throbber

`
`
`
`appropriate nebulizers is a major precondition for a
`use of
`The
`treatment and prevention of Pneumocystis carinii pneumonia
`successful
`with pentamidine aerosol.
`The apparatus
`should supply a
`sufficient
`amount
`of
`pentamidine with
`adequate particle
`size. Using Fisons
`ultrasonic nebulizer FISO Neb, model FZ2V 40 BAMKI, De Vilbiss ultrasonic
`nebulizer, Porta-Sonic, model 8500 GB, and the Marquest Medical Products
`jet nebulizer Respirgard II,
`two pentamidine concentrations (300 mg/6 mi
`and 60 mg/6 ml) were compared by measuring nebulized pentamidine mass
`distribution and particle size distribution under in vitro conditions by
`means of a
`laser
`light-scattering particle sizer of
`the type Malvern
`Master sizer. It was found that there were significant differences among
`nebulizers. Mass distribution experiments with air flow 6 1/min showed
`that using FISO Neb the quantity of nebulized pentamidine was 201.4 mg
`and 36.7 mg, whereas using Porta-Sonic the values found decreased to
`85.2 mg and 23.6 mg. Using Respirgard II the values were 80.0 mg and
`10.64 mg. The measured total duration times of nebulization were 6 - 8.5
`min,
`12 min and 25 min for
`the nebulizers FISO Neb, Porta-Sonic and
`Respirgard II. A decomposition of pentamidine during nebulization in the
`case of ultrasonic nebulizers doesn*‘t
`take place. The measured mass
`median diameters (MMD) were 5.6 - 6.9 pm, 1.96 - 3.04 pm and 1.9 - 2.5
`wm for the nebulizers FISO Neb, Porta-Sonic and Respirgard II. Using 300
`mg pentamidine the nebulized amounts of pentamidine containing particles
`sizes < 2 um predominately available for alveolar deposition were with
`values of about 43 mg markedly higher for Respirgard II and Porta-Sonic
`than the measured 10.5 mg for FISO Neb.
`
`a
`
`|
`
`JOURNAL OF AEROSOL MEDICINE
`Volume 5, Number 2, 1992
`Mary AnnLiebert, Inc., Publishers
`
`Measurementof Particlé and Mass
`Distribution of Pentamidine Aerosol by
`Ultrasonic and Air Jet Nebulizers
`*
`ss
`o
`J. HAGER, K.-H. GOBER,J.-P. LOHR, and M. DURR
`
`’ Rhéne-Poulenc Rorer GmbH, Nattermannallee 1
`5000 Kéln 30
`
`.
`
`ABSTRACT
`
`:
`
`INTRODUCTION
`
`i
`
`te
`
`:
`
`1-
`i
`
`"
`
`n
`
`'s
`
`n s
`
`1
`2
`
`s
`a
`.
`5
`/
`>
`y
`
`Liquidia's Exhibit 1098
`Page 1
`
`'
`
`Pentamidine (Pentacarinat®) is used both in intravenous administra-
`tion and as aerosol
`for the treatment and prophylaxis of Pneumocystis
`carinii pneumonia associated with the acquired immunodeficiency syndrome
`(AIDS).
`In aerosol
`therapy aqueous pentamidine solutions with various
`
`Keywords:Pneumocystis carinii pneumonia, aerosol, nebulizer, pentamidine
`
`65
`
`aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaasaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaasaaaaaaaaaaaaaaaaaaaaasaaaaaaaasaaaaaaaaaaaasasaasaaassasaaaaasaal
`
`Liquidia's Exhibit 1098
`Page 1
`
`

`

`including determination of drug content and decomposition products.
`
`Influence
`pentamidine
`for nebulizing liquids with
`The
`use of an ultrasonic method
`particle diameters < 5 ym showing alveolar deposition is associated with
`a high energy supply to the nebulization liquid. This explains
`the
`possibility of
`substance
`decomposition
`reported in the
`literature
`(Konietzko N., 1989). While temperatures of about 50°C were measured
`transitorily during ultrasonic nebulization of pentamidine
`solutions,
`the stability tests were performed at various temperatures. Nebulization
`liquids with various concentrations of pentamidine were nebulized at
`different
`temperatures by an ultrasonic method. The chemical stability
`of pentamidine was
`investigated according to a new analytical method
`
`4
`
`such as
`drug concentration are applied by means of air jet nebulizers,
`Respirgard II, or by ultrasonic nebulizers.
`Several publications (Dirnagel, 1971; Corkery et al., 1988; Matthys
`and Herceg, 1990; Doherty et al., 1990; Konietzko, 1989; Herer et. al.,
`1989; Kéhler et al., 1983; Baskin et al., 1990; Smith et al.,1989; Wong
`et al., 1990) contain investigations on the particle size of pentamidine
`aerosols. According
`to these
`reports,
`the use of
`a particle size
`exhibiting alveolar deposition is
`a precondition. for a successful
`therapy with pentamidine
`aerosol.
`‘There
`are
`only
`few reports
`on
`pentamidine mass
`distribution
`during
`nebulization
`by
`different
`nebulizers,
`duration of nebulization and
`influence
`of ultrasonic
`nebulization on the chemical stability of pentamidine.
`
`under
`the following investigations was to determine
`The aim of
`various
`conditions
`(nebulizers
`of
`various
`types,
`different
`concentrations of pentamidine, various flow rates)
`the above mentioned
`operating characteristics and to compare
`these results with published
`similar data.
`
`:
`
`METHODS
`
`Pentamidine mass distribution »
`
`"output" can be defined gravimetrically as the loss
`The nebulizer
`in weight of
`the nebulizer or
`the change in volume of
`the nebulizer
`solution over
`time
`(Smaldone
`et al.,
`1988).
`In our
`studies
`the
`pentamidine mass distribution was
`determined
`by gravimetry and
`a
`chemical-analytical method, of pentamidine after completing nebulization
`in conditions
`similar
`to those used
`in clinical practice,
`except
`breathing pattern.of the patient. The results were given with reference
`to the used liquid in ‘percentage, or pentamidine in mg, as
`follows:
`residue in nebulizer,
`residue in safe-vent, nebulized amount/
`amount
`provided for treatment.
`
`Duration of nebulization
`
`investigations was to determine the nebulization
`The aim of our
`time of pentamidine-containing solutions using the nebulizers Respirgard
`II, Porta-Sonic and FISO Neb. The influence of the following parameters
`was tested: concentration of pentamidine and air flow.
`
`of ultrasonic nebulization on the chemical stability of
`
`66
`
`
`
`
`
`Liquidia's Exhibit 1098
`Page 2
`aaaacaaaa
`
`Liquidia's Exhibit 1098
`Page 2
`
`

`

`Particle size distribution
`
`So far
`
`it
`
`has not
`
`been possible
`
`to determine particle
`
`size
`
`in
`i.e.
`inhalation therapy,
`of
`conditions
`the
`distribution under
`a
`For
`patients with
`determination
`of
`pulmonary
`drug deposition.
`comparison of
`results described in the literature we need comparable
`conditions as to measuring methods used, and exact data on particle size
`designations. The aim of the particle size ‘measurements described below
`was to study the subjects hardly dealt with in the literature regarding
`the nebulization of pentamidine-containing solutions:
`influence of
`the
`concentration of pentamidine on particle size distribution; course of
`particle: size distribution during nebulization; comparison of particle
`size obtained -by the air jet.nebulizer and ultrasonic nebulizers. The
`present investigations were done by using Malvern Master sizer with mass
`median diameter for particle ‘size designation.
`
`Experimental protocol
`
`_
`
`The following conditions were fixed for each nebulizer: Respirgard
`II: airpressure 3.5 bar, airflow 6
`l/min, pentamidine
`(Pentacarinat
`300, Rhéne-Poulenc Rorer)
`300 mg or 60 mg/6 ml water
`f£.
`inj. DAB 9;
`total running time 25 min; Porta-Sonic with Safe-Vent: air flow 24 l/min
`and 6 l/min, pentamidine (Pentacarinat~
`300, Rhéne-Poulenc Rorer) 300 mg
`or 60 mg/6 ml water f.
`inj. DAB9
`;
`total
`running time 20 or 21 min.;
`FISO Neb:
`air
`flow 6
`l/min,
`power
`setting ‘“Minimum";
`pentamidine
`(Pentacarinat® 300, Rhéne-Poulenc Rorer) 300 mg/6 ml or 60 mg/6 ml water
`f.
`inj. DAB 9,
`total running time 6.5 min.. These conditions were used
`for measurements of mass distribution as well as for the particle size
`distribution with exception of the air flow value 24 l/min only used in
`determination .of mass distribution.
`In pilot studies with triplicate
`samples it was
`found that
`the measured mass distributions were in the
`range of
`3
`Z
`referring to the mean value of
`each device. For
`the
`particle size measurements
`a maximum deviation of
`4
`Z
`from the mean
`value was determined. Therefore, we considered that single measurements
`were
`sufficient
`to
`demonstrate
`the nebulizing properties
`of
`the
`investigated nebulizers for the conditions described in this paper.
`
`Pentamidine mass distribution and duration of nebulization.
`
`The nebulizers were operated according to Figures 1A, 1B and
`The
`nebulized
`amounts
`of
`pentamidine
`were
`collected
`in
`gaswasher/scrubber filled with 100 ml 1 Z ortho-phosphoric acid.
`
`1C.
`the
`
`'
`
`Influence
`pentamidine
`
`of ultrasonication on
`
`the
`
`chemical
`
`stability of
`
`and 60 mg/6 ml water
`Pentacarinat® 300 at concentrations of 300 mg
`inj. were investigated at 25 °C and 60 °C. The samples were sonicated
`f.
`by the nebulizers Porta-Sonic and FISO Neb and the Branson sonifier type
`250.
`For
`comparison
`samples
`autoclaved
`(121°C,
`20 min)
`were
`investigated. Drug contents and degradation products were analyzed by
`HPLC. The ultrasonication were done by the nebulizers Porta-Sonic and
`FISO Neb and the Branson sonifier type 250. The samples were analyzed by
`HPLC.
`
`-
`
`67
`
`Liquidia's Exhibit 1098
`Page 3
`
`|
`
`iJ
`
`|
`
`.
`-
`
`ch as
`
`thys
`
`al.,
`Wong
`idine
`size
`sful
`3
`on
`arent
`sonic
`
`nder
`erent
`ioned
`ished
`
`/
`+088
`-izer
`the
`doa
`‘tion
`tcept
`rence
`-Ows:
`rount
`
`‘gard
`‘ters
`
`“0
`

`
`rith
`with
`the
`ture
`ured
`ons,
`tion
`1 at
`lity
`thod
`
`!
`
`!
`
`
`
`Liquidia's Exhibit 1098
`Page 3
`
`

`

`
`
`
`
`_==,
`,
`
`2
`
`
`
`=
`
`3
`
`
`
`
`
`spentair
`
`_=,
`
`4
`
`5
`
`1: air compressor
`2: nebulizer
`* 3: gas washer
`4: flow meter
`5: vacuum pump
`
`— air supply
`— spentair
`
`= aerosol
`
`FIGURE 1A: The Respirgard II arrangement
`
`ambientair
`
`—
`
`'
`
`<=
`=
`=
`
`
`
`
`
`
`
`1
`
`2
`
`3
`
`4
`
`spentair
`
`]
`
`1: nebulizer
`2: gas washer
`3: flow meter
`4: vacuum pump
`
`-
`
`—— air supply
`== spentair
`a
`= aerosol
`
`FIGURE 1B: The Porta-Sonic arrangement
`
`ambientair
`sucked by nebulizer
`
`
`
`
`
`
`=—
`=
`=
`
`1
`
`2
`
`3
`
`4
`
`—
`
`spent air
`
`|
`
`1: nebulizer
`2: gas washer
`3: flow meter
`4: vacuum pump
`
`— air supply
`=== Spent air
`= aerosol
`
`FIGURE 1C: The FISO Neb arrangement
`
`Particle size distribution
`
`the Malvern Master Sizer was used. From the
`the measurements
`For
`size distribution curves
`and
`the
`total
`pentamidine mass
`particle
`6
`available for
`inhalation with an air flow of
`l/min the fractions of
`particles < 2 pm were calculated. The liquids used in the measurements
`
`68
`
`Liquidia's Exhibit 1098
`Page 4
`
`
`Liquidia's Exhibit 1098
`Page 4
`
`

`

`
`
`i
`
`COMPOSITION:
`PENTAMIDINE
`WATER f.
`INJ. |
`
`mg
`ml
`
`appearance
`
`colour
`
`pH
`
`density (m/V)
`
`refraction index
`
`viscosity (mPa.s)
`
`6.70.
`
`0.998
`
`1.3329
`
`1.000
`
`superficial tension
`(mN/m)
`
`72.8
`
`.
`
`osm. pressure (Z NaCl
`
`)
`
`0
`
`TABLE 1
`
`Physical data of nebulized liquids
`
`-
`6.0
`
`clear
`‘liquid
`
`300.0
`6.0
`
`ad
`
`clear
`liquid
`
`60.0
`6.0
`
`ad
`
`clear
`liquid
`
`colourless
`
`colourless
`
`colourless
`
`6.19
`
`0.0124
`
`1.3419
`
`1.1830
`
`71.2
`
`0.3806
`
`6.31
`
`1.0014
`
`1.3348
`
`1.0560
`
`68.0
`
`.
`
`not
`measured
`
`
`
`of particle size distribution showed the characteristics represented in
`Table 1, pentamidine (Pentacarinat
`300, Rhéne-Poulenc Rorer).
`
`Points of sizing measurements were:
`Respirgard II: 1, 5, 10 ,15 and 20 min
`Porta-Sonic
`;: 0.5 (10 mg/ml), 1, 5 and 9 min
`FISO Neb
`: 1,
`3 and 5 min
`
`the aerosol
`In case of Porta-Sonic the air supply was used to direct
`into the laser beam. The vacuum pump connected with the filter was only
`used to collect the pentamidine waste.
`
`The nebulizers used were operated according to FIGURES 2A, 2B, 2C
`
`air supply1
`*
`
`1: air compressor
`2: flow meter
`3: nebulizer
`4: laser particlesizer
`5: filter
`6: vacuum pump
`
`aan
`
`— air supply
`.
`== Spentair
`== aerosol
`
`FIGURE 2A: The Respirgard II arrangement
`
`69
`
`the
`mass
`is of
`
`nents
`
`Liquidia's Exhibit 1098
`Page 5
`
`
`Liquidia's Exhibit 1098
`Page 5
`
`

`

`
`
`
`
`
`
`
`spentair
`
`
`
`LJ
`
`— air supply
`
`== spentair
`2 aerosol
`
`
`
`| air supply
`
`1: air compressor
`. 2: flow meter
`3: nebulizer
`4: laser particlesizer
`5:filter
`6: vacuum pump
`
`FIGURE 2B: The Porta-Sonic arrangement
`
`spent air
`
`4
`
`
`
`
`
`
`
`e
`
`ambientair
`sucked by nebulizer
`
`.
`
`-
`
`1: nebulizer
`2: laser particlesizer
`3: filter
`4: vacuum pump
`
`
`
`
`
`|
`
`
`
`
`
`
`
`3
`
`——~ air supply
`=== spentair
`<= aerosol
`
`
`
`
`
`FIGURE 2C: The FISO Neb arrangement
`
`= Particle size measurement principle: laser-light scattering,
`apparatus: Malvern Master Sizer
`
`HPLC analysis
`
`in different solutions
`of pentamidine
`the content
`fo determine
`during this study a new HPLC-assay method was established. The different
`samples
`to be analyzed by HPLC were obtained from starting solutions,
`retaining solutions from nebulizer chambers, outwashing solutions from
`the pipes
`and collected aerosols
`in water. The new HPLC-method was
`statistically validated and the linearity of
`the determinations was
`checked.
`Two analytical HPLC-runs of each sample with two injections
`were made
`to calculate the mean value of
`the different components of
`interest.
`Chromato-
`2500
`D
`LC-System with
`HPLC Merck-Hitachi:
`Apparatus:
`Integrator. Reagents: water,
`HPLC grade; acetonitrile,
`HPLC grade;
`trifluoroacetic-acid,
`puriss;
`methanol,
`p.a..
`The
`quantitative
`determination follows the method ‘external standard". For
`this purpose,
`exactly identical portions
`(10 pl) of each sample
`and standard are
`injected into the
`separating column. Column: Kromasil
`5C8,
`250 mm
`length,
`4 mm internal diameter; mobile phase: A: 0.12 trifluoroacetic-
`
`70
`
`
`
`Liquidia's Exhibit 1098
`Page 6
`
`
`Liquidia's Exhibit 1098
`Page 6
`
`

`

`7B 23 - 100,
`acid, B: acetonitrile; gradient:
`Flow rate: 1 ml/min;
`temperature: 40°C.
`
`7 min; detection: 263 nm,
`
`4
`
`Pentamidine mass distribution
`
`RESULTS
`
`The relationships between pentamidine concentrations in the liquids
`the
`air
`flows,
`the
`types
`of nebulizer
`and
`the
`"static"
`used,
`nebulization without’ motion or
`tapping nebulization with motion or
`tapping and
`the
`repartition of
`the pentamidine mass
`introduced to
`nebulization are shown in TABLE 2, 3, 4, 5, 6, 7 and 8.
`
`TABLE 2
`
`Pentamidine ‘mass distribution with Respirgard II nebulizer
`
`PENTAMIDINE CONCENTR.
`AIR FLOW
`NEBULIZING TIME, TOTAL
`
`300 mg /6 ml
`6 1/min
`25 min
`
`.
`
`nebulizing liquid
`
`Pentamidine
`mg
`desired
`
`Pentamidine
`mg
`actual
`
`retained in Respirgard. II
`
`18.93 Z (m/m)
`
`56.79
`
`109.10
`
`retained in filter and
`tubes
`
`nebulized
`
`40.78 Z (m/m)
`
`40.29 Z (m/m)
`
`122.34
`
`120.87
`
`110.90
`
`80.00
`
`TABLE 3
`
`Pentamidine mass distribution with Respirgard II
`
`PENTAMIDINE CONCENTR.
`AIR FLOW
`NEBULIZING TIME, TOTAL
`
`60 mg/6 ml
`6 1/min
`25 min
`
`nebulizing liquid
`
`Pentamidine
`mg
`desired
`
`Pentamidine
`mg
`actual
`
`retained in Respirgard II
`
`10.17 Z (m/m)
`
`6.10
`
`retained in filter and
`tubes
`
`47.50 2 (m/m)
`
`28.50
`
`10.86
`
`38.50
`
`10.64
`25.40
`42.33 Z (m/m)
`nebulized
`eee
`
`71
`
`
`
`Liquidia's Exhibit 1098
`Page 7
`
`
`Liquidia's Exhibit 1098
`Page 7
`
`

`

`TABLE 4
`;
`
`Pentamidine mass distribution with Porta-Sonic with Safe-Vent
`
`s
`
`PENTAMIDINE CONCENTR.
`AIR FLOW
`NEBULIZING TIME WITHOUT
`MOTION
`.
`NEBULIZING TIME, TOTAL
`
`300 mg/6 mi
`24 L/min
`
`12 min
`21 min .
`nebulizing liquid
`,
`:
`
`retained in Porta-Sonic
`
`retained in Safe-Vent
`
`nebulized
`
`16.45 Z
`
`39.35 Z
`
`44.20 2
`
`Table 5
`
`. Pentamidine Pentamidine
`mg
`mg
`desired
`actual
`
`49.35
`
`118.05
`
`132.60
`
`63.00
`
`119.00
`
`118.00
`
`Pentamidine mass distribution with Porta-Sonic with Safe-Vent
`
`PENTAMIDINE-CONCENTR.
`AIR FLOW
`NEBUL. TIME WITHOUT
`MOTION
`NEBUL. TIME TOTAL
`
`60 mg/6 m2
`24 L/min
`

`
`6 min
`20 win
`
`60 mg/6 mi
`6 l/min
`
`9 min
`21 min
`
`~
`
`Penta
`nebulizing midine
`liquid
`mg
`desired
`
`Penta
`nidine
`ng
`actual
`
`Penta
`nebulizing midine
`liquid
`mg
`desired
`
`Penta
`midine
`ng
`actual
`
`-
`
`retained in Porta-Sonic
`
`retained in Safe-Vent
`
`12.09 %
`(a/n)
`
`28.21 %
`(m/m)
`
`7.25
`
`15.60
`
`16.99
`
`35.80
`
`9.97 %
`(n/m)
`
`40.86 %
`(m/m)
`
`5.98
`
`10.30
`
`24.52
`
`26.10
`
`nebulized
`59.60 %
`35.76
`8.60
`49.17 2%
`29.50
`23.60
`(nfm)
`(m/m)
`
`
`Pentamidine mass distribution with Porta-Sonic with Safe-Vent
`
`TABLE 6
`
`PENTAMIDINE-CONCENTR.
`AIR FLOW
`NEBULIZING TIME WITHOUT
`MOTION
`NEBULIZING TIME, TOTAL
`
`300 mg/6 ml
`6 1/min
`
`12 min
`20 min
`nebulizing liquid
`
`retained in Porta-Sonic
`
`retained in Safe-Vent
`
`nebulized
`
`16.91 2%
`(m/m)
`
`33.01 Z
`(m/m)
`
`50.08 Z
`(m/m)
`
`Pentamidine Pentamidine
`mg
`mg
`desired
`actual
`
`50.73
`
`80.90
`
`99.03
`
`133.90
`
`150.24
`
`85.2
`
`Liquidia's Exhibit 1098
`Page 8
`
`
`Liquidia's Exhibit 1098
`Page 8
`
`

`

`TABLE 7
`
`Pentamidine mass distribution with FISO Neb,
`power setting "Minimum"
`
`PENTAMIDINE CONCENTR.
`AIR FLOW
`NEBULIZING TIME, TOTAL
`
`300 mg/6 mi”
`6 l/min
`6.5 min
`
`-
`
`nebulizing liquid
`
`‘
`
`Pentamidine
`mg
`desired
`
`Pentamidine
`mg
`actual
`
`retained in FISO Neb
`
`16.45 Z (m/m)
`
`49.35
`
`70.80
`
`retained in filter and
`tubes
`.
`10.38 Z (m/m)
`31.14
`27.80
`, 73.17 Z (m/m) 219.51nebulized 201.40
`
`
`
`
`
`
`
`
`
`‘ine
`
`cal
`
`00
`
`00
`
`00
`
`ine
`
`al
`
`90
`
`90
`
`TABLE 8
`
`Pentamidine mass distribution with FISO Neb,
`power setting "Minimum"
`
`PENTAMIDINE CONCENTR.
`AIR FLOW
`NEBULIZING. TIME, TOTAL
`
`60 mg/6 ml
`6 l/min
`6.5 min
`
`nebulizing liquid
`
`Pentamidine
`mg
`desired
`
`Pentamidine
`mg
`actual
`
`retained .in FISO Neb
`
`12.12 Z (m/m)
`
`7.27
`
`retained in filter and
`tubes
`
`19.06 Z (m/m)
`
`11.44
`
`11.60
`
`11.70
`
`
`
`68.82 Z (m/m) 41.29nebulized 36.70
`
`
`
`
`
`comparison of nebulized pentamidine mass in Z%, available for
`The
`PCP inhalation therapy is shown in FIGURE 3.
`
`reveals that with an
`the tested three nebulizers
`A comparison of
`initial amount of 60 or 300 mg pentamidine for nebulization,
`the FISO
`Neb is superior in respect of the quantity of nebulized pentamidine. The
`pentamidine
`concentrations,
`i.e.
`60 mg/6 ml
`and
`300 mg/6
`ml
`respectively, used in aerosol
`therapy hardly influence the proportion of
`nebulized pentamidine. During nebulization concentration of pentamidine
`occurs in the nebulizing chambers.
`
`Duration of nebulization
`
`The results are given in FIGURE 4
`
`73
`
`Liquidia's Exhibit 1098
` Page 9
`
`Liquidia's Exhibit 1098
`Page 9
`
`

`

`ReeeeAotot
`
`m
`
`+
`
`thbeEO
`
`a
`
`[%] of Pentamidine
`nebulized
`
` Péntamidine
`
`
`60 mg/6 mi
`
`Pentamidine
`300 mg/6 ml
`
`FIGURE 3: 2 of nebulized pentamidine mass available for inhalation.
`Air flow:
`6 1/min
`
`Duration of nebulization
`
`The results are given in FIGURE 4
`
`time
`[min]
`30-7
`
`Pentamidine 300 mg/6 mi
`
`
`Respirgard I!
`
`6 Yimin.
`
`air flow
`
`Porta-Sonic
`6 Umin.
`
`FISO Neb
`position min.
`
`FIGURE 4: Duration of nebulization
`
`The results given in FIGURE 4 suggest that the nebulization time is
`shortest with the FISO Neb.
`In the case of Porta-Sonic and Respirgard IT
`the nebulization time was prolonged by moving (shaking,
`tapping)
`the
`nebulizers. Our experiments confirmed the variation in nebulization time
`reported in the literature (Vinciguera, Smaldone, 1990). With regards to
`the Respirgard II and Porta-Sonic it
`is recommended not
`to perform
`nebulization by static procedure, but
`to strive for a prolongation of
`nebulization time by slightly shaking and tapping the nebulizer.
`
`
`
`
`- P
`
`entamidine 60 mg/6 mi
`
`
`
`Liquidia's Exhibit 1098
`Page 10
`
`Liquidia's Exhibit 1098
`Page 10
`
`

`

`
`
`
`
`Influence of ultrasonic nebulization on the chemical stability of
`Pentamidine
`
`concentrations of pentamidine
`Nebulization liquids with various
`were nebulized for 30 minutes at different temperatures by an ultrasonic
`method. Additionally; pentamidine
`solutions were
`autoclaved. Under
`laboratory
`nebulization
`conditions
`‘no
`evidence was
`obtained
`of
`decomposition of pentamidine neither. with the ‘Porta-Sonic nor with the
`FISO Neb nebulizer.
`No
`decomposition was
`detectable
`even
`under
`ultrasonic treatment
`involving a higher energy supply using the Branson
`Sonifier 250 at
`room temperature for
`30 minutes. It was
`shown that
`decomposition reactions
`occur
`only at high
`temperatures
`of
`121°C
`(autoclaving) over a period of 20 minutes, affecting 10 - 15 2 of
`the
`pentamidine. Based
`on
`the
`results,
`the
`stability of
`pentamidine-
`containing liquids used for ultrasonic nebulization can be considered
`guaranteed.
`
`Particle size distribution
`
`each
`aerosols obtained by
`the
`Particle size distribution of
`nebulizer under conditions described in the experimental protocol are
`shown in FIGURE 5.
`.
`
`
`
`WRespirgard 11
`
`GRBFiso Neb
`
`Porta-Sonic
`
`
`
`water
`f. injection
`
`Pentamidine
`60 mg/6 ml
`
`Pentamidine
`300 mg/6 ml
`
`FIGURE 5: Mass median diameter
`
`lowest mass median diameter
`For the tested nebulized liquids the
`values 1.91 - 2.51 pum, were obtained by Respirgard II.
`In the case of
`the Porta-Sonic nebulizer the corresponding values were 1.96 - 3.04 um.
`The values for the FISO Neb nebulizer were 5.8 - 6.93 pm. A comparison
`of the data from FIGURE 5
`shows
`that pentamidine concentrations in the
`nebulization liquid of
`0
`-
`300 mg pentamidine/6 ml hardly influence
`particle diameter. A slight decrease of
`the mass median diameters were
`found for all 3 nebulizers at concentration of 300 mg pentamidine/6 ml.
`All measurements of particle size distribution during nebulization are
`shown in FIGURES 6 and 7.
`
`There was no increase in particle size diameter during nebulization
`for any of the liquids nebulized by the 3 nebulizers under study.
`
`75
`
`Liquidia's Exhibit 1098
`Page 11
`
`Liquidia's Exhibit 1098
`Page 11
`
`

`

`
`
`FIG
`
`int)
`conc
`mas<
`FIS(
`ear]
`dec]
`aerc
`162
`pent
`prin
`and
`obta
`nebu
`(eff
`labo
`nebu
`dete
`trea
`pent:
`in t
`dist:
`
`deper
`Sonic
`nebul
`have
`treat
`relat
`
`case
`corre
`confi
`
`tion
`diffr:
`
`Liquidia's Exhibit 1098
`Page 12
`
` IT rr time
`
`
`
` .
`
`v FiSO Neb
`
`.
`
`e
`
`5
`
`*Porta-Sonic

`
`. Respirgard ile
`
`10
`
`15
`
`20.
`
`[min]
`
`FIGURE 6: Change of MMD during nebulization
`Pentamidine 60 mg/6 ml
`
`MMD
`[om]
`
`77
`of.
`
`+ *
`
`+ FISO Neb
`
`-
`
`|
`
`.

`
`* Porta—-Sonic

`
`* Respirgard Ile
`
`= a
`
`fon
`of
`e
`2
`.
`.
`1
`
`DI tere A tim
`0
`5
`10
`15
`20
`{min}
`
`FIGURE 7: Change of MMD during nebulization
`Pentamidine 300 mg/6 ml
`
`Considering the percentage of pentamidine mass nebulized from 300
`mg pentamidine available for inhalation
`and the part of particle sizes
`S$ 2 pm see FIGURE 8 it’s evident that comparing the nebulizers used,
`the
`Porta-Sonic and the Respirgard II nebulizers deliver the highest amounts
`of pentamidine, with the value of about 43 mg available for alveolar
`deposition. For
`the FISO Neb nebulizer only 10.5 mg of particle sizes <
`2 um were found in spite of the highest total mass output of 201.4 mg.
`Similar relations between total mass output and available particle sizes
`< 2 um were observed for 60 mg pentamidine.
`
`DISCUSSION
`
`the
`already mentioned
`As
`alveolar deposition is
`only
`
`output
`a part
`
`of
`of
`
`pentamidine available for
`the
`amount
`of pentamidine
`
`76
`
`Liquidia's Exhibit 1098
`Page 12
`
`

`

`
`
`
`
`§ nebulized Pentamidine particles < 2 um
`
`total nebulized Pentamidine
`
`
`
`
`
`om 300
`2 sizes
`ad,
`the
`amounts
`lveolar
`sizes <
`1.4 mg.
`2 sizes
`
`le for
`amidine
`
`Pentamidine
`Pentamidine
`60 mg/6 mi
`300 mg/6 mi
`FIGURE 8: Output of total nebulized pentamidine and nebulized
`pentamidine with particle size < 2 pm
`
`laboratory standard
`the
`on
`independent
`introduced in the nebilizer
`conditions or clinical use. Comparisons between the present pentamidine
`masses nebulized found to be 80.0 mg and 201.4 mg for Respirgard II and
`FISO Neb by introducing 300 mg pentamidine into the nebulizers,
`and
`earlier published data for Respirgard II by Smaldone et al.
`(1988)
`declared as efficiency: 4.6 Z,
`by Smith et al.
`(1990) declared as
`aerosol output: 127.3 mg, and for FISO Neb by Smaldone et al.
`(1988):
`16Z and Smith et al.
`(1990): 82.7 mg,
`indicate that output of nebulized
`pentamidine is a sensitive function of various parameters: nebulizing
`principle, characteristics and attachment of nebulizer used, air flow
`and air pressure, duration of nebulization, breathing pattern etc.. For
`obtaining of comparable
`data under
`in vitro conditions concerning the
`nebulizing
`capacity,
`the
`terminology
`of
`nebulized
`pentamidine
`(efficiency, aerosol output, pentamidine mass distribution etc.) and the
`laboratory conditions
`should be
`standardized.
`For determination of
`nebulized
`pentamidine
`output
`in
`conditions
`of
`clinical
`use
`the
`determination
`of
`alveolar
`pentamidine
`deposition
`or
`inhalation
`treatment,
`should
`be correlated with investigations
`concerning the
`pentamidine retained in the nebulizer container,
`tubes and valves. Only
`in this way it will be possible to do a real balance of pentamidine mass
`distribution.
`
`The duration of nebulization determined under laboratory conditions
`depends specially on the nebulizer used: Respirgard II: 25 min, Porta-
`Sonic: 9-12 min, FISO Neb:
`5
`-
`6 min. Considering the duration of
`nebulization as treatment time in clinical uses, patient related factors
`have to be considered.
`In the case of Respirgard II for 24 patients a
`treatment
`time of
`15
`to 60 min with a mean
`(SD) of 29.1 (9.6)
`is
`related.
`(Vinciguera, Smaldone, 1990).
`
`evidence of decomposition of pentamidine was obtained in the
`No
`ultrasonic
`nebulizers
`Porta-Sonic
`and
`FISO Neb.
`Thus,
`of
`case
`corresponding
`data
`in the
`literature
`(Konietzko,
`1989) were
`not
`confirmed.
`
`aerosol particle size distribu-
`The published data of pentamidine
`determined
`by
`laser
`light-scattering
`(LL-S),
`laser
`light-
`tion
`diffraction (LL-D) and cascade-impactor
`(CI) methods are different not
`
`77
`
`Liquidia's Exhibit 1098
`Page 13
`
`
`Liquidia's Exhibit 1098
`Page 13
`
`

`

`
`

`
`only by using different nebulizers, but also using the same nebulizer
`and
`similar
`pentamidine
`concentration. Additional, with regard
`to
`particle size, different designations are used: mass median aerodynamic
`diameter
`(MMAD),
`in wm,
`a fictive size,
`referring to a model particle
`with specific weight of water,
`(Corkery et al.,
`1988; Herer et al.,
`1989; Dirnagel, 1971; Matthys.and Herceg, 1990 a, 1990 b); count median
`diameter
`(CMD),
`in’pm (Huchon, 1990); mass median diameter
`(MMD),
`in pm
`(Simonds et al., 1990); diameter,
`in pm (Dirnagel, 1971): volume median
`diameter D(V) 50 Z-value,
`in‘pm, equal
`to the mass median diameter when
`homogeneous. material of
`the same density is used.
`In literature the
`following data are described for the nebulizers Respirgard II and FISO
`Neb: Respirgard IT,
`(CI), MMAD = 0.76 pm, FISO Neb,
`(CI), MMAD = 2.5 pm
`(Smaldone et al., 1988); Respirgard II,
`(LL-D), MMD = 1.2 pm, FISO Neb,
`(LL-D), MMD = 4:7 pm (Smith et al., 1990); Respirgard II,
`(CI), MMAD =
`0.93 pm (Corkery et al., 1988), Respirgard II with baffle,
`(LL-D), MMD
`1 pm, without baffle,
`(LL-D), MMD = 1.7 pm (Simonds et al., 1989). These
`data indicate that smaller particle sizes are measured according to the
`cascade-impactor, method used specially for particle size measurements of
`powder-aerosols..” Laser
`light methods should be preferred, because they
`are able to measure the real droplet size distribution of nebulized
`pentamidine-containing solutions.
`In our investigations we measured mass
`median diameters between 1.9 - 2.5 pm for the Respirgard II nebulizer.
`This values were slightly,higher than the above mentioned data. Mass
`median
`diameters
`obtained
`from Porta-Sonic were
`comparable with
`Respirgard II, whereas the values measured for FISO Neb (5.8 - 6.9 wm)
`were much higher than expected. Due to this fact,
`the FISO Neb nebulizer
`delivers
`only
`small amounts
`of pentamidine
`available
`for alveolar
`deposition in spite of its high total pentamidine mass output. According
`to Simonds et al., 1990, we
`suggest
`from our results to use in the
`future for characterisation of pentamidine containing liquid-aerosols
`only measurement systems based on laser light-diffraction,
`laser light-
`scattering etc., systems, able’ to measure the droplet size distribution.
`
`ACKNOWLEDGMENTS
`
`The authors would
`technical assistance.
`
`like to thank Ernst Vélkel
`
`for his excellent
`
`REFERENCES
`
`(1988). Characteristics of
`CORKERY, K., LUCE, J.M., MONTGOMERY, A.B.
`Nebulizers used in Aerosolized Pentamidine Studies. Respiratory
`care. Vol. 33, No 10, 916.
`JOHNSON, M.A., TALAEE, N. LEE, C.A.,
`SIMONDS, A.K., NEWMAN, S.P.,
`CLARKE, S.W.
`(1990). Alveolar Targeting of Aerosol Pentamidine.
`Am. Rev. Respir. Dis. 141, 827-829.
`HERER, B., GIRAUD, V., MORDELET-DAMBRINE, M.,COLLIGNON, M.A.,
`CHRETIEN, J., HUCHON, G.J.
`(1989). Pentamidine particle size
`increases during aerosolization. Am. Rev. Respir. Dis.139, A248.
`KOHLER, D., SIMONIDES, R., ROTHFUSS, J., VATSAG, E., DAIKELER, G.,
`MATTHYS, H.
`(1983). Aerosolverteilungsmuster von 16 handelstiblichen
`Inhalationspraparaten. Prax: Klin. Pneumol. 37, 922-924.
`DIRNAGEL, K.
`(1971). Technisch-physikalische Grundlagen der
`Inhalationstheorie mit Aerosolen in ihrer Bedeutung fiir die
`Auswahl von Aerosolgeraten. Dtsch. Med. J. 22, 245-247.
`MATTHYS, H., HERCEG, R.
`(1990). Lungendeposition von 300 und 600 mg
`
`Pentamidin,
`inhaliert mit verschiedenen Verneblern. Med. Klin.
`85, 251-254.
`MATTHYS, H., HERCEG, R.
`(1990). Corporal Deposition of Aerosolized
`
`78
`
`Liquidia's Exhibit 1098
` Page 14
`
`Liquidia's Exhibit 1098
`Page 14
`
`

`

`Pentamidine 300 mg Versus 600 mg Inhaled With Three Different
`Nebulizers. Suppl.
`to JAMA SEA, Vol 6, No 2, 29-32.
`HUCHON, G.
`(1990). Effect of Different Nebulizers on the Particle
`Size of Aerosolized Pentamidine. Suppl.
`to JAMA SEA, Vol. 6,
`No 2, 33-35.
`(1990).
`O'DOHERTY, M.,; ‘THOMAS, S., PAGE, Ch., NUNAN, T., BATEMAN, N.
`Lung Deposition Following Aerosolization of Pentamidine with
`Different Nebulizers -.clinical Perspectives. Suppl.
`to JAMA SEA,
`Vol. 6, No 2, 36-38.
`KONIETZKO, N.
`(1989). Verteilung von Aerosolen in der Lunge.
`Die Pneumocystis carinii Pneumonie (Springer Verlag), 85-93.
`VINCIGUERRA, C., SMALDONE, G.
`(1990). Treatment Time and Patient
`Tolerance for Pentamidine Delivery by Respirgard II and
`AeroTech II. Respiratory care, Vol 35, No 11, 1037-1041.
`SMALDONE, G., PERRY, R., DEUTSCH, D.
`(1988). Characteristics of
`Nebulizers Used in the Treatment of AIDS-Related PCP. Journal of
`aerosol medicine, Vol.1, No 2, 113-126.
`WONG, R., HARDY, D.
`(1990). Focus on aerosolized pentamidine: An agent
`for prophylaxis of Pneumocystis carinii pneumonia. Hosp. Formul.,
`Vol. 25, 1061-1075.
`(1989). Comparison of
`SMITH, D., ERSKINE, D., STEELE, I. et al.
`nebulizer efficiency for aerosolizing pentamidine. Fifth Intern.
`Conference of AIDS, Montreal, Canada.
`(abstract TBP 64).
`BASKIN, M., ANTOINE, G. et al.
`(1990). Regional Deposition of
`Aerosolized pentamidine. Annals of Internal Medicine, Vol. No 113,
`677-683.
`
`Article received on August 12, 1991
`in final form February 26, 1992
`
`Reviewed by:
`Myrna Dolovich
`
`Address reprint requests to:
`Jérg Hager, Dipl. Ing.
`Pharmaceutical Development
`Rhone-Poulenc Rorer GmbH
`Nattermannallee 1
`D-5000 Kdln 30
`
`|
`
`ulizer
`rd
`to
`ynamic
`rticle
`= al.,
`median
`in pm
`median
`r when
`ze
`the
`d FISO
`2.5 pm
`O Neb,
`MMAD =
`MMD =
`These
`to the
`nts of
`e
`they
`ulized
`d mass
`lizer.
`Mass
`with
`-9 pm)
`ulizer
`reolar
`ording
`in the
`‘rosols
`
`light-
`ition.
`
`rllent
`
`then
`
`79
`
`Liquidia's Exhibit 1098
`Page 15
`
`
`Liquidia's Exhibit 1098
`Page 15
`
`

`

`
`
`Liquidia's Exhibit 1098
`Page 16
`
`Liquidia's Exhibit 1098
`Page 16
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket