throbber
EXHIBIT 2133
`
`EXHIBIT 2133
`
`IPR2017—003 17
`CONDITIONAL MOTION TO AMEND
`
`VALENCELL, INC.
`EXHIBIT 2133 — PAGE 1
`
`IPR2017-00317
`CONDITIONAL MOTION TO AMEND
`
`VALENCELL, INC.
`EXHIBIT 2133 - PAGE 1
`
`

`

`lntemat‘ional Conference on Control. Automation and Systems 2007
`Oct. [WEI], 2007 in COEX, Seoul, Korea
`
`Development of a wearable health monitoring device with motion artifact reduced algorithm
`(ICCAS 2007)
`
`Hyonyoung Han. Yunjoo Lee and Jung Kim
`
`I Department of Mechanical Engineering. Korea Advanced Institute of Science and Technology- Daejeon- Korea
`(Tel : r82-4E-869-327l; E-mail: rangel@kaist.ac.kr)
`1 Department of Mechanical Engineering. Korea Advanced Institute of Science and Technology. Daejeolt. Korea
`(Tel : +82-42-869-3270: E-mail: rosela@kaist.ac.kr )
`3 Department of Mechanical Engineering. Korea Advanced Institute of Science and Technology. Daejcon, Korea
`( Tel : +8242-869-323 l; E-mai|:.lungkim@kaist.ac.kr ]
`
`Abstract: In this paper. a real-time. wearable and motion artifact reduced health monitoring device is represented. A
`finger band, wearable health monitoring device,
`is consists of phot‘oplethysmography {PPG} sensor, 3—axis
`accelerometer, microproeessor and wireless module. The PPG sensor acquire distorted heart beat signal due to motion
`artifact. The linger movements are detected using the accelerometer, and major motion directions causing of the noise
`are researched by comparing each directional motion signals and distorted PPG signal. Two directional motions are
`significantly related to noise.
`therefore.
`these two directional active noise cancellation algorithm was applied to
`reconstruct the noise added heart beat signal. Low order t4th order} NLMS (normalized Least Mean Square} adaptive
`filter is employed for small size wearable device. The finger band device is experimented in daily body motion
`condition (1—3 Hz). and reduce distortion rate less than 5‘! by active noise cancellation algorithm. The motion artifact
`reduced finger band sensor can offer continuous health monitoring without daily motion artifact.
`
`Keywords: real-time. wearable, photoplethysmography. motion artifact. active noise cancellation
`
`LINTRODUCTION
`
`Vital signal measurement devices and techniques are
`important research fie|d5 for health care monitoring and
`an emergency health alarm system for patients and the
`aged. These person who has Weak vital signal. should
`monitor their bio-signal, such as pulsation. continuously
`and alarm to others when their signal is weaker. Among
`various vital signals. pulsation is adequate to health
`monitoring device since it can be measured easier and it
`needs simple devices than others.
`Pltotopletltysmography (ping) is a noninvasive heart
`beat. pulsation. measurement
`instrument. which has a
`potential to he developed into a portable device due to
`its relatively small size. Heart beat affeci
`to blood
`pressm'e and it' change the volume of the vesset and rate
`of blood flow. These volume and flow change can be
`detected using near red and infrared wave length light
`source and detector. Using these properties. pulsation,
`which is related to heart heat. can be measured by
`gathering detector‘s signal.
`
`The instrument has been developed for. monitoring
`
`and diagl‘wsis patient not only in the hospital but also at
`home. However. for portable and ubiquitous health care
`
`To reduce the problem many methods were researched
`this hindrance; one is sensorless approaches, which
`
`“Sing
`frequency component
`“mm the pulsation‘s
`[2. 3]
`frequency and time domain Feature analysis
`without
`any other sensors. Another
`is
`the sensor
`approach. which obtain pulsation signal by removing
`the motion noises. using body movement
`information
`me other motion "icclt-IISIII'JI'I
`SEESGFS- The "1051
`well-known method 3'30“ this approach is active "OISE
`cancellation algorithm With an adaptive 5'19" [4? 5]- BL”
`most ot‘their research is otT—line analysis. large program
`size which is not adequate to wearable and portable
`device. and experiments are not mostly wireless system.
`This paper F‘WWMFi a real-time. wireless 311d wearable
`device with emotion artifact reduction algorithm. The
`device has an applicable 51113” pragl'amming size for
`portable devices.
`
`2- SYSTEM SETUP
`2'1 Hardware Description ,
`
`system, the real-time measurement, wireless and motion
`
`artifacts problems should be worked out.
`
`In particular.
`
`motion artifact reducnon '5 the most challenging issue.
`
`Fig. [Finger hand sensor
`The wearable sensor device should be small and light
`
`978-89-950038-6-2—9856tltt17tliIS 3'§_"|CR(IS
`IPR2017—003 17
`CONDITIONAL MOTION TO AMEND
`
`1 58 1
`
`VALENCELL, INC.
`EXHIBIT 2133 — PAGE 2
`
`IPR2017-00317
`CONDITIONAL MOTION TO AMEND
`
`VALENCELL, INC.
`EXHIBIT 2133 - PAGE 2
`
`

`

`
`
`
`
`and attach to body tightly to reduce noise effect and feel
`comfortable
`to wear. Additionally, wireless
`communication device also reduces motion artifact by
`reducing effect of the data cable inertia.
`The finger band sensor, wearable PPG sensor, is
`attached to the finger base, as shown in Fig. 1. The size
`and weight is 25 x 30 mm2 and 16.8g, respectively. The
`PPG sensor is located on the inner layer of the band, and
`the accelerometer is on the circuit.
`The finger band device is divided into two parts; the
`sensor device and the host analysis computer. In detail,
`the sensor device part can be divided into three parts, as
`shown in Fig.2.
`
`
`(cid:1659)
`
`Fig. 21 Block diagram of photoplethysmography
`
`
`The first sensor part consists of a light source, a photo
`sensor for PPG and an accelerometer for motion
`detection. A 940 nm wavelength, a surface mounted
`type Infrared LED, and a photo-diode are used to
`acquire the PPG signal. Although a more detailed
`explanation follow in the next chapter, motion is
`considered to be directly related to noise, so to measure
`and analysis the body motion and a noise source, a
`3-axis accelerometer is used.
`Next part is the pre-signal processing, circuit part,
`which contains an analog signal processing part,
`amplifiers, filters, and analog to digital converters
`(ADC). Since the raw signal on the sensor signal
`(especially the AC component) is so small and distorted,
`signal processing is demanded before being sent to a
`communication system. The raw signal demands a low
`pass filter for reducing high frequency noise and high
`pass filter for rejecting a DC component to enhance the
`AC component. As filters, second order active analog
`high and low pass filters (Sallen-Key Filter) are used.
`Filtering signals are amplified to enhance and acquire
`discriminable signals by a thousand times.
`The last part is digital signal processing. High order
`filtering has good performance to extract wanted signal,
`but more number of components are required to increase
`filtering order. Therefore, digital filtering is employed
`to satisfy both circuit size and filtering performance.
`The filters are designed as a 0.5 – 3 Hz band pass filter,
`
`and totally fourth order analog active filter and digital
`filter are used in this signal processing.
`The following part is the communication system. To
`transmit obtained data from the sensor device to the host
`computer,
`the microcontroller converts data
`into
`communication language, an 8bit digital signal. Then it
`transmits the data through a wireless device, Bluetooth.
`It is a widely used device, especially for portable
`devices such as in MP3 players or cellular phones. The
`device has up to 30m communication distance, and size
`of the device is 20 x 18 mm2. The operation voltage is
`3.7 V.
`Labview software from National Instrument obtains
`transmitted data and display by graph on the host
`computer. Also, digital filtering and analysis algorithm
`are programmed
`
`2.2. Motion artifact reduction algorithm
`Active noise cancellation algorithm is used to reduce
`motion artifact. This is a signal filtering algorithm from
`a noise added pulsation signal and body motion signal
`to noiseless pulsation signal. The distorted PPG signal
`contains motion signal components and using adaptive
`filter, heart beat signal is extracted from various noise
`signal components.
`(cid:1659)
`
`(cid:1659)
`
`Fig. 3. Block diagram of the active noise cancellation algorithm
`
`Fig. 3 shows a block diagram of an active noise
`cancellation algorithm, which reconstructs a raw
`pulsation signal (sk) from the corrupted signal (dk),
`using measurable noise signal (xk). Here, PPG and body
`motion data correspond to dk and xk respectively. This
`research predominantly used 3-axis accelerometer
`signals (xk) for body motion data (nk).
`In this study, NLMS (Normalized Least Mean
`Square) adaptive filters were employed due to their fast
`processing and low order filter coefficients [6]. In the
`equation (1), w(n), the digital filter coefficient is
`computed from products of step size ((cid:541)(n)), input data
`(x(n)) and error data (e(n)). Instead of fixed step size in
`LMS algorithm,
`the step size are changed and
`normalized by the energy of input data vector. Step size,
`
`1582
`
`IPR2017-00317
`CONDITIONAL MOTION TO AMEND
`
`VALENCELL, INC.
`EXHIBIT 2133 - PAGE 3
`
`

`

`almost same number of reference signal.
`
`
`Table 1 Error between signals on various
` frequency conditions
`(cid:1659) Z.C. error (%)
`1.5 (Hz)
`2.0 (Hz)
`2.5 (Hz)
`
`25.00
`
`57.89
`
`74.36
`
`2.78
`
`5.26
`
`2.56
`
`Signal
`comparison
`Reference v.s.
`Measurement
`Reference v.s.
`Reconstructed
`
`
`Table 1 shows error rates between the numbers of
`each signal’s zero-crossing
`in various
`frequency
`conditions. Error rates increase from 25 % to 75 % as
`the frequency is higher. However reconstructed PPG
`signal case have low error rate, averagely lower than 5%.
`Besides it is represented lower than 4 count on beat per
`minute unit. Especially, in the 2.5 Hz motion condition,
`the error rate between reference and measured signal is
`74 %, but between reference and reconstructed signal is
`2.56 %. It means that the designed device and algorithm
`reduce motion artifacts efficiently.
`
`
`4. CONCLUSION
`
`ref.
`
`corrupted
`
`reconstructed
`
`
`
`4
`
`0
`
`0
`
`5
`
`10
`
`15
`
`Time(sec)
`Fig.5. Reference (bottom), corrupted (middle) and reconstructed (top)
`signal in 2.5 Hz hand waving conditions
`
`
`
` A
`
` real-time, wearable and wireless finger band sensor
`is designed with
`the motion artifact
`reduction
`algorithms. We obtain body motion data which is source
`of motion artifact using the accelerometer, and it is
`applied to active noise cancellation algorithm. As
`experiments, one directional hand motions which has
`difference frequency conditions are given. Fig.5 shows
`experimental signals;
`lower periodical signal
`is
`reference pulsation signal, middle complex signal is
`measured corrupted signal, and
`top smooth and
`periodical signal is reconstructed signal. On the graph
`also represent the performance is well-done. As a result,
`counting error of pulsation signal is reduced less than
`
`1583
`
`
`
`(1)
`
`(2)
`
`3. EXPERIMENT
`PPG signals are measured at not only moving left
`finger but also right finger in fixed pose, as a reference
`signal. Pre-experiment resulted that longitudinal axis
`and rotational direction to finger directional movements
`are strongly related to motion artifact. Therefore, two
`directional active noise cancellation algorithms are
`applied to reconstruct the pure PPG signal. The
`experiments are progressed in that various frequency
`movement conditions are experimented during 30 sec in
`finger longitudinally and 20-cm-long hand waving. The
`range of the frequency is similar to daily body motion,
`1.5~2.5 Hz. Zero-crossing (Z.C) peak counting method
`are used as an evaluation method.
`
`
`(cid:52)(cid:71)(cid:72)(cid:71)(cid:84)(cid:71)(cid:80)(cid:69)(cid:71)(cid:3)(cid:50)(cid:50)(cid:41) (cid:47)(cid:71)(cid:67)(cid:85)(cid:87)(cid:84)(cid:71)(cid:70)(cid:3)(cid:50)(cid:50)(cid:41) (cid:52)(cid:71)(cid:69)(cid:81)(cid:80)(cid:85)(cid:86)(cid:84)(cid:87)(cid:69)(cid:86)(cid:71)(cid:70)(cid:3)(cid:50)(cid:50)(cid:41)
`
`(cid:19)(cid:20)(cid:18)
`
`(cid:19)(cid:21)(cid:24)
`
`(cid:27)(cid:18)
`
`(cid:25)(cid:22)
`
`(cid:25)(cid:20)
`
`(cid:25)(cid:24)
`
`(cid:25)(cid:20)
`
`(cid:25)(cid:26)
`
`(cid:25)(cid:24)
`
`(cid:19)(cid:24)(cid:18)
`(cid:19)(cid:22)(cid:18)
`(cid:19)(cid:20)(cid:18)
`(cid:19)(cid:18)(cid:18)
`(cid:26)(cid:18)
`(cid:24)(cid:18)
`(cid:22)(cid:18)
`(cid:20)(cid:18)
`(cid:18)
`
`(cid:60)(cid:71)(cid:84)(cid:81)(cid:15)(cid:37)(cid:84)(cid:81)(cid:85)(cid:85)(cid:75)(cid:80)(cid:73)
`
`(cid:19)(cid:16)(cid:23)
`
`(cid:20)(cid:16)(cid:23)
`
`(cid:1659)
`
`(cid:20)(cid:16)(cid:18)
`(cid:40)(cid:84)(cid:71)(cid:83)(cid:87)(cid:71)(cid:80)(cid:69)(cid:91)(cid:3)(cid:10)(cid:42)(cid:92)(cid:11)
`Fig.4. Pulsation counts from the Z.C. of the signals
`
`
`
`The result, Fig.4, shows that the motion artifact
`reduction algorithm improve performance. The x, y axis
`represent
`frequency
`and
`represent number of
`zero-crossing during a minute, respectively. Lower than
`1Hz condition, zero-crossing error between reference
`and distorted measured PPG signal are nearly zero. The
`number of reference pulsation is 72 to 78 bpm (beat per
`minute), from normal healthy subject, and That of
`measured signal is 90 to 136 bpm in 1.5~2.5 Hz hand
`waving conditions, the error rates increase as movement
`frequency is higher. However, that of the reconstructed
`signal with motion artifact reduced algorithm detected
`
`(cid:541), are computed with the coefficients a, b and input data
`as equation (2), and the role of the coefficients are
`prevent the step size not to fluctuate excessively.
`Thorough this various step size condition, in NLMS
`cases, more flexible and stable signal processing is
`possible, which is appropriate for real-time and wireless
`sensors.
`w
`(
`
`n
`
`1)
`(cid:14) (cid:32)
`
`w
`
`
`
`( )n
`
`(cid:16)
`
`(cid:541)
`
`
`
`( )n
`
`(cid:32)
`
`a
`
`
`
`
`
`
`
`(cid:541)
`
`( ) ( ) ( )n n e nx
`b
`x x
`(cid:14) T
`
`IPR2017-00317
`CONDITIONAL MOTION TO AMEND
`
`VALENCELL, INC.
`EXHIBIT 2133 - PAGE 4
`
`

`

`
`
`
`
`5%.
`The experimental frequency condition is similar to a
`hand’s daily movement. For example as hand motions,
`running, walking [7], and hand gesture [8] have
`approximately 2 to 4 Hz frequency motion, and as
`finger motions, object exploration and texture scanning
`[9] have 0.8 to 2 Hz in table 2. But in hand writing,
`typing and tapping condition (4~7 Hz), they demand
`more high frequency artifact experiments and it remains
`future works.
`
`
`Table 2 Frequency of daily hand movement
`Motion
`Frequency (Hz)
`3
`2
`0 - 4
`0.8 – 1.6
`0.8 – 2.2
`
`Hand motion
`
`Finger motion
`
`4 - 7
`
`Run
`Walk
`Hand gesture
`Object
`exploration
`Texture scan
`Hand writing
`Typing
`Tapping
`
`variable step-size LMS filter," 2002.
`[5] H. Harry Asada, H. H. Jiang, and P. Gibbs,
`"Active noise cancellation using MEMS
`accelerometers for motion-tolerant wearable
`bio-sensors," Conf Proc IEEE Eng Med Biol
`Soc, vol. 3, pp. 2157-60, 2004.
`[6] S. M. K. a. D. Morgan, Active Noise Control
`Systems: Algorithms and DSP Implementations:
`JOHN WILEY \& Sons, Inc., 1995.
`[7] Akiko OGAWA, Yusuke KONISHI, Ryosuke
`SHIBAZAKI, “Identification of human activity
`modes with wearable sensors for autonomous
`human positioning system,”, Geoinformation
`forum Japan 2002, Student forum, 2002.
`[8] Y. X. a. F. Quek, "Hand Motion Gesture
`Frequency Properties and Multimodal Discouse
`Analysis," International Journal of Computer
`Vision, vol. 69, pp. 353-371, 2006.
`[9] E. Kunesch, F. Binkofski, and H. J. Freund,
`"Invariant
`temporal
`characteristics
`of
`manipulative hand movements," Exp Brain Res,
`vol. 78, pp. 539-46, 1989.
`
`
`Humans are moved more than remain fixed, relaxed
`pose on our daily life. Moreover, accidents and
`emergencies occur in movement condition. Therefore,
`wearable and portable health care device with motion
`artifact reduced algorithm techniques are demanded for
`continuous health monitoring. And the proposed device
`in this paper could have a potential to developing this
`portable and wearable device without daily motion
`artifact.
`
`
`ACKNOWLEDGMENT
`
`
`This work was supported by the IT R&D program of
`MIC/IITA. [2005-S-096-02, Development of wearable
`PC interface technology for the disabled ]
`
`
`REFERENCES
`
`
`J.G.Webster, Design of Pulse Oximeters, 1997.
`[1]
`[2] B. S. Kim and S. K. Yoo, "Motion artifact
`reduction
`in photoplethysmography using
`independent component analysis," IEEE Trans
`Biomed Eng, vol. 53, pp. 566-8, 2006.
`[3] Y. S. Yan, C. C. Poon, and Y. T. Zhang,
`"Reduction of motion artifact in pulse oximetry
`by smoothed pseudo Wigner-Ville distribution,"
`J Neuroengineering Rehabil, vol. 2, pp. 3, 2005.
`[4] K. W. Chan and Y. T. Zhang, "Adaptive
`reduction
`of motion
`artifact
`from
`photoplethysmographic
`recordings using a
`
`1584
`
`IPR2017-00317
`CONDITIONAL MOTION TO AMEND
`
`VALENCELL, INC.
`EXHIBIT 2133 - PAGE 5
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket