is displayed.
`Insure that all personnel are wearing appropriate eye protection and then depress the
`Start/Stop switch on the BLU-U® Timer. The System Status Indicator will flash three
`(3) times and goes off. If it does not flash three times, try the remedies in the
`Troubleshooting Table. If the System Status Indicator still does not flash three
`times, do not use the BLU-U® even if the Timer works and the tubes light; system
`output may be incorrect under these circumstances. [See the Levulan® Kerastick® for
`Topical Solution, 20% Package Physician Insert for further instructions on cancellation
`of light treatment.]
`5. Verify that all seven tubes are lit. If one or more does not light, discontinue the light
`treatment and call customer service. See the Levulan® Kerastick® for Topical Solution,
`20% Package Physician Insert for further details on early termination or cancellation of
`light treatment.
`6. Periodically check the System Status Indicator. If the System Status Indicator lights
`during treatment, see “Indicator Lights” in the Controls section.
`7. Take care that the patient does not move during the time the BLU-U® is on as this may
`result in under exposure of the lesion(s).
`8. At the end of the treatment period, the Timer will automatically turn off the BLU-U®.
`
`Following patient treatment:
`1. Remove the patient from the BLU-U® and remove the patient’s eye protection.
`2. Turn the Key Switch on the BLU-U® to the “0” position.
`3. Remove the key from the BLU-U® and store it in a secure location where unauthorized
`personnel cannot use it.
`
`Page 11
`
` MAN-1211 Rev B
`
`

`

`OPERATING MANUAL
`BLU-U® Blue Light Photodynamic Therapy Illuminator
`Model 4170
`When Printed This Document is "For Reference Use Only"
`
`Instructions for Use for the Light Only treatment of Acne:
`Initial Set-Up:
`1.
`Inspect the unit power fuse holder to ensure the correct line voltage has been selected.
`2. Plug the female end of the supplied electrical cord into the mating jack on the floor
`stand base and plug the other end into a standard 120 ∼ outlet.
`3. Press the Main Power Switch to the “l” (on) position.
`
`Set-Up:
`1. Using the key, turn the Key Switch to the “l” (on) position. Verify that the red Timer
`display is active.
`2. Position the eye protection on the patient prior to treatment. Place the eye protection
`over the patient’s eyes and insure the eye protection is secure against the patient’s
`face. Verify that the eye protection does not cover or shadow any area intended for
`treatment.
`3. Place the patient in an upright, sitting position. The patient’s head may be supported
`during treatment. Ensure that the method of head support does not cover or shadow
`any area intended for treatment.
`
`Procedure for treating Acne:
`1. Loosen the knobs on either side of the light unit and rotate it to the vertical position (the
`“U” shaped bulbs stacked vertically). Retighten the knobs to lock the light unit in place.
`2. For treatment of the face, position the BLU-U® around the patient’s head so the entire
`surface area to be treated lies between 2” and 4” from the BLU-U® surface:
`a.) The patient’s nose should be no closer than 2” from the surface
`
`and
`b.) The patient’s forehead and cheeks should be no further than 4” from the surface
`
`and
`c.) The sides of the patient’s face and the patient’s ears should be no closer than 2”
`from the BLU-U® surface
`
`Note:
`The patient’s hair should not cover or shadow the area to be treated.
`However, the patient’s hair may be closer than 2” to the surface of the
`BLU-U® without any deleterious effects.
`3. Set the Timer for the desired treatment time by depressing the Time Select buttons.
`Continue to depress these buttons until the correct time is displayed. The
`recommended exposure time is 16 minutes and 40 seconds per treatment (10
`joules/cm²). Light treatments should be performed two to three times per week until the
`desired clinical results have been achieved with a maximum recommended total
`cumulative light exposure of 320 joules/cm². Insure that all personnel are wearing
`appropriate eye protection and then depress the Start/Stop switch on the BLU-U®
`Timer. The System Status Indicator will flash three (3) times and go off. If it does
`not flash three times, try the remedies in the Troubleshooting Table. If the System
`
`Page 12
`
` MAN-1211 Rev B
`
`

`

`OPERATING MANUAL
`BLU-U® Blue Light Photodynamic Therapy Illuminator
`Model 4170
`When Printed This Document is "For Reference Use Only"
`Status Indicator still does not flash three times, do not use the BLU-U® even if the
`Timer works and the tubes light; system output may be incorrect under these
`circumstances.
`4. Verify that all seven tubes are lit. If one or more does not light, discontinue the light
`treatment and call for service.
`5. Periodically check the System Status Indicator. If the System Status Indicator lights
`during treatment, see “Indicator Lights” in the Controls section.
`6. Take care to minimize patient movement during the time the BLU-U® is in use.
`7. At the end of the treatment period, the Timer will automatically turn off the BLU-U®.
`
`Following Patient Treatment:
`1. Remove the patient from the BLU-U® and remove the patient’s eye protection.
`2. Turn the Key Switch on the BLU-U® to the “0” position
`3. Remove the key from the BLU-U® and store it in a secure location where unauthorized
`personnel cannot use it.
`
`Page 13
`
` MAN-1211 Rev B
`
`

`

`OPERATING MANUAL
`BLU-U® Blue Light Photodynamic Therapy Illuminator
`Model 4170
`When Printed This Document is "For Reference Use Only"
`
`Troubleshooting/Service and Repair:
`
`
`
`There are no service parts in the equipment.
`
`
`
`
`
`Components of the system should not be opened, except by a Qualified Service Person.
`Tampering by Unqualified Persons can harm the person and/or damage the unit.
`
`CAUTION:
`
`The following chart has been included to assist in determining a solution for a problem or error
`Symptom
`Possible Cause
`What To Do
`BLU-U® is not
`No power /
`Verify that the proper line voltage has been
`selected on the fuse holder.
`plugged in.
`Fans not running /
`Verify that the BLU-U® is plugged into a standard
`No Timer display upon
`120 ∼ wall outlet.
`turning Key Switch to the
`“l” (on) position.
`Verify that power is present at the outlet.
`
`
`Verify that the Main Power Switch is in the “l” (on)
`position.
`Verify that the Key Switch is in the “l” (on)
`position by rotating it clockwise 1/4 turn until a
`“click” is felt.
`If the fans now run, but the Timer still does not
`light or lights intermittently, there is an internal
`electrical fault. In this case use of the BLU-U® will
`not be possible. Call for service.
`Check the fuses in the Fused Power Entry
`Module, located next to the socket for the
`electrical cord on the base of the floor stand.
`• Turn the Key Switch and Main Power Switch
`to the “0” (off) position
`• Unplug the unit
`• With a small screwdriver, slide out the fuse
`holder
`• Check the status of the two fuses by using the
`fuse key. If either or both fuses are blown (as
`indicated by a break in the thin wire
`connecting the two metal ends of the fuse),
`replace with 5A fuses (0.5 x 20 mm slo-blo,
`Bussman # GMC-5A or equivalent).
`• Reinsert the fuse holder
`• Plug the unit into a standard 120 ∼ wall outlet
`• Turn the Main Power Switch and Key
`Switch to the “l” (on) position.
`
`
`
`
`
`No power is present
`at the wall outlet.
`Main Power Switch
`is not set to “l” (on).
`Key Switch is not
`fully turned to “l” (on).
`
`One or more fuses in
`the Fused Power
`Entry Module have
`blown.
`
`Page 14
`
` MAN-1211 Rev B
`
`

`

`
`
`Symptom
`
`
`System Status Indicator
`does not flash three (3)
`times when the Start/Stop
`button on the Timer is
`pressed
`System Status Indicator
`rapidly flashing
`Tubes not lit
`System Status Indicator
`slowly flashing
`Tubes lit
`
`System Status Indicator
`on steady or intermittently
`Tubes lit
`Tubes not all lit
`
`F001 Error Code
`Displayed on Timer
`F101 Error Code
`Displayed on Timer
`F002 Error Code
`Displayed on Timer
`F202 Error Code
`Displayed on Timer
`F303 Error Code
`Displayed on Timer
`
`
`OPERATING MANUAL
`BLU-U® Blue Light Photodynamic Therapy Illuminator
`Model 4170
`When Printed This Document is "For Reference Use Only"
`Possible Cause What To Do
`
`If normal operation does not resume or the fuse
`continues to blow, there is an internal electrical
`fault. In this case use of the BLU-U® will not be
`possible.
`Call for service.
`Use of the BLU-U® will not be possible.
`Call for service.
`Discontinue use of the BLU-U®.
`Call for service.
`
`Internal electrical
`fault.
`
`Neon light or control
`circuitry is not
`functioning properly.
`
`Control circuitry is not
`functioning properly.
`
`Use of the BLU-U® will not be possible.
`Call for service.
`
`Power output is
`above specified
`range.
`Control circuitry is not
`functioning properly.
`
`Power below
`specified range.
`
`Discontinue use of the BLU-U®.
`Call for service
`
`Discontinue use of the BLU-U®.
`Call for service.
`Complete treatment of patients.
`Call for service.
`
`Tube(s) cracked or
`broken.
`
`Control circuitry is
`not functioning
`properly.
`Timer error
`
`Timer error
`
`Timer error
`
`Timer error
`
`Timer error
`
`Discontinue use of the BLU-U®.
`Call for service.
`Discontinue use of the BLU-U®.
`Call for service.
`
`With the Key Switch turned to “l” (on), press the
`Start/Stop button to clear the Timer display.
`Call customer service to receive further
`instructions.
`With the Key Switch turned to “l” (on), press the
`Start/Stop button to clear the Timer display.
`Call customer service to receive further
`instructions.
`Call customer service to receive further
`instructions.
`
`Page 15
`
` MAN-1211 Rev B
`
`

`

`OPERATING MANUAL
`BLU-U® Blue Light Photodynamic Therapy Illuminator
`Model 4170
`When Printed This Document is "For Reference Use Only"
`
`Transport and Storage:
`
`1. The BLU-U® can be safely stored in a cool dry area. Care should be taken to avoid
`rough handling or jarring of the unit.
`
`•
`
`2. Storage Conditions:
`-20°C to 60°C
`0 to 95% RH, Non-Condensing
`500hPa to 1600hPa
`
`•
`
`•
`
`Cleaning/Disinfecting:
`
`
`
`WARNING:
`
`Turn the power off and disconnect the power cord before cleaning the machine.
`Failure to do so may result in severe electrical shock or death.
`
`CAUTION:
`
`Never immerse machine in liquids. Do not use abrasive materials to clean the machine.
`Do not allow water to enter this device. Do not clean the inside of this device. Doing so
`will cause damage to this machine.
`
`1. The exterior surface of the BLU-U® may be wiped down with a mild disinfectant or
`isopropyl alcohol. Dry with a clean dry cloth.
`
`2. The outside surface of the plastic shield may be wiped down with a mild disinfectant or
`isopropyl alcohol. Dry with a clean dry cloth.
`
`3.
`
`If goggles are used for eye protection, their surface may be wiped down with a mild
`disinfectant or isopropyl alcohol after each use. Dry with a clean dry cloth.
`
`Page 16
`
` MAN-1211 Rev B
`
`

`

`OPERATING MANUAL
`BLU-U® Blue Light Photodynamic Therapy Illuminator
`Model 4170
`When Printed This Document is "For Reference Use Only"
`
`
`
`Customer Service:
`For service, repair, or calibration of this equipment call:
`
`DUSA Pharmaceuticals, Inc.®
`Phone: 1-877-533-3872
`or
`978-657-7500
`Warranty Coverage and Disclaimers:
`See the Terms and Conditions of your contract for specific information.
`Manufactured For:
`DUSA Pharmaceuticals, Inc.®
`25 Upton Drive
`Wilmington, MA 01887
`www.dusapharma.com
`
`Page 17
`
` MAN-1211 Rev B
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket