throbber
RX-0035.0218
`
`APL_MAS_ITC_00015835
`
`11
`
`1
`
`1
`
`User interface for a pulse oximeter
`
`201
`
`problem. Other applications are to provide feedback to an input from the
`operator, and in some units to codify the patient pulse strength by changing the
`sound pitch accordingly to the strength as defined in the Standards (ASTM 1992).
`Despite these acoustical outputs, the primary output of a pulse oximeter is
`visual. Pulse oximeters can be primarily classified based on the technique used to
`present visual information into two categories:
`
`1. Graphical displays that present analog and digital information.
`2. Numerical displays that only present digital information.
`
`12.2.1 Graphical displays
`
`It is common knowledge that 'a picture is worth thousand words', and pulse
`oximeters are not an exception. Graphs produce a spatial presentation to
`communicate quantitative information to the exterior world, making them very
`flexible (Gillan and Lewis 1994). The displays used in pulse oximeters are
`normally liquid crystal displays (LCDs), although some models from Protocol
`Systems Inc. (Propaq 102/104/106) also have versions with an electroluminescent
`display (ELD). ELD displays perform better when it is necessary to view them
`from long distances. They are aimed toward bedside monitoring, where the units
`can be plugged to a power line source, because of the higher power that these
`displays require. On the other hand, LCD displays are better in direct sunlight
`and require much less power, which increases both the display life and the battery
`discharge cycle (Bosman 1989). Most of the commercially available LCD units
`have a backlight that increases display readability but also dramatically decreases
`the battery operating time. For example, Criticare specifies for its 503 model, a
`battery use time of 20 h when the backlight is turned off, while it decreases to 10
`h when the backlight is turned on.
`Graphical displays present one or more real-time waveforms. Normally, the
`units that incorporate graphical displays are also the ones that acquire more
`physiological signals, so there are more choices for display. All the units with
`graphical displays can simultaneously present different waveforms, although for
`readability it is not convenient to present more than two. The most common
`waveforms are the plethysmographic waveform and the ECG. The model POET
`TE Plus from Criticare also monitors CO2 and can display the capnographic
`waveform. The Propaq models from Protocol Systems, Inc., have different
`modular systems that can measure oxygen saturation, ECG, CO2 consumption,
`and invasive and noninvasive blood pressure. The units from Medical Research
`Laboratories, Inc. can be used as stand-alone systems or as a part of an integrated
`monitoring system as previously described. The model 9500 from Magnetic
`Resonance Equipment Co. is a multigas monitoring system that measures oxygen
`saturation, CO2, N02.02 and invasive and noninvasive blood pressure. The
`model BIOX 3700 from Ohmeda, shown in figure 12.2 has two separated LCD
`displays with different functions for each one. One displays different waveforms,
`while the other displays the values of oxygen saturation and pulse rate.
`In addition to real-time waveforms, displays can also present the trend from
`a past period of time. This feature does not involve a major increase in the
`complexity of the electronic design because it only requires storage of the already
`digitized values and further processing. The length of time that is available for
`display depends on the amount of the memory used in the design, but also on the
`sampling frequency, which is normally user selectable. There is a large variation
`
`218
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01300
`
`

`

`RX-0035.0219
`
`APL_MAS_ITC_00015836
`
`11
`
`202
`
`Design of pulse oximeters
`
`among the length of time that different models store trend display. In the Biox
`3700 from Ohmeda, the length of the trend can be selected between 20 and 60
`min, by pressing a key in the front panel. as shown in figure 12.2. The model N-
`3000 from Nellcor has three different ways of recording data for trend analysis.
`In the first two modes, the unit stores the average of oxygen saturation and heart
`rate measured over a period of 5 or 10 s. with a total duration of I 2 or 24 h
`respectively. In the third mode, the unit stores the maximum and minimum values
`obtained over a period of 20 s, with a total duration of 32 h. The length of the
`recording also changes with presentation. The Propaq models from Proti,col
`Systems, Inc. can display a total of 5 h on the screen and 8 h on a printer with a
`rexolution of 2 min. The data can be presented in graphical or tabular forni.
`
`Wik-r
`
`9t
`ames 9 0000
`
`O-1
`Jt
`
`1
`
`Figure 12.2 Front panel of Ohmeda B iox 3700 pulse oximeter (Courtesy of Ohmeda). The
`display at the left side shows mal-time waveforms and pulse strength, while the display at the right
`side is used for alann settings
`
`In some units, for example the model 504 from Criticare, the memory in
`which the trend data are being stored is not erasable on power-off. When trend
`data that contain periods of time in which the unit was turned off are displayed,
`the time during which the unit was turned off is shown as special characters so
`that the operator can be aware of this situation. This feature allows us to follow a
`patient during a long duration in which constant monitoring is not required. The
`drawback of this feature is that it can acquire the trend from the wrong patient if
`the previous data are not erased before starting to monitor a new patient. The
`trend display also marks times during which alarm set points have been exceeded
`or the pulse has been lost.
`
`219
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01300
`
`

`

`RX-0035.0220
`
`APL_MAS_ITC_00015837
`
`1
`
`User interface for a pulse oximeter
`
`203
`
`Trend graphs incorporate cursors that can be scrolled through the display
`with numeric readouts that normally show the values of the waveform and the
`time. This feature is particularly useful when the screen displays different
`waveforms because they do not incorporate a numerical vertical axis and it is not
`possible to distinguish magnitude by only reading the screen. It is also very
`important to properly label the different waveforms, as the most commonly
`displayed waveforms (pulse rate versus time, and oxygen saturation versus time)
`can present numerical values very similar to each other and confuse the system
`operator. It is also important that the trend display have the capability to use the
`dynamic range available in the screen to more clearly show small changes. For
`example, the model 504 from Criticare displays the trend in oxygen saturation
`between 75% and 100%. Although large changes in oxygen saturation can be
`easily recognized, it is difficult to notice small changes at a glance because most
`of the monitored patients will not have such a large oxygen saturation change.
`A series of menus that appear on the screen normally permit the operator to
`select between the displayed waveforms, cursor displacement, and other function
`controls. The selection keys are placed under the display screen or at its sides,
`and the function of a particular key is automatically changed depending on the
`displayed screen mode.
`The display of pulse strength is mandatory for those pulse oximeters that
`display a normalized pulse waveform (ASTM 1992). The reason for this feature
`is because the amplitude of the plethysmographic signal can be changed by the
`operator in order to achieve a good dynamic range on the screen, and it is
`desirable to have an indication of pulse strength regardless of the operator
`settings. In units with graphical displays, it is commonly done by a graphic bar
`whose amplitude is proportional to the pulse strength, situated on one side of the
`screen, as for example the unit shown in figure 12.2. The display of the pulse
`strength must be accompanied by acoustical signals.
`Other information commonly found in graphical display units is the values at
`which alarms have been set, their status, low battery indication, system
`malfunctions, and other messages of interest to operators.
`
`12.2.2 Numerical displays
`
`1
`
`The majority of the marketed pulse oximeters use only a numerical display made
`of red LED segments. In all the units examined, information on oxygen
`saturation and heart rate is presented. In addition to these variables, the models
`507 and 5070 from Criticare Systems, Inc., that are complex monitoring units,
`also display the values of systolic, diastolic, and mean blood pressure. POET TE
`Plus from Protocol Systems, Inc. displays the values of oxygen saturation and
`C02· Because in some patients, oxygen saturation and heart rate can reach the
`same numerical values, it is highly desirable that the displays incorporate a fast
`and reliable way for the operator to associate the number on the panel with the
`physiological variable of interest. However, only a few units have this feature.
`For example, the model POET TE Plus from Criticare Systems, Inc. uses
`different colors for LED segments to display oxygen saturation and CO2. The
`model 3500 from Magnetic Resonance Equipment Co. and the models 504 and
`504S from Criticare Systems, use different size LED segments to display oxygen
`saturation and pulse rate, and Medical Research Laboratories, Inc. uses larger
`
`220
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01300
`
`

`

`RX-0035.0221
`
`APL_MAS_ITC_00015838
`
`204
`
`Design of pulse oximeters
`
`green LEDs for oxygen saturation display and smaller red LEDs for heart rate
`display. The pulse strength in all the units with numerical output is displayed
`using a LED bargraph
`
`12.3 FUNCTION CONTROLS
`
`Function controls carry out communication from the healthcare professionals to
`the pulse oximeter to achieve the proper monitoring and care for the patient.
`Function controls are basically used to operate alarms (set alarm values, activate,
`deactivate and silence alarms) and displace the cursors along the graphical screen
`in those units with this feature.
`It is possible to distinguish three different function controls: switches,
`turning knobs and keys. They do not all need to exist in the same unit.
`The main function of switches is to turn tile device on or off. In some units
`switches are replaced by keys. Since this is the most basic function in a pulse
`oximeter, it is important that it cannot be turned off accidentally. For this reason.
`some units have the main power switch or key in a lateral panel where it is
`unlikely to turn the power off by accident.
`There are few units that incorporate turning knobs. The model N-200 shown
`in figure 12.3 and model N-3()00 from Nellcor use turning knobs as an intuitive
`and quick way to increase or decrease the alarm settings. The turning knobs are
`placed on the front panel or on the top of the unit, where they are large and thus
`are easier to manipulate without affecting other controls. A function thal uses
`turning knobs for control has to be designed so that a movement upwards, to the
`right or in a clockwise direction increases the control function (ASTM 1992).
`
`e
`
`Figure 12.3 Nellcor Puritan Bennett N-200 pulse oximeter (Reprinted by permission of Nellcor
`Puritan Bennett, Pleasanton, California).
`
`The majority of pulse oximeters use keys as input devices to control the
`instrument. We can distinguish between units that use touch panel keys and units
`
`221
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01300
`
`

`

`RX-0035.0222
`
`APL_MAS_ITC_00015839
`
`User interface for a pulse oximeter
`
`205
`
`that use push buttons. Touch panel keys have the advantage that they are cheaper
`to manufacture and insert during the manufacturing process and can
`accommodate LEDs to indicate thal the function is active. They also contribute to
`a better seal of the unit's front panel. thus making it more suitable for use in
`hostile environments. Figure 12.2 shows a unit that uses these kind of keys for
`front panel functions. On the other hand, push buttons have a better feel and
`require lower pressure to activate. However, they have open spaces around them,
`can permit dust, humidity and other chemical agents to shorten the life of their
`electrical contacts.
`For any kind of keys used, it is desirable that the operator has a feedback
`that the key has been pressed successfully, either by a visual stimulus such as
`turning on a LED in a touch panel key, an audio stimulus by emitting a
`characteristic sound, or tactile feedback from the release of the pressed key
`pressing on the operator's finger (Cakir et al 1980).
`It is important to consider the number of different keys that are available in
`a pulse oximeter. In general, it is best to have as few keys as possible to simplify
`the access to the most common and critical functions. such as setting the alarm
`values. For example, the model N-200 from Nellcor has a very intuitive way of
`setting the alarm values (low oxygen saluralion, high oxygen saturation, low
`pulse rate, and high pulse rate) that consists of pressing a single key to select the
`alarm, and modify the actual value by rotating the turning knob as can be seen in
`figure 12.3. However, this device has only five different keys, so the operator
`needs to press two different keys simultaneously to activate other functions.
`Because the key labeling only refers to the basic function, it can become difficult
`to remember which keys need to be pressed in order to activate the desired
`function, and it is therefore harder to perform. In this particular unit, the
`manufacturer supplies a quick reference card to be placed on the bottom of the
`unit. It provides a helpful reminder to the operator if the operator knows where
`to look.
`On the other hand, the model 504US from Criticare uses the dynamic key
`function and labeling that has been described in previous sections. Wilh only
`lhrce touch panel keys for menu purposes, the operator enters a series of menus
`and submenus, changing the function of the keys according to the menu that is
`active. Allhough this way of controlling the functions has the advantage that the
`operator always knows lhe function of the set of keys, it is very easy to forget the
`depth of the menu entered, in which submenu a particular function of interest is
`located. It can also be time consuming to move between functions located in
`different submenus.
`In the same way that the operator needs feedback to indicate that a particular
`key has been pressed successfully, the operator also needs some feedback that
`indicates that the key, or combination of pressed keys, is valid, and a control
`funclion has been executed. The most common way to produce this feedback is by
`turning on a visual indicator that is related to the function executed, or by
`emitting a characteristic sound in the case of invalid keys.
`11 is also important to pay attention to the layout of displays and indicators
`and their control keys, selecting the position of the controls in a place that is
`consistent with the display. Figure 12.4 shows di fferent examples of good and
`poor relative positions between displays or indicators and controls, based on the
`idea that they have to be laid out in such a way that the relationship between
`controls and their indicators is obvious.
`
`222
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01300
`
`

`

`RX-0035.0223
`
`APL_MAS_ITC_00015840
`
`11
`
`1
`
`206
`
`Design of pulse oximeters
`
`2 1.711-3.1
`8 1-31 III
`
`Poor
`
`D-1 Fil 4-1
`
`Poor
`
`00 ~~
`
`O0 1-5-1 ~
`
`Good
`
`m Fill-31
`000
`
`Good
`
`Figure 12.4 Layout of controls and indicators to ensure good operator interaction. From
`Salvendy (1987).
`
`12.4 ALARM CONTROLS
`
`The alarms communicate the patient to the healthcare professionals, alerting of a
`potentially dangerous situation, Because the alanns are I he most critical functions
`in a pulse oximeter, it is absolutely necessary to be sure of their proper working
`condition, as well as to take extra effort to design them in such a way that they
`cannot be disconnected accidentally.
`The design of alarms and their controls section is by far the most regulated
`by the standards. The most common type of pulse oximeters, the units that display
`the oxygen saturation and heart rate, provide alarms for the following situations:
`
`1. High oxygen saturation.
`2. Low oxygen saturation.
`3. High pulse rate.
`4. Low pulse rate.
`
`Other sections in the ASTM Standard regarding the operation of alarms
`require that the alarm set points be operator adjustable, that the default limits on
`low oxygen saturation be 80% saturation or greater, and the difference between
`the alarm set point and the actual value of arterial oxygen saturation when the
`alarm is activated not exceed 2% of oxygen saturation (ASTM 1992).
`In most of the units, it is possible to deactivate at least the alarm for high
`oxygen saturation, except in the case when the pulse oximeter is configured for
`neonatal monitoring. The pulse oximeter shown in figure 12.2 has deactivated the
`alarms fur high oxygen saturation and high puse rate.
`From these alarms, only the low oxygen saturation alarm is required for the
`pulse oximeter to be qualified as a monitoring device. Those devices without low
`oxygen saturation the alarm shall be marked as 'NOT FOR MONITORING'
`(ASTM 1992). All tile marketed units examined provide these four alarm
`
`223
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01300
`
`

`

`RX-0035.0224
`
`APL_MAS_ITC_00015841
`
`User interface .for a pulse oximeter
`
`207
`
`situations, except the models 8500 and 9500 from Nonin Medical, Inc. These
`units have been designed not for a bedside monitoring situation in a hospital
`where the alarms are used to attract the operator's attention, but in a one-on-one
`working situation where a healthcare professional is always present with the
`patient, using the pulse oximeter to measure the oxygen saturation, for example,
`during ambulance transport.
`The 9500 unit, shown in figure 12.5, is the smallest available in the market.
`With a weight of only 36 g without batteries and an extremely small size, just
`slightly larger than most of the reusable finger probes, it displays heart rate and
`oxygen saturation. The unit 8500 is a hand-held pulse oximeter that has been
`designed to provide 100 h of continuous operation with batteries. Both units have
`been designed for evacuation situations. They both comply with the USAF
`vibration standards for helicopter tlight use, can operate at temperatures below
`freezing, and the manufacturer stresses their use in helicopter evacuation.
`
`. i
`
`Figure 12.5 A small Nonin model 9500 pulse oximeter designed for emergency evacuation
`purposes (courtesy of Nonin Medical Inc.).
`
`The visual and acoustic characteristics of the alarms are also regulated by the
`ASTM Standards, as shown in table 12.1. The ASTM differentiates three kinds of
`alarms based on their priority, assigning different colors and flashing frequency
`to each one.
`
`224
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01300
`
`

`

`RX-0035.0225
`
`APL_MAS_ITC_00015842
`
`IH
`
`208
`
`Design ofpu/se o.runeters
`
`Table 12.1 Aim·m characteristics for pulse oximeters (ASTM 1992).
`
`11
`
`R.ishing
`Alarm category Operator response Audible indicators Indicator color frequency (Hz)
`Red
`immediate
`High priority
`Not medium or
`1.4 to 2.8
`low prioi-ity
`Not high or low Yellow
`priority
`Not high or
`medium priority
`
`Medium priorily Prompt
`
`Low pi·iority
`
`Awareness
`
`Yellow
`
`0.4 to 0.8
`
`Constant
`
`The current ASTM Standard specifies neither the frequency nor the volume
`of the acoustic alarm sounds. Good practice suggests that the frequency of
`warning sounds should be between 150 Hz and 1000 Hz. It should have at least
`four frequency components in order to avoid masking from environmental noise.
`The acoustic level recommended is 15 dB to 16 d B above the masked threshold
`for signals that are triggered by situations that require a rapid response, and
`levels between 6 dB and I O dB above the masked threshold for all other kinds of
`signals, to achieve 100% detectability in controlled situations. In all cases, the
`level should be less than 30 dB above the masked threshold to minimize operator
`annoyance and disruption of communications (Salvendy 1987).
`The alarms in a pulse oximeter can be disconnected or silenced. Temporary
`silencing should be used when the operator has been alerted of the potentially
`dangerous situation and has taken steps in order to solve the problem, The
`Standard specifies that if this feature is provided in the pulse oximeter, it should
`not exceed 120 s, and a visual condition of the alarm has to remain on until the
`condition that triggered the alarm is corrected (ASTM 1992). The reason pulse
`oximeters incorporate a permanent silencing alarm is to avoid nuisance noise
`when the device and probe are being connected to the patient. The permanent
`alarm silencing activation must be designed in such a way that it requires a
`deliberate action for deactivation by the operator to be sure that it is not done in '
`error. It also requires a visual indication of this condition.
`As most of the pulse oximeters monitor heart rate from the
`plethysmographic waveform, they also incorporate alarms in case the pulse is
`lost. This increases security for the patient by monitoring more vital signs. but it
`also triggers false alarms, i n particular due to motion artifacts. To avoid this
`problem, Nellcor has developed what they call Oxismart, which, for loss of pulse,
`aims to distinguish between a real clinical condition and a motion artifact. This
`feature is incorporated in the latest models, such as the N-3000.
`Motion artifacts are detected by processing the plethysmographic waveform
`and before validating a pulse, requiring three different steps. Only the signals that
`pass all the steps are used to calculate Sp02 (Nellcor 1995). To differentiate
`between a loss of pulse due to motion artifact from a loss of pulse due to a
`clinical condition, the system assumes that if the pulse is lost, but the patient is
`moving, the patient has pulse and the loss is due to a motion artifact. Figure 12.6
`illustrates this fact. If the pulse oximeter fails to detect at least one pulse in 10 s,
`it enters into pulse search mode. The operator is aware of this situation because
`the PULSE SEARCH indicator lights, and the display alternates between data and
`dashes. In this condition, the pulse oximeter enters an evaluation period of 50 s.
`[f the patient is moving, each time that the pulse oximeter detects a valid pulse,
`readings for heart rate and oxygen saturation are validated. The device returns to
`
`1
`
`225
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01300
`
`

`

`RX-0035.0226
`
`APL_MAS_ITC_00015843
`
`User interface for a pulse oximeter
`
`209
`
`its normal operation after detecting an adequate sequence of validated pulses. If
`during the 50 s evaluation period, an adequate pulse sequence is not detected, a
`low-priority alarm sounds, and there is a visual indication of this condition as
`shown in table 12.1. On the other hand, if the pulse oximeter does not detect
`motion after 60 s in pulse search mode, a high-priority alarm sounds, and there is
`also a visual indication of this condition. With this feature, it is possible to track
`the oxygen saturation even in patients that produce signals of poor quality, and at
`the same time warning can be given of a potentially dangerous condition.
`
`Lost pulse WITH
`continuous motion
`1
`1
`I PULSE SEARCH indicator lights continuously i
`1
`1
`Sp02 and pulse r~te
`1 zeros
`alternate between previous display
`I and dashes
`
`1 Low-priority
`i alarm sounds
`PULSE SEARCH
`flashes
`
`11
`
` Data displays
`
`11
`
`60
`
`Time (s)
`
`MOTION indicator lights continuously
`
`1
`10
`
`16
`
`1
`
`1 1
`
`Lost pulse WITHOUT
`motion
`
`1i
`
` HIGH-priority alarm sounds
`1
`1 PULSE SEARCH 1
`I indicator lights 1
`I continuously
`
`PULSE SEARCH indicator flashes
`
`1 Sp02 and pulse rate 1
`1
`1 alternate between
`previous values and
`1
`1 flashes
`16
`10
`
`Sp02 and pulse rate flashes zeros
`1
`60
`
`Time (s)
`
`0
`
`0
`
`Figure 12.6 Oxismart© alarm detectors used in some Nellcor units to reduce false alarms due to
`motion artifacts (Nellcor 1995).
`
`12.5 COMMUNICATION FUNCTIONS
`
`Communication functions are not a primary function, but an added value feature
`for a pulse oximeter. Communication functions can be found in all types of
`devices, but they provide a great improvement to the units with only numerical
`display, because it gives them the graphical features that otherwise are missing.
`They are used to send data to a printer or plotter. The most common use is to
`print the trend for both oxygen saturation and heart rate for a patient. This
`feature converts the most simple units into units that act like solid state Holter
`
`226
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01300
`
`

`

`RX-0035.0227
`
`APL_MAS_ITC_00015844
`
`210
`
`Design Of pulse oximeters
`
`monitors, with the clinical advantage associated with the knowledge of trend over
`time. There are few units that incorporate an internal printer, normally a thermal
`one, thus eliminating the need for extra connectors and cables.
`The most common method of communication is using the RS-232 protocol.
`It is also possible to obtain analog signals proportional to the plethysmographic
`and pulse rate waveforms. The voltage output is normally selectable between a
`range of 0 to I V dc and I to 10 V dc.
`
`12.6 CABLES AND CONNECTORS
`
`The cables and connectors are used to transmit power and signals between the
`device and the surrounding accessories and power supplies. We can roughly
`distinguish three levels of communication:
`
`1. Interface with the power source.
`2. Interface with the lead and probe.
`3. Interface with auxiliary equipment.
`
`The power connector is used to transfer Ihe energy required from the power
`source to the unit for its operation. The Standard requires that it should be
`designed so that it protects the patient from human errors (ASTM I 992). This
`means that it has to be clearly different from the connectors that will be attached
`to the patient. Power connectors are used to operate the units when it is turned
`on, and to recharge the battery when the unit is turned off.
`The connector for the lead and probe is usually placed on the front panel,
`and it is usually mechanically incompatible between different manufacturers,
`unless they specify that the probe is compatible. For example, Protocol Systems
`advertises that their Propaq models can use probes l'rom Nellcor. The most
`common types of probe connectors are DB9 and DIN. in all cases, the connectors
`are mechanically designed with physical alignment aids and visual indicators to be
`sure Ihat the lead is inserted the correct way into the connector. It is important
`that the connectors be constructed robustly, because the unit can be subjected to
`severe mechanical stress and vibration. Because most of the units can be
`synchronized with the ECG signal, obtained through a separate module, it is
`common to have an ECG connector.
`The auxiliary connectors are normally located on the side or the back panels,
`and they are normally used for communication functions. The most common ones
`are the transfer of digital data to a printer or analog data for further recording
`or to a graphical plotter. For these auxiliary functions almost every manufacturer
`uses their own set of connectors, voltage levels, and conlmunication protocols that
`make them work only with their own peripheral units.
`
`12.7 OTHER FEATURES
`
`Other indications that need to be displayed in a pulse oximeter are those
`regarding the correct labeling of all inputs, outputs, control knobs, and keys.
`Some models of pulse oximeters are manufactured in different levels of electrical
`isolation. For example, Criticare manufactures the unit 504/504US in BF (body
`floating) and CF (cardiac floating) versions. Because they look externally very
`
`1
`
`227
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01300
`
`

`

`RX-0035.0228
`
`APL_MAS_ITC_00015845
`
`User interface for a pulse oximeter
`
`211
`
`similar, if not the same, it is very important to carefully mark its application on
`the front panel to avoid connecting a patient that needs a CF unit to a BF unit.
`For those units that can be operated using an internal rechargeable battery,
`or disposable batteries, it is important to have an external indication of the
`approximate level of charge of the batteries and the remaining operating time, to
`control their replacement. The units from Medical Research Laboratories, Inc.
`display the charge level on an indicator. Most other units display a low-battery
`warning signal.
`
`12.8 COMPLIANCE REQUIREMENTS
`
`The Electromagnetic Compliance (EMC) requirements for electrical equipment
`in general, and biomedical equipment in particular, are changing at a fast pace.
`Because most of the new regulations have long transition periods during which
`they are not mandatory, it is wise to design products for future compliance with
`those regulations. We do not describe the current applicable regulations and
`standards, but describe their existence and probably future evolution.
`The basic idea behind the set of EMC regulations is to ensure the safety of
`operation of electrical equipment during normal circumstances. This means that a
`particular device should not cause harmful interference to other devices and this
`device should not be affected by interference from other devices. Figure 12.7
`illustrates these effects. They can be summarized as conducted emissions, radiated
`emissions, and immunity from interference generated by other equipment that
`can be either radiated or conducted to the device in question (Gerke and Kimmel
`1994a). For the interference generated in the unit, most o f the problems are
`caused by the radiated emissions, because the use of microprocessors running at
`high clock frequencies is becoming more common in medical devices and these
`generate radiated interference.
`
`Radio frequency
`
`interference 1
`
`Electrostatic
`discharges
`
`1
`
`Radlated
`~-r Interference
`
`Medical device -
`
`Microprocessor
`
`Eiectronics
`
`1
`1
`
`Power -
`disturbances
`
`- Conducted
`interference
`
`Figure 12.7 Different sources of disturbances and interferences for EMC purposes.
`
`The ASTM Standard refers to the IEC 601-1 and IEC 801-2 Standards for
`electromagnetic compatibility requirements in pulse oximeters (IEC 1988, 1990).
`The IEC 601-1 Standard describes a general set of requirements for the safety of
`electrical equipment for medical use. The unit only needs to be tested against
`electrostatic discharges (ESD) for its accessible parts, rather than in the interior
`
`228
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01300
`
`

`

`RX-0035.0229
`
`APL_MAS_ITC_00015846
`
`212
`
`Design Of pulse oximeters
`
`of the device. The Standard justifies this procedure based on the fact that pulse
`oximeters are not life-support devices. but vigilance adjuncts. Therefore. the cost
`to provide immunity against ESD in the interior of the system is not justified
`(ISO 9919, annex L). The same Standard, however, serves as a reminder to
`exercise common sense and provides acceptable work procedures for maintenance
`personnel that require them to open the device.
`However, many times the manufacturers try to expand their market by
`exporting their products to other countries. Therefore the designers must be
`aware of the existence of other EMC regulations, which are generally less strict
`in the US and more strict in European, Asian and most other countries. As a rule
`of thumb, the European Economic Community (EEC) countries have more
`regulations and fewer exceptions to those regulations than the US, where most of
`the regulations are voluntary for most of the medical equipment. However,
`medical regulations are undergoing significant changes, and we may expect
`mandatory EMI regulations in the future, regarding ESD, RF fields and power
`disturbances, driven by the Food and Drug Administration (FDA) and the
`regulations in the EEC. At the present time, there are no mandatory regulations
`in the US, as medical devices are exempted from Federal Communications
`Commission (FCC) emission regulations, and they are covered only by voluntary
`susceptibility requirements. On the other hand, in the EEC countries, the
`equipment is required to be tested for emissions but not for immunity (Gerke and
`Kimmel 1994b). This situation is expected to change soon, and in the future we
`may expect mandatory regulations for RFI, ESD, and power disturbances in the
`US. Because of the need to be competitive in international markets, designers
`should consider that the best way to avoid unnecessary delays, and to lower the
`economic impact of changing a design, is to design for compliance from the first
`stages, without overdesign that implies an increment of cost with no additional
`value.
`
`REFERENCES
`ASTM \991 Standard Specijication for Pulse Oximeters, F!-1!5-1992 (Philade\phia, PA:
`American Society for Testing und Matenals)
`10% \991 Pulse oximeters for medical use-Requirements, ISO 99]9:!992 (E) CGeneva.·
`International Organi

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket