throbber
~ Designation: F2004- 05 (Reapproved 2010)
`~u11 7
`
`INTERNATIONAL
`
`Standard Test Method for
`Transformation Temperature of Nickel-Titanium Alloys by
`Thermal Analysis 1
`
`Thi s standard is issued under the fixed designation F2004; the number immediately following the designation indicates the year of
`original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
`superscript epsilon (E) indicates an editorial change since the last revision or reapproval.
`
`1. Scope
`J. J This test method defines procedures for determining the
`transformation temperatures of nickel-titanium shape memory
`alloys.
`
`1.2 The values stated in SI units are to be regarded as
`standard. No other units of measurement are included in this
`standard.
`1.3 This standard does not purport to address all of the
`safety concems, if any, associated with its use. It is the
`responsibility of the user of this standard to establish appro(cid:173)
`priate safety and health practices and to determine the
`applicability of regulatoty limitations prior to use.
`
`2. Referenced Documents
`2.1 ASTM Standards: 2
`E473 Terminology Relating to Thermal Analysis and Rhe(cid:173)
`ology
`E967 Test Method for Temperature Calibration of Differen(cid:173)
`tial Scanning Calorimeters and Differential Thermal Ana(cid:173)
`lyzers
`E 1142 Terminology Relating to Thermophysical Properties
`F2005 Terminology for Nickel-Titanium Shape Memory
`Alloys
`
`3. Terminology
`3.1 Specific technical terms used in this test method are
`found in Terminologies E473 , E1142, and F2005.
`
`4. Summary of Test Method
`4.1 This test method involves heating and cooling a test
`specimen at a controlled rate in a controlled environment
`through the temperature interval of the phase transformation.
`
`1 T his test method is under the jurisdiction of ASTM Committee F04 on Medical
`and Surgical Materials and Devices and is the direct responsibility of Subcommiuee
`F04.15 on Material Test Methods.
`Current edition approved June I , 2010. Published September 2010. Originally
`approved in 2000. Last previous edition approved in 2005 as F2004- 05. DOl:
`I 0.1520/F2004-05R I 0.
`2 For referenced ASTM standards, visit the ASTM website, www.astrn.org, or
`contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
`Standards volume infonnation, refer Lo the standard's Document Summary page on
`the ASTM website.
`
`The difference in heat flow between the test material and a
`reference material due to energy changes is continuously
`monitored and recorded. Absorption of energy due to a phase
`transformation in the specin1en results in an endothermic peak
`on heating. Release of energy due to a phase transformation in
`the specimen results in an exothermic peak on cooling.
`
`5. Significance and Use
`5.1 Differential scanning calorimetry provides a rapid
`method for determining the transformation temperature(s) of
`nickel-titanium shape memory alloys.
`
`5.2 This test method uses small, stress-free, annealed
`samples to determine whether a sample of nickel-titanium alloy
`containing nominally 54.5 to 56.5 % nickel by weight is
`austenitic or martensitic at a particular temperature. Since
`chemical analysis of these alloys does not have sufficient
`precision to determine the transformation temperature by
`measuring the nickel to titanium ratio of the alloy, direct
`measurement of the transformation temperature of an annealed
`sample of known thermal history is recommended.
`5.3 This test method is useful for quality control, specifica(cid:173)
`tion acceptance, and research.
`5.4 Transformation temperatures derived from differential
`scanning calorimetry (DSC) may not agree with those obtained
`by other test methods due to the effects of strain and load on the
`transformation.
`
`6. Interferences
`6.1 Make sure the material to be tested is homogeneous
`since milligram sample quantities are used.
`
`6.2 Take care in preparing the sample. Cutting and grinding
`can cause cold work, which affects the transformation tempera(cid:173)
`ture. Oxidation during heat treatment can change the thermal
`conductance of the sample.
`
`6.3 Set the gas flow to provide adequate themml conductiv(cid:173)
`ity in the test celL
`
`7. Apparatus
`7.1 Use a differential scanning calorimeter capable of heat(cid:173)
`ing and cooling at rates up to 10°C/min and of automatically
`
`Copyright C ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428·2959. United States
`Copyright by ASTM Int'l (all rights reserved); Wed Aug 13 19:36:06 EDT 2014 I
`D ownloaded/printed by
`Bernard Cowger, Jr (Rothwell, Figg, Ernst Manbeck, P.C.) pursuant to License Agreement. No further reproductions authorized.
`
`GOLD STANDARD EXHIBIT 2018
`US ENDODONTICS v. GOLD STANDARD
`CASE PGR2015-00019
`
`

`
`F2004 − 05 (2010)
`
`recording the differential energy input between the specimen
`and the reference to the required sensitivity and precision.
`7.2 Use sample capsules or pans composed of aluminum or
`other inert material of high thermal conductivity.
`7.3 Use helium gas purge supply. See 10.3.1.
`7.4 Use an analytical balance with a capacity of 100 mg
`capable of weighing to the nearest 0.1 mg.
`
`8. Sampling
`8.1 Use a sample size of 25 to 45 mg. Cut the sample to
`maximize surface contact with the (DSC) sample pan.
`8.2 Anneal the sample at 800 to 850°C for 15 to 60 min in
`vacuum or inert atmosphere, or in air with adequate protection
`from oxidation. Rapidly cool the sample to prevent precipita-
`tion of phases which may change transformation temperature
`of the alloy.
`8.3 Clean the sample of all foreign materials such as cutting
`fluid. If the sample is oxidized in heat treatment, grind, polish,
`or etch the sample to remove the oxide. Take care to avoid cold
`working the sample as this will change its thermal response.
`Slight oxidation is permissible but remove all heavy oxide
`scale.
`
`9. Calibration
`9.1 Calibrate the temperature axis of the instrument using
`the same heating rate, purge gas, and flow rate as those used for
`analyzing the specimen in accordance with Practice E967.
`
`10. Procedure
`10.1 Place the sample on the sample pan and place the pan
`on the test pedestal.
`10.2 Place an empty pan on the reference pedestal.
`10.3 Turn on the purge gas at a flow rate of 10 to 50
`mL/min.
`10.3.1 Use helium as the purge gas for the sample chamber.
`10.3.2 Use a dry air, helium, or nitrogen cover gas. The dry
`gas shall have a dew point below the lowest temperature of the
`cooling cycle.
`
`10.4 Run the cooling and heating program.
`10.4.1 Use the heating and cooling rates of 10 6 0.5°C/min.
`10.4.2 Heat the sample from room temperature to a tem-
`perature of at least Af + 30°C; hold at temperature for a time
`sufficient to equilibrate the sample with the furnace.
`10.4.3 Cool the sample to a temperature of below Mf –
`30°C; hold for a time sufficient to equilibrate the sample with
`the furnace. Then, heat the sample to a temperature of at least
`Af + 30°C.
`10.5 Data Acquisition—Record the resulting curve from the
`heating and cooling program from Af + 30°C to Mf – 30°C.
`
`11. Graphical Data Reduction
`11.1 Draw the baselines for the cooling and heating portions
`of the curve as shown in Fig. 1.
`11.2 Draw the tangents to the cooling and heating spikes
`through the inflection points as shown in Fig. 1. If a computer
`program is used to construct the tangents, care must be taken in
`locating the tangent points.
`11.3 Obtain Ms, Mf, As, and Af as the graphical intersection
`of the baseline with the extension of the line of maximum
`inclination of the appropriate peak of the curve as shown in
`Fig. 1. Ap is the peak minimum of the endothermic curve, and
`Mp is the peak maximum of the exothermic curve. Read Ap and
`Mp directly from the graph as shown in Fig. 1.
`
`12. Report
`12.1 Report the following information with the test results:
`12.1.1 Complete identification and description of the mate-
`rial tested including the specification and lot number.
`12.1.2 Description of the instrument used for the test.
`12.1.3 Statement of mass, dimensions, and geometry.
`12.1.4 Material for the specimen pan and temperature pro-
`gram.
`12.1.5 Description of the temperature calibration procedure.
`12.1.6 Identification of the specimen environment by gas,
`flow rate, purity and composition.
`12.1.7 Results of the transformation measurements using
`the nomenclature in accordance with Terminology F2005.
`Temperature results should be reported to the nearest 1°C.
`
`13. Precision and Bias
`13.1 An interlaboratory study was conducted in accordance
`with Practice E691 in seven laboratories with three different
`materials, with each laboratory obtaining five results for each
`material. There were two rounds of testing. In the first round,
`all the test samples were annealed in one laboratory; in the
`second round, the samples were annealed by the laboratory that
`conducted the test. The details are given in ASTM Research
`Report No. F04–1008.3
`13.2 The results of round one are summarized in Tables 1-6
`for each transformation temperature parameter (Mf, Mp, Ms,
`As, Ap, Af). The values are in degrees Celsius. The terms
`
`FIG. 1 DSC Curve for Nickel-Titanium (NiTi)
`
`3 Supporting data have been filed at ASTM International Headquarters and may
`be obtained by requesting Research Report RR:F04-1008.
`
`2
`
`
`
`Copyright by ASTM Int'l (all rights reserved); Wed Aug 13 19:36:06 EDT 2014
`Downloaded/printed by
`Bernard Cowger, Jr (Rothwell, Figg, Ernst Manbeck, P.C.) pursuant to License Agreement. No further reproductions authorized.
`
`

`
`F2004 − 05 (2010)
`
`Material Mf,
`grand
`mean
`-50.0
`-26.3
`48.5
`
`A
`B
`C
`
`TABLE 1 Precision of Mf
`Repeatability
`Reproducibility
`Repeatability
`Limit
`Standard
`Standard
`Deviation
`Deviation
`0.57
`3.97
`0.95
`2.62
`1.02
`1.54
`
`1.6
`2.7
`3.0
`
`Reproducibility
`Limit
`
`11.1
`7.3
`4.3
`
`Material Mp,
`grand
`mean
`-43.8
`-20.5
`58.1
`
`A
`B
`C
`
`TABLE 2 Precision of Mp
`Repeatability
`Reproducibility
`Repeatability
`Limit
`Standard
`Standard
`Deviation
`Deviation
`0.42
`2.65
`1.10
`2.21
`0.88
`1.05
`
`1.2
`3.1
`2.5
`
`Reproducibility
`Limit
`
`7.4
`6.2
`2.9
`
`Material Ms,
`grand
`mean
`-41.6
`-16.9
`64.8
`
`A
`B
`C
`
`TABLE 3 Precision of Ms
`Repeatability
`Reproducibility
`Repeatability
`Limit
`Standard
`Standard
`Deviation
`Deviation
`0.40
`2.35
`0.95
`1.24
`0.74
`1.15
`
`1.1
`2.7
`2.1
`
`Reproducibility
`Limit
`
`6.6
`3.5
`3.2
`
`Material
`
`A
`B
`C
`
`Material
`
`A
`B
`C
`
`Material
`
`A
`B
`C
`
`As,
`grand
`mean
`-25.3
`-4.8
`72.9
`
`Ap,
`grand
`mean
`-19.4
`5.7
`94.7
`
`Af,
`grand
`mean
`-23.4
`2.5
`91.6
`
`TABLE 4 Precision of As
`Repeatability
`Reproducibility
`Repeatability
`Standard
`Standard
`Limit
`Deviation
`Deviation
`0.30
`1.85
`0.32
`1.58
`0.65
`2.68
`
`0.8
`0.9
`1.8
`
`Reproducibility
`Limit
`
`5.2
`4.4
`7.5
`
`TABLE 5 Precision of Ap
`Repeatability
`Reproducibility
`Repeatability
`Standard
`Standard
`Limit
`Deviation
`Deviation
`0.16
`1.94
`0.57
`1.50
`0.85
`3.18
`
`0.5
`1.6
`2.4
`
`Reproducibility
`Limit
`
`5.4
`4.2
`8.9
`
`TABLE 6 Precision of Af
`Repeatability
`Reproducibility
`Repeatability
`Standard
`Standard
`Limit
`Deviation
`Deviation
`0.23
`2.01
`0.67
`1.39
`0.80
`2.25
`
`0.6
`1.9
`2.2
`
`Reproducibility
`Limit
`
`5.6
`3.9
`6.3
`
`repeatability limit and reproducibility limit are used as speci-
`fied in Practice E177.
`13.3 The results of round two are summarized in Tables
`7-12 for each transformation temperature parameter (Mf, Mp,
`
`TABLE 7 Precision of Mf, When Samples are Annealed by Testing
`Laboratory
`Reproducibility
`Standard
`Deviation
`8.49
`5.93
`1.84
`
`Repeatability
`Limit
`
`Reproducibility
`Limit
`
`5.7
`5.2
`3.4
`
`23.8
`16.6
`5.1
`
`Material Mf,
`grand
`mean
`-49.1
`-27.3
`48.5
`
`A
`B
`C
`
`Repeatability
`Standard
`Deviation
`2.03
`1.85
`1.23
`
`TABLE 8 Precision of Mp, When Samples are Annealed by
`Testing Laboratory
`Repeatability
`Reproducibility
`Repeatability
`Limit
`Standard
`Standard
`Deviation
`Deviation
`1.31
`8.69
`1.67
`7.31
`1.31
`1.47
`
`Reproducibility
`Limit
`
`3.7
`4.7
`3.7
`
`24.3
`20.5
`4.1
`
`Material Mp,
`grand
`mean
`-43.0
`-21.5
`58.1
`
`A
`B
`C
`
`TABLE 9 Precision of Ms, When Samples are Annealed by
`Testing Laboratory
`Repeatability
`Reproducibility
`Repeatability
`Limit
`Standard
`Standard
`Deviation
`Deviation
`2.84
`8.29
`2.38
`6.60
`1.39
`2.29
`
`Reproducibility
`Limit
`
`7.9
`6.7
`3.9
`
`23.2
`18.5
`6.4
`
`Material Ms,
`grand
`mean
`-40.7
`-18.9
`64.8
`
`A
`B
`C
`
`TABLE 10 Precision of As, When Samples are Annealed by
`Testing Laboratory
`Repeatability
`Reproducibility
`Repeatability
`Standard
`Standard
`Limit
`Deviation
`Deviation
`0.94
`5.26
`0.52
`1.93
`0.70
`2.80
`
`Reproducibility
`Limit
`
`2.6
`1.4
`2.0
`
`14.7
`5.4
`7.8
`
`Material
`
`A
`B
`C
`
`As,
`grand
`mean
`-23.2
`-4.1
`72.9
`
`TABLE 11 Precision of Ap, When Samples are Annealed by
`Testing Laboratory
`Repeatability
`Reproducibility
`Repeatability
`Standard
`Standard
`Limit
`Deviation
`Deviation
`0.56
`5.30
`0.65
`2.71
`1.01
`3.37
`
`Reproducibility
`Limit
`
`1.6
`1.8
`2.8
`
`14.8
`7.6
`9.4
`
`Material
`
`A
`B
`C
`
`Ap,
`grand
`mean
`-19.1
`2.2
`89.5
`
`TABLE 12 Precision of Af, When Samples are Annealed by
`Testing Laboratory
`Repeatability
`Reproducibility
`Repeatability
`Standard
`Standard
`Limit
`Deviation
`Deviation
`0.87
`4.95
`0.69
`4.07
`0.91
`2.61
`
`Reproducibility
`Limit
`
`2.4
`1.9
`2.5
`
`13.9
`11.4
`7.3
`
`Material
`
`A
`B
`C
`
`Af,
`grand
`mean
`-16.6
`6.6
`94.7
`
`Ms, As, Ap, Af). The values are in degrees Celsius. The terms
`repeatability limit and reproducibility limit are used as speci-
`fied in Practice E177.
`
`3
`
`
`
`Copyright by ASTM Int'l (all rights reserved); Wed Aug 13 19:36:06 EDT 2014
`Downloaded/printed by
`Bernard Cowger, Jr (Rothwell, Figg, Ernst Manbeck, P.C.) pursuant to License Agreement. No further reproductions authorized.
`
`

`
`F2004 − 05 (2010)
`
`14. Keywords
`14.1 differential scanning calorimeter; DSC; nickel-titanium
`alloy; NiTi; Nitinol; shape memory alloy; TiNi; transformation
`temperature
`
`APPENDIX
`
`(Nonmandatory Information)
`
`X1. RATIONALE
`
`X1.1 This test method uses small, stress-free, annealed
`samples to determine whether a sample of nickel-titanium alloy
`containing nominally 54.5 to 56.5 % nickel by weight is
`austenitic or martensitic at a particular temperature. Since
`chemical analysis of these alloys does not have sufficient
`precision to determine the transformation temperature by
`measuring the nickel to titanium ratio of the alloy, direct
`measurement of the transformation temperature of an annealed
`sample of known thermal history is recommended.
`
`transformation temperature, possibly due to stress effects
`which retain residual martensite. One method of achieving the
`desired cooling rate is to heat treat the test specimens on a foil
`tray and then allow the samples and the foil tray to cool
`together, out of the furnace, in room temperature air.
`
`X1.3 Transformation temperatures derived from differential
`scanning calorimetry (DSC) may not agree with those obtained
`by other test methods due to the effects of strain and load on the
`transformation.
`
`X1.2 It is well known that slow cooling after annealing of
`nickel-rich alloys allows precipitates of the Ni4Ti3 type to
`form, thereby increasing the Ti content of the matrix and the
`transformation temperatures. The practice is to avoid slow
`cooling preserve the “as annealed” transformation.
`It
`is
`possible, however, to cool the samples too quickly, raising the
`
`X1.4 Differences in sample preparation techniques between
`laboratories influenced the reproducibility limit. Differences in
`calibration techniques may have also influenced reproducibil-
`ity. To minimize interlaboratory variations in results, common
`sample preparation and calibration practices must be estab-
`lished.
`
`ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
`in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
`of infringement of such rights, are entirely their own responsibility.
`
`This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
`if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
`and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
`responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
`make your views known to the ASTM Committee on Standards, at the address shown below.
`
`This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
`United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
`address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or
`through the ASTM website
`(www.astm.org). Permission rights to photocopy the standard may also be secured from the ASTM website (www.astm.org/
`COPYRIGHT/).
`
`4
`
`
`
`Copyright by ASTM Int'l (all rights reserved); Wed Aug 13 19:36:06 EDT 2014
`Downloaded/printed by
`Bernard Cowger, Jr (Rothwell, Figg, Ernst Manbeck, P.C.) pursuant to License Agreement. No further reproductions authorized.

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket