`EXPERT REPORT OF ROBERT SINCLAIR, PH.D.
`
`I.
`
`INTRODUCTION
`
`1.
`
`I, Robert Sinclair, Ph.D., have been retained by Plaintiffs Dentsply International
`
`Inc. and Tulsa Dental Products LLC d/b/a Tulsa Dental Specialties (collectively “Plaintiffs” or
`
`“Dentsply”) to serve as an expert witness in the above-captioned matter. I expect to testify
`
`regarding the matters discussed in this report as well as any supplemental report that I may later
`
`submit and to respond to any testimony that Defendant US Endodontics, LLC (“US Endo”) may
`
`offer at the preliminary injunction hearing.
`
`II.
`
`BACKGROUND AND QUALIFICATIONS
`
`2.
`
`I hold a B.A. in Materials Science (1968) and a Ph.D. in Materials Science
`
`(1972), both from Cambridge University. After receiving my Ph.D., I worked from 1973-1977 as
`
`a Postdoctoral Research Engineer at the University of California, Berkeley.
`
`3.
`
`Since 1977, I have been employed at Stanford University in Stanford, California,
`
`where I have successively served as Assistant Professor in Materials Science and Engineering
`
`1 of 7
`
`PGR2015-00019 - Ex. 1019
`US ENDODONTICS, LLC., Petitioner
`
`

`
`37.
`
`The property imputed to the instrument as a result of step (b) in the ‘773 patent is:
`
`“wherein the heat-treated shank has an angle greater than 10 degrees of permanent deformation
`
`after torque at 45º flexion when tested in accordance with ISO Standard 3630-1.”
`
`C.
`
`38.
`
`Independent Testing of the EdgeFiles
`
`I have reviewed and am familiar with the results of the thermal testing (DSC)
`
`conducted on each of the EdgeFiles by Veritas.
`
`39.
`
`Samples of EdgeFiles were provided to Veritas as unopened packages of 3 or 6
`
`files in the form in which they are commercially available. Prior to testing, each file was cut
`
`from the tip of the EdgeFiles for each specimen into small segments using a diamond saw.
`
`Segments of sample material of approximately 5 mm were found to weigh between 5-15
`
`milligrams.
`
`40.
`
`The DSC testing performed by Veritas utilized a PerkinElmer DSC6000 with an
`
`Intracooler 2 instrument. The instrument was calibrated to the manufacturer’s specifications
`
`using indium and zinc standard supplied by the manufacturer. Each sample was placed in the
`
`DSC instrument, and that the instrument was cooled to -65 ºC for 1 minute before increasing the
`
`temperature at 10 ºC per minute until reaching 100 ºC. Samples of each EdgeFile type (X1, X3,
`
`X5, X7, and XR) were tested in triplicate.
`
`41.
`
`The DSC curves showing the results of the thermal analysis of the EdgeFiles were
`
`provided. The table below provides a key to the samples tested and the results. The procedure in
`
`the ASTM standard was utilized to calculate the As, Af, Ms, and Mf data points.
`
`2 of 7
`
`PGR2015-00019 - Ex. 1019
`US ENDODONTICS, LLC., Petitioner
`
`

`
`Product
`
`Size Length
`
`
`
`
`
`
`
`EdgeFile X1 25.06 25 mm
`
`
`
`
`
`
`
`
`
`
`
`
`
`EdgeFile X3
`
`N2
`
`25 mm
`
`
`
`
`
`
`
`
`
`
`
`
`
`R-Phase
`start
`temp
`(Rs)
`
`martensite
`start temp
`(Ms)
`
`R-Phase
`finish
`temp
`(Rf)3
`
`martensite
`finish
`temp (Mf)4
`
`28 °C
`
`28 °C
`
`27 °C
`
`29 °C
`
`29 °C
`
`28 °C
`
`28 °C
`
`-17 °C
`
`-18 °C
`
`-18 °C
`
`-37 °C
`
`-37 °C
`
`-38 °C
`
`-20 °C
`
`-19 °C
`
`21 °C
`
`22 °C
`
`21 °C
`
`21 °C
`
`19 °C
`
`20 °C
`
`22 °C
`
`22 °C
`
`-35 °C
`
`-36 °C
`
`-36 °C
`
`-51 °C
`
`-58 °C
`
`-51 °C
`
`-39 °C
`
`-39 °C
`
`DSC
`Curve2
`22714011
`Sample 1
`22714011
`Sample 2
`22714011
`Sample 3
`20614010
`Sample 1
`20614010
`Sample 2
`20614010
`Sample 3
`81913031
`Sample 1
`81913031
`Sample 2
`81913031
`Sample 3
`31014023
`Sample 1
`31014023
`Sample 2
`31014023
`Sample 3
`22714092
`Sample 1
`22714092
`Sample 2
`22714092
`Sample 3
`
`EdgeFile X5 40.04 25 mm
`
`
`
`
`
`
`
`
`
`
`
`
`
`EdgeFile X7 35.06 25 mm
`
`
`
`
`
`
`
`
`
`
`
`
`
`EdgeFile XR
`
`R4
`
`23 mm
`
`
`
`
`
`
`
`
`
`27 °C
`
`27 °C
`
`29 °C
`
`28 °C
`
`29 °C
`
`
`
`
`
`
`
`-19 °C
`
`-19 °C
`
`-19 °C
`
`-20 °C
`
`3 °C
`
`0 °C
`
`1 °C
`
`22 °C
`
`23 °C
`
`22 °C
`
`22 °C
`
`
`
`
`
`
`
`-38 °C
`
`-37 °C
`
`-37 °C
`
`-36 °C
`
`-38 °C
`
`-45 °C
`
`-42 °C
`
`42.
`
`Veritas also provided a supplemental analysis in which one DCS curve from each
`
`file tested was reproduced with the austenite finish temperature determination included. The
`
`2 The number corresponds to the Lot number on the packaging.
`3 This value was calculated by hand.
`4 This value was calculated by hand.
`
`3 of 7
`
`PGR2015-00019 - Ex. 1019
`US ENDODONTICS, LLC., Petitioner
`
`

`
`table below provides a key to the supplemental DSC curves. The procedure in the ASTM
`
`standard was utilized to calculate the As, Af, Ms, and Mf data points.
`
`Product
`
`Size
`
`Length
`
`Lot
`
`EdgeFile X1
`
`25.06
`
`EdgeFile X3
`
`N2
`
`EdgeFile X5
`
`EdgeFile X7
`
`40.04
`
`35.06
`
`EdgeFile XR
`
`R4
`
`
`
`25 mm
`
`25 mm
`
`25 mm
`
`25 mm
`
`23 mm
`
`22714011
`
`20614010
`
`81913031
`
`31014023
`
`22714092
`
`DSC Curve
`22714011
`Sample 1
`20614010
`Sample 3
`81913031
`Sample 2
`31014023
`Sample 3
`22714092
`Sample 1
`
`Austenite
`finish temp
`(Af)
`
`39 °C
`
`35 °C
`
`37 °C
`
`41 °C
`
`4 °C
`
`43.
`
`I have reviewed and am familiar with mechanical testing conducted on each of the
`
`EdgeFiles by Knight Mechanical. A report on the mechanical bend testing performed by Knight
`
`Mechanical in accordance with ISO Standard 3630-1 was provided. Samples of the EdgeFiles
`
`were provided to Knight Mechanical as unopened packages of 3 or 6 files in the form in which
`
`they are commercially available.
`
`44.
`
`A device was assembled in accordance with ISO Standard 3630-1 and a stiffness
`
`test was conducted in accordance with Section 7.5. After the stiffness test was completed, the
`
`files were measured for any angle of permanent deformation. The table below provides a key to
`
`the samples tested and the results.
`
`Size Length Sample
`Product
`
`
`X1_S1
`
`EdgeFile X1 25.06
`25 mm X1_S2
`
`
`
`X1_S3
`
`
`
`X3_S1
`EdgeFile X3 N2
`25 mm X3_S2
`
`Torque at 45°
`Angular
`Displacement
`(mN·m)
`3.60
`3.53
`3.35
`0.93
`0.91
`
`Post-testing Permanent
`Deformation Angle
`(degrees)
`23.75
`27.00
`27.33
`1.83
`3.58
`
`4 of 7
`
`PGR2015-00019 - Ex. 1019
`US ENDODONTICS, LLC., Petitioner
`
`

`
`
`
`
`
`EdgeFile X5 40.04
`
`
`
`
`EdgeFile X7 35.06
`
`
`
`
`EdgeFile XR R4
`
`
`
`
`X3_S3
`
`X5_S1
`
`25 mm X5_S2
`
`X5_S3
`
`X7_S1
`25 mm X7_S2
`
`X7_S3
`
`XR_S1
`23 mm XR_S2
`
`XR_S3
`
`1.35
`5.46
`4.85
`4.98
`6.49
`7.43
`7.44
`3.76
`3.43
`4.24
`
`2.17
`13.50
`11.75
`19.83
`29.00
`29.08
`26.67
`1.00
`0.25
`0.42
`
`D.
`
`45.
`
`Defendant US Endo’s Heat Treatment Process
`
`In his August 1, 2014 declaration, Mr. Bobby Bennett states that US Endo
`
`receives fluted, nickel titanium endodontic files from a third-party supplier, and then heat-treats
`
`the files in an industrial oven for the manufacture of EdgeFiles X1, X35, X5, and X7. (Bennett
`
`Declaration signed August 1, 2014 at ¶¶ 7, 8, 12, 13).
`
`46.
`
`Documents produced by US Endo outline the processing steps for the heat
`
`treatment of the EdgeFiles in more detail. First, the oven is warmed to a “proprietary
`
`temperature.” See, e.g., USENDO0000006 (Sequence # 4); USENDO0000011 (Sequence # 7).
`
`Next, the files are loaded into a metal tubes (USENDO0000006 (Sequence # 5)), and the tubes
`
`are loaded into the oven. USENDO0000007 (Sequence # 6). The oven door is shut and the tubes
`
`containing the entire files are heated for a period of time, which is recorded on a corresponding
`
`product router document. USENDO0000007 (Sequence # 7-9). The date, tube number, and time
`
`log for each heat treatment is recorded in the US Endo Heat Treat Log. See USENDO0000129-
`
`USENDO0000179. The quantities of files that are heat-treated, as well as the time of the day in
`
`which the tubes containing the files are loaded into and taken out of the oven, are recorded on
`
`US Endo Product Router documents. See, e.g., USENDO0000201.
`
`5 Mr. Bennett states that there is one type of the X3 EdgeFile having a specific taper that is not
`heat-treated; however, he admits that other X3 EdgeFiles are heat-treated.
`
`5 of 7
`
`PGR2015-00019 - Ex. 1019
`US ENDODONTICS, LLC., Petitioner
`
`

`
`E.
`
`Comparison of the EdgeFiles with Claim 1 of the ‘773 Patent
`
`Step 1: “providing”
`
`51.
`
`According to the declaration of Mr. Bobby Bennett of August 1, 2014, EdgeFiles
`
`X1, X5, and X7 are fluted, nickel titanium files that are heat-treated, whereas EdgeFile XR is a
`
`fluted, nickel-titanium file that is not treat-treated.
`
`52.
`
`As shown by the mechanical bend testing results, the EdgeFile XR does not
`
`demonstrate permanent deformation when subjected to the testing in standard ISO 3630-1. See
`
`supra at 54. In other words, when a bending force is applied to EdgeFile XR and subsequently
`
`released, the file will spring back straight rather than remain permanently deformed. Thus, the
`
`EdgeFile XR is superelastic. It follows that EdgeFiles X1, X5, and X7—which are fluted, nickel
`
`titanium files when received by US Endo—are also superelastic.
`
`53.
`
`Accordingly, it follows that US Endo’s process for manufacturing the EdgeFiles,
`
`or at least the EdgeFiles X1, X5, and X7, includes the first step in the ‘773 patent: “(a) providing
`
`an elongate shank having a cutting edge extending from a distal end of the shank along an axial
`
`length of the shank, comprising a superelastic nickel titanium alloy.” In other words, before heat
`
`treatment, the EdgeFiles have a cutting edge and comprise a superelastic nickel titanium alloy.
`
`Step 2: “heat-treating”
`
`54.
`
`The product packaging for Edge Endo’s EdgeFiles X1, X5 and X7 indicate that
`
`the products are “HEAT TREATED NiTi FILES.”
`
`55.
`
`The August 1, 2014 declaration submitted by Mr. Bobby Bennett states that the
`
`EdgeFiles are prepared by heat-treating nickel titanium files.
`
`56.
`
`The procedure outlined in the US Endo documents call for the following steps:
`
`load the files into tubes; load the tubes into the oven; and close the oven door. See
`
`6 of 7
`
`PGR2015-00019 - Ex. 1019
`US ENDODONTICS, LLC., Petitioner
`
`

`
`7 of 7
`
`PGR2015-00019 - Ex. 1019
`US ENDODONTICS, LLC., Petitioner

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket