throbber
Durability/Wear Testing of Heart Valve Substitutes
`Helmut Reul, Klaus Potthast
`
`Helmholtz-Institute for Biomedical Engineering, Aachen, Germany
`
`Background and aims of the study: The current stan-
`dardsfor accelerated heart valve testing have consid-
`erable differences in test conditions. Another prob-
`lem arises from the fact that such test systems are not
`standardizedatall. It was shownearlierthat different
`test systems generate totally different valve loading,
`even if operating at standard conditions. The present
`study aimed to improvethis unsatisfactory situation
`and to develop a new concept whereactual loading of
`valves is measured either in vitro or in vivo under
`physiologic conditions and subsequently to repro-
`duce these conditions during accelerated testing.
`Methods:Integral loading forces at valve closure were
`measuredfor several valve types using a piezoelectric
`force ring within a real-time circulatory mock loop
`under physiologic conditions. This facilitated defini-
`tion of a physiologic loading range. Physiologic load-
`ing was subsequently reproducedina single-chamber
`accelerated test system. Working conditions obtained
`in termsof stroke, bypass flow and compliance served
`as design criteria for a new test chamber and a com-
`
`plete 12-chamberaccelerated testing system.
`Results: The integral loading obtained using the force
`ring showeda correlation with previous in vitro and
`in vivo results of strain-gauged valves. Loading
`forces for mechanical valves are about one order of
`magnitude higher than for bioprosthetic valves and
`are strongly related to cardiac output for both valve
`types. At physiologic loading, however, the differen-
`tial pressures across the valves are considerably
`below those given in FDA guidelines.
`Conclusions: This pilot study demonstrates that phys-
`iologic valve loading is reproducible over a wide
`range under appropriate testing conditions. It also
`showedthat,at the back-pressuresof the currentstan-
`dards, the loading forces during accelerated testing
`exceed the real-time loading forces by far and, thus,
`may provide unrealistically high valve loads. These
`initial findings indicate that amendments of the cur-
`rently valid standards maybe need to be accorded.
`
`The Journal of Heart Valve Disease 1998;7:151-157
`
`In 1996 we publisheda critical review on the state of
`the art of heart valve wearandfatigue testing as repre-
`sented by the current standards of ISO, FDA and CEN
`(1). These standards have considerable differences in
`test conditions as far as pressure differences across the
`test valves and total cycle numbers are concerned.
`Moreover, since the testing devices themselves are not
`standardized, different devices generate totally differ-
`ent loading conditions on the valves, even if the tests
`are carried out under the same standard. This study
`clearly demonstrated that, under the current stan-
`- dards, the conditions of actual, in vivo impact loading
`of a valve cannot be reproduced. Also, the twotest
`devices which were compared, although both operat-
`ing at the given standards conditions, generate totally
`different loading conditions. For one tester loading
`decreases with increasing cycle rate, while for the other
`it increases.Forall testers, time history of loading and
`
`Address for correspondence:
`Prof. Dr.-Ing. H. Reul, Helmholtz-Institute for Biomedical Engineer-
`ing, Pauwelsstr. 20, D-52074 Aachen, Germany
`
`pressure difference across the valve are a function of
`test frequency, compliance and otherfactors.
`To improve this unsatisfactory situation it was
`strongly suggested that the actual loading conditions
`of each valve type be measuredeither in animal mod-
`els in vivo or within a circulatory mock loop which pro-
`vides physiologic loading conditions;
`this loading
`could then be subsequently reproduced during accel-
`erated wear testing. The present study is set out to
`investigate this concept.
`
`Measurementsof valve loading
`under physiologic conditions
`
`Since most valves - and especially bioprostheses - are
`not suitable for the attachmentof strain gauges due to
`the lack of appropriate fixation points which are repre-
`sentative for valve loading, a piezoelectric force mea-
`surement
`ring (Kistler,
`type 906 1A, Winterthur,
`Switzerland) (Fig. 1, top right) was usedto assess load-
`ing at valve closure. This type of force transducer mea-
`sures the integral loading force acting on the valve.
`
`© Copyright by ICR Publishers 1998
`
`PAGE 1 OF 7
`
`WATERS TECHNOLOGIES CORPORATION
`
`EXHIBIT 1014
`
`WATERS TECHNOLOGIES CORPORATION
`EXHIBIT 1014
`
`PAGE 1 OF 7
`
`

`

`152 Durability/weartesting of heart valve substitutes
`H. Reul, K. Potthast
`
`J Heart Valve Dis
`Vol. 7. No. 2
`March 1998
`
`Integral Loading Forcein Vitro
`400|(Piezoelectric Force Ring)™~
`
`
`
`Strut Loading in Vitro
`\ eo
`=
`
`eu e
`
`a
`
`Strut Loading in Vivo
`
`seme Re
`
`;
`
`120
`
`
`
`Force/N—>
`
`ooaoOOo
`
`20
`
`
`
`( NF = 100 kHz )
`
`semiconductor
`pressure transducer
`
`
`inductive
`displacement
`transducer
`
`elektromagnetic
`ACA
`
`vibrator
`
`0
`
`500
`
`2000
`1500
`1000
`Left Ventricular dp/dt / mmHg/s >
`
`2500
`
`Figure 2: In vivo andin vitro loading of a 29 mm BSCC valve
`as a functionof left ventricular dp/dt before valve closure.
`
`methodsgive a good correlation with left ventricular
`dp/dt and, thus, can as well be correlated among each
`other.
`
`Measurements underaccelerated testing condi-
`tions
`
`Figure 1: Schematic of single-chamberaccelerated valvetest
`system with piezoelectric force ring in the upperright.
`
`After establishing the above correlation between inte-
`gral forces and left ventricular dp/dt the valve was
`inserted into a specifically developed single-chamber
`system foraccelerated valve testing (also shownin Fig.
`1).
`
`The next step wasto correlate these integral loading
`forces with previously obtained in vivo and in vitro
`results (2,3) measured bya strain-gauged 29 mm Bjérk-
`An electromagnetic vibrator (Ling Dynamics, type
`Shiley Convexo-Concave (BSCC) mechanical
`tilting
`409, Royston, UK) generates a sinusoidal flow through
`disk valve. For this purpose, the strain-gauged BSCC
`the test valve by compression and extension of a metal-
`valve was mounted onthe force ring and inserted into
`lic bellows. When the valveis closed, fluid from the
`the mitral position of a physiologic circulatory mock
`lower chamber flows to the upper chamber via an
`loop as previously described (4). Both, strut loads and
`adjustable bypass. This bypass serves simultaneously
`integral loading force were measured simultaneously.
`to control the pressure difference across the valve,
`For these measurements the left ventricular dp/dt
`which is measured upstream and downstream of the
`(according to the FDA guidelines averaged over the
`valve by two semiconductorpressure transducers with
`last 20 ms before valve closure) of the model ventricle
`a natural frequency of 100 kHz (Cobe Disposable
`wasvaried between 500 and 1500 mmH¢g/s by increas-
`Transducer, Lakewood, CO, USA). Test fluid is water
`ing the stroke volumeata fixed test rate of 70 per min,
`at room temperature.
`a mean aortic pressure of 100 mmHg,andasystolic
`Thetest rate was increased from 200 to 2000 per min
`duration of 15%.
`and the integral loading forces were measured. The
`The results are shown in Figure 2. The upper curve
`pressure difference across the valve was kept constant
`represents the force ring measurements which range
`at 120 mmHgforall test rates. The results are present-
`from 25 to 110 N integral loading force at the corre-
`ed in Figure3.
`sponding dp/dt values. The two lower curves repre-
`The shaded area represents the range of loading
`sent the strut loading forces obtained by strain gauges
`forces under all potential physiologic conditions as
`in vivo and in vitro for the same valve; these range
`obtained in vivo and in vitro and as depicted in Figure
`from 2 to 20 N strut loading force. The difference in
`2. The central curve showsthe measuredintegral load-
`measured forces between the two methods is obvious
`ing forces under accelerated testing conditions under
`andis related to the fact that the strain gauges measure
`observance of the FDA conditions for accelerated valve
`only loadingof a single strut caused bystrut deflection,
`testing such as full opening andclosing and 120 mmHg
`while the force ring measuresthe integral total loading
`pressure difference. The minimally adjustable loads
`force acting on the closed valve. Nevertheless, both
`
`
`
`PAGE 2 OF 7
`
`PAGE 2 OF 7
`
`

`

`Vol. 7. No. 2
`March 1998
`
`Durability/wear testing of heart valve substitutes
`H.Reul, K. Potthast
`
`153
`
` J Heart Valve Dis
`
`Oo
`
`600
`
`1000 1200 1400 1600 1800 2000 2200
`800
`Test rate / beats / min»
`
`*5
`
`200
`
`400
`
`Figure 3: Integral loading forces within the single chamber
`accelerated test system obtained for a 29 mm BSCCvalveat a
`back-pressure of 120 mmHg (FDA-condition).
`
`are represented by the lower curve. Higher loads can
`be easily generated by adjusting stroke and bypass
`throttle of the test system. These results clearly show
`that physiologic valve loading can be reproduced over
`a wide range under appropriate accelerated testing
`conditions. It is also evident that, under FDA condi-
`tions, only a single loading force which increases with
`highertest rates can be generatedfor a definedtestrate,
`whereastest conditions outside of the current standard
`in termsof pressure differencefacilitate the adjustment
`of a range of physiologic loading conditions at any test
`rate. Thus, valve loadingat resting conditions as well at
`exercise conditions can be adjusted, resulting ina much
`better simulation of the loading history of an implant-
`ed valve.
`
`.
`
`Valve loading forces for alternative valve types
`
`So far, all presented results are only valid for one single
`valve type and size, a 29 mm diameter BSCC tilting
`disk valve. Therefore, in order to develop a generally
`valid concept, it is necessary to extend the above find-
`ings to other valve types andsizes. For this purpose, a
`St. Jude Medical (SJM) 27 mm mechanical bileaflet
`valve and an Jonescu-Shiley 25 mm pericardial bio-
`prostheses were selected as a first step. Both valves
`were inserted into mitral position of our circulatory
`mockloop andintegral loading forces were measured
`underthe following experimental conditions (accord-
`ing to FDA guidelines for pulsatile flow valvetesting):
`
`Cardiac output: 2 1/min and 7 1/min;
`Test rate: 70 per min;
`Meanaortic pressure: 100 mmHg;
`Meanatrial pressure: 10 mmHg; and
`Systolic duration: 35%.
`
`PAGE 3 OF 7
`
`
`
`
`
`IntegralLoadingForce/N—>
`
`= rso
`
`= id°o
`
`= oo
`
`Bure
`
`40
`
`20
`
`Loading under FDA -Conditions
`
`Rangeof Physiologic Conditions
`
`Minimally Adjustable Loading
`
`Force/N—>
`
`LoO
`
`NMwOoOo
`
`Force/N—>
`
`= oO 0
`
`2L/min
`
`7L/min
`
`2L/min
`
`7L/min
`
`Figure 4: Integral loading forces for a 27 mm SJM mechanical
`valve and a 25 mm ISP bioprosthesis measured underreal-
`time conditions within a circulatory mock loop.
`This range of test conditions represents the normal
`physiologic range and can be usedfor the characteriza-
`tion of corresponding loading conditions. For each
`valve type the measured loading forces were averaged
`over 100 cycles.
`The results (Fig. 4) show that loading for the mechan-
`ical valve is about one order of magnitude higher than
`for the bioprosthetic valve, and loading increases with
`a factor of about two when cardiac output changes
`from 2 1/min to 7 1/min.
`
`Design of a new fatigue tester
`
`Theprincipal design criteria for a valve test chamberin
`terms of stroke, displaced volume, bypass flow and
`compliance were obtained by meansof the above-men-
`tioned single-chamber system (see Fig. 1). Based on
`these criteria a new test compartment for general use
`wasdesigned (shown schematically in Fig. 5). For the
`new design, the electromagnetic vibrator was replaced
`by a swashplate with adjustable stroke. Otherwise, the
`operating principle was the sameas already described
`above.
`Larger valves generate higher impact.loading forces
`than smaller ones. Therefore, since the new test com-
`partment is intended for use with all kinds of valve
`sizes, an adjustable compliance chamber wasadded for
`additional control of
`loading forces. The desired
`impact force can then be adjusted by varying the fol-
`lowing four parameters:
`
`- stroke of the swashplate;
`-
`test rate;
`- air compliance; and
`- bypass throttle adjustment.
`
`The FDA guidelines require that for all valves, test-
`ing should be conducted onthree of the largest, medi-
`um and smallest of each valve type. One equivalenttis-
`
`PAGE 3 OF 7
`
`

`

`154 Durability/wear testing of heart valve substitutes
`H.Reul, K. Potthast
`
`Test compartmentof HIA-FT2
`
`eee Observation Tube
`
`__— Test-Compartment
`
`— Air-Compliance
`
`_> Pressure Taps
`
`Compliance Chamber
`
`
`
`open to atmosphere ~
`
`Throttle —
`
`Testvalve
`Bypass
`
`——
`
`Piston Rod
`
`
`
`J Heart Valve Dis
`Vol. 7. No. 2
`
`March 1998
`
`into the new test compartment and the loading forces
`together with the pressure difference across the valves
`were measured within the fatigue tester at a test rate of
`1000 per min for the SJM and 600 per min for the ISP,
`respectively. The results (Figs. 8 and 9) show that for
`both valve types the peak loading forces are within the
`previously determined physiologic range (compare Fig.
`4). The differential pressures, however, are considerably
`below thepressures given in the FDA guidelines, which
`should be adjusted to 120 mmHgfor both valve types.
`An exampleillustrating this controversy for the SJM
`valve (Fig. 10) showsthat, in this case, the pressure dif-
`ference across the valve was adjusted to 120 mmHg by
`reducing the test chamber compliance. Test
`rate,
`bypass flow and stroke were kept constant. As can be
`clearly seen, valve loading forces exceed 50 N and,
`thus, are far above physiologic loading.
`
`Discussion
`
`
`
`——— Metal Bellow
`
`Figure 6: Top view of newly developed 12-chamberaccelerated
`test system (HIA-FT2).
`
`
`
`}-— ——— AdjustableStroke
`
`SwashPlate
`
`Figure 5: Schematic cross-section of newly developed test com-
`partment for accelerated valvetesting.
`
`sue annulus diameterof each type reference valve must
`be tested under identical conditions. This results in a
`total of 12 test valves. Accordingly, a 12-cylinder
`fatigue tester (HIA-FT2) has been designed and manu-
`factured (Fig. 6).
`Thefirst two of the above-listed parameters cannot
`be adjusted individually; they are the same for all 12
`test compartments. Thus, after the correct adjustment
`of one compartmentthe other 11 must be adjusted by
`variation of air compliance and bypassthrottle flow.
`An example (Fig. 7) shows where the peak loading
`forces for a 27 mm SJM valve have been varied by
`changing the air compliance at a constant test rate,
`bypassthrottle and stroke settings. Peak load decreas-
`es linearly with increasing compliance volume and can
`easily be adjusted for physiologic loading conditions.
`
`PAGE 4 OF 7
`
`This pilot study demonstrates that physiologic valve
`loading can be reproduced under appropriate acceler-
`ated testing conditions. However, testing under these
`conditions requires special testing compartments with
`an increased number of control parameters for the
`adjustmentof physiologic loads. It also requires a two-
`step testing approach:first, physiologic loading forces
`have to be determined within a real-time circulatory
`mock loop; and second, this loading has to be repro-
`duced within the accelerated tester. For propertransfer
`
`Inafinal step a 27 mm mechanical SJM valve and a 25 of physiologic real-time loading forces to accelerated
`testing the test valves have to be mounted within a cal-
`mun lonescu-Shiley pericardial valve (ISP) were inserted
`
`Verification of concept within new fatigue tester
`
`PAGE 4 OF 7
`
`

`

`| Heart Valve Dis
`Vol. 7. No. 2
`March 1998
`
`
`
`
`
`
`Force/N-=
`
`
`
`70
`
`100
`90
`80
`Compliance Volume / ml—
`
`110
`
`Physiologic Range
`according to FDA-Guidelines
`
`Figure 7: Integral loading force as a function of air compliance
`volume measured for a 27 mm SJM valve within the new test
`compartment at constant test rate (1000 per min), bypass
`throttle andstroke settings.
`
`ibration compartment, equipped with a force ring
`transducer. Once physiologic loading is adjusted and
`the parametersettings in terms of stroke volume, com-
`pliance volume and bypass flow are obtained,
`the
`valves are mounted within the geometrically similar
`final test compartments and the continuously mea-
`sured pressure difference serves as the single control
`parameterfor long-term studies.
`It could also be shownthatat the back-pressures of
`the current standards the loading forces during accel-
`erated testing exceed thereal-time loading forces by far
`and thus mayprovide unrealistically high valveloads.
`These initial findings indicate that corresponding
`amendments of the currently valid standards may be
`necessary.
`
`25 mm ISP
`
`Physiologic Range
`
`Force/N—
`
`Ap/mmHg—>
`
`0,0
`
`01
`
`03
`0,2
`Time/s—>
`
`04
`
`05
`
`Figure 9: Integral loading force and pressuredifference across
`valve for a 25 mm ISP bioprosthesis at a test rate of 600 per
`min,
`
`PAGE 5 OF 7
`
`Durability/wear testing of heart valve substitutes
`H.Reul, K. Potthast
`
`155
`
`
`
`Force/N—»
`
`27 mm SJM
`
`4‘
`Physiologic Range
`
`Ap/mmHg—> Time /s —>
`
`Figure 8: Integral loading force and pressure difference across
`valve for a 27 mm SJM valveata test rate of 1000 per min.
`
`Durability testing of bioprosthetic heart valves
`requires some additional discussion. The FDA and ISO
`guidelines for the weartesting of tissue valves are very
`similar to those for mechanical valves. Valves are to be
`tested for an equivalent of five years at a peak back-
`pressure of at least 90 mmHgfor aortic and 120 mmHg
`for mitral valves. The valves should open and close
`completely. For stentless valves, the aortic wall should
`be modeled by a compliant tube.
`Unlike mechanical valves, tissue valves may show
`severe damageas a result of the wear test and may
`becomedysfunctional before completion of the experi-
`ment. Possible damage to tissue valves include tissue
`tear, holes, delamination, abrasion, prolapse and stent
`fracture (5,8,9). Table I lists commonly observed dam-
`age in tissue valves, their cause, andcritical test para-
`meters that influence the damage. Wear in mechanical
`valvesis also includedasa reference.
`Tissue tear is typically caused by hightensile stress-
`es. Finite element modeling (6) indicates that the high-
`est tensile stresses are in the region of the commissures
`150
`
`27mm SJM
`
`8
`
`Force/N—>
`
`is
`
`Ap/mmHg—> -100
`
`0,00
`
`0,04
`Time /s —>
`
`0,08
`
`Figure 10: Integral loading force and pressure difference across
`valve for a 27 mm SJM valveat a test rate of 1000 per min,
`but at a preselected pressure difference of 120 mmHg (FDA-
`condition) across the valve.
`
`PAGE 5 OF 7
`
`

`

`156 Durability/weartesting of heart valve substitutes
`H.Reul, K. Potthast
`
`J Heart Valve Dis
`Vol. 7. No. 2
`March 1998
`
`
`
`Table I: Wear and damageobservedin cardiac valve prostheses during accelerated weartesting.
`
`Valve type
`Wear/damage
`Cause
`Critical physical
`Critical test
`
`parameters
`parameter
`
`Mechanical _
`Sliding wear
`Sliding motion
`Range of motion
`Full opening and heart
`
`valve
`closing
`
`Impact wear
`Inertia at impact
`Impactforce
`Ap, dp/dt, impact force
`
`
`
`
`Stented tissue valve__Tissuetearat Hightensile stresses Pressure force at Ap
`
`commissure closure
`
`
`Contact between
`Tissue deflection
`Tissue abrasion at
`commissure
`during valve opening
`tissue and stent during
`
`during valve opening
`Tissue deflection
`Contact between
`Abrasion at the base
`Ap
`
`at closure
`of the stent
`. leaflets and stent
`
`Maximum opening
`
`Delamination of
`Tissue deflection
`High bending
`tissue
`stresses, tissue
`during cardiac cycle
`
`buckling
`
`Ap, maximum opening
`
`Stentless tissue valve
`
`Pressure force at
`Tissue tear at
`AP, Piny aortic wall
`commissure
`compliance
`closure, aortic wall
`
`motion
`
`Hightensile stresses
`
`Abrasion at the base
`Contact between
`AP, Pim aortic wall
`leaflets and inflow
`closure, aortic wall
`compliance
`of cusp
`
`rim
`motion
`
`Leaflet deflection at
`
`Tissue deflection
`Delamination of
`High bendingstresses,
`AP, Piny Maximum
`tissue
`opening,aortic wall
`during cardiac cycle
`tissue buckling
`compliance
`Axial and radial
`Tearin aortic wall
`Hightensile stresses
`AP, Piny
`wall compliance
`pressure loading
`AP: back-pressureacrosstheleaflets; p,,,: transmural pressure acrosstheaortic wall.
`
`in response to the pressure loading at closure. It is
`therefore important to control the closing pressure
`accurately. In the case of stentless valves, the stresses in
`the leaflets may be redistributed by the flexible aortic
`wall which mayactas a shock absorber. Proper model-
`ing of the aortic wall compliance and control of the
`transmural pressure gradients should therefore be con-
`sidered whentesting stentless valves.
`tissue
`Tissue abrasion is caused by rubbing of
`against the stent, stent coveror tissue reinforcement.It
`most likely occurs close to the commissuresor at the
`base of the cusps where excursion ofthe tissue is max-
`imum during the cardiac cycle. Maximum deformation
`of the leaflets in the cusp region occurs at maximum
`back-pressure, while maximum deformation of the
`leaflets at the commissure occurs during full opening.
`Thus, control of both peak back-pressure and maxi-
`mum leaflet opening is important when testing for
`leaflet abrasion. It should be noted that the high cycle
`
`rate in accelerated weartesters may notcreate the same
`leaflet deflection or stresses as in a real-time pulse
`duplicator. Vesely et al. (10) pointed out that porcine
`tissue is viscoelastic,i.e. the stress-strain relationship is
`time-dependent.
`Delamination in the tissue is due to internal shear
`stresses. Vesely and Boughner (7) have shown that
`porcine tissue loses someofits naturalability to shear
`when tanned. As a result, bending of tanned tissue
`leads to high internal shear stresses, and potential
`buckling of the tissue (8). Delamination is typically
`‘observed in the hinge area of porcine valves where the
`tissueis relatively thick.
`Asindicated in TableI, critical test parameters for the
`testing of tissue valves include the maximum back-
`pressure and the maximum valve opening. In some
`commercially available durability testers,
`the back-
`pressureis controlled by a throttle in the flow loop that
`adjusts the bypass flow during closure. As pointed out
`
`PAGE 6 OF 7
`
`PAGE 6 OF 7
`
`

`

`Durability/wear testing of heart valve substitutes
`H. Reul, K. Potthast
`
`157
`
`References
`
`1. Reul H, Eichler M, Potthast K, Schmitz C, Rau G. In
`vitro testing of heart valve wearoutside of the manu-
`facturer’s laboratory-requirements and controversies.
`J Heart Valve Dis 1996;5(Suppl. 1):S97-S104
`2. RauG, Reul H, Eichler M, Schreck S, Wieting DW. The
`effect of left ventricular dp/dt on the in vitro dynam-
`ics of Bjork-Shiley convexo-concave mitral valves. J
`Heart Valve Dis 1995;4(Suppl. I):S17-S20
`3. Schreck S, Inderbitzen R, Chien H,et al. Dynamics of
`Bj6rk-Shiley convexo-concave mitral valves in sheep.
`J Heart Valve Dis 1995;4(Suppl. I):S 21-S25
`4. Knott E, Reul H, Knoch M,Steinseifer U, Rau G. In
`vitro comparison of aortic heart valve prostheses. J
`Thorac Cardiovasc Surg 1988;96(6):S952-5961
`5. Walley VM, Keon WJ. Patterns of failure in Ionescu-
`Shiley bovine pericardial bioprosthetic valves. J Tho-
`rac Cardiovasc Surg 1987;93:S925-S933
`6. Krucinski S, Vesely I, Dokainish MA, Campbell G.
`Numerical simulationof leaflet flexure in bioprosthet-
`ic valves mountedin rigid and expansile stents. J Bio-
`mech 1993;26(8):S929-S943
`7. Vesely I, Boughner DR. Analysis of the bending
`behaviourof porcine xenograftleaflets and of natural
`aortic valve material. Bendingstiffness, neutral axis
`and
`shear
`measurements.
`J
`Biomech
`1989;22(6/7):S655-S671
`8. Vesely I., Boughner DR, Song T, Tissue buckling as a
`mechanism of bioprosthetic valve failure. Ann Thorac
`Surg 1988;46:5302-S308
`9. Thiene G, Bortolotti U, Valente M,et al. Modeoffail-
`ure of the Hancock pericardial xenograft. Am J Car-
`diovasc Surg 1989;63:S129-S133
`10. Vesely I, Boughner DR, Leeson-Dietrich J. Biopros-
`thetic valve tissue viscoelasticity:
`Implications on
`accelerated pulse duplicator testing. Ann Thorac Surg
`1995;60(2):S379-S382
`
`] Heart Valve Dis
`Vol. 7. No. 2
`March 1998
`
`earlier, tissue valves may experience notable damage
`during the experiments which often leads to increased
`backflow leakage. To maintain the required peak back-
`pressure, the bypass resistance has to be increased to
`compensate for leakage through the test valves.It is
`therefore desirable to have individualthrottles for each
`valve in order to maintain the same back-pressure for
`all valves throughoutthe duration of the experiment.
`The second importanttest parameteris the degreeof
`opening of tissue valves. According to the regulatory
`guidelines, full opening should be achieved in the
`durability tester.
`‘Full opening’
`is somewhat
`ill-
`defined, as the degree of opening is a function of the
`flow rate. At Baxter Laboratories, for instance, the
`valve is first tested in a pulse duplicator at a cardiac
`output of 51/min. The degree of opening during peak
`forward flow is recorded with a video camera posi-
`tioned along the axis of the valve. When the valveis
`placedin the durability tester, the stroke of the actuator
`is adjusted to reproduce the same degree of opening as
`in the pulse duplicator. Video images of the valve in
`both systemsare used to verify proper tuning.
`In conclusion, the above discussion addresses some
`of the basic considerations for the testing of tissue
`valves. Because of the complex biomechanical proper-
`ties of tissue, it is unclear if and howthe accelerated
`cycling rate is modifying the stress andstrain distribu-
`tion in the valves (10). Furthermore, biochemical
`degradation andin situ host response are not consid-
`ered. Biochemical degradation or host overgrowth may
`modify the tissue properties and alter the motion and
`stress distribution in the tissue. Results of accelerated
`weartesting of tissue valves should therefore be
`reviewed in the context of the limitations of the test.
`Additional knowledge aboutthe viscoelastic behavior
`of fresh and aged tissue will be necessary to further
`refine the accelerated wear testing of tissue valves.
`
`Acknowledgments
`The authors thank Stefan G. Schreck PhD, Director,
`Engineering Research and Standards at Baxter Health-
`care Corporation, Irvine, CA and Carlos Rios from the
`same group for their most valuable contribution to the
`discussion on specific aspects for accelerated testing of
`bioprostheses.
`
`PAGE 7 OF 7
`
`PAGE 7 OF 7
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket