`Design of pulse oximeters
`
`4.2.3 He,noglobin absorbance spectra
`
`The chemical binding of the different hemoglobin species changes the physical
`properties of the hemoglobin as well. Figure 4.2 shows the extinction coefficients
`of oxyhemoglobin, reduced hemoglobin, methemoglobin and carboxyhemoglobin
`at wavelengths in the range of interest in pulse oximetry.
`The absorbance of light in the red region of the spectrum is much higher for
`reduced hemoglobin than for oxyhemoglobin. The extinction coefficients of both
`hemoglobin species are equal at the point isosbestic point (805 nm). The reduced
`hemoglobin is more transparent to light from the infrared region than
`oxyhemoglobin.
`The extinction coefficient of carboxyhemoglobin is about the same as that of
`oxyhemoglobin at the wavelength of 660 nm while it is almost transparent in the
`infrared region. Methemoglobin absorbs much light in the red region of the
`spectrum and its extinction coefficient remains higher than that of oxyhemoglobin
`in the infrared region.
`
`mothemoolol,In
`
`087.0.0910'Dil
`
`\,iduced
`\Dern<glothn
`
`to.
`
`C
`11
`
`0 0
`
`0C0
`
`2
`XLU
`
`01,
`600
`
`840
`
`1
`680
`
`720
`
`1
`760
`800
`840
`Wavelength (nm)
`
`C[rbo/'llemoolobin
`1
`Bao
`960
`920
`
`1000
`
`Figure 4.2 Extinction coefficients of the four most common hemoglobin species oxyhemoglobin,
`reduced hemoglobin, carboxyhemoglobin, and methemoglobin at the wavelengths of interest in
`pulse oximetry (courtesy of Susan Manson, Biox/Ohmeda, Boulder, CO).
`
`4.3 BEER'S LAW IN PULSE OXIMETRY
`
`Pulse oximeters determine the oxygen saturation of arterial blood by measuring
`the light absorbance of living tissue at two different wavelengths and using the
`arterial pulsation to differentiate between absorbance of arterial blood and other
`absorbers.
`
`61
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0062
`
`APL_MAS_ITC_00015679
`
`Light absorbance in pidse oximetry
`
`45
`
`4.3.1 Criteria for the choice of wavelengths
`Different reasons lead to the most common choice for wavelengths used in pulse
`oximetry. The red skin pigmentation absorbs a great amount of light at
`wavelengths shorter than 600 nm and therefore it is not desirable to measure
`light absorbance in this range. Large differences in the extinction coefficients of
`reduced hemoglobin and oxygenated hemoglobin change the absorbance of ] ight
`significantly, even when the oxygen saturation changes slightly. A good choice
`for a wavelength in the red region is 660 nm because of a large difference in the
`extinction coefficients.
`Another issue for the wavelength choice is flatness of the absorption spectra
`shown in figure 4.2 around the chosen wavelength. Otherwise shifts in the peak
`wavelength of the LEDs (see section 5.3) will result in a larger error. The
`absorbance spectra of reduced hemoglobin and oxygenated hemoglobin are
`relatively flat at 660 and 940 nm (Moyle 1994).
`Mannheimer et al (1997) have shown that sensors fabricated with 735 and
`890 nm emitters read more accurately at low saturations under a variety of
`conditions, while 660 and 990 nm emitters read more accurately at high
`saturations.
`
`4.3.2 Absorbance in hemoglobin solutions
`
`The different species of hemoglobin are the main light absorbers in arterial and
`venous blood. Most of the hemoglobin in human blood is either oxygenated or
`reduced hemoglobin which determine the functional oxygen saturalion SO2
`(equation (4.5)). The concentrations of oxygenated hemoglobin (cilbO2) and
`reduced hemoglobin (cHb) can be expressed as a function of SO2 as a fraction and
`the sum of the concentrations CHb02 and cHb
`cHb02 = SO2~CHbO2 + CIIb )
`CHb - (1 - SC)2)( CHb02 + CHb).
`
`(4.8)
`(4.9)
`According to Beer's law we derive the total absorbance At of a solution
`containing only reduced and oxygenated hemoglobin as absorbing substances
`from equation (4.4)
`
`At =81.-Ib02(A)CHbo: d}lb02 +EHb(A)cHbdHh-
`
`(4.10)
`
`Assuming that the optical path length d is the same for the oxygenated
`hemoglobin (411)02) and reduced hemoglobin (dilb) and using equations (4.8),
`(4.9), and (4.10), we derive
`
`At = ~EHI:>02 (A)SC)2 + £Ht,(A)(1- S02) ~(CHI:> +CHb02 )d. (4.11)
`
`Thus At can be expressed for known concentrations of hemoglobin in terms
`of functional oxygen saturation as a fraction, the extinction coefficients of
`hemoglobin, and the length of the optical path. Values for the extinction
`coefficients of adult reduced hemoglobin (4[b) and adult oxygenated hemoglobin
`
`62
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0063
`
`APL_MAS_ITC_00015680
`
`46
`
`Design of pulse oximeters
`
`(€HbO,) at the two wavelengths most commonly used in pulse oximetry (660 nm
`and 940 nm) have been measured by ZOIstra et al (1991) (see table 4.1).
`
`Table 4.1 Table of extinction coefficients of reduced and oxygenated hemoglobin in adults at the
`wavelengths of 660 nm and 940 nm (values from Zijlstra etal !99]).
`
`Wavelength, nm
`660
`940
`
`Extinction coefficient. L mmol-' cm-1
`Hb
`Hb02
`0.08
`0.81
`0.18
`0.29
`
`Figure 4.3 shows the characteristics of light absorbance for a sample with a
`fixed concentration of total functional hemoglobin (CHboa + CHb) of 1 mmol L-1,
`a fixed path lenglh d of 1 cm and varying functional oxygen saturations. The two
`lines shown in figure 4.4 represent the properties for the two most commonly
`used wavelengths in pulse oximetry (660 nm and 940 nm). The absorbance of
`light at a wavelength of 940 nm increases with an increased oxygen saturation. At
`660 nm the absorbance of light decreases rapidly with an increasing functional
`oxygen saturation (Pologe 1987).
`It is possible to determine the concentrations of hemoglobins in hemoglobin
`solutions or hemolized blood by using a device such as a spectrophotometer (see
`section 3.3).
`
`~----ai 660 nm~
`-S-al-~ nrnl
`
`1
`1
`80
`60
`40
`20
`Functional oxygen saturation (%)
`
`100
`
`0.9 -
`08
`0.7 -
`06-
`0.5 -
`0.4 -
`0,3 -
`0.2 -
`0.1
`0
`
`0
`
`Normalized absorbance
`
`Figure 4.3 Changes in light absorbance in hemoglobin solutions as a function of functional
`oxygen saturation for the wavelengths used in pulse oximetry. Absorbonce decreases rapidly with
`increasing oxygen saturation at 660 nm (dashed line) but increases slightly with increasing oxygen
`saturation 21 940 nm (solid line).
`
`4.3.3 Pulsation of the blood
`
`Light traveling through biological tissue (e.g. the finger or earlobe) is absorbed
`by different absorbing substances. Primary absorbers of light in the region of
`interest are the skin pigmentation, bones, and the arterial and venous blood.
`Instead of measuring the arterial oxygen saturation of the blood in vitro with a
`sample of arterial blood and a spectrophotometer, or at a wide range of different
`wavelengths as with the Hewlett-Packard ear oximeter, pulse oximeters take
`
`63
`
`11 1
`
`1
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0064
`
`APL_MAS_ITC_00015681
`
`Light absorbance in pulse oximetry
`
`47
`advantage of arterial pulsation. Figure 4.5 shows the amount of absorbed and
`transmitted light in living tissue as a function of time.
`The arteries contain more blood during systole than during diastole, and
`therefore, their diameter increases due to increased pressure. This, effect occurs
`only in the arteries and arterioles but not in the veins. The absorbance of light in
`tissues with arteries increases during systole mainly because of the larger amount
`of absorbing substances (hemoglobin), due to the fact that the optical path length
`d in the arteries increases. This alternating part of the total absorbance allows us
`to differentiate between the absorbance due to venous blood, a constant amount of
`arterial blood, and other nonpulsatile components such as skin pigmentation (dc
`component of the total absorbance) and the absorbance due to the pulsatile
`component of the arterial blood (ac component). The alternating part of the light
`absorbed by the living tissue usually does not exceed 1 % to 2% o f the constant
`absorbance of the dc components. The time varying signal of transmitted light is
`refurred to as the plethysmographic (or photoplethysmographic) signal.
`The intensity of the light passing through the tissue during diastole is high
`(IH)· The absorbers that are present during diastole are the DC components. All
`DC components except the nonpulsating arterial blood are collectively
`represented by ED((,11, cDC, and dDC· The diameter of the arterial vessels is
`minimal (dmin) and therefore the absorbance due to arterial hemoglobin is
`minimal and the amount of transmitted light is high ( /H) and has a peak (see
`figures 4.4 and 4.5)
`
`41 = ID e-£De(A)cDC d[)Ce-I£Hb(A)CHb+£HbO2 (A)CHbOG]dmin (4.12)
`
`ft 1 1-
`10«4~- ''9'777-,i//07 Nonpulsaling arterial blood
`
`t
`
`Pulsatingarterial blood
`
`Venous blood
`
`E E
`
`0* Ii' -**~jor-u
`
`[r,cluent ligi,1 4]
`
`1- 1
`
`One cardiac
`cycle
`
`Other tissue
`
`t
`
`Figure 4.4 Absorbed anci transmilted light in living tissue. The amount of absorbed light
`coirclates witli the pulsation of aiterial blood. A constant amount of- light is absorbed by the skin
`pigmentation. bone, other t issue, venous blood and tlic noripulsating part of the arterial blood.
`More blood is present in the aiteries during systole and therefore more light is absorbed. The
`intensity of the transmitted light varies from 41 (maximum) to /L (minimum) within one cardiac
`cycle
`The optical path length in the arteries increases during the systole to dmax. The
`amount of absorbed light reaches a maximum peak and therefore the transmitted
`light reaches the low peak /L
`
`64
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0065
`
`APL_MAS_ITC_00015682
`
`48
`
`Design of pulse oximeters
`
`4. = /O e-EDC(A)CDC £4)Ce-££Hb(A)6·Hb +EHbO2(A)£.Hboildmax
`
`(4.13)
`
`The light intensity I of the light arriving at the photodetector is a function of the
`diameter d of the arteries and arterioles. During one cardiac cycle this diameter
`changes from dmin to dmax· By substituting d with dmin + Ad we derive the
`fullowing expression from Beer's law, where / is expressed as a function of 41
`and Art, the part of the diameter that changes from 0 to dinax - dmin with time
`(4.14)
`
`I =IHe -[8Ht,(A)c·Hb +€HbO2 (A)CHI,02 ]Ad
`
`Figure 4.5 shows these properties in a simplified model.
`
`253<venous,v.AT:.1 ,
`02< Arteriali-1 Photo-
`4ioAM.&*blood, r,~ov
`LED 'r---*pigment.*%' bjood €
`detector
`~----'*ation,
`3.Metc. *455*«
`1
`
`1
`1
`1
`1
`1
`1
`1
`1
`
`Ught
`intensity /
`
`/0
`
`dmin
`
`1-11 1
`1
`lH-- - 4
`1 Ill
`1 1
`.1. 1 .1
`
`dDC
`
`dmax
`
`Figure 4.5 Beer's law in pulse oximetry. The DC components of the tissue (e.g. skin
`pigmen[ation. bone, venous blood and the nonpulsating part of the arterial blood) absorb a constant
`amount of the inciden, light lo. The effective optical path length in the DC components without the
`constant level of arterial blood is represented by dD£· During diastole [hc optical path length
`[hrough the arteries has a minimum length of dmin and the ligh[ in[ensity at [hc photodetector is
`maximal (41)· 'Ilhe optical path length reaches a maximum d~ax during systolc and the hemoglobin
`in the artenes absorbs a maximum amount, causing / to decrease to a minimum level of /L
`
`4.3.4 Measurement of pulse oximeters
`
`The reading of the pulse oximeter SpO2 is an estimation of the arterial oxygen
`saturation Sa02· Measuring at two wavelengths allows us to distinguish the
`concentrations of only two different absorbers (Hb and HbO2). But in humans
`more species of hemoglobin, such as carboxyhemoglobin and methemoglobin, are
`present. These other hemoglobins absorb light as the functional hemoglobiris do
`and therefore influence our measurements. As long as we do not measure at as
`many wavelengths as absorbers are present in the blood, we can not determine the
`concentrations of Hb and HbO2 and therefore the arterial oxygen saturation
`correctly (Barker and Tremper 1987).
`
`65
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0066
`
`APL_MAS_ITC_00015683
`
`Light absorbance in pulse oximetr¥
`
`49
`
`Due to the fact that Hb and HbO2 are the main absorbers, the error may be
`small. Nevertheless, the results of determining either the actual functional o r
`fractional oxygen saturation (see equations (4.5) and (4.7)) of the arterial blood
`are not exact. This problem is also discussed in sections 10.1.1 and 11.1.1. The
`oximeter reading becomes less accurate if the concentrations of dyshemoglobins
`are larger than in normal humans. Section 11.7 deals with the presence of high
`concentrations of dysfunctional hemoglobins.
`
`4.4 SATURATION VERSUS NORMALIZED RATIO
`
`The arterial oxygen saturation can be derived based on Beer' s law as a function
`of the ratio of absorbances at two wavelengths. Due to nonlinearities in the LEDs,
`the photodetector, and light absorbance in the tissue, the absorbances have to be
`normalized in the ratio. This model results in a theoretical calibration curve, but
`it is not used in practice as will be described in the following sections.
`
`4.4.1 Normaliz.ation
`
`The measured light intensities at the different wavelengths have to be normalized
`before they can be compared with each other due to the fact that the light-
`emitting diodes (LEDs) may emit light with different intensities. The absorbing
`characteristics of the DC components and the sensitivity of the photodetector
`differ for the two different wavelengths and the tissue absorption and path length
`varys widely from patient to patient and with the probe site (de Kock and
`Tarassenko 1991). The normalized signal In is calculated by dividing the
`transmitted light intensities (the raw signals) by their individual maximum peaks
`(41,R for the red wavelength and IH,IR for the infrared wavelength). From
`equation (4.14) we derive
`
`In - J_ = e-[€Hb(A)CHb +EHb02 (A)CHI:,02 ]Ad
`IH
`
`(4.15)
`
`This results in normalized signals with the same intensities 41,n during diastole.
`The normalized signals of the transmitted red and infrared light are independent
`of the incident light levels and photodetector nonlinearities as shown in figure
`4.6. The AC components of the normalized signals represent only changes of
`transmitted light caused by the pulsation of blood in the arteries and can be
`compared with each other. They depend on the absorbers present in the arterial
`blood (ideally Hb and HbO2) and the actual optical path length d through the
`volume changing part of the arteries.
`
`4.4.2 Ratio of normalized signals
`
`The absorbance of the light is derived by calculating the natural logarithm of the
`measured and normalized transmitted light level. Dividing the raw signal by the
`iransmitted light during diastole 4i as in equation (4.15) and calculating the total
`absorbance then is comparable to calculating the total absorbance only due to the
`AC components in the pathway. The transmitted light during diastole represents
`the new nonchanging incident light level and the ratio R of these normalized
`
`66
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0067
`
`APL_MAS_ITC_00015684
`
`50
`
`Design of pulse oximeters
`
`absorbances at the red (R) and infrared (IR) wavelengths depends only on the
`light absorbers present in the arterial blood (see equation (4.3))
`
`/H,IR
`
`t
`41.R
`1 .R
`
`t
`41,n
`44,n
`6.,AA
`6.t€h.Op
`646:· Id 19023
`
`Normalized signals
`'Raw' signals
`Figure 4.6 The nonnalization of the signals The [ransmitted light from the red LED (R) and
`from the infrar'ed LED CIR) is divided by its individual DC component. 11,us. both normalized light
`intensities have [lic same magnitude during diastole. The normalized signals determine the basis for
`the calcul•lion al- the arterial oxygen samration.
`
`ALR ln(IL,R /41,R)
`At, IR lnUL, IR /Ill,IR )
`
`By using equation (4.15) the ratio can be derived as
`
`R=
`
`[(£Ht,(AR )CHb +(EHbO, (AR k.'Ht-,02 ~AdR
`[(CHI)(AIR)(Hb + (EnbO2 (kIR)('HbO·, ]Ad]R
`
`(4.16)
`
`(4.17)
`
`Assuming that the optical path lengths d~ for red light and dIR for the infrared
`light are equal, only the arteries change their diameter, and using equation (4.ID
`
`R=
`
`£Hb(AR)+[€HbC)2 (AR 1 -Ellb (AR )]SaO2
`£Hb (AIR ) + [£Hboa C XIR 1 - €111, ( AIR )]S:102
`
`(4.18)
`
`In this form the ratio R is not a function of the optical path length and can be
`derived from the arterial oxygen saturation instead of the concentration of the
`hemoglobins in the blood (see de Kock and Tarassenko 1993).
`
`4.4.3 Theoretic calibration citrve
`
`Equation (4.18) can be rewritten in a form where Sa02 is a function of the
`measured and calculated ratio R
`€Hb (AR)- €Hh (AIR)R
`£Hb (AR ) - £HbO'l (XR ) + ~€Ht,02 (AIR ) - €Hb (AIR )]R
`
`Sa02=
`
`x 100%. (4.19)
`
`Therefore. the functional oxygen saturation in arterial blood can be derived
`theoretically by calculating the ratio R of measured and normalized total light
`
`67
`
`1
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0068
`
`APL_MAS_ITC_00015685
`
`Light absorbance in pulse oximetry
`
`51
`
`absorbances in the red and infrared region and using equation (4.19). Figure 4,7
`plots this relationship as the theoretical calibration curve.
`
`- theoretical I
`Amparl/=1 1
`
`1
`1
`
`t
`2
`3
`Ratio of normalized absorbances
`
`1
`4
`
`1
`5
`
`100 -
`90 -
`80 -
`9 70 -
` 60-
`2 50-
`e
`40 -
` 30 -
`* 20 -
`0
`10 -
`0
`0
`
`C6
`
`0
`
`00
`
`1
`
`Figure 4.7 Calibration curves for pulse oximeters: the solid line is the theoretical curve by Beer's
`law lind the dashed I ine is the empirical curve. The difference between these curves is due mainly
`to light scatte,ing effects. This empirical calibration curve is derived by a second order polynomial.
`
`4.5 VALIDITY OF BEER'S LAW IN PULSE OXIMETRY
`
`Incident light passing through human tissue is not split only into absorbed light
`and transmitted light as proposed by Beer's law. Some parts of the light are
`reflected and others are scattered.
`Light reflection at the skin surface and light absorbance due to tissue other
`than the pulsating arterial blood are overcome by using the plethysinographic
`waveform. However, the skin surface, tissue, muscle, bone and eapecially blood
`cause light scattering which increases the absorbance of light (see following
`section). Blood is a nonhomogeneous liquid, which is capable of nonlinear
`absorbance of light, e.g. as the concentration of hemoglobins varies (Wukitsch et
`al 1988).
`The variation in light absorbance is not entirely due to the increased optical
`path length during systole. If the change in diameter were the only reason, the
`variation would be much less. The reason is a change in the axis of the red blood
`cells, which changes their absorbance as well. Red blood cells have the shape of a
`biconcave disk. Their major diameter is aligned parallel to the direction of blood
`flow during diastile and aligns perpendicular to the direction of flow during
`qystole. l'herefore. the optical path length is larger during systole and increases
`light absorbance. Even the light reflectance changes with the axis of the red blood
`cells, which is important for the use of reflectance probes. As a result of these
`properties, the absorbance and reflectance of blood in motion varys within the
`cardiac cycle and with the velocity of blood flow (Moyle 1994).
`
`68
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0069
`
`APL_MAS_ITC_00015686
`
`1 52
`
`Design of pulse oximeters
`
`4.6 LIGHT SCATTERING
`
`The results of oximetry measurements with whole blood differ from the results
`of the theory based on Beer ' s law . A physical phenomenon called light scattering
`highly increases the absorbance of light. Nevertheless, pulse oximeters read the
`arterial oxygen saturation of the blood accurately enough for clinical use under
`normal circumstances. This is due to the fact that most of the commercial pulse
`oximeters use a calibration curve based on empirical data, because modeling the
`problem of light scattering mathematically for different conditions is very
`complex. Several approaches have been made to create models which describe the
`real process within certain limits of accuracy.
`
`4.6.1 Light absorbance in whole blood
`
`Unfortunately Beer' s law does not apply for whole blood. The absorbance of
`light is not simply proportional to the concentration of hemoglobin or to the
`length of the optical path. Beer's law assumes no light scattering, which is not
`true in whole blood, besides the fact that the LEDs do not emit monochromatic
`light.
`Shymada and Yoshida (1984) verified that the influence of multiple
`scattering can not be overcome be subtracting the DC level as had been expected.
`Kramer et al (1951) stated that the absorbance of light due to oxyhemoglobin and
`reduced hemoglobin is increased in whole blood compared to hemolyzed blood
`by factors of the order of five.
`The reasons for the increased absorbance are mainly scattering and multiple
`scattering . Light scattering causes the deviation of a light beam from its initial
`direction. It occurs when light is refracted by an object of a size similar to the
`magnitude of the wavelength of the light and a change in the index of refraction
`at the interface of this object. The wavelengths of red and infrared light do have
`the same order of magnitude as the geometric dimensions of red blood cells
`(approximately 7 pm in diameter). The discontinuity in the index of refraction at
`the interface between plasma and red blood cells and the great proportion of red
`blood cells in blood yield a highly light scattering medium. Light that is scattered
`once will likely be scattered again by cells and therefore multiple scattering
`occurs (Steinke and Shepherd 1986). Multiple scattering increases the optical path
`length and therefore increases the absorbance.
`The intensity of the light scattered by the tissue depends on such factors as
`the red blood cell concentration in the blood; on the size, shape, orientation, and
`index of refraction of the scattering particles; on the tissue thickness; and on the
`aperture cone of the detector (Fine and Weinreb 1995). The thickness of the
`tissue, the distance between the LED and the photodiode, and the concentration of
`hemoglobin will vary from patient to patient and the shape and orientation of the
`red blood cells is irregular. Thus it is difficult to develop a physical model which
`can be used under different circumstances.
`
`4.6.2 Models for light absorbance including scattering
`
`It would be very useful to find a relationship between SaO2 and the ratio R of
`normalized absorbances for whole blood instead of only for hemoglobin
`solutions. An accurate scattering theory for whole blood could replace the
`
`69
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0070
`
`APL_MAS_ITC_00015687
`
`Light absorbance in pulse oximetry
`
`53
`
`empirical calibration curves used for the SpO2 readings. A few attempts are
`described below.
`
`4. 6. 2. 1 Twersky 's multiple scattering theory. Twersky ( 1962 , 1970a, b) has
`developed an analytical theory to describe the scattering of light by large, low-
`refracting, and absorbing particles. It is based on electromagnetic field theory
`and uses statistical averages to expand the theory for scattering and absorbing
`valid for a single particle, to find a formulation valid for multiple scattering (de
`Kock and Tarassenko 1993).
`The total absorbance of whole blood can be expressed as the sum of
`absorbance as described by Beer's law and a second term representing the
`attenuation of light due to scattering. These two processes can be treated as
`independent processes. The intensity of scattering depends on variables such as
`those mentioned i n section 4.6.1. The theory can be adapted for a special setting
`and will provide accurate results, but once the physiological conditions change,
`recalibration is required (Fine and Weinreb 1995). Hitachi, Ltd uses Twersky's
`approach in one of their US patents (Ito et al 1993).
`
`4. 6 2. 2 Comparison of dijferent models. Steinke and Sheperd (] 986) compared
`Twersky's theory of radiation scattering and photon diffusion equations. They
`found Twersky's original equation to give the best fit for the measured data.
`Marble et al (1994) found the three dimensional photon diffusion theory to
`be useful for modeling tissue optics although the pulse oximeter system violates
`many of the requirements of the model. However, they came to the conclusion
`that this theory can not replace clinical calibration studies.
`De Kock and Tai·assenko ( 1993) also found Twersky's theory to give the best
`fit to the experimen[al data. They compared results of this model with the photon
`diffusion theory and the Kubelka-Munk theory.
`
`4.6.3 Influence of scattering on pulse oximeter readings
`
`Although the assumptions of Beer's law are violated in pulse oximetry, the actual
`readings of the devices show a good correlation between the measurement and the
`actual arterial oxygen saturation.
`Steinke and Sheperd (1986) found that the scattering effects of the light
`passing through whole blood depend on the wavelength of the light and the
`oxygen saturation. The relationship between oxygen saturation and total
`scattering effects {absorbance due to hemoglobin plus multiple scattering) is
`approximately linear and so scattering does not influence the IiIlearily of the pulse
`oximeter in a negative way. In contrast, the total absorbance has a larger slope
`than that due only to the absorbance of hemoglobin following Beer's law.
`Therefore, light scattering increases the sensitivity of the whole blood nximeter.
`Fine and Weinreb (1993, 1995) demonstrate that the ratio of total
`absorbances is a function of the effective blood layer thickness and the
`concentration of hemoglobin. Therefore physiological factors such as
`temperature or peripheral vasoconstriction reduce the accuracy of saturation
`readings. The error increases as the level of arterial oxygen saturation decreases.
`This is dangerous because the clinician has to question the readings of the oxygen
`saturation when it is most critical for the patient.
`
`70
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0071
`
`APL_MAS_ITC_00015688
`
`4
`4
`4
`
`44
`
`444
`
`4
`
`44
`
`4
`
`54
`
`Design of pulse oximeters
`
`4.6.4 Calibration curves used ,for pulse oximeters
`
`Commercial pulse oximeters are calibrated from iii vir,-0 data (see section 10.1 )
`A large set of data obtained in clinical studies is collected containing information
`about the ratio R of' the absorbances calculated by the pulse oximeter and the
`actual arterial oxygen saturation Sa02 measured by a very accurate method such
`as the CO-oximeter (see section 3 3 I. Lookup tables or equations are used to find
`the relatinnship of these two variables for a pulse oximeter reading.
`To relate the measured values of the ratio R to the reading of the pulse
`oximeter, the equation of the theoretical calibration curve based on Beer's law
`can be modified as Mendelson and Kent (1989) described
`
`ki - k2R
`402=4-k#R
`
`(4.20)
`
`In this equation the extinction coefficients from equation (4.19) are replaced by
`constants ki. These constants are determined by clinical studies to give the curve a
`to the in ritro measured data. Another approach for a mathematical
`best fit
`representation is the use of a polynomial such as found for example in the
`Ohmeda 3700 and Radiometer OX100 pulse oximeters (Fine and Weinreb 1995)
`
`Sp02 =4 +k,R+43*2
`
`(4.21)
`
`Figure 4.7 provides an example of a calibration curve used in pulse oximeters in
`comparison to the theoretical calibration curve.
`
`REFERENCES
`Barker S J and Tremper K K 1987 Pulse oximetry : applications and limitations /nt. Anesthesiol.
`Clinics 25 155-75
`Bunn H F 1986 He,nogiobin: Molecular, Genetic, and Clinical Aspects (Philadelphia PA-
`Saunders)
`Fine I and Weinreb A 1993 Multiple scattering effect in transmission oximetry Med. Biol. Eng.
`Comput. 31 5 16-22
`Fine I and Weinreb A 1995 Multiple scattering effect in transmission pulse oximetry Med. Biol.
`Eng. Comput. 33 709-12
`de Kock J P and TarassenkoL 199 ! in vitro investigalion of tlic factors affecting pulse oximetry J.
`Biomed. En,t: 13 6 1 -6
`de Kock J P und Turassenko L !993 Pulse oximetry: [hcoretical and experimental models Med.
`Biol. Eng. Compur. 31 29 1 -300
`Ito Y, Kawaguchi F Yoshida M and Kohida 11 1993 Method and equipment for measuring
`absorptance of light scattering materials using plund wavelengths of light US patent 5.239. 185
`Kramer K, Elam J O. Saxion G A and Elam W N Jr 1951 influence of oxygen saturation.
`crythrocyte concentration and optical depth upon the red and near-infrared light transrnittance of
`whoic blood Am J Physiol. 165 229-46
`Malinheimcr P D, Casciana J R. Fein M E :ind Nierlich S L 1997 Wavelength selection for low-
`s,ituration pulse oximetry IEEETrans. Bionic {L Eng. 44 148-58
`Marble D R, Bums D H and Cheung P W 1994 Diffusion-based model of pulse oxiinetry: in vitro
`und in vivo comparisons Appl Op, 33 1279-85
`Mendelson Y and Kent JCI 989 Variationx in optical absorption spectra of adult und fetal
`hemoglobins and its effect on pulse oximetry /EEE Truns. Bionied Eng. 36 844-8
`Moyle JTB 1994 Pul.ve Oximaers iLondon: BMJ)
`Nellcor 1993 Hemoglobin and the principles of pulse oximetry Referenec Note: Pulse Oximetry
`Nofe Number ! (Picasanton , CA: Nellcor)
`
`71
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0072
`
`APL_MAS_ITC_00015689
`
`Light absorbance in pulse oximetry
`
`55
`Pologe JAI 987 Pulse oximetry : technical aspects of machine design Int. Anestliesio!. Clinics 15
`(3) 137-53
`Shymoda Y und Yoshida I 1984 Effects of multiple scaltering and peripheral circulation un arterial
`oxygen saturation measured with a pulse-type oximeter Med. Bial. Eng. Comptit. 11 475-8
`Steinke .1 M and Shepherd A P 1986 Role of light scatteri ng in whole blood oxi mctry IEEE Trans.
`Binmed. Eng. 33 294-301
`Twersky V 1962 Multiple scaillering of waves and optical phenomena J. Opt. Soc. Am. 52 145-71
`Twersky V 1 970a Interface cffects in multiple scattering by large, low refracting, absorbing
`particles J. Opt. Soc. Ant. 60 908-14
`Twersky V 1970b Absorption and multiple scattering by biological suspensions J. Opt. Soc. Am.
`60 1084-93
`Wukitsch M W, Petterson M T, Tobler D R and Pologe J A 1988 Pulse oximetry: analysis of
`theory , technology, and practice J. C/in Monitoring 4 290-301
`Zijlstra W G, Buursma A and Meeuwsen-van der Roest WP 1991 Absorption spectra of fetal and
`adult oxylicmoglobin, de-oxyhemoglobin , corboxyhemoglobin , and methemoglobin Cim.
`Chem. 37 1633-8
`
`INSTRUCTIONAL OBJECTIVES
`
`4.1 Describe the properties and limitations of Beer's law.
`4.2 Describe different species of hemoglobin and their effect on the oxygenation of blood.
`4.3 Describe the functional and the fractional hemoglobin saturation and their difference.
`4.4 Describe the properties and assumptions of the spectrophotometric method to determine
`oxygen saturation in hemoglobin solutions.
`4.5 Describe the pnociples of pulse oximetry :ind what a pulse oximeter measureN.
`4.6 Describe why and how a pulse oximeter measures the ahsorbance in the arterial blood only.
`4.7 Describe tile normalization of the signals mid the reasons for this normalization.
`4.8 Explain how and why the ratioof the normalized signals is calculated,
`4.9 Explain errors in the spectrophotoinetric method when used for whole blood samples.
`4.10 Describe the different physical phenomena occurring when light travels through tissue and
`blood.
`4.11 Describe what light scattering is and where it occurs in pulse oximetry.
`4.12 Describe the influence of light scattering on the accuracy of a pulse oximeter.
`
`1
`h
`i
`e
`
`0
`ya
`11
`e
`
`11
`
`4.
`
`1
`
`1
`
`72
`
`g,
`/t.
`J'
`
`dn
`
`g15In,of
`
`W-
`tro
`:tal
`
`try
`
`MASIMO 2053
`Apple v. Masimo
`IPR2022-01299
`
`

`

`RX-0035.0073
`
`APL_MAS_ITC_00015690
`
`..
`
`CHAPTER 5
`LIGHT-EMITTING DIODES AND THEIR
`CONTROL
`Brad W J Bourgeois
`
`In order to make pulse oximetry practical in the modern medical environment, a
`light source is required that is powerful enough to penetrate more than a
`centimeter of tissue yet diminutive enough to fit in a small probe. Chapter 4
`shows that it also is desirable for the light source at each desired wavelength to
`have a very narrow emission spectrum, which minimizes error
`in the
`measurement of arterial oxygen saturation (Sa02)· Fortuitously, light-emitting
`diodes (LEDs) fulfill all the requirements for the light source in a pulse
`oxtmeter.
`pulse oximeter designers is how to deal with variations and shifts in the peak
`However, LEDs are not without drawbacks. The primary problem faced by
`wavelength of each LED . Because the main function of a pulse oximeter,
`measuring arterial oxygen saturation, is so heavily dependent upon accurate
`values for the two wavelengths of light. a design which does all it can to
`compensate for LED wavelength changes will outperform its competition.
`This chapter discusses important characteristics of LEDs, a LED driver
`circuit in a pulse oximeter, and various problems with the use of LEDs in pulse
`oximetry.
`
`5.1 AN INTRODUCTION TO LIGHT-EMITTING DIODES
`
`Light-emitting diodes are the light source of choice for all pulse oximeters on the
`market today. Their small size, excellent drive characteristics, and large light
`output over a very narrow ban

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket