throbber
Applied
`Radiation and
`Isotopes
`
`PERGAMON
`
`Applied Radiation and Isotopes 50 (1999) 1015–1023
`
`Manufacture of strontium-82/rubidium-82 generators and
`quality control of rubidium-82 chloride for myocardial
`perfusion imaging in patients using positron emission
`tomography
`
`Teresa M. Alvarez-Diez a, b, *, Robert deKemp b, Robert Beanlands b,
`John Vincent c
`
`aNuclear Medicine Division, Ottawa Civic Hospital, 1053 Carling Ave., Ottowa, Ont., Canada K1Y 4E9
`bOttawa Heart Institute, 1053 Carling Ave., Ottowa, Ont., Canada K1Y 4E9
`cTRIUMF, 4004 Westbrook Mall, Vancouver, B.C., Canada V6T 2A3
`
`Received 13 July 1998; received in revised form 5 September 1998; accepted 8 September 1998
`
`Abstract
`
`We describe a protocol to manufacture 82Sr/82Rb generators and 82RbCl for myocardial imaging with PET. The
`generators are manufactured in 3 stages: (1) preparation of a tin oxide column, (2) leak test of the generator column
`and (3) loading of the generator with 82Sr. The generators produced sterile and non-pyrogenic 82RbCl for i.v.
`injection. No significant 82Sr/85Sr breakthroughs were observed after elution with 20 l of saline. The automated
`system delivered human doses of 82RbCl accurately. # 1999 Elsevier Science Ltd. All rights reserved.
`
`1. Introduction
`
`Conventional diagnostic techniques used to assess
`coronary artery disease and its severity are of limited
`sensitivity. Stress SPECT (single photon emission com-
`puted tomography) using thallium-201 (201Tl) or tech-
`netium-99m-sestamibi (99mTc-sestamibi) has been used
`extensively in clinical practice to determine myocardial
`perfusion. However, this technique is limited by at-
`tenuation artifacts and does not permit in some cases
`the accurate distinction between hypoperfused but
`viable myocardium and infarcted tissues, thus underes-
`timating the areas of viable tissue (Wackers et al.,
`1976; Cloninger et al., 1988; Galli et al., 1988).
`Cardiac positron emission tomography (PET) using
`several short half-life (t1/2) radionuclides has been used
`
`* Corresponding author. Present address: 495 Canotia
`Drive, Orleans, Ont., Canada K4A 2J4. Tel.: +1-613-830-
`4733; e-mail: teresa.ad@sympatico.ca
`
`to characterize myocardial perfusion and metabolism
`non-invasively (Bergmann et al., 1985; Goldstein et al.,
`1986; Brunken et al., 1987; Camici et al., 1989; Demer
`et al., 1989; Saha et al., 1992). It has been demon-
`strated that rubidium-82 (82Rb), a positron emitter
`(t1/2=75 s)
`radionuclide with ultra-short half-life
`(Woods et al., 1987), permits the assessment of myo-
`cardial perfusion with high sensitivity and specificity
`(Goldstein et al., 1983; Gould et al., 1986, 1988; Go et
`al., 1990; Stewart et al., 1991; Grover-McKay et al.,
`1992). The accuracy of 82Rb PET has been shown to
`be superior to 201Tl SPECT imaging (Go et al., 1990;
`Stewart et al., 1991). This is especially important for
`the detection of early coronary artery disease and for
`the evaluation of therapy designed to protect or sal-
`vage myocardium. Some clinical studies have also
`demonstrated the utility of 82Rb as a quantitative mar-
`ker of myocardial necrosis/viability (Gould et al.,
`1991; Dahl et al., 1996).
`There are two other advantages of 82Rb. It has a
`ultra-short t1/2 which allows the sequential perform-
`
`0969-8043/99/$ - see front matter # 1999 Elsevier Science Ltd. All rights reserved.
`PII: S 0 9 6 9 - 8 0 4 3 ( 9 8 ) 0 0 1 7 0 - 5
`
`JUBILANT EXHIBIT 1012
`Jubilant v. Bracco, IPR2018-01449
`
`

`

`1016
`
`T.M. Alvarez-Diez et al. / Applied Radiation and Isotopes 50 (1999) 1015–1023
`
`ance of scans every 10 min and reduces the exposure
`of the patients to radiation. Secondly, 82Rb is a PET
`radiopharmaceutical which is generator-produced from
`its parent radionuclide strontium-82 (82Sr). This makes
`it feasible for institutions to participate in investiga-
`tional and clinical PET studies without the need to
`have expensive on-site cyclotrons.
`Several studies have been reported in the literature
`about
`the manufacture of strontium-82/rubidium-82
`(82Sr/82Rb) generators. Neirinckx et al., 1982 demon-
`strated the e(cid:129)ciency of hydrous tin oxide, a cationic
`exchanger material, for the separation of 82Rb from
`82Sr in a column. A special
`issue of
`this journal
`(Waters and Coursey, 1987) contained 11 papers deal-
`ing with 82Sr production and the physics and chemistry
`of the 82Sr/82Rb generators and 2 papers dealing with
`the clinical applications of 82Rb. However, much of
`these works were taken from on-going projects and
`there are in several cases discrepancies reported. Also,
`our group had previously described the development
`of a 82Rb generator system while focusing on the safe
`82Sr from metallic rubidium targets
`production of
`(Cackette et al., 1993). However, the 82Rb eluted from
`these generators is not suitable for clinical studies but
`only for research applications. In addition, a detailed
`protocol for the manufacture of 82Sr/82Rb generators
`from simple components that include quality control
`procedures and specifications of 82Rb for clinical ima-
`ging has not yet been reported in the literature.
`Thus, we describe in this manuscript a novel and
`simple manufacturing protocol which include the qual-
`ity control procedures for the production of 82Sr/82Rb
`generators and 82Rb chloride doses for i.v. adminis-
`tration in patients. Other nuclear pharmacies could
`produce generators in a similar fashion and participate
`in clinical PET studies. We discuss some of the factors
`that determine the performance of 82Sr/82Rb genera-
`tors for clinical imaging: (1) the type and properties of
`the cationic exchanger material used in the manufac-
`ture of the generator column; (2) the tightness of the
`cationic exchanger packing and (3) the type of eluent.
`We also report the quality control tests and specifica-
`tions of 82Rb chloride suitable for clinical myocardial
`imaging.
`
`2. Materials and methods
`
`All components of the generator column and the
`wrenches used during manufacturing were thoroughly
`washed with laboratory soap and water, rinsed with
`sterile water and then autoclaved. Soaking of the gen-
`erator components in soap and water for at least 20
`min removes the traces of lubricant used during its
`manufacturing. Rinsing the stainless steel components
`with a diluted solution of hydrochloric acid before
`
`autoclaving is not recommended because we observed
`that HCl with repeated autoclaving accelerates the oxi-
`dation of the components. The bu€ers and solutions
`used during the manufacturing of the generator col-
`umn are sterile and pyrogen free. 82Sr/82Rb generators
`are manufactured under aseptic conditions in 3 stages.
`
`2.1. Manufacturing of the generator column
`
`2.1.1. Preparation of hydrous tin oxide
`The generator column is prepared in a clean manu-
`facturing room dedicated to the compounding of
`radiopharmaceuticals. An excess of hydrous tin oxide
`(International Tin Research Institute, Middlesex,
`England) was sieved with a 150 mm stainless steel sieve
`for 10 min. The sieved tin oxide (with particle size
`<150 mm) was then sieved thoroughly with a 75 mm
`sieve in order to separate the fines (tin particles of
`<75 mm) from the tin oxide particles with sizes in the
`interval of interest, 150 mm>x>75 mm. The fines were
`discarded. The tin oxide, approximately 3.5 g, was
`washed with 30 ml of 0.1 N NH4OH/NH4Cl, pH 10
`and incubated overnight with 10 ml of this bu€er in
`order to activate its cationic exchanger properties. The
`tin oxide was kept in the bu€er of incubation until
`loaded in the column subassembly.
`
`2.1.2. Preparation of the generator column assembly
`The generator column assembly consists of two 9.5–
`1.5 mm Swagelok reducing adaptors with nuts and fer-
`rules, one column and two 25 mm filters (frits) (Fig. 1).
`The dimensions of the generator column are: 2.6 cm
`length, 6 mm internal diameter and 0.5 mm wall thick-
`ness. All components are made of stainless steel type
`316. A generator column subassembly is first prepared
`by attaching the column to a reducing adaptor con-
`taining a 25 mm filter. This column subassembly is
`then loaded with about 3.5 g of the wetted a-hydrous
`tin oxide (Sn2O(cid:1)xH2O where x = 1–2) in 10 ml of 0.1
`N NH4OH/NH4Cl bu€er. A 3-cm Teflon reservoir
`containing 0.1 N NH4OH/NH4Cl is connected to the
`upper end of the generator column and a vacuum
`aspirator to the lower end to facilitate the packing of
`the tin oxide into the column. It is important not to
`draw air through the column. An alternative method
`that facilitates the packing of the tin oxide column is
`to gently shake the reducing adaptor of the loaded col-
`umn with a small electrical vibrator (i.e. an engraver).
`When the packing is finished (the resin is leveled with
`the top of the column), a 7.9 mm Swagelok capping
`nut (plug) is screwed onto the column outlet to prevent
`fluid drainage. The column assembly is then completed
`by attaching a second Swagelok reducing adapter con-
`taining a 25 mm filter to the upper end of the column.
`The Swagelok nut is tightened one full turn past finger
`tight.
`
`

`

`T.M. Alvarez-Diez et al. / Applied Radiation and Isotopes 50 (1999) 1015–1023
`
`1017
`
`passing a few ml of the eluent, will indicate that the
`tin oxide of the generator column is in the Na+ form
`and that it is successfully working as a cationic exchan-
`ger material, exchanging some of its Na+ for the H+
`of the eluent. If the pH of the eluate is outside this pH
`interval, the generator column is discarded.
`After this test has been performed the hydrous tin
`oxide is replenished with sodium cations by repeating
`the 2 M NaCl flush.
`
`2.2. Generator assembly and leak test
`
`The generator inlet and outlet lines used are pre-
`formed 1.5 mm stainless steel tubes of 7 and 15.5 cm,
`respectively (Triumf, Vancouver, Canada) (Fig. 1). The
`outlet line is filled with 0.9% NaCl and attached to the
`column outlet using a 7.9 mm Swagelok nut and fer-
`rules. The nut is tightened 3/4 of a turn past finger
`tight. The inlet line is attached to the generator column
`inlet in the same way. This assembly is then attached
`to the lid of the shielding body (Fig. 2). The generator
`
`Fig. 1. Physical components of the strontium-82/rubidium-82
`generator column and infusion line used during generator’s
`manufacture. The components (1–11) are made of stainless
`steel type 316. The generator column assembly consistent of
`two 9.5–1.5 mm Swagelok reducing adaptors (4) with nuts
`and ferrules (1–3), one column (6), two 25 mm filters (frits) (5)
`and the generator inlet (7) and outlet (11) lines which are
`attached to the column using 7.9 mm Swagelok nuts (8) and
`ferrules (9–10). The Teflon infusion line (12) is attached to
`the syringe pump (luer end) and to the column inlet in order
`to infuse the bu€ers through the hydrous tin oxide column.
`
`The column backpressure is measured by passing ap-
`proximately 120 ml of 0.1 N NH4OH through the col-
`umn at flow rates of 10 and 20 ml/m using a syringe
`connected to a syringe pump and an in-line pressure
`gauge.
`The tin oxide exchanger is then saturated with
`sodium cations by passing 120 ml of 2 M NaCl
`through the column at a flow rate of 0.5 ml/m fol-
`lowed by 500 ml of 0.9% saline at a flow rate of 10
`ml/m.
`The cationic exchanger properties of hydrous tin
`oxide were tested by passing through the column
`about 10 ml of sterile water for irrigation USP that
`was acidified with 0.1 N HCl to pH 4. A rapid shift on
`the pH of the eluate from pH 6 to pH 9 to 10, after
`
`Fig. 2. This figure shows the generator column assembled,
`packed with hydrous tin oxide and attached to the lid of the
`shielding body. Both shielding lid and body (not shown) con-
`tain depleted uranium and weigh approximately 17 kg.
`
`

`

`1018
`
`T.M. Alvarez-Diez et al. / Applied Radiation and Isotopes 50 (1999) 1015–1023
`
`lines are passed through the hexagonal holes in the
`shielding lid until the lines protrude at least 1 cm
`above the surface of the lid. Female threaded luer fit-
`tings are attached to the inlet and outlet lines with 7.9
`mm Swagelok nuts and ferrules. The nut is tightened
`1/4 of a turn. The outlet line is then capped with a
`male luer plug. The inlet
`line of
`the generator is
`attached to a research purity argon gas cylinder using
`a 7.9 mm Swagelok nut. The generator column is then
`leak tested by passing a static pressure of 50 psi of
`argon. Leaks are detected by immersing the generator
`column into a 1 l beaker containing a su(cid:129)cient volume
`of sterile water to cover the column assembly (Fig. 3).
`If no leaks are observed, this assembly is detached
`from the argon gas cylinder and the generator column
`inserted into the shielding body. The shielding body is
`a cylinder of 2.8 cm internal diameter, 3 cm wall thick-
`ness and 15.5 cm length (not shown). Both shielding
`lid and body contain depleted uranium and weigh ap-
`
`Fig. 3. Leak test of the generator column. The generator col-
`umn is leak tested by capping the outlet line of the column
`and passing a static pressure of 50 psi of argon through the
`column. The generator column is immersed in a beaker con-
`taining 1 liter of sterile water. If a leak occurs, argon bubbles
`will be observed in the water.
`
`proximately 17 kg in total. The shielding lid is secured
`to the body using lever clamps.
`Spent generators are disassembled and the columns
`stored for decay. Generator shielding assemblies are
`reused to manufacture subsequent generators.
`
`2.3. Loading 82Sr into the generator column
`
`Loading of the generator with 82Sr is performed
`inside a lead castle in the Radiopharmacy laboratory.
`This castle has a front and back wall thickness of 7.5
`cm. Mirrors placed on the inside front and back walls
`of the castle provide an inside view. Lead bricks at the
`top of the side walls can be removed to enable manual
`access during the manufacturing procedure.
`The arrangement for loading the generator with 82Sr
`is shown in Fig. 4. It consists of the assembled genera-
`tor column, a 30 ml vial containing the 82Sr stock sol-
`ution (MDS Nordion, Vancouver, Canada), a syringe
`isolator and a waste container. The stock vial, syringe
`isolator and the waste container are shielded with lead.
`The syringe isolator (Triumf, Vancouver, Canada) con-
`sists of a 20 ml syringe attached by its plunger to the
`plunger of a 60 ml syringe. The plunger of the 60 ml
`syringe was modified to permit attachment of a 20 ml
`syringe. The 60 ml syringe of the isolator is connected
`to a remote 60 ml syringe which can be operated using
`a syringe pump.
`50 mCi of 82Sr in 0.1 N HCL with a concentration
`of <50 mCi/ml (MDS Nordion, Vancouver, Canada)
`was mixed with 15 ml of 0.5 M Tris bu€er pH 7.5.
`Using the remote syringe the 82Sr was withdrawn into
`the shielded syringe of the isolator with a 19G sterile
`spinal needle. The 82Sr solution is then pumped from
`the isolator through the generator column at a flow
`rate of 2 ml/h. After the generator column has been
`loaded with 82Sr, the column is purged with 500 ml of
`0.9% NaCl at a moderate flow rate of 0.5 ml/m.
`Washing the column with a large volume of saline is
`intended to remove any possible impurities contained
`in the 82Sr stock solution.
`
`2.4. Quality control of the 82Sr/82Rb generators
`
`The 82Sr/82Rb generators were eluted with sterile
`and pyrogen free 0.9% NaCl. The initial quality con-
`tests performed on the 82Rb chloride eluate
`trol
`included: visual
`inspection, pH measurement, radio-
`nuclide purity (82Sr and 85Sr breakthrough), chemical
`purity, sterility and pyrogen tests.
`The radionuclide purity of a sample of 82Rb was
`measured by quantitative gamma spectrometry using a
`calibrated and certified multi-channel analyzer (Tracor
`Northen, Middleton, Wisconsin, USA) with an intrin-
`sic germanium lithium detector (Ge(Li)) and computer
`analysis (AccuSpec). (The multi-channel analyzer was
`
`

`

`T.M. Alvarez-Diez et al. / Applied Radiation and Isotopes 50 (1999) 1015–1023
`
`1019
`
`Fig. 4. Arrangement for loading strontium-82 into the generator column. It consists of the assembled generator column (1), a vial
`containing the strontium-82 chloride solution (2), a syringe isolator (3) and a waste container (4).
`
`certified by the Institute for National Measurements
`Standards; National Research Council, Ottawa,
`Canada). The amounts of 82Sr and 85Sr in the eluate
`were measured relative to eluted yield of 82Rb. The
`82Sr/82Rb ratio limit established was 0.02 mCi/mCi.
`The limit of the 85Sr/82Rb ratio was 0.2 mCi/mCi. The
`elution yield of 82Rb was also measured with a dose
`calibrator and corrected for decay of the measured
`82Rb activity to the time of loading.
`A sample of 82Rb was sent to an independent lab-
`oratory for trace metal analysis by inductively coupled
`plasma/atomic emission spectroscopy (ICP/AES).
`A sterility test of 82Rb was performed by inoculation
`in trypticase soy broth and thioglycolate media for the
`detection of aerobic and anaerobic microorganisms.
`The absence in the eluate of pyrogenic substances
`was established using the limulus amebocyte lysate
`(LAL) test.
`82Sr and 85Sr breakthrough measurements using a
`dose calibrator and pyrogen tests of 82Rb were also
`performed daily before administration to the patients.
`The 82Sr and 85Sr breakthroughs relative to the peak
`of 82Rb radioactivity eluted were measured in 20-ml
`samples of 82Rb eluted at a rate of 20 ml/m and which
`were allowed to decay for 1 h after elution. 30 ml of
`the first elution of 82Rb is discarded every day. 82Sr
`breakthrough was
`calculated using the
`following
`
`equations:
`
`82Sr breakthrough (cid:136) 82Sr
`82Rb
`
`(cid:133)1(cid:134)
`
`where 82Sr is the amount of 82Sr in the sample in mCi
`calculated from Eq. (2); 82Rb is the peak of 82Rb
`radioactivity in mCi measured at
`the end of
`the
`elution.
`
`82Sr (cid:136) Radioactivity at one hour
`1 (cid:135) 0:48 (cid:2) R0
`
`:
`
`(cid:133)2(cid:134)
`
`The numerator of Eq. (2) is the radioactivity of the
`sample in mCi measured in a dose calibrator (on the
`82Rb and/or 82Sr setting) at least 1 h after elution (at
`complete decay of 82Rb) and R0 is the 85Sr/82Sr ratio
`on the date of the measurement, calculated from the
`manufacturer specifications. The correction factor 0.48
`is used to compensate for the contribution of the 85Sr
`to the reading. The 85Sr breakthrough was calculated
`by multiplying the 82Sr (mCi) by R0.
`85Sr breakthrough (cid:136)82 Sr Breakthrough (cid:2) R0:
`
`(cid:133)3(cid:134)
`
`A sterility test of a sample of 82Rb was also performed
`every day retrospectively.
`
`

`

`1020
`
`T.M. Alvarez-Diez et al. / Applied Radiation and Isotopes 50 (1999) 1015–1023
`
`2.5. 82Rb automated infusion system
`
`When the initial quality control of the 82Sr/82Rb
`generator is finalized, the generator is connected to an
`automated delivery system (Fig. 5). Automated intra-
`venous (i.v.) infusion of 82Rb chloride eluted from the
`generator into the patients is required due to the short
`half-life of 82Rb (75 s). The delivery system is a com-
`puter controlled peristaltic pump with direct monitor-
`ing of the 82Rb radioactivity eluted from the 82Sr/82Rb
`generator. The primary function of the delivery system
`is to accurately deliver a patient dose of 82Rb during
`the infusion time. This is achieved by varying the infu-
`sion rate of 82Rb to a maximum of 50 ml/m and by
`monitoring the 82Rb radioactivity eluted from the 82Sr/
`82Rb generators. The infusion system also records the
`actual volume of 82Rb administered (<100 ml) as well
`as the infusion backpressure and administered dose
`profiles over time. The backpressure from the genera-
`tor column is limited to 30 psi. An emergency stop
`button can be depressed in the event of patient distress
`or a mechanical malfunction. Sterile tubing and fittings
`are used throughout. A disposable sterile i.v. infusion
`line with a one-way valve on each end is used for each
`
`patient. A sterilizing 0.22 mm Millipore filter is also
`attached to the patient valve outlet line. A bypass line
`permits the flushing of the remaining 82Rb radioac-
`tivity in the patient infusion line, reducing background
`radiation.
`
`3. Results and discussion
`
`The adsorption of 82Sr into the generator column
`was greater than 99.99%.
`82Sr and 85Sr breakthroughs measured by gamma
`spectroscopy were lower than the established limits;
`82Sr/82Rb ratio was <0.02 mCi/mCi and the 85Sr/82Rb
`ratio <0.2 mCi/mCi. The minimum detectable activity
`of the multi-channel analyzer used was 2 (cid:2) 10(cid:255)3 mCi of
`82Sr and 2 (cid:2) 10(cid:255)4 of 85Sr. The 82Sr and 85Sr break-
`through measurements performed daily in a dose cali-
`brator using Eqs. (1) and (3) enabled us to correct the
`contribution of 85Sr to the 82Sr breakthrough reading,
`which could have otherwise resulted in overestimates
`of the 82Sr breakthrough. No significant breakthrough
`was observed either by gamma spectrometry or by
`using a dose calibrator after elution of the generators
`
`Fig. 5. Positron emission tomograph (T) and strontium-82/rubidium-82 generator shielded with depleted uranium and lead rings
`(inside the stainless steel cart) (G) connected to the automated patient’s delivery system (D). The system delivers accurate intrave-
`nous doses of 82RbCl to the patients by pumping 0.9% NaCl (S) through the 82Sr/82Rb generator. The delivery system is connected
`to a controlling computer with a touch screen (C).
`
`

`

`T.M. Alvarez-Diez et al. / Applied Radiation and Isotopes 50 (1999) 1015–1023
`
`1021
`
`with more than 20 l of 0.9% NaCl. No other radio-
`nuclide impurities were detected by gamma spec-
`troscopy.
`The breakthrough of radioactive Sr is a function of
`the type and volume of the adsorbent of the generator
`column and the type and flow rate of the eluent. We
`and others (Brihaye et al., 1987; Kensett et al., 1987;
`Cackette et al., 1993) have previously demonstrated
`that hydrous tin oxide is an e(cid:129)cient cation exchanger
`material for the preparation of reliable 82Sr/82Rb gen-
`erators. Donaldson and Fuller, 1970 demonstrated that
`+ form is an e(cid:129)cient ma-
`hydrous tin oxide in the NH4
`terial for the separation of alkali metals from alkaline
`earth elements in a column. Sr2+ and Ca2+ were com-
`pletely retained on the column. However, the mono-
`valent cations tested (Na+, K+ and Cs+) were easily
`eluted without alkaline earth breakthrough using a
`mobile phase of NH3/NH4NO3. Neirinckx et al., 1982
`for the
`measured the distribution coe(cid:129)cients (KD)
`equilibration of 82Sr (II) and 82Rb (I) between hydrous
`tin oxide and NaCl solutions of di€erent pH and con-
`centrations (mobile phases). They showed that the KD
`values for 82Sr (II) were high and essentially constant
`between pH 7.2 and 11.4 (KD=53,000–56,000). The
`corresponding KD values for 82Rb (I) were 4 orders of
`magnitude lower. The 82Sr breakthroughs reported in
`this study were 5 (cid:2) 10(cid:255)8 and <10(cid:255)9/ml using hydrous
`tin oxide columns of 1- and 2-cc bed sizes, respectively.
`These columns were eluted with 0.9% NaCl at 10 ml/
`m.
`Other adsorbents previously used in the manufactur-
`ing of the 82Sr/82Rb generator column include Al2O3,
`the 82Sr
`ZrO2 and organic adsorbents. However,
`breakthrough values reported using these materials
`(Yano et al., 1979) were higher when compared with
`the breakthrough measurements observed when hy-
`drous tin oxide was employed. In addition, 82Sr/82Rb
`generators manufactured
`using Al2O3
`adsorbent
`require 2% NaCl eluent with pH 8–9 in order to mini-
`mize 82Sr breakthrough. This hypertonic eluent is not
`adequate for clinical use. It needs to be diluted prior
`to i.v. administration which further dilutes the 82Rb
`bolus, thus requiring higher patient infusion rates.
`However, when hydrous tin oxide is used as the
`adsorbent to manufacture 82Sr/82Rb generators,
`it is
`possible to elute 82Rb with physiological saline sol-
`utions of pHr 5.5 if the exchanger in the NH4
`+ form
`has been previously saturated with sodium ions. There
`is no need to neutralize the 82Rb chloride eluted. This
`was done following the method previously reported by
`our group (Cackette et al., 1993) and others (Brihaye
`et al., 1987). An exchange between the H+ ions of
`NaCl solution and the Na+ of the hydrous tin oxide
`occurs resulting in an increase of the pH value of the
`eluate to at least pH 6. At this pH, the separation of
`82Sr and 82Rb is maximized, thus minimizing the Sr
`
`breakthrough in the 82Rb eluted. Organic adsorbents
`have not been extensively utilized because they are un-
`stable to radiation.
`Another factor that determines the performance of
`the generator column is the tightness of the tin oxide
`packing. This factor was assessed by measuring the
`backpressure of the column when passing a volume of
`NH4OH at flow rates of 10 and 20 ml/m. The column
`backpressure ranged from 20 to 30 kPa and from 50
`to 60 kPa for these flow rates, respectively.
`The minimum backpressure established were 20 and
`50 kPa for 10 and 20 ml/m. Backpressures lower than
`20 kPa are the result of loose packing of the tin oxide
`into the column, which may result in low 82Rb yield
`and high 82Sr breakthrough. Columns with backpres-
`sures higher than 150 kPa were not accepted because
`the pressure limit of the infusion pump used is reached
`at approximately this pressure.
`The 82Rb elution yield obtained was at least 80%
`using an elution rate of 20 ml/m.
`The 82Rb infusion system permits the accurate deliv-
`ery of a 82Rb patient dose of 5 to 10 mCi over 1 min.
`This is achieved by setting the dose of 82Rb to be
`administered and the infusion time in the automated
`system and using variable infusion rates of 82Rb (up to
`50 ml/m)
`and
`infusion
`volumes
`(maximum
`volume = 30 ml). However, commercially available
`82Rb delivery systems (Squibb Diagnostic, 1995) do
`not permit slow infusions of 82Rb because the elution
`rate must be maintained constant at 50 ml/m. These
`doses of 82Rb are delivered instead by bolus adminis-
`tration. The infusion of 82Rb versus a bolus injection
`is advantageous for quantification studies in order to
`measure the arterial input function accurately and to
`assess necrosis/viability in patients.
`The chemical purity of samples of 82Rb eluted from
`3 di€erent 82Sr/82Rb generators was measured by ICP/
`AES, (Table 1). Stannous (Sn) was not detected in the
`samples. The detection limit of Sn using this method is
`5 (cid:2) 10(cid:255)2 ppm. 82Rb chloride samples contained 0.12–
`2.67 ppm of NH3, 0–2 (cid:2) 10(cid:255)3 ppm Pb, 0–0.11 ppm of
`Cu, 4 (cid:2) 10(cid:255)2–7 (cid:2) 10(cid:255)2 ppm of Fe and 0–1 ppm of Ca
`and Mg, respectively. However, the amounts of metal
`impurities contained in the maximum patient volume
`of 82Rb administered are lower than the toxic levels
`for i.v. injections (data not shown).
`Titration of Sn and trace metal analysis of samples
`of two batches of hydrous tin oxide were also per-
`formed by ICP/AES and compared with the data pro-
`vided by the manufacturer (Table 2). The samples
`analyzed contained more than 68% of Sn. The first
`column shows a typical analysis of hydrous tin oxide.
`The second and third columns show the actual trace
`metal analysis of the two batches of hydrous tin oxide
`analyzed. We have also indicated on column 5, the
`method detection limit
`(MDL)
`for
`each metal
`
`

`

`1022
`
`T.M. Alvarez-Diez et al. / Applied Radiation and Isotopes 50 (1999) 1015–1023
`
`Table 1
`Trace metal analysis of 82Rb chloride eluted from the 82Sr/
`82Rb generators
`
`During the entire shelf life (2–3 months) of the 82Sr/
`82Rb generator, all samples of 82Rb chloride eluted
`were sterile and pyrogen free.
`
`Trace metals
`
`Concentration (ppm)
`
`Sn
`N–NH3
`Pb
`Cu
`Fe
`Ca
`Mg
`
`NDb
`0.12–2.67
`ND–2 (cid:2) 10(cid:255)3
`ND–0.11
`4 (cid:2) 10(cid:255)2–7 (cid:2) 10(cid:255)2
`ND–1
`ND–1
`
`MDL (ppm)a
`5 (cid:2) 10(cid:255)2
`2 (cid:2) 10(cid:255)2
`2 (cid:2) 10(cid:255)3
`5 (cid:2) 10(cid:255)2
`5 (cid:2) 10(cid:255)2
`1
`1
`
`The second column of this table indicates the minimum
`and maximum concentrations of metals in ppm observed in
`the 82Rb chloride samples analyzed (n = 3). aMinimum detec-
`tion limit (MDL) of inductively coupled plasma/atomic emis-
`sion spectroscopy (ICP/AES). bNot detected (ND).
`
`measured. The MDL reported here are di€erent than
`those reported on Table 1 because these parameters
`varied with the sample matrix. 82Rb chloride is a sol-
`ution, however hydrous tin oxide is a solid. The levels
`of Pb and Cu, measured in the second batch of hy-
`drous tin oxide samples were higher than the typical
`levels reported by the manufacturer. The amounts of
`Ca and Mg on hydrous tin oxide were relatively high
`compared with other metals. This may be due to the
`absorption of these metals during the manufacturing
`process. Hydrous tin oxide is a good adsorbent of
`divalent cations. However, Ca and Mg were not
`detected in the 82Rb eluate. Even when the batch of tin
`oxide that contained relatively high levels of
`trace
`metals was used to manufacture the generator column,
`we did not observe a reduction on its cationic
`exchange e(cid:129)ciency.
`
`Table 2
`Sn assay and trace metal analysis of hydrous tin oxide
`
`Trace
`metals
`
`Typical
`analysisa
`
`Actual analysisb
`
`MDLc
`
`batch 1
`
`batch 2
`
`Sn
`Pb
`Cu
`Fe
`Sb
`Ca
`Mg
`
`66%
`0.1 ppm
`10
`20
`40
`
`68.5%
`NDd
`ND
`4
`ND
`ND
`ND
`
`68.8%
`34 ppm
`70
`100
`1
`458
`159
`
`0.01%
`3 ppm
`1
`1
`1
`25
`25
`
`a Analysis provided by Tin Research Institute, Kingston
`Lane, Middlesex, England. bTrace metal analysis of
`two
`batches of hydrous
`tin oxide performed by inductively
`coupled plasma (ICP)/atomic emission spectroscopy (AES).
`cMethod detection limit.dNot detected.
`
`4. Conclusion
`
`82Sr/82Rb generators manufactured following this
`protocol provide a means to obtain sterile and pyrogen
`free solutions of 82Rb chloride injection. This protocol
`uses simple components and quality control tests com-
`monly used in radiopharmacy.
`Our delivery system permits the infusion of small
`doses of 82Rb into the patients which is suitable for
`quantification studies of myocardial blood flow.
`Other nuclear medicine facilities could use this tech-
`nique to produce on-site 82Sr/82Rb generators and
`82Rb chloride for clinical studies.
`
`References
`
`Bergmann, S.R., Fox, K.A.A., Geltman, E.M., Sobel, B.E.,
`1985. Positron emission tomography of the heart. Prog.
`Cardiovasc. Dis. 28, 165–194.
`Brihaye, Cl., Guillaume, M., O’Brien, H.A., Raets, D., De
`Landsheere, CH., Rigo, P., 1987. Preparation and evalu-
`ation of a hydrous tin (IV) oxide 82Sr/82Rb medical genera-
`tor system for continuous elution. Appl. Radiat. Isot. 38,
`213–217.
`Brunken, R., Schwaiger, M., Grover-McKay, M., Phelps, M.,
`Tillisch, J., Schelbert, H.R., 1987. Positron emission tom-
`ography detects tissue metabolic activity in myocardial seg-
`ments with persistent thallium perfusion defects. JACC 10,
`557–567.
`Cackette, M.R., Ruth, T.J., Vincent, J.S., 1993. Sr production
`from metallic Rb targets and development of an Rb genera-
`tor system. Appl. Radiat. Isot. 44, 917–922.
`Camici, P., Ferrannini, E., Opie, L.H., 1989. Myocardial
`metabolism in ischemic heart disease: basic principles and
`applications to imaging by positron emission tomography.
`Prog. Cardiovasc. Dis. 32, 217–238.
`Cloninger, K.G., DePuey, E.G., Garcia, E.V., Roubin, G.S.,
`Robbins, W.L., Nody, A., DePasquale, E.E., Berger, H.J.,
`1988. Incomplete redistribution in delayed thallium-201
`single photon emission computed tomographic (SPECT)
`images: an overestimation of myocardial scarring. J. Am.
`Coll. Cardiol. 12, 955–963.
`Dahl, J.V., Muzik, O., Wolfe, E.R., Allman, C., Hutchins,
`G., Schwaiger, M., 1996. Myocardial rubidium-82 tissue
`kinetics assessed by dynamic positron emission tomography
`as a marker of myocardial cell membrane integrity and via-
`bility. Circulation 93, 238–245.
`Demer, L.L., Gould, K.L., Goldstein, R.A., Kirkeeide, R.L.,
`Mullani, N.A., Smalling, R.W., Nishikawa, A., Merhige,
`M.E., 1989. Assessment of coronary artery disease severity
`by positron emission tomography. Circulation 79, 825–835.
`
`

`

`T.M. Alvarez-Diez et al. / Applied Radiation and Isotopes 50 (1999) 1015–1023
`
`1023
`
`Donaldson, J.D., Fuller, M.J., 1970. Ion exchange properties
`of tin (IV) materials-III. J. Inorg. Nucl. Chem. 32, 1703–
`1710.
`Galli, M., Bencivelli, W., Pardo, N.F., Tavazzi, L., 1988.
`Underestimation of residual ischemia by 201-Thallium scin-
`tigraphy after myocardial infarction. Chest 94, 876–878.
`Go, R.T., Marwick, T.H., MaIntyre, W.J., Saha, G.B.,
`Neumann, D.R., Underwood, D.A., Simpfendorfer, C.C.,
`1990. A prospective comparison of rubidium-82 PET and
`thallium-201 SPECT myocardial perfusion imaging utilizing
`a single dipyridamole stress in the diagnosis of coronary
`artery disease. J. Nucl. Med. 31, 1899–1905.
`Goldstein, R.A., Mullani, N.A., Marani, S.K., Fisher, D.J.,
`Gould, L.K., O’Brien, H.A., 1983. Myocardial perfusion
`with rubidium-82. II. E€ects of metabolic and pharmaco-
`logic interventions. J. Nucl. Med. 24, 907–915.
`Goldstein, R.A., Mullani, N.A., Wong, W.-H., Hartz, R.K.,
`Hicks, C.H., Fuentes, F., Smalling, R.W., Gould, K.L.,
`1986. Positron imaging of myocardial infarction with rubi-
`dium-82. J. Nucl. Med. 27, 1824–1829.
`Gould, K.L., Goldstein, R.A.

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket