throbber
Case 6:12-cv-00799-JRG Document 124-5 Filed 03/07/14 Page 1 of 18 PageID #: 3763
`Case 6:12—cv—00799—JRG Document 124-5 Filed 03/07/14 Page 1 of 18 Page|D #: 3763
`
`EXHIBIT 5
`
`EXHIBIT 5
`
`
`
`

`
`Case 6:12-cv-00799-JRG Document 124-5 Filed 03/07/14 Page 2 of 18 PageID #: 3764
`
`·· Dec. 23, 1958
`
`Filed Au1. 26, 1954
`
`W. ROTH
`GYROSCOPIC MASS FLOWMETER
`
`2,865~201
`
`.4 Sheets-Sheet 1
`
`FIG. I
`
`FIG. 2
`
`FIG. 4
`
`FIG. 6
`
`FREQUENCY
`
`.
`INVENTOR.
`WILFRED. ROTH
`
`ATTOR~_E.'f.S
`
`1-l' I
`
`MM0634688
`
`

`
`Case 6:12-cv-00799-JRG Document 124-5 Filed 03/07/14 Page 3 of 18 PageID #: 3765
`
`Dec. 23, 1958 ·
`
`Filed Au1. 26, 1954
`
`W. ROTH
`GYROSCOPIC .MASS FLOWMETER
`
`2;865,201
`
`4 Sheets-Sheet 2
`
`C\1
`It)
`
`,
`
`It)
`
`1.()
`
`+
`m
`
`INVENTOR.
`WILFRED ROTH
`BY~ Ed--t
`·~~~~y~
`ATTORNEYS
`
`~JUOU"~,~·
`~
`
`MM0634689
`
`

`
`Case 6:12-cv-00799-JRG Document 124-5 Filed 03/07/14 Page 4 of 18 PageID #: 3766
`
`Dec. 23, 1958
`
`Filed Aui. 26, 1954
`
`FIG. 7
`
`W. ROTH
`GYROSCOPIC MASS FLOWMETER
`
`2,865,201
`
`4 Sheets-Sheet 3
`
`FIG. 7b
`
`6
`7c _ _j
`
`8
`
`FIG. 7c
`
`INVENTOR.
`WILFRED ROTH
`
`FIG. 7a
`
`'
`I
`I .
`I
`
`i
`
`~
`BY
`~ U-.~~-4dJ--.-.QI" ~
`ATTORNEYS
`
`r
`
`'
`
`MM0634690
`
`

`
`Case 6:12-cv-00799-JRG Document 124-5 Filed 03/07/14 Page 5 of 18 PageID #: 3767
`
`Dec. 23, 1958
`
`Filed Aui. 26, 1954
`
`W. ROTH
`GYROSCOPIC MASS FLOWMETER
`
`2,865,201
`
`4 Sheets-Sheet 4
`
`FIG. 8
`
`FREQUENCY
`
`FIG. 10
`
`FIG. 9
`
`FIG. II
`
`FIG. 12
`
`93
`
`. f
`
`(.,
`
`INVENTOR.
`WILFRED ROTH
`
`ATTORNEYS
`
`MM0634691
`
`

`
`Case 6:12-cv-00799-JRG Document 124-5 Filed 03/07/14 Page 6 of 18 PageID #: 3768
`
`United States PaterttOffice
`
`2,865,201
`Patented Dec. 23, 1958
`
`1
`
`2,865,201
`GYROSCOPIC MASS FLOWMETER
`Wilfred Roth, West Hartford, Conn.
`Application August 26, 1954, Serial No. 452,437
`30 Claims. (CI. 73....-194)
`
`5
`
`2
`strument, as described above. While the osCillation re(cid:173)
`moves the need for rotating joints,-the rotating flywheel
`is considered ]Jighly undesirable for the reasons given
`above.
`.
`It is a primary object of the present invention to proc
`vide a mass flowmeter of the gyroscopic type, wherein the
`loop is oscillated so as to avoid the need for rotating
`joints; and in whicli the need for a rotating flywheel is
`avoided. An oscillating instrument is here termed the
`10 A.-C. type. · Certain features _of the inven'tion, however,
`are applicable to a · continuously rotating instrument, here
`termed the D .-C. type. Although the apparatus of the
`invention is patticulariy useful in measuriJ:ig the mass
`flow of fluids, generally speakih.g it .is capable, with suit-
`15 able design paramet:ers; of rrieasuring the mass flow of
`any fluid-like material.: Such materials :include emui(cid:173)
`sions, slurries of solid particles -in a liquid or gaseous
`carrier, multi-phase mixtures of liquids or gases, etc:
`The invention _,viii be explained in conjunction with
`20 the accompanying drawings, and certain features will · in
`part be pointedmit anci in part be evident from the draw(cid:173)
`ings and description thereof.
`.In the drawings:
`Fig. 1 is a side view of an A .-C; mass flowmeter; Figs.
`25 la and lb are details illustrating suitable damping means;
`Fig. 2 is a view at right angles to that of Fig. 1;
`.
`Fig. 3 is a detail showing the arrangement ofthe inlet
`and outlet conduit sectrons;
`Fig. 4 shows curves to explain: the non-resonant opera(cid:173)
`~') tion of the apparatus of Figs. 1-3; .
`Fig. 5 is acircuitdiagrarn of an indicating device which
`may be used with the apparatus of Figs. 1~3;
`Fig. 6 is another embodiment of an A.-C. mass flow(cid:173)
`meter with· simplified indication;
`Fig. 7 is a further embodiment of an A.-C .. mass flow-
`meter;
`Fig. 7 a is a detail showing the iriiet and outlet con-
`duit sections;
`. .
`.
`Figs. 7 b and 7 c are · details of a torque drive which
`
`35
`
`This invention relates to mass flowmeters utilizing
`the gyroscopic principle. The invention is especially di(cid:173)
`rected to the provision of satisfactory A.-C. or oscillat(cid:173)
`ing flowmeters, as <distingl1ished from those of the D.-C.
`or continuously rotating type, although certain features
`are applicable to the latter.
`There is a considerable need in industry for an instru(cid:173)
`ment which will measure mass flow, as distinguished from
`In many industrial processes it is the mass
`volume flow.
`of a reagent that is important, rather than m erely volume.
`Also, it is often advantageous to market fluid-li"ke mate(cid:173)
`rials according to their mass rather than volume. While
`mass .flow is the product of volume flow and density, the
`den·sity may vary . depending upon the exact constituents
`of the material, and usually varies considerably with tem(cid:173)
`perature. Thus the conversion of volume flow to mass
`flow is often difficult. Even when such conversion is
`possible, it is advantageous to have an instrument which
`indicates mass flow directiy.
`It has been suggested to employ the gyroscopic prin(cid:173)
`ciple in order to measure mass flow directly. In such an
`instrument the fluid-like material is caused to flow in a
`curved conduit, specifically a conduit in the fonn of a
`loop. For a given fluid and conduit, the angular mo(cid:173)
`mentum varies with the rate of flow of the fluid thr.ough
`40 may be employed with.the apparatus of Fig. 7;
`the conduit. By virtue of the flowing fluid, the conduit
`Figs. 8 and 9 show curves illustrating the resonant op-
`is equivalent to the rotor of an ordinary gyroscope. if
`eration of the apparatus of Fig. 7;
`.
`.
`the loop is caused to rotate about an axis perpendicular
`Figs. 10 and 11 are details of an alternative form of
`to that of the. angular momentum, a torque will be pro(cid:173)
`torque drive which may be employed in the apparatus of
`If, for ex(cid:173)
`duced about the mutually · orthogonal axis.
`45 Fig; 7; and

`ample, the loop is circular and· is caused to rotate about
`Fig. 12 is a dia:gtain showing a torque feedback sys(cid:173)
`a diameter thereof by a drive source, a torque or couple
`tem in accordance-with the invention.
`will be produced about an axis mutually perpendicular
`. . Referring .now to Fig. l, a fluid conduit:IO is arranged
`to the axis of rotation and the axis of the .loop. T he in(cid:173)
`in the .forhl of a loop and attached to support members
`.stantaneous value of this torque_ will be propm:tional to
`50 31, 31'.· As specifi:cally shown the loop is circular, but
`the instantaneous value of the angular inom'enttlm as de(cid:173)
`other configurations· could-be employed if desired." Inlet
`termined by the tate of mass flow of the fluid, and. the
`and outlet fluid conduit sections 11 and 12 extend. from
`instantaneous value of the angular velocity of the .loop
`adjacent points 13', 13'. of the loop to approximately the
`about the drive axis.

`center of the loop: · As here shown,- conduit sections :u
`. In one instniment of this generaJtype which has been
`55 and 12 are ·of. flexible' hose a:n:d secured to the horizontal
`proposed, continuous :rotation of. the lo_op about one axis
`support member 14 by a band 15: Or, the sections H; 12
`has been employed, and a rotating mass mounted ·con(cid:173)
`can be extensions of:the tubing of'loop 10,. extending in(cid:173)
`centrically with the . axis of the ]oop has been: driven at
`wardly to the loop ax!sin the manner shown but without
`an ar1g~lar velocity .. controlled by· gyroscopic cp.uples pro(cid:173)
`the . restraining band 15; and flexible couplings attached
`duced by the flowing liquid, but in a counter direction, so
`60 to the tube sections near the center of the loop.
`that the an·gular. momentum of the flowing·liquid is coun(cid:173)
`The loop 10 is· mounted- for angular . movement with
`teracted by the angular momentum of tb,e rotating mass.
`This produces a null type instrument. · The use of. a rotat-
`respect'to member 14 by suitable means · which .are her~
`ing mass in this manner is ·considered undesirable be(cid:173)
`shown as short lengths of music wire 16, 16'. Thus, the
`cause of the added weight and complexity involved, to(cid:173)
`65 loop 10 is mounted for: angular movement about im axis
`approximately in the plane- of the loop, and the lengths
`gether withthe need.for careful maintenance. Further(cid:173)
`of music wire. form torsionill springs which produce· a-re(cid:173)
`more; a continuously rotating loop requires :sealed rotat-
`storing moment when the loop 10 is··angularly deflected oh
`ing .bearings which ate relatively expensive, require care-
`either side of the. central position illustrated,
`ful maintenance and may be troublesome with chemically
`The loop· and. its ·associated support member 14 is
`active·fluids: .or) fluids at:high pressure: . . .
`.
`.
`.
`mounted for rotation · about an . axis approximately per(cid:173)
`It has also· been· proposed ·to oscillate the loop, imd
`pendicular to-that of .rnember 14 by a member 17, here
`employ· a rotating .flywheeJ:to .'produce a null type in-
`
`70
`
`MM0634692
`
`

`
`Case 6:12-cv-00799-JRG Document 124-5 Filed 03/07/14 Page 7 of 18 PageID #: 3769
`
`2,865,201
`
`3
`shown as mounted for rotation about a vertical axis by
`bearings 18, 18' in a suitable housing 19. The loop 10
`may be oscillated about the vertical axis by motor 21 and
`eccentric cam 22 bearing against a rod 23 affixed to ver(cid:173)
`tical member 17 by a collar 24. Spring means 25 is at(cid:173)
`tached at one end to rod 23 and at the other end to a
`stationary support 26 so as to hold rod 23 in engagement
`.with eccentric cam 22. For convenience of illustration,
`the point at which the spring 25 is attached to support 26
`is shown lying above the rod, but in practice it will be
`understood that the point of attachment will ordinarily
`be substantially on a line with rod 23. While an eccentric
`cam is specifically illustrated, any other suitable means
`for oscillating the loop about the vertical axis may be
`employed.
`Since a constant frequency of oscillation at the selected
`frequency of operation is desirable for accuracy, motor 21
`is advantageously of the synchronous type. Other types
`may of course be employed if due care is taken to assure
`constant speed under existing operating conditions.
`In operation, fluid is supplied to the loop through one
`of conduit sections 11, 12 and led away from the loop
`through the other section.
`In flowing through loop 10,
`the mass of the fluid creates an angular momentum which
`is proportional to the rate of mass flow of the fluid. When
`the loop is rotated about the vertical axis, a torque is
`developed about the horizontal axis of member 14 which
`is proportional to the vector product of the instantaneous
`value of the angular momentum and the instantaneous
`value of the angular velocity about the vertical axis.
`Both of these quantities have direction as well as magni(cid:173)
`tude. Hence if the direction of fluid flow or angular
`velocity is reversed, the resulting torque will be reversed.
`With a sinusoidal oscillation, as provided by the drive
`motor 21 and cam 22 in Fig. 1, the resulting torque will
`also be sinusoidal.
`The movement produced by this torque is restrained by
`music wires 16, 16' and hence the loop 10 oscillates
`about the horizontal axis at the frequency of the vertical
`oscillation and with an amplitude proportional to the rate
`of mass flow.
`A transducer is associated with the loop which is sensi(cid:173)
`tive to gyro-scopic couples produced by the loop about the
`horizontal axis of 14. As here illustrated, the transducer
`is of the velocity type so as to yield an output proportional
`to the angular velocity of the-loop about the axis of 14.
`In the form illustrated, a coil 27 is attached to the loop 10
`and a portion thereof moves in an air gap of magnet 28.
`The magnet may be of the permanent type or magnet(cid:173)
`ized by a suitable field coil. Connections 29 to the coil
`are provided so ·that the electric potential induced in the
`coil as the position of the coil in the magnetic field varies
`may be supplied to an indicating instrument.
`Although many types of transducers to measure dis(cid:173)
`placement or ·its time derivatives, or stresses, strains and
`the like, can be employed, such as resistance wire strain
`gauges, magnetostrictive strain gauges, piezoelectric strain
`gauges, differential transformers, etc., those of the veloc(cid:173)
`ity typ~ are preferred at the present time.
`In a structure of the type described, the maximum or
`peak angular displacement of the loop from its central
`position varies with the ·rate of mass flow. Also, the
`angular acceleration of the loop is a maximum at maxi(cid:173)
`mum displacement. If a transducer is employed which
`is responsive to either displacement or acceleration the
`peak in~tantaneous values of the output will occur ~hen
`the loop is at its maximum excursion from the zero or
`neutral position. This varies with each value of rate
`of mass flow. Consequently, if an output linear with rate
`of mass flow is desired, the peak displacement of the loop
`must be linear over the desired range of mass flow
`measurements. This may be difficult to achieve in prac(cid:173)
`tice since any non-linearity in the restoring moment of
`torsional springs 16, 16' or non-Iinearities els<;where in
`
`5
`
`4
`the system will impair the linearity of the transducer
`output.
`.On the other hand, the angular velocity of the loop is
`a maximum as the loop passes through the zero or neu-
`tral position. Consequently, if a transdt:cer responsive
`to the velocity of the loop is employed, the peak in(cid:173)
`stantaneous values of the output will always be produced
`at the same point in the oscillating arc of the loop, namely,
`at the zero position thereof, regardless of the rate of mass
`1 o flow therethrough. Thus, the effect of non-linearities in
`the oscillatory motion of the loop is greatly reduced or
`entirely eliminated.
`A velocity-type transducer or pick-up is particularly
`advantageous· when combined with a peak detector cir-
`15 cuit. An example of such a circuit will be described here(cid:173)
`inafter in connection with Fig. 5.
`An important aspect of the present invention lies in
`the arrangement of the conduit sections for leading fluid
`to and from the loop. Among the advantages of the ar-
`20 rangement provided is that of preventing so-called Coriolis
`forces from affecting the accuracy of measurement.
`When a pipe or conduit containing a flowing fluid is
`subjected to angular movement transverse to its axis, the
`walls of the pipe must exert a force on the flowing fluid
`25 to impart angular acceleration thereto. This is known
`as Coriolis force. The force varies with rate of mass flow
`of the fluid in the pipe, and in an apparatus of the type
`herein considered would introduce an error unless the
`force is eliminated or the apparatus designed so that the
`30 force does not affect the output.
`In the apparatus of Fig. 1, conduit section 11 rotates
`about the drive axis 17 and hence, when fluid is flowing
`outwardly from the center, a Coriolis force is present
`which is substantially in a plane perpendicular to the
`35 drive axis (horizontal plane as specifically illustrated)
`and creates a torque about the drive axis. Similarly, a
`Coriolis force is present due to fluid flow in conduit sec(cid:173)
`tion 12, but since the fluid flow is inwardly toward the
`center, the force and resulting torque about the drive
`40 axis opposes that of conduit 11. Hence, the effects of
`the tw() Coriolis forces substantially cancel arid do not
`affect the output indication.
`In the arrangement of Fig. 1 a constant velocity drive
`source is employed, that is, a drive source whose angular
`4.) velocity is relatively unaffected by the load thereon. In
`this case it is not essential that the Coriolis forces cancel,
`so long as they are effective only about the drive axis,
`since only an additional load would be imposed on the
`drive source and the output would be unaffected. This is
`;;u accomplished in Fig. 1 by leading fluid to and from the
`loop 10 by conduits substantially parallel to and closely
`adjacent the horizontal axis 14 about which the loop
`moves to produce an output signal, that is, the output
`axis of rotation. This relationship is helpful in the event
`55 that perfect cancellation of Coriolis forces is not obtained
`by the parallel counterflow feed. Rotation of the feed
`lines by 90° is of course possible with a close parallel
`counterflow arrangement, or where the unbalance in
`Coriolis force is sufficiently small for the intended appli-
`60 cation.
`An added feature is that by connecting the inlet and
`outlet conduit sections 11, 12 to external pipe lines, etc.,
`from points near the intersection of the axes of vertical
`member 17 and horizontal member 14, and by providing
`65 flexibility in the conduit sections near the intersection of
`the axes, any restraint in the freedom of the loop to oscil(cid:173)
`late due to the connections to the external pipe line, etc.
`is reduced to a very small or negligible amount, since the
`In
`moment arm about the center of rotation is small.
`70 Fig. 1 this is accomplished by employing flexible tubes or
`conduit sections 11, 12 and providing substantially right
`angle bends in the sections, as shown in Figs. 2 and 3.
`Fig. 7, to be described hereinafter, shows an alternative
`In both figures, the loop
`arrangement to the same end.
`75 structure can be dynamically balanced by adding or re-
`
`MM0634693
`
`

`
`Case 6:12-cv-00799-JRG Document 124-5 Filed 03/07/14 Page 8 of 18 PageID #: 3770
`
`·s
`plotted in Fig, 4 since it is . the quantity commonly
`measured; it' being -:understood' that fx:equency is
`1
`2ir
`
`times angular velocity.
`When the driving frequency approaches the natural
`resonant frequency f' of the loop for a liquid of given
`density, a very ·targe increase in displacement is ob(cid:173)
`tained, as shown by curves 32, 32'. In Fig. 4 damping
`has been neglected so that curves 32, 32' would intersect
`at infinity. In any practical system, damping is of course
`If the
`present and will affect the shape of the curves.
`driving frequency is increased appreciably beyond the
`natural resonant frequency, the displacement lim drops
`off inversely with frequency.
`With a fluid of lower density, similar curves will be
`obtained but the natural resonant frequency will lie at a
`higher frequency such as shown at f" in Fig. 4. For a
`20 fluid of still lower density the natural resonant frequency
`will be still higher, as indicated at f'".
`As above mentioned, in accordance with one feature
`of the present invention it is contemplated applying a
`driving frequency which will be sufficiently low· com~
`pared with the natural . resonant frequency of the lo{)p
`when filled with fluid of density within a predetermined
`range so that the apparatus will operate within the region
`35 indicated in Fig. 4.
`For a given selected frequency in this range, i. e., for
`
`30
`
`moving .weight·from-member 14:on :the:•opp_osite ~ide of
`the vertiCal. axis from. the. conduit -sections,
`In an arrangementsuch as shown in fig; -1;. the:loop :10
`has a natural.re:sonant ·frequency of oscillation about the
`axis of member .14 due to the moment oLinertia of the 5
`loop and the restoring moment provided by · torsional
`springs 16, 16'.
`In. accordance with well-known · prin,
`ciples of mechanics, the moment of inertia ·of. the loop
`10 will include not only the mass and configuration of
`the conduit itself, but also the mass and location of any 10
`members associated therewith such as. the pickup coil 27
`and the inwardly projecting supports 31, 31'. Also, any
`stiffness of the conduit sections 11, 12 must be taken :into
`account along with the stiffness of the tGrsional springs
`16, 16'. Furthermore, the natural resonant · frequency 15
`of the loop 10 will be . affected by the mass of the fluid
`contained therein. While the volume of the fluid is essen(cid:173)
`tially fixed for a given instrument, the effective mass of
`the loop when filled with fluid wiil . vary with the · density
`of the fluid.
`It has been found very important to select properly
`the frequency of ·oscillation of the loop about the vertical
`axis with resp.ect to the natural resonant frequency of the
`loop about the horizontal axis in order to obtain an accu(cid:173)
`rate indication of mass flow when metering fluids of 25
`varying density, or to meter fluids . of different density
`without changing calibration.
`In accordance with one
`aspect of the invention, it is contemplated to oscillate the
`loop at a frequency which is low compared to the natural
`resonant frequency of the loop.
`.
`.
`The curves of Fig. 4 will be .helpful in understanding
`this condition of operation. Fig. 4 shows three sets.of
`curves 32, 32', 33, 33' and 34, 34' for fluids of different
`density. Frequency of oscillation is plotted along the
`horizontal axis and .peak angular. displacement 11m. about 35
`axis o.f 14 is plotted for the vertical axis. The conditions
`plotted are for equal rates of mass flow.
`Analysis of a system such as that shown· in Fig. l indi•
`cates that the displacement II can be represented .approxi(cid:173)
`mately by the following equation:
`
`small compared to unity, Equation 1 indicates that the
`maxirimm angular displacement will be proportional to
`the mass flow.
`The selection of a particular operating frequency will,
`of course, depend upon the design parameters of the ap(cid:173)
`paratus and the range of fluid densities over which it is
`desired to employ the instrument. It will .be understood
`40 that a portion of the moment of inertia of the loop struc(cid:173)
`ture will be that due to the material of which the loop
`is constructed, and the remainder will be contributed by
`the mass of the fluid therein. Thus, only a fraction of
`the total moment of inertia will be subject to change by
`45 variations in fluid density. The torque per se about the
`axis of member 14 will be unaffected by the moment of
`inertia of the loop structure, since it is a function of rate
`of mass flow through the loop. However, the displace(cid:173)
`merit of the loop abotlt the axis of member 14, and the
`50 resulting angular velocity and acceleration, will be af(cid:173)
`fected by the moment of inertia of the loop structure.
`Thus with a pickup transducer sensitive to . one of these
`quantities, the greater the fixed component of inertia, the
`less the output of the transducer will be affected by
`55 changes in the fluid density; Of course, it is not desirable
`to make the loop structure too massive, since the sensi(cid:173)
`tivity of the instrument \Vill be reduced or more amplifi(cid:173)
`cation required, more driving poWer will be required, a ad
`greater acceleration forces will be encountered.
`
`6° For a given field of use; the variation in density of
`
`(1)
`
`where ll=angular displacement in radians of loop 10
`about axis of 14.
`'
`g=acceleratiori o.f gravity
`W=pounds of material flowing across any cross-section
`!=time in seconds
`.
`R=radius of the loop in feet
`cp=maximuril angular displacement of loop 10 about axis
`of 17 due to the constant velocity driving source in
`radians
`w=angular velocity of the constant velocity driving source
`in radians/se.c.
`.
`. .
`.
`kx=spring constant. of t~e constraint about the· torque
`axis 14 in lbAt./radhirt.
`_
`.. . . . .
`wx=natural resonant an,gular· velocitY .·· of the loop. 10
`about the torque axis 14.
`The maximum or peak angular disphicement;: lim, is

`, -
`.
`given by Equation I when ei•t= I.
`This analysis indicates, · as will be noted from Equatwn
`l, that the maximum angular displacement .of the loop
`10, when driven by a constant velocity source such· as
`·motor 21, varies . linearly with · the driving angular ve(cid:173)
`locity w regardless of .the value of density at· low driving.
`frequencies where the quantity · ·
`
`w2 ·
`
`65
`
`liquids likely to be encountered is not so great as to pre(cid:173)
`clude the selection of a proper driving frequency which
`will give accurate indications and adequately high output
`signal.

`It ·has been mentioned that damping inherent in any
`. practical syste111 will affect the shape of the curves of Fig.
`4. Where desired, damping can be introduced inten(cid:173)
`tionally so as to increase the frequency range of region
`70 35. Such damping can be i'ntroduced at 16, 16' by in(cid:173)
`serting viscous material, by employing dash pots between
`loop 10 and the vertical member i7, by employing elec(cid:173)
`trical damping. such as eddy current damping, etc. These
`and many other forms of damping are well known in the
`art.
`
`'
`
`.
`
`wx_2
`
`'
`
`...
`is small compared to unity. This Is the region 35-sliown
`jn Fig." 4 .
`. Frequency, father .than angular velocity, is 75
`
`~-
`
`MM0634694
`
`

`
`Case 6:12-cv-00799-JRG Document 124-5 Filed 03/07/14 Page 9 of 18 PageID #: 3771
`
`V=kR'do
`dt
`
`30
`
`8
`wave corresponding to the peak values of the signal in coil
`Illustrative of the foregoing, Figs .. la and lb show eddy
`27. If the rate of mass flow is constant, a constant D.-C.
`current and viscous damping, respectively. In Fig. la an
`electrically conductive metal plate 96 is secured to loop
`value wilr"be obtained at 45, and as the rate of mass flow
`varies the voltage at point 45 will likewise vary.
`10 and arranged to oscillate in the field of a magnet 97
`Tile detector output at point 45 is supplied to a thermi-
`which may be of the permanent magnet type or energized 5
`onic vacuum tube 46 connected as a cathode fol'owcr,
`by a suitable coil. Magnet 97 is stationary with respect
`to member 14 and is attached thereto by support mem-
`and the voltage across cathode resistor 47 is supplied to
`one terminal of a micro-ammeter 48. Ammeter 48 may
`ber 98. As the conductive plate 96 moves in the field
`be shunted by a variable resistance 49 for calibration
`of magnet 97, eddy currents are created in plate 96 and
`damp the movement of the loop 10. Advantageously the 10 purposes. In order to make the ind:cator insensitive to
`plate 96 is mounted on loop 10 diametrically opposite the
`line voltage variations and also to provide for setting the
`transducer coil 27 (Fig. I) so as to lie in a plane per-
`zero of meter 48, another tube 51 is provided wh:ch
`pendicular to the plane of the loop and passing through
`has its anode energized from the same B+ sour:;! as tube
`the axis of vertical member 17, as shown.
`46, and its grid supplied with a constant positive bias
`In. Fig. lb a plate 99 is attached to the loop 10 in a 15 from the same B+ source thr-ough the voltage divider
`resistors 52, 53. The cathode circuit of tube 51 includes
`manner similar to plate 96 of Fig. Ia. In this figure, the
`plate 99 is arranged to oscillate in a container Hll . filled
`resistors 54, 54' and potentiometer 55 whose total re-
`with a viscous liquid, such as a viscous oil, and the con-
`sistance is advantageously approximately equal to that
`tainer is attached to member 14 similar to magnet 97 in
`of resistor 47. The other side of meter 48 is then con-
`Fig. la. As plate 99 moves back and forth, it will shear 20 nected to the cathode of tube 51. The lower terminal of
`the viscous liquid and hence dissipate energy.
`x:ectifier 43 is returned to a variable tap on potentiom-
`While variations in the displacement Om with mass flow
`eter 55, as shown, so that an adjustable D.-C. bias can be
`may be measured by a suitable transducer and used to
`applied to the rectifier.
`indicate mass flow, as above pointed out it is advantageous
`Before making a measurement of mass flow, the zero
`to employ a velocity-type transducer. The output of such 25 of meter 48 can be set by adjusting the arm of potenti-
`cmeter 55. This impresses a positive bias on the grid
`a transducer will be proportional to the time rate of
`change of B rather than B directly.
`on tube 46 through rectifier 43 and the re,istance of
`Thus:
`filter 44. The bias is so adjusted that the difference in
`potential between the cathodes of tubes 46 and 51 re(cid:173)
`sults in sufficient current through meter 48 to bring the
`pointer to the zero setting. Thereafter the pcsit:on of
`the pointer will vary with the rate of mass flow. Meter
`38 may be calibrated in arbitrary units or directly in
`terms of rate of ma~s flow for a given instrument.
`As before pointed out, it is advantageous in a m1ss
`flowmeter such as shown in Fig. 1 to employ a trans(cid:173)
`ducer of the velocity type which gives voltage peaks at
`the neutral axis which are proportional to rate of mass
`flow. Thus, the effect Gf possible non-linearities in the
`40 osci!lation of the fluid conduit loop are relatively unim(cid:173)
`portant. Since the indicator of Fig. 5 employs peak
`deteGtion, the meter 48 is sensitive cnly to variations
`in the peak amplitude of the applied wave and v.aria(cid:173)
`tions in the instantaneous voltage between peaks are un-
`,1J important. Thus, with relatively simple instrumentation
`an accurate indication of rate of mass flow is obta'nable.
`The circuit arrangement of Fig. 5 is given merely as
`an example of a suitable arrangement employing peak
`dete·ction which has been found satisfactory in practice.
`50 However, many other circuits are known in the art em(cid:173)
`ploying peak detection and any suitable fo rm can be em(cid:173)
`ployed if · desired. Also, with adequate lin2arity in the
`-oscillation of the fluid conduit loop, or w:th transducers
`other than those of the velocity type, other forms of de-
`55 teeters can, of course, be employed.
`If the integrated mass flow is desired, rather than the
`instantaneous rate of mass flow, a suitable form of inte(cid:173)
`grating indicator 'can be employed in place of meter 48.
`For example, a watt-hour meter with one coil connected
`00 in place of meter 48 and the other coil energized from
`a constant voltage source could be employed.
`For less stringent applications the displacement of the
`loop may be indicated directly, as by means of a pointer,
`rather than employing apickup transducer and associated
`tlJ circuitry such as described above.
`Fig. 6 illustrates such an arrangement, together with a
`loop and mounting arrangement which,. although not pos(cid:173)
`sessing all the advantages of that shown in Fig. 1, may
`nevertheless be employed in some applications. Here the
`iO loop 10 is mounted for rotation about a diameter thereof
`by bearings 56, 56' · carried ·by a U -shaped frame 57.
`Frame 57 is mounted for rotation about an axis perpen(cid:173)
`dicular to that of bearings 56, 56' by shaft 58 rotating
`in bearing housing 58'. The loop may be oscillated about
`75 the axis of 58 in the saine manner as in Fig. 1.
`
`where
`V=voltage output of the transducer
`k=a constant dependent on design of the transducer
`R'=~distance of the pickup from the axis 14 of rotation 35
`Curves similar to Fig. 4 may be drawn which indicate
`· the variation in output with velocity-type pickups by clif(cid:173)
`ferentiating Equation 1 and plotting the results, exclusive
`of the factor etwt. Although the shape of the curves will
`differ from those shown in Fig. 4, the conclusion as to the
`oper1lting region described above will be evident there(cid:173)
`from.
`The amplitude of oscillation in an apparatus like that
`of Fig. 1 can be made quite small. For example in one
`specific construction of an instrument designed to measure
`relatively low rates of mass flow, up to 10 pounds per
`minute, the amplitude of oscillation about the vertical axis
`was ±0.5 degrees. For a full-scale mass flow of 10
`lbs./min. the maximum displacement of the loop about
`the horizontal axis was approximately ±0.005 degree.
`The loop radius was 3.5", the operating frequency was 10
`cycles/second and the natural resonant frequency of the
`loop was 100 cycles/second. A relatively simple velocity
`pickup gave sufficient output for convenient amplification
`and indication.by a circuit similar to that shown in Fig. 5.
`Referring now to Fig. 5, a circuit is shown for receiving
`the output of apparatus such as shown in Fig. 1, and giv(cid:173)
`ing a direct indicatio-n of rate of mass flow on a suitably
`calibrated meter. In Fig. 5. coil 27 is that shown in Figs.
`1 and 2 and supplies a voltage proportional to rate of
`mass flow to the circuit. A rejection filter 41 is advan(cid:173)
`tageously employed to reject any 60-eycle power line fre- ·
`quencies. As here shov;n, it is of the so-called "twin T"
`type, but of course any suitable form of filter can be em(cid:173)
`ployed. The voltage variations from input coil 27 are
`amplified in stages including tubes 42. Any suitable low
`frequency amplifier d

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket