JOURNAL OF ENDODONTICS
`Copyright © 2002 by The American Association of Endodontists
`
`Printed in U.S.A.
`VOL. 28, NO. 8, AUGUST 2002
`
`SCIENTIFIC ARTICLES
`
`Differential Scanning Calorimetric Studies of Nickel
`Titanium Rotary Endodontic Instruments
`
`W. A. Brantley, T. A. Svec, M. Iijima, J. M. Powers, and T. H. Grentzer
`
`Differential scanning calorimetric (DSC) analyses
`were performed between ⴚ130° and 100°C on spec-
`imens prepared from nickel-titanium (NiTi) rotary
`endodontic instruments: ProFile (n ⴝ 5), Lightspeed
`(n ⴝ 4), and Quantec (n ⴝ 3). The ProFile and Light-
`speed instruments were in the as-received condition,
`whereas the Quantec instruments were randomly se-
`lected from a dental clinic and had unknown history.
`The DSC plots showed that the ProFile and Light-
`speed instruments analyzed had the superelastic
`NiTi property, with an austenite-finish (Af) tempera-
`ture of approximately 25°C. Differences in DSC plots
`for the ProFile instruments and the starting wire
`blanks (n ⴝ 2) were attributed to the manufacturing
`process. The phase transformation behavior when
`the specimens were heated and cooled between
`ⴚ130° and 100°C, the temperature ranges for the
`phase transformations, and the resulting enthalpy
`changes were similar to those previously reported
`for nickel-titanium orthodontic wires having super-
`elastic characteristics or shape memory behavior in
`the oral environment. The experiments demon-
`strated that DSC is a powerful tool for materials char-
`acterization of these rotary instruments, providing
`direct information not readily available from other
`analytical techniques about the NiTi phases present,
`which are fundamentally responsible for their clinical
`behavior.
`
`After the pioneering research by Walia et al. (1), which introduced
`nickel-titanium (NiTi) hand files to the endodontic profession, both
`nickel-titanium hand files and particularly rotary instruments have
`achieved widespread popularity. A major reason for their selection
`is the much greater flexibility (i.e. much lower elastic modulus) of
`the nickel-titanium alloy compared with stainless steel (1), which
`offers distinct clinical advantages with curved root canals. Numer-
`ous articles have recently reported the performance of the different
`
`567
`
`brands of nickel-titanium rotary instruments (2– 6) and hand in-
`struments (7–9).
`The manufacture of nickel-titanium instruments for endodontics
`has been discussed in a review article (10). The nickel-titanium
`alloys have the approximate composition of 55% nickel and 45%
`titanium and are based upon the intermetallic compound NiTi (11).
`Originally introduced for orthodontics (12), the nickel-titanium
`archwire alloys are available in nonsuperelastic, superelastic, and
`shape-memory forms (11, 13, 14). The nonsuperelastic nickel-
`titanium alloys have a predominantly heavily cold-worked, stable,
`martensitic NiTi structure, whereas the superelastic and shape
`memory alloys undergo a reversible transformation by twinning on
`the atomic scale (11) from the lower-temperature martensitic NiTi
`structure to the higher-temperature austenitic NiTi structure. The
`superelastic alloys undergo transformation from austenitic NiTi to
`martensitic NiTi with the application of stress. The shape memory
`alloys return to a higher-temperature shape established during
`processing (15), when the temperature is raised above the austen-
`ite-finish (Af) temperature, where the transformation to the auste-
`nitic NiTi structure is completed. The NiTi shape memory alloys
`used in orthodontics have the Af temperature below body temper-
`ature, whereas the superelastic (but not shape memory) orthodontic
`alloys have the Af temperature above that of the oral environment
`(16). The overall phase transformation process can be complex,
`and it has been found that an intermediate R-phase can form during
`the transformation between martensitic and austenitic NiTi (17).
`The structure of the NiTi alloys, which is central to their clinical
`performance, is conveniently studied by differential scanning cal-
`orimetry (DSC), in which the difference in thermal power supplied
`to a test specimen and an inert control specimen heated at the same
`rate is measured very accurately (11, 16). Structural transforma-
`tions in the NiTi alloys are revealed as endothermic peaks on the
`heating DSC curves and as exothermic peaks on the cooling DSC
`curves, and information is obtained about the temperature ranges
`and enthalpy changes for the phase transformations. For NiTi
`alloys, DSC indicates which of the three phases (martensitic NiTi,
`R-phase, or austenitic NiTi) will be present at a given temperature.
`Although X-ray diffraction analysis is a useful complementary
`method to investigate the structure of NiTi orthodontic wires (18),
`this technique only reveals the structure within approximately 50
`␮m of the surface, whereas DSC provides information for the
`overall bulk specimen (19). Electrical resistivity measurements
`
`PGR2015-00019 - Ex. 1038
`US ENDODONTICS, LLC., Petitioner
`
`

`

`568
`
`Brantley et al.
`
`Journal of Endodontics
`
`FIG 1. DSC plot for a ProFile specimen. The solid line is the heating curve, and the dashed line is the cooling curve.
`
`may also be used to study the structural transformations in the
`nickel-titanium alloys (11). The purpose of this study was to utilize
`DSC to investigate the phase relationships within three brands of
`popular nickel-titanium rotary endodontic instruments.
`
`MATERIALS AND METHODS
`
`Three brands of rotary nickel-titanium endodontic instruments
`(ProFile 0.04 taper, Dentsply Tulsa Dental, Tulsa, OK; Lightspeed,
`Lightspeed Technology, San Antonio, TX; Quantec, Analytic/
`Sybron Dental Specialties, Orange, CA) were selected for study,
`along with starting 1.0-mm diameter wire blanks for the ProFile
`instruments that were obtained from the manufacturer. Test spec-
`imens were carefully cut with a water-cooled, slow-speed diamond
`saw to minimize mechanical stresses that might change the pro-
`portions of the austenitic and martensitic NiTi phases from those in
`the as-received instruments and wire blanks. Each test specimen
`consisted of 2 to 4 segments, each approximately 4 to 5 mm in
`length, which were placed in an open aluminum pan; no crimped
`pan top was used to avoid mechanical stresses on the specimens.
`An empty aluminum pan served as the inert control specimen for
`the DSC measurements.
`Five specimens were analyzed for the 25-mm long, ISO size 25
`ProFile instruments. Four specimens were randomly selected from
`a package containing ISO sizes 40, 42.5, 45, and 47.5 for the
`25-mm long, Lightspeed instruments. Three specimens were ana-
`lyzed for the Quantec instruments, which were of varying sizes and
`randomly selected from the Endodontic Clinic at the University of
`Texas-Houston Dental Branch. Two replicate specimens of the
`starting wire blanks for the ProFile instruments were analyzed. The
`ProFile and Lightspeed instruments were in the as-received con-
`dition, whereas the Quantec instruments had an unknown clinical
`history.
`The DSC analyses were conducted (model 2910 DSC, TA
`Instruments, Wilmington, DE) over a temperature range from
`⫺130° to 100°C, using the liquid nitrogen cooling accessory (TA
`
`Instruments) to achieve subambient temperatures. For each anal-
`ysis, the specimen was first cooled from room temperature to
`⫺130°, then heated from ⫺130° to 100°C to obtain the heating
`DSC curve, and subsequently cooled from 100°C back to ⫺130°C
`to obtain the cooling DSC curve. The linear heating or cooling rate
`was a standard 10°C per minute (16), and during each analysis, the
`DSC cell was purged with dry nitrogen at a rate of 50 cm3/min.
`Temperature calibration of the DSC apparatus was performed with
`n-pentane, deionized water, and indium. The DSC plots were
`analyzed by computer software (TA Instruments) to obtain the
`peak temperatures for the phase transformations, along with the
`enthalpy changes associated with these processes. The interpreta-
`tions of the plots were based upon previous DSC studies of NiTi
`alloys in orthodontics (11, 16, 20).
`
`RESULTS
`
`Figures 1 and 2 show the two types of DSC plots that were
`obtained for the ProFile specimens. In all of the DSC plots, the
`heating curve is shown as a solid line at the bottom of the figure,
`and the cooling curve is shown as a dashed line at the top of the
`figure. The lines obtained by extrapolating the baselines adjacent
`to the peaks show the temperature ranges for the transformations.
`The other construction lines are used by the computer software to
`calculate values of enthalpy change (⌬H) and determine onset
`temperatures for the transformations.
`A pair of endothermic peaks can be seen on the heating curve in
`Figure 1, whereas an unresolved smaller endothermic peak exists
`on the left shoulder of the main peak at approximately 10°C. These
`peaks correspond to the initial transformation of martensitic NiTi
`to R-phase at lower temperatures, followed by transformation at
`higher temperatures of R-phase to austenitic NiTi, which is com-
`pleted at approximately 25°C. The area under both peaks repre-
`sents a total enthalpy change (⌬H) of approximately 3.3 J/g for the
`overall transformation from martensitic NiTi to austenitic NiTi. On
`the cooling curve, there is a broad exothermic peak at approxi-
`
`PGR2015-00019 - Ex. 1038
`US ENDODONTICS, LLC., Petitioner
`
`

`

`Vol. 28, No. 8, August 2002
`
`DSC Analysis of NiTi Files
`
`569
`
`FIG 2. DSC plot for a second ProFile specimen. The solid line is the heating curve, and the dashed line is the cooling curve.
`
`mately 5°C; and the overall enthalpy change of approximately 3.4
`J/g also includes a contribution from the unresolved smaller exo-
`thermic peak on the left shoulder, i.e. at lower temperatures than
`the main peak. Note that the vertical scale has been expanded for
`the heating curve, compared with the cooling curve. The main peak
`on the cooling curve corresponds to the initial transformation of
`austenitic NiTi to R-phase, which then transforms at lower tem-
`peratures to martensitic NiTi with a much lower enthalpy change
`than that for the transformation from austenitic NiTi to R-phase.
`In Figure 2, for another ProFile specimen, the smaller endother-
`mic peak at approximately ⫺18°C on the heating curve, corre-
`sponding to the initial transformation from martensitic NiTi to
`R-phase, is now resolved. The second endothermic peak at ap-
`proximately 1°C corresponds to the subsequent transformation
`from R-phase to austenitic NiTi. The enthalpy value of 4.4 J/g is
`applicable to the overall transformation from martensitic NiTi to
`austenitic NiTi when the sample is heated. For this specimen, the
`transformation to austenitic NiTi on heating (Af temperature) is
`again completed at approximately 25°C. The interpretation of the
`cooling curve follows that for Figure 1, with the main exothermic
`peak at approximately ⫺4°C, corresponding to the transformation
`from austenitic NiTi to R-phase. The transformation at lower
`temperatures from R-phase to martensitic NiTi again appears as an
`unresolved exothermic peak on the left shoulder of the main peak
`and has a lower enthalpy change than the transformation from
`austenitic NiTi to R-phase. The enthalpy change of 2.8 J/g corre-
`sponds to the overall transformation from austenitic NiTi to mar-
`tensitic NiTi when the sample is cooled. Note that both the heating
`and cooling curves have the same vertical scale in Figure 2.
`The other three ProFile specimens had DSC plots that were very
`similar to Figures 1 and 2. For the five specimens, the enthalpy
`change for the overall transformation from martensitic NiTi to
`austenitic NiTi on the heating curve ranged from approximately
`3.3 to 4.4 J/g, and the enthalpy change for overall transformation
`from austenitic NiTi to martensitic NiTi on the cooling curve
`ranged from approximately 2.8 to 3.6 J/g. The transformation to
`the austenitic NiTi structure on heating was largely completed for
`all five specimens at approximately 25°C.
`
`Figure 3 shows a DSC plot for a specimen prepared from the
`starting wire blanks used to manufacture the ProFile instruments;
`nearly the same results were obtained for the two specimens that
`were analyzed. The heating curve contains a clearly resolved first
`endothermic peak at approximately ⫺25°C for the transformation
`from martensitic NiTi to R-phase, followed by a second endother-
`mic peak at approximately 2°C for the transformation from R-
`phase to austenitic NiTi, which is completed at approximately
`25°C. The enthalpy change of 4.8 J/g corresponds to the overall
`transformation from martensitic NiTi to austenitic NiTi. The cool-
`ing curve contains a single resolved exothermic peak at approxi-
`mately ⫺6°C, corresponding to the transformation from austenitic
`NiTi to R-phase. As in Figures 1 and 2, the peak for the subsequent
`transformation from R-phase to martensitic NiTi is an unresolved
`left shoulder on the main exothermic peak and has a much lower
`enthalpy change than the initial transformation from austenitic
`NiTi to R-phase. The enthalpy change of 3.4 J/g corresponds to the
`overall transformation from austenitic NiTi to martensitic NiTi.
`(Note that the vertical scale for the heating curve has been sub-
`stantially expanded, compared with the scale for the cooling
`curve.)
`Figure 4 shows a DSC plot for a Lightspeed test specimen; very
`similar results were obtained for the other three samples. The
`heating curve contains two endothermic peaks at approximately
`⫺16°C and 2°C that correspond to the transformation from mar-
`tensitic NiTi to R-phase, followed by the transformation from
`R-phase to austenitic NiTi, respectively. The enthalpy change for
`the overall transformation from martensitic NiTi to austenitic NiTi
`is 8.2 J/g. As in the ProFile specimens in Figures 1 and 2, the
`cooling curve in Figure 4 contains a single exothermic peak at
`approximately ⫺2°C, but the extended left shoulder suggests the
`presence of an unresolved second exothermic peak with a smaller
`enthalpy change. Accordingly, the cooling curve is assumed to
`indicate the transformation of austenitic NiTi to R-phase, followed
`by the transformation at lower temperature from R-phase to mar-
`tensitic NiTi. The enthalpy change of 2.6 J/g arises from the overall
`transformation from austenitic NiTi to martensitic NiTi. For the
`four Lightspeed specimens, the overall enthalpy change on the
`
`PGR2015-00019 - Ex. 1038
`US ENDODONTICS, LLC., Petitioner
`
`

`

`570
`
`Brantley et al.
`
`Journal of Endodontics
`
`FIG 3. DSC plot for a specimen prepared from the NiTi wire blanks used to manufacturer the ProFile rotary instruments. The solid line is the
`heating curve, and the dashed line is the cooling curve.
`
`FIG 4. DSC plot for a Lightspeed specimen. The solid line is the heating curve, and the dashed line is the cooling curve.
`
`heating curves ranged from 7.4 to 8.4 J/g, and the overall enthalpy
`change on the cooling curves ranged from 1.8 to 2.6 J/g. The
`transformation to the austenitic NiTi structure on heating was
`completed at 25°C for all four specimens.
`The three Quantec specimens, which had an unknown clinical
`history, had more variable DSC plots than the ProFile and
`Lightspeed specimens, which were prepared from known as-
`received instruments. Figure 5 shows the DSC plot for one
`Quantec specimen; the general appearance of the plots was
`similar for the other two specimens, although there was sub-
`stantial variation in the peak temperatures. The heating curve in
`
`Figure 5 contains two clearly resolved endothermic peaks at
`approximately ⫺24° and 3°C that correspond to the transfor-
`mation from martensitic NiTi
`to R-phase, followed by the
`transformation from R-phase to austenitic NiTi. The enthalpy
`change of 4.3 J/g is for the overall transformation from mar-
`tensitic NiTi to austenitic NiTi. The cooling curve contains a
`single, broad, exothermic peak at approximately 1°C, which is
`assumed to include both the transformation from austenitic NiTi
`to R-phase at higher temperatures, followed by the transforma-
`tion from R-phase to martensitic NiTi because of the long
`shoulder on the left side of this broad peak. The enthalpy change
`
`PGR2015-00019 - Ex. 1038
`US ENDODONTICS, LLC., Petitioner
`
`

`

`Vol. 28, No. 8, August 2002
`
`DSC Analysis of NiTi Files
`
`571
`
`FIG 5. DSC plot for a Quantec specimen. The solid line is the heating curve, and the dashed line is the cooling curve.
`
`TABLE 1. Properties determined from the differential scanning calorimetric (DSC) plots for the three brands of
`endodontic instruments
`
`Property
`⌬H (martensitic NiTi to austenitic NiTi) on heating (J/g)
`⌬H (austenitic NiTi to martensitic NiTi) on cooling (J/g)
`Peak temperature, martensitic NiTi to R-phase on heating (°C)
`Peak temperature, R-phase to austenitic NiTi on heating (°C)
`Peak temperature, austenitic NiTi to R-phase on cooling (°C)
`Approximate completion temperature for transformation to
`austenitic NiTi (°C)
`
`ProFile*
`
`3.3 – 4.9
`2.8 – 3.6
`⫺25 – 14‡
`1 – 10
`⫺6 –⫹5
`25
`
`Lightspeed
`
`7.4 – 8.4
`1.8 – 2.6
`⫺18 – ⫺16
`2 – 3
`⫺3 – 0
`25
`
`Quantec†
`
`4.3 – 7.7
`1.6 – 2.7
`⫺25 – ⫺9
`3 – 24
`1 – 26
`25 – 50
`
`* Includes data for instruments and blanks.
`† Used instruments.
`‡ Not clearly defined for samples from three instruments, i.e. based on samples from two instruments and two wire blanks.
`
`of 2.7 J/g is likewise assumed to correspond to the overall
`transformation from austenitic NiTi to martensitic NiTi.
`The second Quantec specimen had endothermic peak tempera-
`tures of ⫺12° and 19°C on the heating curve, with an overall
`enthalpy change of 5.8 J/g. On the cooling curve, the exothermic
`peak temperature was 16°C with an overall enthalpy change of 1.6
`J/g. The third Quantec specimen had endothermic peak tempera-
`tures of ⫺9° and 24°C on the heating curve, with an overall
`enthalpy change of 7.7 J/g. On the cooling curve, the exothermic
`peak temperature was 26°C with an overall enthalpy change of 2.6
`J/g. Whereas the transformation to austenitic NiTi on heating was
`nearly completed at 25°C for the Quantec specimen in Figure 5, the
`temperatures were approximately 40° and 50°C for the other two
`specimens.
`Table 1 summarizes the properties determined from the DSC
`measurements on the three brands of NiTi rotary endodontic
`instruments. Ranges of the enthalpy changes are provided for
`the overall forward and reverse transformations between mar-
`tensitic NiTi and austenitic NiTi on heating and cooling, re-
`spectively. Ranges in peak temperatures are listed for the trans-
`formations from martensitic NiTi to R-phase and from R-phase
`to austenitic NiTi on heating and for the transformation from
`austenitic NiTi to R-phase on cooling. (There were no clearly
`
`defined peaks on the cooling DSC plots for the transformation
`from R-phase to martensitic NiTi.) Lastly,
`the approximate
`completion temperature (Af) is given for the transformation to
`austenitic NiTi on heating.
`
`DISCUSSION
`
`The DSC results in Figures 1, 2, and 4 show that the as-received
`ProFile and Lightspeed rotary endodontic instruments analyzed
`would be essentially in the completely austenitic NiTi condition at
`room temperature. These instruments would undergo superelastic
`behavior during clinical usage, where the imposition of stress
`causes transformation to martensitic NiTi and the removal of stress
`(withdrawal of the instrument from the root canal) results in
`reversible transformation to the original austenitic NiTi structure,
`unless more than approximately 10% tensile strain takes place
`(13). By analogy with the DSC results for shape memory NiTi
`orthodontic wires (11, 16), the NiTi alloy for the ProFile and
`Lightspeed rotary instruments would be capable of true shape
`memory under clinical conditions, if the manufacturers performed
`the necessary processing procedures (15). However, the shape
`memory property would not be necessary for the clinical use of
`
`PGR2015-00019 - Ex. 1038
`US ENDODONTICS, LLC., Petitioner
`
`

`

`572
`
`Brantley et al.
`
`Journal of Endodontics
`
`these instruments, provided that any tensile strain generated did not
`exceed approximately 10%.
`Differences in the transformation temperatures and enthalpies of
`the NiTi alloys for the ProFile and Lightspeed instruments (Figs.
`1, 2, and 4) are indicative of some variations in processing, but it
`is evident that the manufacturers have emphasized the importance
`of the Af temperature being near 25°C to permit superelastic
`behavior during clinical use. Differences in Figures 1–3 are attrib-
`uted to the manufacturing processing for converting the starting
`wire blanks to the ProFile instruments, but it is noteworthy that for
`both the wire blanks and the instruments the Af temperature is
`close to 25°C.
`Because of their unknown clinical history, it is not possible to
`comment about the sources of substantial variability in the DSC
`plots for specimens from the three Quantec instruments. This
`variability may have been due to processing differences for the
`as-received instruments or to their subsequent clinical use. A study
`is currently in progress to examine the effect of controlled simu-
`lated clinical usage on the DSC plots for the ProFile and Light-
`speed instruments.
`The enthalpy changes observed for the overall martensitic NiTi
`to austenitic NiTi transformations in the specimens ranged from
`2.8 to 4.4 J/g for the as-received ProFile instruments and from 1.8
`to 8.4 J/g for the as-received Lightspeed instruments. These values
`lie within the range of 1.7 to 19.2 J/g (0.4 – 4.6 cal/g), which has
`been reported (11) for some nickel-titanium orthodontic wires. It is
`tempting to speculate that the higher enthalpy changes of 7.4 to 8.4
`J/g for the overall transformation from martensitic NiTi to auste-
`nitic NiTi in the Lightspeed instruments, compared with values
`ranging from 3.3 to 4.4 J/g for the ProFile instruments, is due to
`lower amounts of stable, work-hardened martensitic NiTi (which
`does not undergo transformation to austenitic NiTi) in the former
`instruments after manufacturing. Although this might suggest that
`the ProFile instruments have a less advantageous microstructure,
`future research is required to provide unambiguous proof of this
`hypothesis.
`
`The authors thank Dr. John C. Mitchell (Department of Geological Sci-
`ences, The Ohio State University) for providing the laboratory facilities to
`prepare the test specimens.
`
`Dr. Brantley is professor, Section of Restorative Dentistry, Prosthodontics
`and Endodontics, and director of the graduate program in Dental Materials
`Science, College of Dentistry, The Ohio State University, Columbus, Ohio. Dr.
`Svec is associate professor, Division of Endodontics, Department of Stoma-
`tology; Dr. Powers is professor and vice chair, Department of Restorative
`Dentistry and Biomaterials, and director, Houston Biomaterials Research
`Center and the graduate program in Dental Materials Science, University of
`Texas-Houston Health Science Center, Dental Branch, Houston, Texas. Dr.
`Iijima is visiting scholar, College of Dentistry, The Ohio State University,
`Columbus, Ohio. His permanent affiliation is instructor, Department of Orth-
`odontics, School of Dentistry, Health Sciences University of Hokkaido, Ish-
`ikari-tobetsu, Japan. Mr. Grentzer is a staff member in Analytical Services &
`
`Technology, Research & Development Department, Ashland Specialty
`Chemical Company, DublinUAddress requests for reprints to Dr. Timothy A.
`Svec, Division of Endodontics, University of Texas-Houston Dental Branch,
`Houston, TX 77030-3402.
`
`References
`
`1. Walia H, Brantley WA, Gerstein H. An initial investigation of the bending
`and torsional properties of nitinol root canal files. J Endodon 1988;14:346 –51.
`2. Thompson SA, Dummer PM. Shaping ability of Lightspeed rotary nick-
`el-titanium instruments in simulated root canals: Part 1. J Endodon 1997;23:
`698 –702.
`3. Thompson SA, Dummer PM. Shaping ability of Quantec Series 2000
`rotary nickel-titanium instruments in simulated root canals: Part 2: Int Endod
`J 1998;31:268 –74.
`4. Mize SB, Clement DJ, Pruett JP, Carnes DL. Effect of sterilization on
`cyclic fatigue of rotary nickel-titanium endodontic instruments. J Endodon
`1998;24:843–7.
`5. Bryant ST, Dummer PM, Pitoni C, Bourba M, Moghal S. Shaping ability
`of .04 and .06 taper ProFile rotary nickel-titanium instruments in simulated
`root canals. Int Endod J 1999;32:155– 64.
`6. Eggert C, Peters O, Barbakow F. Wear of nickel-titanium Lightspeed
`instruments evaluated by scanning electron microscopy. J Endodon 1999;25:
`494 –7.
`7. Pettiette MT, Metzger Z, Phillips C, Trope M. Endodontic complications
`of root canal therapy performed by dental students with stainless-steel K-files
`and nickel-titanium hand files. J Endodon 1999;25:230 – 4.
`8. Schafer E, Lau R. Comparison of cutting efficiency and instrumentation
`of curved canals with nickel-titanium and stainless-steel instruments. J End-
`odon 1999;25:427–30.
`9. Carvalho LA, Bonetti I, Borges MA. A comparison of molar root canal
`preparation using stainless-steel and nickel-titanium instruments. J Endodon
`1999;25:807–10.
`10. Thompson SA. An overview of nickel-titanium alloys used in dentistry.
`Int Endod J 2000;33:297–310.
`11. Brantley WA. Orthodontic wires. In: Brantley WA, Eliades T, eds.
`Orthodontic materials: scientific and clinical aspects. Stuttgart: Thieme, 2001:
`77–103.
`12. Andreasen GF, Morrow RE. Laboratory and clinical analyses of nitinol
`wire. Am J Orthod 1978;73:142–51.
`13. Miura F, Mogi M, Ohura Y, Hamanaka H. The super-elastic property of
`the Japanese NiTi alloy wire for use in orthodontics. Am J Orthod Dentofacial
`Orthop 1986;90:1–10.
`14. Khier SE, Brantley WA, Fournelle RA. Bending properties of super-
`elastic and nonsuperelastic nickel-titanium orthodontic wires. Am J Orthod
`Dentofacial Orthop 1991;99:310 – 8.
`15. Civjan S, Huget EF, DeSimon LB. Potential applications of certain
`nickel-titanium (nitinol) alloys. J Dent Res 1975;54:89 –96.
`16. Bradley TG, Brantley WA, Culbertson BM. Differential scanning calo-
`rimetry (DSC) analyses of superelastic and nonsuperelastic nickel-titanium
`orthodontic wires. Am J Orthod Dentofacial Orthop 1996;109:589 –97.
`17. Otsuka K. Introduction to the R-phase transition. In: Duerig TW, Mel-
`ton KN, Sto¨ ckel D, Wayman CM, eds. Engineering aspects of shape memory
`alloys. London: Butterworth-Heinemann, 1990:36 – 45.
`18. Thayer TA, Bagby MD, Moore RN, De Angelis RJ. X-ray diffraction of
`nitinol orthodontic arch wires. Am J Orthod Dentofacial Orthop 1995;107:
`604 –12.
`19. Eliades G, Brantley WA. Instrumental techniques for study of orth-
`odontic materials. In: Brantley WA, Eliades T, eds. Orthodontic materials:
`scientific and clinical aspects. Stuttgart: Thieme, 2001:49 –76.
`20. Yoneyama T, Doi H, Hamanaka H, Okamoto Y, Mogi M, Miura F.
`Super-elasticity and thermal behavior of Ni-Ti alloy orthodontic arch wires.
`Dent Mater J 1992;11:1–10.
`
`PGR2015-00019 - Ex. 1038
`US ENDODONTICS, LLC., Petitioner
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket