throbber

`
`CENTER FOR DRUG EVALUATION AND RESEARCH
`
`Guidance for Industry
`
`The FDA published Good Guidance Practices in February 1997.
`This guidance was developed and issued prior to that date.
`
`Additional copies are available from:
`Office of Training and Communications
`Division of Communications Management
`Drug Information Branch, HFD-210
`5600 Fishers Lane
`
`Rockville, MD 20857
`
`(Tel) 301 —82 7-45 73
`(Internet) http://www.fda. gov/cder/guidance/index. htm
`
`US. DEPARTMENT OF HEALTH AND HUMAN SERVICES, FOOD AND DRUG ADMINISTRATION
`
`Page 1
`
`CUBIST 2203
`
`AGILA V. CUBIST
`IPR2015-00143
`
`CUBIST 2203
`AGILA v. CUBIST
`IPR2015-00143
`
`

`

`GUIDELINE ON
`
`VALIDATION OF THE LIMULUS AM EBOCYTE LYSATE TEST.
`AS AN END-PRODUCT ENDOTOXIN TEST FOR HUMAN
`AND ANIMAL PARENTERAL DRUGS, BIOLOGICAL PRODUCTS, AND
`MEDICAL DEVICES
`
`US. DEPARTMENT OF HEALTH AND HUMAN SERVICES
`PUBLIC HEALTH SERVICE
`FOOD AND DRUG ADMINISTRATION
`
`Page 2
`
`Page 2
`
`

`

`GUIDELINE ON
`VALIDATION OF THE LIMULUS AMEBOCYTE LYSATE TEST
`AS AN END—PRODUCT ENDOTOXIN TEST FOR HUMAN
`
`AND ANIMAL PARENTERAL DRUGS, BIOLOGICAL PRODUCTS, AND
`MEDICAL DEVICES
`
`December 1987
`
`Prepared by: Center for Drug Evaluation and Research
`Center for Biologic Evaluation and Research
`Center for Devices and Radiological Health
`Center for Veterinary Medicine
`
`Maintained by: Division of Manufacturing and Product Quality (HFN—BZO)
`Office of Compliance
`Center for Drug Evaluation and Research
`Food and Drug Administration
`5600 Fishers Lane
`
`Rockville, MD 20857
`
`Page3
`
`Page 3
`
`

`

`CONTENTS
`
`INTRODUCTION
`
`I.
`
`Background ...... ..... ..................... . ................ .....1
`
`II.
`
`Legal Effect ........................... . ........................ 3
`
`III. Regulatory Requirements ......................................... 4
`
`IV.
`
`Human and Animal Drugs and Biological Products .................. 6
`
`A. Validation of the LAL Test ................................... 6
`
`B. Testing of Drugs by the LAL Test ............................. 8
`
`V.
`
`Medical Devices...... .......................................... 11
`
`A. Validation of the LAL Test.... .......................... ....11
`
`B. Testing of Devices by the LAL Lest .......................... 14
`
`VI.
`
`Appendices ................... . ......... . ....... . ......... . ..... l6
`
`A. Initial Quality Control ..................................... 17
`
`B. USP XXI/NF XIV Bacterial Endotoxins Test .................. ..18
`
`C. Determination of the Relationship Between the Control
`
`Standard Endotoxin and the Reference Standard Endotoxin ..... 19
`
`D. Maximum Valid Dilution Calculation .......................... 22
`
`E. Maximum Dose and Endotoxin Limit Table ...................... 25
`
`Page4
`
`Page 4
`
`

`

`INTRODUCTION
`
`This guideline sets
`
`forth acceptable conditions
`
`for use of
`
`the
`
`Limulus Amebocyte Lysate
`
`test.
`
`It also describes procedures
`
`for
`
`using this methodology as an end—product endotoxin test
`
`for human
`
`injectable drugs (including biological products),
`
`animal
`
`injectable
`
`drugs,
`
`and medical devices.
`
`The procedures may be used in lieu of
`
`the rabbit pyrogen test.
`
`For
`
`the purpose of
`
`this guideline,
`
`the terms "lysate" or "lysate
`
`reagent"
`
`refer only to Limulus Amebocyte Lysate licensed by
`
`the
`
`Center
`
`for Biologic Evaluation and Research.
`
`The
`
`term "official
`
`test" means
`
`that
`
`a
`
`test
`
`is
`
`referenced
`
`in
`
`a United States
`
`Pharmacopeia drug monograph, a New Drug Application, New Animal Drug
`
`Application or a Biological License.
`
`PageS
`
`Page 5
`
`

`

`i.
`
`BACKGROUND
`
`FDA announced that Limulus
`In a notice of January 12, 1973 (38 FR 1404),
`circulating
`blood
`cells
`Amebocyte
`Lysate
`(LAL),
`derived
`from
`(amebocytes) of the horseshoe crab,
`(Limulus polyphemus),
`is a biological
`product.
`As such,
`it
`is subject
`to licensing requirements as provided in
`section 351 of
`the Public Health Service Act
`(42 U.S.C. 262).
`Since
`1973,
`LAL has proved to be
`a sensitive indicator of
`the presence of
`bacterial
`endotoxins
`(pyrogens).
`Because
`of
`this
`demonstrated
`sensitivity, LAL can be of value in preventing the administration or use
`of products which may produce fever, shock, and death if administered to
`or used in humans or animals when bacterial endotoxins are present.
`
`available data and
`1973 notice was published,
`the January 12,
`When
`its adoption. as
`the
`experience with IJU. were not adequate to support
`final pyrogen test
`in place of
`the rabbit pyrogen test, which had been
`accepted and recognized for many years.
`In order
`to establish.
`a data
`base and gain experience with the use of LAL,
`that notice permitted the
`introduction of LAL into the marketplace without
`a
`license.
`This was
`upon the condition that its use be limited to the in—process testing of
`drugs
`and other products,
`that
`the decision to
`use
`it
`be
`reached
`voluntarily by affected firms,
`and the labeling on. LAL state that
`the
`test was not suitable as a replacement for the rabbit pyrogen test.
`
`time, production techniques have been. greatly improved and
`Since that
`standardized so
`that
`they consistently yield LAL with an
`endotoxin
`sensitivity over 100 times greater
`than originally obtained. Moreover,
`it is widely recognized that
`the LAL test is faster, more economical, and
`requires a smaller volume of product
`than does
`the rabbit pyrogen test.
`In addition,
`the procedure is less labor intensive than the rabbit test,
`making it possible to perform many tests in a single day.
`
`In a notice published in the Federal Register of November 4, 1977 (42 FR
`57749),
`FDA described conditions for
`the use of LAL as an end—product
`test
`for endotoxins
`in. human biological products
`and medical devices.
`The notice stated further that
`the application of LAL testing to human
`drug products would
`be
`the
`subject
`of
`a
`future Federal Register
`publication.
`
`for Devices and
`now FDA's Center
`then Bureau of Medical Devices,
`The
`for
`the use of
`Radiologic Health (CDRH),
`issued recommended procedures
`LAL testing as an end—product endotoxin test on March 26, 1979.
`These
`procedures were
`revised as
`a
`result
`of
`the
`comments
`received
`from
`interested parties .
`
`result of CDRH'S experience in approving petitions for the
`a direct
`As
`LAL
`test
`in place of
`the
`rabbit pyrogen
`test,
`several
`use of
`the
`procedures for using the LAL test have evolved and have been adopted for
`devices.
`
`FDA announced
`In the FEDERAL REGISTER of January 18, 1980 (45 FR 3668),
`the availability of
`a draft guideline that set
`forth procedures for use
`of the LAL test as an end~product
`testing method for endotoxins in human
`and animal
`injectable drug products. This draft guideline was made
`_ l ,
`
`Page6
`
`Page 6
`
`

`

`available to interested parties to permit manufacturers, especially those
`who had used the LAL test
`in parallel with the rabbit pyrogen test,
`to
`submit data that could be considered in the preparation, of
`any final
`guideline.
`
`FDA
`received on the January 18 draft guideline,
`In response to comments
`and
`changes
`(i.e. Endotoxin
`limits
`changed
`made
`several
`significant
`deletion of section on. Absence of Non—endotoxin Pyrogenic Substances),
`and many minor editorial changes.
`The agency also determined that
`a
`single document
`should be made available 'covering all
`FDA regulated
`products that may be subject
`to LAL testing.
`Primarily because of
`the
`addition of biological products and medical devices to the guideline,
`the
`the agency made,
`in the FEDERAL REGISTER of March 29, 1983 (43 FR 13096),
`another draft of the guideline available for public comment.
`
`FDA has
`Based on the comments received on the March 29 draft guideline,
`made
`several
`changes
`in this final guideline.
`The
`comments used in
`support of
`these
`changes may be viewed at
`FDA's Dockets Management
`Branch, Room 4—62, 5600 Fishers Lane, Rockville, MD between 9 am and 4 pm
`Monday through Friday. Briefly,
`the significant changes made are:
`
`A.
`
`chromogenic,
`the
`for
`criteria
`validation
`of
`Inclusion
`endpoint—turbidimetric and kinetic-turbidimetric LAL techniques.
`
`can be
`turbidimetric)
`chromogenic or
`B. Any technique (gel—clot,
`for endotoxin. However, if a gel—clot
`used in testing a product
`lysate is used in.
`a different
`technique the results must be
`interpreted using the criteria for the technique being used.
`
`requirement
`the
`C. Elimination of
`rabbit pyrogen testing colony.
`
`to test
`
`the sensitivity of
`
`a
`
`D.
`
`has
`(CDRH)
`and Radiological Health
`for Devices
`The Center
`adopted the USP Endotoxin Reference Standard and revised the
`limit
`expressions
`from ng/mL to EU/mL.
`The
`new limit
`for
`medical devices is 0.5 EU/mL except for devices in contact with
`cerebrospinal
`fluid for which the limit
`is 0.06 EU/mL.
`These
`limits for devices are equivalent
`to those for drugs for a 70 Kg
`man when consideration is given to the following:
`
`1.
`
`in
`In the worst case situation, all endotoxin present
`the combined rinsings of 10 devices could have come from
`just
`one device.
`A wide variation in bioburden
`is
`common to some devices.
`
`2. Published FDA studies
`
`indicate that
`
`less than half of
`
`recovered
`is
`endotoxin
`added
`non—pyrogenic water rinse.
`
`from devices
`
`using
`
`a
`
`E.
`
`for Drug Evaluation and Research (CDER) has added a
`The Center
`the maximum doses
`per
`Kg
`per
`hour
`and
`the
`listing of
`endotoxin
`limits
`for most
`the
`aqueous
`corresponding
`of
`injectable drugs
`and biologics currently on.
`the market.
`This
`listing was added to promote uniformity among
`companies making
`the same product.
`
`Page7
`
`Page 7
`
`

`

`II.
`
`LEGAL EFFECT OF THE GUIDELINE
`
`(21 CFR lO.90(b)) of
`section lO.90(b)
`issued under
`This guideline is
`for use of guidelines
`FDA's administrative regulations, which provides
`to outline procedures or standards of general applicability that are
`acceptable to FDA for
`a subject matter within its statutory authority.
`Although guidelines are not
`legal
`requirements,
`a person who follows an
`agency guideline may be assured that
`the procedures or standards will be
`acceptable to FDA.
`The following guideline has been developed to inform
`manufacturers of human drugs
`(including biologicals), animal drugs,
`and
`medical devices of procedures FDA considers necessary to validate the use
`of LAL as an end—product endotoxin test.
`A manufacturer who adheres
`to
`the guideline would be considered in compliance with relevant provisions
`of
`the applicable FDA current good manufacturing practice regulations
`(CGMP)
`for drugs
`and devices
`and other applicable
`requirements.
`As
`provided in 21 CFR 10.90(b), persons who use methods and techniques not
`provided in the guideline should be able to adequately assure,
`through
`validation,
`that
`the method or
`technique they use is adequate to detect
`the endotoxin limit for the product.
`i
`
`Page 8
`
`Page 8
`
`

`

`III.
`
`REGULATORY PROVISIONS THAT PERMIT INITIATION
`OF END—PRODUCT TESTING WITH LAL
`
`firm must meet before using the LAL test
`The regulatory provisions that
`the same
`for all categories of products
`as an end-product
`test are not
`because
`applicable
`statutory
`provisions
`and
`of
`the
`different
`regulations. These provisions are as follows:
`
`a
`
`A.
`
`Human Drugs
`Abbreviated
`Applications,
`Applications
`AWE.-
`
`or
`gNDAs)
`New Drug Applications
`to
`subject
`New Drug, Applications
`gANDAs), Antibiotic Drug
`and
`animal
`drugs
`subject
`to New Animal Drug
`(NADAs),
`and Abbreviated
`New
`Animal
`Drpg
`
`a
`to submit
`are
`these classes of drugs, manufacturers
`For
`supplemental application to provide for LAL testing.
`However,
`under
`21 CFR 314.70(c)
`for drugs
`for
`human use
`and 21
`CFR
`514.8(d)(3)
`for drugs for animal use various changes may be made
`before FDA approval. Under these sections changes in testing of
`a human or animal drug that give increased assurance that
`the
`drug will have the characteristics of purity it purports or
`is
`represented to possess
`should be placed into effect at
`the
`earliest possible time. Therefore,
`if a firm validates the LAL
`test
`for
`a particular
`drug product
`covered by
`a
`new drug
`application by the procedures
`in this guideline using a
`LAL
`reagent
`licensed by
`the Center
`for Biologic Evaluation and
`Research (OBER)
`for the technique being used,
`the change may be
`made
`concurrently with
`the
`submission
`of
`the
`supplement
`providing for it.
`The supplement should contain initial quality
`control data,
`inhibition/enhancement data
`and
`the
`endotoxin
`limit for the drug product.
`
`Biological products for human use.
`
`in the manufacturing
`changes
`21 CFR 601.12 significant
`Under
`methods of biological products are required to be reported to
`the agency and may not become effective until approved by the
`Director,
`OBER.
`Therefore,
`a manufacturer
`of
`a biological
`product
`shall obtain an
`approved
`amendment
`to
`its
`product
`license before changing to the use of LAL
`in an end—product
`test,
`irrespective of the validation procedure used.
`
`C.
`
`Drugs not subject
`
`to premarket approval.
`
`A manufacturer of an injectable drug for human or animal use
`that
`is not subject
`to premarket approval would be able to use
`the LAL
`test
`as
`an
`end—product
`test
`for
`endotoxins without
`submitting any
`information to the agency.
`CGMPs
`require the
`manufacturer to have data on file to validate the use of
`the LAL
`test for each product for which it is being used.
`
`flaw.
`
`the basis of extensive experience in review of LAL data on
`On
`devices since November 1977, CDRH believes that
`the LAL test,
`_ 4
`_
`
`Page9
`
`Page 9
`
`

`

`least
`at
`is
`guideline,
`this
`to
`according
`validated
`when
`test
`for
`equivalent
`to the rabbit pyrogen test as an end—product
`medical
`devices.
`A manufacturer
`labeling
`a
`device
`as
`non—pyrogenic must validate the LAL test
`for
`that device in the
`test
`laboratory to be used for end—product
`testing before using
`the LAL test as an end—product endotoxin teSt for any device.
`
`this guideline may be
`The data discussed under Section V of
`expressed graphically or
`in tabular
`form and should be on file
`at
`the manufacturing site; no preclearance prior
`to use of
`the
`LAL
`test
`as
`an
`end—product
`test
`is
`required if
`it
`is used
`according to this FDA guideline. Voluntary submission of
`LAL
`validation and inhibition data obtained following issuance of
`this guideline will be accepted for CDRH review and comment.
`
`that
`test procedures
`LAL
`to use
`a manufacturer plans
`When
`deviate
`significantly from the
`LAL guideline,
`a premarket
`notification under section 510(k) of the Federal Food, Drug,
`and
`Cosmetic Act (the Act) or a Premarket Approval Application (PMA)
`supplement under section 515 of
`the Act
`should be
`submitted.
`Significant deviations would 'include—— but not necessarily’ be
`limited to—— higher endotoxin concentration release criteria,
`sampling from fewer
`than three lots for
`inhibition/enhancement
`testing,
`lesser sensitivity to endotoxin,
`rabbit retest when the
`LAL method
`shows
`endotoxin
`above
`the
`recommended
`allowable
`endotoxin dose,
`and
`a device
`rinsing protocol
`resulting in
`greater dilution of
`endotoxin than that
`recommended
`in this
`guideline.
`
`for another deviation from this
`PMA supplement
`notification or
`draft guideline; process control of endotoxin contamination with
`reduced end—product
`testing,
`i.e.,
`a decrease in the number of
`devices
`per
`lot
`undergoing
`end—product
`testing.
`The
`manufacturer must demonstrate adequate control of the production
`process by the use of routine checks for endotoxin at key stages
`of production except where it has been shown that no possibility
`of contamination exists.
`
`of
`providers
`review,
`PMA
`subsequent
`facilitate
`To
`to 21 CFR part 812 or 813 are
`investigational devices subject
`encouraged to use this guideline when a non—pyrogenic device is
`to be manufactured.
`
`PagelO
`
`Page 10
`
`

`

`IV. HUMAN AND ANIMAL DRUGS AND BIOLOGICAL PRODUCTS
`
`GENERAL REQUIREMENT
`
`licensed
`reagent
`LAL
`an
`use
`shall
`Manufacturers
`validation,
`in—process, and end—product LAL tests.
`
`by
`
`CBER
`
`in all
`
`A.
`
`VALIDATION OF THE LAL TEST
`
`for the release of
`the LAL test as an endotoxin test
`Validation of
`human
`and
`animal
`drugs
`includes
`the
`following:
`(1)
`initial
`qualification of
`the laboratory, and (2)
`inhibition and enhancement
`tests.
`
`INITIAL QUALIEICATION OF THE LABORATORY
`
`the detection of
`Various methodologies have been described for
`endotoxin,
`using
`limulus
`amebocyte
`lysate.
`Currently,
`commercially available
`licensed
`lysates
`use
`the
`gel
`clot,
`chromogenic,
`endpoint—turbidimetric
`or
`kinetic—turbidimetric
`techniques.
`Other methods which
`have
`been
`reported
`show
`potential
`increasing further
`the
`sensitivity of
`the LAL
`method.
`
`for
`
`testing
`the
`the variability of
`assess
`should
`Manufacturers
`Each analyst
`tests are performed.
`laboratory before any offical
`using a single lot of LAL and a single lot of endotoxin should
`perform the
`test
`for
`confirmation
`of
`labeled LAL
`reagent
`sensitivity or of performance criteria.
`Appendix A gives
`the
`procedures and test criteria for the current
`licensed techniques.
`
`INHIBITION AND ENHANCEMENT TESTING
`
`the LAL
`enhancement of
`inhibition or
`The degree of product
`procedure should be determined for each drug formulation before
`the LAL
`test
`is used to assess
`the endotoxin content of any
`drug. All validation tests should be performed on undiluted
`drug product or on an appropriate
`dilution. Dilutions should
`not exceed the Maximum Valid Dilution (MVD)
`(see Appendix D).
`At
`least
`three production batches of
`each finished product
`should be tested for inhibition and enhancement.
`
`8) GEL—CLOT TECHNIQUE
`
`testing should be conducted according
`Inhibition/enhancement
`to
`the directions
`in the preparatory section of
`the USP
`Bacterial Endotoxins Test
`(see Appendix B). Briefly,
`the
`a
`method
`involves
`taking
`drug
`concentration containing
`varying concentrations of a standard endotoxin that bracket
`the sensitivity of
`the lysate and comparing it
`to a series
`of
`the same endotoxin concentrations
`in water alone.
`The
`
`then diluted
`"spiked” with endotoxin and
`is
`drug product
`with additional drug product
`(so that
`the drug concentration
`remains constant)
`to
`the
`same
`endotoxin concentrations
`in
`6
`
`_
`
`_
`
`Pagell
`
`Page 11
`
`

`

`water. Results of endotoxin determination in water and the
`
`twofold
`a
`plus/minus
`fall within
`should
`product
`drug
`If the undiluted drug
`dilution of
`the labeled sensitivity.
`product
`shows
`inhibition,
`the drug product
`can be diluted,
`not
`to exceed the MVD, with the same diluent
`that will be
`used
`in
`the
`release
`testing
`and
`the
`above
`procedure
`repeated. Negative controls (diluent plus lysate) should be
`included in all inhibition/enhancement testing.
`
`b)
`
`CHROMOGENIC AND ENDPOINT—TURBIQIMETRIC TECHNIQUES
`
`a
`techniques,
`these
`testing by
`In inhibition/enhancement
`the
`drug concentration containing 4 lambda concentration of
`RSE
`or
`CSE
`(lambda
`is
`equal
`to
`the
`lowest
`endotoxin
`concentration used to generate the standard curve)
`is tested
`in
`duplicate
`according
`to
`the
`lysate manufacturer's
`methodology.
`The standard curve for
`these techniques shall
`consist of at
`least
`four RSE or CSE concentrations in water
`
`If the desired range is
`that extend over the desired range.
`standards concentrations
`greater
`than one
`log, additional
`should
`be
`included.
`The
`standard curve must meet
`the
`
`The
`linearity' as outlined in Appendix A(2).
`criteria for
`be
`detected amount of endotoxin in the spiked drug must
`within plus or minus
`25% of
`the 4 lambda concentration for
`the drug concentration to be considered to neither enhance
`nor
`inhibit
`the assay.
`If the undiluted drug product shows
`inhibition,
`the drug product can be diluted, not
`to exceed
`the MVD, and ti
`test repeated.
`
`An alternate procedure may be performed as described above
`except
`the RSE/CSE
`standard
`curve
`is
`prepared
`in LAL
`negative drug product,
`i.e. no detectable endotoxin,
`instead
`of LAL negative water.
`The standard curve must meet
`the
`test
`for
`linearity,
`i.e.
`r equal
`to or greater than 0.980,
`and in addition the difference between the 0.D.
`readings for
`the
`lowest
`and highest
`endotoxin concentrations must
`be
`greater
`than 0.4 and less than 1.5 0.D. units.
`If
`the
`standard curve
`does not meet
`these criteria,
`the
`drug
`product cannot be tested by the alternate procedure.
`
`c) KINETIC-TURBIDIMETRIC TECHNIQUE
`
`a drug
`testing by this technique,
`In inhibition/enhancement
`the RSE
`concentration containing 4
`lambda concentration of
`endotoxin
`or
`CSE
`(lambda
`is
`equal
`to
`the
`lowest
`concentration used to generate the standard curve) is tested
`in
`duplicate
`according
`to
`the
`lysate manufacturer's
`methodology.
`The standard curve shall consist of at
`least
`four RSE or CSE concentrations.
`If
`the desired range
`is
`greater
`than one
`log,
`additional
`standard concentrations
`should
`be
`included.
`The
`standard curve must meet
`the
`criteria
`outlined in Appendix A(3).
`The calculated Inean
`of
`amount
`endotoxin in
`the
`spiked
`drug
`product,
`when
`referenced to the standard curve, must be within plus or
`minus
`25% to be considered to neither enhance nor
`inhibit
`
`the
`
`assay.
`
`If
`
`the
`
`7,
`
`__
`
`undiluted
`
`drug
`
`product
`
`shows
`
`PagelZ
`
`Page 12
`
`

`

`the drug product can be diluted,
`inhibition or enhancement,
`not
`to exceed the MVD, and the test repeated.
`
`An alternate procedure may be performed whereby the RSE/CSE
`standard
`curve
`is
`prepared
`in drug
`product
`or
`product
`dilution instead of water.
`The drug product cannot have a
`background
`endotoxin
`concentration
`of more
`than
`10%
`(estimated by extrapolation of
`the regression line) of
`the
`lambda concentration (lambda equals the lowest concentration
`used to generate the standard curve).
`The standard curve
`must meet
`the test
`for
`linearity,
`i.e.
`r equal
`to or
`less
`than. —O.980,
`and in addition the slope of
`the regression
`must
`be
`less
`than —O.1
`and greater
`than —l.0.
`If
`the
`standard curve
`does not meet
`these criteria,
`the
`drug
`product cannot be tested by the alternate procedure.
`
`various
`in
`is manufactured
`drug
`the
`instances when
`those
`In
`concentrations of active ingredient while the other components of the
`formulation remain constant, only the highest and lowest concentration
`need be
`tested.
`If there is a significant difference,
`i.e. greater
`than
`twofold,
`between
`the
`inhibition endpoints
`or
`if
`the
`drug
`concentration, per mL,
`in the test solutions is different,
`then each
`remaining concentrations should be tested.
`If the drug product
`shows
`inhibition or enhancement at
`the MVD, when tested by the procedures in
`the above sections, and is amenable to rabbit
`testing,
`then the rabbit
`test will still
`be
`the appropriate test
`for
`that drug.
`If
`the
`inhibiting
`or
`enhancing
`substances
`can
`be
`neutralized without
`affecting the sensitivity of
`the test or
`if the LAL
`test
`is more
`sensitive than the rabbit pyrogen test
`the LAL test can be used.
`For
`those drugs not amenable to rabbit pyrogen testing,
`the manufacturer
`should determine
`the
`smallest quantity of
`endotoxin
`that
`can
`be
`detected. This data should be submitted to the appropriate FDA Office
`for review.
`
`the
`tests must be repeated on one unit of
`The inhibition/enhancement
`product
`if
`the
`lysate manufacturer
`is
`changed.
`If
`the
`lysate
`technique is changed,
`the inhibition and enhancement
`tests must be
`repeated using three batches. When
`the manufacturing process,
`the
`product formulation,
`the source of a particular ingredient of the drug
`formulation, or
`lysate lot
`is changed,
`the positive product control
`can be used to reverify the validity of the LAL test
`for the product.
`Firms
`that are obtaining an ingredient
`from a new manufacturer are
`encourged to include as part of their vendor qualification the rabbit
`pyrogen
`test
`to
`determine
`that
`the
`ingredient
`does
`not
`contain
`non—endotoxin pyrogens.
`
`Routine Testing qwarugs b
`
`the LAL Test.
`
`the
`from
`data
`on
`based
`be
`to
`is
`testing
`End—product
`Samples,
`testing as outlined in Section A(2).
`inhibition/enhancement
`standards, positive product controls and negative controls should be
`tested at least
`in duplicate.
`
`Pagel3
`
`Page 13
`
`

`

`an endotoxin standard series does not
`technique,
`the gel—clot
`For
`to be
`run with each set of
`tests if consistency of
`standard
`have
`endpoints has been demonstrated in the test
`laboratory.
`It should
`be run at least once a day with the first set of
`tests and repeated
`if
`there
`is
`any
`change
`in lysate lot,
`endotoxin lot or
`test
`conditions during the day.
`An endotoxin standard series should be
`run when
`confirming end—product
`contamination.
`Positive product
`controls (
`two lamda concentration of standard endotoxin in product)
`must be positive.
`If your
`test protocols state that you are using
`the USP Bacterial Endotoxin Test,
`remember
`that
`it
`requires
`a
`standard series to be run with each test.
`The above deviation must
`be noted in your test protocol.
`
`an
`techniques,
`endpoint—turbidimetric
`and
`chromogenic
`the
`For
`endotoxin standard series does not have to be run with each set of
`tests if consistency of standard curves has been demonstrated in the
`test
`laboratory.
`It
`should be
`run at
`least once a day with the
`first set of
`tests and repeated if there is any change in lysate
`lot,
`endotoxin lot or
`test conditions during the day. However, at
`least duplicates of a 4 lambda standard concentration in water and
`in each. product
`(positive product control) must be
`included with
`each run of
`samples.
`The mean
`endotoxin concentration of
`the
`standard must be within plus/minus 25% of
`the actual concentration
`and the positive product control must meet
`the same criteria after
`subtraction of
`any
`endogenous
`endotoxin.
`An
`endotoxin standard
`series should be run when confirming end—product contamination.
`If
`the alternate procedure is used,
`a standard in product series must
`be conducted each time the product is tested.
`
`is not necessary to run a
`it
`the kinetic—turbidimetric test,
`For
`standard curve each day or when confirming end product contamination
`if consistency of standard curves has been demonstrated in the test
`laboratory..
`However, at
`least duplicates of a 4
`lambda standard
`concentration in water
`and
`in
`each
`product
`(positive
`product
`control) must
`be
`included with each run of
`samples.
`The mean
`endotoxin concentration of
`the standard when calculated using an
`archived standard curve (See Appendix C), must be within plus/minus
`25% of
`the actual concentration and the positive product control
`must meet
`the
`same criteria after subtraction of
`any endogenous
`endotoxin.
`If
`the alternate procedure
`is used,
`a
`standard
`in
`product series must be conducted each time the product
`is tested.
`
`the labeled sensitivity of
`Before a new lot of lysate is used,
`lysate or
`the performance criteria should be
`confirmed by
`laboratory, using the procedures in Appendix A.
`
`the
`the
`
`The sampling technique selected and the number of units to be tested
`should be based on the manufacturing procedures and the batch size.
`A minimum of
`three units,
`representing the beginning, middle,
`and
`end,
`should
`be
`tested from a
`lot.
`These
`units
`can
`be
`run
`individually or pooled.
`If the units are pooled and any endotoxin
`is detected,
`repeat
`testing can be performed.
`The
`LAL
`test may be
`repeated no more than twice.
`The first repeat consists of twice the
`initial number of replicates of
`the sample in question to examine
`the possibility that extrinsic contamination occurred in the initial
`
`, g _
`
`Pagel4
`
`Page 14
`
`

`

`if any endotoxin is detected in
`On pooled samples,
`assay procedure.
`the
`first
`repeat,
`proceed to second
`repeat.
`The
`second repeat
`consists of an additional 10 units tested individually.
`None of
`the
`10 units tested in the second repeat may contain endotoxin in excess
`of the limit concentration for the drug product.
`
`following should be
`The
`parenteral drugs
`to meet
`end—product endotoxin test:
`
`for all
`endotoxin limit
`considered the
`if
`the
`LAL
`test
`is
`to be used as
`an
`
`1. K/M:
`
`administered
`those
`except
`drug
`any parenteral
`For
`endotoxin is
`for
`endotoxin limit
`intrathecally,
`the
`defined as K/M, which equals
`the amount of
`endotoxin
`(EU) allowed per ng or mL of product.
`K is equal
`to 5.0
`EU/Kg.
`(SEE appendix D for definition of M).
`
`an
`have
`that
`drugs
`For parenteral
`administration, K is equal
`to 0.2 EU/Kg.
`
`intrathecal
`
`route
`
`of
`
`Drugs exempted from the above endotoxin limits are:
`
`l. Compendial drugs
`established.
`
`for which other
`
`endotoxin limits have been
`
`applications,
`drug
`new
`by
`covered
`drugs
`2. Non—compendial
`antibiotic drug applications, new animal drug applications, and
`biological product
`licenses where different
`limits have been
`approved by the agency.
`
`3.
`
`Investigational drugs or biologicals for which an IND or
`exemption has been filed and approved.
`
`INAD
`
`4. Drugs or biologicals which cannot be tested by the LAL method.
`
`A batch which fails a validated LAL release test should not be retested
`
`the high
`to
`Due
`it passes.
`released if
`and
`test
`rabbit
`the
`by
`test
`as
`an
`rabbit
`reproducibility of
`the
`lack of
`variability and
`endotoxin assay procedure, we do not consider
`it an appropriate retest
`procedure for LAL failures.
`
`PagelS
`
`Page 15
`
`

`

`V. MEDICAL DEVICES
`
`General Reguirements
`
`the "HIMA Collaborative Study for
`reviewed the results of
`The CDRH has
`the Pyrogenicity Evaluation of
`a Reference Endotoxin by the USP Rabbit
`Test."
`This study' recommends 0.1 ng/mL (10 mL/kg) of E; coli 055:B5
`endotoxin from Difco Laboratories as the level of endotoxin which should
`
`testing of
`end—product
`test when used for
`be detectable in the LAL
`sensitivity (0.1
`ng/mL given 10 mL/kg)
`is
`medical
`devices.
`This
`LAL
`sufficient
`for
`testing and
`for
`retest
`of devices
`in rabbits.
`According to recent collaborative studies in the rabbit pyrogen and LAL
`tests, one nanogram of E. coli 055:85 endotoxin is similar in potency to
`5 EU of
`the USP Endotoxin Reference Standard.
`The endotoxin limit
`for
`medical devices has been converted to EU and is now 0.5 EU/mL using the
`rinse volume recommended in Section 2 below.
`Liquid devices should be
`more appropriately validated and tested according to the requirements for
`drugs by taking the maximum human dose per kilogram of body weight per
`hour into consideration (See Section IV,B).
`
`a USP
`failures with the LAL test or
`Manufacturers may retest LAL test
`rabbit pyrogen test.
`If the endotoxin level
`in a device eluate has been
`quantitated by LAL at 0.5 EU/mL endotoxin or greater,
`then retest
`in
`rabbits is not appropriate. Medical devices that contact cerebrospinal
`fluid should have
`less
`than 0.06 EU/mL of
`endotoxin.
`These values
`
`correspond to those set by the CDER for intrathecal drugs.
`
`licensed by
`reagent
`LAL
`an
`use
`shall
`Manufacturers
`validation,
`in—process, and end—product LAL tests.
`
`OBRR
`
`in all
`
`A. Validation of the LAL Test
`
`demonstrating
`Data
`1. Sensitivity:
`reproducibility of the LAL test.
`
`the
`
`sensitivity
`
`and
`
`2.
`
`line of devices
`Each product
`Inhibition/Enhancement Testing:
`utilizing different materials or methods of manufacture should
`be checked for inhibition or enhancement of the LAL test.
`
`Further explanation of the above points is given as follows:
`
`1.
`
`SENSITIVITY
`
`A manufacturer must be able to demonstrate a sensitivity of at
`least
`0.5
`EU/mL.
`The
`level
`of
`endotoxin selected as
`the
`
`for evaluating pyrogenicity of products using
`pass/fail point
`the
`LAL
`test must
`be
`equivalent
`to
`or
`below this
`level.
`Manufacturers may
`use
`another
`endotoxin
`if
`a
`reproducible
`correlation between it and the USP Reference Endotoxin Standard
`
`has been demonstrated in their laboratory (see appendix C).
`
`Pagel6
`
`Page 16
`
`

`

`the LAL technique used should be determined
`The sensitivity of
`and
`criteria
`in Appendix A.
`Routine
`by
`the
`procedures
`performance of
`the LAL
`test
`should include standards
`(run in
`duplicate) and a negative control.
`An endotoxin standard series
`is useful
`for checking lysate sensitivity and the competence of
`the technician,
`and for
`identifying other problems such as the
`contamination of glassware.
`
`The stability of the endotoxin standards and appropriate storage
`conditions should also be considered; dilute endotoxin solutions
`are not as stable as more concentrated solutions under certain
`conditions.
`
`INHIBITION AND ENHANCEMENT TESTING
`
`inhibition or enhancement of the LAL test should
`Lack of product
`be shown for each type of device before use of
`the LAL
`test.
`Possible inhibition of different chemical components of similar
`devices
`should be
`considered.
`A manufacturer may ilogically
`divide its device products into groups of products according to
`common
`chemical
`formulation;
`and may
`then qualify only
`a
`representative product
`from each such group.
`Ideally,
`the
`product chosen from each group would be the one with the largest
`surface area contacting body or
`fluid for administration to a
`patient.
`
`type should be
`three production lots of each product
`least
`At
`the
`sampling
`for
`inhibition.
`In general,
`use
`of
`tested
`technique selected should result
`in a
`random sampling of
`a
`finished production lot.
`CDRH recommends
`testing 2 devices for
`lot sizes under 30,
`3 devices
`for
`lot
`sizes
`30—100,
`and
`3
`percent of
`lots above size 100, up to a maximum of 10 devices
`per lot.
`
`or
`pyrogen
`for
`of preparing an eluate/extract
`The process
`Some
`inhibition/enhancement
`testing may vary for each device.
`medical devices can be flushed,
`some may have to be immersed in
`the non—pyrogenic rinse solution, while others may be tested by
`disassembling or by cutting the device into pieces prior
`to
`extraction by immersion.
`In general,
`for device

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket