throbber
Case 1:19-cv-11586-FDS Document 343-12 Filed 03/02/22 Page 1 of 8
`Case 1:19-cv-11586-FDS Document 343-12 Filed 03/02/22 Page 1 of8
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`EXHIBIT AL
`EXHIBIT AL
`
`

`

`Case 1:19-cv-11586-FDS Document 343-12 Filed 03/02/22 Page 2 of 8
`Case 1:19-cv-11586-FDS Document 343-12 Filed 03/02/22 Page 2 of8
`
`OriginalArticle
`
`Checkfor
`|
`updates,
`Page 1 of 7
`
`Assessing the ability of the Fitbit Charge 2 to accurately predict
`VOsmax
`
`Kaitlin A. Freeberg, Brett R. Baughman, Ted Vickey, Jeff A. Sullivan, Brandon J. Sawyer
`
`Departments of Kinesiology and Biology, Point Loma Nazarene University, San Diego, CA, USA
`Contributions: (I) Conception and design: All authors; (II) Administrative support: KA Freeberg, T Vickey, BJ Sawyer; (II) Provision of study
`materials or patients: KA Freeberg, BR Baughman; (IV) Collection and assembly of data: KA Freeberg, BR Baughman; (V) Data analysis and
`interpretation: KA Freeberg, BJ Sawyer; (VI) Manuscript writing: All authors; (VID Final approval of manuscript: All authors.
`Correspondence to: Brandon J. Sawyer, PhD. Point Loma Nazarene University, Department of Kinesiology and Biology, 3900 Lomaland Drive, San
`Diego, CA 92106, USA. Email: bsawyer@pointloma.edu.
`
`Background: Theaim of this study was to assess the ability of the Fitbit Charge 2 (FBC2) to accurately
`estimate VO,,,.. in comparison to both the gold standard VO;,,,,. test and a non-exercise VO),,.. prediction
`equation.
`Methods: Thirty healthy subjects (17 men, 13 women) between the ages of 18 and 35 (age =21.7+3.1 years)
`were given a FBC2to wearfor seven days and followed instructions on how to obtain a cardio fitness score (CFS).
`VOrmax WaS Measured with an incremental test on the treadmill followed by a verification phase. VO,,,.. was
`predicted via a non-exercise prediction model (N-Ex) using self-reported physicalactivity level.
`Results: Measured VO,,,.. was significantly lower than FBC2 predicted CFS (VO,,... =49.91+6.83; CFS
`=52.53+8.43, P=0.03). N-Ex prediction was significantly lower than CFS but notsignificantly lower than
`measured VO3... (N-Ex =48.79+6.32; CFS us. N-Ex: P=0.01; VO. vs. N-Ex: P=0.54). Relationships
`between both VO,,,.. vs. CFS and VOon2, vs. N-Ex were good ICC: VO,... vs. CFS=0.87, VOon. vs. N-Ex
`=0.87); Bland-Altman analysis indicated consistency of CFS measurement and lack of bias. The coefficient
`of variation (CV) and mean absolute percent error (MAPE) were greater with CFS than N-Ex (CV: CFS
`=6.5%+4.1%, N-Ex =5.6%+3.6%; MAPE: CFS =10.2%+6.7%, N-Ex =7.8%+5.0%). Heart rate (HR)
`estimated by the FBC2 was lower than estimated (Est) HR for pace based on HRextrapolation (FBC2
`=155+18 bpm, Est =183+15 bpm, P<0.001). The difference in CFS and VO,,,,, was inversely correlated with
`the difference in FBC2 HR and Estimated HR (r =-0.45, P<0.001).
`Conclusions: The FBC2 shows consistent, unbiased measurement of CFS while overestimating VO>,...
`in healthy men and women. The non-exercise VO,,,.. prediction equation provides a similar, slightly more
`accurate, VO>,,.. prediction than the CFS without the need for an exercise test or purchase ofa Fitbit.
`
`Keywords: Cardiorespiratory; VO2max; exercise; heart; rate; Fitbit; cardio; fitness; score
`
`Received: 22 May 2019; Accepted: 23 August 2019; Published: 23 September 2019.
`doi: 10.2 1037/mhealth.2019.09.07
`
`View thisarticle at: http://dx.doi.org/10.21037/mhealth.2019.09.07
`
`Introduction
`
`VOromex testing is known as the gold standard for measuring
`cardiorespiratory fitness and is frequently used in research
`settings to determine the efficacy of training program
`interventions (1). Exercise physiology laboratories regularly
`use VOome testing to evaluate the cardiorespiratory health
`of individuals as well as develop exercise prescriptions (1).
`
`Furthermore, VO,,,. is a strong predictor of cardiovascular
`disease (CVD)risk and overall CVD mortality (2). Maximal
`exercise testing has become the standard for measuring
`functional capacity, evaluating therapy, estimating risk, and
`organizing transplantation candidacy in patients with heart
`failure (3). Maximal exercise testing is also important in
`diagnosing and assessing coronary artery disease, peripheral
`
`© mHealth.All rights reserved.
`
`mealth 2019;5:39 | http://dx.doi.org/10.21037/mhealth.2019.09.07
`
`PNA-FB0016682
`
`

`

`Case 1:19-cv-11586-FDS Document 343-12 Filed 03/02/22 Page 3 of 8
`Case 1:19-cv-11586-FDS Document 343-12 Filed 03/02/22 Page 3of8
`
`Page 2 of 7
`
`mHealth, 2019
`
`Table 1 Descriptive statistics ofall participants (n=30)
`Subject characteristic Outcome, mean + SD
`
`
`
`Age (yr)
`
`21.743.1
`
`23.542.6
`BMI (kg/m’)
`Body fat (%) 20.5+7.1
`
`
`
`accurately estimate VO;,,., in comparison to both the
`gold standard VO,,,,, test and a non-exercise VOomax
`prediction equation. We hypothesized that the FBC2 would
`overestimate VO,,,, due to its reduced HR monitoring
`accuracy at increased exercise intensities (6-9).
`
`arterial disease, heart failure, valvular heart disease, and
`unexplained exertional dyspnea (3). Use of exercise testing
`by physicians and non-physicians has grown extensively,
`resulting in the administration of millions of tests (4).
`Despite the accuracy and proliferation of maximaltesting,
`there are difficulties involved that make this type of testing
`2max
`less accessible to the general population. VO,,,,,
`testing
`requires maximal effort and thus puts tremendousstrain on
`the body. Furthermore, maximal testing requires access to
`a lab and specific equipment necessary for assessing oxygen
`uptake, single tests of which can be expensive for the general
`population. Fitbit has released the Charge 2 watch [Fitbit
`Charge 2 (FBC2)], which is advertised to predict VOzn.. by
`displaying a userfriendly “cardio fitness score” (CFS). Using
`the relationship between running pace and heart rate (HR),
`the watch calculates a score comparable to one’s VOzmax in
`mL/kg/min. To our knowledge, Fitbit has not released
`research on how the FBC2 specifically predicts VOxm, thus
`the level of prediction accuracy is unclear. The accurate
`prediction of VO
`amex DY a wrist worn device is appealing due
`to the lower cost, less strenuous testing methodology, and
`potential for more widespread awareness of cardiovascular
`health.
`Other companies, such as Garmin, have created wearable
`personal fitness devices to estimate VOx,01. One study
`soughtto validate the use of the Garmin Forerunner 920XT
`watch in VO,,... estimation (5). Sixteen subjects were
`instructed to jog or run for ten minutes around a football
`field wearing HR monitors and the GPS Garmin watch and
`perform a treadmill VO,,,., test 2-5 days later (5). Results
`showednosignificant differences between the mean VOspe
`from the Garmin watch and the treadmill test as well as a
`high Pearson correlation coefficient (r=0.84), suggesting the
`Garmin Forerunner 920XT provides a relatively accurate
`prediction of VO,,.., (5). However, to our knowledge, no
`studies have been performed to evaluate the accuracy of the
`estimation.
`FBC2 in VOsing
`This study aimed to assess the ability of the FBC2 to
`
`Methods
`
`Experimental design
`
`Thirty subjects (17 men, 13 women) were given the FBC2
`to wear for seven days and followed instructions on how
`to obtain a CFS. Subjects came into the laboratory on two
`separate occasions. VOsma, Was predicted on their first visit
`via a non-exercise prediction model (N-Ex) using selt-
`reported physical activity level (10) and subjects performed
`submaximal exercise to become familiar with the maximal
`exercise equipment. VO}... was measured at their second
`visit via an incremental test on the treadmill followed by a
`verification phase. Body composition was also assessed to
`determine accurate subject characteristics. Participants were
`advised to perform their individual runs at least 48 hours
`apart and abstain from physical activity 48 hours prior to
`their measured VO,,,,, test.
`
`Participants
`
`On the basis of previously published data (11), we calculated
`that completing 27 subjects in our study would yield 95%
`powerto detect a 2% difference in VO... between CFS
`and measured VOomex (at a two-tailed alpha level of 0.05).
`Planning for subject attrition, we enrolled 34 subjects.
`‘Two subjects dropped out due to time constraints and two
`subjects were excluded from data analysis due to failure to
`adhere to instructions on how to obtain a CFS,resulting
`in a final sample size of 30. Physical characteristics of the
`participants who completed the study are shown in Table 1.
`Inclusion criteria were healthy, non-sedentary individuals
`aged 18-35 years old. Non-sedentary individuals were those
`who answered above a zero on the self-reported physical
`activity questionnaire (12). The study was approved by the
`university institutional review board; all subjects provided
`written informed consent and completed a Physical Activity
`Readiness Questionnaire (PAR-Q)before initiating the study
`to determine if the subject was healthy enough to exercise.
`Answering “yes” to any questions on the PAR-Q would
`immediately disqualify anyone from participation in the study.
`
`© mHealth.All rights reserved.
`
`mealth 2019;5:39 | http://dx.doi.org/10.21037/mhealth.2019.09.07
`
`PNA-FB0016683
`
`

`

`Case 1:19-cv-11586-FDS Document 343-12 Filed 03/02/22 Page 4 of 8
`Case 1:19-cv-11586-FDS Document 343-12 Filed 03/02/22 Page 4 of 8
`
`miieaith, 2019
`
`Assesswentofbody composition
`
`Body composition measurement was performed using air
`displacement plethysmography (Bod Pod Cosmed, Rome,
`Ttaly) (13). Subjects were fasted and refrained from exercise
`12 hours priorto testing. Wearing minimal clothing(spandex
`shorts or swimsuit) and a swim cap, subjects were weighed
`on a calibrated digital scale and height was recorded from
`a wall-mounted stadiometer (Seca, Birmingham, UK). The
`subject was then instructed to sit quietly within the BOD
`PODchamber for two measurements of body volume, each
`lasting about 45 seconds. Uf these two measurements apreed
`within 150 mL, they were averaged. If the two measurements
`did not agree within 150 mL,a third measurement was taken
`and the two values that were the closest and met criteria for
`apreement were averaged. Usingthe data collected for body
`mass and body volume as well as the predicted thoracic lung
`volume, body density and percent body fat were calculated
`using the Siri equation (14).
`
`Assessment ofsubmaximal HR and equipment
`familiarization
`
`Height and weight measurements were inserted into the
` sing self-reported physical activity level to predict
`VO ome (10). Subjects were equipped with an oronasal
`mask connected to a standard nonrebreathing valve (Hans
`Rudolph, Shawnee, KS, USA) for continous measurement
`of ventilation and respiratory gas exchange data usinga
`previously validated (15) metabolic measurement system
`(Parvo Medics TrueOne 2400; Parvo medics, Sandy, UT,
`USA). A standard 3-point calibration was performed
`before each test or every four hours per manufacturer
`recommendations. While measuring gas exchange and
`HRdata, subjects performed a submaximal treadmill run
`at 60, 70, 80, and 90%of their estimated VO,,,,, (10) to
`become familiar with the equipment. Subjects ran for
`three minutes at each intensity. Usingsteady state HR from
`each running pace, linear regression equations were created
`for each subject using running pace to estimate HR. These
`equations were subsequently used to estimate HR from the
`GPS measured running pace during the independent runs
`while wearing the FBC2. The estimated (Est) HR was then
`compared to the FBC2 measured HR.
`
`Assessment of CFS
`
`Subjects were assigned a FBC2 to wear for seven days. The
`
`Page 3 of 7
`
`FBC2s were updated with the latest firmware at the time
`of the study which was version 22.55.2. During the seven
`days, subjects were asked to complete two independent
`rans on flat terrain with the FBC2. Acceptable locations
`for running were recommended and GPS tracking from
`the watch confirmed participants ran on flat terrain. Hach
`of these runs consisted of a 5-minute warmup at a self-
`selected speed. With GPS and Bluetooth on and paired
`with their Firbir account on their smart phone, subjects
`then performed a 10-minute run. Based on the instructions
`from the manufacturer on how to obtain a CFS, subjects
`were instructed to run at as high of an intensity as could be
`continuously sustained for the full 10 minutes. Subjects then
`synced watch datato their phone application and a CFS was
`calenlated. Screenshots of the CFS, average pace, time, and
`average HR were sent to the primaryinvestigatorafter each
`ofthe tworuns.
`
`Assessinent of VOnas
`
`Subjects were set up with the same metabolic cart and
`procedures as during the familiarization visit. The
`incremental test protocol was chosen using an estimated
`VOand estimated speed and grade that were designed
`to elicit exhaustion in approximately 10 minutes (12). After
`collecting 2 minutes of resting data, subjects warmed up
`for five minutes at a speed of 3.5-4.0 mph and 0% grade
`on the treadmill (Trackmaster, Carrollton, TX, USA).
`After the warm-up phase, the speed increased to a constant
`based on the individualized protocol (4-7 mph) and
`treadmill grade increased continuously by 1% every minute
`until volitional exhaustion. After exhaustion was reached,
`the treadmill speed and grade were immediately reduced to
`2.5 mph and 0% grade for a 10-minute recovery period.
`The verification phase was then performed at 110%
`of peak work rate reached during the initial bout (16).
`VOono, Was confirmed if the verification phase attained
`a VO,max Value within 3% of the incremental test (17). Tf
`the verification phase yielded a VO... which was more
`than 3% below the VO2max value from the incremental
`test, subjects were required to come back and repeat their
`verification phase at the same intensity. If the verification
`phase was more than 3% above the incremental test
`VOomas WOrny Value from subjects were required to do
`test with both incremental and verification
`another VOona.
`
`phases until 3% criterion was achieved. VO,
`max TPOTM
`each test was determined by taking the average of the
`two highest consecutive 15 sec VO, values. Verbal
`
`© mHealth. All rights reserved.
`
`wiblealth 2019;5:39 | btep.//ds.doi.org/10.21037/mhealth.2019.09.07
`
`PNA-FB0016684
`
`

`

`Case 1:19-cv-11586-FDS Document 343-12 Filed 03/02/22 Page 5 of 8
`Case 1:19-cv-11586-FDS Document 343-12 Filed 03/02/22 Page 5of8
`
`Page 4 of 7
`
`mHealth, 2019
`
`Table 2 Average VO.,,...
`
`2max
`
`values from each testing method (n=30)
`
`Measured VO.,,,, Cardio fitness Non-exercise
`
` Subject pool (mL/kg/min) score prediction
`
`
`Combined
`49.91+6.83
`52,.53°'48.43
` 48.79+6,32
`(mean + SD)
`
`Women(n=13)
`45.48+5.90
`46.384+5.97
`44.5643.87
`Men (n=17) 52.02+5.98 53.31+5.50 57.24+6.92
`
`*
`
`
`
`
`
`, significantly higher than measured VOsmax, P=0.03; *,
`significantly higher than Non-Ex, P<0.01.
`
`for bias and consistency in VOon.. estimation by CFS, N-Ex,
`and measured VO,,,,,. Pearson correlations were used to
`examine the relationship between the difference in VOsna.
`measures (measured VO,,,,, - CFS) and the difference in
`HR measured by the FBC2 and estimated by the linear
`regression equations.
`
`Results
`
`VOding, differences
`

`
`aE
`
`3 1
`
`0
`
`‘
`-1.96 SD
`2
`a $158s
`0
`10
`20
`30
`40
`50
`60
`70
`Mean, VOsma, & CFS (mL/kg/min)
`
`‘There was a significant main effect for a difference in
`+1.96 SD
`Ig
`VO,across the three tests (P<0.01). Measured VO,,,,, Was
`iain vereavastecersracrecrcretcea re eserarecrearsra aratenn
`7.61
`>
`.

`significantly lower than CFS (VO,,,,, =4+9.91+6.83 mL/kg/
`5
`x


`e
`min; CFS =52.53+8.43 mL/kg/min, P=0.03) (Iable 2). The
`oa
`wo
`2
`*
`0
`uw
`N-Ex prediction was significantly lower than the CFS
`* Mean
`2 °
`9
`a aa
`but not significantly lower than measured VOjme. (N-Ex
`E
`24

`*
`SG -5
`gas *
`2.92
`%
`=48.7926.32 mL/kg/min; CFS vs. N-Ex: P<0.01; VOomas US.
`>
`+
`oe
`a
`*
`N-Ex: P=0.54. CVs were similar with CFS and N-Ex when
`compared to the gold standard measured VO}... value (CFS
`=6.5%#4.1%; N-Ex =5.6%+3.6%). MAPE waslarger for CFS
`than N-Ex when compared to VOomas (CFS =10.2%+6.7%;
`N-Ex =7.8%+5.0%). Bland-Altman analysis indicated
`consistent, unbiased measurement of CFS (Figure 1). ICCs
`between both VO),,,,, vs. CFS and VO3,,., vs. N-Ex were
`good (VOnmex US. CFS =0.87, VOsmax US. N-Ex =0.87).
`
`o?¢
`
`oa
`
`80
`
`Figure 1 Bland-Altman plot of mean and difference between
`measured VO,,,.. and FBC2 CFS. The solid line represents the
`mean difference of -2.92 mL/kg/min and the dashed lines are the
`95% limits of agreement. FBC2, Fitbit Charge 2; CFS, cardio
`fitness score.
`
`encouragement was given throughout all laboratory VOjn4.
`tests.
`
`Data analysis
`
`All data were analyzed using SPSS Software (SPSS 21.0;
`IBM Corp., Armonk, NY, USA). All data in text, tables, and
`figures are presented as means and standard deviations (SD)
`and significance was set at P<0.05. We tested the outcome
`variables for normality with the Shapiro-Wilk test to assure
`all variables met the assumptionsofthestatistical tests used.
`A repeated measuresanalysis of variance (RMANOVA)with
`a Bonferonni post-hoc test was used to test for ditterences
`measurement
`between the three methods of VO,,,,,
`(VOsinay CFS, and N-Ex). The assumption of sphericity was
`tested before interpreting the results of the RMANOVA.
`Coefficients of variation (CVs) and mean absolute percent
`error (MAPE) were calculated to determine prediction
`accuracy of the CFS and N-Ex. Bland-Altman plots and
`intraclass correlation coefficients (ICCs) were used to test
`
`AR differences
`
`HRestimated by the FBC2 was lower than Est based on
`HRextrapolation (FBC2 =155+18 bpm, Est =183+15 bpm,
`P<0.001) (Figure 2). The difference in CFS and VOonax
`(measured VO. - CFS) was inversely correlated with the
`difference in FBC2 HR and Est HR (Est HR - FBC2 HR)
`(r =-0.45, P<0.01) (Figure 3).
`
`Discussion
`
`Our study found that the FBC2 produces a consistent,
`2max
`unbiased estimate of VO,,,..
`(CFS) while significantly
`overestimating VO}... when compared to the gold-standard
`value obtained from the incremental test with verification.
`Interestingly, the value predicted by the N-Ex model is
`not significantly different from the measured VO... and
`therefore slightly more accurate than the FBC2 CFS
`in predicting VOzn.. This suggests that an individual
`who does not want to perform a maximal exercise test or
`purchase a FBC2 maystill benefit from completing a non-
`
`© mHealth.All rights reserved.
`
`mealth 2019;5:39 | http://dx.doi.org/10.21037/mhealth.2019.09.07
`
`PNA-FB0016685
`
`€
`

`

`Case 1:19-cv-11586-FDS Document 343-12 Filed 03/02/22 Page 6 of 8
`Case 1:19-cv-11586-FDS Document 343-12 Filed 03/02/22 Page 6 of8
`
`mHealth, 2019
`
`Page 5 of 7
`
`220
`pm) 200
`
`===&aaQooOoO
`
`MeasuredHeartRate(b 2=
`Fitbit
`
`=i QoO
`100
`
`120
`
`180
`160
`140
`Extrapolated Heart Rate (bpm)
`
`200
`
`220
`
`Figure 2 HR values from FBC2 and extrapolation. Thelineis the
`line of identity. FBC2, Fitbit Charge2.
`
`
`
`VOsmaxdifference:measuredVO,,..,-CFS
`
`
`
`
`
`= °
`
` R’=0.1815
`
`-20
`
`80
`60
`40
`20
`0
`Heart Rate Difference: Extrapolated-Fitbit Measured
`
`100
`
`fitness watch against a laboratory test of aerobic capacity (19).
`Eighteen college-age students completed a VOzn.. test on the
`treadmill and performed a Polar fitness test (19). The Polar
`fitness test required that the subject report their physical
`activity level from the last three months based on descriptions
`provided by Polar, then lie supine for five minutes while the
`Polar HRstrap recorded data (19). At the end of thetest, the
`watch would display a VOzn.. value based on the subject’s age,
`height, weight, sex, activity level, maximum HR,and seated
`HR (19). The paired samples T-test showed nosignificant
`differences between the Polar VO,,,,, value and the metabolic
`cart value (Polar: 47.67 mL/kg/min vs. Metabolic Cart:
`44.09 mL/kg/min, P=0.111) (19). Both the Garmin and
`Polar wrist worn fitness devices were not significantly
`different from metabolic cart values, suggesting they may be
`appropriate means of measuring aerobic capacity for those
`not requiring the accuracy of laboratory equipment.
`Difficulty with the Fitbit measuring an accurate HR
`during runs may play a crucial role in the accuracy of
`the CFS (8,9). Wallen ez a/. found that among the Apple
`Watch, Fitbit Charge HR, Samsung Gear S and Mio
`Alpha, all devices underestimated HR in comparison
`to electrocardiography (8). However, it is important to
`note that these underestimations are not always clinically
`significant and may only reach significance under certain
`situations. For example, studies show that as exercise
`intensity increases, there is greater underestimation of HR
`(6,9). Our study discovered an inverse relationship between
`the difference in CFS and VO,,,,, and the difference in
`FBC2 HRandindividual subject extrapolated HR. In other
`words, the more the FBC2 underestimated HR, the more
`it overestimated VO)...
`. Thus, if the FBC2 underestimates
`HR during a run then it will most likely overestimate
`VOomaxy assuming the lower measured HRfora given paceis
`evidence of higherfitness level.
`Onestrength of this study was that VO),,,, testing was
`performed with a verification phase, the current gold
`standard methodology for verifying if subjects reach a “true”
`VOrmax (20). All subjects in this study verified their maximal
`values within 3% and were required to repeattheir tests if
`values were not confirmed. Also, subjects performed two
`individual runs and an average CFS was used forstatistical
`analyses to assess intraclass reliability and assure the
`subjects first run did not skew results. Subjects wore the
`same FBC2 for all seven days of the study and the watch
`was worn for at least 2 nights before subjects performed
`their runs in order to allow the FBC2 to get accustomed to
`the individual’s resting HR. Although instruction was given,
`
`Figure 3 Differences between extrapolated HR and FBC2
`Measured HRin relation to differences between measured VOpimax
`and CFS. FBC2, Fitbit Charge 2; CFS, cardio fitness score.
`
`exercise self-reported physical activity questionnaire, which
`predicts VOjn. with good accuracy.
`A similar study was performed on the Garmin Forerunner
`920XT and found that the Garmin watch was highly
`correlated to aerobic capacity measurements obtained via
`open-circuit spirometry (Garmin: r=0.84) (5). Unlike our
`study, however, the Garmin watch was notsignificantly
`different from the measured aerobic capacity (5). This
`difference in significance between studies could be attributed
`to the different software and prediction equations within
`the watches, as Garmin uses a company called FirstBeat
`‘Technologies and Fitbit does not (18). Furthermore, the use
`of a HR monitor strap during the Garmin watch run may
`have provided more accurate HR data than was obtained
`from the Fitbit wrist worn HRsensor.
`
`A recent study tested the accuracy of the Polar RS300X
`
`© mHealth.All rights reserved.
`
`mealth 2019;5:39 | http://dx.doi.org/10.21037/mhealth.2019.09.07
`
`PNA-FB0016686
`
`

`

`Case 1:19-cv-11586-FDS Document 343-12 Filed 03/02/22 Page 7 of 8
`Case 1:19-cv-11586-FDS Document 343-12 Filed 03/02/22 Page 7 of8
`
`Page 6 of 7
`
`mHealth, 2019
`
`subjects were not supervised and no verbal encouragement
`was given during their individual runs. As such, some
`subjects had difficulties obtaining a CFS and may have
`performed better if given encouragement similar to that
`given during the VO,,,,, test. However, these user errors
`are a better depiction of the general population, as the
`average individual would likely not have a personal trainer
`encouraging them and confirming proper use of the FBC2.
`‘To improve our study, measurement of HR during the
`10-minute runs with a chest strap HR monitor would have
`been moreaccurate than extrapolating the data. Despite
`all subjects running on flat terrain, it would have been
`more controlled if individual runs were all recorded at a
`single location. Furthermore, darker skin tones and larger
`wrist circumferences have been associated with decreased
`accuracy of wearable devices (21,22), however, these
`data were not collected. The current study looked at the
`accuracy of VOzmestimation by the FBC2 in a group of
`healthy young men and women; subject race and ethnicity
`were not reported. Therefore, future studies should
`determine the accuracy of the FBC2 for predicting VOone
`in adults varying in age, race, and ethnicity to enhance the
`generalizability of ourresults.
`The results of our study suggest the FBC2 provides a
`consistent, unbiased prediction while overestimating VOomex
`in young, healthy men and women.A non-exercise prediction
`equation provides a similar, slightly more accurate, VOomex
`prediction than the CFS without the need to perform an
`exercise test or purchase a wearable device. The accuracy
`of the FBC2 CFS may belimited by its ability to correctly
`detect exercise HR at increased submaximalintensities.
`
`Acknowledgments
`
`None.
`
`Footnote
`
`Conflicts ofInterest: The authors have no conflicts of interest
`to declare.
`
`Ethical Statement: The authors are accountable for all
`aspects of the work in ensuring that questions related
`to the accuracy or integrity of any part of the work are
`appropriately investigated and resolved. The study was
`approved by the university institutional review board,all
`subjects provided written informed consent.
`
`References
`
`1.
`
`10.
`
`11.
`
`12.
`
`13.
`
`14.
`
`Bassett DR Jr, Howley ET. Limitingfactors for
`maximum oxygen uptake and determinants of endurance
`performance. Med Sci Sports Exerc 2000;32:70-84.
`Katzmarzyk PT, Church TS, Blair SN. Cardiorespiratory
`fitness attenuates the effects of the metabolic syndrome on
`all-cause and cardiovascular disease mortality in men. Arch
`Intern Med 2004;164:1092-7.
`Arena R, Myers J, Williams MA,et al. Assessmentof
`functional capacity in clinical and research settings: a
`scientific statement from the American Heart Association
`
`Committee on Exercise, Rehabilitation, and Prevention
`of the Council on Clinical Cardiology and the Council on
`Cardiovascular Nursing. Circulation 2007;116:329-43.
`GibbonsL, Blair SN, Kohl HW, et al. Thesafety of
`maximal exercise testing. Circulation 1989;80:846-52.
`Kraft GL, Roberts RA. Validation of the Garmin Forerunner
`920 XT Fitness Watch VO 2 peak Test. IntJ Innov Educ Res
`2017;5:61-7.
`Boudreaux BD, Hebert EP, Hollander DB,et al. Validity of
`Wearable Activity Monitors during Cycling and Resistance
`Exercise. Med Sci Sports Exerc 2018;50:624-33.
`Thomson EA, Nuss K, Comstock A, et al. Heart rate
`measures from the Apple Watch, Fitbit Charge HR 2, and
`electrocardiogram across differentexercise intensities. J
`Sports Sci 2019;37:1411-9.
`Wallen MP, Gomersall SR, Keating SE, et al. Accuracy
`of Heart Rate Watches: Implications for Weight
`Management. PLoS One 2016;11:e0154420,
`Wang R, Blackburn G, Desai M,et al. Accuracy of Wrist-
`Worn Heart Rate Monitors. JAMA Cardiol 2017;2:104-6.
`Jackson AS, Blair SN, Mahar MT; et al. Prediction of
`functional aerobic capacity without exercise testing. Med
`Sci Sports Exere 1990;22:863-70.
`Weiglein L, Herrick J, Kirk S, et al. The 1-mile walk testis
`a valid predictor of VO(2max) andis a reliable alternative
`fitness test to the 1.5-mile run in U.S. Air Force males. Mil
`
`Med 2011;176:669-73.
`Sedgeman D, Dalleck L, Clark IE,et al. Analysis of
`square-wave bouts to verify VO2max. Int] Sports Med
`2013;34:1058-62.
`McCrory MA, Gomez TD, Bernauer EM,etal.
`Evaluation of a new air displacement plethysmograph
`for measuring human body composition. Med Sci Sports
`Exere 1995;27:1686-91.
`Siri WE. Body composition from fluid spaces and density:
`
`© mHealth.All rights reserved.
`
`mealth 2019;5:39 | http://dx.doi.org/10.21037/mhealth.2019.09.07
`
`PNA-FB0016687
`
`

`

`Case 1:19-cv-11586-FDS Document 343-12 Filed 03/02/22 Page 8 of 8
`Case 1:19-cv-11586-FDS Document 343-12 Filed 03/02/22 Page8of8
`
`mHealth, 2019
`
`Page 7 of 7
`
`19,
`
`20.
`
`21.
`
`22;
`
`Available online: https://assets.firstbeat.com/firstbeat/
`uploads/2017/06/white_paper_VO2max_30.6.2017.pdf
`Kraft GL, Dow M.Validation of the Polar Fitness Test.
`Int J Innov Educ Res 2018;6:27-34.
`Poole DC, Jones AM. Measurementof the maximum
`oxygen uptake Vo2max: Vo2peakis no longer acceptable. J
`ApplPhysiol 1985 2017;122:997-1002.
`Menghini L, Gianfranchi F, Cellini N, et al. Stressing
`the accuracy: Wrist-worn wearable sensor validation over
`different conditions. Psychophysiology 2019;e13441.
`Shcherbina A, Mattsson CM, Waggott D, et al. Accuracy
`in Wrist-Worn, Sensor-Based Measurements of Heart
`Rate and Energy Expenditure in a Diverse Cohort. J Pers
`Med 2017. doi: 10.3390/jpm7020003.
`
`analysis of methods 1961. Nutrition 1993;9:480-91;
`discussion 480, 492.
`Bassett DR Jr, Howley ET, Thompson DL,etal. Validity
`of inspiratory and expiratory methods of measuring gas
`exchange with a computerized system. J Appl Physiol
`(1985) 2001;91:218-24.
`Midgley AW, Carroll S. Emergence ofthe verification
`phase procedure for confirming “true” VO(2max). Scand J
`MedSci Sports 2009;19:313-22.
`Midgley AW, McNaughton LR, Carroll S. Verification
`phase as a useful tool in the determination of the maximal
`oxygen uptake of distance runners. Appl Physiol Nutr
`Metab 2006;31:541-8.
`Automated Fitness Level (VO2max) Estimation with
`Heart Rate and Speed Data.[cited 2019 Aug 12].
`
`15.
`
`16.
`
`17.
`
`18.
`
`doi: 10.21037/mhealth.2019.09.07
`
`Cite this article as: Freeberg KA, Baughman BR, Vickey
`T, Sullivan JA, Sawyer BJ. Assessing the ability of the Fitbit
`Charge 2 to accurately predict VO,,,.,. mHealth 2019;5:39.
`
`© mHealth.All rights reserved.
`
`mealth 2019;5:39 | http://dx.doi.org/10.21037/mhealth.2019.09.07
`
`PNA-FB0016688
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket