throbber
Case 6:12-cv-00799-JRG Document 124-8 Filed 03/07/14 Page 1 of 18 PageID #: 3880
`Case 6:12—cv—00799—JRG Document 124-8 Filed 03/07/14 Page 1 of 18 Page|D #: 3880
`
`EXHIBIT 8
`
`EXHIBIT 8
`
`
`
`

`
`Case 6:12-cv-00799-JRG Document 124-8 Filed 03/07/14 Page 2 of 18 PageID #: 3881
`
`United States Patent £191
`Kalotay et al.
`
`[11] Patent Number:
`[45] Date of Patent:
`
`5,009,109
`Apr. 23, 1991
`
`[54] FLOW TUBE DRIVE CIRCUIT HAVING A
`BURSTY OUTPUT FOR USE IN A CORIOLIS
`METER
`Inventors: Paul Kalotay, Lafayette; Robert
`Bruck, Boulder; Arnold Emc:h, Estes
`Park; Donald Martella, Louisville, all
`of Colo.
`
`[75]
`
`[73] Assignee: Micro Motion, Inc:., Boulder, Colo.
`[21] Appl. No.: 446,614
`Dee. 6, 1989
`(22] Filed:
`lat. a.s ................................................ GOlF 1/84
`[51]
`[52] u.s. a. .................................. 73/861.38; 331/154
`[58] Field of Search ............. 73/861.37, 861.38, 32 A;
`331/154, 155, 65
`
`[56]
`
`References Cited
`U.S. PATENT DOCUMENTS
`4,912,962 4/1990 Kawaguchi ........................ 73/32 A
`
`Primary Examiner-Herbert Goldstein
`Attorney, Agent, or Firm-Peter L. Michaelson
`
`[57]
`ABSTRACI'
`A drive circuit for providing bursts, rather than contin(cid:173)
`uously alternating amounts, of energy for use in driving
`a flow tube (conduit) in a Coriolis meter and methods
`for use in such a circuit. Specifically, the drive circuit
`provides a pre-defmed burst of energy to a drive coil
`afTued to a flow conduit at an appropriate point during
`a cycle of the oscillatory motion of the conduit in order
`to maintain the peak amplitude of the oscillatory motion
`substantially within a prescribed range. This burst can
`be applied at a pre-defmed point, illustratively the peak.
`in each cycle of the oscillatory motion with no energy
`being applied during that cycle other than when the
`pulse occurs in order to reduce the amount of electrical
`energy applied to the drive coil. Alternatively, to fur(cid:173)
`ther reduce this energy, a burst need not be applied
`during every such cycle but rather only at those pre(cid:173)
`defined points, e.g. the peaks, within those cycles where
`the velocity of the flow conduit is less than a pre(cid:173)
`defmed limit value.
`
`15 Claims, 6 Drawing Sheets
`
`100
`
`195
`
`185
`
`20
`
`26
`
`IETER
`lliCTFOHCS
`
`ANALOG & DIGITAl
`PROCESS OUTPUT SIGNALS
`
`RIGHT VELOCITY SIGNAL
`
`165R 165L
`
`MM0635546
`
`

`
`Case 6:12-cv-00799-JRG Document 124-8 Filed 03/07/14 Page 3 of 18 PageID #: 3882
`
`FIG. 1
`
`20
`
`I METER
`ELECTRONICS
`
`26
`
`I
`
`I
`
`I
`
`•
`
`ANALOG & DIGITAL
`PROCESS OUTPUT SIGNALS
`
`~ • rLJ •
`~ a. n> a
`
`~
`!"'
`N
`~
`"""'
`\C
`\C
`"""'
`
`Cll =(cid:173)(!)
`
`(T) -"""' Q ...,
`
`c:7\
`
`...
`U1
`0
`0
`...
`\0
`1-l
`0
`\0
`
`195
`
`185
`
`100
`RTO SIGNAL
`lc.DRIVE SIGNAL
`lc.LEFT VELOCITY SIGNAL
`![RIGHT VELOCITY SIGNAL
`I
`?
`{
`J
`165R 165L
`
`I
`
`;,~I''-.,
`
`-,., 'w·
`·120'
`
`:5:
`:5:
`0
`0) w
`
`(}1
`(}1
`
`~ ......
`
`

`
`Case 6:12-cv-00799-JRG Document 124-8 Filed 03/07/14 Page 4 of 18 PageID #: 3883
`
`I
`
`r------------, 165
`. L
`I
`LEFT
`I VELOCITY ~
`-=-
`: SENSOR
`I J
`RIGHT
`VELOCITY&:::
`SENSOR
`
`I
`
`I
`
`'
`
`FIG. 2
`
`r------------------------,
`r30
`
`I
`
`I
`
`I I .,
`
`MASS FLOW
`RATE
`CIRCUIT
`
`282
`
`I
`I
`
`40
`
`FLOW TUBE
`
`v20
`I
`I
`L---~~-:~u!~------~~~i~~!~~
`
`:
`
`I
`
`i 185
`tao i
`~~~
`-
`METER I
`L __ --;::r~~SE~~ 'U
`10
`
`:5:::
`:5:::
`0
`Q) w
`
`()1
`()1
`~
`CXl
`
`~ •
`rJl
`•
`
`'-= a ('D = """""
`
`>
`~
`~ .... \0
`....
`
`\0
`
`262
`
`263
`
`ftl
`
`00. =(cid:173)ftl
`....
`e,
`
`~
`
`Q'l
`
`Ul
`
`.. 0
`.. ,...,.
`
`0
`\C)
`
`0
`\C)
`
`

`
`Case 6:12-cv-00799-JRG Document 124-8 Filed 03/07/14 Page 5 of 18 PageID #: 3884
`
`FIG. 3
`FLOW TUBE DRIVE CIRCUIT
`
`447 ....
`
`SYNCH
`
`~449
`
`Un.L JL LL JLLj
`
`-
`
`-
`I
`
`I
`
`435
`NEGATIVE
`EDGE
`DETECTOR
`
`REFERENCE
`LEVEL (Vu~)
`VM-'
`439
`
`I -'1
`
`•I
`
`.I
`
`POSITIVE I ~ DRIVE
`
`LEVEL
`
`_[_ 180
`
`VL r
`
`"'
`
`'
`
`40
`
`VM I "'
`
`FIG. 3A
`tORI~~ I
`DRIVE COIL
`SIGNAL
`-DRIVE
`
`:
`
`00
`•
`
`~ ..
`~ =
`......
`n> =
`......
`
`> ,
`
`!""
`~
`JH
`"""
`\C
`\C
`"""
`
`r:n g-
`
`~
`CN
`Q ....
`
`0\
`
`...
`Cll
`0
`0
`... ......
`\0
`
`0
`\0
`
`~VL
`.190' PHASEI
`
`[431
`
`SHIFT
`
`• I
`I
`LEFT
`160L I
`~f VELOCITY
`I
`SENSOR
`
`:5:::
`:5:::
`0
`Q) w
`
`()1
`()1
`.j::>.
`c.o
`
`

`
`Case 6:12-cv-00799-JRG Document 124-8 Filed 03/07/14 Page 6 of 18 PageID #: 3885
`
`FIG. 4
`
`520
`
`'\
`
`I 5.5~5"\--
`
`Yin
`
`A/D
`
`c
`
`Do r-
`
`;a
`
`~1U
`
`f
`ltflUT SIGNAL
`- COtfliTIONING CIRCUIT
`
`rv
`~
`
`r
`
`160L
`·~
`i I
`...L
`
`r-
`
`~ - ENABLE ~ 54 a
`
`540
`
`GATE
`TIMER/COUNTER
`
`(
`
`OOT
`
`-
`
`~3
`
`:5:::
`:5:::
`0
`Q) w
`
`()1
`()1
`()1
`0
`
`J
`
`I 5~3
`
`555
`~
`
`530
`
`~CJ
`
`f
`
`I -
`
`fJ-P
`-- CHAtffl
`·-------..
`
`: RAM r ~535
`
`~-------·
`
`DMA
`
`c
`
`POWER
`SWITCH
`
`560
`
`TO
`
`DRI VE COIL
`'----· I
`~ I
`I --
`
`1.._""185
`
`DRIVE
`--·· ·-
`
`I
`
`I
`
`I
`I
`I
`
`~ •
`r.ll •
`
`~ a ('t) = """'
`
`>
`~
`
`~
`
`~~ ... :g ...
`
`~
`
`00
`="'
`fD a
`Q ....,
`"'
`
`...
`(II
`0
`0
`... .....
`\C)
`
`0
`\C)
`
`

`
`Case 6:12-cv-00799-JRG Document 124-8 Filed 03/07/14 Page 7 of 18 PageID #: 3886
`
`U.S. Patent
`
`Apr. 23, 1991
`
`Sheet 5 of 6
`
`5,009,109
`
`FIG. 5
`
`ENTER
`
`DRIVE CIRCUIT
`ROUTINE
`600
`
`610
`
`INSTRUCT TIMER/COUNTER 550 TO
`DELIVER INITIAL DRIVE PULSE (~.01 SEC)
`TO FLOW CONDUITS TO BEGIN
`OSCILLATORY MOVEMENT
`
`620
`
`INITIATE DMA TRANSFER FROM I/0 SPACE
`TO MEMORY ARRAY TO TRANSFER A REQUISITE
`NUMBER OF SAMPLES TO CHARACTERIZE ONE
`PERIOD OF FLOW CONDUIT MOTION
`
`645
`
`640
`
`PROCESS •NoN-DRIVE•
`RELATED METER FUNCTIONS
`
`690
`
`650
`
`DETERMINE MAXIMUM & MINIMUM VALUES OF SAMPLES IN ARRAY;
`CALCULATE ABSOLUTE VALUE OF PEAK, IVpeakl;and
`DETERMINE APPROXIMATE RESONANT FREQUENCY OF FLOW
`CONDUIT MOTION AND STORE FOR USE DURING NEXT
`DMA TRANSFER FROM I/0 SPACE
`
`APPLY SUITABLE LEVEL CHANGE TO
`;,..........-._..,. GATE Itf>UT OF TIMER/COUNTER
`550 TO ACTIVATE ITS PULSE
`NIDTH MODULATED
`OUTPUT
`
`675
`
`680
`
`685
`APPLY SUITABLE LEVEL CHANGE TO
`GATE INPUT OF TIMER/COUNTER 550 +----___. ______ _
`TO DEACTIVATE ITS PULSE NIDTH
`MODULATED OUTPUT
`
`MM0635551
`
`

`
`Case 6:12-cv-00799-JRG Document 124-8 Filed 03/07/14 Page 8 of 18 PageID #: 3887
`
`vu
`vPEAK
`2
`FIG. JL
`vREF ---...
`
`I
`
`I
`
`I
`
`r'
`
`"
`"
`
`I
`
`DRIVE
`WINO OW
`
`. .
`
`. .
`
`r--\.
`1 ' \0
`
`I
`
`"
`
`. .
`
`t ·I\
`
`DRIVE
`WINDOW
`
`I
`• • •
`
`I
`
`.
`
`('-LEFT
`
`VELOCITY
`SENSOR
`SIGNAL
`
`t -1\ I~-
`
`- FIG. 7
`
`0 ---·----------------·---------
`TRIANGULAR
`
`HALF
`SINUSOIDAL
`
`STEPPED WITH
`DECAYING TRAILING
`EDGE
`
`HAVERSINE
`
`~
`~
`
`0 m w
`
`01
`01
`01
`1\)
`
`•
`
`~ • 00
`"'C = f"'f'-
`('D =
`
`f"'f'-
`
`>
`'1:J :s
`~
`JN
`
`~
`\C
`\C
`~
`
`00
`:r'
`ft) a
`
`Q\
`~
`Q\
`
`740
`
`(II
`
`0
`\0
`
`.. 0
`.. ... 0
`
`\0
`
`

`
`Case 6:12-cv-00799-JRG Document 124-8 Filed 03/07/14 Page 9 of 18 PageID #: 3888
`
`1
`
`5,009,109
`
`FLOW TUBE DRIVE CIRCUIT HAVING A BURSTY
`OUTPUT FOR USE IN A CORIOLIS METER
`
`BACKGROUND OF THE INVENTION
`l. Field of the Invention
`The invention relates to apparatus for a drive circuit
`that provides bursts, rather than continuously alternat(cid:173)
`ing amounts, of energy for use in driving a flow tube
`(conduit) in a Coriolis meter and to methods for use in 10
`such a circuit.
`2. Description of the Prior Art
`Currently, Coriolis meters are fmding increasing use
`as an accurate way to measure the mass. flow rate and(cid:173)
`/or density of various process fluids in many applica- 15
`tions.
`Generally speaking, a Coriolis mass flow rate meter,
`such as that described in U.S. Pat. No. 4,491,025 (issued
`to J. E. Smith et al on Jan. 1, 1985), contains one or two
`parallel conduits, each typically being aU-shaped flow 20
`conduit or tube. Each flow conduit is driven to oscillate
`about an a.Xis to create a rotational frame of reference.
`For aU-shaped flow conduit, this axis can be termed
`the bending axis. As process fluid flows through each
`oscillating flow conduit, movement of the fluid pro- 25
`duces reactionary Coriolis forces that are orthogonal to
`both the velocity of the fluid and the angular velocity of
`the conduit. These reactionary Coriolis forces, though
`quite small when compared to the force at which the
`conduits are driven, nevertheless cause each conduit to 30
`twist about a torsional axis that, for a U-shaped flow
`conduit, is normal to its bending axis. The amount of
`twist imparted to each conduit is related to the mass
`flow rate of the process fluid flowing therethrough.
`This twist is frequently measured using velocity signals 35
`obtained from magnetic velocity sensors that are
`mounted to one or both of the flow conduits in order to
`provide a complete velocity proflle of the movement of
`each flow conduit with respect to either the other con(cid:173)
`duit or a fixed reference. In dual tube meters, both flow 40
`conduits are oppositely driven such that each conduit
`oscillates (vibrates) as a separate tine of a tuning fork.
`This "tuning fork" operation advantageously cancels
`substantially all undesirable vibrations that might other-
`wise mask the Coriolis force.
`In such a Coriolis meter, the mass flow rate of a fluid
`that moves through the meter is proportional to the
`time interval that elapses between the instant one point
`situated on a side leg of a flow conduit crosses a pre(cid:173)
`determined location, e.g. a respective mid-plane of os- 50
`cillation, until the instant a corresponding point situated
`on the opposite side leg of the same flow. conduit,
`crosses its corresponding location, e.g. its respective
`mid-plane of oscillation. For parallel dual conduit Cori(cid:173)
`olis mass flow rate meters, this interval is equal to the 55
`phase difference between the velocity signals generated
`for both flow conduits at the fundamental (resonant)
`frequency at which these flow conduits are driven. In
`addition, the resonant frequency at which each flow
`conduit oscillates depends upon the total mass of that 60
`conduit, i.e. the mass of the conduit itself, when empty,
`plus the mass of any fluid flowing therethrough. Inas(cid:173)
`much as the total mass varies as the density of the fluid
`flowing through the tube varies, the resonant frequency
`likewise varies with any changes in fluid density and as 65
`such can be used to track changes in fluid density.
`As noted above, these mass flow and density relation(cid:173)
`ships inherent in a Coriolis meter require that each flow
`
`45
`
`2
`conduit in the meter must be driven to resonantly vi(cid:173)
`brate in order for the meter to properly operate. To
`ensure that proper vibratory motion is initiated in, for
`example a dual tube Coriolis meter, and thereafter main-
`S tained during operation of the meter, the meter contains
`an appropriate drive mechanism that is mounted to both
`of the flow conduits typically between corresponding
`extremities of these conduits. Tlie drive mechanism
`fr~quently contains any one of many well known ar(cid:173)
`rangements, such as a magnet mounted to one conduit
`and a coil mounted to the other conduit in an opposing
`relationship to the magnet. A drive circuit continuously
`applies a periodic, typically sinusoidally or square
`shaped, drive voltage to the drive mechanism. Through
`interaction of the continuous alternating magnetic field
`produced by the coil in response to the periodic drive
`signal and the constant magnetic field produced by the
`magnet, both flow conduits are initially forced to vi(cid:173)
`brate in an opposing sinusoidal pattern which is thereaf(cid:173)
`ter maintained. Inasmuch as the drive circuit tightly
`synchronizes the frequency of the drive signal to essen-
`tially match the resonant frequency of the conduits,
`both flow conduits are kept in a state of opposing sub(cid:173)
`stantially resonant sinusoidal motion.
`One known drive circuit currently in use today and
`typified by that disclosed in, for example, U.S. Pat. No.
`4,777,833 (issued to B. L. Carpenter on Oct. 18,
`1988-hereinafter referred to as the '833 Carpenter
`patent-and currently owned by the present assignee)
`utilizes an analog drive circuit. Specifically, this circuit
`utilizes a synchronous analog amplifier to generate a
`continuous square wave with two analog levels that
`each equally change based upon a simultaneously oc-
`curring difference between an analog reference voltage
`and a flow conduit position signal. As the magnitude of
`this difference increases (decreases), based upon de(cid:173)
`creasing (increasing) amplitudes of the oscillatory
`movement of the flow conduits which results from, for
`example, increases (decreases) in the density in the pro(cid:173)
`cess fluid that simultaneously flows through the flow
`conduits, positive and negative drive levels produced
`by the synchronous amplifier corresponding and
`equally increase (decrease) to once again restore the
`amplitude of the oscillatory flow tube movement to its
`pro~r level. Various analog components, such as inter
`alia amplifiers, buffers, a phase shifter and an edge de(cid:173)
`tector, are used · to appropriately determine this differ(cid:173)
`ence based upon the analog reference level and one of
`the velocity sensor signals, typically a left velocity sen(cid:173)
`sor signal, produced within the meter.
`Unfortunately, analog drive circuits used in Coriolis
`meters and typified by that described in the '833 Car(cid:173)
`penter patent suffer from various drawbacks.
`First, analog drive circuits, particularly those which
`provide an alternating square shaped drive signal to the
`coil, do not permit the energy that is applied to the
`drive coil to be precisely controlled by the drive circuit
`itself at any one instant during the signal. With these
`circuits, the drive signal is merely set to alternate be(cid:173)
`tween two levels that are static within any one drive
`cycle. Precise control over the energy supplied to drive
`coil by the drive circuit itself has proven to be particu(cid:173)
`larly important in those applications, such as intended
`use of the meter particularly the mechanical Coriolis
`metering assembly itself in a hazardous environment,
`where a critical need exists to always limit this energy
`to as low a value as is realistically possible. While intrin-
`
`MM0635553
`
`

`
`Case 6:12-cv-00799-JRG Document 124-8 Filed 03/07/14 Page 10 of 18 PageID #: 3889
`
`5,009,109
`
`3
`sic safety barriers are used in these applications to limit
`the energy that would flow to the drive coil located in
`a hazardous area to below a pre-defined maximum
`amount and in doing so perform extremely well, it
`would be preferable to further limit the energy at its S
`source, if possible, i.e. drive circuit, and rely on the
`barrier as a back-up protective device rather than as a
`main mechanism for limiting the energy.
`Second, analog drive circuits generally tend to be
`complex and require a multitude of parts which adds to 10
`the manufacturing cost of the meter electronics.
`Third, discrete analog components, such as those
`used in a drive circuit, may exhibit undesirable tempera(cid:173)
`ture, aging and/or drift characteristics any one of which
`might, over time, cause the output produced by such a 15
`component to vary. These affects can be minimized to a
`certain and usually acceptable extent by using compo(cid:173)
`nents with matched temperature characteristics, appro(cid:173)
`priate temperature compensation circuits and/or suffi(cid:173)
`ciently frequent
`re-calibration. However, use of 20
`matched components further increases the cost of the
`meter electronics, while temperature compensation
`circuits often require additional components which
`increase the parts count as well as the manufacturing
`cost of the drive circuit. Re-calibration disadvanta- 25
`geously increases the costs associated with actual use of
`the meter.
`Therefore, a need exists in the art for a simple and
`inexpensive flow tube drive circuit particularly suited
`for use in a Coriolis meter that provides substantially 30
`accurate control over the amount of energy that is to be
`applied to the drive coil at any instant, has a reduced
`parts count and cost over analog drive circuits kitown
`in the art, and does not appreciably, suffer, if at all, from
`temperature, aging and/or drift affects which are com- 35
`monly associated with analog drive circuits kitown in
`the art.
`
`SUMMARY OF THE INVENTION
`An object of the present invention is to provide a 40
`drive circuit for use in a Coriolis meter that provides
`substantially accurate control over the amount of en(cid:173)
`ergy that is to be applied to the drive coil at any time.
`Another object is to provide such a drive circuit that
`generates a reduced amount of energy to the drive coil, 45
`as compared to that generated by drive circuits kitown
`in the art, but which is nevertheless sufficient to main(cid:173)
`tain the amplitude of the vibratory motion of the flow
`conduits at a desired level.
`Another object is to provide such a drive circuit that 50
`does not appreciably suffer from temperature, drift
`and/or aging affects commonly associated with analog
`drive circuits kitown in the art.
`Another object is to provide such a drive circuit that
`has a relatively low parts count and is relatively simple 55
`and inexpensive to manufacture.
`These and other objects are provided in accordance
`with the teachings of our inventive drive circuit which
`provides a pre-defmed burst of energy to a drive coil
`afflxed to a flow conduit at an appropriate point during 60
`a cycle of the oscillatory motion of the conduit in order
`to maintain the peak amplitude of the oscillatory motion
`substantially within a prescribed range. This burst can
`be applied at a pre-defmed point in each cycle of the
`oscillatory motion with no energy being applied during 65
`that cycle other than when the burst occurs in order to
`reduce the amount of electrical energy applied to the
`drive coil. Alternatively, to further reduce this energy,
`
`4
`a burst need not be applied during every such cycle but
`rather only at those pre-defined points within those
`cycles where the amplitude of the oscillatory motion of
`the flow conduit is less than a pre-defined limit value.
`In accordance with the teachings of a preferred em(cid:173)
`bodiment of our invention, our inventive drive circuit
`periodically samples the left velocity sensor signal
`throughout a single cycle of this signal using a pre(cid:173)
`defmed sampling period. These samples are transferred
`on a direct memory access (DMA) basis, using a well(cid:173)
`kitown cycle stealing technique, from an input/output
`space into a memory array, both situated within random
`access memory in a microprocessor. Transferring sam(cid:173)
`pled data values in this manner does not adversely and
`appreciably affect the throughput of the microproces(cid:173)
`sor. In response to the samples occurring throughout
`this cycle of the signal, the drive circuit, specifically the
`microprocessor contained therein, determines the zero
`crossings and maximum and minimum values of this
`cycle and thereafter calculates the absolute value of the
`peak of the cycle using the maximum and minimum
`values. Using two adjacent zero crossings contained
`within the cycle, the microprocessor also determines
`the approximate . frequency of the velocity signal and
`hence the approximate resonant frequency of the flow
`conduits. Once these operations have occurred, the
`· microprocessor compares the absolute value of the peak
`against a pre-defined limit value, V ref This comparison
`determines whether the amplitude of the vibratory mo(cid:173)
`tion of the flow conduits has decayed to a sufficiently
`low value to warrant the addition of a burst of energy to
`the drive coil and therethrough to the flow conduits in
`order to appropriately restore this amplitude. Specifi(cid:173)
`cally, in the event the absolute value of the peak is less
`than the limit value, then the microprocessor illustra(cid:173)
`tively gates a timer/counter circuit to generate a burst,
`such as a pulse, having a pre-defined shape to the drive
`coil within a specific window during the remainder of
`the cycle. Alternatively, if the absolute value of the
`peak is greater than the limit value, then no such pulse
`is generated by the timer/counter and hence no burst of
`energy is applied to the drive coil. Depending upon
`various mechanical characteristics of the flow tubes and
`the rate at which the density of the process fluid flow(cid:173)
`ing therethrough changes, several, perhaps quite a num(cid:173)
`ber, of cycles of oscillatory flow tube movement may
`elapse until the absolute value of the peak decays to a
`sufficiently low value to cause the drive circuit to apply
`a burst of energy to the flow tubes. In addition, the
`microprocessor, using the approximate value of the
`frequency of the velocity signal, determines the number
`of samples that need to be obtained during the next
`DMA transfer in order to fully characterize the next
`cycle of oscillatory flow tube movement and stores this
`number for use during subsequent initiation of that
`DMA transfer.
`Furthermore, a burst of energy can also be imparted
`to the drive coil at an appropriate point outside the
`window during a cycle(s) in order to remove a fmite
`amount of vibratory energy from the flow conduits and
`thereby effectively retard the peak value of these vibra(cid:173)
`tions, when necessary.
`In accordance with a feature of our invention, the
`drive circuit can adapt its performance to changing
`operating conditions of the Coriolis meter, such as
`changes in the density of the process fluid flowing
`through the meter, while imparting relatively minimal
`amounts of energy to the drive coil that are nevertheless
`
`MM0635554
`
`

`
`Case 6:12-cv-00799-JRG Document 124-8 Filed 03/07/14 Page 11 of 18 PageID #: 3890
`
`5,009,109
`
`5
`6
`tion, mass flow rate information is also provided in
`sufficient to sustain the flow tubes in resonant oscilla(cid:173)
`analog 4-20 rnA form over leads 26 for easy connection
`tory motion with a pre-defmed peak value. Specifically,
`to downstream process control and/or measurement
`the limit value can be changed, e.g. increased, when(cid:173)
`equipment.
`everthe rate of change in the absolute value of the peak
`Coriolis meter assembly 10, as shown, includes a pair
`is sufficiently high so that bursts of mechanical energy S
`of manifolds 110 and 110'; tubular member 150; a pair of
`can be rapidly added to the vibrating flow conduits,
`parallel flow conduits (tubes) 130 and 130'; drive mech(cid:173)
`such as over a larger number of successive cycles than
`anism 180; a pair of velocity sensing coils 160L and
`would otherwise occur. Adding bursts of energy in this
`160R; and a pair of permanent magnets 170L and 170R.
`fashion quickly compensates for increased attenuation
`10 Conduits 130 and 130' are substantially U-shaped and
`that occurs in the peak of the vibratory amplitude of the
`have their ends attached to conduit mounting blocks
`flow tubes, caused by large rapid increases in the fluid
`120 and 120', which are, in tum, secured to respective
`density. Moreover, whenever the absolute value of the
`manifolds 110 and 110'. Both flow conduits are free of
`peak amplitude reaches or exceeds the increased limit
`pressure sensitivejoints.
`value, the limit value can be appropriately decreased to
`15 With the side legs of conduits 130 and 130' fixedly
`a normal value in order to reduce the rate at which
`attached to conduit mounting blocks 120 and 120' and
`mechanical energy wil.l be imparted to the vibrating
`these blocks, in tum, fixedly attached to manifolds 110
`flow conduits.
`and 110', as shown in FIG. 1, a continuous closed fluid
`BRIEF DESCRIPTION OF THE DRAWINGS
`path is provided through Coriolis meter assembly 10.
`The teachings of the present invention may be clearly 20 Specifically, when meter 10 is connected, via inlet end
`understood by considering the following detailed de-
`101 and outlet end 101', into a conduit system (not
`shown) which carries the fluid that is being measured,
`scription in conjunction with the accompanying draw-
`ings, in which:
`fluid enters the meter through an orifice in inlet end 101
`FIG. 1 is an overall diagram of Coriolis mass flow
`of manifold 110 and is conducted through a passageway
`rate metering system 5;
`25 therein having a gradually changing cross-section to
`FIG. 2 depicts a block diagram of meter electronics
`conduit mounting block 120. There, the fluid is divided
`20 shown in FIG. 1;
`and routed through flow conduits 130 and 130'. Upon
`FIG. 3 is a block diagram of a prior art embodiment
`exiting flow conduits 130 and 130', the fluid is recom-
`of flow tube drive circuit 40;
`bined in a single stream within conduit mounting block
`FIG. 3A depicts various waveforms associated with 30 120' and is thereafter routed to manifold 110'. Within
`manifold 110', the fluid flows through a passageway
`drive circuit 40 shown in FIG. 3;
`FIG. 4 is a block diagram of a preferred embodiment
`having a similar gradually changing cross-section to
`of drive circuit 40 constructed in accordance with the
`that of manifold 110-as shown by dotted lines 105-to
`teachings of our present invention;
`an orifice in outlet end 101'. At end 101' the fluid reen-
`FIG. 5 depicts a flowchart of drive circuit routine 35 ters the conduit system. Tubular member 150 does not
`600 executed by microprocessor 530 shown in FIG. 4 to
`conduct any fluid. Instead, this member serves to axially
`generate a drive signal in accordance with the teachings
`align manifolds 110 and 110' and maintain the spacing
`of our invention;
`therebetween by a pre-determined amount so that these
`FIG. 6 is a waveform depicting two illustrative cy-
`manifolds will readily receive mounting blocks 120 and
`cles of the left velocity signal and the temporal relation- 40 120' and flow conduits 130 and 130'.
`ship between this velocity signal and the occurrence of
`U-shaped flow conduits 130 and 130' are selected and
`drive signal bursts produced by our inventive drive
`appropriately mounted to the conduit mounting blocks
`so as to have substantially the same moments of inertia
`circuit; and
`FIG. 7 depicts various illustrative waveforms each of
`and spring constants about bending axes W-W and
`which can be used to produce a drive signal burst.
`45 W'-W', respectively. These bending axes are perpen-
`To facilitate understanding, identical reference nu-
`dicularly oriented to the side legs of the U-shaped flow
`merals have been used, where appropriate, to designate
`conduits and are located near respective conduit mount-
`identical elements that are common to the figures.
`ing blocks 120 and 120'. The U-shaped flow conduits
`extend outwardly from the mounting blocks in an essen(cid:173)
`DETAILED DESCRIPTION
`SO tially parallel fashion and have substantially equal mo-
`ments of inertia and equal spring constants about their
`After reading the following description, those skilled
`respective bending axes. Inasmuch as the spring con-
`in the art will readily appreciate that our inventive drive
`stant of the conduits changes with temperature, resistive
`circuit can be utilized with nearly any Coriolis meter
`temperature detector (RTD) 190 (typically a platinum
`regardless of whether that meter is measuring mass flow
`rate, density or other parameter(s) of a process fluid. 55 RTD device) is mounted to one of the flow conduits,
`Nevertheless, for purposes of brevity, the inventive
`here conduit 130', to continuously measure the tempera-
`drive circuit will be discussed in the context of a meter
`ture of the conduit. The temperature of the conduit and
`that specifically measures mass flow rate.
`hence the voltage appearing across the RTD, for a
`FIG. 1 shows an overall diagram of Coriolis mass
`given current passing therethrough, will be governed
`flow rate metering system 5.
`60 by the temperature of the fluid passing through the flow
`As shown, system 5 consists of two basic compo-
`conduit. The temperature dependent voltage appearing
`nents: Coriolis meter assembly 10 and meter electronics
`across the RTD is used, in a well known method, by
`20. Meter assembly 10 measures the mass flow rate of a
`meter electronics 20 to appropriately compensate the
`desired process fluid. Meter electronics 20, connected
`value of the spring constant for any changes in conduit
`to meter assembly 10 via leads 100, illustratively pro- 65 temperature. The RTD is connected to meter electron-
`ics 20 by lead 195.
`vides mass flow rate and totalized mass flow informa-
`tion. Mass flow rate information is provided over leads
`Both of these flow conduits are sinusoidally driven in
`26 in frequency form and in scaled pulse form. In addi-
`opposite directions about their respective bending axes
`
`MM0635555
`
`

`
`Case 6:12-cv-00799-JRG Document 124-8 Filed 03/07/14 Page 12 of 18 PageID #: 3891
`
`5,009,109
`
`7
`and at essentially their common resonant frequency. In
`this manner, both flow conduits will vibrate in the same
`manner as do the tines of a tuning fork. Drive mecha(cid:173)
`nism 180 supplies the sinusoidal oscillatory driving
`forces to conduits 130 and 130'. This drive mechanism S
`can consist of any one of many well known arrange(cid:173)
`ments, such as a magnet mounted to illustratively flow
`conduit 130' and an opposing coil mounted to illustra~
`tively flow conduit 130 and through which an alternat(cid:173)
`ing current is passed, for sinusoidally vibrating both 10
`flow conduits at a common frequency. A suitable con(cid:173)
`tinuous alternating drive signal is applied by meter elec(cid:173)
`tronics 20, via lead 185, to drive mechanism 180.
`With fluid flowing through both conduits wJille these
`conduits are sinusoidally driven in opposing directions, 15
`Coriolis forces will be generated along adjacent side
`legs of each of flow conduits 130 and 130' but in oppo(cid:173)
`site directions, i.e. the Coriolis force generated in side
`leg 131 will oppose that generated in side leg 131'. This
`phenomenon occurs because although the fluid flows 20
`through the flow conduits in essentially the same paral-
`lel direction, the angular velocity vectors for the oscil(cid:173)
`lating (vibrating) flow conduits are situated in opposite
`though essentially parallel directions. Accordingly,
`during one-half of the oscillation cycle of both flow 25
`conduits, side legs 131 and 131' will be twisted closer
`together than the minimum distance occurring between
`these legs produced by just the oscillatory movement of
`the conduits generated by drive mechanism 180. During
`the next half-cycle, the generated Coriolis forces will 30
`twist the side legs 131 and 131' further apart than the
`maximum distance occurring between these legs pro(cid:173)
`duced by just the oscillatory movement of the conduits
`generated by drive mechanism 180.
`During oscillation of the flow conduits, the adjacent 35
`side legs, which are forced closer together than their
`counterpart side legs, will reach the end point of their
`travel, where their velocity crosses zero, before their
`counterparts do. The time interval which elapses from
`the instant one pair of adjacent side legs reaches their 40
`end point of travel to the instant the counterpart pair of
`side legs, i.e. those forced further apart, reach their
`respective end point is proportional to the total mass
`flow rate of the fluid flowing through meter assembly
`10. The reader is referred to U.S. Pat. No. 4,491,025 45
`(issued to J. E. Smith et. al. on Jan. 1, 1985) for a more
`detailed discussion of the principles of operation of
`parallel path Coriolis flow meters than that just pres(cid:173)
`ented.
`To measure the time interval, ~t, coils 160L and 160R so
`are attached to either one of conduits 130 and 130' near
`their free ends and permanent magnets 170L and 170R
`are also attached near the free ends of the other one of
`the conduits. Magnets 170L and 170R are disposed so as
`to have coils 160L and 160R located in the volume of SS
`space that surrounds the respective permanent magnets
`and in which the magnetic flux fields are essentially
`uniform. With this configuration, the electrical signal
`outputs generated by coils 160L and 160R provide a
`velocity profile of the complete travel of the conduit 60
`and can be processed, through any one of a number of
`well known methods, to determine the time interval
`and, in turn, the mass flow rate of the fluid passing
`through the meter. In particular, coils 160L and 160R
`produce. the left and right velocity signals that appear 65
`on leads 165L and 165R, respectively.
`As noted, meter electronics 20 accepts as input the
`R TD signal appearing on lead 195, and left and right
`
`8
`velocity signals appearing on leads 165L and 165R, re(cid:173)
`spectively. Meter electronic

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket