throbber
Review
`The Progress of Glucose Monitoring—A Review of
`Invasive to Minimally and Non-Invasive Techniques,
`Devices and Sensors
`
`Wilbert Villena Gonzales *
`
`, Ahmed Toaha Mobashsher
`
`and Amin Abbosh
`
`School of Information Technology and Electrical Engineering, The University of Queensland, St Lucia,
`Brisbane 4072, Australia; a.mobashsher@uq.edu.au (A.T.M.); a.abbosh@uq.edu.au (A.A.)
`* Correspondence: w.villena@uq.edu.au; Tel.: +61-07-3365-8354
`
`Received: 31 October 2018; Accepted: 22 January 2019; Published: 15 February 2019
`
`Abstract: Current glucose monitoring methods for the ever-increasing number of diabetic people
`around the world are invasive, painful, time-consuming, and a constant burden for the household
`budget. The non-invasive glucose monitoring technology overcomes these limitations, for which
`this topic is significantly being researched and represents an exciting and highly sought after market
`for many companies. This review aims to offer an up-to-date report on the leading technologies
`for non-invasive (NI) and minimally-invasive (MI) glucose monitoring sensors, devices currently
`available in the market, regulatory framework for accuracy assessment, new approaches currently
`under study by representative groups and developers, and algorithm types for signal enhancement
`and value prediction. The review also discusses the future trend of glucose detection by analyzing the
`usage of the different bands in the electromagnetic spectrum. The review concludes that the adoption
`and use of new technologies for glucose detection is unavoidable and closer to become a reality.
`
`Keywords: glucose; non-invasive; minimally-invasive; spectroscopy; continuous monitoring; MARD;
`FDA; ISO 15197; plasmon resonance; fluorescence; ultrasound; metabolic heat conformation
`
`1. Introduction
`
`Diabetes is a disease that results from abnormal levels of insulin in the body, due to either
`a malfunction of the pancreas not producing enough insulin or the cells in the body not using it
`adequately [1]. Insulin is a hormone that regulates the level of glucose by allowing cells to absorb
`it from the bloodstream to obtain energy or store it for future use. However, if the level of glucose
`in the blood remains very low or very high for long periods of time, it could cause hypoglycemia or
`hyperglycemia, respectively, leading to severe medical conditions, including tissue damage, stroke,
`kidney failure, blindness and heart disease, among others, and finally, death if left untreated [2].
`Deficient production of insulin in the pancreas leads to diabetes type 1, characterized by the sudden
`drop of glucose levels. On the other hand, ineffective use of insulin leads to diabetes type 2, which is
`characterized by high levels of glucose. Both conditions do not have a cure, meaning that regular
`glucose monitoring in diabetic people is necessary for the rest of their lives.
`Unfortunately, the issue of regularly checking the blood glucose for most diabetic people is
`not very pleasant. Conventional devices for glucose monitoring use the electrochemical method [3],
`which requires a small amount of blood to be drawn out of the body by either finger-pricking or a
`thin lancelet implanted subcutaneously. The difference between both is that the first only provides
`a snapshot of the glucose level at one specific point in time and does not require professional
`assistance, so it is called self-monitoring blood glucose (SMBG) monitoring device. The second
`provides continuous monitoring, and thus it is called continuous-glucose-monitoring device (CGM).
`
`Sensors 2019, 19, 800; doi:10.3390/s19040800
`
`www.mdpi.com/journal/sensors
`
`sensors
`
`# & !*-
`0+/ .
`
`ABBOTT Exhibit 2013
`Dexcom v. Abbott Diabetes Care, IPR2023-01409
`
`

`

`Sensors 2019, 19, 800
`
`2 of 45
`
`However, both of them not only cause discomfort and pain after repeated use but also pose risks
`of potential infection and tissue damage [4], resulting in poor patient compliance for daily assigned
`measurements [5]. Consequently, since the end of last century, there has been a continuous effort
`for developing non-invasive (NI) devices, i.e., no need of bloodletting, and minimally-invasive (MI),
`aimed at reducing some of the issues connected with the traditional invasive methods.
`The development of a genuinely non-invasive device for glucose measurement would represent a
`life-changing factor for millions of patients around the world, allowing them to monitor their glucose
`level confidently and receiving quick treatment if necessary. It also represents a vast potential market.
`According to the World Health Organization (WHO), currently there are around 450 million cases of
`diabetes in the world, and the number could potentially reach 700 million by 2045 [6], with an increase
`to 39.7 million by 2030 and 60.6 million in 2060 in the United States alone [7].
`Current developments try to exploit the characteristics of the glucose molecule at different
`frequencies in the spectrum, from DC and ultrasound, all the way to the near-infrared (NIR) and
`visible regions. However, it is in these last two where most of the promising technologies have emerged
`and even been used in the development of some commercial devices. Many are no longer existent due
`to low accuracy, selectivity and sensitivity of the measurement [8], whereas those already available,
`still have not reached accuracies comparable to the traditional methods. This situation leaves the
`issue of NI glucose monitoring still open to many possibilities, including the combination of several
`techniques, which could finally lead to the development of a reliable and cost-effective device for
`glucose monitoring.
`Many prominent publications have already reviewed several NI glucose technologies and devices,
`some of which are mentioned hereafter. Chen et al. for example provide a comprehensive description
`of the current state of MI and NI technologies for CGM analysis [9]. Lin et al. reviewed not only some of
`the past and current NI devices, but also discuss the main challenges associated with NI detection [10].
`Van Enter and von Hauff reviewed the physical and chemical properties of the glucose molecule and
`analyze their effect on the accuracy and effectiveness of NI technologies [11]. Uwadaira and Ikehata
`not only provided a comprehensive list of technologies employed for non-invasive glucose detection
`but also summarize their main advantages and limitations. Khalil provided an excellent account on
`the properties and characteristics of the glucose molecule and tissue at different NIR wavelengths,
`and then compares and analyzes the accuracy and sensitivity of glucose measurements in in-vitro,
`ex-vivo and in-vivo samples [12].
`It is upon such previous works that the present review is limited to the topics shown in Figure 1
`to provide an update on the technologies behind the current development of sensors for minimal
`and noninvasive glucose monitoring and to expand our understanding of how current regulatory
`framework and data processing algorithms shape such development. Sections 2 and 3 describe the
`currently accepted methods, standards and regulatory bodies to understand the way they have been
`shaping the evolution of non-invasive glucose detection. Section 4 provides basic descriptions of
`past and current leading technologies and associated instrumentation, putting particular emphasis on
`the most recent and promising technologies, from our perspective. Section 5 lists currently existent
`devices and those with high potential of coming out to the market shortly (NI and MI), along with
`the technology and main characteristics associated with each of them. Section 6 lists the techniques
`currently under research and development in different universities and research institutions. Then,
`Section 7 provides a brief view of the different types of algorithms used not only to improve the
`quality of the detected signals but also to predict future values for better treatment of diabetic patients.
`Section 8 brings together all the technologies, methods, devices and research described in the previous
`sections, shows them in a graphical chart and discusses the current progress and prospects for further
`development. Finally, the conclusions presented in Section 9 connect the previous sections and provide
`a further view into the future development of glucose monitoring.
`
`ABBOTT Exhibit 2013
`Dexcom v. Abbott Diabetes Care, IPR2023-01409
`
`

`

`Sensors 2019, 19, 800
`
`3 of 45
`
`2. Conventional Methods for Glucose Monitoring
`
`Figure 1. Structure of the paper.
`
`Glucose concentration can be determined using either whole blood, plasma, or serum samples,
`although the last two are preferred because readings from whole blood are usually 15% lower due
`to the additional water content in the blood cells [13]. As such, standard methods require a certain
`amount of blood, meaning they are invasive.
`Initially, glucose measurement could be performed only in laboratories by taking advantage of the
`reducing and condensation properties of glucose, but issues associated with non-specificity, toxicity
`and cross-reaction with other agents quickly phased them out from clinical practice [13]. Hence,
`present techniques rely on enzymatic and hexokinase methods. Both present high degrees of accuracy,
`specificity and minimum cross-reaction, but while laboratories use both of them, point-of-care and
`home monitoring prefer the enzymatic method due to its simplicity and relative affordability.
`
`2.1. Laboratory Techniques
`
`Enzymatic-amperometric and hexokinase are the preferred methods for measuring blood glucose
`concentrations at laboratories. Table 1 shows some of the equipment based on such methods.
`
`Table 1. Representative equipment used for accurate glucose measurement in the laboratory.
`
`Method
`
`Equipment
`
`YSI 2700
`
`YSI 2950D
`
`Enzymatic
`
`Biosen C-Line/S-Line
`
`Beckman Coulter DxC 800
`Abbott ARCHITECT
`c8000/c16000
`Hitachi 917
`Cobas c 701/702
`
`Hexokinase
`
`Intended Use
`Laboratory
`Point-
`of-care
`Laboratory
`
`Laboratory
`Point-of-care
`
`Laboratory
`
`Blood
`Plasma
`Serum
`CSF
`
`Sample Type
`•
`•
`•
`•
`•
`•
`•
`
`Blood
`Plasma
`Serum
`
`•
`•
`•
`•
`
`Plasma
`Serum
`Urine
`CSF
`
`Range
`
`0–2500 mg/dL
`
`5–2500 mg/dL
`
`Ref.
`
`[14]
`
`[15]
`
`9–900 mg/dL
`
`[16]
`
`5–700 mg/dL
`
`1–800 mg/dL
`2–750 mg/dL
`2–750 mg/dL
`
`[17]
`
`[18]
`[18]
`[19]
`
`
`
`ABBOTT Exhibit 2013
`Dexcom v. Abbott Diabetes Care, IPR2023-01409
`
`

`

`Sensors 2019, 19, 800
`
`4 of 45
`
`All of them possess a high level of specificity, sensitivity and can detect a broad range of glucose
`concentrations, including under hypoglycemic and hyperglycemic scenarios. Such characteristics
`mean they can be used as the standard of reference to measure the performance of less accurate
`instruments, i.e., SMBG, CGM and future non-invasive and minimally-invasive devices. Besides, it is
`important to point out that most of the laboratory equipment can detect and measure other sugars,
`and chemical compounds, including lactose, methanol and hydrogen peroxide by using other reagents
`or combined methods, however, this is not discussed in this publication.
`
`2.1.1. Enzymatic-Amperometric Method
`
`Considering that the enzyme glucose oxidase (GOx) is specific to glucose, in this method,
`the oxidation of glucose takes place in the presence of GOx, oxygen (O2) and water (H2O) to form
`gluconic acid and hydrogen peroxide (H2O2). The hydrogen peroxide is then electrochemically
`oxidised at the anode of an electrochemical probe, producing an amperometric signal (current)
`proportional to the concentration of glucose in the sample (Figure 2).
`
`Figure 2. Enzymatic-amperometric method for measurement of glucose concentration in-vitro.
`
`A popular glucose analyser using such technology is the blood gas analyser, which contains a
`solution of GOx between the gas permeable membrane of a pO2 electrode and an outer semipermeable
`membrane. Through diffusion, the glucose crosses the semipermeable membrane and reacts with GOx.
`Once the hydrogen peroxide is oxidised, the reaction consumes the oxygen near the surface of the pO2
`electrode, then the consumption rate is measured. The loss of electrons and the rate of decrease in pO2
`is directly proportional to the concentration of glucose in the sample [20].
`
`2.1.2. Hexokinase Method
`
`Hexokinase method, also known as photometric method, consists of a series of chemical reactions,
`as shown in Figure 3. In the first stage, the glucose reacts with the enzyme hexokinase, in the
`presence of adenosine triphosphate (ATP) and magnesium ions, to produce glucose-6-phosphate
`(G6P) and adenosine diphosphate (ADP). In the second stage, G6P and nicotinamide adenine
`dinucleotide (NAD) go through oxidation with glucose-6-phosphate dehydrogenase until being
`reduced to 6-phosphogluconate and nicotinamide-adenine-dinucleotide-reduced (NADH), respectively.
`The amount of NADH is proportional to the amount of glucose in the sample, and it has the property
`of absorbing light at 340 nm. The amount of absorption is proportional to the amount of NADH,
`meaning that glucose can be measured using standard spectrophotometric techniques [21].
`
`glucose
`
`GOx
`
`Gluconic acid
`
`H2O + O2
`2H+
`
`H2O2
`
`2e-
`
`Electrode
`(anode)
`
`
`
`ABBOTT Exhibit 2013
`Dexcom v. Abbott Diabetes Care, IPR2023-01409
`
`

`

`Sensors 2019, 19, 800
`
`5 of 45
`
`Figure 3. Hexokinase method for measurement of glucose concentration in-vitro.
`
`2.1.3. Clinical Significance
`
`Both methods are highly specific, accurate and sensitive. As such, and depending on the specific
`technology developed by each manufacturer, some models are used as reference gold-standards for
`calibration of other glucose meters and at central laboratories. Also, given the small size of some
`models, it is possible to use them in a point-of-care environment, meaning that medical staff can take
`quick analyses of ill patients, especially in emergency and intensive care units (ICU).
`The main disadvantages associated with laboratory methods is the inherent invasiveness since all
`these methods need to be done in-vitro, i.e., blood samples taken from patients; the need of trained
`laboratory personnel, leading to additional costs; and extended waiting periods of time until receiving
`laboratory results. Besides, not all laboratory equipment is highly accurate, as revealed by a recent
`study of Liang et al., showing some blood gas analyzers not complying with the new requirements set
`by the FDA 2014 draft and 2016 final guidance [22], or not providing accurate readings in hypoglycemia
`cases, especially in patients with unstable hemodynamics [23].
`
`2.2. Home-Monitoring Techniques
`
`There are two types of devices intended for personal use and self-assessment: Non-continuous
`monitoring (NCGM), and continuous glucose monitoring (CGM). As the name implies, NCGM devices
`(commonly known as self-monitoring blood glucose SMBG devices) are used to monitor glucose levels
`only at specific points during the day, with a frequency dependent on diabetes type, diet, medication
`dosage and clinical condition of the person. On the other hand, CGM devices can monitor glucose
`levels every few minutes automatically, making possible to monitor rapid changes and trends missed
`by SMBG testing. Nevertheless, the accuracy and reliability of both systems are suitable in point-of-care
`and self-assessment situations.
`
`2.2.1. Self-Monitoring Blood Glucose—SMBG
`
`SMBG devices are the typical glucometers requiring finger pricking with a lancet to access the
`capillary blood. The glucose measurement method is then fundamentally the same electrochemical
`technique previously described. The main difference, however, is that the complete reaction and
`detection takes place in a glucose test strip connected to a meter. After putting a drop of the blood
`sample on the test strip, the glucose oxidizes in the presence of an enzyme to produce a certain
`amount of current proportional to the glucose level. The electrons then travel to the meter containing a
`current-to-voltage converter to provide a voltage proportional to the level of glucose.
`The test strip contains the enzyme and an arrangement of three electrodes (Figure 4): the working
`electrode, which senses the actual current of the reaction; the reference electrode, holding a voltage
`always constant respect to the working electrode to aid with the chemical reaction; and the
`counter electrode, supplying the current to the working electrode [24]. However, new designs
`only need the working and reference electrodes. Also, depending on the model, some devices use
`glucose oxidase GOx as the enzyme, while others use glucose dehydrogenase (GDH) attached to a
`coenzyme, pyrroloquinoline-quinone PQQ or flavin adenine dinucleotide FDA, although inaccuracy
`
`glucose
`
`Hexokinase
`
`G6P
`
`Hexokinase
`
`6-phosphogluconate
`
`340 nm
`
`Wavelength
`
`
`
`Absorption
`
`ATP
`
`ADP
`
`NAD
`
`NADH
`
`ABBOTT Exhibit 2013
`Dexcom v. Abbott Diabetes Care, IPR2023-01409
`
`

`

`Sensors 2019, 19, 800
`
`6 of 45
`
`and specificity issues with GDH-PQQ, due to interference with other sugars, is well known, as reported
`by several studies [25–28].
`
`Figure 4. Block diagram of the device for glucose measurement with finger-pricking method.
`
`2.2.2. Continuous Glucose Monitoring—CGM
`
`CGM devices consist of three essential parts: a wireless receiver, a transmitter and a sensor.
`The receiver has a monitor displaying the glucose reading. The transmitter is attached to the sensor
`and transmits the measurements to the receiver via RF waves. The sensor is a tiny sensing device
`inserted into the subcutaneous tissue, extending just far enough to get access to the interstitial fluid
`(ISF). Then, by the electrochemical technique, the sensor uses GOx to oxidize the glucose present in the
`ISF, just as a test strip in SMBG devices does. The resulting peroxide reacts with platinum, to produce
`the electrical current, which travels along a thin wire to the transmitter, located outside of the skin.
`Once the receiver gets the data from the transmitter, it processes the information and calculates the
`glucose level.
`
`2.2.3. Clinical Significance
`
`Even though the apparent advantage of CGM devices is the ability to measure glucose levels
`continuously, they cannot be considered as the best option for blood glucose monitoring, since they still
`need calibration at least twice a day with the standard finger-pricking method. Besides, CGM devices
`measure glucose from the ISF, implying the existence of a lag-time between 6 and 12 min [29], meaning
`that the ISF readings are not a reflection of the actual glucose level in the blood. Additionally, there are
`issues associated with inaccurate readings due to mishandling and poor self-monitoring technique,
`problems related with the insertion of the sensor under the skin, skin irritation and discomfort when
`securing the device to the body [30–32].
`Finger-pricking SMBG devices are currently the most reliable and accurate devices for
`self-monitoring due to the relative simplicity of the measuring procedure, and their reliance on
`capillary blood to obtain accurate glucose readings [31]. Unfortunately, the pain and discomfort caused
`by regular finger-pricking several times per day, and the costs associated with the constant purchase of
`test strips, deter many patients from checking their glucose level regularly, as proven by some studies,
`including one performed by Ward et al in 2015, showing that 50% are willing to measure their glucose
`levels “occasionally, as needed” [33].
`
`Working electrode
`coated with enzyme
`
`Reference electrode
`
`Counter electrode
`
`Test strip
`
`VREF
`
`Current-to-voltage
`converter
`
`I
`
`VOUT
`
`Meter
`
`
`
`ABBOTT Exhibit 2013
`Dexcom v. Abbott Diabetes Care, IPR2023-01409
`
`

`

`Sensors 2019, 19, 800
`
`7 of 45
`
`2.3. Laboratory Techniques vs Home-Monitoring Techniques
`
`As shown in Table 2, despite the accuracy of both techniques, laboratory and home-monitoring,
`the invasiveness and time associated are their main limitations. These, not only cause discomfort on
`patients but also represent potential risks on late diagnostics and sample contamination. However,
`their level of accuracy and sensitivity still make them the most trusted options for glucose monitoring.
`
`Table 2. Comparison between laboratory and self-monitoring techniques.
`
`Characteristics
`Accuracy
`Sensitivity
`Measurement time
`Trained laboratory personnel
`Sample type
`Blood extraction method
`
`Laboratory
`Very good
`Very good
`Long
`Yes
`Blood, serum, plasma, urine
`Invasive
`
`Self-Monitoring
`Good
`Good
`Quick
`No
`Blood, ISF
`Invasive
`
`Laboratory methods are the most accurate and sensitive. Hence they are used as the reference
`technique to calibrate other devices. Home-monitoring devices are not as accurate as their laboratory
`counterpart, but still, provide results quickly and accurate enough for personal and point-of-care uses.
`
`3. Accuracy Assessment Tools and Standards
`
`In order to test the accuracy and effectiveness of glucose monitoring devices, there is a set of tools,
`guidelines and standards. The mean-average-relative-measurement (MARD) and the error grids are
`metric measurements to evaluate accuracy. The standard ISO 15197 provides the quality guidelines,
`requirements and specifications that glucose measuring devices should comply with to guarantee their
`suitability for human use. As such, many countries around the world use ISO’s guidelines, through
`their national agencies, to assess whether each device is suitable for commercialization in their territory
`or not. Nevertheless, there are exceptions like the United States that has its own set of assessment
`guidelines. Knowing the metrics and standards is essential to understand the reason why developers
`and researchers are focusing on certain technologies while leaving behind others, as well as the level
`of accuracy they intend to achieve.
`
`3.1. Mean Absolute Relative Difference—MARD
`
`MARD is currently the most widely accepted metric measurement to evaluate the performance
`and accuracy of glucose detection devices, including CGM, SBGM, MI and NI devices, due to its
`simplicity [34]. Its calculation is quite straightforward, as it consists of taking the average of all the
`absolute errors between the measured points and those set as the reference. As a result, MARD consists
`of a single number, expressed as a percentage, which represents the closeness of the measured data to
`the real value. A small number indicates the capability of the device to take accurate measurements,
`while large numbers are an indication of considerable inaccuracies.
`The standard way to calculate the MARD is by using two sets of data, taken at the same time,
`during clinical trials. One set of data is the blood glucose concentration measured by the device under
`test, while a standard laboratory method provides the second data set (e.g., YSI-2700). Then both
`measurements are compared [35].
`Unfortunately, the value of MARD is heavily dependent on the characteristics and details of the
`study. Therefore, comparing the MARD of different devices may lead to misinterpretations [36]. As a
`result, MARD should not be taken blindly as an absolute indicator of accuracy, but rather, as stated by
`Reiterer et al. [35], as an “indication with some uncertainty”.
`
`ABBOTT Exhibit 2013
`Dexcom v. Abbott Diabetes Care, IPR2023-01409
`
`

`

`Sensors 2019, 19, 800
`
`8 of 45
`
`3.2. Error Grids–Clarke, Parkes and Surveillance
`
`Error grids evaluate the clinical accuracy of glucose measuring devices. They provide a qualitative
`approach by describing the clinical outcome of basing a treatment decision on the result of the
`measurement method under evaluation [37]. They consist of a two-dimension grid, divided in a set
`of ‘risk’ zones’, where results from both, the glucose measuring device and the reference method, are
`plotted against each other. By analyzing the distribution of paired data points in the grid, it is possible
`to determine the percentage of points contained in each zone, permitting to categorize each device
`according to the degree of risk that an adverse measurement would represent due to an inaccurate
`measurement of the glucose level.
`Clarke error grid (CEG), Parkes or Consensus error grids (PEG) for diabetes types 1 and 2,
`and Surveillance error grid (SEG) are currently the four main types of error grids, each of them divided
`into five distribution risk zones identified with letters A to E (CEG and PEG) or a color-coded pattern
`(SEG). The first one to appear was CEG, but limitations on its assessment method gave way to the
`development of the PEG which comprises two types of grids, one for each diabetes type, given the
`higher tolerance of type 2 patients to larger margins of error in the accuracy of the reading than type
`1. However, with new regulatory ISO and FDA guidelines, the clinical community becoming more
`aware of the severe consequences of inaccurate readings [38], and traditional out-of-date medical
`practices, CEG and PEG are falling out of use, in addition to their inability of identifying clinical states
`in which tight glycemic control is necessary [39]. As such, in 2014 authors from academia, industry
`and regulatory agencies introduced the Surveillance Error Grid. Contrary to CEG and PEG, SEG uses
`different colors from green, for no risk, to red indicating extreme risk of hypo or hyperglycaemia.
`Table 3 summarizes the meaning of each zone from a clinical accuracy point of view. In general
`lines, zones A, B and Green represent accurate or acceptable glucose results, while the ones in the
`opposite end represent potential dangerous situations if not taking appropriate corrective measures.
`
`Table 3. Standard error grids for clinical accuracy assessment of glucose detection.
`
`Clarke Error Grid
`[40]
`A to E
`
`Parkes Error Grid
`Parkes Error Grid
`Type 2 Diabetes [41]
`Type 1 Diabetes [41]
`A to E
`
`Surveillance Error
`Grid [38]
`Green to Dark-red
`
`Clinically correct
`decisions
`
`No effect on clinical action
`
`No risk
`
`Clinically uncritical
`decisions
`
`Altered clinical action or little or no effect on
`clinical outcome
`
`Mild risk
`
`Overcorrections
`that could lead to a
`poor outcome
`
`Altered clinical action: likely to affect clinical
`outcome
`
`Moderate risk
`
`Dangerous failure
`to detect and treat
`
`Altered clinical action: potential significant
`medical risk
`
`Erroneous
`treatment
`
`Altered clinical action: potential dangerous
`consequences.
`
`High risk
`
`Extreme risk
`
`Risk zones
`A
`
`Green
`B
`
`G/Y
`C
`
`Y/R
`D
`
`Red
`E
`
`Dark Red
`
`3.3. ISO 15197 Standard
`
`The International Standards Organization (ISO) is an independent and non-governmental
`organization that defines and develops specifications for procedures, services and production of
`high quality, reliable and safe products in a wide range of industries, including medical devices,
`food safety, environmental management and Information technology among others [42]. Currently,
`ISO comprises 162 national standard bodies of high technical and expertise levels, and influences
`regulations of several government agencies worldwide.
`
`
`
`Green
`
`
`
`G/Y
`
`
`
`Y/R
`
`
`
`Red
`
`
`
`Dark
`
`ABBOTT Exhibit 2013
`Dexcom v. Abbott Diabetes Care, IPR2023-01409
`
`

`

`Sensors 2019, 19, 800
`
`9 of 45
`
`ISO 15197:2013 is the newest standard, released in 2013, for glucose monitoring devices and
`systems for self-testing. Compared to its ancestor, ISO 15197:2003, the new standard has tighter
`accuracy requirements that new devices will have to follow. Nevertheless, adhering to the new
`guidelines will provide greater confidence to patients and clinicians that glucose readings are reliable
`and sufficiently accurate on a day-to-day basis.
`The new standard requires that, compared to a reference laboratory method, 95% of the blood
`glucose results have to be within ±15 mg/dL for glucose concentrations less than 100 mg/dL or ±15%
`at glucose concentrations of 100 mg/dL or more. Additionally, 99% of the readings have to be inside
`of zones A and B of the Parkes (Consensus) Error Grid for diabetes type 1 [43].
`In 2015, ISO released a harmonized version called EN ISO 15197:2015 to be used by the European
`Union. This version, however, did not introduce any change to the requirements for the performance
`evaluation of glucose meters [44,45].
`
`3.4. Approval Agencies
`
`Some agencies have their guidelines for the approval of medical devices in their own countries,
`while others, such as the European Medicines Agency follow the guidelines given by ISO 15197:2013
`(devices fulfilling the ISO requirements can get the CE mark) [46]. However, currently there are no
`specific standards for non-invasive glucose monitors (NIG), as such, manufacturers of NIG devices
`follow the general guidelines, created for invasive methods, to design their devices and comply with
`national regulations.
`Table 4 summarizes the evaluation criteria for the acceptance of glucose monitoring devices in
`certain countries. In the case of countries following the ISO standard, it is important to mention that
`ISO 15197:2013 is the new standard that new products should comply with if they are released in
`territories already using the 2013 version. Complying with the requirements from the 2003 version is
`still accepted in many places.
`
`Table 4. Guidelines for approval of glucose monitoring devices in some countries.
`
`Agency
`
`Country Guideline/StandardRelease
`Year
`
`Device Type
`
`UCM 380325
`
`UCM 380327
`
`EN ISO
`15197
`
`2016
`
`2015
`
`BGMS
`
`SMBG
`
`ISO 15197
`
`2013
`
`BGMS/SMBG
`
`Food & Drug
`Administration
`(FDA)
`
`European
`Medicines
`Agency (EMA)
`
`Health Canada
`
`Agência
`Nacional de
`Vigilância
`Sanitária
`(ANVISA)
`China Food &
`Drug
`Administration
`(CFDA)
`Pharmaceuticals
`and Medical
`Devices Agency
`(PMDA)
`
`Therapeutic
`Goods
`Administration
`TGA
`
`USA
`[47,48]
`
`EU [49]
`
`Canada
`[50]
`
`Brazil
`[51]
`
`China
`[52]
`
`Japan
`[53,54]
`
`Australia
`[55–57]
`
`Glucose
`Concentration
`≥75 mg/dL
`
`<75 md/dL
`
`Entire range
`≥100
`mg/dL
`
`Criteria
`95% within ±12%
`98% within ±15%
`95% within ±12
`mg/dL
`98% within ±15
`mg/dL
`95% within ±15%
`99% within ±20%
`95% within ±15%
`
`<100 mg/dL
`
`95% within ±15
`mg/dL
`
`Entire range
`(Type 1
`Diabetes)
`
`99% within Zones A
`& B of Parkes Error
`Grid
`
`ABBOTT Exhibit 2013
`Dexcom v. Abbott Diabetes Care, IPR2023-01409
`
`

`

`Sensors 2019, 19, 800
`
`10 of 45
`
`4. Minimally-Invasive and Non-Invasive Technologies
`
`Technologies for glucose detection without the invasiveness, pain, discomfort and risks associated
`with standard methods, have been the focus of intensive research. Thus, we can classify them in two
`major groups: minimally-invasive (MI) and non-invasive (NI). MI technologies are those that need
`to extract some form of fluid from the body (ex. tears and interstitial fluid) to measure the glucose
`concentration through an enzymatic reaction. NI technologies rely solely on some form of radiation
`without the need of accessing to any body fluid.
`Likewise, technologies for glucose detection can be classified in four sub-groups: Optical, thermal,
`electrical and nanotechnology methods (see Figure 5). Optical, in a broad sense, comprise all the
`techniques developed to work in the infrared and optical bands of the spectrum, since they take
`advantage of the reflection, absorption and scattering properties of light when passing through
`biological media. Thermal methods monitor glucose by detecting physiologic indices related to
`metabolic heat generation proper of the glucose molecule, as such, they work in the far-infrared
`band. Electric methods take advantage of the dielectric properties of glucose at low frequencies using
`small amounts of electromagnetic radiation, current and ultrasound. Finally, there is the new field of
`nanotechnology. Currently, only two techniques have started exploring such new venue extensively
`(SPR and fluorescence), in combination with optical techniques. However, there are several other
`potential techniques that can be developed, such as carbon nanotubes and plasmonics [58–62], but they
`are still in a very early stage of development, with most of the progress happening in the theoretical
`side. As such the authors will not consider them in the present review. Nevertheless, it is important to
`note that regardless of the type of technology, they all aim at minimizing the influence of physiological
`variability and various environmental conditions.
`
`Figure 5. Technologies under development for minimally-invasive and non-invasive glucose detection
`(SPR-surface plasmon resonance, OP-optical polarimetry, OCT-optical coherence tomography, TOF-time
`of flight, THz-TDS-Terahertz time domain spectroscopy, TES-thermal spectroscopy, MHC-metabolic
`heat conformation, PAS-photo-acoustic spectroscopy, mmW-millimeter wave, µm-Microwave,
`EMS–Electromagnetic sensing, BS-Bioimpedance spectroscopy).
`
`
`
`ABBOTT Exhibit 2013
`Dexcom v. Abbott Diabetes Care, IPR2023-01409
`
`

`

`Sensors 2019, 19, 800
`
`11 of 45
`
`4.1. Surface Plasmon Resonance (SPR)
`
`Surface plasmon resonance is the point at which the collective coherent charge-density waves,
`called surface plasmon polaritons (SPPs), are excited by an electr

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket