throbber
3
`
`Particle size analysis by image analysis
`
`3.1 Introduction
`
`A microscope examination should always be carried out whenever a sample
`is prepared for particle size analysis. Such an examination allows an
`estimate of the particle size range of the powder under test and its degree of
`dispersion. If the dispersion is incomplete it can be determined whether
`this is due to the presence of agglomerates or aggregates and, if
`agglomeration is present, may
`indicate the need for an alternative
`dispersing procedure.
`Microscopy is often used as an absolute method of particle size analysis
`since it is the only method in which the individual particles are observed
`It is particularly useful in aerosol science where
`and measured [1-3].
`particles are often collected in a fonn suitable for subsequent optical
`examination. It is useful not only for particle size measurement but also for
`particle shape and texture evaluation, collectively called morphology, with
`sensitivity far greater than other techniques. Reports have also been
`presented of the use of microscopy to relate particle size to processing
`characteristics of valuable mineral ores [4,5]. Particle shape may be defined
`either qualitatively or quantitatively. The fonner includes the use of such
`tenns as acicularity, roundness and so on. The latter, more definitively,
`compares perpendicularly oriented diameters, for example, to obtain shape
`factors. The introduction of automatic image analyzers al lows for the
`factors which were previously
`detennination of complex shape
`unobtainable. These factors are of great value in defining crystal
`morpho logy and relating this to operating (attrition during conveying,
`compaction and so on) and end-use properties.
`A size analysis by number is simpkr to perform than an analysis by
`mass since, in the former, the statistical reliability depends solely on the
`number of particles measured. For a mass analysis the omission of a single
`I 0 µm particle leads to the same error as the omission of a thousand I µm
`
`

`

`Image analysis
`
`143
`
`particles since they both have the same volume. For a particle size analysis
`by mass, 25 particles in the largest size category have to be counted, in
`order to obtain an estimated standard error of less than 2%.
`If all the
`particles in this area were counted the final count would run into millions.
`It is obvious therefore that the area to be examined must decrease with
`decreasing particle size and the results obtained must be presented as
`particles per unit area.
`The problem may be likened to determining the size distribution of a
`number of differently sized homogeneous balls in a container. If the balls
`are of size 2, 2'12, 4, 4'12·· · ·· ··64, 64'12, in line with the size ratios often
`adopted in optical microscopy, and the relative frequency of the top size
`category is found to be 8 in 1000 particles this can be readily converted to a
`mass frequency when the number of balls in the other size categories is
`known. If the estimated mass of the 25 particles is 10% of the sample then
`If, on
`the forecasted percentage standard error is ( 10/'125) i.e. 2%.
`completing the analysis, the mass percentage of the coarsest fraction is
`greater than 10% then it is necessary to count more coarse particles in order
`to maintain this level of accuracy.
`The errors in converting from a number to a volume (mass) distribution
`are greatest when the size range is wide. For a narrowly classified powder,
`ranging in size from say I 0 to 30 µm, it is necessary to use an arithmetic
`grading of sizes, probably a 2 µm interval in this case, but the same rules
`still apply and the direct conversion of a number distribution to a weight
`distribution can still give rise to considerable error at the coarse end of the
`size distribution. Using closer size intervals adds little to analytical
`accuracy but can greatly increase the computation time.
`The images may be viewed directly or by projection. Binocular
`eyepieces are preferred for particle examination but monoculars for
`carrying out a particle size analysis since, by using a single eyepiece, the
`tube length can be varied to give stepwise magnification. Most
`experienced operators prefer direct viewing but projection viewing, less
`tiring to the eye, is often used for prolonged counting. Projection may be
`front or back. With the former the operation is carried out in a darkened
`room due to the poor contrast attainable. Back projection gives better
`illumination but image definition is poor; this can be rectified by using a
`system whereby two ground-glass screens are placed with their faces in
`contactt and one is moved slowly relative to the other [6].
`Some automatic counting and sizing devices work from photographic
`negatives or positives. The principle objection that can be leveled against
`photographic methods is that only particles in focus can be measured
`accurately and this can lead to serious bias. Although photographic
`
`

`

`144 Powder sampling and particle size determination
`
`methods are often convenient and provide a permanent record, the
`processing time may well offset any advantage obtained. This is
`particularly true when a weight count is required since, from statistical
`considerations, a large number of fields of view are required for accurate
`results.
`Light microscopy is best suited for the size range 0.8 to 150 µm, with a
`resolution of around 0.2 µm depending on the wavelength of the light
`source. Scanning ele-ctron microscopy (SEM) operates in the size range
`from 0.1 µm to I 000 µm with a resolution of I 0 nm and transmission
`electron microscopy (TEM) from 0.0 l µm to I 0 µm with a resolution of
`5 nm. Back scattered electrons and x-rays contain information on the
`chemistry and average atomic number of the material under the beam.
`Groen et. al. [7] determined the optimum procedure for automatic
`focusing of a microscope. Kenny [8] examined the errors associated with
`detecting the edge of the particle image and outlined a technique, suitable
`for automatic image analysis for minimizing this error.
`The shape and texture of construction aggregates are important
`parameters that have a direct bearing on the strength and durability of their
`asphalt and concrete end products. Typically, a batch of material is
`rejected if more than a specific fraction of particles have elongation and
`flatness ratio that exceed some limit.
`Jn the ASTM procedure [9] the
`measurements are carried out on l 00 particles using specially designed
`calipers. More recently this has been replaced by image analysis which
`reduces the measurement time to less than 10 minutes [IO]. In addition,
`is capable of conducting other useful particle
`this procedure
`characterization measurements without the need for additional image
`processing time. One such measurement incorporated into the design is
`roughness defined as '"surface irregularity" and 'jaggedness".
`Examples of determining both a number and a mass distribution are
`given below. Although the examples relate to manual counting, the
`conditions also govern size analyses by automatic image analyzers.
`
`3.2 Standards
`
`Relevant national standards are available covering particle size analysis by
`microscopy. BS 3406 Part 4 [11] is the British Standard guide to optical
`microscopy. The American standard ASTM E20 was discontinued in 1994
`[12). ASTM 175-82 [13]
`is a standard defining terminology for
`microscope related applications. ASTM E766-98 [14] is a standard
`practice for calibrating the magnification of an SEM. NF X 11-661 [ 15] is
`the French standard for optical microscopy. NF XI 1-696 [16] covers
`
`

`

`Image analysis
`
`145
`
`ISO/CD 13322 [I 7] is a draft
`general image analysis techniques.
`international standard on image analysis methods.
`
`3.3 Optical microscopy
`
`Optical microscopy is most often used for the examination of particles from
`about 3 µm to 150 µm in size, although a lower limit of 0.8 µm is often
`quoted. Above 150 µm a simple magnifying glass is suitable.
`The most severe limitation of optical transmission microscopy is its
`small depth of focus, which is about 10 µmat a magnification of lOOx and
`about S µm at I OOOx. This means that, for a sample having a wide range of
`sizes, only a few particles are in focus in any field of view. Further, in
`optical transmission microscopy, the edges of the particles are blurred due
`to diffraction effects. This is not a problem with particles larger than about
`5 µm s ince they can be studied by reflected light, but only transmission
`microscopy, with which silhouettes are seen, can be used for smaller
`particles.
`A two dimensional array of latex spheres is often used for measuring
`more or less uniformly sized lattices. Hartman [18-20] investigated the
`errors
`in
`this method which comprise focusing,
`image distortion,
`misreading of photomicrographs, distortions in the photographic material,
`anisotropy, other array defects, non-uniformity of particle size, coating of
`solutes on the lattices and contact deformation. Hartman introduced a new
`method, the center finding technique in which the latex spheres acts as
`lenses enabling the center-to-center distance to be determined with high
`accuracy (10 ± 0.4) µm for 10 µm particles. The National Physical
`Laboratory [21] introduced an NPL certified stage graticule [22] to test
`linearity over the complete image field.
`
`3.3.1 Upper size limit for optical microscopy
`
`The method is preferably limited to sub-200 mesh sieve size (75 µm) but
`larger particles may be counted and sized provided their fractional weight
`is less than 10% of the total weight of the powder. When the fractional
`oversize weight exceeds I 0%, these particles should be removed and a
`sieve and microscope analyses merged. Alternatively such large particles
`can be sized using a simple magnifying glass.
`
`

`

`146 Powder sampling and particle size determination
`
`3.3.2 Lower size limit/or optical microscopy
`
`The theoretical limit of resolution of an optical microscope is expressed by
`the fundamental formula:
`
`d _JA.
`l - NA
`
`(3.1)
`
`where dl is the limit ofresolution, i.e. particles in closer proximity than this
`appear as a single particle, A, is the wavelength of the illuminant, the
`numeriical aperture of the objective NA = µsin(} whereµ is the refractive
`index of the immersion medium, 0 is the angular aperture of the objective
`and f is a factor of about 0.6 to allow for the inefficiency of the system.
`For A,= 0.6 µm the resolving power is a maximum with NA = 0.95 (dry)
`and NA= 1.40 (wet) giving lower size limits, dmin = 0.38 µm and 0.26 µm
`respectively. The images of particles having a separation of less than these
`limits merge to form a single image.
`The resolution of the human eye is around 0.3 mm, therefore the
`maximum effective magnification with white light is:
`
`30mm~ 1000
`28µm
`
`Particles smaller than the limit of resolution appear as diffuse circles;
`image broadening occurs, even for particles larger than dmin.> and this
`results in oversizing. Some operators routinely size down to this level but
`the British Standard BS 3406 Part 4 [l l] is probably correct in stipulating a
`minimum size of 0.8 µm and limited accuracy from 0.8 to 2.3 µm.
`Powders containing material smaller than this are usually imaged by
`transmission or scanning electron microscopy and the resulting negatives or
`prints examined.
`Charmain (23) in an investigation into the accuracy of sizing by
`transmission optical microscopy, showed
`that for
`two-dimensional
`silhouettes greater than I µm in diameter, the estimated size under ideal
`conditions was about 0.13 µm too high; a 0.5 µm silhouettes gave a visual
`estimate of 0.68 µm and all silhouettes smaller than 0.2 µm appeared to
`have a diameter of 0.5 µm (Figure 3. I). The measurements were made
`with the circular discs immersed in oil. Due to less precise focusing with
`three-dimensional particles, real particles are subject to greater errors.
`
`

`

`Image analysis
`
`147
`
`Rowe [24] showed that wide differences m particle sizing can occur
`between operators because of this effect.
`
`3.4 Sample preparation
`
`Great care has to be taken in slide or grid preparation since the
`measurement sample is so small that it is difficult to make it representative
`of the bulk. Many particulate systems contain agglomerates and aggregates
`and, if it is necessary that they retain their integrity, the dispersing
`procedure needs to be very gentle. Further, since it is usually impossible to
`measure every particle in the measurement sample it is necessary that it be
`dispersed uniformly. Small regions selected at random or according to
`some predetermined plan must therefore be representative of the w hole.
`The ana lysis is suspect if the regions in o ne area of the measurement
`sample give a very different size distribution to those in another area.
`The simplest procedure
`is
`to extract samples from an agitated
`suspension; for less robust materials a procedure detailed in reference [3]
`may be used in which an air j et circulates the suspension through a
`sampling tube that can be closed and withdrawn to provide samples for
`analysis.
`S lides may be of three main types: dry, temporary and
`permanent. For very easily dispersed materia l, the particles may be
`
`3.0
`Visual
`estimate
`(µm)
`
`2.0
`
`1.0
`
`, , ,
`
`0.0
`
`0
`
`0.5
`
`1.5
`2
`1
`True disc diameter (µm)
`
`2.5
`
`3
`
`Fig. 3.1 Oversizing of small discs by optical m icroscopy [23)
`
`

`

`148 Powder sampling and particle size determination
`
`shaken from a fine brush or the end of a spatula on to a slide. Humphries
`[25) describes a microsample splitter that assists the fre.e flow of grains in
`order to provide the very small samples needed for microscopy: A diagram
`of the device is reproduced in a book by Hawkins [26]. Hawkins also
`describes a moving pavement version of the spinning riffler designed for
`preparation of representative samples of free flowing particles on
`microscope slides [27). A novel method of mounting particles on regularly
`spaced adhesive circles has also been developed. This method of mounting
`results in an ordered array rather than the random chaos of usual methods
`and greatly facilitates particle analysis [28,29].
`Some acceptable procedures for easily dispersed powders are described
`by Green [30] and Dunn [31]. For a temporary slide the powder can be
`incorporated into a viscous liquid, such as glycerin or oil, in which it is
`known to disperse completely. Some operators work the powder into the
`liquid with a flexible spatula; others roll it in with a glass rod. Either of
`these procedures can cause particle fracture and a preferable alternative is
`to use a small camelhair brush. A drop of this liquid can then be
`transferred to a microscope slide and a cover slip gently lowered over it.
`Rapid pressing of the cover slip must be avoided as it causes preferential
`transfer of the larger particles to the edge of the cover slip. It is undesirable
`for liquid to spread outside the limits of the cover slip; improved spreading
`is best effected with highly viscous liquids by pre-warming the microscope
`slide. Sealing the cover slip with amyl acetate (nail varnish is a good
`substitute) makes the slide semi-permanent.
`If low viscosity liquids are
`used it is necessary to have a well, or depression, on the slide to contain the
`dispersion.
`The method of Orr and Dallevalle [32) for the production of permanent
`slides is to place a small representative sample of the powder to be
`analyzed in a I 0 ml beaker, add 2 to 3 ml of a solution containing about 2%
`colloidon in butyl acetate, stir vigorously and place a drop of suspension on
`the still surface of distilled water in a large beaker. Prior to adding the
`suspension the surface is cleaned by allowing a drop of butyl acetate to fall
`on it. As the resulting film expands, it sweeps any particles on the surface
`to the walls of the beaker. As the drop of suspension spreads, the volatile
`liquid evaporates and the resulting film may be picked up on a clean
`microscope slide and completely dried. A dispersing agent may be added
`to prevent flocculation.
`Pennanent slides may also be produced by using the alternative
`combinations of Canada balsam or polystyrene in xylol, dammar in
`turpentine, gum arabic in glycerin, styrex in xylene, rubber in xylene and
`gelatin in water [33). With a l % solution this may be formed by dropping
`
`

`

`Image analysis
`
`149
`
`it on to the cleaned surface of distilled water; with a 0.5% solution it may
`be cast directly on to a microscope slide; spreading is accelerated if the
`slide is first washed in a detergent.
`Dullien and Mehta [34] use Cargill's series H compound, having a
`refractive index of 2.0, as a mounting medium for salt particles. This gives
`a transparent yellow background for the particles and, since it has a higher
`refractive index than salt, the particles appear as dark spots. A range of
`systems is necessary in order to select one where the difference in
`refractive index gives an easily detectable image.
`MiHipore recommend filtering a dilute suspension through a 0.2 µm
`PTFE membrane filter that is then placed on a dry microscope slide. The
`slide is then inverted over a watch glass half filled with acetone, the vapors
`of which render the filter transparent after two to three minutes.
`Harwood [35) describes two methods for dispersing difficult powders.
`One involves the use of electrical charges to repel the particles then fixing
`the aqueous solution with a gelatin-coated slide to overcome Brownian
`motion. The other, for magnetic materials, involves heating the sample to a
`temperature above the Curie point then dispersing it and fixing it on a slide
`to cool.
`Allen [36] mounted the powder directly into clear cement, dispersing it
`by using sweeping strokes of a needle and spreading the film on a
`microscope slide to dry. Lenz [37] embedded particles in solid medium
`and examined slices of the medium.
`Particles may also be suspended in a filtered agar solution that is poured
`on to a microscope slide where it sets in seconds [38]. Variations in
`analyses between these procedures may occur due to particles settling on
`the slide with preferred orientations. Ellison showed that if particles were
`allowed to fall out of suspension on to a microscope slide they would do so
`with a preferred orientation. Also, if the dispersing is not complete, the
`presence of floes will give the appearance of coarseness [39]. Pidgeon and
`Dodd [ 40], who were interested in measuring particle surface area using a
`microscope, developed methods for preparing slides of particles in random
`orientation. For sieve size particles, a thin film of Canada balsam was
`spread on the slide and heated until the liquid was sufficiently viscid,
`determined by scratching with fine wire until there was no tendency for the
`troughs to fill in. Particles sprinkled on the slide at this stage were held in
`random orientation. After a suitable hardening time, a cover glass coated
`with glycerol or wann glycerol jelly, was placed carefully on the slide.
`Sub-sieve powders were dispersed in a small amount of melted glycerol
`jelly. When the mixture started to gel a small amount was spread on a dry
`slide. After the mount had set, it was protected by a cover slip coated with
`
`

`

`I 50 Powder sampling and particle size determination
`
`glycerol jelly. With this technique, it is necessary to refocus for each
`particle since they do not lie in the same plane. Several of these techniques
`were examined by Rosinski et. al. [41) in order to find out which gave the
`best reproducibility.
`The sizing of fibrous particles by microscopy presents serious problems
`including overlapping. In order to minimize this it is necessary to work
`with only a few particles in the field of view at any one time. Timbrell
`[ 42,43] showed that certain fibers showed preferred orientations in a
`magnetic field, e.g. carbon and amphibole asbestos. He dispersed the fibers
`in a 0. 5% solution of colloidon in amyl acetate and applied a drop to a
`microscope slide, keeping the slide in a magnetic field until the film had
`dried. For SEM examination an aqueous film may be drained through a
`membrane filter held in a magnetic field. In order to reduce overlapping to
`an acceptable level it is necessary to use a far more dilute suspension than
`for more compact particles.
`Various means of particle identification are possible with optical
`identification of
`microscopy. These include dispersion staining for
`asbestos particles [ 44] and the use of various mounting media [ 45]. Proctor
`et. al. (46,47) dispersed particles in a solidifying medium of Perspex
`monomer and hardener. This was poured into a plastic mold that was
`slowly rotaled to ensure good mixing. Microscope analyses were carried
`out on thick sections; a lower size limit of 5 µm was due to contamination.
`Zeiss [48] describes a method for measuring sections of milled ferrite
`powder. The powder was mixed in 40:60 volume ratios wirth epoxy resin
`using a homogenizing head rotating at 25,000 rpm. The mixture was then
`poured into a 0.5 in diameter mold and cured al 60°C and 1000 psi to
`eliminate a ir bubbles. The casting was then polished in a vibratory polisher
`using 0.3 and 0.5 µm a lumina in water. A photomicrograph of the polished
`section was used for s11Jbsequent analysis.
`Automatic and quantitative microscopes tend to give erroneous results
`for transparent particles. To overcome this problem Amor and Block [49) a
`silver staining technique to make the particles opaque. The particles are
`dry-mounted on to a thin fi lm of tacky colloidon on a microscope slide.
`Si lver is then deposited from solution using the silver mirror reaction.
`Preliminary sensitizing the crystalline surface ensures that much more
`silver is deposited on the particles than on the colloidon. A method of
`sta ining particles in aqueous solution prior to deposition on a membrane
`fi lter for analysis is also given.
`Hamilton and Phelps [50] adapted the metal shadowing technique for
`the preparation of transparent profiles of dust particles. The process
`consisted of evaporating in vacuo a thin metal film in a direction normal to
`
`

`

`Image analysis
`
`151
`
`a slide containing particles. The particles are then removed by a jet of air
`or water, leaving sharp transparent profiles.
`
`3.5 Measurement of plane sections through packed beds
`
`When the size distribution of particles embedded in a continuous solid
`phase is required, the general approach is to deduce the distribution from
`the size of particle cross-section in a plane cut through the particle bed.
`The problem has occupied the attention of workers in diverse fields of
`science, who have tended to work in isolation and this has led to much
`duplication of effort. The historical development of this technique has been
`reviewed by Eckhoff and Enstad [51] and the relevant theory of Schei I by
`Dullien et. al. [52]. A theoretical analysis [53] has been criticized on
`several grounds [54].
`Dullien et. al. [55-57] examined salt particles embedded in a matrix of
`Wood's metal using the principles of quantitative stereology. They then
`leached out the salt particles and examined the matrix using mercury
`porisimetry. Poor agreement was obtained and this they attribute to the
`mercury porosimetry being controlled by neck diameter. Nicholson
`[58]considered the circular intersections of a Poisson distribution of
`spherical particles to estimate the particle size distribution. Saltzman et. al.
`[59] generated a computer based i.maging system for slices through a
`packed bed and found good experimental agreement.
`
`3.6 Particle size
`
`The images seen in a microscope are projected areas whose dimensions
`dependl on the particles' orientation on the slide. Particles in stable
`orientation tend to present their maximum area to the microscopist, that is
`the smaller dimensions of the particles are neglected, hence the sizes
`measured by microscopy tend to be greater t han those measured by other
`methods. Any one particle has an infinite number of linear dimensions
`hence, if a chord length is measured at random, the length wm depend upon
`the particle orientation on the slide.
`These orientation dependent
`measurements are known as statistical diameters, acceptable only when
`determined in such numbers as to typify a distribution. They are measured
`parallel to some fixed direction and are acceptable only when orientation is
`random; i.e. the distribution of diameters measured parallel to some other
`direction must give the same size distribution. They are representative of
`the two largest particle dimensions, since the smallest is perpendicular to
`the viewing plane if the particles are in stable orientation.
`
`

`

`152 Powder sampling and particle size determination
`
`Acceptable statistical diameters are:
`
`•
`
`•
`
`•
`
`•
`
`•
`
`Martin's diameter (dM) is the length of the line which bisects the area
`of the particle's projected area. The line may be in any direction,
`which must be maintained constant throughout the analysis [ 60,61].
`Feret's diameter (dF) is the distance between two tangents on
`opposite sides of the particle parallel to some fixed direction [62].
`Longest dimension. A measured diameter equal to the maximum
`value of Feret's diameter.
`Perimeter diameter (de) is the diameter of a circle having the same
`circumference as the particle.
`Projected area diameter (d0
`) takes into account both dimensions of
`the particle in the measurement plane, being the diameter of a circle
`having the same projected area as the particle.
`It is necessary to
`differentiate between this diameter and the projected area diameter
`for a particle in random orientation (dp) since, in this case, the third
`and smallest dimension of the particle 1s also included.
`
`The easiest diameter to measure is the Feret diameter but this is
`significantly larger than the other two diameters for most powders. It is
`probably best to reserve this diameter for comparison purposes and for
`rounded particles. Of the other two diameters, the projected area diameter
`is preferred since two dimensions are included in one measurement and the
`projected area is easier to estimate using globe and circle graticules than the
`length of the chord that bisects the image.
`It has been shown {63,64] that the relationship between specific surface
`and Martin's diameter is:
`
`s =~
`v d
`M
`
`(3.2)
`
`Since the surface-volume diameter is inversely proportional to Sv, the
`constant of proportionality being a minimum of six for spherical particles,
`Martin's diameter is systematically different to the surface-volume
`diameter. Experiments confirm that, on the whole, dM<da<dF. The ratios
`of these three diameters remains fairly constant for a given material and
`may be expressed as. a shape function. For example dpldM = 1.2 for
`Portland cement and 1.3 for ground glass [65].
`Heywood measured crushed sandstone which had passed through a
`I J/8 in. square aperture sieve and been retained on a 1 in. square aperture
`
`

`

`Image analysis
`
`153
`
`sieve. He determined the projected area with a planimeter and calculated
`the mean projected diameter; he next estimated the diameter using both the
`opaque and transparent circles on a globe and circle graticule and also
`determined Feret and Martin diameters. His conclusion, based on an
`examination of 142 particles, was that the Feret diameter was greatly
`different to the other diameters for elongated particles, but that the Martin
`and projected area diameters are sufficiently in agreement for all practical
`purposes. This was disputed by Walton [66] who showed that the Feret
`diameter, averaged over all particle orientations, was equal to the other
`diameters. Herdan [67] examined Heywood's data more rigorously and
`found that:
`
`(a) The Feret diameter was significantly different from the other four
`diameters.
`(b) The Martin diameter showed significant difference from that
`obtained using the globe and circle graticule if the planimeter data
`were accepted as standard.
`
`He concluded that there was no definite advantage to be gained by
`laboriously measuring profiles. As one might expect, the projected area
`diameters gave the best estimate of the true cross-sectional areas of the
`particles. This does not rule out the use of the other diameters if they are
`conveniently measured, since the cross sectional-area diameter of a particle
`is not necessarily its optimum dimension.
`
`3. 7 Calibration
`
`It is necessary to use a calibrated eyepiece scale when carrying out a
`microscope analysis. The simplest form consists of a glass disc that is
`fitted on to the field stop of the ocular. Engraved upon the disc is a scale
`that is calibrated against a stage micrometer placed in the object plane;
`typically this is a microscope slide on which is engraved a linear scale. The
`image of the scale is brought into coincidence with the ocular scale by
`focusing. With a single tube microscope the magnification be varied
`somewhat by racking the tube in or out. The stage graticule is then
`replaced by the microscope slide containing the sample. The microscope
`slide is made to traverse the eyepiece scale and particles are sized as the
`cross tlhe reference line. Linear eyepiece graticules labeled 0 to I 00 may be
`used to scan the sample so that the linear dimensions yields a size
`distribution as a function of the Martin or Feret diameter. Special
`graticules are also available containing globes (opaque images) and circles
`
`

`

`I 54 Powder sampling and particle size determination
`
`(transparent images). The former are designed for the sizing of opaque
`images and the latter for transparent images.
`
`3. 7.1 Linear eyepiece graticules
`
`These are linear scales, typically 10 mm, divided into I 00 divisions of
`100 µm, or 2 mm divided into I 00 divisions of 20 µm each. They are
`placed in the focus of the microscope eyepiece so that they are coincident
`with the image of the microscope slide on the microscope stage.
`Calibration is effected using a stage graticule, 10 mm (100 x 100 µm),
`I mm (JOO x 10 µm) or 100 µm (50 x 2 µm), which is placed in the object
`plane.
`Kohler illumination should be used [ 11] to give uniform illumination of
`the viewing plane. Using an oil immersion objective, it is possible to
`resolve down to about 1 µm, although a 15% oversizing is to be expected at
`this level due to diffraction effects.
`Ocular graticules having a
`linear scale are satisfactory for the
`measurement of linear dimensions of particles. Particle sizes obtained with
`a linear eyepiece graticule are best classified arithmetically hence it is most
`suited to particles having a narrow size range.
`
`3. 7.2 Globe and circle graticules
`
`Linear eyepiece graticules have been criticized on the grounds that the
`dimensions measured are greater than those determined by other methods.
`To overcome this objection, grids inscribed with opaque and transparent
`circles have been developed. For best results, opaque images are measured
`using the (opaque) globes while transparent images are best measured using
`the (transparent) circles. This permits direct comparisons between the
`projected areas of the particles and the areas of the circles. According to
`Cauchy the projected area is a quarter of the surface area for a random
`dispersion of convex particles (68] hence this measurement is fundamental
`to the properties of the powder.
`The earliest of these graticules by Patterson and Cawood has 10 globes
`and circles ranging in diameter from 0.6 to 2.5 µm when used with a
`+2 mm I OOx objective-eyepiece combination and is suitable for thermal
`precipitator work [69].
`Fairs [70] designed graticules covering a size range of 128: 1 using
`reference circles with a root two progression in diameter except for the
`smaller sizes. He considered this system to be superior to the Patterson-
`
`

`

`Image analysis
`
`155
`
`Cawood where the series is much closer. He also described a graticule,
`having nine circles in a '12 progression of sizes, for use with the projection
`microscope [71]. This was incorporated in a projection screen instead of
`being
`in the eyepiece and was adopted by the British Standards
`Organization [ 1 l].
`Watson [72] developed a graticule designed specifically to measure
`particles in the 0.5 to 5 µm (respirable dust) size range.
`May's graticule, [73] covers 0.25 to 32 µm in a root two progression of
`sizes (the lower limit is highly suspect).
`
`7
`
`4 •
`
`3 •
`
`2 •
`
`1
`•
`
`I
`I
`
`I
`I
`
`4
`0
`
`3
`
`0
`
`2
`0
`
`1
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket