throbber
Journal of Pharmaceutical and Biomedical Analysis
`19 (1999) 877–882
`
`Analysis of benzalkonium chloride and its homologs: HPLC
`versus HPCE1
`
`Shelly J. Prince a,*, Hei-Jen McLaury b, Loyd V. Allen c, Phil McLaury d
`a Southwestern Oklahoma State Uni6ersity School of Pharmacy, 100 Campus Dri6e, Weatherford, OK 73096, USA
`b Parke-Da6is Pharmaceuticals, 170 Tabor Road, Morris Plains, NJ 07950, USA
`c Uni6ersity of Oklahoma College of Pharmacy, 1110 N. Stonewall, Oklahoma City, OK 73117, USA
`d No6artis Pharmaceutical Corporation, 59 Route 10, East Hano6er, NJ 07936, USA
`
`Received 18 November 1996; accepted 23 June 1998
`
`Abstract
`
`Benzalkonium chloride (BAK) is a mixture of alkylbenzyldimethylammonium chloride homologs with n-C12H25,
`n-C14H29, and n-C16H33 comprising a major portion of the alkyl groups present. An analytical method for BAK must
`differentiate and quantitate the homologs in the BAK mixture. Reversed-phase high performance liquid chromatog-
`raphy (HPLC) separates compounds based on their affinity for a nonpolar column, which is a direct correlation to
`the compounds’ polarity. High performance capillary electrophoresis (HPCE), however, separates compounds in an
`electric field according to their charge and size. The BAK homologs are suitable for separation by either of these
`methods because their polarity and sizes differ significantly. The HPLC method employed a mobile phase of 60%
`acetonitrile and 40% 0.1 M sodium acetate buffer pH 5 pumped at 1.0 ml min (cid:28) 1, a 4.6(cid:29)250 mm cyano column with
`5 mm packing, and UV detection at 254 nm. The HPCE method utilized a run buffer of 30% acetonitrile and 70%
`0.05 M sodium phosphate pH 3.06, a 50 mm(cid:29)20 cm open silica capillary, 7.5 kV electric field and UV detection at
`214 nm. Both HPLC and HPCE demonstrated good linearity in the range of 0.025 to 0.8 mg ml (cid:28) 1 with r 2 values of
`approximately 0.99. The HPLC method produced good separation of the homolog peaks with a total analysis time
`of 25 min. HPCE run time was less than 5 min and demonstrated good separation of the three homologs. The HPLC
`method, however, was superior to HPCE in the areas of sensitivity and precision. The HPLC has been extensively
`used in the routine quantitation and qualitation of benzalkonium chloride concentrations in various products;
`however, long analysis times make this method inefficient. The HPCE method produced comparable results to the
`HPLC method but with much shorter analysis times. An HPCE analysis method, as presented here, may prove to be
`a much more useful and efficient method for the analysis of benzalkonium chloride and its homologs. © 1999 Elsevier
`Science B.V. All rights reserved.
`
`Keywords: Benzalkonium chloride; High performance liquid chromatography; High performance capillary elec-
`trophoresis
`
`* Corresponding author. Tel.: (cid:27)1 580 7743105; fax: (cid:27)1 580 7747020; e-mail: princes@swosu.edu
`1 Presented at the 1997 AAPS Annual Meeting.
`
`0731-7085:99:$ - see front matter © 1999 Elsevier Science B.V. All rights reserved.
`PII S0731-7085(98)00187-3
`
`Innopharma EX1018, Page 1
`
`

`

`878
`
`S.J. Prince et al. :J. Pharm. Biomed. Anal. 19 (1999) 877–882
`
`1. Introduction
`
`Benzalkonium chloride (BAK) is a bactericidal
`antimicrobial agent used as a preservative in over
`65% of the ophthalmic products currently avail-
`able on the market [1]. It is used to preserve otic,
`nasal and parenteral formulations as well. It pos-
`sesses surfactant properties and is also used as a
`topical antiseptic and medical equipment disinfec-
`tant. BAK is an unusual compound in that it is
`composed of a mixture of straight chain homologs
`which possess different physical, chemical and
`microbiological properties. Therefore, the propor-
`tions of these homologs in the mixture determine
`its effectiveness as a preservative and disinfectant.
`The BAK is a mixture of alkylbenzyldimethyl-
`ammonium chlorides with the general formula
`[C6H5CH2N(CH3)2R]Cl where R(cid:30)n-C8H17
`to
`n-C19H39. The C12, C14 and C16 chains comprise
`the major portion of the alkyl mixture. The ho-
`mologs do not possess identical bactericidal activ-
`ity; therefore, the USP:NF sets forth guidelines
`for the content of Benzalkonium Chloride NF as
`follows [2]:
`
`On the anhydrous basis, the content of the
`n-C12H25 homolog is not less than 40%, and the
`content of the n-C14H29 homolog is not less
`than 20%, of
`the total alkylbenzyldimethyl-
`ammonium chloride content. The amounts of
`the n-C12H25 and n-C14H29 homolog compo-
`nents comprise together not less than 70% of
`the total alkylbenzyldimethylammonium chlo-
`ride content.
`
`In general, the C12 homolog is most effective
`against yeast and fungi, the C14 homolog against
`gram-positive bacteria and the C16 homolog
`against gram-negative bacteria [3].
`The BAK is a difficult compound to analyze
`due to its multicomponent nature. The assay used
`must be both quantitative and qualitative to be
`able to identify and distinguish the different ho-
`molog components and the level of each in the
`mixture. The method should be sensitive and effi-
`cient enough to detect all of the homologs present
`in a reasonable amount of time. High perfor-
`mance liquid chromatography (HPLC) has been
`
`used extensively and successfully to separate and
`quantitate the benzalkonium chloride homologs
`[2,4–14]. The analysis times, however, typically
`range from 15 to 30 min. These lengthy analysis
`times can be quite troublesome to those doing
`routine QA:QC checks of ophthalmic products
`containing BAK where there is a large number of
`samples. High
`performance
`capillary
`elec-
`trophoresis (HPCE) is an analytical method which
`separates compounds on the basis of their charge
`and size. This method will produce adequate sepa-
`ration of the homologs because of their incremen-
`tal sizes and because the quaternary ammonium
`molecules carry a positive charge, as shown by
`Althia et. al. in a study of the analysis of BAK
`and histamine acid phosphates mixture by using
`HPCE [15]. The purpose of this study was to
`compare the analysis of BAK using HPLC to
`analysis using HPCE.
`
`2. Materials and methods
`
`2.1. Sol6ents and chemicals
`
`The BAK was supplied as a 50% aqueous solu-
`tion from Spectrum (Gardena, CA). The C12, C14,
`and C16 homologs were purchased from Sigma
`(St. Louis, MO). The acetonitrile and sodium
`acetate were purchased from Fisher Scientific
`(Fair Lawn, NJ), and the sodium phosphate,
`monobasic was purchased from Mallinckrodt
`(Paris, KY). All other chemicals used in the ana-
`lytical methods were HPLC grade. Water was
`purified by using a Milli-Q water system from
`Millipore (Bedford, MA).
`The composition of
`the BAK sample from
`Spectrum Chemicals used for HPLC analysis was
`64.64% C12, 27.49% C14, and 7.86% C16 homolog.
`In comparison, the BAK sample prepared in the
`laboratory from the individual homologs used for
`HPCE analysis consisted of 41.84% C12, 32.65%
`C14, and 25.51% C16 homolog. A sample concen-
`tration of 0.1 mg ml(cid:28) 1 was used for determina-
`tion of precision, and the concentrations for
`determination of linearity were 0.025, 0.05, 0.1,
`0.2, and 0.4 mg ml(cid:28) 1, with the exception of the
`BAK sample used in the HPLC method. The
`
`Innopharma EX1018, Page 2
`
`

`

`S.J. Prince et al. :J. Pharm. Biomed. Anal. 19 (1999) 877–882
`
`879
`
`BAK concentration used for HPLC precision was
`0.2 mg ml (cid:28) 1, and the concentrations used for
`linearity were 0.05, 0.1, 0.2, 0.4, and 0.8 mg ml(cid:28) 1
`due to the low concentration of the C16 homolog
`in the BAK mixture.
`
`2.2. Equipment
`
`The HPLC system consisted of a Waters M-
`6000A Solvent Delivery System, Micromeritics
`728 Autosampler, Rheodyne 7010 Electrically Ac-
`tuated Valve, Waters Model 441 Fixed Wave-
`length Absorbance Detector and Waters 745 Data
`Module. A Beckman P:ACE System 2100 was
`used for the HPCE analysis and Beckman System
`Gold software was used for data collection.
`
`2.3. Experimental conditions
`
`The column used for the HPLC method was a
`250(cid:29)4.6 mm i.d. stainless steel column filled with
`Bakerbond cyano 5 mm packing and was kept at
`room temperature. The mobile phase consisted of
`acetonitrile–sodium acetate (pH 5.0; 0.1 M)
`(60:40, v:v), and the pH of the buffer was ad-
`justed by adding glacial acetic acid. Elution was
`performed at a flow rate of 1.0 ml min(cid:28) 1 and the
`injection volume was 20 ml. The absorbance was
`monitored at 254 nm.
`The capillary used for the HPCE method was a
`Beckman P:ACE System:eCAP 50 mm i.d.(cid:29)26
`cm open silica capillary with a 20 cm effective
`length maintained at 20°C. The run buffer was
`acetonitrile-sodium phosphate, monobasic (pH
`3.06 adjusted using phosphoric acid; 0.05M)
`(30:70, v:v), and the applied voltage was 7.5 kV.
`All samples were injected into the silica capillary
`by 5 s of pressure at 0.5 c, a volume of approxi-
`mately 12.5 nl, and detected by absorbance at 214
`nm.
`
`3. Results and discussion
`
`Both the HPLC and HPCE methods used for
`analysis of benzalkonium chloride solutions pro-
`duced adequate resolution of
`the three major
`homolog peaks corresponding to the C12, C14, and
`
`C16 homologs (Figs. 1 and 2). The HPLC method
`produced an analysis time of 25 min for all three
`homologs, whereas the HPCE method produced
`an analysis time of only 5 min. The sensitivities of
`the HPLC and HPCE methods were approxi-
`mately 2 and 5 mg ml (cid:28) 1, respectively. The HPCE
`method was less sensitive due to a shorter path
`length in the detector and a smaller injection
`volume (12.5 nl for HPCE vs. 20 ml for HPLC).
`
`Fig. 1. Chromatogram of benzalkonium chloride 0.2 mg ml (cid:28) 1.
`
`Innopharma EX1018, Page 3
`
`

`

`880
`
`S.J. Prince et al. :J. Pharm. Biomed. Anal. 19 (1999) 877–882
`
`Fig. 2. Electropherogram of benzalkonium chloride 0.1 mg ml (cid:28) 1.
`
`The sensitivity for both methods is adequate,
`however, because the minimum concentration of
`BAK in ophthalmic formulations is 40 mg ml(cid:28) 1
`(0.004%).
`For both HPLC and HPCE systems, the preci-
`sion was determined by the intraday variation of
`five injections of each homolog and BAK mixture
`and the interday variation for 5 days (Table 1).
`The precision for HPLC was higher than for
`HPCE in most instances. This was probably due
`to problems encountered with capillary degenera-
`tion, which was somewhat resolved by recondi-
`tioning the column after every five sample runs.
`The day 5 HPCE results for the C16 homolog and
`BAK samples are not reported because the capil-
`lary became unusable and reconditioning efforts
`failed; therefore, the data could not be obtained.
`This problem may have been due to adsorption of
`the C16 homolog to the silica capillary. The main
`causes of adsorption to the silica capillary are
`ionic interactions between cationic solutes and the
`negatively charged wall and hydrophobic interac-
`
`tions [16]. Because the C16 homolog is more hy-
`it
`drophobic than the C12 or C14 homologs,
`exhibited a higher degree of hydrophobic interac-
`tions. Furthermore, all benzalkonium chloride ho-
`mologs are large cations, so all demonstrated
`ionic interactions with the negatively charged wall
`to various degrees. Adsorption to the silica capil-
`lary walls can be minimized by increasing buffer
`concentration which will reduce solute interac-
`tions, by using buffers at the extremes of pH and
`also by coating the capillary wall [16]. The 0.05 M
`sodium phosphate buffer system used in this study
`was at low pH (3.06), but a higher buffer concen-
`tration exceeded the maximum suggested operat-
`ing current.
`The results for the standard curves obtained by
`the HPLC and HPCE methods are given below:
`HPLC
`C12 Homolog: peak area(cid:30)5132(cid:27)314897 con-
`centration (r 2(cid:30)0.9988)
`C14 Homolog: peak area(cid:30)5230(cid:27)330413 con-
`centration (r 2(cid:30)0.9991)
`
`Innopharma EX1018, Page 4
`
`

`

`S.J. Prince et al. :J. Pharm. Biomed. Anal. 19 (1999) 877–882
`
`881
`
`Table 1
`Intraday and interday variation for benzalkonium chloride and its homologs using HPLC and HPCE
`
`Day 1
`
`HPLC
`2.38a
`3.31
`4.87
`2.59
`
`HPCE
`11.06
`0.96
`6.63
`4.67
`
`C12
`C14
`C16
`BAK
`
`C12
`C14
`C16
`BAK
`
`a RSD%, n(cid:30)5.
`
`Day 2
`
`Day 3
`
`Day 4
`
`Day 5
`
`Interday
`
`2.26
`2.77
`3.73
`1.97
`
`12.46
`1.82
`6.47
`2.30
`
`2.78
`5.91
`2.99
`0.99
`
`3.37
`4.02
`6.55
`3.40
`
`1.74
`2.92
`2.53
`3.30
`
`8.10
`5.19
`9.20
`5.90
`
`3.85
`3.86
`3.34
`1.13
`
`8.30
`9.37
`—
`—
`
`0.50
`1.16
`1.62
`0.93
`
`3.57
`19.86
`9.01
`21.01
`
`C16 Homolog: peak area(cid:30)6768(cid:27)310245 con-
`centration (r 2(cid:30)0.9942)
`BAK mixture: peak area(cid:30)2874(cid:27)303956 con-
`centration (r 2(cid:30)1.0000)
`
`HPCE
`C12 Homolog: peak area(cid:30) (cid:28)6.371(cid:29)10 (cid:28) 3(cid:27)
`6.205(cid:29)10(cid:28) 3 concentration (r 2(cid:30)0.9995)
`C14 Homolog: peak area(cid:30) (cid:28)1.795(cid:29)10 (cid:28) 3(cid:27)
`5.062(cid:29)10(cid:28) 3 concentration (r 2(cid:30)0.9981)
`C16 Homolog: peak area(cid:30) (cid:28)2.654(cid:29)10 (cid:28) 2(cid:27)
`5.400(cid:29)10(cid:28) 3 concentration (r 2(cid:30)0.9960)
`BAK mixture: peak area(cid:30)4.537(cid:29)10 (cid:28) 3(cid:27)
`5.125(cid:29)10(cid:28) 3 concentration (r 2(cid:30)0.9997)
`The linearity for both the HPLC and HPCE
`methods is comparable; therefore, both methods
`are useful for the accurate quantitation of benzal-
`konium chloride as well as the individual ho-
`mologs in the mixture.
`
`4. Conclusion
`
`A high performance capillary electrophoresis
`assay for qualitation and quantitation of benzal-
`konium chloride and its homologs has been inves-
`tigated. The primary advantage of the HPCE
`method, when compared to HPLC,
`is a much
`shorter analysis time. The linearity and sensitivity
`of both methods is comparable; however, the
`precision for the HPLC method was better than
`
`for HPCE. This problem may have been due in
`part to sorption of the C16 homolog to the capil-
`lary wall.
`In summary, HPCE has the potential to be a
`more efficient and useful method for BAK analy-
`sis than the established HPLC methods.
`
`References
`
`[1] G. Hecht, in: A.R. Gennaro (Ed.), Remington: The Sci-
`ence and Practice of Pharmacy, Mack, PA, 1995, pp.
`1563–1573.
`[2] United States Pharmacopeia 23:National Formulary 18,
`United States Pharmacopeial Convention, Inc., Mary-
`land, 1995, pp. 2218–2219.
`[3] J.J. Merianos, in: S.S. Block (Ed.), Disinfection, Steriliza-
`tion, and Preservation, Lea and Febiger, PA, 1991, pp.
`225–255.
`[4] M.R. Euerby, J. Clin. Hospital Pharm. 10 (1985) 73–77.
`[5] A. Bettero, A. Semenzato, C.A. Benassi, J. Chromatogr.
`507 (1990) 403–407.
`[6] L.J. Cohn, V.J. Greely, D.L. Tibbetts, J. Chromatogr.
`321 (1985) 401–405.
`[7] T.Y. Fan, G.M. Wall, J. Pharm. Sci. 82 (1993) 1172–
`1174.
`[8] A. Nakae, K. Kunihiro, G. Muto, J. Chromatogr. 134
`(1977) 459–466.
`[9] D.F. Marsh, L.T. Takahashi, J. Pharm. Sci. 72 (1983)
`521–525.
`[10] G. Ambrus, L.T. Takahashi, P.A. Marty, J. Pharm. Sci.
`76 (1987) 174–176.
`[11] R.C. Meyer, J. Pharm. Sci. 69 (1980) 1148–1150.
`[12] L. Elrod Jr., T.G. Golich, J.A. Morley, J. Chromatogr.
`625 (1992) 362–367.
`
`Innopharma EX1018, Page 5
`
`

`

`882
`
`S.J. Prince et al. :J. Pharm. Biomed. Anal. 19 (1999) 877–882
`
`[13] A. Gomez-Gomar, M.M. Gonzalez-Aubert, J. Garces-
`Torrents, J. Costa-Segarra, J. Pharm. Biomed. Anal. 8
`(1990) 871–876.
`[14] G. Parhizkari, R.B. Miller, C. Chen, J. Liq. Chromatogr.
`18 (1995) 553–563.
`
`[15] K.D. Altria, J. Elgey, J.S. Howells, J. Chromatogr. B 686
`(1996) 111–117.
`[16] D.N. Heiger, High Performance Capillary Electrophore-
`sis—An Introduction, Hewlett-Packard, France, 1992,
`pp. 32–35.
`
`.
`
`Innopharma EX1018, Page 6
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket