S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`REMARKS
`
`The content of new claims 72, 77—80, 82, 83, 86, 92, 93 and 96 is disclosed in
`
`US. Application No. 15/099,274. US. Application No. 15/151,012 is incorporated by
`
`reference in the present application. US. Application No. 15/151,012 (at 0003)
`
`incorporates US. Application No. 15/099,274 by reference.
`
`Support for the change to 00103 is at 0006 of 15/099,274.
`
`New paragraph 0015.01 is incorporated by reference from paragraph 0011 of
`
`14/789,156.
`
`New paragraph 0037.1
`
`is incorporated by reference from paragraph 0048 of
`
`15/099,274.
`
`New paragraphs 0051.1 to 0051.60 are incorporated by reference from
`
`paragraphs 48—108 of 15/099,274. The Figure and elements numbers in these
`
`paragraphs are change for consistentcy with the present application.
`
`New Figures 7, 8A-8G and 9-11 are the same as Figs. 1, 2A-2G and 3-5 of US
`
`Application No. 15/099,274.
`
`66964-8046.USOO/1370589951
`
`23
`
`

`

`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`Response to objected documents by the Examiner
`
`Sokolowski
`
`
`
`We fully agree with the Examiner that Sokolowski teaches a reduction of fat tissue in
`
`abdomen, buttocks and thighs. Further Sokolowski teaches a contactless application
`
`of the time-varying magnetic field to the patient.
`
`[Sokolowski: Abstract, par. 5]
`
`Heating the tissue
`
`We would like to point out that the Examiner incorrectly interpreted Sokolowski. The
`
`Examiner Matthews
`
`rejected Claims 41, 56 or 62 because she interpreted
`
`Sokolowski as applying heat
`
`to the patient.
`
`[based on Sokolowski: par. 50].
`
`Following the description Sokolowski teaches in par. 50: “Also shown is the power
`
`unit (14) belonging to the voltage supply of the device, which is configured and
`
`installed separately on account of the high heat levels produced, with output
`
`voltages of between 500 and 1000 V and the associated connecting cables (15)”.
`
`We interpret [Sokolwski: par. 50] in the sense that the heat is generated in the power
`
`unit (14). The power unit (14) is remote from the device due to the generated heat
`
`hence we presume that the heat is not applied to the patient. Due to this teaching we
`
`hope that the Examiner would reconsider her opinion.
`
`Further the heat may be generated in a coil. On the other hand, Sokolowski teaches
`
`various methods of
`
`removing the heat
`
`from the coil.
`
`[Sokolowski: par. 11-17]
`
`66964-8046.USOO/1370589951
`
`24
`
`

`

`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.U800
`
`Sokolowski teaches “The cross-section of the conductor is 1 cm2 to about 2 cm2.
`
`This large cross-section of the coil conductor, by comparison with that of coils
`
`used in site-selective magnetic field stimulation (focussing), reduces the ohmic
`
`resistance, thus reducing the Joule's heat”. [Sokolowski: par. 13] Referring to the
`
`cross-section of the conductor it equals a conductor diameter in a range of 11.28 to
`
`15.96 mm.
`
`Based on par. 13 Sokolowski teaches away from using conductors of smaller cross-
`
`section area since smaller cross-section area causes higher ohmic losses. It is given
`
`by the following equation:
`
`where PR is the power loss heat dissipation (W);p is the resistance (Q-m);
`
`/ is the
`
`length of wire (m); S is the surface area (m2);
`
`I is the current (A) and m is 1 kg of
`
`wire material.
`
`The larger cross-section area of the conductor the higher reduction of Ohmic losses
`
`is.
`
`Additionally, Sokolowski
`
`teaches “the coil may also be produced from high-
`
`frequency wires, by conventional technology, to avoid skin and proximity effects.
`
`However, an estimation of the skin depth shows that this is not necessary with a
`
`pulse duration of about 200 ps” [Sokolowski: par. 14]
`
`Based on this paragraph, we suppose that Sokolowski teaches away from using the
`
`high-frequency wire for the coil since it is not necessary for the invention.
`
`Continual surveillance of an operator
`
`Sokolowski teaches a feedback link including integrated camera for determining
`
`muscle contractions and optimizing treatment parameters based on the muscle
`
`contractions evaluated by a computer. [Sokolowski: par. 25]
`
`On the other hand, the term continual surveillance of an operator claimed should be
`
`interpreted in the sense that
`
`the continual surveillance is necessary for safe
`
`treatment. [0103]. The safe treatment means protecting the patient and/or the device
`
`from heat damage.
`
`66964-8046.USOO/1370589951
`
`25
`
`

`

`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`We hope that the Examiner would reconsider her position since such a feature can’t
`
`be evaluated by the device taught by Sokolowski. Further Sokolowski doesn’t protect
`
`the patient and/or the device. Sokolowski optimizes treatment parameters, there is no
`
`safety function.
`
`Marchitto
`
`
`
`We fully agree with the Examiner that Marchitto generates RF waves for treatment of
`
`a patient. [Marchitto: Abstract...]
`
`Marchitto teaches application of high-frequency alternating magnetic field to promote
`
`a production of new collagen fibers. [Marchitto: par. 15, 19...]
`
`Muscle contraction
`
`We disagree with the examiner that Marchitto teaches inducing a muscle contraction.
`
`Marchitto teaches a word muscle several times — in par. 33, 34, 70 and 71 in the
`
`sense that in Figs. 6 and 7 is a muscle. The nearest reference to the muscle is in par.
`
`51: “Treatment of skin wrinkles can sometimes employ botulism toxin, whereupon an
`
`injection of toxin in or around the nerves associated with the wrinkle temporarily relax
`
`the muscle leading to reduction in the appearance of the wrinkle.”
`
`66964-8046.USOO/1370589951
`
`26
`
`

`

`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`Further Marchitto teaches “These currents encounter resistance and the result is the
`
`generation of heat” [Marchitto: par. 47] The muscle contraction can’t be induced by
`
`heat
`
`We hope that the Examiner would reconsider her position with respect to inducing the
`
`muscle contraction by the high-frequency magnetic field since Marchitto has no
`
`reference of such a feature.
`
`Adipose cells reduction
`
`We disagree with the Examiner that Marchitto reduces adipose cells by RF waves.
`
`Marchitto teaches: “A scaffold or lattice structure may be placed within tissues to
`
`provide support. The structure may be fixed or fused in place using methods
`
`described herein. For example,
`
`localizing a scaffold made of polylactic acid, or a
`
`similar polymer in proximity to fat layers found in or under skin could prove beneficial
`
`in the treatment of cellulite fat
`
`By increasing the number of attachment points, or
`
`by fixing a mesh-like substrate in place to minimize bulge, or by heating the fat to
`
`achieve melting and flowing into the scaffold, cosmesis may conceivably be
`
`improved. The devices and methods described herein may be used to heat and melt
`
`fat layers.” [Marchitto: par. 49]
`
`There is no reference teaching that the fat cells are reduced and/or damaged by the
`
`heat. Melting the fat and flowing into the scaffold doesn’t cause any reduction.
`
`It
`
`causes redistribution of the fat. The redistribution is not the reduction.
`
`Furthermore Marchitto teaches a scaffold/lattice made of polylactic acid or similar
`
`polymer in the same paragraph. The polylactic acid (PLA) is biodegradable polymer
`
`which is not produced by a human body. PLA is used as biocompatible material.
`
`Hence based on this fact we can only guess how PLA is involved in the method
`
`taught by Marchitto. Further we can guess how the scaffold/lattice is inserted to the
`
`proximity of fat layer.
`
`Additionally, Marchitto teaches in par. 47: “The device may be used to induce
`
`changes in tissue by applying alternating magnetic fields to the tissue such that
`
`currents are induced within the tissue (eddy currents). These currents encounter
`
`resistance and the result is the generation of heat. These eddy currents form
`
`most efficiently where there is a moist, polar environment, thus enabling electron
`
`66964-8046.USOO/1370589951
`
`27
`
`

`

`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`displacement or ion flow within the target. Thus, for example, in skin, the generation
`
`of eddy currents in the moist underlying dermis is favored over the superficial
`
`dehydrated epidermis and stratum corneum layers, and the underlying, more
`
`non-polar adipose layer. This results in preferable heating of conductive
`
`tissues over non- or weakly-conductive tissues.”
`
`Marchitto teaches in par. 14: “Thus, the prior art is deficient in methods and devices
`
`for non-invasive and safer methods of skin rejuvenation induced by high frequency
`
`alternating magnetic fields that inductively heat skin tissues. The prior art is also
`
`deficient
`
`in the use of alternating magnetic fields to inductively heat
`
`conductive dermal tissues with a high degree of specificity. The prior art is also
`
`deficient in methods and devices to target during the skin rejuvenation process
`
`only conductive dermal tissue and have little effect on fat and skin surface
`
`fissues?
`
`Based on these paragraphs and the entire description of Marchitto every person
`
`having ordinary skills in the art would recognize that Marchitto teaches treatment of
`
`shallow conductive layer of a skin,
`
`i.e. dermis including water to promote collagen
`
`production.
`
`Furthermore,
`
`in these paragraphs Marchitto teaches that non—polar and weakly
`
`conductive tissues are not treated, i.e. adipose cells. Particularly in par. 47 Marchitto
`
`teaches
`
`that adipose layer
`
`is non-polar and non-conductive. Moreover,
`
`in
`
`Background of the invention Marchitto teaches that current methods lack sufficient
`
`specificity to treat only the dermal layer.
`
`Treatment area
`
`We disagree with the Examiner that Marchitto teaches at least one treatment area of
`
`thighs, saddlebacks, buttocks, abdomen, hips or arms.
`
`We accept
`
`that Marchitto teaches application of
`
`radiofrequency treatment
`
`to
`
`back/dorsa. [Marchitto: par. 72, 73]
`
`We assume this is the reason of rejection claimed treated area of torso.
`
`Continual surveillance of an operator
`
`Marchitto teaches: “the heating effect
`
`in tissue tends to change the electrical
`
`properties of the tissue. As the tissue alters the electrical
`
`impedance that
`
`the
`
`radiofrequency generator sees,
`
`it
`
`is possible to use this change to monitor the
`
`heating process. For example, as the impedance of the skin changes, the impedance
`
`66964-8046.USOO/1370589951
`
`28
`
`

`

`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`mismatch between the radiofrequency generator and the tissue/applicator increases
`
`and so the power reflected back into the radiofrequency generator increases.”
`
`[Marchittoz par. 43]
`
`The feedback based on Marchitto optimizes energy transfer to the patient,
`
`i.e_
`
`treatment parameters. The treatment parameters are optimized in order to maximal
`
`the energy transfer.
`
`On the other hand, the term continual surveillance of an operator claimed by 8046
`
`should be interpreted in the sense that the continual surveillance is not necessary for
`
`safe treatment.
`
`[8046: par. 103] The safe treatment means protecting the patient
`
`and/or the device from heat damage.
`
`We hope that the Examiner would reconsider her position since such a feature can’t
`
`be evaluated by the device taught by Marchitto. Marchitto optimizes treatment
`
`parameters, there is no safety function.
`
`Notwithstanding we explained above the differences of the objected documents and
`
`we changed the claims to overcome the objected documents.
`
`66964-8046.USOO/1370589951
`
`29
`
`

`

`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.U800
`
`CLAIM CHANGES
`
`Energy storage device — herein after as capacitor
`
`Magnetic field generating device — herein after as coil
`
`Switching device — herein after as switch
`
`Claim 70
`
`Claim 70 describes a voltage drop between two successive peak amplitudes
`
`output from the energy storage device,
`
`is not higher than 21%. This limitation was
`
`found to be patentable in Application No. 15/073,318 and in US Patent No.
`
`9,586,057, both assigned to the Applicant.
`
`Claim 71
`
`Claim 71 describes a device which includes a conductor of a diameter less
`
`than 3 mm, generating the time-magnetic field by the magnetic field generating
`
`device with a magnetic flux density in a range of 0.5 to 7 T, applying the time-varying
`
`magnetic field to at least one of thighs, saddlebags, buttocks, abdomen, hips,
`
`love
`
`handles, torso and/or arms of the patient; and generating the time-varying magnetic
`
`field with a treatment duty cycle is higher than 10 %. This combination of limitations is
`
`not suggested in the prior art
`
`Claim 72
`
`Claim 72 includes discharging energy from the energy storage device to the
`
`magnetic field generating device in order to generate a time-varying magnetic field
`
`with a magnetic flux density in a range of 0.1 to 7 T, an impulse duration in a range of
`
`10 to 900 ps and with a repetition rate in a range of 1 to 700 Hz; 0. generating
`
`the
`
`time-varying magnetic field with a treatment duty cycle is higher than 10 %; and
`
`applying the time-varying magnetic field to at
`
`least one of thighs, saddlebags,
`
`buttocks, abdomen, hips, love handles, torso and/or arms of the patient.
`
`Such combination of treatment parameters is beneficial for the treatment since the
`
`parameter may be fitted for large spectrum of patients, e.g. obese patient requires
`
`higher magnetic flux density compared to thin patient. It is due to a thickness of the
`
`fat layer.
`
`66964-8046.USOO/1370589951
`
`30
`
`

`

`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`The prior art does not disclose a duty cycle higher than 10 %. The duty cycle
`
`provides benefits to the treatment results.
`
`Exemplary treatment duty cycle of 10 % is illustrated below. The rectangles represent
`
`a number of pulses of some repetition rate such as 10 Hz, i.e. 2 seconds of continual
`
`stimulation (20 pulses total) and 18 seconds of no stimulation.
`
`
`
`The treatment duty cycle is important for the patient due to relaxation of the treated
`
`muscle. The muscle needs time for regeneration after a contraction. If the stimulation
`
`is continual for long time period then the muscle can’t relax and the muscle is
`
`exhausted in short
`
`time period which is
`
`insufficient
`
`for
`
`the treatment
`
`result.
`
`Furthermore a lactate accumulates in the muscle hence the muscle is painful after
`
`the treatment in the case of continual stimulation for a long time period.
`
`Targeting magnetic field to claimed body region is beneficial since these regions are
`
`prone to cellulite and/or adipose cells accumulation. The majority of patient wants
`
`improve the visual appearance of especially these body regions.
`
`Claim 73 — serial connection of energy storage device and magnetic field
`
`generating device
`
`Sokolowski teaches “The selected charge voltage of this reservoir capacitor
`
`(Ci) essentially determines the stimulation power and hence the amplitude level of
`
`the stimulation pulse that is to be generated. The actual energy store,
`
`the pulse
`
`66964-8046.USOO/1370589951
`
`31
`
`

`

`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`capacitor (C2),
`
`is charged up and recharged by means of the thyristor circuits A and
`
`B” [Sokolowski: par. 62] Sokolowski doesn’t teach such topology.
`
`m ‘N
`
`“mm
`
`‘i
`53“;\
`WW“...-
`mg
`f MN
`
`r
`
`M
`5.)
`
`‘u
`
`54,21“;
`
`a
`
`.
`a
`\xv.
`
`\
`
`g
`‘
`mi...“
`~ ”xx: t:
`.
`MW
`f
`E
`\“wnwmni
`
`(:1
`
`3

`
`l
`
`;‘ 7'
`v:
`»»‘\
`
`f:
`

`
`g
`
`-
`"
`i3“:
`j?3
`: £1
`i
`:22:
`mm.
`"x.
`K
`:‘é:
`x
`:.,:.:
`'
`"
`3*
`.
`2i“?
`i T
`a
`.


`g
`i
`Wm '
`m¢n\w\mn$n\\\wm\&v\Wm‘nxmwngékw“‘3:.3
`
`5-
`
`12
`
`’
`
`w:
`
`5
`
`{m
`
`Oppositely, the claimed serial connection of the capacitor and the coil is beneficial for
`
`the treatment device since this topology contributes to reducing energy losses.
`
`This topology enables discharging the capacitor directly to the coil. Further the
`
`topology enables using of a switch which is parallel to the serial connection of the
`
`capacitor and the coil.
`
`Claim 74 — controlled shorting
`
`Such a switch enables protecting the energy source from the high voltage load at the
`
`terminals of the energy source 19 during the second half-period of LC resonance.
`
`[0047]. Necessity of protective usage is avoided. The energy losses caused by
`
`protective losses are reduced and the device is more efficient. Further a number of
`
`components is reduced and manufacturing price is lower. [0048].
`
`Claim 75 — patient position
`
`These patients positions differ from Sokolowski since Sokolowski teaches a standing
`
`position of the patient.
`
`[Sokolowski: Fig.
`
`1] Further teaching of height variable
`
`pedestal also tends to interpret Sokolowski to treat the standing patient. [Sokolowski:
`
`Fig. 23, 49...]
`
`The claimed positions are more comfortable for the patient. Further in lying and/or
`
`seating position the patient may better relax the treated muscle and the treatment
`
`acceptability is higher. See above arguments concerning the duty cycle (Claim 72).
`
`66964-8046.USOO/1370589951
`
`32
`
`

`

`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`Claim 76 — litz-wire
`
`Litz-wire enables using thinner conductor while the total diameter of the complete
`
`conductor is equal. The thinner conductor is the energy loss caused by eddy currents
`
`is reduced. Using litz—wire in the coil highly reduces energy losses caused by eddy
`
`currents. The litz—wire
`
`is highly important for the device.
`
`It reduces the power
`
`consumption.
`
`Claim 77 — sensor
`
`Using such a sensor differs from Sokolowski which teaches a camera,
`
`i.e. optic
`
`sensor. [Sokolowski: par. 25, 50...] The sensor claimed measures different physical
`
`quantity. Further the claimed sensor measures the physical quantity of the device.
`
`Oppositely, Sokolowski
`
`teaches camera for capturing images of
`
`the patient.
`
`[Sokolowski: par. 50+52]
`
`Claim 78 — unintended event and notification
`
`Such feature overcomes Sokolowski and Marchitto since none of the objected
`
`documents teaches such features.
`
`These features are beneficial because they highly improve safe operation of the
`
`device and protect the patient from burns and/or movement of the metal object within
`
`proximity of the treatment device. The movement of the metal object may harm the
`
`patient. Hence the patient’s safety is highly improved. [0051.05]
`
`Claim 79 — disabling the treatment
`
`Such feature overcomes Sokolowski since Sokolowski teaches a sensor for adjusting
`
`treatment parameters based on the feedback.
`
`On the other hand, the application uses the sensor for safety reasons. The feedback
`
`from the sensor highly improves safe operation of the device and protects the patient
`
`from burns and/or movement of the metal object within proximity of the treatment
`
`device. The movement of the metal object may harm the patient. Hence the patient’s
`
`safety is highly improved. [0051.05]
`
`Furthermore, Sokolowski doesn’t enable disabling the treatment based on the
`
`camera.
`
`66964-8046.USOO/1370589951
`
`33
`
`

`

`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`Claim 80 — providing maximal treatment parameter
`
`Such a feature isn’t taught by any objected document.
`
`The benefit of this feature is providing non-interrupted operation of the treatment
`
`device during the treatment.
`
`It overcomes the state of art because commercially
`
`available devices stops the treatment in the case of maximal temperature is reached
`
`without any notification. The problem is that the operator waits until the temperature
`
`is lower and then the treatment may continue.
`
`The maximal treatment parameter suggestion avoids unpredictable stopping of the
`
`treatment. It uses operator friendly solution.
`
`Claim 81 — blower on circumference
`
`Such a position of the blower is highly beneficial. The blower is not influenced by the
`
`generated magnetic field. Eddy currents are not induced in the blower. The blower is
`
`not heated,
`
`it
`
`is not prone to service issues and the lifespan of the blower is
`
`enhanced.
`
`Claim 82 — parallel flow of cooling media
`
`Sokolowski teaches: “The air conveying channel (20) in the hollow frame (16) of the
`
`stand (1) comprises, at an end remote from the patient, the fan (10) with air outlet
`
`openings (34) which aspirates air through the air inlet openings (22) in the plastic
`
`housing (2) of the stimulation coil
`
`(3),
`
`through filter mats (23), past air guidance
`
`elements (not shown), the windings (24) of the stimulation coil (3) and the connecting
`
`leads (18, 19),
`
`in order to cool it” [Sokolowski: par. 55] Only the upper side of the coil
`
`(24) is cooled. The air direction is represented by red lines.
`
`66964-8046.USOO/1370589951
`
`34
`
`

`

`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`
`
`
`1;”;
`x";-
`
`
`2%
`17-
`EEK ..
`
`
`
`' fi‘é. 3%“ ""‘waz‘f 9»
`‘-
`
`
`
`{EEE
`
`Claim 83 — directing cooling upper over upper and lower side
`
`It is similar argument with Claim 82.
`
`Further directing the cooling media over both sides of the coil is beneficial since the
`
`cooling surface of the coil
`
`is the most achievable. The heat dissipation is improved
`
`and the coil temperature can be maintained lower. The lower the temperature is the
`
`higher
`
`reduction of energy losses caused by resistive losses is because the
`
`resistance corresponds with a temperature described by the equation below.
`
`R(t) = Ro[1 + 010‘ — to)]
`
`where R(t) is resistance at temperature t; R0 is resistance at temperature to; d is
`
`temperature coefficient of resistance and to is a fixed reference temperature.
`
`Claim 84- — applying RF waves and heating adipose cells
`
`Marchitto teaches away from using the RF waves for adipose cells heating because
`
`Marchitto teaches that adipose cells are not heat etc. Further Marchitto selectively
`
`heats dermis. See p. 6 and 7.
`
`Claim 85 — balun and transmatch
`
`Marchitto doesn’t teach such features.
`
`Balun is used for transformation of unbalance impedance to balanced impedance.
`
`Due to the ability the homogeneity of the heating is improved and the adipose cells
`
`may be selectively heated. [8046: par. 96]
`
`66964-8046.USOO/1370589951
`
`35
`
`

`

`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`Claim 86 — HW components and sensor, treatment parameters, body regions
`
`None of the objected documents teaches such a sensor.
`
`Further a combination of treatment parameters is beneficial for the treatment since
`
`the parameter may be fitted for
`
`large spectrum of patients, e.g. obese patient
`
`requires higher magnetic flux density compared to thin patient. It is due to a thickness
`
`of the fat layer.
`
`Moreover, Sokolowski teaches maximal magnetic flux density 0.1 T. The claimed
`
`value is 0.5 T.
`
`Claim 87 — serial connection of energy storage device and magnetic field
`
`generating device
`
`It is similar argument with Claim 73.
`
`Claim 88 — controlled shorting
`
`It is similar argument with Claim 74.
`
`Claim 89 — patient position
`
`It is similar argument with Claim 75.
`
`Claim 90 — litz-wire
`
`It is similar argument with Claim 75.
`
`Claim 91 — conductor diameter
`
`Sokolowski teaches away from low conductor diameter since Sokolowski is focused
`
`on resistive losses. See p. 4 and 5.
`
`Further
`
`the conductor diameter
`
`less than 3 mm was found as allowable in
`
`combination with the duty cycle.
`
`66964-8046.USOO/1370589951
`
`36
`
`

`

`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`Claim 92 — unintended event and notification
`
`It is similar argument with Claim 78.
`
`Claim 93 — disabling the treatment
`
`It is similar argument with Claim 79.
`
`Claim 94 — providing maximal treatment parameter
`
`It is similar argument with Claim 80.
`
`Claim 95 — blower on circumference
`
`It is similar argument with Claim 81.
`
`Claim 96 — parallel flow of cooling media
`
`It is similar argument with Claim 82.
`
`Claim 97 — applying RF waves and heating adipose cells
`
`It is similar argument with Claim 84.
`
`Claim 98 — balun and/or transmatch
`
`It is similar argument with Claim 85.
`
`Claim 99 — directing the radiofrequency waves
`
`It is similar argument with Claim 85.
`
`Further Marchitto is unable to provide such step because Marchitto doesn’t teach the
`
`balun. Due to functionality of the balun the radiofrequency waves are directed
`
`between the electrodes because balun generate “virtual ground”. On the other hand,
`
`Marchitto teaches no balun hence Marchitto directs the radiofrequency waves
`
`between the electrode and the ground. Marchitto is unable to selectively heat a layer
`
`including adipose cells but Marchitto is able to heat dermal layer. [par. 14; Summary]
`
`66964-8046.USOO/1370589951
`
`37
`
`

`

`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`Application 15/601,719 has claims similar to the claims listed above.
`
`In a
`
`6/27/2017 Office Action (Examiner Dorna, AU3735) rejected claims in 15/601,719 as
`
`anticipated by Sokolowski US 2015/0157873, and as obvious over Sokolowski
`
`in
`
`view of Phillips US 2013/0137918.
`
`In view of the foregoing,
`
`it
`
`is submitted that the claims are in condition for
`
`allowance. A Notice of Allowance is requested.
`
`Dated: September 25, 2017
`
`Respectfully submitted,
`
`Customer No. 34055
`Perkins Coie LLP
`
`Patent - LA
`PO. Box 1247
`
`Seattle, WA 98111—1247
`Phone:
`(310) 788-9900
`Fax:
`(206) 332-7198
`Email: KOhriner@perkinscoie.com
`
`PERKINS COIE LLP
`
`By:
`
`/KennethHOhriner/
`Kenneth H. Ohriner
`
`Reg. No. 31,646
`
`66964-8046.USOO/1370589951
`
`38
`
`

`

`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`Appendix
`
`Summary of newly added Claims relative to the outline for the Examiner Interview
`
`provided to the Examiner on September 20, 2017.
`
`Claim No. for
`
`Claim No. proposed
`
`Response
`
`for interview
`
`
`
`Additional new feature
`
`78+80 —
`70
`
`71
`78+81+83+84
`
`
`
`72
`
`73
`
`78+83+84
`
`78+83+84+77
`
`impulse duration+repetition
`
`rate
`
`
`
`
`
`independent
`independent
`
`independent
`
`independent
`
`82
`78+83+84+86
`
`
`83
`78+83+84+75
`
`
`84
`
`86
`
`78+83+84+41
`
`78+83
`
`sensor+impu|se duration+
`
`
`
`repetition rate
`
`92
`78+89
`unintended event
`
`
`disabling treatment
`78+90
`93
`
`
`66964-8046.USOO/1370589951
`
`39
`
`

`

`S/N 15/473,390
`
`Attorney Dkt. No. 066964—8046.USOO
`
`
`
`Claim No. for
`
`Claim No. proposed
`
`
`
`Response
`
`94
`
`95
`
`96
`
`97
`
`98
`
`99
`
`
`
`for interview
`
`Additional new feature
`
`78+91
`
`78+92+99
`
`78+93+86
`
`78+94+41
`
`78+95
`
`78+96
`
`maximal treatment parameter
`
`directing radiofrequency waves
`
`balun and/0r transmatch
`
`66964-8046.USOO/1370589951
`
`40
`
`

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.

We are unable to display this document.

PTO Denying Access

Refresh this Document
Go to the Docket