throbber
RX-0035.0181
`
`APL_MAS_ITC_00015798
`
`ill 164
`
`1
`
`Design of pulse oximeters
`
`10.2.1 Simulators using blood
`
`Several simulators have been proposed that need whole blood to teSt the
`functionality of the pulse oximeter These simulators are all based on the concern
`of being able to simulate the absorbance of human tissue (normally tile finger)
`between the LEDs and the photodiode of the pulse oximeter under test. Since few
`substances have been found that simulate the optical properties of blood, these
`types of systems typically provide the most accurate simulation.
`10. 2. 1 . 1 Reynolds system. The system described in section 10 . 1
`.2 function.
`equally well as a simulator to test the functionality of pulse oximeters. In fact this
`system has been used to compare ten commercially available oximeters (Reynolds
`et at 1992), and has been used to evaluate the effects of dyshemoglobins on pulse
`oximeter accuracy (Reynolds et al 1993a,b), However, this in vitro test system iN
`not practical in a hospital setting where most pulse oximeters are used. The
`system requires a laboratory setting, is not portable, uses oxygenated whole blood
`and needs a CO-oximeter for comparison. However. this instrument is generally
`considered the gold standard for calibrating and testing a pulse oximeter over its
`complete range.
`10. 2. 1. 2 Vellfors system. The Vegfors system is similar to the Reynolds system
`but with a focus on the artificial finger or 'finger phantom' used. Vegfors et d
`( 1993) describe a system where their artificial fingers consist of silicone rubber
`tubes inserted in plastic Delrin cubes. The tubing system chosen was based on its
`characteristics of tubing diameter, wall elasticity, and blood flow velocity to
`simulate normal physiological characteristics of blood in motion, Delrin was used
`because it has similar optical scattering properties to human tissue. Figure ] O.3
`shows three models. Two different finger models, one with one tube and anothei
`with five tubes were tested along with a third artificial finger consisting of 15
`silicone rubber tubes mounted in silicone rubber in the form of glue. The „bject
`was to develop an optical model which simulated the arterial bed of the human
`finger containing blood vessels and surrounding tissue. The results of these
`different finger configurations detennined that physical dimensions of the
`artificial bed are of minor significance for pulse oximeter readings.
`10.2. 1 . 3 Single wedge system . Several other
`less complicated simulators using
`whole blood have been proposed. In one system, proposed by Yount { 1989), a
`light-absorbing wedge shaped vessel containing blood of known oxygen saturation
`level is placed in the pulse oximeter's optical path. If the wedge (figure I 0.4) is
`moved repetitively back and forth perpendicular to this optical path, either
`manually or with the aid of a mechanical device, both the pulse rate and shape of
`the pulse can be altered. Pulse rate can be simulated by changing the frequency at
`which the wedge is moved across the optical path. The shape of the pulse can be
`changed by altering the speed at which the wedge is moving.
`
`1 0.2.1.4 Dual wedge system. In another arrangement o f the system, two wedges
`are used. One is filled with 10093 oxygen saturated blood and the other with
`completely unsaturated blood. The wedges are placed as shown in figure 10.5 und
`by varying the position along the optical path of this arrangement, virtually any
`saturation level can be obtained. Note however that with this second arrangement,
`
`181
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0182
`
`APL_MAS_ITC_00015799
`
`1
`
`1
`
`11
`
`Calibration
`
`165
`
`an additional external device is needed to obtain a pulsatile variation in the
`simulator. figure 10.6 shows the polarization filter system proposed by Yount
`(1989) to achieve this pulsatile variation needed. A pair of polarizing disks
`simulate the changes in transmittance expected by the pulse oximeter. A stepper
`motor controls the motion of one disk. This changes the angle of polarization
`between the two disks, and therefore the amount of light transmitted. By varying
`the rate of angle change, this system can simulate both the shape and pulse rate
`seen by the pulse oximeter. This particular system also has several glass windows.
`This allows for multiple samples to be loaded on the same disk so different
`oxygen saturation levels can be simulated by rotating the appropriate sample into
`the probe.
`
`, .
`
` 2
`
`Blood flowIl i
`
`.f
`
`Blood flow /*'
`
`0,0,
`
`Blood flow ~-1~
`
`Figure 10.3 Block diagram of various artificial fingers as proposed by Vegfors et at (1993).
`
`One limitation of these wedge systems is that i f blood is used as the medium
`in the wedge, the samples either need to be prepared shortly before use or steps
`need to be taken to stabilize the blood.
`
`182
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0183
`
`APL_MAS_ITC_00015800
`
`166
`
`Design of pulse oximeters
`
`Blood
`
`Light path
`
`'4 Wedge movement
`%
`22*GRY~hlj/
`3
`/5
`'~ ? ilitli ftilll/il/t
`-4/Wi7 Tl·41 I i- ' . . . . i ' :ie
`l
`
`Constant absorber
`
`Figure 10.4 Block diagram of a wedge system as proposed by Yount ( I 989).
`
`0% oxygenated blood
`
`Light path
`
`.7-r-9»»~-w~"--ip--*-44
`
`7%~ 1 -
`
`Constant absorber
`
`100% oxygenated blood
`
`1
`
`Figure 10.5 Block diagram of the dual wedge system as proposed by Yount (1989),
`
`10. 2. 1 . 5 Bulb device. Volgyesi
`( 1989) proposed a simple mechanical design to
`simulate a pulsing finger. Figure 10.7 shows the tube and bulb type device. It
`requires a 0.5 to I mL blood sample for each saturation level to be tested. A
`piece of silicone rubber tubing is placed inside a disposable plastic test tube which
`contains a blood specimen. The operator then manually squeezes the bulb at
`regular intervals which causes the silicone rubber tubing and the blood in the
`annular space between the silicone rubber tubing and the test tube to deform or
`pulse. Samples o f heparinized blood are externally altered to different saturation
`levels so different levels of oxygen saturation can be tested. With a variety of
`oxygen saturation level samples prepared in individual test tubes, the pulse
`oximeter can be applied to the device. After the operator is able to rhythmically
`squeeze the bulb for a consistent plethysmograph (rate and amplitude), a reading
`is recorded from the pulse oximeter and the sample is sent to a CO-oximeter for
`a comparison reading. The main advantage of this system is its simple
`implementation. The disadvantage is that the pulsatile nature of the system is
`operator dependent and samples of known oxygen saturation levels of blood need
`to be prepared.
`
`183
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0184
`
`APL_MAS_ITC_00015801
`
`Drobe
`
`Support
`Motor
`
`CO
`
`Calibration
`
`167
`
`LEDs
`
`Samples
`
`rn I
`1 1'
`11
`
`Side view
`
`- Metal outer disk
`Fixed polarized filter
`
`Rotating polarized filter
`
`Metal outer disk
`
`Glass window in disk
`
`9
`
`Top view of metal disk
`
`Figure 10.6 Schematic diagram of polarization system (adapted from Yount 1989).
`
`Tubing
`
`Air
`
`2 of
`
`Bulb
`
`- Plastic cap
`
`Silicone
`» 3
`rubber 2~
`tubing
`
`~> Annular spacers
`
`1- lti'
`
`Blood
`
`/ 12*&5/
`
`Plastic test tube
`
`Pulse oximeter probe
`
`Silicone rubber plug
`
`Figure 10.7 Block diagram of tube and bulb device.
`
`184
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0185
`
`APL_MAS_ITC_00015802
`
`168
`
`Design of pulse oximeters
`
`1 0.2.2 Nonblood simulators
`
`Nonblond simulators. like simulators that use blood, are also based on the concept
`of being able to simulate the absorbance of human tissue (normally the finger)
`between the LEDs ancl the photodiode of the pulse oximeter under test. These
`devices use colored materials to simulate blood. These simulators use a variety of
`mechanical and electrical devices to achieve the desired variations in absorbance.
`The more difficult aspect is simulating the scattering properties of whole blood.
`One of the most successful studies in this area (Marble et al 1994) used a
`combination of nondairy creamer mixed with solutions of red and green dye.
`
`10. 2. 2. ] Billb device. The bulb device described in section 10 . 2 . I above can also
`be used with liquids having differing optical absorbance properties corresponding
`to oxyhemoglobin. A commercial version of this device is currently being
`marketed by Nonin under the trade name finger phantom. This product
`INonin
`1995 ) pr{,vides three translucent white arti.ficial fingers that simulate ariel-ial
`blood at naminally 80%. 90%, and 97% saturation levels. The operator gently
`presses the finger phantom about once every second to generate a pulse. The
`typical infrared percent modulation when squeezed is 0 to 5%.
`10. 2. 2. 2 Wedge device. The wedge device described in section 10 . 2 . 1 above can
`also be used with liquids other than blood having optical absorbance properties
`corresponding to those of the human finger.
`10. 2. 2. 3 Polvester resin device. Figure 10. 8 shows a simple test object proposed
`by Munley er al ( 1989). This device consists of a piece of polyester resin that is
`f<,rmed in the shape of a finger. The resin is adapted to allow a core to be placed
`inside the artificial finger. At the end of the core, in the area exposed to the pulse
`oximeter LED's light path, a slotted piece of suitably colored Plexiglas is placed.
`As the device handle is rotated, the slot allows varying levels of LED light to
`reach the pulse oximeter photodiode. Speed of rotation of the crank will
`determine the pulse rate that
`the oximeter
`reads. Changing the color
`characteristics of Plexiglas will change the oxygen saturation reading that the
`pulse oximeter registers. This device was also shown to produce similar oxygen
`saturation readings among multiple devices of the same make and model of pulse
`oximeter.
`10.2, 2. 4 Colored colloid simulator. Leuthner ( 1994 ) proposed the pulse oximeter
`development system shown in figure 10.9. A transparent bag is filled with a
`colored colloid solution. The color determines the extinction coefficients at the
`two wavelengths of interest. This system uses a water-gelatin mix which is heated
`and colored with red and black ink. To simulate different oxygen saturation
`levels, multiple bags with varying ratios of red and black dye need to be
`prepared. The bag is positioned between two acrylic disks. The disks and bag are
`then rotated by a stepper motor under microcontroller control. With this
`configuration, both the DC and AC absorbance ratio can be adjusted. Increasing
`the angle between the two plates increases change of absorbance over each
`rotation for an increase in relative AC signal. The simulated pulse shape is
`determined by speed of the disk rotation and the pulse rate is determined by the
`rotation frequency. A constant absorber material is placed on top of the disk to
`simulate the constant light absorbance of fingers of different people. In practice,
`
`185
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0186
`
`APL_MAS_ITC_00015803
`
`Calibration
`
`169
`
`it can vary by a factor of four. Generally a piece of white paper of varying
`thickness is used as the constant absorber. Two optic fibers are integrated into an
`artificial finger which then plugs into the finger probe of the pulse oximeter. The
`other ends of fibers are connected opposite each other near the rotating plates. I f
`testing is done using different waveforms, the angular velocity of the rotation has
`to change and as such is controlled through the stepper motor via microcontroller
`control. The whole system is enclosed in a box to prevent disturbances from
`ambient light.
`
`1
`
`Plexiglas
`
`r\«\IN
`
`/1
`
`Handle
`
`11
`
`Poiyester resin finger
`
`Figure 10.8 Polyester resin system proposed by Munley et al (1989).
`
`The physical behavior of this system can be almost totally described using
`Beer's law, but the system cannot be used for finding the calibration table of a
`pulse oximeter. The main reason is that the scattering effect in whole blood is not
`present in this system. However this system can be used for a rough calibration
`table of a new instrument and to test an existing pulse oximeter for the response it
`gives when different colored bags are used.
`
`Screws to adjust
`AC/DC absorbance ratio
`
`l
`
`Probe clip
`
`Photodiode I
`
`t ~==rzccc:A>'
`
`..,1
`1
`
`LEDs
`
`Constant absorber
`
`Plastic bag
`with absorbing -
`colloid
`
`Acrylic disk
`
`l
`
`Stepper motor
`
`1 11
`Ef
`
`PI
`
`Figure 10.9 Leuthner's (1994) colored colloid disk system.
`
`186
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0187
`
`APL_MAS_ITC_00015804
`
`Design of pulse oximeters
`
`170
`10. 2.2.5 Liquid crystal retarder simulator. Zhou et al ( 1992 ) developed a device
`for generating test signals for pulse oximeters based on a voltage-controlled
`liquid-crystal light valve. In the first system, the pulse oximeter's LEDs are
`separated by an optical filter, modulated by a light valve. and recombined before
`detection by the probe's photodiode. The newer system does not require
`wavelength separation and its associated hardware as shown in figure 10.10, The
`transmittance characteristics are varied by taking advantage of the intrinsic
`wavelength dependence of a twisted-nematic liquid-crystal retarder (LCR).
`Polarizers are used to generate optical density variations that can be made to
`resemble blood perfused tissue. The intensity transmitted through the optical
`system can be adjusted by varying the voltage on the LCR. To simulate a pulsatile
`change in transmittance, the attenuation is initially made a constant DC value. A
`small AC voltage is then superimposed on top of the DC voltage to provide a
`pulsatile component. The transmittance at both the red and IR wavelengths varies
`depending on the voltage amplitude applied to the LCR. This allows the AC/DC
`ratios to be controlled by adjusting the amplitude of the voltage applied to the
`LCR. The polarizers are required because the angle of polarization strongly
`affects the range of variation of the red/IR ratio and its sensitivity to the applied
`voltage. Zhou et at are continuing work on this concept to provide the capability
`of simulating the shape of the plethysmographic waveform applied to the LCR.
`
`AC Voltage i DC Voltage
`
`Light path ~ ~ ~U
`
`UFixed polarizing filter
`
`Rotated polarizing filter
`
`Figure 10. 10 Diagram of the liquid crystal retarder (LCR) system proposed by Zhou et d
`(1992)
`tisSILe model, A device based on the same general principles as
`10.2. 2. 6 Aovagi
`the wedge system been proposed by Aoyagi etal ( 1994). Figure 1 0.1 I shows that
`a static tissue model having absorption characteristics similar to a human finger is
`inserted into a pulse oximeter probe. A blood model having blood absorption
`characteristics similar to a specified oxygen saturation level is moved within the
`tissue model to simulate pulsatile motion and pulse rate. By altering the geometry
`of the blood model and/or the rate of motion of the blood model in and out of the
`tissue model, both the pulsatile waveform and pulse rate can be simulated.
`10.2. 2.7 Optoelectronic device. A number of
`relatively simple easy-to-use
`simulators have begun to appear on the market based on optoelectronic
`
`1
`
`1
`
`187
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0188
`
`APL_MAS_ITC_00015805
`
`Calibration
`
`171
`
`principles. Figure 10.12 shows a block diagram for one of these types of
`simulators. First, the user selects the parameter(s) to be simulated. The pulse
`oximeter probe is then attached to the device and a signal is received from the
`pulse oximeter probe's LEDs by the simulator. Pulse separator and timer
`circuitry convert the red and infrared light pulses from the pulse oximeter probe
`into electric signals. These signals are modulated with the appropriate level of
`AC/DC ratio (under computer control) and then converted back to light pulses,
`via the LED bar, to the probe's photodiode. Finally, the pulse oximeter responds
`to the converted light pulses as it would to light pulses modulated by living tissue.
`LEDs
`
`Pulse oximeler probe
`
`Blood model -_-1_-1 NO' 4*-f.*'.64
`
`11I
`
`Photodede
`
`~%-2177~1//1 1/ll//1 ,
`
`Tissue model
`
`1
`
`Figure 10.11 Block diagram of system as proposed by Aoyagi et al (1994).
`
`These systems can test the probe and oximeter over the complete specified
`range of the oximeter. Also, simulation of a wide range of conditions is possible.
`The modulated signal can vary plethysmographic amplitude and wave shape to
`simulate a variety of ambient light conditions, motion arti facts, and arrhythmias.
`At least one system (Clinical Dynamics 1995) also includes a probe analyzer
`capability which independently tests LED and photodiode continuity and
`sensitivity. These types of simulators are primarily used by pulse oximeter
`manufacturers during final assembly and checkout of their products. In addition,
`their capability to generate automatic test sequences help document JCAHO (Joint
`Commission on Accreditation of Healthcare Organizations) testing requirements.
`
`10.2.3 Electronic simulators
`
`Electronic simulators have limited usefulness since they only simulate electronic
`signals to and from the probe. Usually these relatively simple devices are
`provided by the pulse oximeter manufacturer and only check a small number of
`values. These devices typically plug into the probe port on the pulse oximeter and
`
`188
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0189
`
`APL_MAS_ITC_00015806
`
`172
`
`Design of pulse oximeters
`
`use the drive current of the probe LEDs to generate a simulated photodiode
`signal back to the pulse oximeter using the device. Figure I (). 13 shows an
`example of such a device. In remote mode. the LEDs just drive an amplifier and
`the output shows up on the sensor output. This is useful for simple continuity
`testing. In local mode, these devices are able to electronically simulate a discrete
`number of simulated oxygen saturation levels, pulse rates and plethysmographic
`waveform strengths. In addition the calibration resistor value reading capability
`of the pulse oximeter can be checked. These simulators are good for functional
`checks of the pulse oximeter's internal circuitry, but because they bypass ihe
`pulse oximeter's probe, are of limited usefulness.
`
`>X\LEDs©4
`
`1
`Pulse separator
`and timer
`1
`RED
`IR
`1
`r-_-- DC multip'lers I
`1
`Conputer control
`1
`L--74 AC multipliers
`1
`1
`44,4- - .
`
`Userinput
`variables
`
`Pulse oximeter
`components
`
`0 Simulator
`
`components
`
`IR Switch
`
`i\\\\\\\I
`\Photodio{le
`\\\\\\:\
`1 Photodiode
`
`1
`
`R + IR
`Timing
`
`Amplifier --- LED BAR
`
`Figure 10.12 Block diagram electro-optic simulator system developed by Merrick and Haas
`(1994).
`
`LED output monitor
`
`10.3 STANDARDS
`
`Although the pulse oximeter has been on the market since 1977 (Santamaria and
`Williams 1994), surprisingly little standardization has been documented to this
`point. Statements like 'machines and probes are interchangeable with less than
`0.5% difference', 'warm-up time factor of 0.5% to 1.0%' and 'the low perfusion
`light on the Ohmeda oximeter indicates the oximeter's microprocessor has low
`confidence level in the data' can be found in the literature. Several standards do
`exist, but their value from the designer's point of view is limited at best.
`
`189
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0190
`
`APL_MAS_ITC_00015807
`
`Calibration 173
`
`0%
`
`Local
`
`Remote
`
`1
`
`DC Level
`
`AC Level
`
`Pulse
`Conditioning
`
`Pulse
`Generator
`
`Sensor
`Output
`
`LED drive
`inputs from
`oximeter
`n
`
`I Amplifier
`
`L
`
`- Power --- +V
`Conversior
`
`CC
`
`Calibration
`Resistor
`
`Resistor input
`
`Figure 10.13 Block diagram of an electronic simulator that replaces the pulse oximeter probe 1
`(used with permission (Nellcor 1994) Pulse oximeter tester Model SRC-2).
`
`10.3.1 ASTM F1415
`
`The ASTM F1415 standard (ASTM 1992) contains requirements for the pulse
`oximeter designer in regard to marking and documenting the system, electrical
`safety concerns, electromagnetic interference and alarms. No specific information
`is provided regarding specific design requirements of the parts of the system
`discussed in the preceding chapters. In addition, no specific information is
`provided in regard to calibration or testing of these devices.
`
`10.3.2 ISO 9919
`
`This standard mentions a few requirements regarding calibration. These include
`requiring manufacturers to provide:
`
`1. The calibration range of the pulse oximeter.
`2. Whether the pulse oximeter is calibrated to display functional or fractional
`saturation.
`3. The accuracy and range of HbO2 saturation level displayed.
`4. Whether the calibration was functional or fractional saturation.
`5. Test methods for calibration need to be available from manufacturer upon
`request.
`
`The ISO 9919 (International Organization for Standardization 1992) also offers
`this disclaimer in Annex L:
`
`Values derived from the pulse oximeter are not a measurement of blood or tissue oxygen
`tension and therefore pulse oximetry provides no direct indication of oxygen delivery to or
`
`consumption by, tissues. At present there is no widely accepted direct in vitro calibration
`
`1
`
`190
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0191
`
`APL_MAS_ITC_00015808
`
`1
`
`It
`
`174
`
`Design of pulse oximeters
`
`method for pulse oximeters. The only accepted in viiro test method for correlation of the
`reading from a pulse oximeter (Sp02) is bench-type oximetry employing more than two
`wavelengths of light or other methods using blood samples drawn from human subjects
`
`Although work ts progressing on the development of direct in \,it,·o calibration methods ,
`
`present techniques still require the use of human subjects. To include test methods in
`standards that require the use of human subjects, has, through past experience, been found to
`be unacceptable, and therefore in vit'o test methods are not included in this International
`Standard.
`
`10.3.3 Other smndards
`
`American Society of Anesthesiologists. Standards for Basic Intra-Operative
`Monitoring, 1986 (0696-ASA).
`American Society of Anesthesiologists. Standards for Post-Anesthesia Care, 1989
`(0697-ASA).
`European Committee for Standardization. Drafting European norm for pulse
`oximeters.
`
`REFERENCES
`
`Ackerman S W and Weith P 1995 Knowing your pulse oximetry monitors Med. Electron. 26 ( 1 )
`82--6
`ASTM 1992 Standard Specification for Pulse Oximeters F/415-92 (Philadelphia PA: American
`Society for Testing and Materials)
`Aoyagi T, Fuse M, Shindo Y and Keto M 1994 Apparatus for calibrating pulse oximeters US
`patent 5,278,627
`Cheung PW, Gauglitz KF, Hunsaker SW, Prosser SJ, Wagner DO and Smith RE1993
`Apparatus for the automaticcalibration ofsignalsemployed in oximetry USpatria 5.259,381
`Clinical Dynamics 1995 Technical sales brochm·e (Wallingford, CT: Clinical Dynamics)
`International Organization for Standardization 1992 Pulse Oximeters for Medical Use-
`Requiremenfs ISO9919: 1992(E)
`Lcuthner T 1994 Development system for pulse oximetry Med. Biol. Eng. CompuL 32 596-8
`in vitro
`Marble D R, Bums D H and Cheung P W 1994 Diffusion-based model of pulse oximetry :
`and in vivo comparison Appl Op' 33 1279-85
`Merrick E B and Haas P 1994 Simulation for pulse oximeter US Patent 5, 348,005
`Munley A J , Sik M J and Shaw A 1989 A test object for assessing pulse oximeters Lancet 1048-9
`Nellcor 1994 Pulse Oximeter Tester Model SRC-2 (Pleasanton. CA: Nellcor)
`Nonin Medical 1995 Nonin finger phantom Technical Nok (Plymouth , MN : Nonin Medical)
`Moyle JT B 1994 Pulse Oximelry ( London: BMG)
`Payne J P and Severinghaus J W ( eds) 1986 Pulse Oximetry (New York : Springer)
`Pologe J A 1989 Functional saturation versus fractional saturation: what does the pulse oximeter
`read J. Ck Monit. 5 288-9
`Reynolds KJ, del<ock JP, Tarssenko L and Moyle JTB 1991 Temperature dependence of LED
`and its theoretical effect on pulse oximetry Brit. J. Anaesthesiot. 67 638-43
`Reynolds KJ,Moyle J TB, Gale LB, Sykes MK and Hahn CEW 1992 In vitro performance
`test system for pulse oximeters Med. Biol. Eng. Comput. 30 629-35
`Reynolds KJ, Moyle JTB. Sykes MK and Hahn CEW 1993a Responses of 10 pulse oximeters
`to an in vitro test system Brit. J. Anaestliesiot. 6% 265-9
`Reynolds KJ, Palayiwa E, Moyle JTB, Sykes MK and Hahn CEW 1993b The effects of
`dyshacmoglobins on pulse oximetry J Clin. Monit. 9 81-90
`Santamaria T and Williams J S 1994 Device focus: pulse oximetry Med. Device Res. Rep. 1 ( 2) 8-
`I 0
`Severinghaus J W, Naifeh K H and Koh S O 1989 Errors in 14 pulse oximeters during profound
`hypoxia J. Clin. Monit. 5 72-81
`Vegfors M, Lindberg L G, Oberg P A and Lennmarken C 1993 Accuracy of pulse oximetry al
`various haematocrits and during hacmolysis in and in vitromodel Med. Biol Eng. Comput. 3 1
`135-41
`
`191
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0192
`
`APL_MAS_ITC_00015809
`
`Calibration
`
`175
`
`Volgyesi G A 1992 Method of testing the accuracy of pulse oximeters and device therefor US
`patent 5,166,517
`Wukitsch M W, Petterson M T, Tobler D R and Pologe J A 1988 Pulse oximetry: analysis of
`theory , technology, and practice J Clin. Monit. 4 290-301
`Yount J E 1989 Device and procedures for in vitro calibration of pulse oximetry monitors US
`patent 4,834,532
`Zhou GX, Schmitt JM and Walker E C 1992 Electro-optical simulator for pulse oximeters Med.
`Biol. Eng. Comput. 31 534-9
`
`INSTRUCTIONAL OBJECTIVES
`
`10.1 Describe how R curves are determined through in vivo testing.
`10.2 Explain the role that LED temperature plays in oxygen saturation level determination.
`10.3 Explain why the term Sp02 is necessary when referring to oxygen saturation levels.
`10.4 Explain the reason why different R curves may be needed for a manufacturer's pulse
`oximeter system.
`10.5 Describe how oxygen saturation level is altered through an in vitro test system.
`10.6 Explain why pulse oximeters are less accurate for SPO2 saturation levels below 60%.
`10.7 Describe the operation of an optoelectronic simulator system.
`10.8 Describe the operation of an colored colloid simulator system.
`10.9 Describe the operation of polyester resin device simulator system.
`10.10 Describe the operation of a wedge simulator system.
`10.11 Describe the operation of the tube and bulb simulator system.
`10.12 Explain the limitations of the electronic simulators used for testing pulse oximeters.
`
`t
`
`1
`
`192
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0193
`
`APL_MAS_ITC_00015810
`
`1
`
`CHAPTER 11
`
`ACCURACY AND ERRORS
`Supan Tung/itkusolmun
`
`Continuous assessment of arterial oxygen saturation (Sa'2) is important in
`clinical management of critically ill patients. Pulse oximeters have been widely
`used as blood oxygen monitoring devices since the early 1980.. Currently, pulse
`oximeters can be found in virtually every operating room, recovery room, and
`intensive care unit. The advantages of pulse oximetry include noninvasiveness,
`ease of use, portability, and patient comfort. A light source generated by two
`LEDs, with wavelengths at approximately 660 nm and 940 nm, and a photodiode
`are mounted in a probe of a pulse oximeter. Circuit control. saturation
`calculation, and display are managed by a microprocessor instrument as described
`in chapter 8. Unlike earlier techniques such as the in vive eight-wavelength
`oximeter (chapter 3), no heating or arterialization techniques are required in
`pulse oximetry.
`All pulse oximeters work using absorption spectrophotometry, however,
`considerable differences exist in the way different manufacturers obtain and
`process the data. These differences occur in the light-emitting diodes, sampling
`frequency, microprocessor algorithms. and the constants used in the calculations,
`or the look-up tables. Since the technique has come into wide clinical use over the
`past decade, it is important to examine circumstances where its reliability may be
`questioned. The objective of this chapter is to describe several sources of error in
`pulse oximetry which may cause hazardous consequences to the patients.
`Recognizing the limitations described in this chapter and applying appropriate
`corrective interventions are essential to optimize the clinical use of pulse
`oximeters.
`
`11.1 EVALUATION OF PULSE OXIMETERS
`
`The gold standard measurement ot arterial oxygen saturation is the CO-oximeter ,
`described in chapter 3. A comparison of the pulse oximeters' readings and CO-
`oximeters' readings is thus required to verify the reliability of the pulse oximetry
`technique. Comparisons between pulse oximeters' arterial oxygen saturation
`values and the CO-oximeters' readings, as well as the HP eight-wavelength ear
`oximeter will be discussed in this section.
`
`176
`
`193
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0194
`
`APL_MAS_ITC_00015811
`
`r
`
`1
`
`Accuracy and errors
`
`177
`
`11.1.1 Accuracy, bias, precision, and confidence limit
`
`Accuracy is a measure of systemic error or bias; the greater the error, the less
`accurate the variable. The accuracy of a measurement is the degree to which it
`actually represents what it is intended to represent. The location of the mean
`errors reflects the accuracy of the measurement. The accuracy of pulse ox imeter
`oxygen saturations can usually be tested by comparing with the reference
`techique, CO-oximetry. Parameters frequently used to represent the degree of
`accuracy are bias, and absolute mean errors. Bias, in this case, is defined as the
`mean of the differences between the pulse oximeter readings and the CO-
`oximeter readings, which can be expressed as
`
`NI xi
`i=1
`bias =-=.F
`N
`
`(11.1)
`
`where xi is calculated by subtracting the ith CO-oximeter measurement from the
`corresponding oximeter saturation displayed by a pulse oximeter. N is the total
`number of measurements. Units are percent saturation.
`Precision is a measure of variation of random error, or degree of
`reproducibility. The dispersion of points around the mean reflects the precision
`of the measurement. Precision is often described statistically using the standard
`deviation (SD) of the differences between the pulse oximeter readings and the
`CO-oximeter readings of repeated measurements (Nickerson et al 1988) as in
`equation (11.2). Units are percent saturation.
`
`1N
`9
`12(xi -x)-
`precision = SD = 11 i=l
`1 N-1
`Some researchers frequently use a 95% confidence limit, which for a normal
`distribution is equal to 1.96 times SD:
`
`(11.2)
`
`95% confidence limit = 1.96 x SD = 2 x SD.
`
`(11.3)
`
`Example l
`
`The results from an experiment to compare pulse oximeter and CO-oximeter
`readings are shown in table 11.1. Ten measurements were made.
`
`From table 11.1,
`
`10
`24i=1
`15---1.5%
`bias =x=--10
`10
`
`194
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0195
`
`APL_MAS_ITC_00015812
`
`178
`
`Design of pulse oximeters
`
`I 0I (ri - i)2
`i=I
`
`10-1
`
`&0.5
`= 1-= 1.51%,
`N 9
`
`precision = li
`
`Table 11.1 Comparison of pulse oximeter and CO-oximeter readings.
`
`Measurement CO-oximeter Pulse oximeter
`(i)
`readings (%) readings (g,)
`9 7
`100
`98
`99
`9
`3
`92
`9I
`98
`96
`4
`99
`97
`5
`93
`90
`6
`90
`89
`1
`98
`95
`8
`90
`9
`88
`93
`92
`10
`
`3
`1
`-1
`2
`2
`3
`1
`3
`9
`
`X, (56) .ri - .7C
`
`1.5
`-0.5
`-2.5
`0.5
`0.5
`1.5
`-0.5
`1.5
`05
`-2.5
`
`and
`
`95% confidence limit =2x 1.5 1% = 3.02%.
`
`The bias of 1.5% means that the test pulse oximeter tends to overestimate the
`oxygen saturation level (postive bias). A 95% confidence limit of 3.02% means
`thal the pulse oximeter will give an outcome in the range between 1.5 - 3.02%
`and 1.5% + 3.02%. or between -1.52% and 4.52% from the true value (the CO-
`oximeter reading) with a probability of 0.95.
`The use of bias and precision is helpful in getting a clear picture of a pulse
`oximeter's performance and how thiE compares to other units or other studies. A
`unit may be very precise. so that the results are highly reproducible with a low
`scatter, but have a high bias so that the results are not centered on the true values,
`In contrast, a unit may have a very low bias, but have poor precision. with values
`swinging widely from side to side of the true value. In clinical practice, a 95%
`confidence I imit o f less than *3% is considered acceptable for most cases.
`Other statistical terms from the regression analysis (correlation coefficient,
`positive error, intercept, and slope) are also used in several studies c Yelderman
`and New 1983, Taylor and Whitwam 1988).
`
`11.1.2 What do pulse oximeters really measure?
`
`Pulse oximeters only measure a ratio of transmitted red and infrared light
`intensities, and relate this to a look-up table of empirical oxygen saturation values
`(see chapter 9). The values in the table depend on the manufacturer's purpose of
`estimating functional or fractional oxygen saturation, but will in reality be
`neither of these unless the dyshemoglobin tdysfunctional hemoglohin) levels, and
`the pH levels in a subject's arterial blood are exactly the same as the average
`values of those used in the empirical calibration to create the look-up table. Choe
`er al ( 1989) found that the measured oxygen saturations in two instruments
`(Ohmeda Biox 3700 and Radiometer Pulse Oximet

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket