|NOvent delivery system
`Operation and Maintenance Manual
`CGA Variant
`
`om —
`
`001
`
`

`
`
`
`
`
` 
`User Responsibility
`
`This Product will perform in conformity with the description thereof contained in
`           
`this operating manual and accompanying labels and/or inserts, when
`        
`assembled, operated, maintained, and repaired in accordance with the
`        
`instructions provided. This Product must be checked periodically. A defective
`         
`Product should not be used. Parts that are broken, missing plainly worn,
`           
`distorted, or contaminated should be replaced immediately. Should repair or
`         
`replacement become necessary, Datex-Ohmeda recommends that a telephonic
`       
`or written request for service advice be made to the nearest Datex-Ohmeda
`           
`Field Service Support center. This Product or any of its parts should not be
`             
`repaired other than in accordance with written instructions provided by Datex-
`          
`Ohmeda and by Datex-Ohmeda trained personnel. The Product must not be
`          
`altered without the prior written approval of Datex-Ohmeda’s Quality Assurance
`         
`Department. The user of this Product shall have the sole responsibility for any
`            
`malfunction which results from improper use, faulty maintenance, improper
`        
`        
`repair, damage, or alteration by anyone other than Datex-Ohmeda.
`
`A WARNING
`
`U.S. Federal and Canadian law restrict this device to sale by or on the or-
`               
`der of a licensed medical practitioner. Outside the U.S.A. and Canada,
`          
`check local laws for any restriction that may apply.
`        
`
`Inhaled Nitric Oxide mixtures must be handled and stored in compliance
`          
`with federal, state and local regulations.
`     
`
`         
`Datex-Ohmeda products have unit serial numbers with coded logic which
`indicates a product group code, the year of manufacture and a sequential unit
`            
`number for identification.
`  
`
`AAA A 12345
`  
`
`P This alpha character indicates the year of product
`
`       
`manufacture and when the serial number was
`      
`assigned; “B” = 1998, “C” = 1999, “D” = 2000, etc.
`          
`‘‘I’’ and “O” are not used.
`     
`
`02/08/00
`
`
`1 605-0014-000
`
`
`002
`
`002
`
`

`
`Contents
`
`1/|nt'-Qduction
`
`Definitions and Abbreviations
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`Symbols used in this manual or on the system .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.1—3
`
`.1—4
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. .2—2
`
`2/Genera] |nfo|-[nation
`
`Overview .
`
`Configuration options .
`
`.
`
`Main component views
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.2—3
`
`.2—4
`
`. 2-?
`
`Display and user controls .
`
`Theory of operation .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`NO Flow from the cylinder to the patient
`breathing circuit.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`
`.
`
`Effect of inserting the |NOvent into a
`breathing circuit.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.2—11
`
`. 2-11
`
`.
`
`.
`
`.2—14
`
`.
`
`.3—2
`
`NO delivery system connections .
`
`Mounting .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.3—2
`
`.3—3
`
`System connections .
`
`Electrical connections
`
`Setup menu .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.3—8
`
`.3—1 0
`
`.3—10
`
`Cylinder concentration .
`
`Hourmeter display .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`To set the alarm loudness .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.3—10
`
`.3—10
`
`.3—11
`
`To set the display brightness .
`
`.
`
`.
`
`.
`
`.
`
`4/Cafibration
`
`Calibration information .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`NO, NO; And O2 Sensor low range
`calibration (daily, room air) .
`.
`.
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. .4—2
`
`.
`
`.
`
`.4—3
`
`.4—3
`
`Procedure .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`NO, N02 And O2 Sensor high range calibration .
`
`Procedure .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. .4—5
`
`.
`
`.4—5
`
`1605-0014-000
`
`02KJ8lU0
`
`003
`
`

`
`|NOvent delivery system
`
`5/P;-e.|_|se
`P|'0CedU|'e5
`
`1. Initial connections and leak test .
`.
`.
`2. System purge and performance test
`
`.
`
`.
`.
`
`.
`.
`
`.
`.
`
`.
`.
`
`.
`.
`
`.
`.
`
`.
`.
`
`.
`.
`
`.
`.
`
`.
`.
`
`.
`.
`
`.
`.
`
`.
`.
`
`.
`.
`
`.
`.
`
`.
`.
`
`. 5-2
`. 5-3
`
`3. Manual N0 delivery system purge and performance test
`
`. 5-5
`
`3/operation
`
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`Before operation .
`Connection to an ICU ventilator circuit .
`
`.
`.
`
`.
`.
`
`Connection to a high frequency oscillatory
`ventilator circuit
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`
`Connection to a circle anesthesia system .
`
`.
`.
`
`.
`
`.
`
`.
`.
`
`.
`
`.
`
`.
`.
`
`.
`
`.
`
`.
`.
`
`.
`
`.
`
`.
`.
`
`.
`
`.
`
`.
`.
`
`.
`
`.
`
`.
`.
`
`.
`
`.
`
`.
`.
`
`.
`
`.
`
`.
`.
`
`.
`
`. 6-2
`. 6-4
`
`. 6-8
`
`. 6-11
`
`. 6-14
`
`Connection to a transport ventilator circuit .
`
`Connection to bagging systems .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. 6-16
`
`. 6-23
`
`Operation .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`Setting or changing the delivered NO concentration. . 6-23
`
`During operation .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. 6-25
`
`. .. 6-25
`
`Pausing NO flow .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`Changing NO therapy cylinders and purging the
`regulator assembly .
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`
`Electronic delivery shutdown .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. 6-26
`
`. 6-28
`
`. 6-29
`
`Monitoring the environment .
`
`.
`
`.
`
`.
`
`7/A|a[-ms
`
`General alarm information .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`High, medium and low priority alarms .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. .. 7-2
`
`.
`
`.
`
`.
`
`.
`
`. 7-2
`
`. 7-3
`
`Alarm silencing .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`Changing the alarm settings .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. 7-4
`
`. 7-5
`
`Alarm adjustment range .
`
`.
`
`Alarm adjustment procedure .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`Clearing resolved alarm messages .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. 7-5
`
`. 7-7
`
`. 7-9
`
`Alarm message table .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`Manual N0 delivery system description .
`3/Manua| N0 de|iVe|-y
`system Manual N0 delivery system connections .
`
`Manual N0 delivery system operation .
`
`.
`
`.
`
`.
`.
`
`.
`
`.
`.
`
`.
`
`.
`.
`
`.
`
`9/T;-oub|e3hooting
`
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`Troubleshooting procedure .
`If the system fails to operate properly .
`
`.
`.
`
`.
`
`.
`.
`
`.
`.
`
`.
`
`.
`.
`
`.
`.
`
`.
`
`.
`.
`
`.
`.
`
`.
`
`.
`.
`
`.
`.
`
`.
`
`.
`.
`
`.
`.
`
`.
`
`.
`.
`
`.
`.
`
`.
`
`.
`.
`
`.
`.
`
`.
`
`.
`.
`
`.
`.
`
`.
`
`.
`.
`
`.
`.
`
`.
`
`.
`.
`
`.
`.
`
`.
`
`.
`.
`
`. 8-2
`. 8-3
`
`. 8-4
`
`. 9-2
`. 9-2
`
`02.-"08.-"00
`
`1605-0014-000
`
`004
`
`

`
`Troubleshooting guide index .
`
`Troubleshooting guide .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`If the problem can’t be corrected using the
`above suggestions .
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`It the |NOvent delivery system must be returned
`for servicing .
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. .9-3
`
`. .9-4
`
`. .9-13
`
`. .9-13
`
`.
`
`.
`
`. .10-2
`
`10/Maintenance
`
`User maintenance schedule .
`
`.
`
`.
`
`.
`
`|NOvent delivery system cleaning .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`Injector Module sterilizing and disinfecting .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. .10-4
`
`. .10-5
`
`. .10-5
`
`Autoclave sterilizing .
`
`.
`
`High level disinfecting .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`Patient Circuit Adapters & Sample Line .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. .10-5
`
`. .10-6
`
`. .10-7
`
`Monthly System Checkout .
`
`.
`
`.
`
`.
`
`1. Initial connections leak tests .
`
`.
`
`.
`
`2. Purge and system alarms tests.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`3. Calibration and monitoring alanns .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`4. |NOvent delivery system performance .
`
`5. Manual NO delivery system purge and
`performance .
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. .10-7
`
`. .10-10
`
`. .10-12
`
`. .10-14
`
`. .10-15
`
`. .10-17
`
`Emptying the Fluid Trap Bottle .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`Replacing the Fluid Trap Filter Cartridge .
`
`Replacing the N02 Sensor .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. .10-18
`
`. .10-19
`
`. .10-20
`
`Replacing the N0 Sensor .
`
`Replacing the O2 Sensor .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`Cleaning or replacing the Cooling Fan Filter .
`
`.
`
`.
`
`Fuse replacement and line voltage selection .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. .10-22
`
`. .1023
`
`. .10-24
`
`Replacing the high pressure hose CGA626 connector tip . .1026
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. .11-2
`
`1 1/Parts and
`Accessories
`
`Standard accessories
`
`Optional accessories .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. .11-2
`
`. .11-3
`
`12/Ap
`pendix A -
`Specifications
`
`Replaceable parts .
`
`.
`
`Functional
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`Ventilator compatibility .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. .12-2
`
`. .12-2
`
`1605-0014-000
`
`02208100
`
`005
`
`

`
`INOvent delivery system
`
`Injector Module .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`Manual N0 delivery system .
`
`NO delivery .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. .. 12-3
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. 12-3
`
`. 12-4
`
`. 12-4
`
`Maximum NO delivery .
`
`Gas monitoring .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`Calibration Gas cylinders .
`
`Calibration Gas regulator .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. 12-5
`
`. 12-5
`
`. 12-5
`
`. 12-5
`
`NO Delivery shutdown .
`
`Physical .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`Dimensions .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. 12-6
`
`. .. 12-6
`
`.
`
`.
`
`. 12-?
`
`Environmental .
`
`Electrical .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`13/Appendix B -
`Cylinder |nf0|'mati0n
`
`Preparing the NO therapy gas cylinder for use .
`Replacing the N0 therapy gas cylinder
`.
`.
`.
`.
`.
`.
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`.
`
`.
`
`.
`
`.
`
`.
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. 12-8
`
`.
`.
`
`.
`
`.
`.
`
`.
`
`.
`.
`
`.
`
`.
`.
`
`.
`
`.
`.
`
`.
`
`.
`.
`
`.
`
`. 13-2
`. 13-3
`
`. 13-4
`
`NO therapy gas cylinder leak check .
`
`Cylinder information .
`
`Warnings .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. 13-5
`
`. 13-6
`
`. 14-1
`
`C - Pre-
`Use Procedures Card
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`Description .
`15/Appendix D -
`Battery operation information .
`T|'a|'|3P0|'t |NoVe|'|t
`defivery system Setup .
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`
`.
`.
`.
`
`Regulator connection diagrams .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`.
`.
`
`.
`
`.
`
`.
`.
`.
`
`.
`
`.
`
`.
`.
`.
`
`.
`
`.
`
`.
`.
`.
`
`.
`
`.
`
`.
`.
`.
`
`.
`
`.
`
`.
`.
`.
`
`.
`
`.
`
`.
`.
`.
`
`.
`
`.
`
`.
`.
`.
`
`.
`
`.
`
`.
`.
`.
`
`.
`
`.
`
`.
`.
`.
`
`.
`
`.
`
`.
`.
`.
`
`.
`
`.
`
`.
`.
`.
`
`.
`
`.
`
`.
`.
`.
`
`.
`
`.
`
`.
`.
`.
`
`.
`
`.
`
`. 15-2
`.
`.
`. 15-2
`.
`.
`. .. 15-4
`
`.
`
`.
`
`.
`
`.
`
`. 15-4
`
`. 15-5
`
`Operation .
`
`.
`
`.
`
`.
`
`Maintenance .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`User maintenance schedule .
`
`.
`
`.
`
`Replaceable parts and accessories .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. 15-6
`
`. 15-6
`
`. 15-?
`
`. 15-8
`
`Specifications .
`
`.
`
`.
`
`.
`
`Dimensions .
`
`Electrical .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. .. 15-8
`
`.
`
`.
`
`. 15-8
`
`02.-"08.-"00
`
`1605-0014-000
`
`006
`
`

`
`Cart mount .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. .16-2
`
`. .16-4
`
`16/Appendix E -
`Optional Mounting
`
`Shelf or table mount .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`Optional Cylinder Mount Fiegulator and
`Hose assembly .
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`In ’Section 3fSetup’ of this manual .
`
`Cylinder concentration .
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. .16-4
`
`.
`
`.
`
`.
`
`.
`
`. .17-2
`
`. .17-2
`
`17/Appendix F -
`Alternate Cylinder
`Concentrations
`
`In ’Section 5!Pre-Use Procedures’ of this manual .
`
`2. System purge and performance test .
`
`.
`
`3. Manual NO delivery system purge and
`performance test .
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`
`In ’Section 10fMaintenance’ of this manual
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. .17-3
`
`. 17- 3
`
`. .17-3
`
`. .17-4
`
`. .17-4
`
`Monthly System Checkout .
`
`.
`
`.
`
`.
`
`4. |NOvent delivery system performance .
`
`.
`
`Warranty
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. .17-4
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. ..W-1
`
`1605-0014-000
`
`02KJ8lU0
`
`007
`
`

`
`|NOvent delivery system
`
`02.-"08.-"00
`
`1605-0014-000
`
`008
`
`

`
`
`
`Introduction
`
`
`AWARNING If an alarm occurs, safeguard the patient first before
`troubleshooting or repair procedures.
`
`Use only pharmaceutical grade NOIN2.
`
`In this section
`
`Definitions and Abbreviations .
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`. 1-3
`                          
`
`Symbols used in this manual or on the system .
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`. 1-4
`             
`
`1605-0014-000
`0
`
`02208200
`
`009
`
`

`
`|N0vent delivery system
`
`|mpo|-[ant
`
`Before using the |N0vent delivery system, read through this manual
`and the related manuals provided. Read through the manuals for the
`ventilator, humidifier and any other accessory items used. Follow the
`manual instructions and obey Warnings and Cautions.
`
`Know the proper setup and operation of the ventilator and the
`humidifier you are using.
`
`Know the proper information provided in this Operation and
`Maintenance manual before operating the |N0vent delivery system.
`
`Keep this manual readily available to answer questions.
`
`Read through each step or procedure to understand it before doing it.
`
`All specifications in this manual are nominal.
`
`Read the User Responsibility statement on the inside front cover of
`this manual; it describes what the user must do to maintain this
`product.
`
`
`A WARNING The |N0vent delivery system must only be used in accordance
`with the indications, usage, contraindications, warnings and
`precautions described in the nitric oxide drug package inserts
`and labeling. Drug package inserts and labeling are packaged
`with each drug cylinder. Refer to this material prior to use.
`
`The use of devices which radiate high-intensity electrical fields
`may affect the operation of the delivery system. Do not drape the
`leads of these devices across the delivery system. Constant
`surveillance of all monitoring and life support equipment is
`mandatory whenever interfering devices are in operation on or
`near a patient.
`
`02r'08.-"00
`
`1605-001 4-000
`
`010
`
`

`
`Definitions and abbreviations
`
`Introduction
`
`°/o vfv:
`
`Button:
`
`% volumefvolume
`
`A pushbutton control used to select a function.
`
`Control wheel:
`
`Rotary control used to change and confirm
`settings or actions.
`
`Display:
`
`HFOV:
`
`Menu:
`
`The electronic information panel on the front of
`the delivery system.
`
`High frequency oscillating ventilator.
`
`A list of available choices for an operation.
`
`Menu area:
`
`Area in the display where choice selections are
`shown.
`
`Message area:
`
`Area in the display where information is shown.
`
`N2:
`
`N0:
`
`N02:
`
`Nitrogen.
`
`Nitric oxide.
`
`Nitrogen dioxide.
`
`NOIN2:
`
`Nitric oxide (NO) and nitrogen (N2) gas mixture.
`
`Numeric area:
`
`Area in the display where values and settings
`are shown.
`
`02:
`
`Oxygen.
`
`Pop-up box:
`
`Menu or message which appears on the display.
`
`Resolved alarm:
`
`An alarm whose cause has been corrected.
`
`Set NO:
`
`The concentration of N0 delivered by the
`system as set by the user.
`
`1605-0014-000
`
`02i‘08lD0
`
`011
`
`

`
`|NOvent delivery system
`
`Symbols used in the manual or on the system
`
`WARNINGS and CAUTIONS tell you about dangerous conditions that
`can occur if you do not obey all of the instructions in this manual.
`
`Warnings tell you about a condition that can cause injury to the
`operator or the patient.
`
`Cautions tell you about a condition that can cause damage to the
`equipment. Fiead and obey all warnings and cautions.
`
`Other symbols replace words on the equipment or in this manual.
`These symbols include:
`
`
`I On (power.)
`
`
`fl This way up.
`
`Dangerous voltage.
`
`Input.
`
`
`
`Output.
`
`Stock number.
`
`Serial number.
`
` 
`
`O Off.
`
`d) Standby.
`
`~ Alternating current.
`
` 
`
`J,— Earth ground.
`
`(9 Protective earth ground.
`
`fig Alarm Silence.
`
`Sample gas outlet.
`
`
`Vgy Eqwpotenflal‘
`
`O2 Sensor connector.
`
`+ Plus’ positive polarity‘
`
`NOIOZ Manual delivery system.
`
`- Minus, negative polarity.
`
`2
`NO Sensor
`
`.
`
`02r'08.-"00
`
`1605-001 4-000
`
`012
`
`

`
`Introduction
`
`
`
`Movement In one direction.
`
`NO Sens“
`
`Autoclavable.
`
`O2 Sensor_
`
`
`
`Not autoclavable.
`
`Product “Use by" information (indicating
`warranty expiration date as YYYY—M M).
`
`
`
`Type B equipment‘
`
`Replace fuse only as marked.
`
`
`
`Type BF equipment‘
`
`Indicates operating-on-battery.
`
`
`
`Warning and Caution symbol.
`
`(E European Union Representative
`
`
`
`Attention, consult accompanying
`documents, IEC 601-1.
`
`A system with this mark agrees with the
`European Council Directive (93/42/EEC)
`for Medical Devices when it is used as
`
`specified in the Operation and
`Maintenance Manual. The xxx:-< is the
`
`certification number of the Notified Body
`used by Datex-Ohmeda's Quality
`Systems.
`
`160 5-001 4-000
`
`02308300
`
`013
`
`

`
`|NOvent delivery system
`
`02!'08.-"00
`
`1605-001 4-000
`
`014
`
`

`
`
`
`
`
`
`
`
`
`
`
`2/General Information
`
`In this section
`
`Overview .
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`. 2-2
`                                          
`
`                                
`Configuration options .
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`. 2-3
`
`Main component views .
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`. 2-4
`                               
`
`Display and user controls.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`. 2-7
`                             
`
`Theory of operation .
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`. 2-11
`                                 
`
`NO Flow from the cylinder to the patient breathing
`circuit .
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`. 2-11
`                                      
`
`Effect of inserting the lN0vent into a breathing
`circuit .
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`.
`
`.
`
`.
`
`.
`
`.
`
`.
`
`. 2-14
`
`1605-0014-000
`0
`
`02208100
`
`015
`
`

`
`lN0vent delivery system
`
`Overview
`
`The lN0vent delivery system delivers nitric oxide therapy gas into
`the inspiratory limb of the patient breathing circuit in a way that
`provides a constant concentration of NO, as set by the user, to the
`patient throughout the inspired breath.
`
`The lN0vent delivery system uses a specially designed Injector
`Module which enables the tracking of ventilator waveforms and the
`delivery of a synchronized and proportional dose of NO.
`
`The lN0vent delivery system may be used with most ventilators.
`
`The lN0vent delivery system provides continuous on—|ine integrated
`monitoring of delivered 02, N02, and NO, and a comprehensive
`alarm system.
`
`The interactive control panel and display provide a simp|e—to—
`understand interface for operation of the lN0vent delivery system.
`
`The battery system provides up to 30 minutes of uninterrupted nitric
`oxide delivery in the absence of an extemal power source.
`
`The lN0vent delivery system provides an integrated Manual NO
`delivery system for administration of a fixed concentration of NOIOQ
`therapy with a resuscitator bag.
`
`02:08.-"00
`
`1605-001 4-O00
`
`016
`
`

`
`Configuration Options
`
`2!Genera| Information
`
`The |NO\rent delivery system is available in two configuration options:
`
`- Shelf or table option and
`
`- Mounted on a transport cart which holds two N0 therapy gas
`cylinders.
`
`The options let you choose the setup which meets the needs of the
`work environment.
`
`1605-0014-000
`
`02lU8lU0
`
`017
`
`

`
`|NOvent delivery system
`
`Main component views
`
`Measured Values llnsillrerll
`N02
`(Wm
`
`0 . 0SJ]
`
`02
`(*2)
`
`40
`
`AA.69.034
`
`. Display and operating panel
`. Manual NO Delivery System Activity Indicator
`. NOINZ Injector Tube connector
`. Injector Module Electrical Cable connector
`. Fluid Trap Bottle
`. Inspiratory Gas Sample Line connector
`. Inspiratory Gas Sample Line
`. Fluid Trap Filter Cartridge
`
`1 2 3 4 5 6 7 8
`
`  
`Figure 2-1 -
`
`INOvent delivery system, front View
`
`02:08.-"00
`
`1605-001 4-O00
`
`018
`
`

`
`2;’Genera| Information
`
`
`
`I‘-V
`5:‘L0
`<1:<t
`
`1. NO Sensor
`
`. N02 Sensor
`. Nurse Call connector
`.Seria| Port connector
`
`. Power Entry Module
`. Power Cord connector
`
`. External Ground connection point
`.ON!STANDBY switch
`
`. Cooling Fan and Filter
`10. O2 Sensor
`11. Sample gas outlet
`12. Regulator Low Pressure Hose with Quick Connector (2)
`13. NOfNg Input Hose Connector (2)
`14. Resuscitator Bag 02 Tube connector
`15. Connector to the O2 Flowmeter
`
`  
`Figure 2-2 -
`
`|NOvent delivery system, rear View
`
`1605-0014-O00
`
`O2f08a"00
`
`019
`
`

`
`|NOvent delivery system
`
`1
`
`\:§D/
`
`1. NOINZ injector tube connection to the INOvent delivery system
`2. Inspiratory limb breathing circuit connection; to humidifier
`3. Inspiratory limb breathing circuit connection; from ventilator
`4. Electrical cable connection to the lNOvent delivery system
`
`  
`Figure 2-3 -
`
`Injector Module, side View
`
`02:08.-"00
`
`1605-001 4-O00
`
`020
`
`

`
`Display and user controls
`
`2!Genera| Information
`
`The |NOvent delivery system display and control panel has a control
`wheel, buttons, and an easy-to-view electroluminescent display. All
`the |NOvent delivery system controls and monitoring are in one place
`which reduces the need for extensive visual scanning.
`
`The buttons and control wheel on the |NOvent delivery system
`control panel perform an assortment of functions and access various
`features.
`
`The steps for performing any operation, or accessing any function, on
`the |NOvent delivery system require the same interaction with the
`control panel, namely:
`
`1. Select (press) a button associated with the desired function or
`operation,
`
`2. Rotate the control wheel to select a menu item or value,
`
`3. Confirm the selection by pressing either the control wheel or the
`function button.
`
`The control panel display, buttons and control wheel are used to:
`
`- Set the concentration of delivered NO,
`
`- Adjust alarms,
`
`- Pause NO delivery,
`
`- Silence alarms,
`
`Review alarm history,
`
`Define setup options,
`Calibrate the monitors.
`
`A single tone indicates:
`
`- Your choice of an action with a button or the control wheel,
`- An invalid action with a button or the control wheel or
`
`- A message box appearance on the display.
`
`In normal operation, a typical display presents these major sources of
`information:
`
`- The Set NO value,
`
`- The measured values of three inspired gases 02, N02 and NO,
`- Alarm information.
`
`Menus guide you through the steps needed to deliver NO. Display
`and user controls are also discussed in the sections of this manual
`
`where system operations are explained.
`
`1605-0014-000
`
`02K)8iU0
`
`021
`
`

`
`|NOvent delivery system
`
`
`
`4
`
`5
`
`Measured Values (Inspired)
`
`0?
`(7%
`40
`
`1D!)
`
`N902
`N0
`(ppm)
`3.0
`10
`canmmmI G ‘
`nlnu-augg
`
`21
`
`I
`
`Oil
`
`15
`
`D
`
`Alarms
`
`Setup
`
`I
`
`15
`
`14
`
`13
`
`12
`
`11
`
`10
`
`9
`
`1. Alarm Silence button
`2. Alarm Silenced Countdown Timer
`
`3. Alarm message area
`4. Alarm History indication are
`5. Alarm History button
`6. NO Bargraph
`7. Control Wheel
`8. Set NO button
`9. Pause Flow button
`
`10. Inspired NO Measured Value
`11.CaIibration button
`
`12. Inspired N02 Measured Value
`13. Setup button
`14. Inspired O2 Measured Value
`15. Alarms button
`
`16. Mains Power light
`
`  
`Figure 2-4 - User controls and indicators on the front
`
`Following are brief descriptions of display and user controls shown in
`figure 2-4. See the other areas in the manual for more information.
`
`1. Alarm Silence
`buflon
`
`Pushing the Alarm Silence button silences an audible alarm and
`starts the alarm silence countdown timer in the upper left location of
`figure 2-4 (item 2). The counter shows the remaining silence time. A
`new alarm will sound even if the Alarm Silence function has
`
`previously been activated.
`
`2. Alarm Silenced
`Countdown Timer
`
`The Alarm Silenced Countdown Timer shows a silenced bell and the
`
`remaining silence time when the Alarm Silence button is pushed. The
`timer counts down from 120 seconds to zero for high and medium
`priority alarms.
`
`02:08.-"00
`
`1605-001 4-O00
`
`022
`
`

`
`2!Genera| Information
`
`3. Alarm message area
`
`The highest priority most recent alarm message appears in the top
`part of this alarm message area first. On high priority alarms, the
`elapsed-time counter counts up. Alarm messages are also
`accompanied by audio signals. Refer to section 7!Alarms for
`additional information on alarms.
`
`4. Alarm History
`indication area
`
`This normally blank area displays “Alarm History” if there are resolved
`alarms. See section 7l'Alarms for information on resolved alarms.
`
`5. Alarm History button
`
`Pushing the Alarm History button shows the Alarm History menu
`listing the resolved alarms. There is an option to clear the displayed
`resolved alarms.
`
`6. NO Bargraph
`
`The NO Bargraph shows the Set N0 value in ppm.
`
`7. control wheel
`
`The Control Wheel’s function is defined by the information on the
`display. When a menu is displayed, turning the wheel in either
`direction selects a choice or adjusts device settings on the menu.
`Pressing the wheel then selects the current menu item or confirms the
`adjusted setting.
`
`8. Set NO button
`
`Pushing the Set NO button allows adjustment of the Set NO bargraph.
`
`9. Pause Flow button
`
`Pushing the Pause Flow button displays the “Pause F|ow” message
`box but does not pause the NO flow. To pause NO flow, you must
`push the Pause Flow button a second time.
`
`10. Inspired N0
`Measured Value
`
`The measured value of inspired NO (in ppm) is displayed in large
`numbers. Low and high alarm settings for inspired N0 appear below
`these numbers.
`
`11. Calibration button
`
`Pushing the Calibration button displays the menu for calibrating the
`high and low range of the sensors. Flefer to section 4!Calibration for
`information on calibration.
`
`12. Inspired N02
`Measured Value
`
`The measured value of inspired N02 in ppm is displayed in large
`numbers. The high alarm setting for inspired N02 appears below
`these numbers.
`
`13. Setup button
`
`Pushing the Setup button displays the Setup menu for setting the
`alarm volume and display brightness or returning to the normal
`display. See section 3fSetup and Calibration for additional
`information.
`
`1605-0014-000
`
`02lU8iU0
`
`023
`
`

`
`|NOvent delivery system
`
`14. Inspired 0; Measured
`Value
`
`The inspired 02 measured value in °/o is displayed in large numbers.
`The low and high alarm settings for inspired 02 appear below these
`numbers.
`
`15. Alarms button
`
`Pushing the Alarms button displays the menu for setting the user-
`adiustable alarm limits. High limits can be set for 02, N02 and NO.
`Low limits can be set for O2 and NO. See section 7!Alarms for
`additional information on alarms.
`
`16. Mains Power light When the green Mains Power light is on, the system is connected to a
`functioning mains power outlet.
`
`02:08.-"00
`
`1605-001 4-O00
`
`024
`
`

`
`2;’Genera| Information
`
`Theory of Operation
`
`The lNOvent delivery system delivers nitric oxide therapy gas into the
`inspiratory limb of the patient breathing circuit in a way that provides a
`constant concentration of NO to the patient throughout the inspired
`breath.
`
`The |NOvent delivery system uses a specially designed Injector
`Module which enables the tracking of ventilator waveforms and the
`delivery of a synchronized and proportional dose of NO.
`
`NO Flow from the
`Cylinder to the
`Patient Breathing
`Circuit
`
`‘
`
`:;9g“..';':,:5"'e
`
` 
`
`
`
`
`
`
`
`
`
`
`
`
`
`_
`s
`-
`
`E
`2
`'
`M°"it°ri"9
`
`and
`
`Alarm Board
` 
`Controls and Display
`  
`
`Inspired Gas
`Sample Port
` 
` 
`
`Quick Connec

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket