Field Testing of a Wireless Wearable Reflectance Pulse Oximeter
`R.J. Duckworth, Ph.D.†, Y. Mendelson, Ph.D.*, G. Comtois, B.Sc.*
`†Department of Electrical and Computer Engineering, *Department of Biomedical Engineering,
`Worcester Polytechnic Institute, Worcester, MA 01609.
`
`ABSTRACT
`
`EVALUATIONS
`
`A wireless wearable pulse oximeter has been developed based on a small forehead-mounted sensor.
`The battery-operated device employs a lightweight optical reflectance sensor and incorporates an
`annular photodetector to reduce power consumption. The system also has short range wireless
`communication capabilities to transfer arterial oxygen saturation, heart rate, body acceleration, and
`posture information to a PDA carried by military and civilian first responders.
`The PDA can monitor multiple wearable pulse oximeters simultaneously and allows medics to collect
`vital physiological information to enhance their ability to extend more effective care to those with the
`most urgent needs. The system can be programmed to alert on alarm conditions, such as sudden
`trauma, or physiological values out of their normal range. It also has the potential for use in combat
`casualty care, such as for remote triage, and for use by first responders, such as firefighters. This
`paper will describe the field testing and clinical utility of this wearable system, and compare its
`performance with that of conventional pulse oximeters.
`
`§ Reference HR and SpO2 measurements were obtained
`simultaneously from the wearable and Masimo SET(cid:210) pulse
`oximeters.
`
`§ Two reflectance-mode sensors were positioned using an elastic
`headband on each side of the forehead above the eyes.
`
`§ Readings were recorded from healthy volunteers during rest,
`walking, and stair climbing.
`
`§ Recordings were acquired during voluntary breath holding
`maneuvers.
`
`SYSTEM CONFIGURATION
`
`§ Sensor Module
`§ Small battery-operated forehead-mounted optical reflectance sensor.
`§ RF transmitter to transmit data to a belt-worn receiver module.
`
`RESULTS
`
`Medic
`
`
`
`Resting Resting
`
`
`
` Walking Walking
`
`
`
` Resting Resting
`
`
`
`RestingResting
`
`
`
`Climbing Stairs Climbing Stairs
`
`
`
` Resting Resting
`
`Simultaneous recordings of HR (upper trace) and SpO2 (lower trace) from the wearable (blue) and Masimo
`SET(cid:210) (red) pulse oximeters for different activities.
`
`DISCUSSION
`
`§ Wearable mobile platform:
`§ Implementing a short-range wirelessly-linked personal area network.
`§ Transmitting vital information to command units and remote off-site base stations.
`§ Compact, low-power, small, light-weight, unobtrusive, and wearable.
`
`§ Preliminary results show that SpO2 and HR readings are within an acceptable
`clinical range; response time of the wearable pulse oximeter is adequate for detecting
`hypoxic episodes.
`
`§ Advantages:
`§ Monitor vital signs during training exercises and combat missions.
`§ Help emergency first-responders operating in harsh and hazardous environments.
`§ Helping to save lives.
`§ Enables medics to commence triage within moments after soldier injury
`§ Medics can receive accurate information about severity and progression of injury
`§ Effective intervention: optimize available treatment and evacuation.
`
`§ Receiver Module
`§ Receives digital data from sensor module.
`§ Processes data in real-time by a low-power microcontroller.
`§ Transmits arterial oxygen saturation (SpO2), heart rate (HR), body
`acceleration and posture information to a PDA.
`
`§ PDA Module
`§ Can monitor multiple wearable sensors.
`§ Allows medics to collect vital physiological information and extend more effective care.
`§ Can be programmed to alert medics when physiological conditions are out of normal range.
`
`90 mm
`
`˘ = 30 mm
`
`50 mm
`
`Receiver Module
`
`Sensor Module
`
`Simultaneous hypoxic recording of SpO2 (upper trace) and HR (lower trace)
`from the wearable (blue) and Masimo SET(cid:210) (red) pulse oximeters.
`
`SUMMARY
`
`PDA INTERFACE
`
`§ Displays SpO2, HR, body
`acceleration and posture information.
`
`§ Carried by military and civilian first
`responders.
`
`§ Alerts carrier during alarm conditions,
`such as sudden trauma, or when
`physiological values are out of normal
`range.
`
`§ Can be used to monitor multiple
`casualties
`
`Preliminary testing of a new wearable reflectance pulse oximeter show that it
`could be used to monitor accurate SpO2 and HR from a forehead sensor. The
`system would allow medics to collect vital physiological information in the
`field. This information could enhance the ability of first responders to extend
`more effective medical care, thereby saving the lives of critically injured
`persons. The system is presently undergoing additional field testing to further
`evaluate its functionality and to determine future improvements.
`
`ACKNOWLEDGMENT
`
`This work is supported by the U.S. Army Medical Research and Material
`Command under Contract No. DAMD17-03-2-0006. The views, opinions
`and/or findings are those of the author and should not be construed as an
`official Department of the Army position, policy or decision unless so
`designated by other documentation.
`
`Simultaneous 4s time-series recordings of HR (upper trace) and SpO2 (lower trace) from the wearable (blue)
`and Masimo SET(cid:210) (red) pulse oximeters. Recording were repeated every 2hr over a 14hr period.
`
`MASIMO 2005
`Apple v. Masimo
`IPR2022-01292
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket