`
`Attorney Dkt. No. 066964—8046.USOO
`
`REMARKS
`
`The content of new claims 72, 77—80, 82, 83, 86, 92, 93 and 96 is disclosed in
`
`US. Application No. 15/099,274. US. Application No. 15/151,012 is incorporated by
`
`reference in the present application. US. Application No. 15/151,012 (at 0003)
`
`incorporates US. Application No. 15/099,274 by reference.
`
`Support for the change to 00103 is at 0006 of 15/099,274.
`
`New paragraph 0015.01 is incorporated by reference from paragraph 0011 of
`
`14/789,156.
`
`New paragraph 0037.1
`
`is incorporated by reference from paragraph 0048 of
`
`15/099,274.
`
`New paragraphs 0051.1 to 0051.60 are incorporated by reference from
`
`paragraphs 48—108 of 15/099,274. The Figure and elements numbers in these
`
`paragraphs are change for consistentcy with the present application.
`
`New Figures 7, 8A-8G and 9-11 are the same as Figs. 1, 2A-2G and 3-5 of US
`
`Application No. 15/099,274.
`
`66964-8046.USOO/1370589951
`
`23
`
`
`
`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`Response to objected documents by the Examiner
`
`Sokolowski
`
`
`
`We fully agree with the Examiner that Sokolowski teaches a reduction of fat tissue in
`
`abdomen, buttocks and thighs. Further Sokolowski teaches a contactless application
`
`of the time-varying magnetic field to the patient.
`
`[Sokolowski: Abstract, par. 5]
`
`Heating the tissue
`
`We would like to point out that the Examiner incorrectly interpreted Sokolowski. The
`
`Examiner Matthews
`
`rejected Claims 41, 56 or 62 because she interpreted
`
`Sokolowski as applying heat
`
`to the patient.
`
`[based on Sokolowski: par. 50].
`
`Following the description Sokolowski teaches in par. 50: “Also shown is the power
`
`unit (14) belonging to the voltage supply of the device, which is configured and
`
`installed separately on account of the high heat levels produced, with output
`
`voltages of between 500 and 1000 V and the associated connecting cables (15)”.
`
`We interpret [Sokolwski: par. 50] in the sense that the heat is generated in the power
`
`unit (14). The power unit (14) is remote from the device due to the generated heat
`
`hence we presume that the heat is not applied to the patient. Due to this teaching we
`
`hope that the Examiner would reconsider her opinion.
`
`Further the heat may be generated in a coil. On the other hand, Sokolowski teaches
`
`various methods of
`
`removing the heat
`
`from the coil.
`
`[Sokolowski: par. 11-17]
`
`66964-8046.USOO/1370589951
`
`24
`
`
`
`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.U800
`
`Sokolowski teaches “The cross-section of the conductor is 1 cm2 to about 2 cm2.
`
`This large cross-section of the coil conductor, by comparison with that of coils
`
`used in site-selective magnetic field stimulation (focussing), reduces the ohmic
`
`resistance, thus reducing the Joule's heat”. [Sokolowski: par. 13] Referring to the
`
`cross-section of the conductor it equals a conductor diameter in a range of 11.28 to
`
`15.96 mm.
`
`Based on par. 13 Sokolowski teaches away from using conductors of smaller cross-
`
`section area since smaller cross-section area causes higher ohmic losses. It is given
`
`by the following equation:
`
`where PR is the power loss heat dissipation (W);p is the resistance (Q-m);
`
`/ is the
`
`length of wire (m); S is the surface area (m2);
`
`I is the current (A) and m is 1 kg of
`
`wire material.
`
`The larger cross-section area of the conductor the higher reduction of Ohmic losses
`
`is.
`
`Additionally, Sokolowski
`
`teaches “the coil may also be produced from high-
`
`frequency wires, by conventional technology, to avoid skin and proximity effects.
`
`However, an estimation of the skin depth shows that this is not necessary with a
`
`pulse duration of about 200 ps” [Sokolowski: par. 14]
`
`Based on this paragraph, we suppose that Sokolowski teaches away from using the
`
`high-frequency wire for the coil since it is not necessary for the invention.
`
`Continual surveillance of an operator
`
`Sokolowski teaches a feedback link including integrated camera for determining
`
`muscle contractions and optimizing treatment parameters based on the muscle
`
`contractions evaluated by a computer. [Sokolowski: par. 25]
`
`On the other hand, the term continual surveillance of an operator claimed should be
`
`interpreted in the sense that
`
`the continual surveillance is necessary for safe
`
`treatment. [0103]. The safe treatment means protecting the patient and/or the device
`
`from heat damage.
`
`66964-8046.USOO/1370589951
`
`25
`
`
`
`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`We hope that the Examiner would reconsider her position since such a feature can’t
`
`be evaluated by the device taught by Sokolowski. Further Sokolowski doesn’t protect
`
`the patient and/or the device. Sokolowski optimizes treatment parameters, there is no
`
`safety function.
`
`Marchitto
`
`
`
`We fully agree with the Examiner that Marchitto generates RF waves for treatment of
`
`a patient. [Marchitto: Abstract...]
`
`Marchitto teaches application of high-frequency alternating magnetic field to promote
`
`a production of new collagen fibers. [Marchitto: par. 15, 19...]
`
`Muscle contraction
`
`We disagree with the examiner that Marchitto teaches inducing a muscle contraction.
`
`Marchitto teaches a word muscle several times — in par. 33, 34, 70 and 71 in the
`
`sense that in Figs. 6 and 7 is a muscle. The nearest reference to the muscle is in par.
`
`51: “Treatment of skin wrinkles can sometimes employ botulism toxin, whereupon an
`
`injection of toxin in or around the nerves associated with the wrinkle temporarily relax
`
`the muscle leading to reduction in the appearance of the wrinkle.”
`
`66964-8046.USOO/1370589951
`
`26
`
`
`
`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`Further Marchitto teaches “These currents encounter resistance and the result is the
`
`generation of heat” [Marchitto: par. 47] The muscle contraction can’t be induced by
`
`heat
`
`We hope that the Examiner would reconsider her position with respect to inducing the
`
`muscle contraction by the high-frequency magnetic field since Marchitto has no
`
`reference of such a feature.
`
`Adipose cells reduction
`
`We disagree with the Examiner that Marchitto reduces adipose cells by RF waves.
`
`Marchitto teaches: “A scaffold or lattice structure may be placed within tissues to
`
`provide support. The structure may be fixed or fused in place using methods
`
`described herein. For example,
`
`localizing a scaffold made of polylactic acid, or a
`
`similar polymer in proximity to fat layers found in or under skin could prove beneficial
`
`in the treatment of cellulite fat
`
`By increasing the number of attachment points, or
`
`by fixing a mesh-like substrate in place to minimize bulge, or by heating the fat to
`
`achieve melting and flowing into the scaffold, cosmesis may conceivably be
`
`improved. The devices and methods described herein may be used to heat and melt
`
`fat layers.” [Marchitto: par. 49]
`
`There is no reference teaching that the fat cells are reduced and/or damaged by the
`
`heat. Melting the fat and flowing into the scaffold doesn’t cause any reduction.
`
`It
`
`causes redistribution of the fat. The redistribution is not the reduction.
`
`Furthermore Marchitto teaches a scaffold/lattice made of polylactic acid or similar
`
`polymer in the same paragraph. The polylactic acid (PLA) is biodegradable polymer
`
`which is not produced by a human body. PLA is used as biocompatible material.
`
`Hence based on this fact we can only guess how PLA is involved in the method
`
`taught by Marchitto. Further we can guess how the scaffold/lattice is inserted to the
`
`proximity of fat layer.
`
`Additionally, Marchitto teaches in par. 47: “The device may be used to induce
`
`changes in tissue by applying alternating magnetic fields to the tissue such that
`
`currents are induced within the tissue (eddy currents). These currents encounter
`
`resistance and the result is the generation of heat. These eddy currents form
`
`most efficiently where there is a moist, polar environment, thus enabling electron
`
`66964-8046.USOO/1370589951
`
`27
`
`
`
`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`displacement or ion flow within the target. Thus, for example, in skin, the generation
`
`of eddy currents in the moist underlying dermis is favored over the superficial
`
`dehydrated epidermis and stratum corneum layers, and the underlying, more
`
`non-polar adipose layer. This results in preferable heating of conductive
`
`tissues over non- or weakly-conductive tissues.”
`
`Marchitto teaches in par. 14: “Thus, the prior art is deficient in methods and devices
`
`for non-invasive and safer methods of skin rejuvenation induced by high frequency
`
`alternating magnetic fields that inductively heat skin tissues. The prior art is also
`
`deficient
`
`in the use of alternating magnetic fields to inductively heat
`
`conductive dermal tissues with a high degree of specificity. The prior art is also
`
`deficient in methods and devices to target during the skin rejuvenation process
`
`only conductive dermal tissue and have little effect on fat and skin surface
`
`fissues?
`
`Based on these paragraphs and the entire description of Marchitto every person
`
`having ordinary skills in the art would recognize that Marchitto teaches treatment of
`
`shallow conductive layer of a skin,
`
`i.e. dermis including water to promote collagen
`
`production.
`
`Furthermore,
`
`in these paragraphs Marchitto teaches that non—polar and weakly
`
`conductive tissues are not treated, i.e. adipose cells. Particularly in par. 47 Marchitto
`
`teaches
`
`that adipose layer
`
`is non-polar and non-conductive. Moreover,
`
`in
`
`Background of the invention Marchitto teaches that current methods lack sufficient
`
`specificity to treat only the dermal layer.
`
`Treatment area
`
`We disagree with the Examiner that Marchitto teaches at least one treatment area of
`
`thighs, saddlebacks, buttocks, abdomen, hips or arms.
`
`We accept
`
`that Marchitto teaches application of
`
`radiofrequency treatment
`
`to
`
`back/dorsa. [Marchitto: par. 72, 73]
`
`We assume this is the reason of rejection claimed treated area of torso.
`
`Continual surveillance of an operator
`
`Marchitto teaches: “the heating effect
`
`in tissue tends to change the electrical
`
`properties of the tissue. As the tissue alters the electrical
`
`impedance that
`
`the
`
`radiofrequency generator sees,
`
`it
`
`is possible to use this change to monitor the
`
`heating process. For example, as the impedance of the skin changes, the impedance
`
`66964-8046.USOO/1370589951
`
`28
`
`
`
`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`mismatch between the radiofrequency generator and the tissue/applicator increases
`
`and so the power reflected back into the radiofrequency generator increases.”
`
`[Marchittoz par. 43]
`
`The feedback based on Marchitto optimizes energy transfer to the patient,
`
`i.e_
`
`treatment parameters. The treatment parameters are optimized in order to maximal
`
`the energy transfer.
`
`On the other hand, the term continual surveillance of an operator claimed by 8046
`
`should be interpreted in the sense that the continual surveillance is not necessary for
`
`safe treatment.
`
`[8046: par. 103] The safe treatment means protecting the patient
`
`and/or the device from heat damage.
`
`We hope that the Examiner would reconsider her position since such a feature can’t
`
`be evaluated by the device taught by Marchitto. Marchitto optimizes treatment
`
`parameters, there is no safety function.
`
`Notwithstanding we explained above the differences of the objected documents and
`
`we changed the claims to overcome the objected documents.
`
`66964-8046.USOO/1370589951
`
`29
`
`
`
`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.U800
`
`CLAIM CHANGES
`
`Energy storage device — herein after as capacitor
`
`Magnetic field generating device — herein after as coil
`
`Switching device — herein after as switch
`
`Claim 70
`
`Claim 70 describes a voltage drop between two successive peak amplitudes
`
`output from the energy storage device,
`
`is not higher than 21%. This limitation was
`
`found to be patentable in Application No. 15/073,318 and in US Patent No.
`
`9,586,057, both assigned to the Applicant.
`
`Claim 71
`
`Claim 71 describes a device which includes a conductor of a diameter less
`
`than 3 mm, generating the time-magnetic field by the magnetic field generating
`
`device with a magnetic flux density in a range of 0.5 to 7 T, applying the time-varying
`
`magnetic field to at least one of thighs, saddlebags, buttocks, abdomen, hips,
`
`love
`
`handles, torso and/or arms of the patient; and generating the time-varying magnetic
`
`field with a treatment duty cycle is higher than 10 %. This combination of limitations is
`
`not suggested in the prior art
`
`Claim 72
`
`Claim 72 includes discharging energy from the energy storage device to the
`
`magnetic field generating device in order to generate a time-varying magnetic field
`
`with a magnetic flux density in a range of 0.1 to 7 T, an impulse duration in a range of
`
`10 to 900 ps and with a repetition rate in a range of 1 to 700 Hz; 0. generating
`
`the
`
`time-varying magnetic field with a treatment duty cycle is higher than 10 %; and
`
`applying the time-varying magnetic field to at
`
`least one of thighs, saddlebags,
`
`buttocks, abdomen, hips, love handles, torso and/or arms of the patient.
`
`Such combination of treatment parameters is beneficial for the treatment since the
`
`parameter may be fitted for large spectrum of patients, e.g. obese patient requires
`
`higher magnetic flux density compared to thin patient. It is due to a thickness of the
`
`fat layer.
`
`66964-8046.USOO/1370589951
`
`30
`
`
`
`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`The prior art does not disclose a duty cycle higher than 10 %. The duty cycle
`
`provides benefits to the treatment results.
`
`Exemplary treatment duty cycle of 10 % is illustrated below. The rectangles represent
`
`a number of pulses of some repetition rate such as 10 Hz, i.e. 2 seconds of continual
`
`stimulation (20 pulses total) and 18 seconds of no stimulation.
`
`
`
`The treatment duty cycle is important for the patient due to relaxation of the treated
`
`muscle. The muscle needs time for regeneration after a contraction. If the stimulation
`
`is continual for long time period then the muscle can’t relax and the muscle is
`
`exhausted in short
`
`time period which is
`
`insufficient
`
`for
`
`the treatment
`
`result.
`
`Furthermore a lactate accumulates in the muscle hence the muscle is painful after
`
`the treatment in the case of continual stimulation for a long time period.
`
`Targeting magnetic field to claimed body region is beneficial since these regions are
`
`prone to cellulite and/or adipose cells accumulation. The majority of patient wants
`
`improve the visual appearance of especially these body regions.
`
`Claim 73 — serial connection of energy storage device and magnetic field
`
`generating device
`
`Sokolowski teaches “The selected charge voltage of this reservoir capacitor
`
`(Ci) essentially determines the stimulation power and hence the amplitude level of
`
`the stimulation pulse that is to be generated. The actual energy store,
`
`the pulse
`
`66964-8046.USOO/1370589951
`
`31
`
`
`
`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`capacitor (C2),
`
`is charged up and recharged by means of the thyristor circuits A and
`
`B” [Sokolowski: par. 62] Sokolowski doesn’t teach such topology.
`
`m ‘N
`
`“mm
`
`‘i
`53“;\
`WW“...-
`mg
`f MN
`
`r
`
`M
`5.)
`
`‘u
`
`54,21“;
`
`a
`
`.
`a
`\xv.
`
`\
`
`g
`‘
`mi...“
`~ ”xx: t:
`.
`MW
`f
`E
`\“wnwmni
`
`(:1
`
`3
`§
`
`l
`
`;‘ 7'
`v:
`»»‘\
`
`f:
`
`§
`
`g
`
`-
`"
`i3“:
`j?3
`: £1
`i
`:22:
`mm.
`"x.
`K
`:‘é:
`x
`:.,:.:
`'
`"
`3*
`.
`2i“?
`i T
`a
`.
`§
`§
`g
`i
`Wm '
`m¢n\w\mn$n\\\wm\&v\Wm‘nxmwngékw“‘3:.3
`
`5-
`
`12
`
`’
`
`w:
`
`5
`
`{m
`
`Oppositely, the claimed serial connection of the capacitor and the coil is beneficial for
`
`the treatment device since this topology contributes to reducing energy losses.
`
`This topology enables discharging the capacitor directly to the coil. Further the
`
`topology enables using of a switch which is parallel to the serial connection of the
`
`capacitor and the coil.
`
`Claim 74 — controlled shorting
`
`Such a switch enables protecting the energy source from the high voltage load at the
`
`terminals of the energy source 19 during the second half-period of LC resonance.
`
`[0047]. Necessity of protective usage is avoided. The energy losses caused by
`
`protective losses are reduced and the device is more efficient. Further a number of
`
`components is reduced and manufacturing price is lower. [0048].
`
`Claim 75 — patient position
`
`These patients positions differ from Sokolowski since Sokolowski teaches a standing
`
`position of the patient.
`
`[Sokolowski: Fig.
`
`1] Further teaching of height variable
`
`pedestal also tends to interpret Sokolowski to treat the standing patient. [Sokolowski:
`
`Fig. 23, 49...]
`
`The claimed positions are more comfortable for the patient. Further in lying and/or
`
`seating position the patient may better relax the treated muscle and the treatment
`
`acceptability is higher. See above arguments concerning the duty cycle (Claim 72).
`
`66964-8046.USOO/1370589951
`
`32
`
`
`
`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`Claim 76 — litz-wire
`
`Litz-wire enables using thinner conductor while the total diameter of the complete
`
`conductor is equal. The thinner conductor is the energy loss caused by eddy currents
`
`is reduced. Using litz—wire in the coil highly reduces energy losses caused by eddy
`
`currents. The litz—wire
`
`is highly important for the device.
`
`It reduces the power
`
`consumption.
`
`Claim 77 — sensor
`
`Using such a sensor differs from Sokolowski which teaches a camera,
`
`i.e. optic
`
`sensor. [Sokolowski: par. 25, 50...] The sensor claimed measures different physical
`
`quantity. Further the claimed sensor measures the physical quantity of the device.
`
`Oppositely, Sokolowski
`
`teaches camera for capturing images of
`
`the patient.
`
`[Sokolowski: par. 50+52]
`
`Claim 78 — unintended event and notification
`
`Such feature overcomes Sokolowski and Marchitto since none of the objected
`
`documents teaches such features.
`
`These features are beneficial because they highly improve safe operation of the
`
`device and protect the patient from burns and/or movement of the metal object within
`
`proximity of the treatment device. The movement of the metal object may harm the
`
`patient. Hence the patient’s safety is highly improved. [0051.05]
`
`Claim 79 — disabling the treatment
`
`Such feature overcomes Sokolowski since Sokolowski teaches a sensor for adjusting
`
`treatment parameters based on the feedback.
`
`On the other hand, the application uses the sensor for safety reasons. The feedback
`
`from the sensor highly improves safe operation of the device and protects the patient
`
`from burns and/or movement of the metal object within proximity of the treatment
`
`device. The movement of the metal object may harm the patient. Hence the patient’s
`
`safety is highly improved. [0051.05]
`
`Furthermore, Sokolowski doesn’t enable disabling the treatment based on the
`
`camera.
`
`66964-8046.USOO/1370589951
`
`33
`
`
`
`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`Claim 80 — providing maximal treatment parameter
`
`Such a feature isn’t taught by any objected document.
`
`The benefit of this feature is providing non-interrupted operation of the treatment
`
`device during the treatment.
`
`It overcomes the state of art because commercially
`
`available devices stops the treatment in the case of maximal temperature is reached
`
`without any notification. The problem is that the operator waits until the temperature
`
`is lower and then the treatment may continue.
`
`The maximal treatment parameter suggestion avoids unpredictable stopping of the
`
`treatment. It uses operator friendly solution.
`
`Claim 81 — blower on circumference
`
`Such a position of the blower is highly beneficial. The blower is not influenced by the
`
`generated magnetic field. Eddy currents are not induced in the blower. The blower is
`
`not heated,
`
`it
`
`is not prone to service issues and the lifespan of the blower is
`
`enhanced.
`
`Claim 82 — parallel flow of cooling media
`
`Sokolowski teaches: “The air conveying channel (20) in the hollow frame (16) of the
`
`stand (1) comprises, at an end remote from the patient, the fan (10) with air outlet
`
`openings (34) which aspirates air through the air inlet openings (22) in the plastic
`
`housing (2) of the stimulation coil
`
`(3),
`
`through filter mats (23), past air guidance
`
`elements (not shown), the windings (24) of the stimulation coil (3) and the connecting
`
`leads (18, 19),
`
`in order to cool it” [Sokolowski: par. 55] Only the upper side of the coil
`
`(24) is cooled. The air direction is represented by red lines.
`
`66964-8046.USOO/1370589951
`
`34
`
`
`
`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`
`
`
`1;”;
`x";-
`
`
`2%
`17-
`EEK ..
`
`
`
`' fi‘é. 3%“ ""‘waz‘f 9»
`‘-
`
`
`
`{EEE
`
`Claim 83 — directing cooling upper over upper and lower side
`
`It is similar argument with Claim 82.
`
`Further directing the cooling media over both sides of the coil is beneficial since the
`
`cooling surface of the coil
`
`is the most achievable. The heat dissipation is improved
`
`and the coil temperature can be maintained lower. The lower the temperature is the
`
`higher
`
`reduction of energy losses caused by resistive losses is because the
`
`resistance corresponds with a temperature described by the equation below.
`
`R(t) = Ro[1 + 010‘ — to)]
`
`where R(t) is resistance at temperature t; R0 is resistance at temperature to; d is
`
`temperature coefficient of resistance and to is a fixed reference temperature.
`
`Claim 84- — applying RF waves and heating adipose cells
`
`Marchitto teaches away from using the RF waves for adipose cells heating because
`
`Marchitto teaches that adipose cells are not heat etc. Further Marchitto selectively
`
`heats dermis. See p. 6 and 7.
`
`Claim 85 — balun and transmatch
`
`Marchitto doesn’t teach such features.
`
`Balun is used for transformation of unbalance impedance to balanced impedance.
`
`Due to the ability the homogeneity of the heating is improved and the adipose cells
`
`may be selectively heated. [8046: par. 96]
`
`66964-8046.USOO/1370589951
`
`35
`
`
`
`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`Claim 86 — HW components and sensor, treatment parameters, body regions
`
`None of the objected documents teaches such a sensor.
`
`Further a combination of treatment parameters is beneficial for the treatment since
`
`the parameter may be fitted for
`
`large spectrum of patients, e.g. obese patient
`
`requires higher magnetic flux density compared to thin patient. It is due to a thickness
`
`of the fat layer.
`
`Moreover, Sokolowski teaches maximal magnetic flux density 0.1 T. The claimed
`
`value is 0.5 T.
`
`Claim 87 — serial connection of energy storage device and magnetic field
`
`generating device
`
`It is similar argument with Claim 73.
`
`Claim 88 — controlled shorting
`
`It is similar argument with Claim 74.
`
`Claim 89 — patient position
`
`It is similar argument with Claim 75.
`
`Claim 90 — litz-wire
`
`It is similar argument with Claim 75.
`
`Claim 91 — conductor diameter
`
`Sokolowski teaches away from low conductor diameter since Sokolowski is focused
`
`on resistive losses. See p. 4 and 5.
`
`Further
`
`the conductor diameter
`
`less than 3 mm was found as allowable in
`
`combination with the duty cycle.
`
`66964-8046.USOO/1370589951
`
`36
`
`
`
`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`Claim 92 — unintended event and notification
`
`It is similar argument with Claim 78.
`
`Claim 93 — disabling the treatment
`
`It is similar argument with Claim 79.
`
`Claim 94 — providing maximal treatment parameter
`
`It is similar argument with Claim 80.
`
`Claim 95 — blower on circumference
`
`It is similar argument with Claim 81.
`
`Claim 96 — parallel flow of cooling media
`
`It is similar argument with Claim 82.
`
`Claim 97 — applying RF waves and heating adipose cells
`
`It is similar argument with Claim 84.
`
`Claim 98 — balun and/or transmatch
`
`It is similar argument with Claim 85.
`
`Claim 99 — directing the radiofrequency waves
`
`It is similar argument with Claim 85.
`
`Further Marchitto is unable to provide such step because Marchitto doesn’t teach the
`
`balun. Due to functionality of the balun the radiofrequency waves are directed
`
`between the electrodes because balun generate “virtual ground”. On the other hand,
`
`Marchitto teaches no balun hence Marchitto directs the radiofrequency waves
`
`between the electrode and the ground. Marchitto is unable to selectively heat a layer
`
`including adipose cells but Marchitto is able to heat dermal layer. [par. 14; Summary]
`
`66964-8046.USOO/1370589951
`
`37
`
`
`
`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`Application 15/601,719 has claims similar to the claims listed above.
`
`In a
`
`6/27/2017 Office Action (Examiner Dorna, AU3735) rejected claims in 15/601,719 as
`
`anticipated by Sokolowski US 2015/0157873, and as obvious over Sokolowski
`
`in
`
`view of Phillips US 2013/0137918.
`
`In view of the foregoing,
`
`it
`
`is submitted that the claims are in condition for
`
`allowance. A Notice of Allowance is requested.
`
`Dated: September 25, 2017
`
`Respectfully submitted,
`
`Customer No. 34055
`Perkins Coie LLP
`
`Patent - LA
`PO. Box 1247
`
`Seattle, WA 98111—1247
`Phone:
`(310) 788-9900
`Fax:
`(206) 332-7198
`Email: KOhriner@perkinscoie.com
`
`PERKINS COIE LLP
`
`By:
`
`/KennethHOhriner/
`Kenneth H. Ohriner
`
`Reg. No. 31,646
`
`66964-8046.USOO/1370589951
`
`38
`
`
`
`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`Appendix
`
`Summary of newly added Claims relative to the outline for the Examiner Interview
`
`provided to the Examiner on September 20, 2017.
`
`Claim No. for
`
`Claim No. proposed
`
`Response
`
`for interview
`
`
`
`Additional new feature
`
`78+80 —
`70
`
`71
`78+81+83+84
`
`
`
`72
`
`73
`
`78+83+84
`
`78+83+84+77
`
`impulse duration+repetition
`
`rate
`
`
`
`
`
`independent
`independent
`
`independent
`
`independent
`
`82
`78+83+84+86
`
`
`83
`78+83+84+75
`
`
`84
`
`86
`
`78+83+84+41
`
`78+83
`
`sensor+impu|se duration+
`
`
`
`repetition rate
`
`92
`78+89
`unintended event
`
`
`disabling treatment
`78+90
`93
`
`
`66964-8046.USOO/1370589951
`
`39
`
`
`
`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`
`
`Claim No. for
`
`Claim No. proposed
`
`
`
`Response
`
`94
`
`95
`
`96
`
`97
`
`98
`
`99
`
`
`
`for interview
`
`Additional new feature
`
`78+91
`
`78+92+99
`
`78+93+86
`
`78+94+41
`
`78+95
`
`78+96
`
`maximal treatment parameter
`
`directing radiofrequency waves
`
`balun and/0r transmatch
`
`66964-8046.USOO/1370589951
`
`40
`
`