throbber
Medication Nebulizer Performance*
`Effects Of Diluent Volume, Nebulizer Flow, and
`Nebulizer Brand
`
`Dean Hess, PhD, RRT; Daniel Fisher, BS, RRT; Purris Williams, BS, RRT;
`Sharon Pooler, RRT; and Robert M. Kacmarek, PhD, RRT
`
`Background: Medication nebulizers are commonly used to delivery aerosolized medications to pa(cid:173)
`tients with respiratory disease. We evaluated output and respirable aerosol available to the patient
`(inbaled mass) for I7 medication nebulizers using a spontaneous breathing lung model.
`Methods: Three nebulizer fill volumes (3, 4, and 5 mL containing 2.5 mg of albuterol) and 3 oxygen
`flows (6, 8, and 10 Umin) were evaluated using the I7 nebulizers. A cotton plug at the nebulizer
`mouthpiece was used to trap aerosol during simulated spontaneous breathing. Following each trial,
`the amount of albuterol remaining in the nebulizer and the amount deposited in the cotton plug were
`determined spectrophotometrically. Aerosol particle size was determined using an ll-stage cascade
`impactor.
`Results: Increasing fill volume decreased the amount of albuterol trapped in the dead volume
`(p<O.OOI) and increased the amount delivered to the patient (p<O.OOI). Increasing flow increased
`the mass output of particles in the respirable range of I to 5 pm (p=0.004), but the respirable mass
`delivered to the patient was affected to a greater extent by nebulizer brand (p<O.OOI) than flow.
`Although 2.5 mg of albuterol was placed into the nebulizers, less than 0.5 mg in the respirable range
`of I to 5 pm was delivered to the mouthpiece.
`Conclusions: The performance of medication nebulizers is affected by fill volume, flow, and nebu(cid:173)
`lizer brand. When they are used for research applications, the nebulizer characteristics must be
`evaluated and reported for the conditions used in the investigation. (CHEST 1996; 110:498-505)
`
`Key words: aerosol therapy; inhaled bronchodilator administration; nebulizers
`
`Abbreviations: GSD =geometric standard deviation; MMAD =mass median aerodynamic diameter
`
`D espite the common use of metered-dose inhalers
`and the availability of dry powder inhalers, aero(cid:173)
`solized medications are still frequently administered by
`nebulizer. Nebulizers are commonly used for inhaled
`
`For editorial comment see page 316
`
`bronchodilator administration to patients with reactive
`airways, including the perioperative and postoperative
`treatment of these patients. Advantages of nebulizers
`include the ability to use tl1em with patients who can(cid:173)
`not coordinate the use of a m etered-dose inhaler1 and the
`ability to conveniently administer a large (or continuous)
`
`*From the Department of Respiratory Care, Massachusetts Gen(cid:173)
`eral HosP.ital , and Harvard M edical School, Boston.
`Presentea, in part, at the annual meeting of the American Associ(cid:173)
`ation for Respiratory Care, Las Vegas, December 1994.
`Supported, in part, by Puritan-Bennett, Hudson RCI, Marquest,
`Professional M edical, SIMS.
`Manuscript received October 20, 1995; revision accepted F ebruary
`26, 1996.
`Reprint requests: Dr. Hess, Respiratory Care, Ellison 401 , Massa(cid:173)
`chusetts General Hospital, Boston, MA 02114
`
`dose into the lungs.2 Important characteristics of nebu(cid:173)
`lizer p erformance include the drug output, the aerosol
`particle size generated, the nebulization time, and the
`amount of drug delivered to the patient. Factors that have
`been shown to affect nebulizer performance include de(cid:173)
`vice construction (ie, manufacturer), fill volume, flow,
`temperature, and humidity of the driving gas.1
`A common feature of nebulizers is dead volume,
`which i s the volume of solution that remains in the
`nebulizer cup after aerosol production ends . Previous
`studies h ave typically evaluated dead volume by serial
`weighing.3-5 This method does not adequately charac(cid:173)
`terize drug output and amount of drug in the dead vol(cid:173)
`ume due to reconcentration in the nebulizer cup.6•7 He(cid:173)
`concentration occurs because of evaporation owning to
`the low relative humidity of the gas powering the nebu(cid:173)
`lizer. Nebulizer output should be determined more ap(cid:173)
`propriately by measuring the amount of medication that
`remains after a erosol production is complete.
`Particle size is an important characteristic of nebu(cid:173)
`lizer performance. Particles too large do not reach the
`
`498
`
`Laboratory and Animal Investigations
`
`IPR2021-00406
`United Therapeutics EX2031
`Page 1 of 8
`
`

`

`A
`
`o,
`
`ring
`stand
`
`c:
`"' E
`t:
`~
`-E
`"'o
`~()
`
`c: .. E
`t: """
`
`c:c.
`·;;: E
`·~ 0
`UU
`
`training test lung
`with lift bar
`
`8
`
`nebulizer
`
`o,
`
`ring
`stand
`
`vacuum
`
`FIGURE l. Top, A: experimental setup used to evaluate nebulizer
`output and mhaled mass. Bottom, B: experimental setup to evalu(cid:173)
`ate aerosol particle size output of the nebulizer.
`
`lower respiratory tract, whereas particles too small are
`exhaled.8 It has been shown that smaller particles are
`produced at higher nebulizer flows.9 Methods that
`determine particle sizes from nebulizers should classify
`them according to aerodynamic diameter and not
`physical diameter. A cascade impactor, unlike the op(cid:173)
`tical laser particle counter, allows quantification of
`drug delivery in terms of aerodynamic size character(cid:173)
`istics of the aerosol. 10 Aerodynamic diameter accounts
`for the density and irregular shape of drug particles and
`more accurately predicts the behavior of the aerosol as
`it is delivered to the patient.
`Loffert et al4 introduced the concept of respirable
`rate, which combines the effects of nebulizer output,
`nebulization time, and percent of particles in the
`respirable range. However, they determined nebulizer
`output by serial weighing rather than measurement of
`the amount of nebulized medication. A more useful
`index of nebulizer function might be respirable mass(cid:173)
`the amount of aerosolized drug in the respirable range.
`In this study, we extended the concept of Loffert
`et al4 by quantifying not only the respirable aerosol
`mass from the nebulizer, but also the respiratory
`aerosol mass made available to the patient at a specific
`ventilatory pattern.
`
`Table !-Nebulizer Brands Evaluated
`
`Nebulizer Brand; Manufacturer; Location
`
`Whisper Jet Nebulizer System; Marquest Medical Products
`Inc; Englewood, Colo
`Ava-Neb Nebulizer; Hudson Respiratory Care Inc;
`Temecula, Calif
`Raindrop Medication Nebulizer; Puritan-Bennett; Lenexa,
`Kan
`Vix One Nebulizer; Westmed Inc; Tucson, Ariz
`Sidestream Nondisposable; Inspired Medical Products;
`Pagham, West Sussex, UK
`T-Updraft II Neb-U-Mist Nebulizer; Hudson Respiratory
`Care Inc; Temecula, Calif
`Fan Jet Nebulizer; Westmed Inc; Tucson, Ariz
`One-No 8900 TG; Salter Labs; Arvin, Calif
`Airlife Misty Neb; Baxter Healthcare Corp; Valencia, Calif
`Hospitak; Lindenhurst, NY
`T Up-Draft; Hudson Respiratory Care Inc; Temecula, Calif
`Sidestream Disposable; Inspired Medical Products; Pagham,
`West Sussex, UK
`Intertech lnspiron; Intertech Resources Inc; Lincolnshire, Ill
`Betamist2 Medication Nebulizer; Professional Medical
`Products Inc; Greenwood, SC
`Micro-Mist; Hudson Respiratory Care Inc; Temecula, Calif
`Ventstream; Inspired Medical Products; Pagham, West
`Sussex, UK
`B & F Medical Products Inc; Toledo, Ohio
`
`A.
`
`B.
`
`C.
`
`D.
`E.
`
`F.
`
`G.
`H.
`I.
`J.
`K.
`L.
`
`M.
`N.
`
`0.
`P.
`
`Q.
`
`It has been suggested that nebulizers be character(cid:173)
`ized by the amount of medication that is delivered to
`the patient. Smaldone11 introduced the term inhaled
`rrwss, which he defined as that mass of drug actually
`delivered by a given nebulizer for a defined breathing
`pattern and period of time. Inhaled mass is affected not
`only by the performance of the nebulizer, but also by
`the breathing pattern chosen. For a given breathing
`pattern, inhaled mass should allow comparison of the
`quantity of drug delivered by different nebulizer sys(cid:173)
`tems and adjustment of the drug dose accordingly. To
`our knowledge, evaluation of inhaled mass has not
`been reported for nebulizers designed primarily for'
`delivery of bronchodilators.
`We conducted this study to evaluate medication
`nebulizer performance addressing those issues de(cid:173)
`scribed above. Dead volume, the amount of drug re(cid:173)
`maining in the dead volume, nebulization time, and
`aerosol available to the patient (inhaled mass) were
`evaluated for 17 nebulizers, 3 fill volumes, and 3 flows.
`We also evaluated particle size for 3 flows with the 17
`nebulizers at a single fill volume.
`
`MATERIALS AND METHODS
`
`Nebulizers Evaluated
`We evaluated 17 commercially available nebulizers (Table l) .
`Nebulizers were provided by their manufacturers. All units were
`from the same lot number and the same packaging case. The neb(cid:173)
`ulizers were provided by the manufacturer from their saleable stock
`none were prototypes or otherwise prepared specifically for thi~
`study, and all were used as supplied from the manufacturer.
`
`CHEST/110/2/AUGUST, 1996
`
`499
`
`IPR2021-00406
`United Therapeutics EX2031
`Page 2 of 8
`
`

`

`e-1 .4
`~1 .2
`n;
`_,[1 .0
`
`"E s 0.8
`~
`ill 0 . 6
`..2
`~ 0.4
`16
`~0.2
`
`0 .0
`
`~
`'5
`~ 1 .5
`
`3 mL 4 mL 5 mL
`6 L/min
`
`3 mL 4 mL 5 mL
`8 L/min
`
`3 mL 4 mL s mL
`10 L/min
`
`0>
`.§.
`E g 1 .0
`E ., ...
`
`E
`:::1
`~0. 5
`-m
`
`"0
`
`0 .0
`
`J K
`I
`A B C D E F G H
`nebulizer brand
`
`l M N 0 P Q
`
`FIGURE 2. Top: effect of volume ( p<0.001) and flow (p=0.02) on
`amount of albuterol trapped in the dead volume. Data are pooled
`from all nebulizers for each flow and volume setting. Bottom: effect
`of nebulizer brand on amount of albuterol trapped in dead volume
`(p <0.001). Data for each nebulizer brand are pooled from all vol(cid:173)
`ume and flow settings.
`
`Evaluation of Dead Volume and Aerosol Available to the Patient
`The experimental system is shown in Figure 1. The nebulizer was
`placed in a clamp and attached to a ting stand in the vetiical posi(cid:173)
`tion. A double-sided t est lung (Michigan Instruments; Grand Rap(cid:173)
`ids, Mich) was used to simulate spontaneous breathing. One side of
`the test lung was attached to a ventilator (Putitan-Bennett 7200
`ventilator; Puritan-Bennett; Carlsbad, Calif) and that side of the test
`lung lifted the contralateral side to simulate spontaneous breathing.
`The lung model was set to simulate breathing at 12 breaths/min,
`fraction of total respiratmy cycle inspiration (Ti!Ttot) approximately
`0.4, peak inspiratory flow approximately 0.3 to 0.4 Us with a sine
`wave pattern, and tidal volume approximately 0.45 to 0.5 L. A pi(cid:173)
`lot evaluation showed that this closely simulated the breathing pat(cid:173)
`tern of a normal volunteer breathing through a mouthpiece with(cid:173)
`out nose clips (imitating mouthpiece breathing witl1 a nebulizer).
`The mouthpiece of the nebulizer was replaced with a step-down
`adapter and a cotton plug was placed into the adapter to trap aero(cid:173)
`sol. Pilot testing using several cotton plugs showed that a single plug
`was 100% effective as a filter of aerosol from the nebulizer. The test
`chamber of the lung model (simulating a spontaneously breathing
`patient) was attached to the adapter containing tl1e cotton plug.
`Sa]jne solution diluent volumes of 2.5, 3.5, and 4.5 mL and 2.5
`mg (0.5 mL of0.5%) of albuterol (Proven til; Schereing; Kenilworth,
`NJ) were placed into tl1e nebulizer cup to produce fill volumes of
`3, 4, and 5 mL. Saline solution and albuterol were precisely mea(cid:173)
`sured using ca]jbrated syringes. Oxygen flows of 6, 8, and 10 Umin
`were used from a ca]jbrated flowmeter (Timeter; Lancaster, Pa)
`connected to the hospital bulk oxygen supply. Three new nebuliz(cid:173)
`ers of each type were evaluated at each combination of diluent
`volume and oxygen flow. All evaluations were conducted at ambi-
`
`ent room temperature (approximately 22°C) and humiruty (approx(cid:173)
`imately 40 to 50% relative humidity).
`ebu(cid:173)
`The nebulizer was tapped periorucally during each trial.
`lization time was determined b y a stopwatch and was considered
`complete w hen there was no visible or auruble evidence of nebu(cid:173)
`lization for a period of 30 s. The nebulizer was weighted empty
`( Ohaus 311 Cent-0-Gram Balance; Carolina Biological; Burlington,
`NC), after it was filled with merucation and diluent, and at tl1e end
`of the trial. The percentage of solution that was nebulized was cal(cid:173)
`culated from these mass values.
`At the end of each trial, the amount of drug remllining in the
`nebulizer cup was determined by washing the inside of the nebu(cid:173)
`lizer cup with 10 mL of sa]jne solution and spectrophotometrically
`detennining the amount of albuterol remllimng in the nebulizer
`cup. The drug present in the cotton plug was extracted u sing 20 mL
`of saline solution and gentle agitation by vortex. The resulting so(cid:173)
`lution was centrifuged at 5,000 g for 10 min to remove all cotton
`fibers from the solution and the amount of albuterol was then d e(cid:173)
`·with other studies using
`termined spectrophotometrically. As
`methods similar to ours, we assumed that all albuterol was extracted
`from the cotton. 12·13
`
`Particle Size Determination
`The ex1)erimental setup used to determine particle size is shown
`in Figure 1. During evaluation, the nebulizer was placed in a clamp
`and attached to a ring stand in the vertical position. Albuterol (0.5
`mL of 0.5%) was placed into the nebulizer cup and ruluted with 2.5
`mL of saline solution. Particle sizes were dete1mined at oxygen
`flows of 6, 8, and 10 Umin. Three new nebulizers of each type were
`evaluated at each oxygen flow.
`Aerosol particle size produced by the nebulizer was determined
`using an 11-stage cascade impactor (Intox; Albuquerque, NM) \vith
`cutoff stages of 12, 9.52, 7.56, 6, 5, 4, 3, 1.8, 1, 0.4, and 0.25 pm.
`Aerosol was sampled 5 em from the outlet of the nebulizer at a flow
`of 2 Umin to the impactor for 2 min. The albuterol deposited on
`each stage of the impactor was collected on plates, washed \vith
`saline solution, and the amount of albuterol was determined spec(cid:173)
`trophotomehically. The cascade impactor was calibrated by the
`manufacturer and used per manufacturer's specifications. Mass
`meruan aerodynamic ruameter (MMAD) and geometric standard
`deviation (GSD) were determined from the calibration curves
`provided b y the manufacturer. Cumulative deposition data were
`plotted agllinst stage cutoff ruameter, and fitted with a logarithmic
`regression curve to determine the particle size at 50% of the accu(cid:173)
`mulated d eposition (MMAD). This relationship wa~ ummodal for
`all nebulizers and R2 for this relationship is typically greater than 0.9
`in our laborato ry. GSD was calculated as the MMAD ruvided by the
`particle size at 16% deposition. In adrution to MMAD and GSD,
`the percentage of particles in the respirable range of 1 to 5 pm was
`determined. Although both larger and smaller particles may have
`clinical benefit, we defined respirable particles as 1 to 5 pm for
`purposes of describing nebulizer performance in the laboratory.
`The same particle size range (1 to 5 pm) has been used by others
`to desctibe nebulizer performance 4
`
`Respirable Mass
`We also calculated respirable mass to combine the effects of drug
`output and the percentage of particles in the respirable range. Drug
`output was calculated b y subtracting the amount of merucation in
`the dead volume from the amount of merucation placed into the
`nebulizer at the beginning of the trial (2.5 mg). Respirable mass
`output of the nebulizer wa~ then calculated b y multiplying the drug
`output times the percentage of particles in the respirable range of
`1 to 5 pm. Respirable mass available to the patient was calculated
`by multiplying the inhaled mass times the percentage of particles
`in the respirable range. Thus, respirable mass output of the nebu-
`
`500
`
`Laboratory and Animal Investigations
`
`IPR2021-00406
`United Therapeutics EX2031
`Page 3 of 8
`
`

`

`0.5 e
`~
`,Q 0.4
`"' C)
`E "-E 0.3
`5
`~ 'E 0.2
`.!!!
`n;
`<>.
`0.1
`
`0.0
`
`eo.e
`~
`.0 n; §:0.6
`c :::>
`~ 0.4
`"' c
`"' 8_o.2
`
`0.0
`
`_15.0
`c:
`§.
`
`i c 10.0
`.2 i
`
`~ 5.0
`
`3 mL 4 mL 5 mL
`6 L/min
`
`3 mL 4 mL 5 mL
`8 L/min
`
`3 mL 4 mL 5 mL
`10 L/min
`
`J K L M N 0 P a
`I
`A B C D E F G H
`nebulizer brand
`
`0 .0
`
`14
`
`12
`
`10
`
`8
`
`6
`
`4
`
`2
`
`0
`
`c
`:§.
`"' E
`=
`
`3 mL 4 mL 5 mL
`6 L/mln
`
`3 mL4 mL 5 mL
`8 L/min
`
`3mL4mL5mL
`10 L/mln
`
`J K L M N 0 P a
`I
`A B C 0 E F G H
`nebulizer brand
`
`FIGURE 3. Top: effect of volume (p<O.OOl) and flow (p=0.02) on
`amount of albuterol delivered to the patient. Data are pooled from
`all nebulizers for each flow and volume setting. Bottom: effect of
`nebulizer brand on amount of albuterol delivered to the patient
`(p<O.OOl). Data or each nebulizer brand are pooled from all volume
`and flow settings.
`
`lizer described the aerosol production of the device, whereas
`respirable mass available to the patient described aerosol available
`at the mouthpiece of the device. Because particle size was deter(cid:173)
`mined only for a single nebulizer volume, these calculations were
`conducted only for a nebulizer volume of 3 mL.
`
`Spectrophotometric Analysis of Albuterol
`
`A stock solution of albuterol (0.05 mglmL) was prepared from
`powdered drug (Sigma; St. Louis) and a standard curve was
`constructed from serial dilutions. An absorbance peak was found at
`278 nm and all absorbance measurements were made at this
`wavelength. Our spectrophotometric scan of the reference drug
`solution agreed well with published spectral properties of al(cid:173)
`buterol.14 The spectrophotometer absorbance was adjusted to zero
`with a saline solution solvent (or saline solution solvent treated with
`cotton as appropriate) before each measurement was made. The
`amount of drug in test solutions was determined from the standard
`curve.
`
`Statistical Analysis
`
`Summary statistics are reported and mean±SE. Differences
`between groups were determined by one-way and multifactorial
`analysis of variance as appropriate. Post hoc analysis was conducted
`using the Scheffe procedure . Statistical significance was set at
`p<0.05.
`
`RESULTS
`
`Dead Volume
`The amount of drug in the dead volume measured
`by serial weighing (gravimetric method) was signifi-
`
`FIGURE 4. Top: effect of volume (p<O.OOl) and flow (p<O.OOl) on
`nebulization time. Data are pooled from all nebulizers for each flow
`and volume setting. Bottom: effect of nebulizer brand on nebuliza(cid:173)
`tion time (p<O.OOl). Data for each nebulizer brand are pooled from
`all volume and flow settings.
`
`cantly less than that determined by measuring the
`amount of drug remaining in the dead volume (spec(cid:173)
`trophotometric method) (0.81±0.01 mg vs 1.14±0.01
`mg; p<0.001). The effects of flow, diluent volume, and
`nebulizer brand on dead volume are shown in Figure
`2. There was a small but significant (p=0.02) decrease
`in dead volume amount with an increase in flow from
`6 to 10 Umin. There was no significant difference
`(p>0.05) between 8 and 10 Umin and 6 and 8 Umin.
`With an increase in diluent volume, there was a
`significant decrease in dead volume amount (p<O.OOl);
`this effect was significant among all levels of diluent
`volume. There were also significant differences in dead
`volume among nebulizer brands (p<0.001).
`
`Aerosol Mass Available to the Patient (Inhaled
`Mass)
`The effects of flow, fill volume, and nebulizer brand
`on the amount of aerosol available at the mouthpiece
`are shown in Figure 3. There was a small but signifi(cid:173)
`cant (p=0.02) decrease in the amount of drug delivered
`to the mouthpiece (and thus available to the patient)
`with an increase in flow from 6 to 10 Umin. There was
`no significant difference (p>0.05) between 8 and 10
`Umin and 6 and 8 Umin. There was a significantly
`greater amount of drug delivered to the mouthpiece
`with an increase in fill volume (p<0.001) from 3 to 4
`
`CHEST /110/2/ AUGUST, 1996
`
`501
`
`IPR2021-00406
`United Therapeutics EX2031
`Page 4 of 8
`
`

`

`50%
`
`~40%
`:::;;
`~30%
`'E
`fl
`iii 20%
`Cl.
`
`10%
`
`0%
`
`60%
`
`0 so%
`< :::;;
`::E 40o/o
`u;>
`
`~ 30o/o
`0
`
`&20%
`
`10%
`
`0%
`
`6
`
`8
`nebulizer flow
`
`10
`
`I J K L M N 0
`H
`A B C D E F G
`nebulizer brand
`
`p a
`
`FIGURE 6. Top: effect of flow on percentage of particles in the re(cid:173)
`spirable range of l to 5 pm (p=0.004). Data are pooled from all
`nebulizers for each flow. Bottom: effect of nebulizer brand on per(cid:173)
`centage of particles in the respirable range of l to 5 pm (p<O.OOl).
`Data for each nebulizer brand are pooled from all flow settings.
`
`Scheffe post hoc analysis, there was a significant
`difference in both MMAD and particles in the respi(cid:173)
`rable range between flows of 6 and 8 Umin and
`between 6 and 10 Umin, but no difference between 8
`and 10 Umin.
`
`Respirable Mass
`Respirable mass output of the nebulizers and respi(cid:173)
`rable mass available to the patient is shown in Figure
`7. With an increase in flow, there was a small but in(cid:173)
`significant increase in respirable mass output (p=0.07)
`and respirable mass available to the patient (p=0.43).
`For nebulizer brands, there were significant differ(cid:173)
`ences in respirable mass output and respirable mass
`available to the patient (p<0.001 in each case). There
`was significantly more respirable aerosol mass output
`from each nebulizer than was available to the patient
`(p<0.001).
`
`DISCUSSION
`In this study, we have demonstrated that medication
`nebulizer function is affected by diluent volume, flow,
`and nebulizer brand. Increasing diluent volume de(cid:173)
`creased the amount of albuterol trapped in the dead
`volume and increased the amount delivered to the
`patient. Increasing diluent volume also increased
`
`6
`
`8
`nebulizer flow
`
`10
`
`6
`
`5
`
`w 4
`e 0
`§. 3
`0 < :::;;
`:::;;
`
`2
`
`0
`
`A B C 0
`
`J K L M N 0 P a
`I
`E F G H
`nebulizer brand
`
`FIGURE 5. Top: effect of flow on MMAD (p<O.OOl). Data are
`pooled from all nebulizers for each flow. Bottom: effect of nebulizer
`brand on MMAD (p<O.OOl). Data for each nebulizer brand are
`pooled from all flow settings.
`
`mL and from 3 to 5 mL. However, there were sig(cid:173)
`nificant differences in albuterol delivered to the mouth(cid:173)
`piece between 4 and 5 mL. There were also significant
`differences among nebulizer brands in the amount of
`drug delivered to the mouthpiece (p<O.OOl).
`
`Nebulization Time
`The effects of flow, fill volume, and nebulizer brand
`on nebulization time are shown in Figure 4. There was
`a significant increase in nebulization time \vith an in(cid:173)
`crease in volume or a decrease in nebulizer flow
`(p<0.001 in each case). These differences were signif(cid:173)
`icant between all levels of volume and nebulizer flow
`(p<0.05 by Scheffe analysis). There were also signifi(cid:173)
`cant differences in nebulization time among nebulizer
`brands (p<0.001).
`
`Particle Size
`The effects of flow and nebulizer brand on particle
`size are shown in Figure 5. There was a significant
`decrease in MMAD \vith an increase in nebulizer flow
`(p<0.001). There were also significant differences in
`MMAD among nebulizer brands (p<0.001). As shown
`in Figure 6, there was a small, but significant (p=0.004)
`increase in the mass of particles in the respirable range
`\vith an increase in nebulizer flow. There were signif(cid:173)
`icant differences among nebulizer brands in the mass
`of particles in the respirable range (p<0.001). By
`
`502
`
`Laboratory and Animal Investigations
`
`IPR2021-00406
`United Therapeutics EX2031
`Page 5 of 8
`
`

`

`nebulization time, the effect of which was offset by
`increased flow. Compared with the effect of diluent
`volume, flow had a lesser effect on nebulizer perfor(cid:173)
`mance. Increasing flow increased the mass of particles
`in the respirable range of l to 5 pm; the respirable mass
`delivered to the patient was affected to a greater extent
`by nebulizer brand than flow. The effect of nebulizer
`brand was relatively greater than the effect of diluent
`volume or flow, with large differences among devices
`of different manufacturers.
`A large fraction of the medication placed into the
`nebulizer cup remains trapped within the device and
`in not made available to the patient. Previous studies
`have evaluated this dead volume by serial weighing.3-5
`However, such gravimetric methods do not account for
`the effect of reconcentration within the nebulizer. We
`found that the amount of albuterol trapped in the
`nebulizer is nearly 50% greater than that estimated by
`gravimetric methods. These results agree with those of
`O'Callaghan et al,7 who also found an overestimation
`of nebulizer output by 50% when serial weighing was
`used rather than direct measurement of nebulizer
`output. This can be explained by the fact that many of
`the particles produced within the nebulizer are baffled
`and return to the nebulizer cup. During this process,
`evaporation occurs so that particles which return to the
`nebulizer solution increase the concentration in the
`dead volume. This effect is accentuated when a cold
`dry gas is used to power the nebulizer, 15 as was the case
`in this study.
`Increasing the diluent volume increased the nebu(cid:173)
`lizer output. Malone et al16 used high-performance
`liquid chromatography to measure the amount of
`albuterol remaining in the nebulizer cup with 3 fill
`volumes (1.5, 2.5, and 3.5 mL). Similar to our findings,
`they found that the amount of albuterol trapped in the
`dead volume increased with smaller fill volumes. It is
`difficult to make direct comparisons between our re(cid:173)
`sults and those of Malone et al. 16 They used a
`compressor (relative humidity approximately 50% and
`flow approximately 6 Umin) rather than dry oxygen to
`power the nebulizer, they used different fill volumes,
`and they used a nebulizer different than any of those
`in our study (DeVilbiss 646). However, they report
`nebulizer outputs at 2.5 mL and 3.5 mL fill volumes
`that are similar to what we found with fill volumes of
`3 mL and 4 mL, respectively. Malone et al16 also
`evaluated the concentration of albuterol in the nebu(cid:173)
`lizer over the nebulization time and found nearly a
`100% increase for smaller diluent volumes, but a
`smaller increase in concentration occurred for the
`larger diluent volume.
`Studies that used gravimetric methods to evaluate
`nebulizer performance have also reported increased
`nebulizer output with increasing fill volume,3·5·17J 8 as
`
`0.6
`
`o;o.s
`E..
`"' ~0.4
`E
`<1>
`~0.3
`
`·c. e 0.2
`
`0.1
`
`0
`
`1 .2
`1 . 1
`
`_0.9
`"" go.s
`iQ 0.7
`E 0.6
`<1>
`~0 .5
`·g-0.4
`~ 0.3
`0.2
`
`0.1
`
`0
`
`6 L/min
`
`nebulizer flow
`
`,I,
`
`II,
`
`II,
`
`f-r-,
`
`f-r-, F,
`
`1-r,
`
`F,
`
`J K L M N 0 P 0
`I
`A B C D E F G H
`nebulizer
`
`FIGURE 7. Top: effect of flow on respirable mass. Solid bars
`represent respirable output of the nebulizer (p=0.07) and open bars
`represent respirable mass available to the patient (p=0.43). Data are
`pooled from all nebulizers for each flow. Bottom: effect of nebulizer
`brand on the respirable mass output of the nebulizer (solid bars,
`p<O.OOl) and respirable mass available to the patient (open bars,
`p<O.OOl ). Data for each nebulizer brand are pooled from all flow
`settings.
`
`well as an increased nebulizer output with increased
`flow.3 In contrast to these findings, we found that flow
`had a relatively small effect on albuterol output from
`the nebulizer. Using three jet nebulizer brands and
`sodium fluoride as a tracer, Dennis et al17 reported that
`nebulizer output was relatively unaffected by flow un(cid:173)
`less very high flows were used. Increased gravimetric
`output with little change in drug output as flow is in(cid:173)
`creased suggests that evaporation and reconcentration
`are greater for higher flows .
`Although flow had little effect on nebulizer output,
`it did affect nebulization time and particle size. A
`higher flow resulted in a shorter nebulization time,
`which could be used clinically to offset the effect of a
`larger diluent volume on nebulization time. Increasing
`the flow decreased the size of the particles generated
`and increased the proportion of particles in the respi(cid:173)
`rable range of l to 5 pm. The effect of flow on parti(cid:173)
`cle size was greater between 6 and 8 Umin than be(cid:173)
`tween 8 and lO Umin.
`It is interesting to note that an increased flow did not
`increase the amount of drug available to the patient.
`When the effects of flow on both particle size and
`amount delivered to the patient are considered (ie ,
`
`CHEST I 110 I 2 I AUGUST, 1996
`
`503
`
`IPR2021-00406
`United Therapeutics EX2031
`Page 6 of 8
`
`

`

`respirable mass), there was little effect of changing
`flow. Increasing flow increases the number of particles
`in the respirable range but also increases the amount
`of waste during the expiratory phase so that respirable
`mass remains relatively constant. This is supported by
`previous studies. Although Clay et al9 reported a
`decrease in aerosol particle size when nebulizers were
`powered with a higher flow, others showed no differ(cid:173)
`ence in bronchodilator response when higher nebu(cid:173)
`lizer flows were used.19·20 The lack of change in clin(cid:173)
`ical response with changes in flow is likely due to our
`findings that respirable mass available to the patient is
`unchanged with changes in flow. It would thus appear
`that flow primarily affects nebulization time and thus
`higher flows are favored for convenience rather than
`improved drug delivery. These results also indicate that
`the amount of aerosol available to the patient should
`not be affected by use of low-flow devices like com(cid:173)
`pressors that are commonly used in the home.
`Differences in performance among nebulizers were
`greater than the differences observed for changes in
`diluent volume or flow. The respirable mass delivered
`by some nebulizers was twice that of others. From our
`results, it is apparent that some nebulizers have been
`engineered for a performance that is superior to oth(cid:173)
`ers. The reasons for this are not readily apparent by
`inspection of the devices and likely relates to a num(cid:173)
`ber of factors, the identification of which was beyond
`the scope of this study. Our in vitro data are consistent
`with the in vivo data of Johnson et al. 21 They compared
`deposition and physiologic response to albuterol in
`patients with stable asthma using a nebulizer that
`produced small particles (MMAD, 3.3 pm) and a
`nebulizer that produced large particles (MMAD, 7.7
`pm). Deposition and physiologic response (ie, FEV1)
`were greater with the nebulizer that produced smaller
`particles. Presumably, the respirable mass was greater
`with the nebulizer that produced smaller particles.
`Nebulizers can be powered only during the inspira(cid:173)
`tory phase to minimize drug waste.22·23 A system has
`also been described that uses a collection bag to trap
`aerosol particles during exhalation and delivers them to
`the patient on the next inspiration.24 These systems are
`not in common use with spontaneously breathing pa(cid:173)
`tients owing to their additional cost and increased
`complexity. Another approach to decrease waste is to
`boost the nebulizer output during inspiration, thus
`decreasing the relative amount of drug lost during the
`expiratory phase, and N ewnham and Lipworth25 found
`that a nebulizer using this feature produced a twofold
`increase in the delivery of salbutamol to the lungs as
`compared with a conventional nebulizer.
`Although a standard dose of 2.5 mg was placed into
`the nebulizer in this study, less than 0.5 mg of respi(cid:173)
`rable mass was delivered to the patient. A number of
`
`the nebulizers that we evaluated delivered only a
`respirable mass of approximately 0.2 mg. It is of inter(cid:173)
`est to note that this is similar to the standard albuterol
`dose delivered by metered-dose inhaler (90 pg per
`actuation; 2 actuations=0.18 mg). This may explain why
`a number of studies have shown that clinical response
`using a nebulizer or metered-dose inhaler are virtually
`equivalent. 1 When the metered-dose inhaler is used,
`patient performance is a major determinant of aerosol
`delivery. With nebulizer use, performance of the neb(cid:173)
`ulizer is a major determinant of aerosol delivery.
`Alvine et al26 found considerable variability in neb(cid:173)
`ulizer function, not only among nebulizer brands, but
`also among nebulizers within specific brands. This is
`not particularly surprising, because nebulizers are
`produced in bulk and are typically short-term single(cid:173)
`patient use devices. Our data do not allow confirmation
`of the results of Alvine et

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket