`Recent advances in dental optics – Part I: 3D intraoral scanners
`for restorative dentistry
`Silvia Logozzo a,n, Elisabetta M. Zanetti a, Giordano Franceschini a, Ari Kilpelä b,
`Anssi Mäkynen b
`a Department of Industrial Engineering, University of Perugia, Via Duranti, 93 06125 Perugia, Italy
`b Department of Electrical Engineering, University of Oulu, P.O.BOX 8000, FI-90014 Oulu, Finland
`
`a r t i c l e i n f o
`
`a b s t r a c t
`
`Available online 17 September 2013
`
`Keywords:
`Intraoral scanner
`Digital dental impressions
`Triangulation
`Confocal microscopy
`Optical coherence tomography
`
`Intra-oral scanning technology is a very fast-growing field in dentistry since it responds to the need of an
`accurate three-dimensional mapping of the mouth, as required in a large number of procedures such
`as restorative dentistry and orthodontics. Nowadays, more than 10 intra-oral scanning devices for
`restorative dentistry have been developed all over the world even if only some of those devices are
`currently available on the market. All the existing intraoral scanners try to face with problems and
`disadvantages of traditional impression fabrication process and are based on different non-contact
`optical technologies and principles. The aim of this publication is to provide an extensive review of
`existing intraoral scanners for restorative dentistry evaluating their working principles, features and
`performances.
`
`& 2013 Elsevier Ltd. All rights reserved.
`
`1. Background
`
`Three-dimensional scanning of the mouth is required in a large
`number of procedures in dentistry such as restorative dentistry
`and orthodontics. The aim of the 3D mapping of the oral cavity is
`to create digital impressions.
`Restorative dentistry is of course the main field that require the
`application of very accurate 3D intraoral scanners. For the realiza-
`tion of any dental prosthesis it is necessary to realize three-
`dimensional mathematical models of the dentition, performing a
`reverse engineering procedure. Then the prosthesis can be realized
`by means of CAD/CAM systems.
`At present, according to the traditional work flow, this procedure
`starts at the dentist's office, and the steps leading to prosthesis's
`creation are as follows:
`
`Abbreviations: AFI, accordion fringe interferometry; AWS, active wave-front
`sampling; CAD/CAM, computer aided design/computer aided manufacturing;
`CLSM or LSCM, confocal laser scanning microscopy; HIPAA, health insurance
`portability and accountability act; LASER, light amplification by stimulated
`emission of radiation; LED, light emitting diode; MEMS, micro electro-mechanical
`system; NA, numerical aperture; OCT, optical coherence tomography; OBJ, alias
`wavefront technologies file format; PLY, polygon file format or Stanford triangle
`format; PMT, photo-multiplier tube; SLA, stereo-lithography; S/N or SNR,
`signal-to-noise ratio; USB, universal serial bus
`n Corresponding author. Tel.: +39 348 7142939.
`E-mail addresses: sililog@hotmail.com (S. Logozzo),
`elisabetta.zanetti@unipg.it (E.M. Zanetti), giordano.franceschini@unipg.it
`(G. Franceschini), arik@ee.oulu.fi (A. Kilpelä), anssi.makynen@ee.oulu.fi
`(A. Mäkynen).
`
`0143-8166/$ - see front matter & 2013 Elsevier Ltd. All rights reserved.
`http://dx.doi.org/10.1016/j.optlaseng.2013.07.017
`
` the dentist captures the traditional impression by means of
`impression trays and impression materials;
` the dentist sends the impression tray to the dental laboratory;
` the laboratory's technician pours plaster inside the tray;
` after the hardening he scans the plaster model to have the 3D
`virtual digital model of the full arch;
` the technician can design the prosthesis by means of CAD/CAM
`systems and send the file to a milling machine;
` the milling machine produces the prosthesis;
` the prosthesis is applied by the dentist and refined inside the
`patient's mouth to verify and adjust the occlusion.
`
`Basically, the 3D digital model is used to design the prosthesis
`and as an input to the program of the milling machine referring to
`CAD/CAM systems. It can also be used to perform surgery simula-
`tions or to build plastic models of the teeth by means of rapid
`prototyping techniques.
`The whole traditional process is often slow and affected by
`errors. Furthermore, although the traditional impression taking
`process is very cheap, it is certainly bothering for the patient and,
`at the present state of the art, definitively obsolete.
`By means of devices here described, the dentist can scan the
`teeth in vivo and he can directly create the virtual 3D model of the
`dentition. This allows bypassing the dental laboratory for a lot
`of steps.
`According to the state of the art, there are three kinds of
`workflows in restorative dentistry. The traditional workflow has
`been described above; it is the oldest and is illustrated in Fig. 1.
`
`Align Ex. 1028
`U.S. Patent No. 9,962,244
`
`203
`
`

`

`204
`
`S. Logozzo et al. / Optics and Lasers in Engineering 54 (2014) 203–221
`
`Fig. 1. Traditional workflow for dental impressions.
`
`Fig. 2. Former digital workflow for dental impressions.
`
`Sometimes the plaster pouring can be skipped because the
`impression tray is directly scanned.
`The second kind of workflow is the ‘former digital workflow’.
`The term ‘former’ is used to distinguish this method from the
`newest one, that is mentioned here as ‘rapid digital workflow’. The
`‘former’ digital workflow can be followed by a clinician who owns
`a standalone intraoral scanner, which is not equipped with a
`milling unit. The former digital workflow is reported in Fig. 2.
`According to the former digital workflow, the steps for pros-
`thesis creation are as follows:
` the dentist captures the digital impression by means of an
`intraoral scanning device;
` the dentist sends the digital prescription to a laboratory;
` the lab downloads the digital file and uses customized software
`to digitally cut the die and mark the margins;
` the SLA model is generated by using CAD/CAM systems;
` the technician can proceed with his preferred finishing
`technique:
`○ hand layered porcelain,
`
`○ pressing with wax patterns,
`○ digitally designed and milled full contour glass ceramic
`restoration by means of CAD/CAM systems (the technician
`must also design the program for the milling machine by
`means of CAM systems);
` the final restoration is then sent to the doctor for seating.
`
`The third kind of workflow is the rapid digital workflow. This
`workflow can be followed when the clinician owns an intraoral
`scanner equipped with an in-office milling unit. The rapid digital
`workflow is shown in Fig. 3.
`According to the rapid digital workflow, the steps for prosthesis
`creation are as follows:
` the dentist captures the digital impression by means of an
`intraoral scanning device;
` the dentist designs the restoration and the software automa-
`tically generates the program for the milling unit;
` the final restoration is milled in a few minutes;
` the doctor applies the restoration.
`
`

`

`S. Logozzo et al. / Optics and Lasers in Engineering 54 (2014) 203–221
`
`205
`
`Fig. 3. Rapid digital workflow for dental impressions.
`
`As a result of the application of 3D intraoral scanners, some
`of the disadvantages related to the traditional workflow can be
`overcome, as:
` Mould instability;
` Mould transport and packaging;
` Plaster pouring and solidification;
` Delamination;
` Lacerations on margins;
` Contact between the tray and the teeth;
` Geometrical and dimensional
`inconsistencies between the
`plaster model and the real teeth.
`
`the following
`
`3D Intraoral scanners can therefore entail
`advantages:
` Implementation of highly accurate models;
` Traditional workflow simplification;
` Possibility to create and periodically update a database of
`dentitions for future interventions;
` Possibility to simulate surgery interventions on the digital model;
`
`Overcoming all the prior disadvantages.
`
`2.
`
`Introduction
`
`The application of CAD/CAM methodologies to the dental field
`was the brainchild of Dr. Francois Duret in his thesis, presented at
`the Université Claude Bernard, Faculté d′Odontologie, in Lyon,
`France in 1973, and entitled ‘Empreinte Optique’ (Optical Impres-
`sion). In detail, he developed and patented a CAD/CAM device in
`1984. The developed system was presented at the Chicago Mid-
`winter Meeting in 1989, and was able to fabricate a dental crown
`in 4 h [2,3].
`Digital
`impressions have been introduced, and successfully
`used, for a number of years, in orthodontics, as well, including
`Cadent's IOC/OrthoCad, DENTSPLY/GAC's OrthoPlex, Stratos/Ora-
`metrix's SureSmile, and EMS’RapidForm, but the introduction of
`the first digital intraoral scanner for restorative dentistry was
`in the 1980s by a Swiss dentist, Dr. Werner Mörmann, and an
`Italian electrical engineer, Marco Brandestini, who developed the
`
`fundamentals for CERECs by Sirona Dental Systems LLC (Charlotte,
`NC), introduced in 1987, as the first commercially available CAD/
`CAM system for dental restorations [2,4]. Ever since, research and
`development by a lot of companies have improved the technolo-
`gies and created in-office intraoral scanners, which are increas-
`ingly user-friendly
`and produce precisely fitting dental
`restorations. These systems are capable of capturing three-
`dimensional virtual
`images of tooth preparations; restorations
`may be directly fabricated from such images (using CAD/CAM
`systems) or the same images can be used to create accurate master
`models, for the restorations in a dental laboratory [2].
`Nowadays, more than 10 intra-oral scanning devices for restora-
`tive dentistry exist all over the world. In this paper all these devices
`are mentioned and 11 are described and analysed. Existing devices
`are based on different non-contact optical technologies such as
`confocal microscopy, optical coherence tomography, active and
`passive stereovision and triangulation, interferometry and phase
`shift principles. Basically, all these devices combine more than one of
`the cited imaging techniques to minimize the noise arising when
`scanning inside an oral cavity as, for example: noise related to the
`optical features of the target surfaces (translucency and the different
`reflectivity of target materials such as teeth, gums, preparations,
`resins, etc.), to wetness and to random relative motions. Also several
`typologies of structured light sources and optical components are
`employed. The analysed intra-oral scanning devices for restorative
`dentistry are listed below:
`
`(1) CERECs – by Sirona Dental System GmbH (Germany)
`(2) iTero – by CADENT Ltd (Israel)
`(3) E4D – by D4D TECHNOLOGIES, Llc (USA)
`(4) Lava™C.O.S. – by 3M ESPE (USA)
`(5) IOS FastScan – by IOS TECHNOLOGIES, Inc. (USA)
`(6) MIA3d™ – by Densys3D Ltd (Israel)
`(7) DPI-3D – by DIMENSIONAL PHOTONICS INTERNATIONAL,
`Inc. (USA)
`(8) 3D Progress – by MHT S.p.A. (Italy) and MHT Optic Research
`AG (Switzerland)
`(9) directScan – by HINT – ELS GmbH (Germany)
`(10) trios – by 3SHAPE A/S (Denmark)
`(11) Bluescans-I – ATRON3Ds GmbH (Austria)
`Only some of these are already commercially available. As
`already mentioned, even if a lot of advantages in taking
`
`

`

`206
`
`S. Logozzo et al. / Optics and Lasers in Engineering 54 (2014) 203–221
`
`Fig. 4. Confocal microscopy principle [6].
`
`digital impressions are achievable, also some disadvantages
`subsist. For example, it is often necessary to apply some
`coatings on the teeth (to minimize the noise of the measure-
`ment) and to rest the camera wand on a tooth to get a steady
`focus. Moreover, the 3D virtual model is often reconstructed
`by post-processing single images (acquired from a single
`perspective); accordingly the reconstruction is not performed
`in real time with a continuous data capture. Furthermore,
`data concerning the accuracy of the available instruments are
`often missing [1].
`Some other new intra-oral scanners have been recently
`presented at International Dental Show 2013 in Cologne:
`(12) Planscan – Planmeca Oy (Finland)
`(13) Condor – Remedent Inc. (Belgium)
`(14) CS 3500 – Carestream Health, Inc. (USA)
`(15) DigImprint – Steinbichler Optotechnik GmbH (Germany)
`
`3. Confocal laser scanner microscopy and devices
`
`Confocal laser scanning microscopy (CLSM or LSCM) is a technique
`to acquire in-focus images from selected depths, a process known as
`optical sectioning (high-resolution optical images with depth selecti-
`vity) [5]. Images are acquired point-by-point and reconstructed by a
`computer. By using this technique, one can reconstruct the surface
`profile of opaque specimens and obtain the interior imaging of non-
`opaque specimens.
`A conventional microscope sees as far into the specimen as the
`light can penetrate, whereas a confocal microscope only images
`one depth level at a time.
`The CLSM achieves a controlled and highly limited depth
`of focus.
`The principle of confocal microscopy was originally patented by
`Marvin Minsky in 1961 [6], but it took another 30 years and the
`development of lasers for CLSM to become a standard technique,
`toward the end of the 1980s.
`In a CLSM technique a laser beam passes through an aperture
`(14 in Fig. 4) and then is focused by an objective lens (11 in Fig. 4)
`into a small focal volume, within or on the surface of a specimen;
`in biological applications the specimen may be fluorescent. Scat-
`tered and reflected laser light, as well as any fluorescent light from
`the illuminated spot, is then re-collected by the objective lens.
`A beam-splitter (17 in Fig. 4) separates off some portion of the
`light into the detection apparatus (28 in Fig. 4).
`This apparatus, in fluorescence confocal microscopy, has also a
`filter, which selectively passes the fluorescent wavelengths while
`blocking the original excitation wavelength. After passing a pin-
`hole (24 in Fig. 4), the light intensity is detected by a photo-
`detection device (usually a photomultiplier
`tube (PMT) or
`
`Fig. 5. iTero digital impression system [2].
`
`avalanche photodiode (APD)), transforming the light signal into
`an electrical one which is recorded by a computer [7].
`The limited detector aperture obstructs the light which is not
`coming from the focal point. The out-of-focus light is suppressed:
`most of the returning light is blocked by the pinhole, which results
`in sharper images than those from conventional fluorescence
`microscopy techniques and permits to obtain images of planes
`located at various depths within the sample (sets of such images
`are also known as ‘z stacks’) [5].
`The detected light, originating from an illuminated volume
`element within the specimen, represents one pixel in the resulting
`image. The brightness of a resulting image pixel corresponds to the
`relative intensity of the respective detected light. The beam is
`scanned across the sample in the horizontal plane by using one or
`more servo-controlled oscillating mirrors.
`Slower scans provide a better signal-to-noise ratio, resulting in
`better contrast and higher resolution. Information can be collected
`from different focal planes by raising or lowering the microscope
`stage or objective lens. The computer can generate a three-
`dimensional picture of a specimen by assembling a stack of these
`two-dimensional images, from successive focal planes [5].
`
`3.1.
`
`iTero by CADENT LTD (IL)
`
`The Cadent iTero digital impression system by Cadent LTD, IL
`(Fig. 5) came into the market in early 2007. iTero system employs a
`parallel confocal imaging technique (Fig. 6) [8]. As shown in Fig. 7,
`an array of incident red laser light beams (36), passing through a
`focusing optics (42) and a probing face, is projected onto the teeth.
`
`

`

`S. Logozzo et al. / Optics and Lasers in Engineering 54 (2014) 203–221
`
`207
`
`Fig. 6. iTero's wand [2,9].
`
`Fig. 7. iTero scanning system [8].
`
`This technique allows iTero capturing all structures and mate-
`rials in the mouth, without the need to apply any coating to the
`patient's teeth [9]. The complete three-dimensional representation
`of the entire structure can be obtained assembling surface topol-
`ogies of adjacent portions, taken at two or more different angular
`perspectives [10]. The iTero camera capability of scanning without
`the need of coating powders is advantageous, however it requires
`the addition of a colour wheel to the acquisition unit itself (Fig. 8),
`resulting in a camera with a larger scanner head, compared to the
`other systems [2]. In fact, also a two-dimensional (2D) colour
`image of the 3D structure of teeth is taken at the same angle and
`orientation with respect to the structure. As a consequence, each
`X–Y point on the 2D coloured image corresponds to one point on
`the 3D scan, having the same relative X–Y values. The coloured
`image (Fig. 8) is obtained illuminating the target surface with
`three beams having three complementary colours (red, green or
`blue light), and combining the respective monochromatic images
`to create a full colour image. The three beams are obtained from
`the same white light source, with colour filters. The filters are
`arranged as sectors of a rotatable disc, coupled to a motor [11].
`Capturing the digital impression requires following a series of
`steps for every impression which the operator is guided through.
`These include five scans of the prepared area: occlusal, lingual,
`buccal, and interproximal contacts of the adjacent teeth [2], and
`require approximately 15 or 20 s per prepared tooth. Then buccal
`and lingual 451 angle views of the remaining teeth in the quadrant
`or arch and opposing arch are obtained. When these scans (at least
`
`Fig. 8. iTero colour imaging system [11].
`
`The focusing optics defines one or more focal planes beyond the
`probing face, in a position which can be tuned by a motor (72).
`The beams generate illuminated spots on the structure and the
`intensity of returning light rays is measured for various positions
`of the focal plane. The topology of the three dimensional structure
`of the teeth is reconstructed on the basis of spot-specific positions
`yielding a maximum intensity of the reflected light beams (Fig. 7)
`[8,10].
`
`

`

`208
`
`S. Logozzo et al. / Optics and Lasers in Engineering 54 (2014) 203–221
`
`21) are complete, the patient is asked to close into centric
`occlusion and a virtual registration is scanned. Overall, complete
`upper and lower quadrant scans and the virtual bite registration
`can take less than 3 min time [2]. When the digital impression has
`been completed, the clinician can have a series of diagnostic tools
`to evaluate the preparation and to complete the impression itself.
`For example, a margin line tool
`is available to assist in the
`identification of clearly defined margin [12]. The completed digital
`impression is sent via a HIPAA-compliant wireless system to the
`Cadent facility and to the dental laboratory. Upon ratification of
`the laboratory, the digital file is output to a model by Cadent.
`Finally, the model is milled from a proprietary blended resin [2].
`
`3.2. 3D Progress by MHT S.p.A. (IT) & MHT optic research AG (CH)
`
`3D Progress, produced by MHT (Medical High Technologies) S.p.A
`(IT) and created by MHT Optic Research AG (CH), is a light-weight,
`portable, digital impression system which can be interfaced to a PC via
`USB 2.0 cable (Fig. 9). MHT Optic Research AG and MHT S.p.A were
`founded in 1995 by Markus Berner, a Swiss engineer, and Carlo
`Gobbetti, an Italian businessman and entrepreneur.
`Besides being available for purchasing, in North America it will
`be also made available for a low monthly rental fee and commer-
`cialized by Clōn 3D Employee, as ‘Progress IODIS (an acronym
`of
`Intra Oral Digital
`Impression System)’. Another authorized
`distributor is Oratio BV company, from Netherlands, which will
`commercialize this device as ‘CYRTINAs Intraoral Scanner’. 3D
`Progress performs the digital impression, taking less than 1/10th
`of a second for a single scan, with an average scanning speed equal
`to 14 scan/second (depending on the PC); therefore, it can scan a
`
`Fig. 9. 3D Progress portable system [13].
`
`full-arch in under 3 min. The scanner will not usually require
`powdering of the translucent surfaces, with the exception of
`highly reflective surfaces as, for example, implant scan abutments
`and markers. Scans are output first as a cloud of points and then,
`as a final output, as a usual STL format surface file, compatible
`with most CAD platforms. The main technical features of the 3D
`Progress components are as follows: a smart Pixel Sensor (which
`enables fast and accurate scanning), an automatic real time
`stitching of each single scan, the possibility to pause/stop the scan
`whenever required, automatic (or semi-automatic) margin line
`detection, a USB 2.0 PC connection. 3D Progress works as a
`confocal microscope combined with Moireé effect detector
`[14,1]. The focal plane is shifted translating a movable lens, located
`as far distal as possible to maximize the miniaturization of the
`optical system. Unlike the prior optical system, based on confocal
`microscope shown in Fig. 10, in 3D Progress (Fig. 11) it is the first
`lens (4), distal from the object, to be moved in three different
`positions (each identified as 4a, 4b and 4c) in order to shift the
`focal plane (7) on the object (6), to positions identified as (7a), (7b)
`and (7c), respectively. The light rays, generated by the illumination
`pattern (1) and reflected at each focal plane (7a), (7b) and (7c),
`pass through the lens assembly (4), (5) and the beam guidance
`means (8) and are deflected by the beam splitter (2), in the
`direction of the detector (3), where the image of the object 6 is
`detected in the focal plane 7.The movable lens is aspherical in
`order to guarantee the necessary imaging quality for all focal
`planes (7a), (7b) and (7c), thus the focal plane is not actually a
`plane, but a curved surface and the scanned surfaces appear
`distorted: flat surfaces and straight lines appear curved, and the
`magnification and curvatures are different for different positions
`in the image. These distortions can be compensated because the
`theoretical distortions are known, having been computed on a
`reference image. The curvatures can be well approximated by an
`analytical function, such as for example, a polynomial [14].
`
`3.3. TRIOS™ by 3Shape A/S (DK)
`
`In December 2010, 3Shape announced the launch of a new
`patient-friendly and high-performance intraoral scanning solu-
`tion, named TRIOS™ (Fig. 12).
`The TRIOS™ system works according to the principle of con-
`focal microscopy, with a fast scanning time. The light source
`provides an illumination pattern producing a light oscillation on
`the object. This variation/oscillation may be spatial and/or it may
`be time varying. The system produces a variation of the focus
`plane of the pattern as well, over a range of focus plane positions,
`while maintaining a fixed spatial relation of the scanner to the
`
`Fig. 10. Prior optical system based on confocal microscope [14].
`
`

`

`S. Logozzo et al. / Optics and Lasers in Engineering 54 (2014) 203–221
`
`209
`
`Fig. 11. Optics for a confocal microscope like 3D Progress [14].
`
`Fig. 12. Trios’ wand and cart [15].
`
`When the focus plane coincides with the scanned surface at
`certain pixel position, the pattern is projected onto the surface
`point in-focus and it has high contrast, thereby giving rise to a
`large amplitude variation of the pixel values over time. It is thus
`possible to identify specific settings of the focusing plane, which
`make each pixel to be in focus. In other words, it is possible to
`transform the contrast information vs. position of the focus plane
`into a 3D surface information, pixel by pixel. The third dimension
`of the scanned object is determined by finding the plane corre-
`sponding to a peak in the correlation measured for each sensor
`belonging to the camera's sensor array.
`A peculiarity of this system is the variation of the focal plane
`position without moving the scanner in relation to the object
`being scanned. The focal plane should be continuously varied in a
`periodic fashion with a predefined frequency, while the pattern
`generation means, the camera, the optical system and the object
`being scanned fixed in relation to each other. Further, the 3D
`surface acquisition time should be small enough to reduce the bias
`produced by involuntary relative movements between the probe
`and the teeth [16]. The scanning system has the property of
`telecentricity in the space of the object being scanned and this
`property as well as magnification scale can be maintained while
`shifting the focal plane.
`
`Fig. 13. Trios scanning system [16].
`
`object (Fig. 13). When a time varying pattern is applied, a single
`sub-scan is actually the collection of a certain number of 2D
`images, corresponding to different positions of the focus plane and
`to their respective different time instances of the illumination
`pattern.
`
`

`

`210
`
`S. Logozzo et al. / Optics and Lasers in Engineering 54 (2014) 203–221
`
`4. Triangulation techniques and devices
`
`Triangulation is a non-contact technique for digitally collecting
`data of the shape of a 3D object and constructing digital 3D
`models, for a wide variety of applications. Both passive and active
`triangulation techniques may be used.
`In active triangulation
`methods, a light radiation is projected onto the scene, and its
`reflection is acquired in order to calculate the position of the target
`object. In passive triangulation methods no kind of radiation is
`emitted by the scanning device itself and the system is based on
`detecting reflected ambient radiation.
`Passive triangulation is also called passive stereovision and uses
`photogrammetric algorithms. This technique is based on processing
`of two stereo images, obtained from two cameras, whose respective
`positions and angulations are known (Fig. 15). This information is
`needed in order to identify points with corresponding features on
`the two images and to apply triangulation, with respect to the same
`
`corresponding points on the epipolar line; the algorithms are based
`on the epipolar geometry (Fig. 14).
`Passive triangulation provides the highest accuracy among
`vision systems of this type. However, only high contrast targets
`and well defined edges can be measured with high accuracy. Using
`three cameras the ambiguity can be reduced.
`Untargeted, or featureless, surfaces may not be measured at all.
`In addition, the ambient light affects significantly the ability of the
`system to successfully extract all desired features, unless con-
`trolled lighting is used [17]. The main advantage is the low cost of
`these systems, made of few and cheap components; furthermore,
`also the working principle is simple, the same as in the human eye.
`In active triangulation, a light beam generated by a laser is
`deflected by a mirror and scanned on the target object. Fig. 16 shows
`a block diagram of a 2D active triangulation system. A camera,
`composed of a lens and a position sensitive photo-detector, mea-
`sures the location of the image of the illuminated point on the
`object. The laser dot appears at different places in the camera's field
`of view, depending on how far away the laser strikes the surface,
`(Fig. 17). This technique is called triangulation because the laser dot,
`the camera and the laser emitter form a triangle. The distance d
`between the camera and the laser emitter is known, it is called
`baseline distance and it corresponds with one side of the triangle.
`The angle θ of the laser emitter corner is also known. The angle Φ of
`the camera corner can be determined by looking at the location of
`the laser dot in the camera's field of view. These three data fully
`determine the shape and size of the triangle and gives the location
`(X, Y, Z coordinates) of the laser dot corner of the triangle, by simple
`trigonometric calculations [17]. In most cases, a laser stripe, a grid or
`
`Fig. 14. Passive triangulation method.
`
`Fig. 16. 2D active triangulation method [17].
`
`Fig. 15. Stereo images and cameras.
`
`

`

`S. Logozzo et al. / Optics and Lasers in Engineering 54 (2014) 203–221
`
`211
`
`Fig. 17. Camera's field of view in active triangulation method.
`
`a series of patterns, instead of one single laser dot, are swept across
`the object to speed up the acquisition process.
`The measuring accuracy of this method depends on several
`parameters. The accuracy error in the estimation of the distance
`(Z) is inversely proportional to both the distance between the laser
`and the position detector (baseline d) and the distance between
`the effective position of the lens and the position detector (f0).
`Unfortunately, neither f0 nor d can be made large. Distance d is
`limited by the mechanical structure of the optical setup and by the
`arising of shadow effects [18]. The accuracy also improves increas-
`ing the number of pixels (photosensitive elements) of the camera
`and for shorter measuring distances [19]: the error of accuracy in
`the estimation of the distance Z is directly proportional to the
`square of the distance itself [18]. The total error of accuracy can
`be roughly estimated to be proportional to the working volume
`(a typical value is 1000:1).
`A major problem affecting all triangulation methods is occlu-
`sion; it takes place whenever an area of the target surface is
`invisible to both, or either of, the laser (laser occlusion) and the
`camera (camera occlusion). A theoretical solution is keeping the
`triangulation angle as small as possible, but the result would lose
`in terms of accuracy. So a good balance between minimal occlu-
`sion and good accuracy must be reached. The measure accuracy is
`also affected by the surface reflectivity of measured objects [19].
`
`4.1. CERECs by Sirona dental system GMBH (DE)
`
`CERECs (an acronym for Chairside Economical Restoration of
`Esthetic Ceramics) was introduced by Sirona Dental System GMBH
`(DE) in 1987 and it has undergone a series of technological improve-
`ments, culminating in the CEREC ACs, powered by BlueCams,
`launched in January 2009.
`The latest versions of the CERECs system (Fig. 18) are capable
`of producing inlays, onlays, crowns, laminate veneers, and even
`bridges and combine a 3D digital scanner with a milling unit, to
`create in-office dental restorations from commercially available blocks
`of ceramic or composite material, in one single appointment [2].
`The latest version of the milling centre, CEREC inLabs MC XL is
`capable of milling a crown in as short as 4 min. CERECs systems
`
`may be described as measurement devices which operate accord-
`ing to the basic principles of confocal microscopy [4,24], and
`according to the active triangulation technique [4,25] and [26]. A
`camera projects a changing pattern of blue light onto the object
`(Fig. 18), using projection grids, which have a transmittance
`random distribution, and which are formed by sub-regions, con-
`taining transparent and opaque structures [27].
`Moreover, it is possible, for each acquired profile, to define a
`specific relationships between the light characteristic and the optical
`distance of the image plane from the imaging optics [4,24], thanks to
`elements designed for varying the length of the optical path.
`A light source (3) (Fig. 19) produces an illumination beam (7.1,
`7.2, 7.3), that is focused onto the surface of the target object (2). An
`image sensor (6) receives the observation beam (9.1, 9.2, 9.3)
`reflected by the surface of the target object. A focusing system
`(5) focuses the observation beam onto the image sensor (6). The
`light source (3) is actually made of various units (3.1, 3.2, 3.3),
`which can be independently regulated in terms of light intensity
`[26]. Thus, the intensity of the light detected by each sensor
`element is a direct measure of the distance between the scan head
`and a corresponding point on the target object [4]. A critical aspect
`of the system is that the triangulation technique requires a
`uniform reflective surface, and different materials (as dentin,
`amalgam, resins, gums) reflect light differently. Therefore, it is
`necessary to coat the teeth with suitable powders, before the
`scanning stage, to make the reflectivity of the surfaces uniform.
`The earlier versions of CERECs employed an acquisition camera
`with an infrared laser light source. The latest version employs blue
`light-emitting diodes (LEDs) (Fig. 18); the shorter-wavelength,
`intense, blue light allows reaching a greater accuracy. The images
`are distortion-free, even at the periphery, so that multiple images
`(e.g. of a complete quadrant) can be stitched together with great
`accuracy. The CERECs AC Bluecam boasts an automatic shake
`detection system, which let the images to be acquired only if the
`camera is absolutely still. It is possible to capture a complete half
`arch in less than a minute. The new CERECs AC Bluecam offers
`image stabilization systems. It implies that the practitioner does
`not have to rest the camera wand on a tooth to get a steady fo

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket