throbber
IPR2020-01649
`Paper No. 18
`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`
`____________
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`____________
`
`EDWARDS LIFESCIENCES CORPORATION and
`EDWARDS LIFESCIENCES LLC,
`Petitioners,
`
`v.
`
`COLIBRI HEART VALVE LLC,
`Patent Owner.
`
`____________
`
`Case IPR2020-01649
`Patent 9,125,739
`____________
`
`Before the Honorable ERICA A. FRANKLIN, JAMES A. TARTAL,
`and ERIC C. JESCHKE,
`Administrative Patent Judges.
`
`
`PETITIONERS’ REPLY
`
`IN SUPPORT OF INTER PARTES REVIEW OF
`
`U.S. PATENT NO. 9,125,739
`
`
`
`
`
`
`
`
`
`

`

`TABLE OF CONTENTS
`
`IPR2020-01649
`Paper No. 18
`
`Page
`
`I.
`
`GROUNDS 2-5 OF THE PETITION EACH RENDER CLAIMS 1-5
`OF THE ’739 UNPATENTABLE .................................................................. 1
`
`A. Grounds 2-5: Bessler Teaches Nearly All Claimed Limitations
`of the ’739 and Would Be Relied Upon by a POSITA as a Base
`Reference in an Obviousness Analysis ................................................. 1
`
`B. Grounds 2-5: It Would Have Been Obvious to Modify Bessler
`to Include a Trumpet-Like Configuration in View of the Teachings of
`Leonhardt and Teitelbaum .................................................................... 4
`
`C. Grounds 2-5: Under Any Interpretation of “Onto the Pusher
`Member,” It Would Have Been Obvious to Modify the Delivery
`System of Bessler to Include a Pusher Member as Claimed ................ 8
`
`D. Grounds 4-5: Claim 5’s “Controlled Release Mechanism”
`Limitation is Obvious .......................................................................... 15
`
`II. GROUND 1: PANIAGUA IS PRIOR ART AND ANTICIPATES
`CLAIMS 1-5 .................................................................................................. 17
`
`In the ’739, “the leaflet portion of the valve itself” must be
`A.
`“formed from a single piece of biocompatible material” in order to
`attain “the novel and improved qualities of the present invention” .... 17
`
`III. CLAIM CONSTRUCTION .......................................................................... 19
`
`“wherein the prosthetic heart valve is collapsed onto the pusher
`A.
`member” .............................................................................................. 19
`
`“flares at both ends in a trumpet-like configuration” & “valve
`B.
`means” ................................................................................................. 23
`
`IV.
`
`IN JANUARY 2002, PETITIONERS’ EXPERT, DR. GOLDBERG,
`WAS A PERSON OF ORDINARY SKILL IN THE ART; PO’S
`EXPERT, DR. DASI, WAS NOT ................................................................. 24
`
`
`
`ii
`
`

`

`IPR2020-01649
`Paper No. 18
`V. A DECISION BY THE BOARD IS CONSTITUTIONALLY VALID ....... 27
`
`VI. CONCLUSION .............................................................................................. 27
`
`
`
`
`
`
`
`
`iii
`
`

`

`IPR2020-01649
`Paper No. 18
`
`TABLE OF AUTHORITIES
`
`
`Cases
`Apogee Lighting Holdings, LLC v. Autronic Plastics, Inc.,
`IPR2020-01024, Paper 11 (PTAB Dec. 1, 2020) ............................................... 25
`Chung v. Bed Bath & Beyond, Inc.,
`No. EP-09-330-FM, 2011 WL 6967991 (W.D. Tex. Dec. 20, 2011) .......... 21, 22
`ClassCo, Inc. v. Apple, Inc.,
`838 F.3d 1214 (Fed. Cir. 2016) .......................................................................... 16
`Edwards Lifesciences Corp. v. Boston Scientific SciMed, Inc.,
`IPR2017-00060, Paper 64 (PTAB Mar. 23, 2018) ............................................. 24
`In re Antor Media Corp.,
`689 F.3d 1282 (Fed. Cir. 2012) ............................................................................ 7
`In re Etter,
`756 F.2d 852 (Fed. Cir. 1985) ............................................................................ 15
`In re Mouttet,
`686 F.3d 1322 (Fed. Cir. 2012) ............................................................................ 2
`In re Nordt Dev. Co.,
`881 F.3d 1371 (Fed. Cir. 2018) .......................................................................... 23
`Infineum USA L.P. v. Chevron Oronite Co.,
`844 F. App’x 297 (Fed. Cir. 2021) ....................................................................... 2
`Microsoft Corp. v. Parallel Networks Licensing, LLC,
`715 F. App’x 1013 (Fed. Cir. 2017) ..................................................................... 7
`Neutrino Dev. Corp. v. Sonosite, Inc.,
`410 F. Supp. 2d 529 (S.D. Tex. 2006) .......................................................... 25, 26
`Norgren Inc. v. ITC,
`699 F.3d 1317 (Fed. Cir. 2012) .......................................................................... 25
`Rivera v. ITC,
`857 F.3d 1315 (Fed. Cir. 2017) .......................................................................... 19
`
`
`
`iv
`
`

`

`Schott Gemtron Corp. v. SSW Holding Company, Inc.,
`IPR2013-00358, Paper No. 106 (PTAB Aug. 20, 2014) .................................... 26
`Wellman, Inc. v. Eastman Chem. Co.,
`642 F.3d 1355 (Fed. Cir. 2011) .......................................................................... 23
`
`IPR2020-01649
`Paper No. 18
`
`
`
`
`
`
`
`
`v
`
`

`

`EXHIBIT LIST
`
`IPR2020-01649
`Paper No. 18
`
`EXHIBIT 1001 U.S. Patent No. 9,125,739 (Paniagua et al.)
`EXHIBIT 1002
`File History for U.S. Patent No. 9,125,739 (Paniagua et al.)
`EXHIBIT 1003 Andersen et al., Transluminal Implantation of Artificial Heart
`Valves. Description of a New Expandable Aortic Valve and
`Initial Results with Implantation by Catheter Technique in
`Closed Chest Pigs, European Heart Journal 13:704-708 (1992)
`EXHIBIT 1004 U.S. Patent No. 5,411,552 (Andersen et al.)
`EXHIBIT 1005 Bailey, Percutaneous Expandable Prosthetic Valves, Textbook
`of Interventional Cardiology, Chapter 75 (1994)
`EXHIBIT 1006 U.S. Patent No. 5,855,601 (Bessler et al.)
`EXHIBIT 1007 U.S. Patent No. 5,332,402 (Teitelbaum)
`EXHIBIT 1008 Moazami et al., Transluminal Aortic Valve Placement A
`Feasibility Study with a Newly Designed Collapsible Aortic
`Valve, ASAIO J., M381-385 (Sept.-Oct. 1996)
`EXHIBIT 1009 U.S. Patent Application Publication No. 2002/0032481 (Gabbay)
`EXHIBIT 1010 U.S. Patent No. 6,425,916 (Garrison et al.)
`EXHIBIT 1011 U.S. Patent No. 6,652,578 (Bailey et al.)
`EXHIBIT 1012 U.S. Patent No. 5,957,949 (Leonhardt et al.)
`EXHIBIT 1013
`Ebeid et al., Use of Balloon-Expandable Stents for Coarctation
`of the Aorta: Initial Results and Intermediate-Term Follow-Up,
`30 J. Am. C. Cardiology 7, 1847-52 (Dec. 1997)
`Edwards Receives European Approval for Advanced SAPIEN 3
`Valve (Jan. 27, 2014)
`EXHIBIT 1015 U.S. Patent Application Publication No. 2005/0113910
`(Paniagua et al.)
`EXHIBIT 1016 Rösch et al., The Birth, Early Years, and Future of
`Interventional Radiology, J. Vasc. Interv. Radiol., 14:841-853
`(2003)
`EXHIBIT 1017 U.S. Patent No. 6,908,481 (Cribier)
`EXHIBIT 1018
`File History for U.S. Patent No. 8,308,797 (Paniagua et al.)
`EXHIBIT 1019 U.S. Patent Application Publication No. 2001/0044633 (Klint)
`EXHIBIT 1020 Declaration of Steven L. Goldberg, M.D.
`
`EXHIBIT 1014
`
`
`
`vi
`
`

`

`IPR2020-01649
`Paper No. 18
`
`EXHIBIT 1028
`
`EXHIBIT 1029
`
`EXHIBIT 1021 U.S. Patent No. 8,308,797 (Paniagua et al.)
`EXHIBIT 1022
`Infringement Chart from Colibri Heart Valve LLC v. Medtronic
`CoreValve LLC, No. 8:20-cv-847-DOC-JDE, D.I. 30-3 (C.D.
`Cal. June 12, 2020) (Exhibit A to First Amended Complaint).
`EXHIBIT 1023 U.S. Patent No. 6,730,118 (Spenser et al.)
`EXHIBIT 1024
`Press Release, Edwards Lifesciences Receives CE Mark for
`Edwards SAPIEN Transcatheter Heart Valve Edwards
`Lifesciences (Sept. 5, 2007)
`EXHIBIT 1025 Colibri’s Opposition to Medtronic’s Motion to Dismiss First
`Amended Complaint [DKT. 38], Colibri Heart Valve LLC v.
`Medtronic CoreValve LLC, No. 8:20-cv-847-DOC-JDE, D.I. 46
`(C.D. Cal. Aug. 14, 2020)
`EXHIBIT 1026 U.S. Patent No. 8,900,294 (Paniagua et al.)
`EXHIBIT 1027 CoreValve Receives CE Mark Approval for its ReValving™
`System and Announces Plans to Initiate Expanded Clinical
`Evaluation, Business Wire (May 16, 2007)
`Edwards SAPIEN XT Transcatheter Valve and Delivery Systems
`Receive CE Mark (Mar. 2, 2010)
`for First
`Edwards Lifesciences Receives FDA Approval
`Catheter-Based Aortic Heart Valve in the U.S. (Nov. 2, 2011)
`EXHIBIT 1030 Medtronic CoreValve® System Obtains Early FDA Approval on
`Exceptional Clinical Performance (Jan. 17, 2014)
`EXHIBIT 1031 Bonhoeffer et al., Percutaneous Replacement of Pulmonary
`Valve in a Right-Ventricle to Pulmonary-Artery Prosthetic
`Conduit with Valve Dysfunction, Lancet 356:1403-05 (2000)
`Pavcnik et al., Aortic and Venous Valve for Percutaneous
`Insertion, Min Invas Ther & Allied Technol 9(3/4) 287-92
`(2000)
`EXHIBIT 1033 Declaration of Megan E. Dellinger in Support of Petitioners’
`Motion for Admission Pro Hac Vice of Megan E. Dellinger
`Pursuant to 37 C.F.R. § 42.10
`Final Written Decision from Edwards Lifesciences Corp. et al. v.
`Boston Scientific Scimed, Inc., IPR2017-00060, Paper 64 (PTAB
`March 23, 2018)
`EXHIBIT 1035 USPTO implementation of an interim Director review process
`following Arthrex, available at
`https://www.uspto.gov/patents/patent-trial-and-appeal-
`board/procedures/uspto-implementation-interim-director-review
`
`EXHIBIT 1032
`
`EXHIBIT 1034
`
`
`
`vii
`
`

`

`IPR2020-01649
`Paper No. 18
`
`EXHIBIT 1038
`
`(last accessed Sept. 9, 2021)
`EXHIBIT 1036 Reply Declaration of Steven L. Goldberg, M.D.
`EXHIBIT 1037 Cribier, The Development of Transcatheter Aortic Valve
`Replacement (TAVR), Glob Cardiol Sci Pract., Dec. 30, 2016
`FDA Clears Aortic, Mitral Valve-In-Valve Procedures for
`Sapien 3 TAVR Valve, available at
`https://www.dicardiology.com/article/fda-clears-aortic-mitral-
`valve-valve-procedures-sapien-3-tavr-valve (last accessed
`Sept. 9, 2021)
`EXHIBIT 1039 Deposition Transcript of Dr. Lakshmi Prasad Dasi (Aug. 25,
`2021)
`
`
`
`
`
`
`
`
`
`viii
`
`

`

`IPR2017-00060
`Paper No. 18
`PO’s attempt to salvage Claims 1-5 of the ’739 should be rejected. PO’s
`
`arguments mischaracterize the prior art, misapply the law, and rely heavily upon
`
`conclusory and unsupported assertions by its declarant that are contradicted by the
`
`underlying record. The Challenged Claims are unpatentable.
`
`I.
`
`GROUNDS 2-5 EACH RENDER CLAIMS 1-5 OF THE ’739
`UNPATENTABLE
`A. Grounds 2-5: Bessler Teaches Nearly All Claimed Limitations
`of the ’739 and Would Be Relied Upon by a POSITA as a Base
`Reference in an Obviousness Analysis
`
`
`
`PO does not dispute that Bessler teaches nearly all claimed limitations of the
`
`’739. It argues that a POSITA would not have relied on Bessler as a starting point,
`
`when the ’739 copied heavily from Bessler’s teachings (Petition, 4-11, 47-66). PO
`
`and its expert argue without support that a POSITA “would have recognized that
`
`Bessler’s valve device design was inferior to the state of the art in January 2002.”
`
`(Patent Owner Response (“POR”), 25). This, according to PO, is because “a sine-
`
`wave or zig-zag configuration stent design … was not a desirable design” as
`
`compared to a stent with “a network pattern.” (Id.). But PO’s arguments are
`
`contrary to the law and the express teachings of Bessler and otherwise unsupported
`
`by the record.
`
`First, PO’s argument is legally wrong because the Federal Circuit has
`
`“rejected the notion that a patent challenger seeking to demonstrate obviousness
`
`must prove that a [POSITA] would have been motivated to select one prior art
`
`
`
`1
`
`

`

`IPR2020-01649
`Paper No. 18
`disclosure over another.” Infineum USA L.P. v. Chevron Oronite Co., 844 F.
`
`App’x 297, 305 (Fed. Cir. 2021); see also In re Mouttet, 686 F.3d 1322, 1334 (Fed.
`
`Cir. 2012) (“just because better alternatives exist in the prior art does not mean that
`
`an inferior combination is inapt for obviousness purposes.”). A challenger just
`
`needs to show a POSITA would have been motivated to modify Bessler’s
`
`teachings with a reasonable expectation of success to arrive at the claimed
`
`invention.
`
`Second, PO’s arguments are at best a criticism of Bessler’s preferred
`
`embodiment. Not once in the 67-page Response or 105-page Dasi Declaration is it
`
`acknowledged that Bessler is not limited to a sine-wave or zig-zag configuration,
`
`and that Bessler, as an alternative, expressly teaches a “mesh configuration or a
`
`similar configuration which will allow the stent to be readily collapsible and self-
`
`expandable.” (EX1006, 6:7-11 (emphasis added)). Dr. Dasi ignored this express
`
`teaching—“[t]o my knowledge, there’s no mesh pattern stent disclosed in Bessler.”
`
`(EX1039, 35:15-36:6.) This disclosure couldn’t have been lost on PO as it cherry-
`
`picks the sine-wave and zig-zag configurations from the same portion of Bessler.
`
`(POR, 25 (citing Bessler, 5:15-6:18)). And Bessler’s description of the various
`
`stent forms that can be adopted was copied word-for-word into the ’739’s
`
`specification. (See Petition, 5-6). It is this copied “mesh” disclosure that supports
`
`
`
`2
`
`

`

`IPR2020-01649
`Paper No. 18
`the “network pattern” Dr. Dasi touts as the most-preferred stent design in January
`
`2002. EX1039, 29:2-30:5.1
`
`Third, Dr. Dasi opines that “Bessler’s single-wire designs would not have
`
`been expected to supply the uniform radial force needed to force the native aortic
`
`leaflets out of place.” (EX2019, ¶108). In addition to ignoring Bessler’s
`
`alternative teaching of a mesh-type stent, Dr. Dasi also ignores Bessler’s teaching
`
`that the diseased native valve can be removed prior to implantation. (EX1006,
`
`2:63-67, 4:27-52, 6:32-50, 7:1-25).2 Any suggestion that the zig-zag and sine-
`
`wave embodiments of Bessler lack sufficient radial force to withstand calcified,
`
`native aortic leaflets is a red herring because (1) the leaflets can be removed prior
`
`to implantation, and (2) the ’739’s claims are not limited to the treatment of aortic
`
`“Network pattern” is not a term used to describe any stent pattern in the
`
`1
`
`’739. “Mesh” is the term that appears in the specification, which Dr. Dasi agrees is
`
`a type of “network pattern.” (See, e.g., EX2019, ¶51; EX1039, 30:1-31:22.) The
`
`references Dr. Dasi cites for the proposition that a “network pattern” was the
`
`“most-preferred stent design” in January 2002 actually take no position on
`
`preferred stent designs. (See EX2019, ¶51 (citing EX2023, EX2024)).
`
`2
`
`Largely copying the disclosure of Bessler, the ’739 also discloses that “[t]he
`
`defective heart valve is removed by a suitable modality ….” (EX1001, 10:60-66;
`
`see EX1006, 4:30-36; Petition, 11).
`
`
`
`3
`
`

`

`IPR2020-01649
`Paper No. 18
`stenosis, and more broadly encompass treatment of insufficient (i.e., regurgitant)
`
`valves where no calcification is present. (EX1039, 10:10-12:5, 95:5-96:15; see
`
`also EX1034, 31 (“Patent Owner’s argument is premised on the false notion that
`
`claim 1 is directed to an aortic replacement valve that may only be placed where
`
`there is a ‘hard, calcified aortic annulus.’”)). Moreover, PO ignores that Dr.
`
`Bessler successfully implanted a wireform stent structure in the aortic position in
`
`animal studies, and that other wireform structures, including Andersen and
`
`Pavcnik, were successfully implanted in the aortic position. (EX1008 at 1, 4-5
`
`(Bessler) (Petition, 14-15), EX1032 (Pavcnik); EX2021 (Andersen)). PO’s bald
`
`assertions to the contrary should be rejected.
`
`B. Grounds 2-5: It Would Have Been Obvious to Modify Bessler
`to Include a Trumpet-Like Configuration in View of the
`Teachings of Leonhardt and Teitelbaum
`
`
`
`PO does not dispute that both Leonhardt and Teitelbaum teach stents with
`
`trumpet-like configurations. (POR, 51-63). PO argues only that a POSITA would
`
`not have been motivated to modify the stent structure of Bessler in view of the
`
`teachings of Leonhardt and Teitelbaum. (Id.). But PO rests its argument on the
`
`false premise that Bessler is limited to a sine-wave or zig-zag stent structure, and
`
`on the erroneous assertion that this prior art would not be combined by a POSITA
`
`because Bessler is limited to an aortic valve prosthesis and Leonhardt and
`
`Teitelbaum are limited to mitral valve prostheses.
`
`
`
`4
`
`

`

`IPR2020-01649
`Paper No. 18
`First, PO argues that a POSITA would not flare the ends of Bessler’s single-
`
`
`
`wire embodiment because it “would increase paravalvular leakage.” (POR at 51
`
`(citing EX2019, ¶¶124, 131)). But Paragraphs 124 and 131 of the Dasi Declaration
`
`are entirely conclusory, cite no support, and blatantly ignore the “mesh”-type stent
`
`structure explicitly taught by Bessler. See supra 2-3. PO concedes that a
`
`“preferred” mesh stent would increase sealing and decrease leakage. (POR at 53
`
`(“selecting Bessler’s [sine-wave or zig-zag] stent design would have decreased
`
`sealing and increased leakage of blood around the valve device—in contrast with
`
`the preferable network pattern a POSA would have chosen.” (emphasis in bold
`
`added))). Moreover, no matter which Bessler stent structure is adopted, PO
`
`ignores that Bessler teaches the use of a cuff that may cover at least a portion of the
`
`stent frame to aid in sealing the prosthesis against the surrounding tissue, thus
`
`undermining any argument that an increase in spacing between adjacent struts
`
`necessarily leads to an increase in paravalvular leaks. (See, e.g., EX1006, Claim 1
`
`(“wherein the cuff portion is configured to position the valve snugly and sealingly
`
`at a valve site”); Petition, 13).
`
`
`
`5
`
`

`

`IPR2020-01649
`Paper No. 18
`Second, PO’s argument that a POSITA would not combine teachings of an
`
`
`
`aortic valve prosthesis with teachings of a mitral valve prosthesis (or vice versa)3
`
`are erroneously based on the preferred embodiments of Bessler, Leonhardt, and
`
`Teitelbaum. None of these prior art references is limited to a prosthesis for a
`
`particular heart valve, which Dr. Dasi does not dispute. (See, e.g., EX1006, 2:26-
`
`29 and Claim 1; EX1012, 4:56-65, 9:64-67; EX1007, 4:38-44 and Claims 1, 5;
`
`EX1039, 98:5-108:13; EX1036, ¶7). Thus, any suggestion that a POSITA would
`
`not combine Bessler with either Leonhardt or Teitelbaum because they are directed
`
`at prostheses for different heart valves is belied by the broad disclosures of each of
`
`these prior art references. (EX1036, ¶7). Even if so-limited, the prior art
`
`recognizes that a prosthesis can be readily adapted for use in multiple valve
`
`locations (see, e.g., EX1012, 4:55-65 (Leonhardt recognizing that “one basic stent
`
`shape” could be adopted, which “may be manufactured with more or less outward
`
`force to accommodate other placement positions”)), and a POSITA looking to
`
`Contrary to Dr. Dasi’s opinion that “[e]ven today,” a POSITA “would not
`
`3
`
`take a transcatheter prosthetic mitral valve device and implant it in the aortic
`
`valve” or vice versa (which implies hindsight), (EX2019, ¶43), it is, in fact, done
`
`today: Edwards’ SAPIEN 3—the only FDA approved THV for mitral valve-in-
`
`valve procedures—is also approved for use in the aortic valve. (EX1036, ¶7 n.1;
`
`EX1038).
`
`
`
`6
`
`

`

`IPR2020-01649
`Paper No. 18
`make improvements to these just-emerging technologies4 would have considered
`
`any potentially beneficial features in prosthetic valves designed for any location
`
`(EX1036, ¶6).5
`
`
`
`Third, PO argues that “a POSA would be concerned that the flared portion
`
`of the stent would push the calcified, native leaflets into the sinus in a way that
`
`occludes the coronary arteries.” (POR, 57). But PO again ignores that Bessler
`
`teaches that the diseased native valve leaflets may be excised prior to implantation
`
`of the prosthesis, thus alleviating any purported concern of blocking the coronaries
`
`with the native valve leaflets. Supra 3; see also EX1039, 95:19-96:2. Even if the
`
`native leaflets are not removed, PO ignores Dr. Goldberg’s testimony that a
`
`Teitelbaum-like stent design “might actually reduce the risk of coronary
`
`Neither an aortic or mitral THV had been successfully implanted in a patient
`
`4
`
`as of January 2002; the first-in-human aortic THV implant was not until April
`
`2002. (EX1037).
`
`5
`
`Dr. Dasi’s cursory, unsupported assertions that certain teachings would not
`
`be a starting point for a POSITA or would not work in certain valve locations
`
`(EX2019, ¶¶42-44, 102-112, 119-22, 124, 128, 130-31, 133-35), cannot overcome
`
`the presumption that these teachings are enabled. Microsoft Corp. v. Parallel
`
`Networks Licensing, LLC, 715 F. App’x 1013, 1021 (Fed. Cir. 2017) (citing In re
`
`Antor Media Corp., 689 F.3d 1282, 1289, 1292 (Fed. Cir. 2012)).
`
`
`
`7
`
`

`

`IPR2020-01649
`Paper No. 18
`obstruction by pushing the leaflets down and away from the coronary arteries
`
`which are up higher.” (EX2020, 59:6-13). PO also ignores Petitioners’ argument
`
`that Cribier (EX1017), the Wallstent (EX1016, 7; EX1018, 167, 169), Garrison
`
`(EX1010), Gabbay (EX1009), and Bailey (EX1011) each provide additional
`
`motivation to combine Bessler with Leonhardt and Teitelbaum as each teach the
`
`use of trumpet-like stent configurations. Petition, 16-21, 51, 53. Cribier, for
`
`example, explicitly teaches the use of a stent with trumpet-like ends (i.e., “a
`
`concave shape”) “aimed at reinforcing the embedding and the locking of the
`
`implantable valve in the distorted aortic orifice.” (EX1017, 5:64-67; Petition, 20-
`
`21). Thus, any suggestion that a prosthesis with trumpet-like ends would not be
`
`desirable in the aortic position is countered by this volume of uncontested evidence
`
`and contrary to the understanding of a POSITA at the time of the alleged invention.
`
`C. Grounds 2-5: Under Any Interpretation of “Onto the Pusher
`Member,” It Would Have Been Obvious to Modify the
`Delivery System of Bessler to Include a Pusher Member as
`Claimed
`
`As set forth in the Petition, even if “onto” is defined narrowly such that it
`
`requires the pusher member to pass through the interior of the prosthetic valve, it
`
`would have been obvious to a POSITA to modify Bessler in view of the teachings
`
`of Klint, Leonhardt, or Teitelbaum to arrive at the ’739’s claimed delivery system.
`
`Petition, 61-63, 66-70. PO’s arguments mischaracterize the prior art and are wrong
`
`on the law.
`
`
`
`8
`
`

`

`1.
`
`PO Mischaracterizes Klint
`
`IPR2020-01649
`Paper No. 18
`
`PO first argues that “Petitioners’ argument relies on Klint’s Example 6,
`
`which includes two embodiments of pusher members, neither of which includes a
`
`prosthetic heart valve device collapsed on to the pusher member.” (POR, 41). But
`
`Klint explicitly teaches that “[t]he prosthesis may be of the radially compressible,
`
`self-expandable type such as … a valve,” (EX1019, ¶[0091]; Petition, 67; EX1036,
`
`¶11), and that “the engagement means [of the pusher member] can also be
`
`designed as an elongate member that extends coaxially inside the radially
`
`compressed prosthesis and engages the prosthesis at several locations along the
`
`length thereof … ,” (id., ¶[0094]; Petition, 68-69; EX1036, ¶10).
`
`Next, PO argues that “a POSA would understand that Klint’s second pusher
`
`member embodiment is deployed by being pushed through an already collapsed
`
`prosthesis when that prosthesis is in place and ready to be released from its
`
`receptacle” because there is only a “single disclosed method for deploying the
`
`pusher member for Klint’s Example 6.” (POR, 44; EX1036, ¶8). But there is
`
`nothing in Klint that limits its second embodiment of a pusher member to a
`
`particular method of use. (See also EX2020, 66:18-67:2; EX1036, ¶9). Even if
`
`there was, Claims 1-5 of the ’739 are device claims, not method claims, and
`
`require only that the prosthesis is in a collapsed state on the pusher member.
`
`Moreover, PO’s interpretation is at odds with the explicit disclosure of Klint,
`
`
`
`9
`
`

`

`IPR2020-01649
`Paper No. 18
`which plainly states that the engagement means “extends coaxially inside the
`
`radially compressed prosthesis.” (EX1019, ¶[0094]). Regardless of the method
`
`used with Klint’s second embodiment, Dr. Dasi’s own mock-up confirms a
`
`POSITA would understand that the valve prosthesis is collapsed onto the
`
`engagement means of the pusher member. (EX2019, ¶175; EX1039, 113:21-114:5
`
`(Dr. Dasi agreeing that, even in his interpretation of Klint’s second embodiment,
`
`the outflow end of the valve prosthesis is in a collapsed state on the pusher
`
`member)).
`
`2.
`A POSITA Would Have Been Motivated to Combine
`Bessler’s Heart Valve Device with Klint’s Pusher Member
`
`PO contends that “[e]ven if … Klint’s second embodiment is considered to
`
`disclose a valve device collapsed onto the pusher member before deployment, a
`
`POSA would not have combined Klint’s second embodiment with Bessler’s heart
`
`valve device.” (POR, 45). According to PO, “a POSA would have understood that
`
`the pusher of Klint’s second embodiment would have risked damaging the valve
`
`device’s leaflets.” (POR, 46-50). But PO’s argument is premised on Klint’s
`
`teaching of a pusher member having “‘radial projections, hooks, ridges,’ or the
`
`like” (POR, 47); PO ignores that “another kind of engagement means such as a
`
`high friction material” could be employed in any areas where a POSITA may be
`
`concerned that a radial projection, hook, or ridge may damage the tissue of the
`
`prosthesis (EX1019, ¶[0094]). See EX1036, ¶12.
`
`
`
`10
`
`

`

`IPR2020-01649
`Paper No. 18
`Even if radial projections, hooks, or ridges are used, the fact that Klint
`
`recognized that these projections are located at “several locations along the length”
`
`of the prosthesis teaches the obvious—that a POSITA could simply employ
`
`projections at locations where there is no valve tissue to minimize the risk of
`
`damaging the tissue. EX1036, ¶12. For example, a POSITA employing the
`
`teachings of Klint’s second embodiment of a pusher member with Bessler Figure 5
`
`would position pusher-member projections along the outflow end of the device
`
`where no valve tissue is located:
`
`(EX1006, Figs. 4 & 5 (red annotations added); see EX1036, ¶12). This is exactly
`
`what is depicted in Dr. Dasi’s modified Klint Figure 16, where the pusher member
`
`includes hooks that are engaging the outflow end of the valve prosthesis:
`
`
`
`
`
`11
`
`

`

`IPR2020-01649
`Paper No. 18
`
`
`
`(EX2019, ¶175; EX1039, 112:8-15 (Dr. Dasi agreeing that his drawing includes
`
`hooks along the outflow end of the valve device)).
`
`3.
`Leonhardt Teaches Colibri’s Interpretation of a Delivery
`System “Wherein the Prosthetic Heart Valve Is Collapsed onto
`the Pusher Member”
`According to PO and its expert, the “pusher member 420” in Figure 8 of the
`
`
`
`’739 is not limited to the nose cone feature labelled as 420, but instead includes a
`
`portion of the catheter that passes through the prosthesis. POR, 13-14; EX1039,
`
`46:20-47:6, 49:7-14, 61:14-22, 66:10-68:20. PO relabels this portion of the
`
`catheter “pusher member 420” (POR, 13-14), notwithstanding that Figure 8
`
`identifies this portion as “the catheter lumen 430” and the specification provides
`
`that the prosthesis is “collapsed over the tube,” which refers to “flexible hollow
`
`tube catheter 410” in Figure 8, not pusher member 420. (EX1001, 12:9-12).
`
`According to Dr. Dasi, the pusher member of the ’739 “can include portions [of the
`
`
`
`12
`
`

`

`IPR2020-01649
`Paper No. 18
`delivery system] that are adjacent to the stent structure on either side of the stent”
`
`(EX1039, 68:12-20) and the friction between the valve prosthesis and the tube it is
`
`collapsed onto is part of what supplies the pushing force to deploy the device (id.,
`
`66:20-67:14).
`
`
`
`Reading Leonhardt in light of PO’s interpretation of the ’739, Leonhardt
`
`discloses the same elements that Dr. Dasi identifies as the “pusher member” in
`
`Figure 8 of the ’739. (See also Petition, 61-62.) Leonhardt has a “nose cone”
`
`(“[t]apered head 156” (EX1012, 8:10-14)) and a tube the valve device is collapsed
`
`onto (inner catheter 110 and valve stent 20 (id., 7:11-20), not pictured below, see
`
`Figs. 9A-D of Leonhardt), and another portion of the “pusher member” behind the
`
`stent (visible within the catheter to the left of Leonhardt Figure 5), identified as the
`
`“push rod 112” (id., 6:43-49).
`
`
`
`13
`
`

`

`IPR2020-01649
`Paper No. 18
`
`
`
`That Leonhardt also includes this “push rod” element behind the valve device does
`
`not negate its disclosure of a “pusher member” akin to the one that PO has
`
`annotated in Figure 8 of the ’739—per Dr. Dasi, the pusher member is not limited
`
`to the portion of the delivery system under the prosthesis and “can include portions
`
`that are adjacent to the stent structure on either side” (EX1039, 68:12-20), and the
`
`frictional force that is part of the pushing force for deployment can be provided by
`
`
`
`14
`
`

`

`IPR2020-01649
`Paper No. 18
`the portion of the catheter tube on which the prosthesis is collapsed (id., 66:20-
`
`67:14).
`
`D. Grounds 4-5: Claim 5’s “Controlled Release Mechanism”
`Limitation Is Obvious
`
`PO does “not dispute that Bessler discusses a restraining method that would
`
`keep a valve device from popping out and hold it for controlled release using
`
`sutures and threads.” (POR, 65-66; EX2019, ¶188). PO argues only that
`
`physically substituting Klint’s pusher member for Bessler’s pusher member would
`
`require removal of Bessler’s controlled release mechanism. (POR, 65). This is
`
`wrong on the law. It has been long established that the obviousness inquiry does
`
`not ask “whether the references could be physically combined but whether the
`
`claimed inventions are rendered obvious by the teachings of the prior art as a
`
`whole.” In re Etter, 756 F.2d 852, 859 (Fed. Cir. 1985). There can be no dispute
`
`that, as a whole, the teachings of Bessler and Klint render the “controlled release
`
`mechanism” requirement obvious. (Petition, 8-9, 66-70).
`
`Even if Klint’s pusher legally had to be physically combined with Bessler’s
`
`delivery system, there is no suggestion that Bessler’s pusher member and
`
`controlled release sutures could not be combined with Klint’s pusher member, or
`
`that the resulting system could not be adapted to include locations to secure the
`
`sutures taught by Bessler if Bessler’s pusher member is replaced by Klint’s pusher
`
`member. (See POR, 65-66.) PO’s entire argument appears to be anchored to Klint
`
`
`
`15
`
`

`

`IPR2020-01649
`Paper No. 18
`not having holes in its pusher member. (POR, 65). But a POSITA in January 2002
`
`wanting to combine Bessler and Klint and to include the advantages of Bessler’s
`
`controlled release mechanism would have used their ordinary creativity to include
`
`holes (or another attachment means) in the pusher member taught by Klint so that
`
`Bessler’s sutures could be looped through. See ClassCo, Inc. v. Apple, Inc., 838
`
`F.3d 1214, 1219 (Fed. Cir. 2016) (“[A POSITA] is also a person of ordinary
`
`creativity, not an automaton…. The rationale of KSR does not support ClassCo’s
`
`theory that a [POSITA] can only perform combinations of a puzzle element A with
`
`a perfectly fitting puzzle element B.” (internal quotation marks, citations omitted));
`
`EX1036, ¶¶13-19. Indeed, Klint itself explains that “projections, hooks, ridges, or
`
`another kind of engagement means such as a high friction material” (EX1019,
`
`¶[0094] (emphases added); EX1036, ¶¶15-16) can be used, and another
`
`embodiment of Klint’s invention (for an embolization device), explicitly describes
`
`use of “two mating parts held together by a thread that can be pulled out for
`
`detachment of the embolization device” and explains that “[t]hese kinds of
`
`connection means are well known in the art, e.g., [collecting patents] ….”
`
`(EX1019, ¶[0113], ¶[0110]; EX1036, ¶¶15-16). Other references known by
`
`January 2002 confirm that a POSITA would know how to use threads or sutures to
`
`temporarily secure a device to a delivery system, and provide additional motivation
`
`
`
`16
`
`

`

`IPR2020-01649
`Paper No. 18
`to combine the teachings of Bessler and Klint. See, e.g., EX1009, ¶[0112];
`
`EX1010, 11:26-34; EX1012, 9:6-24, 11:37-63, Figs. 7a-7b; EX1036, ¶¶17-19.
`
`II. GROUND 1: PANIAGUA IS PRIOR ART AND ANTICIPATES
`CLAIMS 1-5
`
`There is no dispute that if Paniagua is prior art to the ’739, then it anticipates
`
`Claims 1-5. The only dispute is whether the specification as of January 4, 2002
`
`adequately described the full scope of Claims 1-5. It did not. (Petition, 34-37, 42-
`
`43).
`
`A.
`
`In the ’739, “The Leaflet Portion of the Valve Itself” Must Be
`“Formed from a Single Piece of Biocompatible Material” in
`Order to Attain “the Novel and Improved Qualities of the
`Present Invention”
`
`PO’s response relies on portions of t

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket