throbber
8r‘24r'2015
`
`General Ctraaters: <521> CHROMATOGRAPHY
`
`us. PHARMACOPEIA
`
`CTC Syringes
`
`Search USP29
`
`L-MARK® GC autosampler
`syringes for CTC PAL®
`autosamplers
`O 0
`
`{ 621 3 CHROMATOGRAPHY
`
`INTRODUCTION
`
`This chapter defines the terms and procedures used in chromatography and provides general information. Specific
`
`requirements for chromatographic procedures for drug substances and dosage forms, including adsorbent and developing
`
`solvents, are given in the individual monographs.
`
`Chromatography is defined as a procedure by which solutes are separated by a dynamic differential migration process in a
`
`system consisting of two or more phases, one of which moves continuously in a given direction and in which the individual
`
`substances exhibit different mobilities by reason of differences in adsorption, partition, solubility, vapor pressure, molecular
`
`size, or ionic charge density. The individual substances thus separated can be identified or determined by analytical
`
`procedures.
`
`The general chromatographic technique requires that a solute undergo distribution between two phases, one of them fixed
`
`(stationary phase), the other moving (mobile phase). It is the mobile phase that transfers the solute through the medium
`
`until it eventually emerges separated from other solutes that are eluted earlier or later. Generally, the solute is transported
`
`through the separation medium by means of a flowing stream of a liquid or a gaseous solvent known as the “eluant.” The
`
`stationary phase may act through adsorption, as in the case of adsorbents such as activated alumina and silica gel, or it
`
`may act by dissolving the solute, thus partitioning the latter between the stationary and mobile phases. In the latter
`
`process, a liquid coated onto an inert support, or chemically bonded onto silica gel, or directly onto the wall of a fused silica
`
`capillary, serves as the stationary phase. Partitioning is the predominant mechanism of separation in gas—liquid
`
`chromatography, paper chromatography, in forms of column chromatography and in thin-layer chromatography designated
`
`as liquid-liquid separation. In practice, separations frequently result from a combination of adsorption and partitioning
`
`effects. Other separation principles include ion exchange, ion-pair formation, size exclusion, hydrophobic interaction, and
`
`chiral recognition.
`
`The types of chromatography useful in qualitative and quantitative analysis that are employed in the USP procedures are
`
`column, gas, paper, thin-layer, (including high-perforrnance thin-layer chromatography), and pressurized liquid
`
`chromatography (commonly called high-pressure or high-perfonnance liquid chromatography). Paper and thin-layer
`
`chromatography are ordinarily more useful for purposes of identification, because of their convenience and simplicity.
`
`Column chromatography offers a wider choice of stationary phases and is useful for the separation of individual
`
`compounds, in quantity, from mixtures. Modern high-perforrnance thin-layer chromatography, gas chromatography, and
`
`pressurized liquid chromatography require more elaborate apparatus but usually provide high resolution and identify and
`
`quantitate very small amounts of material.
`
`Use of Reference Substances in Identity Tests— In paper and thin-layer chromatography, the ratio of the distance (this
`
`htlpJ'M'ww.pl1armaoopeia.crIN29240lusp29nf24s0_c621_viewa||.htm|
`
`PARAGON — EXHIBIT 2010 1126
`
`PARAGON - EXHIBIT 2010
`
`

`

`8r‘24r'2015
`
`General Chaaters: <621> CHROMATOGRAPHY
`
`distance being measured to the point of maximum intensity of the spot or zone) traveled on the medium by a given
`
`compound to the distance traveled by the front of the mobile phase, from the point of application of the test substance, is
`
`designated as the R,_- value of the compound. The ratio between the distances traveled by a given compound and a
`
`reference substance is the RR value. RR values vary with the experimental conditions, and thus identification is best
`
`accomplished where an authentic specimen of the compound in question is used as a reference substance on the same
`
`chromatogram.
`
`For this purpose, chromatograms are prepared by applying on the thin-layer adsorbent or on the paper in a straight line,
`
`parallel to the edge of the chromatographic plate or paper, solutions of the substance to be identified, the authentic
`
`specimen, and a mixture of nearly equal amounts of the substance to be identified and the authentic specimen. Each
`
`sample application contains approximately the same quantity by weight of material to be chromatographed. If the substance
`
`to be identified and the authentic specimen are identical, all chromatograms agree in color and RR value and the mixed
`
`chromatogram yields a single spot; i.e., RR is 1.0.
`
`Location of Components— The spots produced by paper or thin-layer chromatography may be located by: (1) direct
`
`inspection if the compounds are visible under white or either short-wavelength (254 nm) or long-wavelength (360 nm) UV
`
`light, (2) inspection in white or UV light after treatment with reagents that will make the spots visible (reagents are most
`
`conveniently applied with an atomizer), (3) use of a Geiger-Muller counter or autoradiographic techniques in the case of the
`
`presence of radioactive substances, or (4) evidence resulting from stimulation or inhibition of bacterial growth by the placing
`
`of removed portions of the adsorbent and substance on inoculated media.
`
`In open-column chromatography, in pressurized liquid chromatography performed under conditions of constant flow rate,
`
`and in gas chromatography, the retention time, t, defined as the time elapsed between sample injection and appearance of
`
`the peak concentration of the eluted sample zone, may be used as a parameter of identification. Solutions of the substance
`
`to be identified or derivatives thereof, of the reference compound, and of a mixture of equal amounts of these two are
`
`chromatographed successively on the same column under the same chromatographic conditions. Only one peak should be
`
`observed for the mixture. The ratio of the retention times of the test substance, the reference compound, and a mixture of
`
`these, to the retention time of an internal standard is called the relative retention time RR and is also used frequently as a
`
`parameter of identification.
`
`The deviations of RR, RR, or t values measured for the test substance from the values obtained for the reference compound
`
`and mixture should not exceed the reliability estimates determined statistically from replicate assays of the reference
`
`compound.
`
`Chromatographic identification by these methods under given conditions strongly indicates identity but does not constitute
`
`definitive identification. Coincidence of identity parameters under three to six different sets of chromatographic conditions
`
`(temperatures, column packings, adsorbents, eluants, developing solvents, various chemical derivatives, etc.) increases
`
`the probability that the test and reference substances are identical. However, many isomeric compounds cannot be
`
`separated. Specific and pertinent chemical, spectroscopic, or physicochemical identification of the eluted component
`
`combined with chromatographic identity is the most valid criterion of identification. For this purpose, the individual
`
`components separated by chromatography may be collected for further identification.
`
`PAPER CHROMATOGRAPHY
`
`In paper chromatography the adsorbent is a sheet of paper of suitable texture and thickness. Chromatographic separation
`
`may proceed through the action of a single liquid phase in a process analogous to adsorption chromatography in columns.
`
`Since the natural water content of the paper, or selective imbibition of a hydrophilic component of the liquid phase by the
`
`paper fibers, may be regarded as a stationary phase, a partitioning mechanism may contribute significantly to the
`
`separation.
`
`Alternatively, a two-phase system may be used. The paper is impregnated with one of the phases, which then remains
`
`stationary (usually the more polar phase in the case of unmodified paper). The chromatogram is developed by slow passage
`
`ltlpJ'M'ww.pl1armaoopeia.crIN2924{llusp29nf24s0_c621_viewa||.html
`
`2r'26
`
`

`

`8r‘24r'2015
`
`General Chaaters: <621> CHROMATOGRAPHY
`
`of the other, mobile phase over the sheet. Development may be ascending, in which case the solvent is carried up the
`
`paper by capillary forces, or descending, in which case the solvent flow is also assisted by gravitational force.
`
`Differences in the value of RF have been reported where chromatograms developed in the direction of the paper grain
`
`(machine direction) are compared with others developed at right angles to the grain; therefore, the orientation of paper grain
`
`with respect to solvent flow should be maintained constant in a series of chromatograms. (The machine direction is usually
`
`designated by the manufacturer on packages of chromatography paper.)
`
`In descending chromatography, the mobile phase flows downward on the chromatographic sheet.
`
`Apparatus— The essential equipment for descending chromatography consists of the following:
`
`Descending Chromatography
`
`A vapor-tight chamber provided with inlets for addition of solvent or for releasing internal pressure. The chamber is
`
`constructed preferably of glass, stainless steel, or porcelain and is so designed as to permit obsen/ation of the progress of
`
`the chromatographic run without opening of the chamber. Tall glass cylinders are convenient if they are made vapor-tight
`
`with suitable covers and a sealing compound.
`
`A rack of comosion-resistant material about 5 cm shorter than the inside height of the chamber. The rack serves as a
`
`support for solvent troughs and for antisiphon rods which, in turn, hold up the chromatographic sheets.
`
`One or more glass troughs capable of holding a volume of solvent greater than that needed for one chromatographic run.
`
`The troughs must also be longer than the width of the chromatographic sheets.
`
`Heavy glass antisiphon rods to be supported by the rack and running outside of, parallel to, and slightly above the edge of
`
`the glass trough.
`
`Chromatographic sheets of special filter paper at least 2.5 cm wide and not wider than the length of the troughs are cut to a
`
`length approximately equal to the height of the chamber. A fine pencil line is drawn horizontally across the filter paper at a
`
`distance from one end such that, when the sheet is suspended from the antisiphon rods with the upper end of the paper
`
`resting in the trough and the lower portion hanging free into the chamber, the line is located a few centimeters below the
`
`rods. Care is necessary to avoid contaminating the filter paper by excessive handling or by contact with dirty surfaces.
`
`Procedure— The substance or substances to be analyzed are dissolved in a suitable solvent. Convenient volumes,
`
`delivered from suitable micropipets, of the resulting solution, normally containing 1 to 20 pg of the compound, are placed in
`
`6- to 10-mm spots not less than 3 cm apart along the pencil line. If the total volume to be applied would produce spots of a
`
`diameter greater than 6 to 10 mm, it is applied in separate portions to the same spot, each portion being allowed to dry
`before the next is added.
`
`The spotted chromatographic sheet is suspended in the chamber by use of the antisiphon rod, which holds the upper end of
`
`the sheet in the solvent trough. The bottom of the chamber is covered with the prescribed solvent system. Saturation of the
`
`chamber with solvent vapor is facilitated by lining the inside walls with paper that is wetted with the prescribed solvent
`
`system. It is important to ensure that the portion of the sheet hanging below the rods is freely suspended in the chamber
`
`without touching the rack or the chamber walls or the fluid in the chamber. The chamber is sealed to allow equilibration
`
`(saturation) of the chamber and the paper with the solvent vapor. Any excess pressure is released as necessary. For large
`
`chambers, equilibration overnight may be necessary.
`
`A volume of the mobile phase in excess of the volume required for complete development of the chromatogram is saturated
`
`with the immobile phase by shaking. After equilibration of the chamber, the prepared mobile solvent is introduced into the
`
`trough through the inlet. The inlet is closed and the mobile solvent phase is allowed to travel the desired distance down the
`
`paper. Precautions must be taken against allowing the solvent to run down the sheet when opening the chamber and
`
`removing the chromatogram. The location of the solvent front is quickly marked, and the sheets are dried.
`
`The chromatogram is observed and measured directly or after suitable development to reveal the location of the spots of
`
`the isolated drug or drugs. The paper section(s) predetermined to contain the isolated drug(s) may be cut out and eluted by
`
`an appropriate solvent, and the solutions may be made up to a known volume and quantitatively analyzed by appropriate
`
`ltlpJ'M'ww.pl1armaoopeia.crIN2924{ltusp29nf24s0_c621_viewa||.html
`
`3126
`
`

`

`8l24l2015
`
`General Chaaters: <621> CHROMATOGRAPHY
`
`chemical or instrumental techniques. Similar procedures should be conducted with various amounts of similarly spotted
`
`reference standard on the same paper in the concentration range appropriate to prepare a valid calibration curve.
`
`Ascending Chromatography
`
`In ascending chromatography, the lower edge of the sheet (or strip) is dipped into the mobile phase to permit the mobile
`
`phase to rise on the chromatographic sheet by capillary action.
`
`Apparatus— The essential equipment for ascending chromatography is substantially the same as that described under
`
`Descending Chromatography.
`
`Procedure— The test materials are applied to the chromatographic sheets as directed under Descending Chromatography,
`
`and above the level to which the paper is dipped into the developing solvent. The bottom of the developing chamber is
`
`covered with the developing solvent system. If a two-phase system is used, both phases are added. It is also desirable to
`
`line the walls of the chamber with paper and to saturate this lining with the solvent system. Empty solvent troughs are
`
`placed on the bottom of the chamber, and the chromatographic sheets are suspended so that the end on which the spots
`
`have been added hangs free inside the empty trough.
`
`The chamber is sealed, and equilibration is allowed to proceed as described under Descending Chromatography. Then the
`
`developing solvent (mobile phase) is added through the inlet to the trough in excess of the solvent required for complete
`
`moistening of the chromatographic sheet. The chamber is resealed. When the solvent front has reached the desired height,
`
`the chamber is opened and the sheet is removed and dried.
`
`Quantitative analyses of the spots may be conducted as described under Descending Chromatography.
`
`THIN-LAYER CHROMATOGRAPHY
`
`In thin-layer chromatography, the adsorbent is a relatively thin, unifonn layer of dry, finely powdered material applied to a
`
`glass, plastic, or metal sheet or plate, glass plates being most commonly employed. The coated plate can be considered
`
`an “open chromatographic column” and the separations achieved may be based upon adsorption, partition, or a combination
`
`of both effects, depending on the particular type of stationary phase, its preparation, and its use with different solvents.
`
`Thin-layer chromatography on ion-exchange layers can be used for the fractionation of polar compounds. Presumptive
`
`identification can be effected by obsenration of spots or zones of identical RF value and about equal magnitude obtained,
`
`respectively, with an unknown and a reference sample chromatographed on the same plate. A visual comparison of the size
`
`or intensity of the spots or zones may serve for semiquantitative estimation. Quantitative measurements are possible by
`
`means of densitometry (absorbance or fluorescence measurements), or the spots may be carefully removed from the plate,
`
`followed by elution with a suitable solvent and spectrophotometric measurement. For two-dimensional thin-layer
`
`chromatography, the chromatographed plate is turned at a right angle and again chromatographed, usually in another
`
`chamber equilibrated with a different solvent system.
`
`Apparatus— Acceptable apparatus and materials for thin-layer chromatography consist of the following.
`
`A TLC or HPTLC plate. The chromatography is generally carried out using precoated plates or sheets (on glass, aluminum,
`
`or polyester support) of suitable size. It may be necessary to clean the plates prior to separation. This can be done by
`
`migration of, or immersion in, an appropriate solvent. The plates may also be impregnated by procedures such as
`
`development, immersion, or spraying. At the time of use, the plates may be activated, if necessary, by heating in an oven
`
`at 120': for 20 minutes. The stationary phase of TLC plates has an average particle size of 10-15 pm, and that of HPTLC
`plates an average particle size of 5 pm. Commercial plates with a preadsorbant zone can be used if they are specified in a
`
`monograph. Sample applied to the preabsorbant region develops into sharp, narrow bands at the preabsorbant-sorbent
`
`interface. Alternatively, flat glass plates of convenient size, typically 20 cm X 20 cm can be coated as described under
`
`Preparation of Chromatographic Plates.
`
`A suitable manual, semiautomatic, or automatic application device can be used to ensure proper positioning of the plate
`
`and proper transfer of the sample, with respect to volume and position, onto the plate. Alternatively, a template can be used
`
`to guide in manually placing the test spots at definite intervals, to mark distances as needed, and to aid in labeling the
`
`ltlpJ'M'ww.pl1armaoopeia.crIl\r2924{ltusp29nf24s0_c621_viewa||.html
`
`#26
`
`

`

`8r‘24.i2015
`
`General Chaaters: <621> CHROMATOGRAPHY
`
`plates. For the proper application of the solutions, micropipets, microsyringes, or calibrated disposable capillaries are
`recommended.
`
`For ascending development, a chromatographic chamber made of inert, transparent material and having the following
`
`specifications is used: a flat bottom or twin trough, a tightly fitted lid, and a size suitable for the plates. For horizontal
`
`development, the chamber is provided with a reservoir for the mobile phase, and it also contains a device for directing the
`
`mobile phase to the stationary phase.
`
`Devices for transfer of reagents onto the plate by spraying, immersion, or exposure to vapor and devices to facilitate any
`
`necessary heating for visualization of the separated spots or zones.
`
`A UV light source suitable for observations under short (254 nm) and long (365 nm) wavelength UV light.
`
`A suitable device for documentation of the visualized chromatographic result.
`
`Procedure— Apply the prescribed volume of the test solution and the standard solution in sufficiently small portions to
`
`obtain circular spots of 2 to 5 mm in diameter (1 to 2 mm on HPTLC plates) or bands of 10 to 20 mm by 1 to 2 mm (5 to 10
`
`mm by 0.5 to 1 mm on HPTLC plates) at an appropriate distance from the lower edge—during chromatography the
`
`application position must be 3 mm (HPTLC) to 5 mm (TLC) above the level of the developing solvent—and from the sides
`
`of the plate. Apply the solutions on a line parallel to the lower edge of the plate with an interval of at least 10 mm (5 mm on
`
`HPTLC plates) between the centers of spots or 4 mm (2 mm on HPTLC plates) between the edges of bands, and allow to
`
`dry.
`
`Ascending Development— Line at least one wall of the chromatographic chamber with filter paper. Pour into the
`
`chromatographic chamber a quantity of the mobile phase sufficient for the size of the chamber to give, after impregnation of
`
`the filter paper, a level of depth appropriate to the dimension of the plate used. For saturation of the chromatographic
`
`chamber, close the lid, and allow the system to equilibrate. Unless otherwise indicated, the chromatographic separation is
`
`performed in a saturated chamber.
`
`Place the plate in the chamber, ensuring that the plate is as vertical as possible and that the spots or bands are above the
`
`surface of the mobile phase, and close the chamber. The stationary phase faces the inside of the chamber. Remove the
`
`plate when the mobile phase has moved over the prescribed distance. Dry the plate, and visualize the chromatograms as
`
`prescribed. For two-dimensional chromatography, dry the plates after the first development, and cany out a second
`
`development in a direction perpendicular to that of the first development.
`
`Horizontal Development— Introduce a sufficient quantity of the developing solvent into the reservoir of the chamber using a
`
`syringe or pipet. Place the plate horizontally in the chamber, connect the mobile phase direction device according to the
`
`manufacturer's instructions, and close the chamber. If prescribed, develop the plate starting simultaneously at both ends.
`
`Remove the plate when the mobile phase has moved over the distance prescribed in the monograph. Dry the plate, and
`
`visualize the chromatograms as prescribed.
`
`For two-dimensional chromatography, dry the plates after the first development, and carry out a second development in a
`
`direction perpendicular to that of the first development.
`
`Detection— Observe the dry plate first under short-wavelength UV light (254 nm) and then under long-wavelength UV light
`
`(365 nm) or as stated in the monograph. If further directed, spray, immerse, or expose the plate to vapors of the specified
`
`reagent, heat the plate when required, obsewe, and compare the test chromatogram with the standard chromatogram.
`
`Document the plate after each obsenration. Measure and record the distance of each spot or zone from the point of origin,
`
`and indicate for each spot or zone the wavelength under which it was observed. Detennine the RF values for the principal
`
`spots or zones (see Glossary of Symbols).
`
`Quantitative Measurement— Using appropriate instrumentation, substances separated by TLC and responding to ultraviolet-
`
`visible (UV-Vis) irradiation prior to or after derivatization can be detennined directly on the plate. While moving the plate or
`
`the measuring device, the plate is examined by measuring the reflectance of the incident light. Similarly, fluorescence may
`
`be measured using an appropriate optical system. Substances containing radionuclides can be quantified in three ways: (1)
`
`directly by moving the plate alongside a suitable counter or vice versa; (2) by cutting the plates into strips and measuring
`
`l'r:lp33M-'ww.pl1armaoopeia.crIN2924{llusp29nf24s0_o621_viewa||.html
`
`5l26
`
`

`

`8i24i2015
`
`General Chapters: <a21> CHROMATOGRAPHY
`
`the radioactivity on each individual strip using a suitable counter; or (3) by scraping off the stationary phase, dissolving it in
`
`a suitable scintillation cocktail, and measuring the radioactivity using a liquid scintillation counter (see Radioactivity ( g }
`).
`
`The apparatus for direct quantitative measurement on the plate is a densitometer that is composed of a mechanical device
`
`to move the plate or the measuring device along the x-axis and the y-axis, a recorder, a suitable integrator or a computer;
`
`and, for substances responding to UV-Vis irradiation, a photometer with a source of light, an optical device capable of
`
`generating monochromatic light, and a photo cell of adequate sensitivity, all of which are used for the measurement of
`
`reflectance. In the case where fluorescence is measured, a suitable filter is also required to prevent the light used for
`
`excitation from reaching the photo cell while permitting the emitted light or specific portions thereof to pass. The linearity
`
`range of the counting device must be verified.
`
`For quantitative tests, it is necessary to apply to the plate not fewer than three standard solutions of the substance to be
`
`examined, the concentrations of which span the expected value in the test solution (e.g., 80%, 100%, and 120%).
`
`Derivatize with the prescribed reagent, if necessary, and record the reflectance or fluorescence in the chromatograms
`obtained. Use the measured results for the calculation of the amount of substance in the test solution.
`
`Preparation of Chromatographic Plates-
`
`Apparatus-
`
`Flat glass plates of convenient size, typically 20 cm x 20 cm.
`
`An aligning tray or a flat surface upon which to align and rest the plates during the application of the adsorbent.
`
`A storage rack to hold the prepared plates during drying and transportation. The rack holding the plates should be kept in a
`
`desiccator or be capable of being sealed in order to protect the plates from the environment after removal from the drying
`oven.
`
`The adsorbent consists of finely divided adsorbent materials, normally 5 to 40 pm in diameter, suitable for chromatography.
`
`It can be applied directly to the glass plate or can be bonded to the plate by means of plaster of Paris [calcium sulfate
`
`hemihydrate (at a ratio of 5% to 15%)] or with starch paste or other binders. The plaster of Paris will not yield as hard a
`
`surface as will the starch, but it is not affected by strongly oxidizing spray reagents. The adsorbent may contain fluorescing
`
`material to aid in the visualization of spots that absorb UV light.
`
`A spreader, which, when moved over the glass plate, will apply a uniform layer of adsorbent of desired thickness over the
`
`entire surface of the plate.
`
`Procedure— [NOTE—|n this procedure, use Purified Water that is obtained by distillation.] Clean the glass plates
`
`scrupulously, using an appropriate cleaning solution (see Cleaning Glass Aggaratus ( 1051 ) ), rinsing them with copious
`quantities of water until the water runs off the plates without leaving any visible water or oily spots, then dry. It is important
`
`that the plates be completely free from lint and dust when the adsorbent is applied.
`
`Arrange the plate or plates on the aligning tray, place a 5- x 20-cm plate adjacent to the front edge of the first square plate
`
`and another 5- x 20-cm plate adjacent to the rear edge of the last square, and secure all of the plates so that they will not
`
`slip during the application of the adsorbent. Position the spreader on the end plate opposite the raised end of the aligning
`
`tray. Mix 1 part of adsorbent with 2 parts of water (or in the ratio suggested by the supplier) by shaking vigorously for 30
`
`seconds in a glass-stoppered conical flask, and transfer the slurry to the spreader. Usually 30 g of adsorbent and 60 mL of
`
`water are sufficient for five 20- X 20-cm plates. Complete the application of adsorbents using plaster of Paris binder within 2
`
`minutes of the addition of the water, because thereafter the mixture begins to harden. Draw the spreader smoothly over the
`
`plates toward the raised end of the aligning tray, and remove the spreader when it is on the end plate next to the raised end
`
`of the aligning tray. (Wash away all traces of adsorbent from the spreader immediately after use.) Allow the plates to remain
`
`undisturbed for 5 minutes, then transfer the square plates, layer side up, to the storage rack, and dry at 105° for 30
`minutes. Preferably place the rack at an angle in the drying oven to prevent the condensation of moisture on the back sides
`
`of plates in the rack. When the plates are dry, allow them to cool to room temperature, and inspect the uniformity of the
`
`lt1pJ'M'ww.pl1armaoopeia.crIN2924{Iiusp29nf24s0_c621_viewa||.html
`
`6l26
`
`

`

`8r‘24r'2015
`
`General Chaaters: <621> CHROMATOGRAPHY
`
`distribution and the texture of the adsorbent layer; transmitted light will show uniformity of distribution, and reflected light will
`
`show uniformity of texture. Store the satisfactory plates over silica gel in a suitable chamber.
`
`COLUMN CHROMATOGRAPHY
`
`Apparatus— The apparatus required for column chromatographic procedures is simple, consisting only of the
`
`chromatographic tube itself and a tamping rod, which may be needed to pack a pledget of glass wool or cotton, if needed, in
`
`the base of the tube and compress the adsorbent or slurry uniformly within the tube. In some cases a porous glass disk is
`
`sealed at the base of the tube in order to support the contents. The tube is cylindrical and is made of glass, unless another
`
`material is specified in the individual monograph. A smaller-diameter delivery tube is fused or otherwise attached by a
`
`leakproof joint to the lower end of the main tube. Column dimensions are variable; the dimensions of those commonly used
`
`in pharmaceutical analysis range from 10 to 30 mm in uniform inside diameter and 150 to 400 mm in length, exclusive of
`
`the delivery tube. The delivery tube, usually 3 to 6 mm in inside diameter, may include a stopcock for accurate control of
`
`the flow rate of solvents through the column. The tamping rod, a cylindrical ram firmly attached to a shaft, may be
`
`constructed of plastic, glass, stainless steel, or aluminum, unless another material is specified in the individual monograph.
`
`The shaft of the rod is substantially smaller in diameter than the column and is not less than 5 cm longer than the effective
`
`length of the column. The ram has a diameter about 1 mm smaller than the inside diameter of the column.
`
`Column Adsorption Chromatography
`
`The adsorbent (such as activated alumina or silica gel, calcined diatomaceous silica, or chromatographic purified siliceous
`
`earth) as a dry solid or as a slurry is packed into a glass or quartz chromatographic tube. A solution of the drug in a small
`
`amount of solvent is added to the top of the column and allowed to flow into the adsorbent. The drug principles are
`
`quantitatively removed from the solution and are adsorbed in a narrow transverse band at the top of the column. As
`
`additional solvent is allowed to flow through the column, either by gravity or by application of air pressure, each substance
`
`progresses down the column at a characteristic rate resulting in a spatial separation to give what is known as the
`
`chromatogram. The rate of movement for a given substance is affected by several variables, including the adsorptive power
`
`of the adsorbent and its particle size and surface area; the nature and polarity of the solvent; the hydrostatic head or
`
`applied pressure; and the temperature of the chromatographic system.
`
`If the separated compounds are colored or if they fluoresce under UV light, the adsorbent column may be extruded and, by
`
`transverse cuts, the appropriate segments may then be isolated. The desired compounds are then extracted from each
`
`segment with a suitable solvent. If the compounds are colorless, they may be located by means of painting or spraying the
`
`extruded column with color-forming reagents. Chromatographed radioactive substances may be located by means of
`
`Geiger-Muller detectors or similar sensing and recording instruments. Clear plastic tubing made of a material such as nylon,
`
`which is inert to most solvents and transparent to short-wavelength UV light, may be packed with adsorbent and used as a
`
`chromatographic column. Such a column may be sliced with a sharp knife without removing the packing from the tubing. If
`
`a fluorescent adsorbent is used, the column may be marked under UV light in preparation for slicing.
`
`A "flowing" chromatogram, which is extensively used, is obtained by a procedure in which solvents are allowed to flow
`
`through the column until the separated drug appears in the effluent solution, known as the "eluate.” The drug may be
`
`detennined in the eluate by titration or by a spectrophotometric or colorimetric method, or the solvent may be evaporated,
`
`leaving the drug in more or less pure fonn. If a second drug principle is involved, it is eluted by continuing the first solvent
`
`or by passing a solvent of stronger eluting power through the column. The efficiency of the separation may be checked by
`
`obtaining a thin-layer chromatogram on the individual fractions.
`
`A modified procedure for adding the mixture to the column is sometimes employed. The drug, in a solid form, and, as in the
`
`case of a powdered tablet, without separation from the excipients, is mixed with some of the adsorbent and added to the
`
`top of a column. The subsequent flow of solvent moves the drug down the column in the manner described.
`
`Column Partition Chromatography
`
`In partition chromatography the substances to be separated are partitioned between two immiscible liquids, one of which,
`
`the immobile phase, is adsorbed on a Solid Support, thereby presenting a very large surface area to the flowing solvent or
`
`ltlpJ'M'ww.pl1armaoopeia.crIN2924{llusp29nf24s0_c621_viewa||.html
`
`7126
`
`

`

`8i24l2015
`
`General Chaaters: <621> CHROMATOGRAPHY
`
`mobile phase. The exceed

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket