throbber
laboratory and animal investigations
`Comparison of Breath-Enhanced to
`Breath-Actuated Nebulizers for Rate,
`Consistency, and Efficiency*
`
`Kitty Leung, BSc; Emily Louca, BSc, RRT; and
`Allan L. Coates, B Eng(Elect), MDCM
`
`Objectives: To evaluate differences between three new-generation nebulizers—Pari LC Star
`(Pari Respiratory Equipment; Mississauga, ON, Canada), AeroEclipse (Trudell Medical Interna-
`tional, London, ON, Canada), and Halolite (Medic-Aid Limited, West Sussex, UK)—in terms of
`rate and amount of expected deposition as well as the consistency of the doses delivered.
`Methods: The in vitro performance characteristics were determined and then coupled to the
`respiratory pattern of seven patients with cystic fibrosis (age range, 4 to 18 years) in order to
`calculate expected deposition. The Pari LC Star and AeroEclipse were characterized while being
`driven by the Pari ProNeb Ultra compressor (Pari Respiratory Equipment) for home use, and by
`a 50-psi medical air hospital source. The Halolite has its own self-contained compressor.
`Algorithms for the rate of output for the inspiratory flow were developed for each device. Patient
`flow patterns were divided into 5-ms epochs, and the expected deposition for each epoch was
`calculated from the algorithms. Summed over a breath, this allowed the calculation of the
`estimated deposition for each patient’s particular pattern of breathing.
`Results: The rate of deposition was highest for the Pari LC Star and lowest for the Halolite. Rate
`of deposition was independent of respiratory pattern for the Pari LC Star and AeroEclipse, but
`proportional to respiratory rate for the Halolite. The differences between the Pari LC Star and
`AeroEclipse were less when driven by the 50-psi source. The AeroEclipse had the least amount
`of drug wastage. As designed, the Halolite delivered a predetermined amount of drug very
`accurately, whereas expected deposition when run to dryness of the other two devices had
`significant variations.
`Conclusions: To minimize treatment time, the Pari LC Star would be best. To minimize drug
`wastage, the AeroEclipse would be best. To accurately deliver a specific drug dose, the Halolite
`would be best.
`(CHEST 2004; 126:1619 –1627)
`
`Key words: aerosols; asthma; breath-actuated nebulizers; breath-enhanced nebulizers; cystic fibrosis; pediatrics
`
`Abbreviations: CF ⫽ cystic fibrosis; CI ⫽ confidence index; Ot ⫽ total drug output; RF ⫽ respirable fraction;
`UV ⫽ ultraviolet; Vr ⫽ residual volume
`
`J et nebulization is one of the mainstays of treat-
`
`ment for cystic fibrosis (CF), where it is used to
`deliver medications ranging from antibiotics1 to mu-
`
`colytics,2,3 and is also commonly used to deliver
`bronchodilators
`for
`the emergency department
`treatment of asthma.4 From previous studies,5–7
`
`*From the Division of Respiratory Medicine and Lung Biology
`Research, Hospital for Sick Children, Research Institute, To-
`ronto, ON, Canada.
`The nebulizers studied were provided through the generosity of
`PARI Respiratory Equipment Inc., Trudell Medical International
`Inc., and Medic-Aid Limited.
`Supported from a grant from the Hospital for Sick Children’s
`Foundation, made possible by a generous donation from Arnold
`and Lynn Irwin for cystic fibrosis research.
`
`Manuscript received September 5, 2003; revision accepted May
`28, 2004.
`Reproduction of this article is prohibited without written permis-
`sion from the American College of Chest Physicians (e-mail:
`permissions@chestnet.org).
`Correspondence to: Allan L. Coates, B Eng(Elect), MDCM,
`Division of Respiratory Medicine, The Hospital for Sick Children,
`555 University Ave, Toronto, ON, Canada, M5G 1X8; allan.
`coates@sickkids.ca
`
`www.chestjournal.org
`
`CHEST / 126 / 5 /NOVEMBER, 2004
`
`1619
`
`Liquidia's Exhibit 1103
`Page 1
`
`

`

`breath-enhanced nebulizers are more efficient than
`unvented nebulizers, but not all breath-enhanced
`nebulizers have the same efficiency,8 with differ-
`ences in residual volume (Vr) and particle size
`resulting in significant differences in expected pul-
`monary deposition. There is a new generation of jet
`nebulizers that are breath actuated, producing med-
`ication only during inspiration, which makes them
`potentially even more efficient than the breath-
`enhanced devices. At present, there are little com-
`parative data available to help the clinician choose
`between devices for specific applications.
`All jet nebulizers have a nebulizing chamber con-
`taining liquid medication and a high-pressure, high-
`velocity jet of gas that creates a partial vacuum at the
`exit orifice of the jet, resulting in the medication
`being drawn up toward the high-velocity orifice,
`where shear forces fragment the liquid into a poly-
`disperse aerosol. The aerosol passes around a series
`of baffles, where larger particles are removed by
`inertial impaction and fall back into the reservoir for
`renebulization. Particles that escape the baffles ei-
`ther leave the nebulizer, or “rain out” and fall back
`into the medication chamber under the influence of
`gravity. Simplistically, the major difference between
`unvented and breath-enhanced nebulizers is that the
`patient’s inspiratory flow is entrained into the device,
`and particles that would otherwise rain out are swept
`along into the patient during inspiration.5,9 Hence,
`the rate of output of breath-enhanced nebulizers
`increases with increasing inspiratory flow and falls
`back to baseline during expiration when no flow is
`entrained. Furthermore, since inertial impaction of
`droplets on the baffles is in part dependent on
`velocity of the particle, increases in entrained flow
`increases the likelihood that larger particles will
`impact on the baffles. This may give rise to a smaller
`particle size distribution during inspiration as the
`inspiratory flow increases.8 Particles between 1 ␮m
`and 5 ␮m in diameter are ideal for pulmonary drug
`delivery, in that they are small enough so as not to be
`removed by inertial impaction at the posterior phar-
`ynx, but large enough to carry a significant amount of
`drug. Given that particle volume is proportional to
`the third power of the radius, particles ⬍ 1 ␮m carry
`little drug. The fraction of the volume of the nebu-
`lizer output carried in particles with a diameter ⱕ 5
`␮m is defined as the respirable fraction (RF).10 –13
`In terms of the appropriate choice of device, a
`number of factors come into play. Clearly, the ability
`to produce a high-density aerosol with a large RF
`during the inspiratory phase is the basic principle,
`but other factors such as Vr at end nebulization are
`an issue, especially if the medication is very expen-
`sive.14 Since one of the challenges in the treatment
`of CF is patient adherence to recommended treat-
`
`Figure 1. A schematic of the breath-enhanced nebulizer Pari
`LC Star.
`
`ment regimens, devices that reduce treatment time
`would be expected to offer advantages to the already
`very time-consuming daily multifaceted treatment
`activities of these patients.15–17 The devices should
`therefore be evaluated on the expected pulmonary
`deposition of a specific dose, and the delivery time
`required. The breath-enhanced nebulizer, the Pari
`LC Star (Pari Respiratory Equipment; Mississauga,
`ON, Canada) [Fig 1], has been shown to be one of
`the more efficient breath-enhanced nebulizers.8
`Breath-actuated nebulizers, such as the AeroEclipse
`(Trudell Medical International, London, ON, Can-
`ada) [Fig 2] and Halolite (Medic-Aid Limited, West
`Sussex, UK) [Fig 3] have recently been developed.
`The Halolite uses an adaptive aerosol delivery system
`that can adapt the drug delivery to each patient’s
`breathing pattern. Table 1 provides a functional
`comparison of all three devices.
`The purpose of this study was to compare the
`three devices in terms of in vitro performance,
`expected in vivo rate of deposition, and in vivo
`efficiency using the respiratory pattern of patients
`with CF breathing through a nebulizer. Significant
`end points are considered to be the percentage of
`the initial dose that would be delivered to the lungs,
`the time required to deliver a “target” dose, and the
`ability to deliver a precise pulmonary dose. It is
`recognized that the importance of these variables
`depends on the expense of the drug being delivered,
`the value in terms of possible greater adherence to
`recommended therapy from rapid delivery, and the
`therapeutic safety profile of the drug in terms of
`accurately delivering a specific amount.
`
`Methods and Materials
`
`Device Operation
`
`The nebulizers and compressors used in this study were the
`Pari LC Star nebulizer driven by the Pari Proneb Ultra compres-
`
`1620
`
`Laboratory and Animal Investigations
`
`Liquidia's Exhibit 1103
`Page 2
`
`

`

`Figure 2. A schematic of the breath-enhanced, breath-actuated nebulizer (BAN) AeroEclipse.
`
`sor (Pari Respiratory Equipment); the AeroEclipse nebulizer,
`which was also driven by the Pari compressor, as no specific
`compressor was recommended; and the Halolite nebulizer with a
`built-in compressor. The Halolite is a microprocessor-controlled
`
`Figure 3. A schematic of the breath-actuated nebulizer Halolite,
`which uses adaptive aerosol delivery technology. Insert shows
`breath tracing; dark areas represent device activation.
`
`device that activates the compressor on each inspiration. Three
`examples of both the Pari LC Star and the AeroEclipse were
`studied, but only a single Halolite device was available. The test
`drug was 2.5 mg (0.5 mL) of albuterol (Ventolin Respirator
`Solution; GlaxoSmithKline; Mississauga, ON, Canada) diluted
`with 3.5 mL of saline solution. This was chosen because it lends
`itself to ultraviolet (UV) spectrophotometry for quantification of
`output.8 The Pari LC Star and the AeroEclipse were also
`evaluated using compressed dry air (hospital air, 50-psi source) at
`8 L/min, which is the same flow recommended by the manufac-
`turer for the AeroEclipse. Flow from the compressor was
`measured by a flow calibration instrument (Timeter RT200;
`Allied Health Care Products, St. Louis, MO), and the flowmeters
`on the hospital air line were calibrated to adjust for “back
`pressure,”12 so as to deliver the expected driving nebulizing flow.
`When driving either the AeroEclipse or the Pari LC Star, the
`output of the ProNeb Ultra compressor was 4.9 L/min.
`
`Particle Size Distribution and Determining Nebulizer Output
`
`Both the Pari LC Star and the AeroEclipse were characterized
`in terms of particle size distribution and rate of output during
`steady-state conditions. Briefly, the device was mounted to allow
`aerosol to pass through the laser beam of a Malvern Mastersizer
`X (Malvern Instruments; Worcestershire, UK), and particle size
`was measured using the Mie theory for transparent droplets.
`Care was taken to avoid vignetting.18 This method has been
`described in detail elsewhere.19 Measurements were made after
`2 min of nebulization, which allowed the nebulizer to attain a
`steady-state temperature,19 after which particle size distribution
`and RF were calculated. In order to mimic entrained flow, air at
`40% relative humidity was added at the point of the inspiratory
`valve in flow increments of 5 L/min up to a maximum of 35
`
`www.chestjournal.org
`
`CHEST / 126 / 5 /NOVEMBER, 2004
`
`1621
`
`Liquidia's Exhibit 1103
`Page 3
`
`

`

`Table 1—A Functional Comparison of the Pari LC Star, AeroEclipse, and Halolite
`
`Pari LC Star
`
`AeroEclipse
`
`Halolite
`
`Breath enhanced
`
`Breath enhanced
`Breath actuated
`
`Inspiratory valve allows air to entrain
`into the chamber during inspiration
`when the flow of patient is greater
`than nebulizing flow
`
`When entrained flow is ⬎ 8 L/m, a
`unique spring-loaded mechanism
`allows the actuator piston to be
`pulled down onto the jet and
`nebulization commences
`Aerosol is only produced during the
`inspiratory phase, making it
`potentially very efficient
`
`Expiratory valve on mouthpiece prevents
`exhaled gases from entering the
`nebulizer
`Treatment complete when device
`sputters
`
`Expiratory valve on mouthpiece
`prevents exhaled gases from
`entering the nebulizer
`Treatment complete when device
`sputters
`
`Breath actuated
`Uses adaptive aerosol delivery system,
`which adapts drug delivery to each
`individual patient’s breathing pattern
`Device has two operating buttons; each
`is designed to deliver the
`manufacturer’s preset volume
`Aerosolization begins when the patient
`pushes the appropriate button
`(albuterol for this study) and begins
`breathing
`Halolite analyses the first three breaths
`of the patient to determine the
`breathing pattern
`A pulse of drug is delivered every
`subsequent breath only during the
`first 50% of inspiration
`No entrainment of flow on inspiration
`Output is constant for each pulse and
`independent of the inspiratory flow
`Valves divert ventilation around
`nebulizing chamber
`
`Treatment is complete when the preset
`dose has been delivered
`
`L/min, and particle size distributions were measured in each
`situation. For the AeroEclipse, the first level of entrained flow
`was 8 L/min because the spring-loaded valve only opens when
`entrained flow reaches this level. The microprocessor control of
`the Halolite makes conventional particle sizing difficult since the
`device is not designed to run continuously. This intermittent
`operation results in differences in temperature of the aerosol
`when nebulized continuously vs pulsed. The increased accuracy
`of 2,000 sweeps during data gathering by the Malvern Master-
`sizer X for particle size distribution calculations in “continuous”
`mode offsets the limited data achieved from a “pulse,” even with
`differences in temperature of the aerosol being particle sized. To
`create a continuous mode, the device was dismantled and the
`back pressure created by the compressor when driving the
`Halolite handset, which contains the nebulizing device and
`microprocessor, was measured as 28 to 30 psi. The compressor
`uses an elastic reservoir that allows pressure to increase during
`expiration, and contributes to the compressor output during the
`pulse of aerosol. This resulted in a driving pressure that is
`considerably higher than that which would have occur if the
`compressor were driving the nebulizer continuously. The micro-
`processor within the Halolite handset was dismantled, and the
`nebulizer was driven by a dry air gas source at a flow matching
`the back pressure previously measured from the Halolite com-
`pressor, which resulted in an output flow of 5.4 L/min. The
`mouthpiece of the handset was positioned to send a continuous
`stream of aerosol across the laser beam. Since there is no
`entrained flow, only one measurement condition was necessary.
`Prior to the particle size measurements, devices were weighed
`empty (for the Halolite, this was only the medication chamber),
`filled, and reweighed using an electronic balance (BL150; Sarto-
`rius Corporation; Edgewood, NY). After 4 min of steady-state
`output, the devices were reweighed. Changes in drug concentra-
`tion due to evaporative losses were assessed initially by changes in
`UV spectrophotometry and water vapor pressure osmolarity
`
`(Advanced Micro-Osmometer 3300; Advanced Instruments;
`Norwood, MA). Eventually, only osmolarity was used since the
`simpler technique gives identical results to the more complex UV
`spectrophotometry. The drug output over the nebulization pe-
`riod was calculated from the Vr and the changes in concentra-
`tion, as seen in Appendix 1.
`For each 4-min run under each condition of entrained flow, the
`total rate of output and that in the RF was calculated, and the
`mean taken for the three examples of both the Pari LC Star and
`the AeroEclipse. Polynomial curve-fitting techniques were used
`to create the algorithm for the rate of output—total and within
`the RF— over the range of entrained flow. Finally, both the Pari
`LC Star and the AeroEclipse were run to dryness, defined as the
`absence of mist for at least 10 s,8,12 to allow the calculation of the
`total output of the device. The details of these techniques have
`been described.8,19,20 Output data for the Halolite were collected
`by connecting it to a modified Harvard pump (Model 613;
`Harvard Apparatus; Holliston, MA) that delivered two half-sine
`waves, with an inspiratory time/total time of respiratory cycle
`(Ti/Ttot) of 0.4, a tidal volume of 500 mL, and a respiratory rate
`of 20 breaths/min. These settings approximate the tidal volumes
`and timing of actual patient flow traces (see below). The Harvard
`pump was run until the Halolite sensed that the preset volume
`had been delivered. The output was calculated from the drug
`remaining in the nebulizer cup via gravimetric techniques and
`changes in osmolarity, which had complete agreement with UV
`spectrophotometry. When the output multiplied by the RF is
`divided by the number of breaths, the result is the expected
`deposition per breath.
`
`Calculation of Estimated Pulmonary Deposition
`
`From a previous study,21 digitized breath tracings of seven
`patients with CF (age range, 4 to 18 years) breathing through a
`nebulizer (Table 2) were used. Patients with FEV1 values ⬎ 60%
`
`1622
`
`Laboratory and Animal Investigations
`
`Liquidia's Exhibit 1103
`Page 4
`
`

`

`Table 2—Demographic Information on the Seven
`Patients Whose Breathing Patterns Were Used To
`Calculate Estimated Pulmonary Deposition and In
`Vivo Efficiency*
`
`Patient
`No.
`
`Age,
`yr
`
`Height,
`cm
`
`Weight,
`kg
`
`FEV1, %
`Predicted
`
`Respiratory Rate,
`Breaths/min
`
`138
`11
`1
`141
`11
`2
`122
`7
`3
`174
`18
`4
`122
`7
`5
`142
`14
`6
`159
`4
`7
`*From Coates et al.21
`
`28
`34
`22
`55
`23
`31
`42
`
`84
`76
`68
`113
`78
`26
`55
`
`17.1
`18.7
`31.4
`18.9
`41.6
`40.7
`24.6
`
`predicted had essentially normal patterns of breathing, although
`the younger ones tended to be a bit tachypneic when breathing
`on the nebulizer. The child with the worse lung function (FEV1
`⬍ 30% predicted) was tachypneic at rest. The respiratory wave-
`forms were broken into 5-ms epochs and were used to calculate
`the expected deposition. Three breaths were chosen from a
`pattern that showed regular respiration, and the same three
`breaths were used to calculate expected deposition for each
`apparatus. Entrained flow was calculated by subtracting the
`nebulizer driving flow from the inspiratory flow. When this
`resulted in a negative number it was defined as zero, since the
`one-way inspiratory valve would be closed. The spring-loaded
`valve on the AeroEclipse does not open until the entrained flow
`reaches 8 L/min; output was considered zero until this occurred.
`From the algorithms of the total rate of output and that in the RF
`for the Pari LC Star and the AeroEclipse, the output in each 5
`ms-epoch for the specific entrained flow of the epoch was
`calculated and summed over the entire breath. These calculations
`are illustrated in Appendix 2. The results are reported as the
`mean of three breaths for each patient. This allows the in vivo
`efficiency, defined as the output during inspiration in the RF
`divided by total output over the entire respiratory cycle,8 to be
`calculated. For the Halolite, in vivo efficiency was equal to the
`output in the RF during inspiration since there is no expiratory
`drug loss.
`
`Validation of Assumptions
`
`To test the assumption that the output of the Pari LC Star and
`the AeroEclipse that was determined under steady-state condi-
`tions were valid under dynamic conditions, they were connected
`to the Harvard pump with the settings described above and run
`for 3 min. Total drug output (Ot) was calculated as described
`above. The two half waves from the Harvard pump were known
`mathematically and were entered as the “patient’s” breathing
`pattern. Using the algorithm for rate of drug output, the output
`over 3 min was calculated and compared to the measured output.
`Device evaluation and comparison included the expected
`pulmonary drug deposition per breath and per minute, in vivo
`efficiency, overall efficiency in terms of expected deposition in
`relation to the initial charge in the nebulizer, and for the Halolite
`the accuracy of the device to deliver a preset amount of drug. The
`calculated output and expected pulmonary deposition of the Pari
`LC Star and AeroEclipse, as well as the length of time to run to
`dryness were compared to the Halolite. To have comparable data,
`the time to deliver a selected predetermined dose was calculated
`for each device. The predetermined dose was defined as the dose
`delivered by the Halolite after four button presses, which was
`
`found to be essentially equivalent to its point of dryness. The time
`difference between the devices for each of the seven patients was
`calculated. The results are expressed as means ⫾ 95% confidence
`limits. Differences in patient size and device performance were
`explored by regression analysis.
`
`Results
`
`The steady-state in vitro assessment of both the
`total rate of output and that in the RF for the Pari
`LC Star and AeroEclipse is shown in Figure 4. With
`increasing entrained flow, the Pari LC Star increases
`both the Ot and that in the RF. The AeroEclipse
`begins producing aerosol when the entrained flow
`reaches 8 L/min (patient inspiratory flow is 13 L/min
`when the compressor driving flow is taken into
`consideration), and there is a slight fall off in Ot with
`increasing entrained flow, but there is an initial small
`increase in the RF, indicating a smaller particle size
`distribution with increasing flows. Given the design,
`the Halolite provides a constant output of 0.0029 mg
`per breath when it is activated.
`When the mathematically predicted output of the
`Pari LC Star and the AeroEclipse for the two half
`sinusoidal waveforms for the Harvard ventilator are
`compared to the actual output, there is no difference
`between the two (0.0089 ⫾ 0.0001 mg per breath
`vs 0.0090 ⫾ 0.0000 mg per breath, and 0.0046 ⫾
`0.0001 mg per breath vs 0.0045 ⫾ 0.0002 mg per
`breath for the Pari LC Star and the AeroEclipse,
`respectively [mean ⫾ 95% confidence index (CI)].
`
`Figure 4. Rate of output (total and in RF) in relation to
`entrained flow for the Pari LC Star and AeroEclipse while being
`driven by the compressor.
`
`www.chestjournal.org
`
`CHEST / 126 / 5 /NOVEMBER, 2004
`
`1623
`
`Liquidia's Exhibit 1103
`Page 5
`
`

`

`This gives credibility to the use of the mathematical
`model derived from steady-state data for the predic-
`tion of drug deposition during dynamic conditions.
`In other words, the quadratic equations that charac-
`terize the device performance during steady-state
`conditions can be applied during dynamic in vivo
`conditions.
`The rate of deposition as a function of respira-
`tory rate (Fig 5) is greatest for the Pari LC Star; it
`is intermediate for the AeroEclipse, and not re-
`lated to respiratory rate. In contrast, the Halolite
`has the lowest rate of deposition with most respi-
`ratory rates, but shows a linear increase with
`increasing rates until ⬎ 40 breaths/min, where the
`expected deposition is comparable to the other
`two devices. As expected, the opposite is true with
`tidal volume since those subjects with the greatest
`lung disease had the highest respiratory rates and
`the lowest tidal volumes. The mean rate of depo-
`sition ⫾ 95% CI was highest for the Pari LC Star
`(0.093 ⫾ 0.0084 mg/min with compressor
`vs
`0.120 ⫾ 0.0122 mg/min for dry air),
`lowest for
`the Halolite (0.055 ⫾ 0.016 mg/min, compressor
`only), and the AeroEclipse is in between (0.075 ⫾
`0.0064 mg/min with compressor, and 0.108 ⫾
`0.014 mg/min for dry air). The difference between
`the Pari LC Star and the AeroEclipse was signif-
`icant, but there was variability among the patients,
`resulting in overlap in the 95% CI between the
`AeroEclipse and the Halolite. The in vivo efficien-
`cies of Pari LC Star and AeroEclipse range from
`51 to 55% and 71 to 77%, respectively, for the
`older group of children. For the younger group,
`
`Figure 5. Rate of deposition as a function of respiratory rate for
`the Pari LC Star, AeroEclipse, and Halolite.
`
`the in vivo efficiencies are found to be 52 to 54%
`and 68 to 73%. The Halolite has an RF of 80%,
`which is equivalent to the in vivo efficiency. For
`the breath-actuated devices, the in vivo efficiency
`was essentially the RF since no drug is lost during
`expiration. For the Halolite, this is constant and
`independent of the subject, which is not the case
`for the AeroEclipse since increasing entrained
`flow played a role by increasing the RF (Fig 4). If
`the devices are run to dryness or “four button
`presses” for the Halolite,
`total expected drug
`deposition is greatest for the AeroEclipse (1.3032 ⫾
`0.0296 mg, compressor; 1.4719 ⫾ 0.0204 mg, dry
`air), least for the Halolite (0.8400 ⫾ 0.0000 mg), and
`intermediate for the Pari LC Star (0.9421 ⫾ 0.0307
`mg, compressor; 0.9719 ⫾ 0.0395 mg, dry air) and all
`independent of size of the subject. There is virtually
`no variation (95% CIs ⬍ 0.00005 mg of the initial
`dose of 2.5 mg of albuterol) in the expected dose
`delivered by the Halolite, despite large differences in
`size and breathing patterns. There was no relation-
`ship between the rate of deposition and the size of
`the subject, either in height or in weight. The larger
`(taller) subjects would have received less drug on a
`milligram per kilogram basis than the smaller sub-
`jects. This was most pronounced for the Aero-
`Eclipse, in which the 115-cm-tall subject would have
`received almost three times the amount in milli-
`grams per kilogram body weight than the subject 174
`cm in height. Furthermore, dosing differences due
`to device performance are greatest for the smaller
`subjects with much less discrepancy for the larger
`ones. With increasing entrained flow, the output and
`deposition are also increased, but the rate of Ot
`starts to level off at approximately 20 L/min for the
`AeroEclipse and approximately 30 L/min for the Pari
`LC Star.
`When evaluating the devices in terms of time to
`deliver a dose of medication, the Halolite consis-
`tently results in an expected pulmonary deposition of
`0.8400 mg with four button presses, which is there-
`fore selected as the comparing dose. Using the
`compressor, the Pari LC Star is fastest with expected
`pulmonary delivery in 9.2 ⫾ 0.8 min (mean ⫾ 95%
`CI) with the AeroEclipse taking 2.2 ⫾ 0.4 min
`longer, and the Halolite requiring 8.0 ⫾ 4.2 min
`longer. The much larger CIs with the Halolite are
`explained by the relationship of rate of output and
`respiratory rate with this device, whereas the other
`two devices are much less dependent on the respi-
`ratory pattern of the child. When driving the devices
`with hospital dry compressed air, the Pari LC Star
`delivers the dose in 7.1 ⫾ 0.7 min, with the Aero-
`Eclipse requiring 0.9 ⫾ 0.6 min longer.
`
`1624
`
`Laboratory and Animal Investigations
`
`Liquidia's Exhibit 1103
`Page 6
`
`

`

`Discussion
`
`This study demonstrates that three of the new
`generation of nebulizers each has particular strengths
`and weaknesses. In terms of rapid drug delivery, a
`factor that may shorten treatment time and improve
`adherence with recommended therapy in a disease
`like CF,16,17 the Pari LC Star appears to be the
`superior device when using a compressor. If the
`nebulizers are driven at 8 L/min from a compressed
`air source, as would be likely in a hospital setting, the
`rate of output for the Pari LC Star increases 29%,
`but 44% for the AeroEclipse, making the perfor-
`mance equivalent for the two devices. In terms of
`maximizing drug delivery, an important factor if the
`drug is very expensive, the AeroEclipse is the supe-
`rior device. From the perspective of a drug with a
`narrow therapeutic safety margin, the Halolite is
`much more predictable for drug delivery. Regardless
`of device, if a specific dose based on milligrams of
`drug per kilogram of body weight is desired, the
`initial dose put into the nebulizer will have to be
`individualized for the size of the patient. If the drug
`being nebulized is an antibiotic and minimal envi-
`ronmental contamination is desired, neither the
`AeroEclipse nor the Halolite allows significant (or
`any) antibiotic to leave the device except that which
`is exhaled by the patient.
`There are some potential limitations to this study.
`The most obvious is that in vitro data are combined
`with in vivo respiratory patterns to estimate, as
`opposed to measure,13 pulmonary deposition. While
`there is no doubt that nuclear medicine techniques
`would have to be considered the “gold standard,”
`comparison studies would mean multiple exposures
`to radioactive material that would not be allowed
`under the current regulations for ethical research in
`children. The comparisons are estimates, and the
`following issues are the potential sources of error in
`the calculations. The first is whether or not output
`data derived under steady-state conditions can be
`applied to the dynamic situation of regular breath-
`ing. In terms of total rate of output, the agreement
`between the data generated by the modified Harvard
`pump, which produces perfect half-sine waves, and
`the mathematical data using the model of the device
`output coupled to the mathematical expression of
`the sine waves suggests that steady-state data could
`be applied with no loss of accuracy. The second issue
`that is not addressed in this study is whether or not
`particle size distribution measured under steady-
`state conditions would apply to the dynamic situa-
`tion. In a previous study13 using the Pari LC Jet
`nebulizer in normal adults, very close agreement was
`found between the RF measured by laser diffraction
`and the in vivo RF measured by scintigraphy. How-
`
`ever, in children, the definition of an RF as the mass
`of aerosol carried in particles ⱕ 5 ␮m could be
`questioned. Recalculating the data of Wildhaber
`et al,22 it would appear that in smaller children a
`definition of ⱕ 5 ␮m for the RF is too large. Support
`for this comes from Geller and colleagues.23 who did
`not find evidence that smaller children received
`more tobramycin per kilogram of body weight from
`a Pari LC Jet nebulizer. They suggested that the
`lower RF in the small children limited pulmonary
`deposition. Since no specific data exist to give a valid
`estimate of RF based on size, this is a potential
`limitation that can be acknowledged but not scien-
`tifically corrected. Finally, the breaths used in the
`mathematical model assume a stable breathing pat-
`tern, by both the choice of where on the ventilatory
`pattern that the representative breaths are chosen,
`and the expression of the results as a mean of the
`three representative breaths. Such a stable pattern is
`frequently not the case in children. The device that
`would be most affected by an irregular pattern would
`be the Halolite because the timing of the output
`pulse is based on the previous three breaths. In a
`situation when breathing is irregular, the intended
`delivery of the pulse during the first half of inspira-
`tion may be mistimed. However, as a comparative
`study, the performance of each nebulizer is calcu-
`lated on the same three breaths from each patient,
`thereby minimizing physiologic variations in device
`performance. Another theoretical issue is that the
`manufacturer of the Halolite suggests that a better
`distribution of drug will occur if the pulse is deliv-
`ered only during the first part of inspiration, as
`compared to throughout the inspiratory cycle for the
`Pari LC Star and whenever inspiratory flow is ⬎ 13
`L/m in the case of the AeroEclipse. There is no
`specific comparative data to support or refute this
`claim, but a comparison between the Halolite and
`the Pari LC Jet, a less efficient precursor of the Pari
`LC Star,24 did not support more uniform distribution
`in patients with CF.
`There are differences in the in vitro performance
`between the Pari LC Star and the AeroEclipse. As in
`all breath-enhanced nebulizers, the entrained flow
`does two things: one is the reduction of rainout,
`thereby increasing output during the inspiratory
`phase; and the other is increasing the flow around
`the baffles, thereby increasing inertial impaction of
`the large droplets and reducing the particle size
`distribution. For the Pari LC Star, the predominant
`factor is the reduction of rainout so both the Ot and
`that in the RF increases with increasing entrained
`flow. In contrast, the Ot of the AeroEclipse fell,
`although the amount in the RF increases initially and
`then falls (Fig 4). As a result, the Pari LC Star is
`more efficient at rapid drug delivery than the Aero-
`
`www.chestjournal.org
`
`CHEST / 126 / 5 /NOVEMBER, 2004
`
`1625
`
`Liquidia's Exhibit 1103
`Page 7
`
`

`

`Eclipse; and both, due to their breath-enhanced
`design, are much faster than the unvented Halolite.
`In conclusion, each of the three devices tested has
`strengths and limitations. Both the choice of the device
`and the amount of drug placed in it should be made on
`the basis of the targeted pulmonary dose sought cou-
`pled with the desire for rapid delivery, the minimiza-
`tion of drug waste, and the need for precision of the
`pulmonary dose delivered. Hence,
`for prescribing
`medication to be given by aerosol, the initial dose, the
`device, and the patient must all be considered together
`if predicable results

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket