throbber
United States Patent
`
`[191
`
`[11] Patent Number:
`
`5,233,990
`
`Barnea
`
`[45] Date of Patent:
`
`Aug. 10, 1993
`
`||||||ll||||||||lllllllllllllllllllllllllllllllllllllllllllllllllllllllllll
`
`Usoo523399oA
`
`[54] METHOD AND APPARATUS FOR
`DIAGNOSTIC IMAGING IN RADIATION
`THERAPY
`
`[76]
`
`Inventor: Gideon Barnea, 7887 E. Uhl St., No.
`410, Tucson, Ariz. 85710
`
`[2]] App]. No.: 819,957
`
`[22] Filed:
`
`Jun. 13, 1992
`
`Int. Cl.5 .............................................. .. A61B 5/05
`[51]
`[52] U.S. Cl. .................................. .. 128/653.1; 378/65
`[58] Field of Search ...................... .. 128/653.1; 378/65
`
`[56]
`
`References Cited
`U.S. PATENT DOCUMENTS
`
`l/I974 Pavltovich .......................... .. 378/65
`3,783,251
`4,123,660 l0/l978 Horwitz .............................. .. 378/65
`4,930,509 6/1990 Brisson ........................... .. 128/653.1
`
`OTHER PUBLICATIONS
`
`Droege, R. T. et al., “Influence of Metal Screens on
`Contrast in Megavoltage X—Ray Imaging,“ Med. Phys.
`6, 487-492, 1979.
`Lutz, W. R. et al., “A Test Object for Evaluation of
`Portal Film" Int. J. Radiat. Oncol. Biol. Phys. 11,
`631-634, 1985.
`Munro. P. et al., “Therapy Imaging: A Signal-to-Noise
`Analysis of Metal Plate/Film Detectors,” Med. Phys.
`14, 975-984, 1987.
`Mark, J. E. et al., “The Value of Frequent Treatment
`Verification Films in Reducing Localization Error in
`the Irradiation of Complex Fields," Cancer,
`37,
`2755-2761, 1976.
`I-luizenga et al. “Accuracy in Radiation Field Align-
`ment in Head and Neck Cancer: A Prospective Study,"
`Rad. Oncol. 11, 181-187, 1988.
`Pearcy et al., “The Impact of Treatment Errors on
`Post—0perative Radiotherapy for Testicular Tumors,”
`Br. J. Radiol. 58, 1003-1005, l985.
`Rabinowitz et al. “Accuracy of Radiation Field Align-
`ment in Clinical Practice," Int. J. Radiat. Oncol. Biol.
`Phys. 11, 1857-1867, 1985.
`
`Lam et al. “On-Line Measurement of Field Placement
`Errors in External Beam Radiotherapy," Br. J. Radiol.
`60, 361-367, 1987.
`Baily et al. “Fluoroscopic Visualization of Megavoltage
`Therapeutic X—Ray Beams," Int. J. Radiat. Oncol. Biol.
`Phys. 6, 935-939, 1980.
`Herk et al., “A Digital Imaging System for Portal Veri-
`fication,” in The Use of Computers in Radiation Ther-
`apy, I. Brunvis Ed., North Holland, 371-373, 1987.
`(List continued on next page.)
`
`Primary Examiner—Lee S. Cohen
`Assistant Examiner—Samuel Gilbert
`Attorney, Agent, or Firm—Ant0nio R. Durando; Harry
`M. Weiss
`
`[57]
`
`ABSTRACT
`
`An apparatus for diagnostic and verification imaging in
`radiation therapy that consists of attachments for stan-
`dard radiotherapy equipment comprising an x-ray tube
`and an x-ray detector placed on opposite sides of a
`patient along the main axis of the beam produced by the
`treatment unit. The detector is placed on a plane or-
`thogonal to the axis of the treatment beam and between
`the beam source and the patient, while the x-ray tube is
`placed on the other side of the patient, coaxially with
`the treatment beam and facing the detector. As a result
`of this configuration, the radiographic View of the x-ray
`beam, as seen on the detector, is equivalent to the view
`produced on the same detector by the therapeutic beam.
`varied only by parallax deviations that can be corrected
`by geometrical calculations. Accordingly, x-ray expo-
`sures and real-time verification of the position of a pa-
`tient can be obtained with the same unit used for treat-
`ment and without requiring movement of either patient
`or equipment. In addition, the apparatus enables a user
`to produce diagnostic images that can be used directly
`to manufacture shielding blocks in conventional shield-
`ing-block cutters.
`
`10 Claims, 2 Drawing Sheets
`
`4
`
`/20
`
`Page 1 of 10
`
`Elekta Exhibit 1012
`
`

`
`5,233,990
`
`Page 2
`
`OTHER PUBLICATIONS
`
`Shalev et al. “Video Techniques for On—Line Portal
`Imaging," Comp. Med. Imag. Graph. 13, 2l7—226, 1989.
`Munro et al. "A Digital Fluoroscopic Imaging Device
`for Radiotherapy Localization,” Int. J. Radiat. Oncol.
`Biol. Phys. 18, 641-649, 1990.
`Durham et al. “Portal Film Quality: A Multiple Institu—.
`tional Study,” Med. Phys. 11, 555-557, 1984.
`Biggs et al. “A Diagnostic X-Ray Field Verification
`Device for a 10 MV Accelerator," Int. J. Radiat. Oncol.
`Biol. Phys. 11, 635-643. 1985.
`Marks J. E. et al. “Localization in the Radiotherapy of
`
`I-Iodgkins Disease and Malignant Lymphoma with Ex-
`tended Mantle Fields,” Cancer 34, 83-90, 1974.
`Marks J. E. et al. “Dose—Response Analysis for Nasc-
`pharyngeal Carcinoma: An Historical Perspective,”
`Cancer 50, 1042-1050 1982.
`
`White J. E. et al. “The Influence of Radiation Therapy
`Quality Control on Survival, Response and Sites of
`Relapse in Oat Cell Carcinoma of the Lung, ” Cancer
`50, 1084-1090, 1982.
`
`Kinzie J. J. et al. “Pattems of Care Study: Hodgkins
`Disease Relapse Rates and Adequacy of Portals, " Can-
`cer 52, 2223-2226, 1983.
`
`Page 2 of 10
`
`

`
`U.S. Patent
`
`Aug. 10, 1993
`
`Sheet 1 of 2
`
`5,233,990
`
`(PRIOR ART)
`
`Page 3 of 10
`
`

`
`U.S. Patent
`
`Aug. 10, 1993
`
`Sheet 2 of 2
`
`5,233,990
`
`2
`
`- THERAPY RAD/AT/ON
`SOURCE
`
`2’ rHs£APmanrA7/o~
`
`SOURCE
`
`" ANATOM {CAL
`1. ANDMAR K
`
`INSH/ELD/N6-_...._
`
`' ' ANATOIWICA L
`LANDMAEK
`
`._..J__
`
`
`
`DIA GHOST/C
`N RADW./Oflsoume
`
`I
`22 \
`
`DIAGNOSTIC
`_ RADIATION SOURCE
`
`.
`'
`F13 3
`
`THERAPY
`
`FI’
`-7
`
`. 4
`
`RAD/AT’/ON
`souzcs
`HR
`:13. ............... - _ -
`VIDEO
`/ ’
`CAMERA
`
`, ’
`
`/
`
`MIRROR
`
`, ’
`
`
`FLUORESCENT
`oe.-rec:-oz PLANE
`
` PATIENT
`
`.
`
`F'_/9. 5
`
`DIAGNOSTC
`
`RAD/ATIOIIV 5ouRcE
`
`Page 4 of 10
`
`

`
`1
`
`5,233,990
`
`l0
`
`15
`
`35
`
`45
`
`50
`
`METHOD AND APPARATUS FOR DIAGNOSTIC
`IMAGING IN RADIATION THERAPY
`
`BACKGROUND OF THE INVENTION
`
`1. Field of the Invention
`
`This invention is related to the general field of radia-
`tion imaging for medical applications. In particular, the
`invention provides a new method and apparatus for
`producing a diagnostic image of the portion of the body
`affected by a tumor, so that the required dosage of
`radiation can be accurately delivered to the prescribed
`target volume.
`2. Description of the Prior Art
`The main object of radiotherapy is to deliver the
`prescribed dosage of radiation to a tumor in a patient
`while minimizing the damage to surrounding, healthy,
`tissue. Since very high energy radiation (produced at 4
`to 25 million volts, typically generated by a linear accel-
`erator) is normally used to destroy tumors in radiother-
`apy, the high energy is also destructive to the normal
`tissue surrounding the tumor. Therefore, it is essential
`that the delivery of radiation be limited precisely to the
`prescribed target volume (i.e., the tumor plus adequate
`margins), which is accomplished by placing appropri-
`ately constructed shielding blocks in the path of the
`radiation beam. Thus, the goal is to accurately identify
`the malignancy within the body of the patient and to
`target the prescribed dosage of radiation to the desired
`region on the immobilized patient.
`To that end, the ideal procedure requires the identifi-
`cation of the exact anatomical location of the tumor and
`
`the corresponding accurate positioning of the radiation
`field during treatment. This could be easily achieved if
`it were possible to locate and treat the tumor at the same
`time. In practice, though, this is not possible because the
`equipment used to identify the tumor (x-ray machine,
`computed tomography equipment, or the like) is sepa-
`rate from the equipment used for the therapeutical irra-
`diation of the patient, requiring the movement and repo-
`sitioning of the patient from one piece of equipment to
`the other. As illustrated in schematic form in FIG. 1, a
`conventional treatment unit 10 consists of a linear accel-
`erator (linac) head 2 mounted on a gantry 4 so that its
`collimated high-energy emissions HR irradiate a patient
`P lying on a gumey 6 directly below through shielding
`blocks 8 attached to the head. A bracket 12 supporting
`a detector 14 may be mounted on the opposite side of
`the head within the field of radiation in order to take
`
`radiographs of the patient being treated. The gantry 4 is
`movable around a pivot 16 to permit the rotation of the
`head (and of the detector) around the patient to afford
`different views of the area to be treated (“multiple
`fields” treatment). The normal procedure involves the
`use of a diagnostic simulator, which is a diagnostic x-ray
`machine with the same physical characteristics of the
`radiation therapy machine (schematically also repre-
`sented by FIG. 1, where a diagnostic x-ray head re-
`places the linac head 2), so that the field of view of the
`low-energy x rays emitted in the simulator is the same as
`that of the high-energy radiation emitted in the radia-
`tion therapy machine. Prior to treatment, the patient is
`radiographed using the simulator and an image of the
`target area is obtained with low-energy radiation (in the
`order of 100 kVp), which yields good image quality.
`The exact target volume is then delineated on the radio-
`graph by a physician and matching shielding blocks are
`constructed to limit the field of view of the irradiating
`
`Page 5 of 10
`
`2
`machine to the region so delineated. A mold of the
`shielding blocks is first cut out of plastic material (nor-
`mally polystyrene) with a shielding-block cutter, a ma-
`chine that reproduces exactly the relative positions of
`the linac head, the shielding blocks and the detector as
`they stand in the treatment unit. By using mechanical
`means, the shielding blocks are cut so that the field of
`irradiation from the treatment unit will corresponds
`exactly to the area delineated by the physician on the
`diagnostic radiograph. The final shielding blocks are
`then made from the mold with lead alloys that attenuate
`considerably the propagation of radiation. Thus, the
`shielding blocks function as templets that limit the radi-
`ation treatment to the areas left open within the contour
`of the shielding blocks. In addition, it is common prac-
`tice to mark the skin of a patient with reference mark-
`ings that are used in aligning the position of the patient
`with the field of emission of the radiation therapy ma-
`chine.
`
`These apparently sound procedures in fact suffer
`from serious practical shortcomings. Errors in position-
`ing the shielding blocks between the radiating source
`and the patient, as well as incorrect beam alignment and
`patient movement, all have a cumulative effect reducing
`the accuracy of the procedure. Even the markings on
`the skin of the patient may be the cause of alignment
`problems because of shifting of the skin with respect to
`the patient‘s internal anatomy as a result of body motion
`or, over a period of time, even of body changes. Thus,
`the area actually irradiated during the therapeutic ses-
`sion often does not correspond to the area delineated in
`the radiograph generated by the simulator.
`Positioning errors during irradiation have been found
`to have very serious consequences for the successful
`prognosis of the treatment. For example, researchers
`have been able to correlate the recurrence of lymphoma
`to such positioning errors ( J. E. Marks, A. G. Haus, H.
`G. Sutton and M. L. Griem, “Localization Error in the
`Radiotherapy of Hodgkin‘s Disease and Malignant
`Lymphoma with Extended Mantle Fields," Cancer 34,
`83-90, 1974); and it has been found that improved tumor
`control of nasopharingeal carcinomas can be related to
`greater accuracy in the delivery of calculated dosages
`of radiation (J. E. Marks, J. M. Bedwinek, F. Lee, J. A.
`Purdy and C. A. Perez, “Dose-Response Analysis for
`Nasopharyngeal Carcinoma: An Historical Perspec-
`tive," Cancer 50, 1042-1050, 1982). Similarly,
`it has
`been found that shielding inaccuracies have resulted in
`significantly lower primary tumor control and survival
`of patients of oat cell lung cancer (J. E. White, T. Chen,
`J. McCracken, P. Kennedy, H. G. Seydel, G. Hartman,
`J. Mira, M. Khan, F. Y. Durrance and 0. Skinner, “The
`Influence of Radiation Therapy Quality Control on
`Survival, Response and Sites of Relapse in Oat Cell
`Carcinoma of the Lung." Cancer 50, 1084-1090, 1982);
`and that the local recurrence of Hodgkin's disease was
`significantly higher when the radiation field did not
`adequately cover the tumor (J. J. Kinzie, G. E. Hanks,
`C. J. Maclean and S. Kramer, “Patterns of Care Study:
`Hodgkin’s Disease Relapse Rates and Adequacy of
`Portals,” Cancer 52, 2223-2226, 1983).
`The only technique widely used today to check the
`accuracy of the radiation field is by imaging with the
`radiotherapy beam itself at the time of treatment. Prior
`to treatment, a “portal" image is obtained by using the
`therapy beam (at high energy) and the resulting expo-
`sure is visually compared with that taken with the simu-
`
`

`
`3
`lator (at low energy). This technique is therefore known
`as "portal imaging” or “therapy verification," and is
`repeated periodically during the period of radiation
`treatment. Unfortunately. though, because of the high-
`energy radiation emitted by the treatment beam (pro-
`duced at 4-25 million volts), the resulting portal images
`have poor resolution and show very poor contrast be-
`tween soft tissues and bones, often making the images
`totally unsuited for verification by comparison with the
`low-energy images produced by the simulator. See, for
`example, R. T. Droege and B. J. Bjarngard, “Influence
`of Metal Screens on Contrast in Megavoltage X-Ray
`Imaging," Med. Phys. 6, 487-492, 1979; L. E. Reinstein,
`M. Durham, M. Tefft, A. Yu and A. S. Glicksman,
`“Portal Film Quality: A Multiple Institutional Study,”
`Med Phys. 11, 555-557, 1984; W. R. Lutz and B. E.
`Bjarngard, “A Test Object for Evaluation of Portal
`Film,” Int. J. Radiat. Oncol. Biol. Phys. 11, 631-634,
`1985; and P. Munro, J. A. Rawlinson and A. Fenster,
`“Therapy Imaging: A Signal-to-Noise Analysis of 20
`Metal Plate/Film Detectors," Med. Phys. 14, 975-984,
`1987. Indeed, positioning errors occur very frequently
`in spite of the use of portal images. See J. E. Marks, A.
`G. Hans, H. G. Sutton and M. L. Griem, “The Value of
`Frequent Treatment Verification Films in Reducing 25
`Localization Error
`in the Irradiation of Complex
`Fields,” Cancer 37, 2755-2761, 1976; R. W. Byhardt, J.
`D. Cox, A. Homburgh and G. Lierrnann, “Weekly
`Localization Films and Detection of Field Placement
`Errors," Int. J. Radiat. Oncol. Biol. Phys. 4, 881-887,
`1978; Huizenga, P. C. Lenendag, P. M. Z. R. De Porre
`and A. G. Visser, “Accuracy in Radiation Field Align-
`ment in Head and Neck Cancer: A Prospective Study,"
`Rad. Oncol. 11, 181-187, 1988; R. G. Pearcy and S. E.
`Griffiths, “The Impact of Treatment Errors on Post-
`Operative Radiotherapy for Testicular Tumors," Br. J.
`Radiol. 58, 1003-1005, 1985; I. Rabinowitz, J. Broom-
`berg, M. Goitein, K. McCarthy and J. Leong. “Accu-
`racy of Radiation Field Alignment in Clinical Prac-
`tice," Int. J. Radiat. Oncol. Biol. Phys. 11, 1857-1867, 40
`1985; and W. C. Lam, M. Partowmah, D. J. Lee, M. D.
`Wharam and K. S. Lam, “On-Line Measurement of
`Field Placement Errors in External Beam Radiother-
`
`apy," Br. J. Radiol. 60, 361-367, 1987. Imaging devices
`other than X-ray film have been used in an attempt to
`improve the quality of the image produced during ther-
`apy verification. These include metal and fluorescent
`screens in contact with conventional film, and non-film
`imaging processes and devices such as xeroradiogra-
`phy, liquid ionization chambers, fluoroscopic imaging,
`linear diode arrays, photostimulable phosphors, and
`others. In addition, various image processing techniques
`(both analog and digital) have been used to enhance the
`quality of the final verification image; but all
`these
`methods and devices have resulted only in a limited
`success in yielding a good quality, and therefore useful,
`diagnostic image. Real-time portal imaging using video
`techniques has also been proposed, so that patient
`movement can be monitored during treatment. Because
`they all use the high-energy therapy beam as the source
`of radiation, though, the quality of the image remains
`poor. See N. A. Baily, R. A. Horn and T. D. Kampp,
`"Fluoroscopic Visualization of Megavoltage Therapeu-
`tic X-Ray Beams,” Int. J. Radiat. Oncol. Biol. Phys.6.
`935-939, 1980; M. V. Herk and H. Meertens, “A Digital
`Imaging System for Portal Verification,” in “The Use
`of Computers in Radiation Therapy," I. Brunvis Ed.,
`North Holland, 371-373, 1987; S. Shalev, T. Lee, K.
`
`Page 6 of 10
`
`5,233,990
`
`4
`Leszczynski, S. Cosby and T. Chu, “Video Techniques
`for On-Line Portal
`Imaging,” Comp. Med.
`Imag.
`Graph. 13, 217-226, 1989; and P. Munro, J. A. Rawlin-
`son and A. Fenster, “A Digital Fluoroscopic Imaging
`Device for Radiotherapy- Localization,” Int. J. Radiat.
`Oncol. Biol. Phys. 18, 641-649, 1990. A survey of 23
`different radiotherapy departments shows that at each
`of eight institutions (i.e., 35 percent of the 23 institutions
`sampled) more than i of the submitted portals were
`evaluated as poor in quality. Furthermore, it shows that
`approximately one-half of the institutions were produc-
`ing poor-quality films at a rate of at least 50 percent. See
`Reinstein, L. E., M. Durham, M. Tefft, A. Yu, and A. S.
`Glicksman, “Portal Film Quality: A Multiple Institu-
`tional Study," Med. Phys. 11, 555-557, 1984.
`Another, logical, approach to obtaining diagnostic
`quality portal films has been by mounting an x-ray tube
`on the head of the treatment unit as close to the linac
`
`gantry as possible. See P. J. Biggs, M. Goitein and M.
`D. Russell, “A Diagnostic X-Ray Field Verification
`Device for a 10 MV Accelerator," Int. J. Radiat. Oncol
`Biol. Phys. 11, 635-643, 1985. The x-ray tube is aligned
`with the linac emission field so that, to the extent possi-
`ble within the physical constraints of both devices, the
`x~ray emissions have the same field of view of the high-
`energy radiation. As a result. the image received on a
`film placed on a detector tray on the opposite side of the
`patient by exposure to either source of radiation is theo-
`retically almost exactly the same. In order to implement
`this approach, though, a special shielding-block holder
`coupled to the gantry has to be made, disabling the
`normal rotation of the linac’s collimator and limiting the
`adjustment capabilities of the equipment. Thus,
`the
`complexity of the procedure, the oblique view of the
`diagnostic beam and the increased time required for
`each treatment have prevented this technique from
`gaining widespread acceptance. Furthermore,
`this
`method is unsuited for real-time portal imaging.
`A similar approach has been followed by placing an
`x-ray tube at a fixed angle with respect to the axis of the
`therapeutic beam, so that the x-ray beam and the thera-
`peutic beam have coinciding isocenters corresponding
`to the location of the radiation target. By rotating the
`gantry of the radiation unit by that angle, the target can
`be irradiated from the same point either with a treat-
`ment beam or an x-ray beam, with every other variable
`remaining unchanged. Therefore, verification can be
`obtained simply by rotating the gantry and switching
`from one mode of operation to the other. The main
`problem with this approach is the inevitable angular
`error introduced during the rotation of the gantry. In
`addition, because of the alternative use of either mode
`of operation, this equipment is also not suitable for real
`time verification.
`
`Therefore, it would be very desirable to have a sim-
`pler and more accurate verification imaging system for
`radiation therapy verification, especially for real time
`applications. This invention relates to the use of a con-
`ventional x-ray tube and conventional imaging devices
`in a novel geometric configuration to produce such an
`improved verification imaging system.
`
`BRIEF SUMMARY OF THE INVENTION
`
`One objective of this invention is the development of
`therapy verification apparatus that produces verifica-
`tion images of the same quality obtained with diagnostic
`apparatus.
`
`

`
`5,233,990
`
`6
`
`DETAILED DESCRIPTION OF THE
`INVENTION
`
`5
`Another objective of the invention is an imaging
`apparatus that can be implemented as an accessory to
`existing radiation-therapy treatment units.
`A further goal of the invention is an imaging appara-
`tus and technique that are suitable for on-line, real time,
`applications in conjunction with radiation treatments.
`Still another objective of the invention is an apparatus
`that, as a result of the position of the x-ray emission
`source, produces an image corresponding to the same
`held of view of the linac beam used in conjunction with 10
`It.
`
`The heart of this invention lies in the recognition that
`5 verification of the correct position of a patient during
`treatment can be achieved by placing a detector be-
`tween the treatment beam source and the patient, and
`placing an x-ray tube along the axis of the treatment
`beam on the opposite side of the patient. Given the
`position of the various components, a fixed geometrical
`relationship exists that permits the direct construction
`of verification images for immediate use during treat-
`ment.
`
`A final objective of this invention is the realization of
`the above mentioned goals in an economical and com-
`mercially viable manner. This is done by utilizing com-
`ponents and methods of manufacture that are either
`already available in the open market or can be devel-
`oped at competitive prices.
`According to these and other objectives, the present
`invention consists of attachments for standard radio-
`therapy equipment comprising an x-ray tube and an
`x-ray detector placed on opposite sides of a patient
`along the main axis of the beam produced by the treat-
`ment unit. The detector is placed on a plane orthogonal
`to the axis of the treatment beam and between the beam
`
`source and the patient, while the x-ray tube is placed on
`the other side of the patient, coaxially with the treat-
`ment beam and facing the detector. As a result of this
`configuration, the radiographic view of the x-ray beam,
`as seen on the detector, is equivalent to the view pro-
`duced on the same detector by the therapeutic beam,
`varied only by parallax deviations that can be corrected
`by geometrical calculations. Accordingly, x-ray expo-
`sures and real-time verification of the position of a pa-
`tient can be obtained with the same unit used for treat-
`ment and without requiring movement of either patient
`or equipment.
`Various other purposes and advantages of the inven-
`tion will become clear from its description in the specifi-
`cation that follows and from the novel features particu-
`iarly pointed out in the appended claims. Therefore, to
`the accomplishment of the objectives described above,
`this invention consists of the features hereinafter illus-
`
`trated in the drawings, fully described in the detailed
`description of the preferred embodiment and particu-
`larly pointed out in the claims. However, such drawings
`and description disclose but one of the various ways in
`which the invention may be practiced.
`
`BRIEF DESCRIPTION OF THE DRAWINGS
`
`FIG. 1 is an elevational schematic representation of a so
`typical radiation therapy unit.
`FIG. 2 is an elevational schematic representation of
`the apparatus of the present invention, illustrated as an
`attachment to the typical radiation therapy unit shown
`in FIG. 1.
`
`FIG. 3 is a diagrammatic representation of the geo-
`metrical relationship of the various points of interest in
`the apparatus of FIG. 2.
`FIG. 4 shows, with reference to the geometry of the
`therapeutic unit represented in FIG. 3, the distance at
`which a diagnostic image must be positioned from the
`simulated radiation source’s position in a shielding-
`block cutter in order to have perfect correspondence
`with a picture produced with a simulator.
`FIG. 5 illustrates in schematic form the us of a fluo-
`rescent screen detector in conjunction with a mirror
`and a video camera to produce real-time verification
`images with the apparatus of the invention.
`
`Page 7 of 10
`
`Referring to the drawings, wherein like parts are
`identified with like symbols and numerals throughout
`this‘ specification, FIG. 2 illustrates in schematic eleva-
`tional representation the verification apparatus 20 of
`this invention, shown as an attachment to a standard
`radiation treatment unit (as illustrated in FIG. 1). An
`x-ray tube 22 is mounted on a support bracket 24 on the
`bottom side of the gantry 4, possibly replacing the de-
`tector 14 shown in FIG. 1. The x-ray tube is positioned
`facing up along the axis A of the beam radiated by the
`linac head 2, so that the axis of the x-ray beam is coaxial
`with that of the treatment beam. At the same time, a
`bracket 25 supporting a detector 26 is mounted on the
`gantry (or otherwise placed in the same position) be-
`tween the patient P and the linac head, as close to the
`patient as practicable. Thus, the detector 26 may be
`exposed either to the high-energy beam HR produced
`by the linear accelerator head 2 or to the low-energy
`beam LR (x rays) produced by the x-ray tube 22, or to
`both beams at the same time.
`
`FIG. 3 is a diagrammatic representation of the geo-
`metrical relationship of the various points of interest in
`the apparatus of FIG. 2. Point 2' represents the location
`of the source of high-energy radiation HR emitted by
`the linac head 2 and contained by the boundary 8’ of the
`shielding blocks 8; similarly, point 22’ corresponds to
`the location of the coaxial source of low-energy radia-
`tion LR emitted by the x-ray tube 22; and the line A’
`corresponds to the axis common to the two beams. Line
`26' represents the location of the detector 26 and point
`P’ represents a point at the boundary of the target vol-
`ume. S, and S4 are the distances of the high-energy
`radiation source (therapeutic) and the low-energy radia-
`tion source (diagnostic), respectively, from the detec-
`tor. Obviously, the view of point P’ on the detector 26',
`as projected by the x-ray beam,
`is different from the
`view seen on the same plane by the treatment beam.
`Point P’ is projected a distance Xdfrom the axis by the
`diagnostic beam (LR), but it is seen (back-projected) at
`a distance X; from the axis by the treatment beam (HR).
`The relationship between Xdand X, is a function of the
`distance d between the detector 26' and the point P’, the
`two being obviously the same for d=0. Simple trigo-
`nometry permits the determination of the following
`general relationship between these variables:
`
`Xd=IXg(I+d/S1)/(1 -d/SJ).
`
`(l)
`
`where S, and S4 are the distances of the high-energy
`radiation source and the low-energy radiation source,
`respectively, from thedetector.
`Therefore, for a given physical configuration of the
`equipment (i.e., for given values of S4, S1 and d), the
`relationship between any point in the image created by
`the x-ray tube and the location of the same point in the
`
`

`
`7
`corresponding image seen on the detector by the thera-
`peutic beam is linear and fixed. That is, one can be
`obtained from the other by a simple parallax correction
`according to the equation given above. Therefore, the
`field of view of the therapeutic beam on the detector
`can be correlated to the image projected on the plane by
`the diagnostic beam by a simple computation, and the
`two fields of view can be compared for verification
`purposes.
`In practice, shielding blocks will have been manufac-
`tured as outlined above and the verification task before
`
`the treatment
`is directed at ensuring that
`treatment
`beam, as attenuated by the blocks,
`is irradiating the
`target area delineated in the original diagnostic picture.
`If that picture was taken using a conventional simulator
`(i.e., an x-ray machine with the exact same geometry of
`the treatment unit). the shielding blocks are developed
`by conventional techniques so that the outline of the
`permitted field of view of the treatment beam coincides
`with the area delineated by the physician on the diag-
`nostic image. If, for example, point P’ corresponds to a
`point in the diagnostic image outlined by the physician
`as a boundary for radiation, the shielding blocks are
`shaped to permit the treatment beam to irradiate the
`area between the axis A’ and point P’ only. and the
`object of verification is to see whether that point is
`indeed at the boundary of the therapeutic beam during
`treatment. As mentioned above, current practice in-
`volves taking a picture with the high-energy treatment
`beam and a detector below the patient, so that the new
`image can in theory be compared with the diagnostic
`image. Unfortunately, though, the poor quality of the
`image obtained with the high-energy treatment beam
`renders it, in many cases, nearly useless in practice.
`The present invention addresses this problem by pro-
`ducing a coaxial image taken with an x-ray machine. By
`creating an x-ray image on the detector 26 through
`exposure to a low-energy beam from the tube 22, a good
`quality image of a patient‘s anatomy is obtained that can
`be corrected by the parallax relationship given above to
`produce an exact replica of the corresponding image
`seen by the therapeutic beam. Therefore, the resulting
`corrected picture is in the same scale of. and can be
`directly compared with, the original diagnostic image
`outlined for targeting the treatment area. As this can be
`done just before treatment, when the patient is posi-
`tioned for therapeutic irradiation, it can be an extremely
`useful
`tool for therapy verification. In addition, the
`verification image can be superimposed on an image
`obtained by exposing the detector on the plane 26’ to
`the shielded therapeutic beam, which in practice will
`only show the contour of the shield (i.e., the boundary
`of the radiation field) because of the high-energy radia-
`tion. These two pictures combined correspond exactly
`to the field of radiation currently targeted by the treat-
`ment beam. Therefore, they provide the radiotherapist
`with a current verification of the area targeted for treat-
`ment.
`
`According to another method of use of the apparatus
`of this invention, it is possible to use the x-ray tube 22
`also for the initial diagnostic image (normally taken on
`film); that is, it may be used as a substitute for the con-
`ventional simulator. In that case,
`the position of the
`diagnostic film in the shielding-block cutter can be ad-
`justed to permit its use to cut the shielding blocks as if
`the diagnostic image had been produced by a simulator.
`As would be obvious to one skilled in the art, FIG. 4
`shows that for a given geometry of the therapeutic unit,
`
`Page 8 of 10
`
`5,233,990
`
`8
`
`taken for example as illustrated in FIG. 3, there exists a
`distance Y at which the film must be positioned from
`the simulated high-energy radiation source's position in
`the shielding-block cutter in order to have perfect cor-
`respondence with a picture produced with a simulator.
`Again, simple trigonometry shows that distance is given
`by the following equation:
`
`Y= S,(l + d/S,)/( I —d/S4).
`
`(2)
`
`Thus, this procedure eliminates the need for a simulator
`as a separate piece of equipment. Moreover, since both
`diagnostic and verification images are taken with the
`same equipment, the x-ray source 22, no correction of
`the verification images is required for comparison with
`the target area in the diagnostic image. The two sets of
`pictures are automatically available in the same scale.
`taken from the same point and perspective, and of the
`same acceptable quality. The only requirement is the
`parallax correction of the radiation field boundary pro-
`duced by the treatment beam HR as a superimposed
`image on the detector, so that it is converted to the same
`scale of the verification image. Equation 1 above is used
`for this correction.
`
`Finally, because this invention does not require
`movement of the patient or of the treatment unit, it is
`suitable for on-line, real-time, application, which ren-
`ders it particularly valuable for radiation therapy. Many
`detectors exist that produce a real-time image. either
`directly or through computerized image enhancement
`processes. For example, FIG. 5 illustrates in schematic
`form the use of a fluorescent screen detector in conjunc-
`tion with a mirror and a video camera to produce real-
`tirne verification images with the apparatus of the in-
`vention. By appropriately positioning the mirror (such
`as, for example, at a 45 degree angle with the detector),
`the video camera can be placed outside the therapeutic
`field of radiation (at a 90 degree angle with the detec-
`tor), so that the image created on the fluorescent screen
`detector by the diagnostic beam is received by the video
`camera during the treatment session without interfer-
`ence. If the patient moves, the image on the detector
`will immediately reveal any change in the area being
`irradiated, so that immediate steps can be taken to mini-
`mize any localization error.
`Note that radiation dosages during therapy are usu-
`ally within the range of 3-10 cGy (rads), while a normal
`dose for diagnostic x-ray imaging purposes is usually
`within the range of 0.0]-0.1 cGy (rads). Therefore,
`detector sensitivity at diagnostic energies is about two
`orders of magnitude higher th

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket