throbber

`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`
`__________________________________
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`__________________________________
`
`INGURAN, LLC d/b/a SEXING TECHNOLOGIES,
`
`Petitioner
`
`v.
`
`PREMIUM GENETICS (UK) LTD.,
`
`Patent Owner
`
`
`
`__________________________________
`
`Case PGR: Unassigned
`
`Patent No. 8,933,395
`
`__________________________________
`
`DECLARATION OF GIACOMO VACCA, PH.D.
`
`
`
`
`
`
` Exhibit No. 1002
` PGR of U.S. Patent 8,933,395
`
`

`

`
`
`TABLE OF CONTENTS
`
`I.
`
`II.
`
`III.
`
`IV.
`
`INTRODUCTION ...............................................................................................................1
`
`QUALIFICATIONS AND BACKGROUND ......................................................................1
`
`BACKGROUND OF TECHNOLOGY ...............................................................................5
`A.
`Flow Cytometry .......................................................................................................6
`B.
`Cell Sorting ..............................................................................................................9
`C.
`Laser Killing .......................................................................................................... 11
`D.
`Sperm Sorting ........................................................................................................12
`E.
`Bulk Cell Separation ..............................................................................................14
`
`THE ‘395 PATENT ............................................................................................................16
`A.
`‘395 Patent Specification .......................................................................................16
`B.
`‘395 Patent Claims .................................................................................................21
`C.
`‘395 Patent Prosecution History ............................................................................29
`
`V.
`
`LEGAL PRINCIPLES APPLIED ......................................................................................31
`
`VI.
`
`PERSON OF ORDINARY SKILL IN THE ART ..............................................................34
`
`VII. CLAIM CONSTRUCTION AND INDEFINITENESS ....................................................34
`
`VIII. THE SPECIFICATION DOES NOT ENABLE THE SUBJECT MATTER OF
`CLAIMS 1-14 OF THE ‘395 PATENT .............................................................................37
`A.
`Non-Enablement of Claim 1 of the ‘395 Patent.....................................................37
`B.
`Non-Enablement of Claim 2 of the ‘395 Patent.....................................................53
`
`IX.
`
`X.
`
`XI.
`
`SUMMARY OF THE PRIOR ART ...................................................................................78
`A.
`Mueth – U.S. Patent 7,355,696 B2 ........................................................................78
`B.
`Durack – PCT Publication No. WO 2004/088283 .................................................83
`C.
`Frontin-Rollet – PCT Publication No. WO 2005/075629 A1 ................................90
`D. Wada – U.S. Patent No. 6,506,609 B1 ...................................................................93
`Kachel – J. Hystochem. Cytochem. 25, 774 (1977) ...............................................98
`E.
`
`CLAIMS 1-13 OF THE ‘395 PATENT ARE ANTICIPATED BY MUETH ...................102
`
`CLAIM 14 OF THE ‘395 PATENT IS OBVIOUS IN VIEW OF MUETH,
`EITHER ALONE, OR IN LIGHT OF DURACK ...........................................................136
`
`XII. CLAIM 1 OF THE ‘395 PATENT IS ANTICIPATED BY FRONTIN-ROLLET ...........142
`
`XIII. CLAIM 1 OF THE ‘395 PATENT IS ANTICIPATED BY DURACK ............................150
`
`XIV. CLAIMS 2-14 OF THE ‘395 PATENT ARE OBVIOUS IN VIEW OF WADA
`AND KACHEL AND DURACK ....................................................................................167
`A.
`Motivation to Combine Wada with Kachel and Durack ......................................167
`
`XV.
`
`SECONDARY CONSIDERATIONS ..............................................................................208
`
`XVI. CONCLUSION ................................................................................................................208
`
`
`
`
`i
`
`

`

`I.
`
`INTRODUCTION
`
`
`
`1. My name is Dr. Giacomo Vacca. I have been retained by Inguran
`
`LLC, dba Sexing Technologies (“Petitioner”) as an independent expert consultant
`
`in this proceeding before the United States Patent and Trademark Office. Although
`
`I am being compensated at my standard consulting rate of $300 per hour for the
`
`time I spend on this matter, I have no personal financial interest in any of the
`
`entities involved in this proceeding, and my compensation does not depend in any
`
`way on my testimony, my conclusions, or the outcome of my analysis.
`
`2.
`
`I have been asked to consider and provide my opinions regarding the
`
`prior art in relation to the claims of U.S. Patent No. 8,933,395 (“‘395 patent”). My
`
`opinions and the bases for my opinions are outlined below.
`
`3.
`
`The testimony I may provide about scientific principles relating to my
`
`analysis and opinions, includes subjects relating to, for example, optics, photonics,
`
`particle analysis, microfluidics, device and system design, and their application to
`
`flow cytometry and particle selection and sorting.
`
`II. QUALIFICATIONS AND BACKGROUND
`
`4.
`
`Attached as Exhibit 1003 is my curriculum vitae, incorporated by
`
`reference within this declaration. The following is a brief summary of my
`
`qualifications.
`
`
`
`1
`
`

`

`
`
`5.
`
`I received a Bachelor of Arts (B.A.) degree and a Master of Arts
`
`(M.A.) degree, both from Harvard University, in 1991, and a Doctorate (Ph.D.) in
`
`Applied Physics from Stanford University in 2001. As part of my doctoral
`
`dissertation work, I invented a technique to use light scattering to probe fluid
`
`phenomena on very short timescales, and designed and built a laboratory and all
`
`necessary custom equipment to demonstrate it. While pursuing my doctoral
`
`studies, I also worked as a Graduate Teaching Assistant, leading laboratory and
`
`discussion sessions and grading student assignments for physics courses.
`
`6.
`
`From 1991 to 1994, I was Associate Physicist at Exxon Research &
`
`Engineering Company in Clinton, New Jersey. This facility was at the time the
`
`corporate research headquarters of Exxon Corporation. There I conducted X-ray
`
`scattering experiments to analyze and characterize complex fluids, thin films, and
`
`composite materials, and designed and built equipment to study the flow of
`
`multiphase fluids in porous media.
`
`7.
`
`From 2000
`
`to 2002, I was Design Physicist at Lightwave
`
`Microsystems Corp. in San Jose, California. I was a co-inventor of several issued
`
`patents at the intersection of optics and microfluidics, for applications from
`
`telecommunication to biotechnology. I set up a new laboratory for research and
`
`development of microfluidics-based optical devices, and prototyped early device
`
`concepts. I also designed optical integrated circuits (the equivalent of electronic
`
`
`
`2
`
`

`

`
`
`integrated circuits, using light instead of electricity), modeled device behavior and
`
`analyzed fabrication data sets to improve yield and performance.
`
`8.
`
`From 2002 to 2005, I took on the successive roles of Optical
`
`Engineer, Project Leader, and Product Marketing Manager at Picarro, Inc., in
`
`Sunnyvale, California, a company which at the time was developing and
`
`manufacturing solid-state lasers and laser-based instrumentation for applications in
`
`flow cytometry, microscopy, and spectroscopy. I was involved in the development,
`
`transfer to manufacturing, and, later, marketing, of the Cyan laser product line—a
`
`compact and robust solid-state 488-nm laser light source aimed at the flow
`
`cytometry market. I also led the development of a tunable solid-state infrared laser
`
`for spectroscopy applications.
`
`9.
`
`From 2005 to 2011, I was Program Development Manager in the New
`
`Product Introduction department of the Hematology Business Unit, Diagnostics
`
`Division, of Abbott Laboratories in Santa Clara, California. I later took on
`
`additional, concurrent, roles of Intellectual Property Manager in the same business
`
`unit; and Member of the New Technology Group in the Diagnostics Division. I
`
`was responsible for generating, proving, developing, and ultimately launching new
`
`technology concepts
`
`in
`
`flow-cytometry-based semi-automated hematology
`
`analysis, encompassing all core aspects of reagent design, fluidic control, optical
`
`interrogation, signal processing, and cell identification algorithms. In 2010, in
`
`
`
`3
`
`

`

`
`
`recognition of my research and innovation efforts, I was inducted as a Research
`
`Fellow into Abbott Laboratories’ Volwiler Scientific Society—a scientific honor
`
`bestowed on less than 0.04% of the company’s then worldwide employee base of
`
`more than 80,000. I also received several company awards for both research and
`
`research management.
`
`10. Since 2011, I have been leading and growing the company I founded,
`
`Kinetic River Corp., a biophotonics design, consulting, and product development
`
`firm based in California’s Silicon Valley. Through my company, I have been
`
`consulting for firms ranging from one-person startups
`
`to Fortune 500
`
`multinationals; our client engagements cover a wide range of project types, from
`
`design and prototyping to intellectual property analysis, competitive landscaping,
`
`and technical due diligence in Mergers & Acquisitions (M&A). The focus of my
`
`company’s efforts is the intersection of light and biology: flow cytometry is the
`
`core competency, and most of our projects involve cell analysis.
`
`11.
`
`In 2013, I co-founded BeamWise, Inc., a company developing
`
`software tools for the automation of optical system design, based in San Jose,
`
`California. I am Chief Scientific Officer at BeamWise, a role I carry out
`
`concurrently to my role as President of Kinetic River Corp.
`
`12.
`
`I have published a number of papers in various areas of optics; most
`
`recently I have written flow-cytometry-related research and technical articles in
`
`
`
`4
`
`

`

`
`
`peer-reviewed papers, conference proceedings, and professional publications. I am
`
`the inventor or co-inventor on 12 issued United States patents, 12 pending United
`
`States patent applications, and a number of international patent applications. I
`
`have presented my original work regularly at flow cytometry and optics
`
`conferences and workshops, and I have been invited to chair sessions, deliver talks,
`
`lead workshops, and participate in panel discussions on topics from flow cytometry
`
`instrumentation to microfluidics to optical system design. A full list of my
`
`publications, patents and patent applications, conference presentations and invited
`
`talks is included in Exhibit 1003.
`
`13. A list of the documents and material I have considered in developing
`
`my opinion can be found in the Appendix below.
`
`III. BACKGROUND OF TECHNOLOGY
`
`14. The technology relevant to the ‘395 patent includes several broad
`
`fields. One of them is bulk separation of particles in fluids; another is single-cell
`
`analysis and sorting. The field of single-cell analysis is generally known as flow
`
`cytometry (flow cytometry actually encompassing more broadly the analysis of
`
`particles, not only cells), while the field of single-cell sorting is simply known as
`
`cell sorting. Below, I provide some relevant background about this technology as
`
`it stood in January, 2014 based on my recollection and experience.
`
`
`
`5
`
`

`

`
`
`A.
`
`Flow Cytometry
`
`15. Flow cytometry has roots that go back more than five decades. For a
`
`comprehensive treatise on the technology, its history, the principles of operation,
`
`and many other related topics, see generally, H. Shapiro, Practical Flow Cytometry
`
`4th ed., (Wiley, 2003).
`
`16. Flow cytometry as a discipline has practitioners in most academic
`
`institutions and life-science research laboratories. They include cell biologists,
`
`immunologists, cancer researchers, and researchers in many other related fields
`
`that rely on analysis of cells. As a field of industry, flow cytometry is very well
`
`established in clinical laboratories worldwide through the many thousands of cell
`
`analyzers (as flow cytometers are sometimes called) installed for clinical use,
`
`mainly for HIV/AIDS monitoring but also for aiding in cancer diagnosis and
`
`prognosis (e.g., leukemia/lymphoma). A related class of instruments called
`
`hematology analyzers, which are in fact flow cytometers, forms an even larger
`
`installed base of many tens of thousands of instruments worldwide that perform
`
`routine blood cell counting. Additionally, flow cytometers are in widespread use in
`
`commercial research laboratories, especially at pharmaceutical companies, where
`
`they are used extensively in the drug discovery and development process. There
`
`are also classes of instruments that qualify as flow cytometers on the basis of their
`
`principle of operation, but that are applied to non-life-science-related problems—
`
`
`
`6
`
`

`

`
`
`for example, quality assurance in commercial production of milk and other foods,
`
`and industrial process monitoring, such as characterization of particle size
`
`distribution in paint. In what follows I will discuss flow cytometers as applied to
`
`life science problems based on my recollection and experience in the field, with the
`
`understanding that the principles are very similar to other applications.
`
`17. Flow cytometers measure cells. The vast majority of flow cytometers
`
`are based on optical interrogation and detection of cells. Other techniques are also
`
`in use, most notably electrical impedance measurements (based on the Coulter
`
`principle); however, these tend to be either implemented in entry-level, fewer-
`
`featured instruments, e.g., simple cell counters; or they are incorporated as adjunct
`
`measurements on optical flow cytometers. In what follows I will discuss
`
`exclusively optical flow cytometers and refer to them as simply flow cytometers.
`
`18.
`
`In my experience, a flow cytometer is typically based on the following
`
`elements: (i) sample fluid; (ii) sheath fluid; (iii) confinement of flowing sample
`
`fluid by flowing sheath fluid; (iv) means to optically interrogate cells (or particles)
`
`in the sample; (v) means to detect optical signals from the cells (or particles) in the
`
`sample. In the system design common to a large majority of all flow cytometers,
`
`the sheath fluid (typically an aqueous buffer solution) envelops concentrically the
`
`sample fluid at the injection point; the combined concentric flows are then forced
`
`through a funnel-shaped section that reduces the cross-section of both by factors of
`
`
`
`7
`
`

`

`
`
`about 100 in each cross-sectional dimension, while at the same time accelerating
`
`the fluid by a factor of about 10,000 (this is called hydrodynamic focusing); the
`
`resulting high-speed, sheath-enclosed narrow sample stream, now squeezed to the
`
`degree that the cells in the sample are in single file, is passed through an optical
`
`interrogation section, where light from one or more sources (typically lasers) is
`
`focused onto the cells, which in turn scatter a portion of the light and (mainly if
`
`dyes are present) absorb another portion of the light to re-emit it as fluorescence;
`
`optical detectors situated at various angles around the interrogation point collect
`
`the optical signals, which are then processed by electronics and are conveyed to a
`
`computer for additional user analysis or storage.
`
`19. This basic framework has exceptions. For example, there are
`
`“sheathless” flow cytometers where single-cell interrogation is not a consequence
`
`of hydrodynamic focusing by the sheath fluid, but of pre-dilution of the sample
`
`fluid. There are also flow cytometers (such as a large portion of hematology
`
`analyzers) which do not interrogate using fluorescence, but only scattered light;
`
`others which only use fluorescence; and others yet which use multiple
`
`interrogation paths at different angles from different directions. Generally,
`
`however, in my experience, the framework described above is both very common
`
`and useful for understanding the principles of flow cytometry even in those cases
`
`where there is a departure from it.
`
`
`
`8
`
`

`

`
`
`20. Some of the main advantages of flow cytometry relevant to this
`
`discussion are: (a) the ability to interrogate individual cells at very high rates (10-
`
`20,000 events per second are routine), as compared, e.g., with microscopy and (b)
`
`the ability to collect multiparameter information on each cell simultaneously, such
`
`as, e.g., using multiple fluorophores conjugated to different antibodies to probe cell
`
`surface antigen expression. One of the drawbacks, again as compared with
`
`microscopy, is the lack of an image to associate with each detected cell (with the
`
`exception of one line of commercial instruments currently available); however, in
`
`many applications such a drawback is tolerable and in some even inconsequential.
`
`B. Cell Sorting
`
`21. Almost as soon as flow cytometers were invented, their function was
`
`extended beyond simply measuring cells to enable the sorting of cells. A flow
`
`cytometer’s ability to analyze single cells over multiple parameters means that
`
`individual cells satisfying certain selection criteria can be identified out of large
`
`populations of other cells; with cell sorting, those cells can be segregated from the
`
`rest of the population for further analysis or use.
`
`22. Based on my experience, until recently, overwhelmingly cell sorting
`
`was performed using the jet-in-air approach, where the sheathed sample is ejected
`
`out of a nozzle as a jet and immediately subjected to optical interrogation in air; the
`
`jet then proceeds to break up into fine droplets through the action of a piezoelectric
`
`
`
`9
`
`

`

`
`
`actuator operating at high frequency; just around the break-off point, a droplet is
`
`given either an electrical charge or no charge depending on whether, based on real-
`
`time analysis of the detected optical signals, the droplet is found to contain a cell
`
`satisfying the selection criteria; the droplets continue on, but depending on their
`
`electrical charge they take different paths due to the application of a large
`
`electrostatic voltage potential to two deflection plates on either side of the droplet
`
`stream. Therefore the streams (there can be two or more, with some systems
`
`performing an 8-way sort by applying up to 8 different levels of electrical charge
`
`according to corresponding selection criteria) take separate paths to end up in
`
`different collection receptacles.
`
`23.
`
`In my opinion, jet-in-air cell sorting has been engineered to a very
`
`high degree, resulting in droplet formation rates of hundreds of kilohertz and sort
`
`rates of tens of kilohertz. Nozzle, piezoelectric actuator, and fluid delivery designs
`
`are capable of delivering extremely stable jets and very reproducible droplet
`
`formation. Trade-offs include the sorting rate (throughput), the purity of the sorted
`
`cells (to what degree unwanted cells contaminate the desired sorted population),
`
`the recovery (to what degree wanted cells are lost in the unwanted stream), and the
`
`yield (to what degree cells are lost or destroyed in the process of sorting).
`
`24. Beginning in early 2000’s, novel sorting designs that do not rely on
`
`the jet-in-air approach started to appear with regularity in the literature. Instead,
`
`
`
`10
`
`

`

`
`
`they are based on microfluidic chips, in which channels are created in a planar
`
`substrate by lithography or other fabrication means; the resulting fluidic circuits
`
`can be much more sophisticated than traditional flow cytometers and sorters,
`
`combining multiple injection paths and other functionality beside interrogation and
`
`selection. Additionally, the fluid path can be fully enclosed, eliminating the
`
`formation of potentially hazardous aerosols. The generally lower pressures and
`
`speed also tend to be gentler on the cells, potentially increasing yield. On the other
`
`hand, actuation of sorting mechanisms on microfluidic chips is a relatively young
`
`field and the task is challenging, and so far microfluidic sorters have not come
`
`close to matching the performance of jet-in-air systems in terms of throughput.
`
`Also it is challenging to incorporate into microfluidic chips three-dimensional
`
`hydrodynamic structures like the round funnel that is one of the mainstays of flow
`
`cytometry.
`
`C. Laser Killing
`
`25. Cell sorting allows cells defined by certain criteria to be segregated
`
`out of a larger population at relatively high efficiency with potentially extremely
`
`high purity. This is key for many fields where contamination of the desired sorted
`
`stream by unwanted cells is extremely undesirable or even disastrous—for
`
`example, when follow-on analysis involves molecular analysis of the cells’ DNA.
`
`However, for certain applications it is not important that the sorted stream contain
`
`
`
`11
`
`

`

`
`
`only the wanted cells—it is enough that the sorted stream contain only live wanted
`
`cells. For example, it is my understanding that if the follow-on analysis is a cell
`
`culture, then if the sorted stream contains also dead unwanted cells it is
`
`inconsequential—it is the same as if the unwanted cells had been removed.
`
`26. Based on my understanding of the technology, laser killing offers such
`
`a solution. By operating at the same rates as the optical analysis, laser killing can
`
`be used to damage or destroy individual cells while still in the flow channel,
`
`somewhat downstream of the optical analysis to give the electronics time to
`
`process the detected signals and generate a logic verdict on whether to kill a given
`
`cell or not. The advantages of laser killing are several: first of all it removes the
`
`need for droplet sorting, with all the associated hardware, software, and
`
`electronics. It also allows the sorting process to occur in the confinement of a flow
`
`channel—eliminating aerosols and the potentially cell-damaging process of droplet
`
`impact.
`
`D.
`
`Sperm Sorting
`
`27.
`
`In my opinion, of the many applications of cell sorting, sperm sorting
`
`is one that has tended to push the throughput envelope the most, at least so far.
`
`Unlike research applications of cell sorting, where collection of a few thousand
`
`desired cells in any given experiment is sufficient for many analytical purposes, in
`
`commercial animal husbandry (and particularly in the bovine industry) the need is
`
`
`
`12
`
`

`

`
`
`there to process vast quantities of bull sperm to select the gender of the offspring.
`
`With typical bull ejaculates amounting to several milliliters and billions of sperm
`
`cells, bovine sperm cell gender selection by flow cytometry and cell sorting
`
`requires enormous throughputs to be commercially viable on a herd scale.
`
`Therefore, much effort has been expended optimizing techniques and technologies
`
`for this purpose.
`
`28. Based on my understanding of the technology, sperm sorting offers
`
`some unique challenges as compared to flow cytometric analysis and sorting of,
`
`say, peripheral white blood cells. One difference is that multiparametric analysis is
`
`not important for sperm sorting; for gender selection, one generally only wants to
`
`know whether a sperm cell has the X chromosome or the Y chromosome. This can
`
`in principle be done with a single detection channel tuned to fluorescence from a
`
`dye that binds to DNA. However, the unique morphology of sperm cells make this
`
`harder than it might seem. Generally shaped as oblong flat discs, sperm cells
`
`interact with
`
`the
`
`interrogating
`
`light beam
`
`in flow cytometers
`
`in rather
`
`unpredictable ways as compared to more or less round white blood cells. In
`
`particular, random orientation of sperm cells with respect to the incoming beam
`
`can cause large differences in both scattered light and emitted fluorescence,
`
`washing out the smaller differences between X- and Y-chromosome-bearing cells
`
`due to their different DNA content. See Johnson, Ex. 1010. In an early patent
`
`
`
`13
`
`

`

`
`
`relating to sperm sorting, Larry Johnson with the USDA attempted to solve
`
`problems caused by rotational orientation of the sperm cells by using two photo
`
`detectors. The first determines whether the sperm cells are properly oriented,
`
`while the second takes a measurement that is used to classify the sperm as having
`
`an X or Y chromosome.
`
`29. Another way to solve this problem is to stably and uniformly align the
`
`sperm cells as they are funneled to the interrogation region so they are uniformly
`
`oriented when they interact with the light beam. Then the variations among
`
`individual cells due to orientation differences are minimized, and the variations due
`
`to DNA content can be detected.
`
`30. Another aspect of the sperm-sorting application is that the unwanted
`
`type of sperm cell (be it X or Y chromosome bearing) can be damaged or killed.
`
`See, e.g., Durack, Ex. 1005, 144:18-21 (discussing photo-damage sorting
`
`techniques). Therefore, sperm sorting admits of approaches like laser killing.
`
`E.
`
`Bulk Cell Separation
`
`31. While flow cytometry and cell sorting are based on a single-cell
`
`approach (thereby giving those techniques great flexibility and allowing a high
`
`degree of precision in selecting specific subpopulations of cells from a mixture),
`
`there have long been other approaches of cell separation that tackle the problem in
`
`altogether different ways. Cell sorting is a serial approach: each cell is measured
`
`
`
`14
`
`

`

`
`
`individually, one after another. By contrast, bulk cell separation is a parallel
`
`approach: all cells are processed simultaneously, generally in batches. One
`
`example of bulk cell separation is centrifugation: By subjecting a fluid suspension
`
`(such as blood) to high acceleration in a spinning geometry, the natural process of
`
`sedimentation can be sped up and cells of different densities made to stratify in a
`
`fraction of the time it would normally take them to settle completely. Other
`
`approaches differ from centrifugation, but have in common with it the reliance on
`
`parallel processing of multiple cells based on intrinsic cell features.
`
`32. For what they offer in parallelism, bulk cell separation methods
`
`generally lack severely in precision. Sedimentation, for example, is mediated by
`
`gravity on the basis of density differences, size differences, and other generally
`
`cell-intrinsic factors; the resulting segregation of cell subpopulations is much
`
`coarser and imprecisely defined than can be achieved using, say, flow cytometry
`
`and cell sorting on the basis of single-cell analysis, even with few detection
`
`parameters. Accordingly, bulk cell separation methods have their place but, in my
`
`experience, they are not generally thought of as direct alternatives to single-cell
`
`analysis and sorting.
`
`
`
`15
`
`

`

`
`
`IV. THE ‘395 PATENT
`
`A.
`
`‘395 Patent Specification
`
`33. The ‘395 patent is entitled “Multiple laminar flow-based particle and
`
`cellular identification” and issued on January 13, 2015. It is generally directed to
`
`“techniques and systems for separation of particulate or cellular materials such as
`
`blood, semen and other particles or cells into their various components and
`
`fractions, using multiple laminar flows which further may be coupled with laser
`
`steering such as holographic optical tapping [sic] and manipulation” (Ex. 1001,
`
`1:65-2:3).
`
`34. The Summary of the Invention in the ‘395 patent states:
`
` “[A]n exemplary method of separating blood into components
`
`includes providing a first flow having a plurality of blood
`
`components; providing a second flow; contacting the first flow with
`
`the second flow to provide a first separation region; and differentially
`
`sedimenting a first blood cellular component of the plurality of blood
`
`components into the second flow while concurrently maintaining a
`
`second blood cellular component of the plurality of blood components
`
`in the first flow. The second flow having the first blood cellular
`
`component is then differentially removed from the first flow having
`
`the second blood cellular component”
`
`(Id., 6:9-19).
`
` “…[T]he first blood cellular component is a plurality of red blood
`
`cells and a plurality of white blood cells, and the second blood cellular
`
`
`
`16
`
`

`

`
`
`component is a plurality of platelets. For the first blood cellular
`
`component, the plurality of white blood cells may be holographically
`
`separated (through laser steering) from the plurality of red blood
`
`cells”
`
`(Id., 6:24-30).
`
`35. The ‘395 patent specification is focused on methods and systems for
`
`physical separation of particles, especially cells, in suspensions. In particular, the
`
`patent teaches separation by differential sedimentation, optionally supplemented by
`
`optical trapping. The applications mentioned in the specification vary, e.g., from
`
`separation of blood components to dialysis; however, the technologies and methods
`
`described in the patent are all based on physical particle separation in suspension
`
`flows. The emphasis on gravity-driven sedimentation is evident in many of the
`
`figures, such as, e.g., Fig. 1, where the differential settling, or sedimentation, of
`
`various components of the blood sample entering the top input channel is shown:
`
`
`
`17
`
`

`

`
`
`
`
`Figure 1 of the '395 patent illustrating physical particle separation by
`differential sedimentation.
`
`36. Likewise, the emphasis of the specification of the ‘395 patent on
`
`optical trapping is evident in many other figures, such as, e.g., Fig. 13, where a set
`
`of optical traps acts to dynamically drag selected components out of the sample
`
`flow into the adjacent separation flow:
`
`Figure 13 of the '395 patent illustrating physical particle separation by
`optical trapping.
`
`
`
`
`
`
`18
`
`

`

`
`
`The specification discloses a few other cell manipulation methods that are used in
`
`conjunction with the multiple flow techniques. For example, it discloses a
`
`spinning disc-based sorter that uses the rotational motion of a disc combined with
`
`optical trapping to sort cells, a Fluorescence-Activated Cell Sorting mechanism,
`
`and the use of lasers. However, each of these alternate sorting techniques is used
`
`in conjunction with the optical trapping apparatus discussed above, physical
`
`separation of cells, or the other cell manipulation techniques discussed in the
`
`specification, and thus only in context of physically separating particles, especially
`
`cells, in suspensions. For example, with respect to killing cells with lasers, the
`
`specification explicitly states that the dead cells are physically removed (Ex. 1001,
`
`40:10-14) and/or used in conjunction with holographic optical traps (Ex. 1001,
`
`40:19-23).
`
`37. As an indication of the focus of the ‘395 patent, the word
`
`“sedimentation” appears 37 separate times in the specification; the phrase “optical
`
`trapping” appears 42 separate times in the specification (not counting citations).
`
`By contrast, the words “analysis,” “analyze,” or “analyzed” only appear 7 times in
`
`the specification; the phrase “flow cytometry” only appears in a citation; and
`
`neither the phrase “cell analysis” nor the phrase “particle analysis” nor the words
`
`“examination” or “examine” appear at all.
`
`
`
`19
`
`

`

`
`
`38. The fact that the ‘395 patent is directed principally to physical particle
`
`separation in suspension flow (whether by bulk sedimentation, optical trapping, or
`
`combinations thereof) and not to particle or cell analysis or examination is further
`
`supported by inspection of the Figures. All of the 27 Figures of the specification
`
`are directed explicitly to illustrating methods and technologies for particle
`
`separation; and none of the Figures are principally directed to illustrating methods
`
`or technologies for particle analysis. Fig. 5 includes elements for illumination and
`
`imaging via a CCD camera; however, the only description of these elements is the
`
`following: “For imaging, an illumination source 503 is provided above the
`
`objective lens 504 to illuminate the sample 506.” (Id., 32:67-33:2)
`
`39. Fig. 12 mentions a system for imaging and trapping and a system for
`
`control and analysis; however,

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket