`
`General Ctraaters: <521> CHROMATOGRAPHY
`
`us. PHARMACOPEIA
`
`CTC Syringes
`
`Search USP29
`
`L-MARK® GC autosampler
`syringes for CTC PAL®
`autosamplers
`O 0
`
`{ 621 3 CHROMATOGRAPHY
`
`INTRODUCTION
`
`This chapter defines the terms and procedures used in chromatography and provides general information. Specific
`
`requirements for chromatographic procedures for drug substances and dosage forms, including adsorbent and developing
`
`solvents, are given in the individual monographs.
`
`Chromatography is defined as a procedure by which solutes are separated by a dynamic differential migration process in a
`
`system consisting of two or more phases, one of which moves continuously in a given direction and in which the individual
`
`substances exhibit different mobilities by reason of differences in adsorption, partition, solubility, vapor pressure, molecular
`
`size, or ionic charge density. The individual substances thus separated can be identified or determined by analytical
`
`procedures.
`
`The general chromatographic technique requires that a solute undergo distribution between two phases, one of them fixed
`
`(stationary phase), the other moving (mobile phase). It is the mobile phase that transfers the solute through the medium
`
`until it eventually emerges separated from other solutes that are eluted earlier or later. Generally, the solute is transported
`
`through the separation medium by means of a flowing stream of a liquid or a gaseous solvent known as the “eluant.” The
`
`stationary phase may act through adsorption, as in the case of adsorbents such as activated alumina and silica gel, or it
`
`may act by dissolving the solute, thus partitioning the latter between the stationary and mobile phases. In the latter
`
`process, a liquid coated onto an inert support, or chemically bonded onto silica gel, or directly onto the wall of a fused silica
`
`capillary, serves as the stationary phase. Partitioning is the predominant mechanism of separation in gas—liquid
`
`chromatography, paper chromatography, in forms of column chromatography and in thin-layer chromatography designated
`
`as liquid-liquid separation. In practice, separations frequently result from a combination of adsorption and partitioning
`
`effects. Other separation principles include ion exchange, ion-pair formation, size exclusion, hydrophobic interaction, and
`
`chiral recognition.
`
`The types of chromatography useful in qualitative and quantitative analysis that are employed in the USP procedures are
`
`column, gas, paper, thin-layer, (including high-perforrnance thin-layer chromatography), and pressurized liquid
`
`chromatography (commonly called high-pressure or high-perfonnance liquid chromatography). Paper and thin-layer
`
`chromatography are ordinarily more useful for purposes of identification, because of their convenience and simplicity.
`
`Column chromatography offers a wider choice of stationary phases and is useful for the separation of individual
`
`compounds, in quantity, from mixtures. Modern high-perforrnance thin-layer chromatography, gas chromatography, and
`
`pressurized liquid chromatography require more elaborate apparatus but usually provide high resolution and identify and
`
`quantitate very small amounts of material.
`
`Use of Reference Substances in Identity Tests— In paper and thin-layer chromatography, the ratio of the distance (this
`
`htlpJ'M'ww.pl1armaoopeia.crIN29240lusp29nf24s0_c621_viewa||.htm|
`
`PARAGON — EXHIBIT 2010 1126
`
`PARAGON - EXHIBIT 2010
`
`
`
`8r‘24r'2015
`
`General Chaaters: <621> CHROMATOGRAPHY
`
`distance being measured to the point of maximum intensity of the spot or zone) traveled on the medium by a given
`
`compound to the distance traveled by the front of the mobile phase, from the point of application of the test substance, is
`
`designated as the R,_- value of the compound. The ratio between the distances traveled by a given compound and a
`
`reference substance is the RR value. RR values vary with the experimental conditions, and thus identification is best
`
`accomplished where an authentic specimen of the compound in question is used as a reference substance on the same
`
`chromatogram.
`
`For this purpose, chromatograms are prepared by applying on the thin-layer adsorbent or on the paper in a straight line,
`
`parallel to the edge of the chromatographic plate or paper, solutions of the substance to be identified, the authentic
`
`specimen, and a mixture of nearly equal amounts of the substance to be identified and the authentic specimen. Each
`
`sample application contains approximately the same quantity by weight of material to be chromatographed. If the substance
`
`to be identified and the authentic specimen are identical, all chromatograms agree in color and RR value and the mixed
`
`chromatogram yields a single spot; i.e., RR is 1.0.
`
`Location of Components— The spots produced by paper or thin-layer chromatography may be located by: (1) direct
`
`inspection if the compounds are visible under white or either short-wavelength (254 nm) or long-wavelength (360 nm) UV
`
`light, (2) inspection in white or UV light after treatment with reagents that will make the spots visible (reagents are most
`
`conveniently applied with an atomizer), (3) use of a Geiger-Muller counter or autoradiographic techniques in the case of the
`
`presence of radioactive substances, or (4) evidence resulting from stimulation or inhibition of bacterial growth by the placing
`
`of removed portions of the adsorbent and substance on inoculated media.
`
`In open-column chromatography, in pressurized liquid chromatography performed under conditions of constant flow rate,
`
`and in gas chromatography, the retention time, t, defined as the time elapsed between sample injection and appearance of
`
`the peak concentration of the eluted sample zone, may be used as a parameter of identification. Solutions of the substance
`
`to be identified or derivatives thereof, of the reference compound, and of a mixture of equal amounts of these two are
`
`chromatographed successively on the same column under the same chromatographic conditions. Only one peak should be
`
`observed for the mixture. The ratio of the retention times of the test substance, the reference compound, and a mixture of
`
`these, to the retention time of an internal standard is called the relative retention time RR and is also used frequently as a
`
`parameter of identification.
`
`The deviations of RR, RR, or t values measured for the test substance from the values obtained for the reference compound
`
`and mixture should not exceed the reliability estimates determined statistically from replicate assays of the reference
`
`compound.
`
`Chromatographic identification by these methods under given conditions strongly indicates identity but does not constitute
`
`definitive identification. Coincidence of identity parameters under three to six different sets of chromatographic conditions
`
`(temperatures, column packings, adsorbents, eluants, developing solvents, various chemical derivatives, etc.) increases
`
`the probability that the test and reference substances are identical. However, many isomeric compounds cannot be
`
`separated. Specific and pertinent chemical, spectroscopic, or physicochemical identification of the eluted component
`
`combined with chromatographic identity is the most valid criterion of identification. For this purpose, the individual
`
`components separated by chromatography may be collected for further identification.
`
`PAPER CHROMATOGRAPHY
`
`In paper chromatography the adsorbent is a sheet of paper of suitable texture and thickness. Chromatographic separation
`
`may proceed through the action of a single liquid phase in a process analogous to adsorption chromatography in columns.
`
`Since the natural water content of the paper, or selective imbibition of a hydrophilic component of the liquid phase by the
`
`paper fibers, may be regarded as a stationary phase, a partitioning mechanism may contribute significantly to the
`
`separation.
`
`Alternatively, a two-phase system may be used. The paper is impregnated with one of the phases, which then remains
`
`stationary (usually the more polar phase in the case of unmodified paper). The chromatogram is developed by slow passage
`
`ltlpJ'M'ww.pl1armaoopeia.crIN2924{llusp29nf24s0_c621_viewa||.html
`
`2r'26
`
`
`
`8r‘24r'2015
`
`General Chaaters: <621> CHROMATOGRAPHY
`
`of the other, mobile phase over the sheet. Development may be ascending, in which case the solvent is carried up the
`
`paper by capillary forces, or descending, in which case the solvent flow is also assisted by gravitational force.
`
`Differences in the value of RF have been reported where chromatograms developed in the direction of the paper grain
`
`(machine direction) are compared with others developed at right angles to the grain; therefore, the orientation of paper grain
`
`with respect to solvent flow should be maintained constant in a series of chromatograms. (The machine direction is usually
`
`designated by the manufacturer on packages of chromatography paper.)
`
`In descending chromatography, the mobile phase flows downward on the chromatographic sheet.
`
`Apparatus— The essential equipment for descending chromatography consists of the following:
`
`Descending Chromatography
`
`A vapor-tight chamber provided with inlets for addition of solvent or for releasing internal pressure. The chamber is
`
`constructed preferably of glass, stainless steel, or porcelain and is so designed as to permit obsen/ation of the progress of
`
`the chromatographic run without opening of the chamber. Tall glass cylinders are convenient if they are made vapor-tight
`
`with suitable covers and a sealing compound.
`
`A rack of comosion-resistant material about 5 cm shorter than the inside height of the chamber. The rack serves as a
`
`support for solvent troughs and for antisiphon rods which, in turn, hold up the chromatographic sheets.
`
`One or more glass troughs capable of holding a volume of solvent greater than that needed for one chromatographic run.
`
`The troughs must also be longer than the width of the chromatographic sheets.
`
`Heavy glass antisiphon rods to be supported by the rack and running outside of, parallel to, and slightly above the edge of
`
`the glass trough.
`
`Chromatographic sheets of special filter paper at least 2.5 cm wide and not wider than the length of the troughs are cut to a
`
`length approximately equal to the height of the chamber. A fine pencil line is drawn horizontally across the filter paper at a
`
`distance from one end such that, when the sheet is suspended from the antisiphon rods with the upper end of the paper
`
`resting in the trough and the lower portion hanging free into the chamber, the line is located a few centimeters below the
`
`rods. Care is necessary to avoid contaminating the filter paper by excessive handling or by contact with dirty surfaces.
`
`Procedure— The substance or substances to be analyzed are dissolved in a suitable solvent. Convenient volumes,
`
`delivered from suitable micropipets, of the resulting solution, normally containing 1 to 20 pg of the compound, are placed in
`
`6- to 10-mm spots not less than 3 cm apart along the pencil line. If the total volume to be applied would produce spots of a
`
`diameter greater than 6 to 10 mm, it is applied in separate portions to the same spot, each portion being allowed to dry
`before the next is added.
`
`The spotted chromatographic sheet is suspended in the chamber by use of the antisiphon rod, which holds the upper end of
`
`the sheet in the solvent trough. The bottom of the chamber is covered with the prescribed solvent system. Saturation of the
`
`chamber with solvent vapor is facilitated by lining the inside walls with paper that is wetted with the prescribed solvent
`
`system. It is important to ensure that the portion of the sheet hanging below the rods is freely suspended in the chamber
`
`without touching the rack or the chamber walls or the fluid in the chamber. The chamber is sealed to allow equilibration
`
`(saturation) of the chamber and the paper with the solvent vapor. Any excess pressure is released as necessary. For large
`
`chambers, equilibration overnight may be necessary.
`
`A volume of the mobile phase in excess of the volume required for complete development of the chromatogram is saturated
`
`with the immobile phase by shaking. After equilibration of the chamber, the prepared mobile solvent is introduced into the
`
`trough through the inlet. The inlet is closed and the mobile solvent phase is allowed to travel the desired distance down the
`
`paper. Precautions must be taken against allowing the solvent to run down the sheet when opening the chamber and
`
`removing the chromatogram. The location of the solvent front is quickly marked, and the sheets are dried.
`
`The chromatogram is observed and measured directly or after suitable development to reveal the location of the spots of
`
`the isolated drug or drugs. The paper section(s) predetermined to contain the isolated drug(s) may be cut out and eluted by
`
`an appropriate solvent, and the solutions may be made up to a known volume and quantitatively analyzed by appropriate
`
`ltlpJ'M'ww.pl1armaoopeia.crIN2924{ltusp29nf24s0_c621_viewa||.html
`
`3126
`
`
`
`8l24l2015
`
`General Chaaters: <621> CHROMATOGRAPHY
`
`chemical or instrumental techniques. Similar procedures should be conducted with various amounts of similarly spotted
`
`reference standard on the same paper in the concentration range appropriate to prepare a valid calibration curve.
`
`Ascending Chromatography
`
`In ascending chromatography, the lower edge of the sheet (or strip) is dipped into the mobile phase to permit the mobile
`
`phase to rise on the chromatographic sheet by capillary action.
`
`Apparatus— The essential equipment for ascending chromatography is substantially the same as that described under
`
`Descending Chromatography.
`
`Procedure— The test materials are applied to the chromatographic sheets as directed under Descending Chromatography,
`
`and above the level to which the paper is dipped into the developing solvent. The bottom of the developing chamber is
`
`covered with the developing solvent system. If a two-phase system is used, both phases are added. It is also desirable to
`
`line the walls of the chamber with paper and to saturate this lining with the solvent system. Empty solvent troughs are
`
`placed on the bottom of the chamber, and the chromatographic sheets are suspended so that the end on which the spots
`
`have been added hangs free inside the empty trough.
`
`The chamber is sealed, and equilibration is allowed to proceed as described under Descending Chromatography. Then the
`
`developing solvent (mobile phase) is added through the inlet to the trough in excess of the solvent required for complete
`
`moistening of the chromatographic sheet. The chamber is resealed. When the solvent front has reached the desired height,
`
`the chamber is opened and the sheet is removed and dried.
`
`Quantitative analyses of the spots may be conducted as described under Descending Chromatography.
`
`THIN-LAYER CHROMATOGRAPHY
`
`In thin-layer chromatography, the adsorbent is a relatively thin, unifonn layer of dry, finely powdered material applied to a
`
`glass, plastic, or metal sheet or plate, glass plates being most commonly employed. The coated plate can be considered
`
`an “open chromatographic column” and the separations achieved may be based upon adsorption, partition, or a combination
`
`of both effects, depending on the particular type of stationary phase, its preparation, and its use with different solvents.
`
`Thin-layer chromatography on ion-exchange layers can be used for the fractionation of polar compounds. Presumptive
`
`identification can be effected by obsenration of spots or zones of identical RF value and about equal magnitude obtained,
`
`respectively, with an unknown and a reference sample chromatographed on the same plate. A visual comparison of the size
`
`or intensity of the spots or zones may serve for semiquantitative estimation. Quantitative measurements are possible by
`
`means of densitometry (absorbance or fluorescence measurements), or the spots may be carefully removed from the plate,
`
`followed by elution with a suitable solvent and spectrophotometric measurement. For two-dimensional thin-layer
`
`chromatography, the chromatographed plate is turned at a right angle and again chromatographed, usually in another
`
`chamber equilibrated with a different solvent system.
`
`Apparatus— Acceptable apparatus and materials for thin-layer chromatography consist of the following.
`
`A TLC or HPTLC plate. The chromatography is generally carried out using precoated plates or sheets (on glass, aluminum,
`
`or polyester support) of suitable size. It may be necessary to clean the plates prior to separation. This can be done by
`
`migration of, or immersion in, an appropriate solvent. The plates may also be impregnated by procedures such as
`
`development, immersion, or spraying. At the time of use, the plates may be activated, if necessary, by heating in an oven
`
`at 120': for 20 minutes. The stationary phase of TLC plates has an average particle size of 10-15 pm, and that of HPTLC
`plates an average particle size of 5 pm. Commercial plates with a preadsorbant zone can be used if they are specified in a
`
`monograph. Sample applied to the preabsorbant region develops into sharp, narrow bands at the preabsorbant-sorbent
`
`interface. Alternatively, flat glass plates of convenient size, typically 20 cm X 20 cm can be coated as described under
`
`Preparation of Chromatographic Plates.
`
`A suitable manual, semiautomatic, or automatic application device can be used to ensure proper positioning of the plate
`
`and proper transfer of the sample, with respect to volume and position, onto the plate. Alternatively, a template can be used
`
`to guide in manually placing the test spots at definite intervals, to mark distances as needed, and to aid in labeling the
`
`ltlpJ'M'ww.pl1armaoopeia.crIl\r2924{ltusp29nf24s0_c621_viewa||.html
`
`#26
`
`
`
`8r‘24.i2015
`
`General Chaaters: <621> CHROMATOGRAPHY
`
`plates. For the proper application of the solutions, micropipets, microsyringes, or calibrated disposable capillaries are
`recommended.
`
`For ascending development, a chromatographic chamber made of inert, transparent material and having the following
`
`specifications is used: a flat bottom or twin trough, a tightly fitted lid, and a size suitable for the plates. For horizontal
`
`development, the chamber is provided with a reservoir for the mobile phase, and it also contains a device for directing the
`
`mobile phase to the stationary phase.
`
`Devices for transfer of reagents onto the plate by spraying, immersion, or exposure to vapor and devices to facilitate any
`
`necessary heating for visualization of the separated spots or zones.
`
`A UV light source suitable for observations under short (254 nm) and long (365 nm) wavelength UV light.
`
`A suitable device for documentation of the visualized chromatographic result.
`
`Procedure— Apply the prescribed volume of the test solution and the standard solution in sufficiently small portions to
`
`obtain circular spots of 2 to 5 mm in diameter (1 to 2 mm on HPTLC plates) or bands of 10 to 20 mm by 1 to 2 mm (5 to 10
`
`mm by 0.5 to 1 mm on HPTLC plates) at an appropriate distance from the lower edge—during chromatography the
`
`application position must be 3 mm (HPTLC) to 5 mm (TLC) above the level of the developing solvent—and from the sides
`
`of the plate. Apply the solutions on a line parallel to the lower edge of the plate with an interval of at least 10 mm (5 mm on
`
`HPTLC plates) between the centers of spots or 4 mm (2 mm on HPTLC plates) between the edges of bands, and allow to
`
`dry.
`
`Ascending Development— Line at least one wall of the chromatographic chamber with filter paper. Pour into the
`
`chromatographic chamber a quantity of the mobile phase sufficient for the size of the chamber to give, after impregnation of
`
`the filter paper, a level of depth appropriate to the dimension of the plate used. For saturation of the chromatographic
`
`chamber, close the lid, and allow the system to equilibrate. Unless otherwise indicated, the chromatographic separation is
`
`performed in a saturated chamber.
`
`Place the plate in the chamber, ensuring that the plate is as vertical as possible and that the spots or bands are above the
`
`surface of the mobile phase, and close the chamber. The stationary phase faces the inside of the chamber. Remove the
`
`plate when the mobile phase has moved over the prescribed distance. Dry the plate, and visualize the chromatograms as
`
`prescribed. For two-dimensional chromatography, dry the plates after the first development, and cany out a second
`
`development in a direction perpendicular to that of the first development.
`
`Horizontal Development— Introduce a sufficient quantity of the developing solvent into the reservoir of the chamber using a
`
`syringe or pipet. Place the plate horizontally in the chamber, connect the mobile phase direction device according to the
`
`manufacturer's instructions, and close the chamber. If prescribed, develop the plate starting simultaneously at both ends.
`
`Remove the plate when the mobile phase has moved over the distance prescribed in the monograph. Dry the plate, and
`
`visualize the chromatograms as prescribed.
`
`For two-dimensional chromatography, dry the plates after the first development, and carry out a second development in a
`
`direction perpendicular to that of the first development.
`
`Detection— Observe the dry plate first under short-wavelength UV light (254 nm) and then under long-wavelength UV light
`
`(365 nm) or as stated in the monograph. If further directed, spray, immerse, or expose the plate to vapors of the specified
`
`reagent, heat the plate when required, obsewe, and compare the test chromatogram with the standard chromatogram.
`
`Document the plate after each obsenration. Measure and record the distance of each spot or zone from the point of origin,
`
`and indicate for each spot or zone the wavelength under which it was observed. Detennine the RF values for the principal
`
`spots or zones (see Glossary of Symbols).
`
`Quantitative Measurement— Using appropriate instrumentation, substances separated by TLC and responding to ultraviolet-
`
`visible (UV-Vis) irradiation prior to or after derivatization can be detennined directly on the plate. While moving the plate or
`
`the measuring device, the plate is examined by measuring the reflectance of the incident light. Similarly, fluorescence may
`
`be measured using an appropriate optical system. Substances containing radionuclides can be quantified in three ways: (1)
`
`directly by moving the plate alongside a suitable counter or vice versa; (2) by cutting the plates into strips and measuring
`
`l'r:lp33M-'ww.pl1armaoopeia.crIN2924{llusp29nf24s0_o621_viewa||.html
`
`5l26
`
`
`
`8i24i2015
`
`General Chapters: <a21> CHROMATOGRAPHY
`
`the radioactivity on each individual strip using a suitable counter; or (3) by scraping off the stationary phase, dissolving it in
`
`a suitable scintillation cocktail, and measuring the radioactivity using a liquid scintillation counter (see Radioactivity ( g }
`).
`
`The apparatus for direct quantitative measurement on the plate is a densitometer that is composed of a mechanical device
`
`to move the plate or the measuring device along the x-axis and the y-axis, a recorder, a suitable integrator or a computer;
`
`and, for substances responding to UV-Vis irradiation, a photometer with a source of light, an optical device capable of
`
`generating monochromatic light, and a photo cell of adequate sensitivity, all of which are used for the measurement of
`
`reflectance. In the case where fluorescence is measured, a suitable filter is also required to prevent the light used for
`
`excitation from reaching the photo cell while permitting the emitted light or specific portions thereof to pass. The linearity
`
`range of the counting device must be verified.
`
`For quantitative tests, it is necessary to apply to the plate not fewer than three standard solutions of the substance to be
`
`examined, the concentrations of which span the expected value in the test solution (e.g., 80%, 100%, and 120%).
`
`Derivatize with the prescribed reagent, if necessary, and record the reflectance or fluorescence in the chromatograms
`obtained. Use the measured results for the calculation of the amount of substance in the test solution.
`
`Preparation of Chromatographic Plates-
`
`Apparatus-
`
`Flat glass plates of convenient size, typically 20 cm x 20 cm.
`
`An aligning tray or a flat surface upon which to align and rest the plates during the application of the adsorbent.
`
`A storage rack to hold the prepared plates during drying and transportation. The rack holding the plates should be kept in a
`
`desiccator or be capable of being sealed in order to protect the plates from the environment after removal from the drying
`oven.
`
`The adsorbent consists of finely divided adsorbent materials, normally 5 to 40 pm in diameter, suitable for chromatography.
`
`It can be applied directly to the glass plate or can be bonded to the plate by means of plaster of Paris [calcium sulfate
`
`hemihydrate (at a ratio of 5% to 15%)] or with starch paste or other binders. The plaster of Paris will not yield as hard a
`
`surface as will the starch, but it is not affected by strongly oxidizing spray reagents. The adsorbent may contain fluorescing
`
`material to aid in the visualization of spots that absorb UV light.
`
`A spreader, which, when moved over the glass plate, will apply a uniform layer of adsorbent of desired thickness over the
`
`entire surface of the plate.
`
`Procedure— [NOTE—|n this procedure, use Purified Water that is obtained by distillation.] Clean the glass plates
`
`scrupulously, using an appropriate cleaning solution (see Cleaning Glass Aggaratus ( 1051 ) ), rinsing them with copious
`quantities of water until the water runs off the plates without leaving any visible water or oily spots, then dry. It is important
`
`that the plates be completely free from lint and dust when the adsorbent is applied.
`
`Arrange the plate or plates on the aligning tray, place a 5- x 20-cm plate adjacent to the front edge of the first square plate
`
`and another 5- x 20-cm plate adjacent to the rear edge of the last square, and secure all of the plates so that they will not
`
`slip during the application of the adsorbent. Position the spreader on the end plate opposite the raised end of the aligning
`
`tray. Mix 1 part of adsorbent with 2 parts of water (or in the ratio suggested by the supplier) by shaking vigorously for 30
`
`seconds in a glass-stoppered conical flask, and transfer the slurry to the spreader. Usually 30 g of adsorbent and 60 mL of
`
`water are sufficient for five 20- X 20-cm plates. Complete the application of adsorbents using plaster of Paris binder within 2
`
`minutes of the addition of the water, because thereafter the mixture begins to harden. Draw the spreader smoothly over the
`
`plates toward the raised end of the aligning tray, and remove the spreader when it is on the end plate next to the raised end
`
`of the aligning tray. (Wash away all traces of adsorbent from the spreader immediately after use.) Allow the plates to remain
`
`undisturbed for 5 minutes, then transfer the square plates, layer side up, to the storage rack, and dry at 105° for 30
`minutes. Preferably place the rack at an angle in the drying oven to prevent the condensation of moisture on the back sides
`
`of plates in the rack. When the plates are dry, allow them to cool to room temperature, and inspect the uniformity of the
`
`lt1pJ'M'ww.pl1armaoopeia.crIN2924{Iiusp29nf24s0_c621_viewa||.html
`
`6l26
`
`
`
`8r‘24r'2015
`
`General Chaaters: <621> CHROMATOGRAPHY
`
`distribution and the texture of the adsorbent layer; transmitted light will show uniformity of distribution, and reflected light will
`
`show uniformity of texture. Store the satisfactory plates over silica gel in a suitable chamber.
`
`COLUMN CHROMATOGRAPHY
`
`Apparatus— The apparatus required for column chromatographic procedures is simple, consisting only of the
`
`chromatographic tube itself and a tamping rod, which may be needed to pack a pledget of glass wool or cotton, if needed, in
`
`the base of the tube and compress the adsorbent or slurry uniformly within the tube. In some cases a porous glass disk is
`
`sealed at the base of the tube in order to support the contents. The tube is cylindrical and is made of glass, unless another
`
`material is specified in the individual monograph. A smaller-diameter delivery tube is fused or otherwise attached by a
`
`leakproof joint to the lower end of the main tube. Column dimensions are variable; the dimensions of those commonly used
`
`in pharmaceutical analysis range from 10 to 30 mm in uniform inside diameter and 150 to 400 mm in length, exclusive of
`
`the delivery tube. The delivery tube, usually 3 to 6 mm in inside diameter, may include a stopcock for accurate control of
`
`the flow rate of solvents through the column. The tamping rod, a cylindrical ram firmly attached to a shaft, may be
`
`constructed of plastic, glass, stainless steel, or aluminum, unless another material is specified in the individual monograph.
`
`The shaft of the rod is substantially smaller in diameter than the column and is not less than 5 cm longer than the effective
`
`length of the column. The ram has a diameter about 1 mm smaller than the inside diameter of the column.
`
`Column Adsorption Chromatography
`
`The adsorbent (such as activated alumina or silica gel, calcined diatomaceous silica, or chromatographic purified siliceous
`
`earth) as a dry solid or as a slurry is packed into a glass or quartz chromatographic tube. A solution of the drug in a small
`
`amount of solvent is added to the top of the column and allowed to flow into the adsorbent. The drug principles are
`
`quantitatively removed from the solution and are adsorbed in a narrow transverse band at the top of the column. As
`
`additional solvent is allowed to flow through the column, either by gravity or by application of air pressure, each substance
`
`progresses down the column at a characteristic rate resulting in a spatial separation to give what is known as the
`
`chromatogram. The rate of movement for a given substance is affected by several variables, including the adsorptive power
`
`of the adsorbent and its particle size and surface area; the nature and polarity of the solvent; the hydrostatic head or
`
`applied pressure; and the temperature of the chromatographic system.
`
`If the separated compounds are colored or if they fluoresce under UV light, the adsorbent column may be extruded and, by
`
`transverse cuts, the appropriate segments may then be isolated. The desired compounds are then extracted from each
`
`segment with a suitable solvent. If the compounds are colorless, they may be located by means of painting or spraying the
`
`extruded column with color-forming reagents. Chromatographed radioactive substances may be located by means of
`
`Geiger-Muller detectors or similar sensing and recording instruments. Clear plastic tubing made of a material such as nylon,
`
`which is inert to most solvents and transparent to short-wavelength UV light, may be packed with adsorbent and used as a
`
`chromatographic column. Such a column may be sliced with a sharp knife without removing the packing from the tubing. If
`
`a fluorescent adsorbent is used, the column may be marked under UV light in preparation for slicing.
`
`A "flowing" chromatogram, which is extensively used, is obtained by a procedure in which solvents are allowed to flow
`
`through the column until the separated drug appears in the effluent solution, known as the "eluate.” The drug may be
`
`detennined in the eluate by titration or by a spectrophotometric or colorimetric method, or the solvent may be evaporated,
`
`leaving the drug in more or less pure fonn. If a second drug principle is involved, it is eluted by continuing the first solvent
`
`or by passing a solvent of stronger eluting power through the column. The efficiency of the separation may be checked by
`
`obtaining a thin-layer chromatogram on the individual fractions.
`
`A modified procedure for adding the mixture to the column is sometimes employed. The drug, in a solid form, and, as in the
`
`case of a powdered tablet, without separation from the excipients, is mixed with some of the adsorbent and added to the
`
`top of a column. The subsequent flow of solvent moves the drug down the column in the manner described.
`
`Column Partition Chromatography
`
`In partition chromatography the substances to be separated are partitioned between two immiscible liquids, one of which,
`
`the immobile phase, is adsorbed on a Solid Support, thereby presenting a very large surface area to the flowing solvent or
`
`ltlpJ'M'ww.pl1armaoopeia.crIN2924{llusp29nf24s0_c621_viewa||.html
`
`7126
`
`
`
`8i24l2015
`
`General Chaaters: <621> CHROMATOGRAPHY
`
`mobile phase. The exceed