`
`Espacenet- Bibliographic data
`
`
`
`
`
`meantt>22.”mFeee,aailyteesc_»=5=—)eded3Ge7WH:4“eenOF
`taete,af,=<oOqeac<aa,G,8Gbe ¢5odOBOhae*
`ames| ™2teo.BDa3..~6-©goiern,=.abaie,ayr5°C~=aeoiie=.gi .u86%ooMwZeg=oPoesavin|“atoOoO42oe&WHEEL:teddc=a—2i2oeeaSeth,beoOa.82o£°@..ah:
`vols@aape&SemonSweyoeOGerOREmwus<LomBetytOIitoONGsfLy
`vealoepm,LTOorFOOgexmartixaroeoPSae&:MntorchSSooiSeBrutOYfed,ot
` certeae<i=pePAEZ?.+|wedandOFGLI6oehyyT>wO—ooiSgwieo:oeee-EamewO&S|24oetid=z<0SSola3ogOsaghtayiF“entUYicolcotteSS“10tyCie,moss4eaS3o6BUEgi:oezOFSFSteeoSo&OO,Beg:isO.:osai&&¢i--SatnaggAcaOdeefsrEoO|oe<cage.teslagas2r#29onSSoo|e22227826§Sa¥eexgq25e(0BE6GSSODBRStaaomAh.aeowsOsFOQZEmomAsaodu«2oywdFOOm9SsBFeestainSoNaOoey@yettNOyr..WHSSBOEBTaNpeookar:usysOnCeqdalcd—NooOGs
`
`apas299109020eeiZ@ar
`
`7%uwawidBB£4:thes$04a(te
`eomaO83oo}oyi=Oon
`PreiyWLresy~~
`
`
`ZHLHwe£8.Goaz46pee:$va0.08<{©PXteeOgteraPronoyByeeUeoO®K&,iOO>€E32No|zoi)aOoO99O0TZO2>”3;ooaSoheGonOF<+Y'TfzSo°SOOa42imSCS5¥006§2aGegitl&6AbemHEO=EeSoeoOoocoybee£=22oiES*:GifonIZLOO®oO©ayoFwef:th,©zx“OoACA=es-Fee
`
`
`_-Luaia©ui8O>~Ly—Wu292Zzsr(8BASoO=aaisaS208ovegeOuarcsooesG70§udeoSsneQO©otenzaoFmoYaeeygeen—Lueh1cgi4264553B9295,8&‘fit,4H“gen
`ogthetotimS242
`fonfen
`
`
`https://worldwide.espacenet.com/publicationDetails/biblio?II
`
`O0&ND
`
`3&adjacent
`
`=true&locale=en_EP&FT
`
`D&date=20090617&CC
`
`EP&NR=2069...
`
`1/2
`
`
`
`
`
`4/23/24, 2
`
`10 PM
`
`Espacenet- Bibliographic data
`
`py,
`
`tapdyee
`
`wsbeeshae
`
`
` wwteodgofee4fataAbode
`
`pedtyan
`
`aneft
` tersBerscetgiaeft~,@
`
`amaa
`
`weeaeEe
`
`OTSstt
`
`https
`
`Ica
`/iworldwide.espacenet.com/pubili
`
`ionti
`
`Details/biblio?II
`
`O&ND=3&adjacen
`
`t=true&locale=en_EP&FT:
`
`D&date=2009061 7&CC=EP&NR=2069.
`
`2/2
`
`
`
`
`
`
`
` {13} Vergifentlichungsaummer:
`
`{11} Publication number:
`
`{13} Numére de publication:
`
`EP 2 069 0714 AG
`
`internationale Anmeldung veréfientlicht durch die
`
`Weltorganisation fur geistiges Eigentum unter der Nummer:
`
`WO 2008/043999 (Art. 153(3) EPU).
`
`international application published by the World
`
`intellectual Property Organization under number:
`
`WO 2008/043999 (Art. 153(3) EPC).
`
`Demande internationale publiée par POrganisation
`
`Mondiale de la Propriéte Intellectuelle sous le numero:
`
`WO 2008/043999 (art. 153(3) CBE).
`
`
`
`(12) INTERNATIONAL APPLICATION PUBLISHED UNDER TRE PATENT COOPERATION TREATY (PCT)
`
`
`
`
`
`(10) Intermational Publication Neamber
`WO 2008/043999 A2
`
`(19) World Intellectual Property Organization
`international Bureau
`‘
`
`(43) International Publication Date
`17 April 2008 17.04.2008)
`
`(54) Tatermational Patent Classification:
`AGEN 5/62 (2006.01)
`AGIB 18/48 2005.01)
`
`(2%) Taterrational Application Number:cr
`AQ97
`4,
`PCT/GB2007003827
`(22) International Fling Date: 10 October 2007 (18.10.2007)
`
`(25) Filing Language:
`
`(26) Publication Language:
`.
`Dyak
`Cds
`sp
`tecne
`(30) Priority Data:
`O620064.6
`717030.4
`
`English
`
`English
`
`10 October2606 (10.10.2006)
`;
`31 August 2007 (31.08.2007)
`
`GB
`GB
`
`(7a) Agents: JOHNSGN, Richard et al.; Mewburn Hitis LLP,
`York House, 23 Kingsway, London Greater London WC2B
`GHP (GB).
`:
`:
`:
`a
`.
`(gy) Designated States (uniess otherwise indicated, for every
`kind of national protection availabley AE, AG, AL. AM,
`AT, AU, AZ, BA, BB, BG, BH, BR, BW, BY, BZ, CA, CH,
`CN, CO, CR, CU, CZ, DE, DK, DM, DO, DZ, EC, EE, BG,
`ES, FL, GB, GD, GE, GH, GM, GT, HIN, HR, HU, ID, IL,
`IN, 18, JP, KE, KG, KM, KN. KE KR, K%, LA, LC, LK,
`DR, LS, UP LU, LY, MA, MD, ME, MG, ME, MN, MW,
`MX, MY, MZ, NA, NG, NE, NO, NZ, OM, PG, PHL PL,
`:
`:O,
`RS,
`RU, SC,
`SP, SE, SG, SK, SL,
`SM,
`SY, SY,
`PE, RO, RS, RU, SC, SD, SH, SG, SK, SL, SM, S¥, SY,
`ETM, PN, PR, ‘PL UY, UA, UG, US, U4 VC, VN, ZA,
`EM, ZW.
`
`(72) Applicant Gor olf designated States except US}: MED-
`ICAL DEVICE INNOVATIONS LIMITED [GB/GH],
`Daresbury Innovation Centre Limited, Daresbury Science
`and Innovation Campus, Plalton Cheshire WA4 4B(GB).
`
`72}
`)
`
`
`
`Inventors; and
`BANCOCK,
`US only:
`(for
`Inventors/Applicants
`Christopher Paul [GB/GE!; 61 Bryants Hull, St George,
`Bristol Avon BSS 807% (GB). WHITE, Maicobn
`[GBAGKE: ‘Lhe Arches, Phe Grove, Coleford, Radstock,
`Bath And North Bast Somerset BA3 SLW (GB). BISHOP,
`Published:
`Jebu [GB/GRB]; 12 Metford Road, Redland, Bristol Avon
`5S6 7LE (GB). BOOTOR, Martie Wynford (GR/GB], —- without international search repart and to be republished
`Cardamine, Dulcote, Wells Somerset BAS 3P7, (GR).
`upon receipt of that report
`
`(84) Designated States (uniess otherwise indicated, forevery
`
`Kind of regional provection available}: ARIPO (BW, GH,
`OM, KE, LS, MW, M4, NA, SD, SL, S54, PZ, UG, 34M,
`ZW), Burasian CAM, AZ, BY, RG, KZ, MD, RU, TS, TM),
`Duropean (AT, BE, BG, CH, CY, CZ, DE, DE, EEL ES, Te,
`PROGR. GR, AU, LEIS, TP UP LO, LV, MO, MP NL, PL,
`FT, RO, SH, ST, SK, TR), OAPI GE, BI, CE CG, CL, CM,
`GA, GN, GQ, GW, ML, MR, NE, 5N, TD, TG).
`
`($4) Title: APPARATUS POR TREATING TISSUE WITH MICROWAVE RADIATION AND ANTUNNA CALIBRATION SYS-
`TEM AND METHOD
`
`
`
`ey
`<f
`
`
`
`Hae
`aB
`
`(87) Abstract: A calibration method and apparatus for surgical antennas which are arranged to deliver microwave radiation (e.g.
`having a treatment frequency of SOQ MHz to 100 GHz) into biological tissue is disclosed. An emitting region of the antenna is
`exposed to a plurality of calibration standards each having a different complex impedanceat the treatment frequency. In one embod-
`iment the calibration standards are created in a short-circuit-terminated waveguide cavity of variable length. In another embodiment,
`cLeaY
`each calibration standard is a different mixture of two or moreliquids. Measurementofthe magnitude and phase ofsignals reflected
`omthe emitting region when exposed to the calibration standard can permit calibration ofthe antenna, ¢.g. by generating a mapping
`function based on the measured values and knownor reference values forthe calibration standards, Also disclosedis tissue treatment
`
`
`apparatus having an ablation channel for conveying microwave radiation fo a surgical antenna at a high power level and a separate
`measurement channel for conveying microwave radiationto a surgical antenna at a low powerlevel, wherein the measurement chan-
`nel bypasses noisy corsponents on the ablation channel. A surgical antenna having an impedance transformerfor matching an e.g.
`coaxial feed structure which terminates in one or more radiating elements wilh Ossue to be treated is also disclosed.
`
`y=©— 3
`
`sS=aW
`
`o:
`
`
`
`WO 2608/043999
`
`POCT/GB2007/803827
`
`APPARATUS FORTREATING TISSUEWITH MICROWAVE
`
`ANTENNA CALIBRATION SYSTEM AND METHOD
`
`TECANICAL FIELD
`
`The invention relates to the treatment of biological
`
`tissue using microwave radiation.
`
`In particular aspects,
`
`the
`
`invention concerns a surgical antenna for delivering microwave
`
`radiation to tissue, a tissue treatment system for carrying
`
`out ablation or measurement of tissue using microwave
`
`radiation from such an antenna, and a system and method of
`
`calibrating an antenna for use in such a system.
`
`An electrosurgical system that is arranged to
`
`controllably ablate a tumour and/or measure information
`
`concerning the tumour and surrounding healthy tissue is known.
`
`Such a system may use two channels: a first channel
`
`toa perform
`
`controllacd tissue ablation, and a second channel
`sensitive tissue state {dielectric} measurements.
`
`to perform
`The general
`
`principles relating to the operation of such a system are
`
`disclosed in WO 2004/047659 and WO 2005/115235.
`
`This disclosure comprises three main aspects.
`
`The first
`
`aspect relates to a system and method for calibrating surgical
`antennas at
`the point of radiation (the aerial), and in
`
`particular to performing the calibration routine automatically
`
`when the calibration systems are used in conjunction with an
`
`electrosurgical system @.g. of the known type.
`
`The second
`
`aspect relates to further improvements to the known treatment
`system, which improvements offer significant advantages in
`terms of enhanced measurement sensitivity and reduced power
`
`levels required in the measurement mode due to the use of a
`
`cn
`
`ts
`
`(si
`
`tal
`
`sad on
`
`
`
`WO 2603/043999
`
`PCT/GB2007/003827
`
`10
`
`L5
`
`separate low power transmitter and receiver {transceiver}.
`
`The third aspect relates to surgical antennas that may be used
`
`with the calibration system to enable said antennas to be
`
`calibrated at the distal and (the aerial) thereby to enable
`Said antennas to be used to perform tissue state measurements
`or to be used to ablate tissue where it is desirable to
`
`perform dynamic impedance matching between the distal tip of
`
`the antenna (the aerial) and the biclogical tissue lead.
`
`Calibration System and Procedure
`
`At its most general,
`
`the first aspect of the invention
`
`May provide calibration apparatus for an antenna that is
`
`arranged to emit microwave radiation from an emitting region
`
`thereof,
`
`the apparatus having:
`
`a loading arrangement adapted
`
`to subject the emitting region of the antenna to a plurality
`of impedances, each impedance having a known value for a
`predetermined frequency of microwave radiation, a detector
`
`arranged to measure the magnitude and phase of microwave
`radiation having the predetermined frequency that is emitted
`
`from the antenna and reflected from the loading arrangement,
`
`and a processing unit configured to generate calibration data
`
`for the antenna, wherein,
`
`if the antenna is used subsequently
`
`to measure magnitude and phase of microwave radiation having
`
`the predetermined frequency with an unknown load at the
`
`emitting region of the antenna,
`
`the calibration data is usable
`
`to convert the measured magnitude and phase to be
`
`representative of the unknown load.
`
`Praferably,
`
`the loading arrangement
`
`includes a
`
`wa
`
`ubstantially lossless waveguide cavity between a first end
`
`adapted to receive the emitting region of the antenna and a
`
`second end, and wherein a distance between the first end and
`
`the second end is variable.
`
`For example,
`
`the second end may
`
`be slidable relative to the first end, e.g. under the action
`
`of a linear actuator.
`
`
`
`WO 2603/043999
`
`PCT/GB2007/003827
`
`Preferably,
`
`the cavity is electrically connectable to the
`
`antenna and the second end electrically connected to the
`
`cavity.
`
`Phe electrical connection between the cavity and
`
`antenna and/or between the cavity and the second end may be
`
`through a radio frequency (RF) choke.
`
`Preferably,
`the plurality of impedances include 9 ©
`{short circuit) and ~ Q {open circuit}.
`Preferably,
`the loading arrangement is adapted to permit
`
`generation of calibration data for two or more different
`
`frequencies of microwave radiation.
`
`the first aspect of the invention
`in another expression,
`may provide a combination of calibration apparatus according
`to any preceding claim and an antenna arranged to emit
`
`microwave radiation from an emitting region thereof, wherein
`
`at least the emitting region of the antenna and the loading
`
`arrangement are packaged together in a sterile environment.
`
`In yet another expression,
`
`the first aspect of the
`
`invention may provide a method of calibrating an antenna that
`is arranged to emit microwave radiation from an emitting
`region thereof,
`the method comprising: subjecting the emitting
`region to a plurality of impedances, each impedance having a
`
`known value for a predetermined frequency of microwave
`
`for each impedance: emitting microwave radiation
`yadiation,
`having the predetermined frequency through the antenna;
`measuring the magnitude and phase of the emitted microwave
`
`radiation that is reflected from the loading arrangement; and
`
`generating calibration data for the antenna from the magnitude
`and phase measured for each of the plurality of impedances,
`whereby,
`if the antenna is used subsequentiy to measure
`magnitude and phase of microwave radiation having the
`predetermined frequency with an unknown load at the emitting
`region of the antenna,
`the calibration data is usable to
`convert the measured magnitude and phase to be representative
`of the unknown load.
`
`The ability to effectively perform antenna calibration at
`the emitting region (e.g. distal tip) may enable efficient
`
`wn
`
`16
`
`bo Cit
`
`30
`
`Ww FV
`
`
`
`WO 2603/043999
`
`PCT/GB2007/003827
`
`Microwave energy transfer into biological tissue, where the
`impedance presented to the distal tip of the surgical anterna
`
`changes as the treatment process progresses. Once the antenna
`
`structure has been calibrated, 1t is then possible to perform
`
`accurate dynamic tuning adjustment to enable the distal tip of
`
`the antenna structure to be impedance matched with the
`
`changing impedance of the biclogical tissue. Fhe ability to
`
`perform impedance matching between the distal tip of the
`fen
`
`surgical antenna and the biological
`
`tissue can prevent
`
`reflection of energy due to impedance mismatch, which can
`
`cause excessive heating of antenna and cable assemblies and
`
`increase the time required to perform ablation of a volume of
`
`tissue.
`
`In applications where the antenna is used in minimally
`
`imvasive surgery this heating may cause te collateral damage
`
`i
`
`5
`
`to healthy tissue structures.
`
`A further advantage is that the
`
`dosage of energy delivered into biological tissue can be
`controlled with greater accuracy than that possible using a
`
`system where unquantifiable reflections due to impedance
`
`mismatches cannot be compensated for.
`
`For the implementation
`
`of this feature it is preferable for this invention to be used
`
`with a method of performing dynamic impedance matching.
`
`A
`
`system to perform such impedance matching using a three stub
`
`waveguide cavity tuner, where the stubs are automatically
`
`moved using three linear motors and a suitable control system
`
`has been disclosed in WO 2004/047659.
`
`Fhe ability to effectively perform antenna calibration at
`the distal tip may also enable the surgical antenna to be used
`as a useful tool for measurement of dielectric information
`
`relating to
`
`the
`
`properties of the biological tissue. Effective
`
`calibration at the distal tip of the surgical antenna enables
`
`the measurement reference plane to be moved to the exact site
`
`{or location} where the measurement is to be performed,
`
`for
`
`example at the periphery between healthy tissue and cancerous
`
`tissue, er inside the cancerous tissue. The ability to
`
`calibrate surgical antennas in this way can enable optimal
`
`measurement sensitivity to be achieved.
`
`2
`
`5
`
`Ud
`
`Co
`
`on
`
`
`
`WO 2603/043999
`
`PCT/GB2007/003827
`
`in
`
`Therefore,
`
`the current invention can be used to enable
`
`dynamic impedance matching to be performed, and can be used to
`
`enable sensitive and repeatable dielectric measurements to be
`
`made.
`
`The current invention relates primarily to the
`
`5
`
`calibration of surgical antennas, but the invention is not
`
`Limited to calibrating these devices.
`
`‘The calibration system disclosed herein effectively
`
`conmects the distal tip of the antenna to a plurality of load
`impedances between open circuit (infinite impedance) and short
`circuit
`(zero impedance} values to enable the antenna to have
`
`10
`
`the capability of being able to measure or be sensitive to a
`
`range of impedances between the two extremes.
`
`The distal tip
`
`of the antenna may be automatically subjected to a range of
`
`~
`
`impedances. Methods of automating the measurements are
`
`15
`
`disclosed herein.
`
`Preferably,
`
`the antenna calibration system uses a sliding
`
`short with the antenna fixed in position. RF chokes may be
`
`included to enable the antenna and the sliding short to bs
`
`loosely fitted inside the cavity.
`
`26
`
`In this specification microwave means the frequency range
`
`of between 506 MHz and 100 GHz. However, frequencies between
`
`14 GHz and 15 GHz are preferred, and a spot frequency of 14.5
`
`GHz is used in some embodiments discussed below.
`
`Although the primary purpose of the invention is to
`
`25
`
`calibrate antennas for use in tissue ablation and measurement,
`
`the invention is not limited to this application.
`
`Indeed,
`
`the
`
`invention. may be used wherever the measurement location is at
`
`the distal tip of an antenna arrangement. Put
`
`in another way,
`
`the invention may permit all systematic errors that are
`
`30
`
`present between the distal tip of the antenna (the aerial} and
`
`the digital signal processing unit to be cancelled out,
`
`thus
`
`enabling the tissue load to be effectively connected ta the
`
`digital signal processing unit. This may take into account
`
`variations in the components within the microwave transceiver,
`
`35
`
`for example,
`
`thermal noise or short noise produced by low
`
`
`
`WO 2603/043999
`
`PCT/GB2007/003827
`
`noise amplifiers, driver amplifiers, microwave frequency .
`
`mixers and PIN diode switches and attenuators,
`
`it may be desirable to calibrate at more than one
`
`frequency where a first frequency is used for controlled
`
`ablation and a second frequency is used to perform dielectric
`
`measurements (tissue type/s tate, etc). For example, it may be
`
`desirable to calibrate an antenna structure over a frequency
`
`range of +/- 50 MHz around a spot frequency of 14.5 GHz or ait
`
`
`may be desirable to use other frequencies within the microwave
`or RF region of the electromagnetic spectrum. The dimensions
`ef the calibration system {or assembly} can be adjusted to
`
`accommodate any practicable frequency {or range of
`
`frequencies) where the underlying theory related to the
`
`current
`
`invention remains valid. Bue to the dynamic tuning
`
`mechanism used in the system (e.g. of the type described in WO
`
`2004/047659}, it may be necessary to calibrate at the ablation
`
`frequency.
`
`In this mode of operation,
`
`information concerning
`
`the state of the ablated tissue is used to automatically
`
`the impedance matching mechanism to ensure that
`control
`maximum power
`(or the demanded power)
`is delivered into the
`
`tissue Load seen by the distal tip of the antenna.
`In a
`particular arrangement given in WoO 2004/0 7659,
`four
`directional couplers are used to measure the forward and
`
`reflected power signals and this information is used as the
`
`basis upon which the dynamic impedance matching method is
`
`implemented. It may,
`
`for example, be desirable to perform
`
`tissue ablation at 10 GHz and make dielectric measurements at
`
`pod 6 GHz.
`
`In the instance where it is necessary to calibrate the
`fy
`oD 7)
`ry gu Oo ps o ct a fe ct
`system at two or more frequencies, it is desi
`
`bw
`
`the frequencies can propagate inside the microwave structure,
`
`i.e.
`
`in the case of using rectangular or cylindrical
`
`waveguide,
`
`the waves
`
`should not be cut-off. Tt may also be
`
`esirable to include more than one transceiver unit in the
`
`system when more than one frequency is used and there is a
`_large difference between the two frequencies, for example a 10
`GHz difference. Higher order modes may also propagate when the
`
`an
`
`i
`
`8
`
`the
`
`3
`
`iw
`
`st
`
`
`
`WO 2603/043999
`
`PCT/GB2007/003827
`
`wavelength of the calibration frequency signais are small
`
`compared with the size of the waveguide or co-axial
`
`calibration assembly.
`
`These effects may be taken into account
`
`by performing a system analysis to model
`
`the effects of higher
`
`ry
`
`order modes set up in the cavity. This may not be needed when
`
`the same frequency is used for ablation and measurement
`
`because a standard waveguide can be used which enables the
`
`relevant frequency to propagate unimpaired (e.g.
`
`for 14 GHz,
`
`WRE?2
`
`(WG17) may be used}.
`(WGLS} or WRYS
`The tissue ablation/measurement systems disclosed e.g.
`
`in
`
`WG 2004/047659 and WO 2005/115275 provide for impedance (or
`
`energy) matching between the surgical treatment antenna and
`
`the tissue load, and far the antenna to be capable of
`
`measuring smali changes in complex impedance to enable
`
`characterisation of various tissue types,
`tissue states and/or
`stages associated with the growth of cancerous tumours to be
`
`measured. The present invention aims to improve the
`
`operational efficiency of those systems by providing for the
`
`calibration of the surgical antenna to take place at the
`
`distal tip where the antenna will radiate energy into tissue.
`
`In order to ensure that the widest possible range of
`
`impedances can be measured, it is desirable for the
`
`calibration system to be capable of locally presenting the
`
`antenna with a range
`
`of known impedances between open circuit
`
`and short circuit conditions.
`
`Tt is further desirable for the
`
`calibration procedure to be automated.
`
`The antenna may be a surgical antenna or any other type
`of antenna structure, or other device. The invention is of
`
`particular use where the design of said antenna (or other
`
`device} dees not
`
`lend itself to be connected
`
`to a standard co-
`
`a@ co-axial {or
`for example,
`axial calibration arrangement,
`waveguide) 50 Q lead, and/or a co-axial {or waveguide} short
`cireuit, andfor a co-axial
`(or waveguide}, open-circuit,
`
`and/or a co-axial {or waveguide) sliding load. Phe calibration
`
`system described herein enables the reference plane for
`a mu te
`
`ibration to be moved to the distal tip of the antenna and
`
`re un
`
`an
`vas)
`
`Nr cn
`
`Lad =>
`
`35
`
`
`
`WO 2603/043999
`
`PCT/GB2007/003827
`
`wm
`
`takes into account
`
`the shape and geometry .of the antenna
`
`structure to be calibrated.
`
`Preferably, a range of calibration positions can be
`
`obtained by moving a sliding short circuit, connected to an
`electromechanical linear actuator,
`in such a manner that the
`Sliding short
`(or plunger} starts at the short circuit
`position on the Smith chart and as the plunger is withdrawn
`from the waveguide cavity,
`the impedance moves around the
`
`outer circle {assuming a lossless cavity) where one or more
`
`calibration points can be measured.
`
`If the cavity is
`
`substantially lossless then there may be no real component
`
`present and so the complex impedance will either be an
`inductive reactance or a capacitive reactance. Such an
`
`arrangement forms a particular aspect of the current invention
`
`and is addressed in detail below.
`
`‘The calibration can be
`
`represented on a Smith chart or in other forms, ¢.g. a polar
`
`chart, phase/maqgnitude plots or another suitable measurement
`
`plane.
`
`Due to the need to accurately measure both the phase and
`
`magnitude of information in terms of impedance seen at the
`
`distal end of the treatment /measurement antenna, it is
`
`desirable to calibrate the surgical antenna at the point where
`
`tissue state (dielectric) measurements are to be made. TF the
`
`antenna is not calibrated at the distal tip (where it connects
`
`to the tissue load)
`
`then it is more difficult to measure the
`
`value of impedance presented to the distal tip, and it will be
`
`‘Gifficult to make valid and repeatable measurements of tissue
`
`impedances and/or be able to differentiate between various
`
`tissue. types. Phase and magnitude variations associated with
`
`the components that form the transmission path between the
`
`measurement
`
`instrumentation (generator) and the distal tip of
`
`the antenna make it difficuit to theoretically determine the
`exact phase and magnitude of the signal at the distal end of
`
`the antenna seen at the generator end, where the microwave
`
`transceiver and signal
`
`processing circultry is located. The
`
`components in the path may include: microwave connectors and
`
`=
`
`Nm
`
`bo un
`
`tad
`
`
`
`WO 2603/043999
`
`PCT/GB2007/003827
`
`interconnects, a Flexible cable assembly, a Length of rigid
`co-axial cable that forms a part of the antenna (this is
`
`inserted inside the body},
`
`the antenna (aerial) itself,
`
`the
`
`tuning unit,
`
`the microwave signal mixers, various co-axia
`
`couplers,
`
`semi rigid or flexible semi rigid assemblies,
`
`low
`
`noise amplifiers, drive amplifiers, microwave circulators and
`
`other components within the microwave transceiver line-up. Due
`
`to the short wavelengths associated with microwave
`
`frequencies, it is very difficult to calculate or quantify the
`
`be oo
`
`phase at the distal end of the antenna,
`for example the free
`space wavelength at 14.5 GHz is 20.69 mm,
`thus a variation of
`imm caused,
`for example, by a connector not fully tightened,
`
`will cause a phase variation of approximately 17 degrees.
`
`Also, due to limitations on possible manufacturing tolerances,
`
`it may be impossible to build transmission line assemblies of
`
`several thousands of millimetres in length with less than 1 mm
`
`for less than 0.1%) yield variation,
`
`If the above assembly is connected to the measurement
`
`instrumentation (generator) and calibration is performed with
`
`20
`
`the reference plane at the distal tip of the antenna,
`
`then the
`
`a@ifficulties described above can be reduced or eliminat
`
`Therefore,
`
`the invention addresses problems associated with
`
`inferring information regarding a remotely located tissue load
`
`connected to the distal tip.of the antenna. Preferably,
`
`the
`
`able assembly used between the generator and the distal tip
`199)
`fe
`oss
`and a small
`
`of the antenna exhibits a low insertion
`
`variation in phase with random flexure. It is preferable for
`the insertion loss of the complete assembly (flexible cable
`assembly and rigid co-axial antenna)
`to be less than 1.5 dB,
`
`30
`
`and for the phase variation to be less than 2 degrees for any
`
`possible physical
`y
`im
`ssembi
`het
`
`random variation of the flexible cable
`
`The flexible cable assembly may be a co-axial cable
`
`assemoly or a waveguide assembly or a combination of the two,
`for example, one metre of low loss waveguide assembly could be
`ee)
`attached to the cutpur of the electrosurgical unit and a
`
`
`
`WO 2603/043999
`
`PCT/GB2007/003827
`
`10
`
`second metre of flexible co~axial cable could be attached to
`
`the distal end of the waveguide assembly and be used as the
`
`flexible section to ease antenna manipulation. This
`
`arrangement could be particularly suitable where it is
`
`desirable to implement a structure where the treatment antenna
`
`is attached to a mechanical arm that forms a permanent
`structure.
`In this arrangement, a rotary joint may be employed
`to move the fixed arm in one plane. Said waveguida assembly
`
`may be flexible,
`
`twistable, or a combination of the two. The
`
`‘advantages of using a waveguide assembly are the high power
`
`handiing capability and low insertion loss.
`
`In one aspect,
`
`the invention may therefore relate to a
`
`calibration system to enable a surgical antenna arrangement to
`
`be calibrated at the distal tip. Such a calibrated antenna
`
`arrangement may be used to make repeatable measurements of the
`complex impedance of biclogical tissue for the purpose of the
`determination of the type of biclogical tissue and/or the
`
`state of said tissue, and/or for differentiating between
`
`healthy and cancerous tissue. Alternatively or additionaily,
`the calibrated antenna arrangement may be dynamically
`
`impedance matched into a load represented by the changing
`
`state of said bislegical tissue during the treatment
`
`(or
`
`ablation} process to ensure that said energy is efficiently
`launched for matched} inte said bislogical tissue load, hence
`providing a controlled and efficient method of causing tissue
`
`ablation which can avoid the drawbacks associated with
`
`conventional ablation systems.
`
`According to the invention, calibration takes place at
`
`the reference plane where
`i.e.
`the distal tip of the antenna,
`said calibration is to be made is positioned at the distal tip
`of the surgical antenna. During calibration, it is desirable
`
`for the distal tip of the antenna be exposed to a large range
`of impedances. Ideally the range should span from an open
`circuit, where the impedance is infinity,
`to a short circuit,
`where the impedance is zero,
`to make it possible
`for as much
`information as possibl
`regarding the state of the biclogical
`
`i5
`
`20
`
`ho an
`
`
`
`10
`
`Me wn
`
`awaM
`
`WO 2008/043999
`
`PCT/GB2007/003827
`
`roy feat
`
`tissue to be available for capture. The calibration system
`
`described in this invention may enable both the calibration
`plane to be located at the distal tip of the antenna and for
`the impedance measurement range to be maximised. It is
`
`reguired to be able to repeatably measure smali changes in
`
`magnitude anc/or phase in order to increase the chance of
`
`being able to differentiate between healthy and cancerous
`
`tissue or between types of cancerous tissue. It is therefore
`preferable to optimise the measurement sensitivity or
`capability. Described generally,
`this can be achieved using a
`
`sliding short circuit
`
`for plunger)
`
`inserted inside a wavequide
`
`cavity that is of large enough geometry to prevent the wave
`
`from being cut-off. The exact location of the antenna and the
`
`overall physical geometry of the calibration system may be
`
`optimised based on the theory set out herein e.g. by using
`
`Computer Simulation Technology (CST) Microwave Studio®
`
`electromagnetic field simulation tool. However,
`
`the design of
`
`the calibration system is not limited to using this simulation
`
`package.
`
`Other suitable electromagnetic field simulation
`
`packages that may be used include Ansoft HFSS and Flomerics
`
`Microstripes.
`
`in an alternative arrangement for performing a multipoint
`
`calibration, a plurality of fixed loads designed to Fit to the
`
`distal tip of the surgical antenna that take into account the
`
`non 50 G impedance environment created by the radiating tip
`
`.(the aerial) could be used. However,
`
`this arrangement may
`
`require each load to be physically connected to the distal tip
`of the antenna. This may
`be particularly important in a
`surgical environment where possible operator error is
`preferably minimised and the time
`available to perform
`surgical procedures is limited. Other possible calibration
`arrangements include:
`a threaded arrangement whereby the load
`is moved by twisting a rod inside a cavity containing a moving
`load (short), or an arrangement whereby a load is moved using
`
`a biased, 6.q. spring loaded, ratchet mechanism (e.g. similar
`
`to that used in retractable pens}.
`
`In the latter arrangement,
`
`
`
`WO 2603/043999
`
`PCT/GB2007/003827
`
`12
`
`for example,
`
`the pressed state (where the spring is
`
`compressed} may cause the movable load to be in the short
`
`circuit position, and the released state (where the spring is
`released) may cause the movable load to be in the open circuit
`position, i.e. the movement between the pressed and released
`
`states may be an exact multiple of an odd number of quarter
`wavelengths at. the frequency of operation to enable the
`impedance transformation from an open circuit to a short
`
`circuit or vice versa.
`
`In another aspect,
`
`the invention may relate to a method
`
`of connecting a transmission assembly to a calibration system
`
`to perform an automated calibration using measurement
`
`instrumentation located away from the calibration site.
`
`In a
`
`preferred ambodiment,
`
`the transmission assembly is 1.62 metres
`
`15
`
`in length and comprises:
`
`a co-axial connector
`
`(preferably N-
`
`type}, a flexible transmission cable of length 1.5 metres, and
`an antenna assembly of Length @.12 metres comprising itself of
`
`a section of rigid co-axial cable with an outer jacket made
`
`from a composition of stainless steel and copper
`
`or silver
`
`{the inner of the outer jacket shail be copper
`
`(or silver
`
`plated)
`
`to provide low conductor loss for the electromagnetic
`
`field), and a distal tip made from a low loss ceramic material
`
`that forms an impedance transformation circuit, and also
`
`provides the desired hardness and sharpness to enable the
`antenna to be directly inserted through human tissue. This
`
`invention is not Limited to the use of co-axial transmission
`
`assemblies, or co~asial anterna structures. For example,
`
`the
`
`transmission line may comprise a waveguide assembly (solid,
`
`flexible or flexible/twistable} and a ceramic antenna
`
`structure may be coupled directly into said waveguide
`
`Gy
`structure. The transmission assembly may be greater than 1.5
`
`The length may be Limited by the insertion
`metres in length.
`5
`ce orwo @
`&Os
`be
`3 of the cable assembly and the antenna structure since
`
`Q of the system is limited by the insertion loss and it. is
`
`required for the 9 to be as high as possible to enable 4
`resonant cavity to be set-up between the tuner and the digital
`
`25
`
`30
`
`
`
`WO 2603/043999
`
`PCT/GB2007/003827
`
`13
`
`tip of the antenna.
`
`For conventional
`
`low less co-axial cable
`
`the length may be limited to 3 metres.
`
`In yet another aspect,
`
`the invention may relate to a
`
`method of operating the sliding short to enable the antenna to
`
`be connected to a plurality of calibration impedances.
`
`An
`
`example of calibrating to a specific impedance between the
`
`open anc short
`
`load is as follows: if the sliding short is
`
`moved an electrical distance of three eighths of a wavelength
`
`at the frequency of operation from the short circuit {the
`generator)
`towards the tissue (the load)
`then, assuming that
`
`the impedance
`the transmission loss along said path is zero,
`Will comprise an inductive reactance of value equal
`to the
`characteristic impedance of the transmission line. It should
`be noted that this analysis assumes that a perfect short
`
`circuit
`exists when the sliding short is at the distal tip of
`the antenna. The advantage of using the sliding load
`
`arrangement is that
`
`the varlable

Accessing this document will incur an additional charge of $.
After purchase, you can access this document again without charge.
Accept $ ChargeStill Working On It
This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.
Give it another minute or two to complete, and then try the refresh button.
A few More Minutes ... Still Working
It can take up to 5 minutes for us to download a document if the court servers are running slowly.
Thank you for your continued patience.

This document could not be displayed.
We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.
You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.
Set your membership
status to view this document.
With a Docket Alarm membership, you'll
get a whole lot more, including:
- Up-to-date information for this case.
- Email alerts whenever there is an update.
- Full text search for other cases.
- Get email alerts whenever a new case matches your search.

One Moment Please
The filing “” is large (MB) and is being downloaded.
Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!
If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document
We are unable to display this document, it may be under a court ordered seal.
If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.
Access Government Site