throbber
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL 45, NO 7, JULY 1998
`
`867
`
`A Study of Printed Spiral Coils for Neuroprosthetic
`Transcranial Telemetry Applications
`
`Maulik R. Shah, Richard P. Phillips, and Richard A. Normann,* Member, IEEE
`
`Abstract— We have explored the use of printed spiral coils
`(PSC’s) for neuroprosthetic transcranial telemetry applications.
`We fabricated two-dimensional PSC’s on a thin (25 m) poly-
`imide substrate using copper (35 m) as a conducting material.
`All the coils had a fixed inner diameter of 1.0 cm. We fabricated
`two sets of coils. One set of coils consisted of 2- to 5-turn
`circular and square spiral coils and had different trace widths
`(WW ), different spacings (S) between adjacent traces, and different
`outer diameters. The other set of coils consisted of 5-turn circular
`spiral coils and had fixed inner and outer diameters but different
`sRs and
`pRp)
`W to SS ratios. We measured loss resistances (
`and quality factors (Q) of these coils at different resonating
`frequencies in the range of 5–40 MHz. Over this frequency range,
`we observed that for fixed inner and outer diameters, the coil
`with the largest WW achieved the lowest Rs and the highest
`pRp and Q. These electrical properties and the fact that these
`coils can conform to the complex convoluted cortical surface
`suggest that a PSC [15] can provide a viable alternative to a
`conventional wire-wound coil for neuroprosthetic transcranial
`telemetry applications.
`
`losses, neuroprosthesis,
`Index Terms— Coil efficiency, coil
`printed spiral coil, transcranial telemetry system.
`
`I. INTRODUCTION
`
`individual electrodes and inject a safe level of current through
`them in order to elicit a pattern of phosphenes [15]. Further,
`in order to access the implanted electrodes and the associated
`electronics, and to provide the required level of current for the
`electrical stimulation of the visual cortex, a suitable interfacing
`link (the link that connects the external power supply and
`the external electronics to the implanted visual prosthesis) is
`required.
`In the previous designs of prototype visual prostheses, two
`types of interfacing links were implemented [2], [7]. These
`interfacing links were a telemetry system implemented by
`Brindley and his research group [2], and a multiple-lead wire-
`percutaneous connector system implemented by Dobelle et al.
`[7]. There are distinct advantages associated with the lead
`wire-percutaneous connector system: 100% efficient power
`transfer (providing for a long life of external batteries), wide
`bandwidth for data transfer, and no need for electronic circuitry
`to be attached with the implanted electrodes. However, the
`lead wire-percutaneous connector that breaches the skin and
`the skull, increases the chances of infection. As the lead wires
`are fastened to both the connector and the array, there is
`an increased chance of lead wire failure due to motion of
`the brain with respect to the cranium. This relative motion
`can cause forces on the array due to its being tethered
`to the connector, which increases the chance of lead wire
`failure and/or displacement of the implanted array. Further,
`patients are unlikely to accept a skull mounted connector for
`a chronic application. Considering these issues for a cortical
`neuroprosthesis (such as a visual prosthesis), a transcranial
`telemetry system across a closed scalp may offer advantages
`over a lead wire-percutaneous system. By eliminating a direct
`connection between internal and external components of the
`neuroprosthesis, the telemetry system substantially reduces
`the likelihood of infection and the consequences of tethering
`forces. However, the design of such a telemetry system offers
`more challenges than a lead wire-percutaneous connector
`system. Further, the use of implanted electronics mandates
`the use of hermetic sealing of the prosthesis to protect the
`circuitry from the corrosive environment of the cerebral spinal
`fluid. Considering these issues, we are beginning to implement
`a transcranial telemetry system to be used in the design of
`the visual prosthesis. The schematic diagram of the telemetry
`system, its associated components, and its integration with the
`electrode array and the demultiplexing chip are shown in Fig. 1
`[19].
`Designs of transcutaneous/transcranial telemetry systems for
`various neuroprosthetic applications have been reported in
`0018–9294/98$10 00 ª
`
`RESEARCHERS have demonstrated that by electrically
`
`stimulating the visual cortex of blind volunteer, spots of
`light (termed phosphenes) could be evoked where a sighted
`individual would normally have his/her visual field [2], [7].
`If patterns of electrically evoked phosphenes can be sys-
`tematically organized to convey useful spatial information,
`then a functional visual prosthesis could become a reality.
`These considerations, coupled with recent advances in the
`fields of silicon micromachining and microelectronics, have
`caused researchers to reexamine the possibility of developing
`a cortically based visual prosthesis for the blind. It is likely
`that silicon-based electrode arrays will become an integral
`part of such a visual prosthesis for accessing neurons of
`the visual cortex [20]. We have developed such an electrode
`array, fabricated on a 4.2 mm 4.2 mm monolithic silicon
`substrate and containing 100 microneedles that project out of
`the substrate [14]. An advanced, silicon-based demultiplexing
`scheme has also been designed to select each of the 100
`
`Manuscript received January 4, 1995; revised January 14, 1998 Asterisk
`indicates corresponding author
`M R Shah is with the Department of Bioengineering, The University of
`Utah, Salt Lake City, UT 84112 USA
`R P Phillips is with Link Research, Salt Lake City, UT 84111 USA
`*R A Normann is with the Department of Bioengineering, The University
`of Utah, Salt Lake City, UT 84112 USA (e-mail: normann@m cc utah edu)
`Publisher Item Identifier S 0018-9294(98)04381-X
`
`1998 IEEE
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Page 1 of 10
`
`SAMSUNG EXHIBIT 1014
`
`

`

`868
`
`IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL 45, NO 7, JULY 1998
`
`Fig 2 A 5-turn circular spiral coil (W = S = 313 m) fabricated on a
`flexible and biocompatible polyimide substrate
`
`appropriate to the skin depth at the frequency at which the
`) and the inner
`coil is operated [4], [25]. The trace width (
`trace spacing (
`) can be optimized to reduce the effect of
`proximity that could result due to the influence of current
`in the adjacent traces of the coil [4]. Further, the substrate
`upon which the coil is fabricated could provide an excellent
`insulating and encapsulating layer and could be used to mount
`discrete microelectronic components and provide interconnect
`traces for the electrode array and the demultiplexing chip.
`These aspects prompted us to fabricate and study the electrical
`properties of PSC’s that could be useful in our neuroprosthetic
`transcranial telemetry application.
`Researchers have described the electrical properties of cylin-
`drical, closely packed wire-wound coils in the frequency range
`of 0.2–20 MHz, as well as the dependence of series (
`)
`and parallel (
`) loss resistances of such coils on resonating
`frequency and the number of turns in the coils [8]. Others
`have reported on electrical properties of printed thin film
`coils (fabricated on glass and ceramic substrates) designed
`to be operated in the frequency range of 100 MHz–1 GHz
`for nonbiological applications [4]. In this reported study, the
`optimum quality factor (
`) (for a coil with an outer diameter
`of 0.25 cm) was obtained when the
`equaled to the
`. Since
`no data were provided on the electrical properties of PSC’s
`in the frequency range of 1–100 MHz, our study focused on
`the frequency range judged most appropriate for a transcranial
`telemetry application.
`We fabricated PSC’s on a thin, conformable, and biocom-
`patible (polyimide) substrate using copper as a conducting
`material (Fig. 2). We fabricated two sets of coils. One set
`of coils consisted of 2- to 5-turn circular and square spiral
`coils and had a variety of
`,
`, and outer diameters. The
`other set of coils were 5-turn circular spiral coils with fixed
`inner and outer diameters and different
`to
`ratios (
`to
`ratios ranging from
`to
`). We studied
`electrical properties of these PSC’s in the frequency range
`of 5–40 MHz. Over this frequency range, we observed that
`for fixed inner and outer diameters, the coil with the largest
`achieved the lowest
`and the highest
`and
`. These
`electrical properties and the fact that these coils are very thin
`and can conform to the complex convoluted cortical surface
`suggest that a PSC coil can provide a viable alternative to a
`
`Fig 1 Different components of the visual prosthesis and the telemetry
`system An external
`transmitting coil will send power and data to the
`implantable receiving coil The receiving coil will provide power and data to
`the demultiplexing chip and the electrodes The component interconnection is
`made clear in the side view of the prosthesis, which shows the demultiplexing
`chip [built into a very large scale integration (VLSI) chip] connected to the
`electrode array in the plane of the receiving coil The top view of the prosthesis
`shows the geometrical layout of the PSC and its interconnection with the
`electrode array and the demultiplexing chip
`
`the literature [2], [5], [8], [9], [13], [18], [21], [27]. Recent
`advances in silicon micromachining and silicon microelec-
`tronics [very large scale integration (VLSI) technology] have
`also enabled researchers to develop miniature neuroprosthetic
`devices that contain RF telemetry systems [16], [19], [26].
`Despite the design variations in these telemetry systems,
`most use a multiple-turn, three-dimensional (3-D), wire-wound
`receiving coil as an internal component of the neuroprosthesis.
`In a conventional wire-wound coil, different turns of the
`coil are stacked one upon another to achieve a multiple-
`turn coil with a “doughnut geometry.” If such 3-D coils are
`to be used in a visual prosthesis application, they may not
`be able to conform to the convoluted surface of the cortex.
`The thickness of the wire-wound coil can be substantially
`reduced by orienting successive turns of the coil in a spiral
`geometry [two-dimensional (2-D)] rather than the doughnut
`configuration (3-D). Conformability of a wire-wound coil
`could be achieved by bonding a very thin wire in a spiral
`configuration to a thin and flexible substrate. These issues sug-
`gest that a thin, conformable, and biocompatible coil could be
`fabricated using simple photolithographic and microfabrication
`techniques. Further, as such a printed spiral coil (PSC) would
`be produced using standard photolithographic techniques, the
`geometric features of the coil can be precisely specified to
`address issues such as the skin and proximity effect on coil
`efficiency. The issue of skin effect (at high frequency) can
`be addressed by adjusting the thickness of the coil to be
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Page 2 of 10
`
`

`

`SHAH et al : PRINTED SPIRAL COILS FOR NEUROPROSTHETIC TRANSCRANIAL TELEMETRY APPLICATIONS
`
`869
`
`TABLE I
`COIL DIMENSIONS OF DIFFERENT CIRCULAR AND SQUARE SPIRAL
`COILS WITH FIXED INNER AND VARIABLE OUTER DIAMETERS
`
`conventional wire-wound coil for neuroprosthetic transcranial
`telemetry applications.
`
`II. MATERIALS AND METHODS
`
`A. Coil Materials and Fabrication
`Polyimide was chosen as the substrate of the PSC’s de-
`scribed herein because of its demonstrated biocompatibility
`and because it has been widely used as an insulating material
`in a variety of neuroprosthetic applications [12], [17], [22].
`Copper was selected as the model material because of its high
`conductivity and its commercial availability. However, since
`copper is not a biocompatible material, copper coils should
`be avoided in systems intended for chronic implantation. All
`our coils were fabricated from a commercially available cured
`polyimide ( Kapton)-based copper foil (Rogers Corporation,
`Chandler, AZ) using photolithographic techniques. The overall
`thickness of the PSC was approximately 90 m, the thickness
`of copper was 35 m, and the thickness of polyimide was 25
`m. The remaining thickness was due to the adhesive between
`polyimide and copper.
`As the PSC was intended to be used along with the
`silicon-based Utah Intracortical Electrode Array [14], the inner
`diameter of all the PSC’s was chosen to be 1.0 cm, a size
`large enough to contain the demultiplexing chip, the electrode
`array (on the back side), and a few discrete microelectronic
`components in the center of the coil (see Fig. 1). We fabricated
`two sets of coils. In one set of the coils, the outer diameter
`of the PSC was variable and was dependent on the number of
`turns and
`and
`of the coils. These coils were a set of 16
`circular spiral PSC’s and also a set of 16 square spiral PSC’s
`with the number of turns varying from 2–5 having different
`and
`. We fabricated these coils in order to explore the
`electrical properties of PSC’s and specifically to study the
`effect of coil geometry, operating frequency, number of turns,
`, and
`on the coil efficiency. Table I shows the dimensions
`of square and circular spiral coils.
`The other set of coils were 5-turn circular spiral coils with
`fixed inner and outer diameters, but with a variety of
`to
`ratios (
`). These coils were fabricated such that the
`
`center to center inner and outer diameters of all the coils
`were 1.0 and 1.625 cm, respectively. In between the outer
`and inner perimeters of the coil, a variety of 5-turn coils
`were fabricated such that the sum of
`and
`remained
`approximately constant. However, individual values of
`and
`varied from coil to coil providing different
`to
`ratios
`for the coils (see Table II). Further, depending on the width
`(
`) associated with an individual coil (in this set of coils),
`the minimum and maximum inner and outer diameters varied
`approximately by 5% from the center to center inner and outer
`diameters of all the coils. We fabricated these coils in order
`to study the possible proximity effect due to different
`to
`ratio (
`) as reflected in coil efficiency. Table II shows the
`values of
`,
`, and
`to
`ratios of these 5-turn circular
`spiral coils.
`
`B. Analysis of Electrical Properties of Coils
`In a telemetry system, the transmitting and receiving coils
`are inductively coupled to each other in order to transfer the
`required power and data from the transmitter to the receiver
`(see Fig. 1). In order to deliver the required level of power
`and data from the transmitter to the receiver efficiently, in
`most of the cases, the transmitting and the receiving coils
`operate at a particular frequency (most commonly the coil
`is operated at the resonant frequency determined by the coil
`and a parallel capacitor). The circuit in which the coil and a
`capacitor are connected to each other such that they resonate at
`a particular frequency is called a resonant circuit (also called
`a resonant tank circuit). There are two main configurations of
`the resonant circuits. One configuration is the series resonant
`circuit in which the coil is resonated by connecting a capacitor
`in series with it, and in another configuration, known as a
`parallel resonant circuit, the coil is resonated by connecting a
`capacitor in parallel with it. The losses associated with the
`resonant circuit are divided into coil losses and capacitive
`losses. However, using capacitors that have negligible losses,
`the resonant tank losses can be attributed to the coil losses.
`The coil, acting as an inductor in either the series or parallel
`resonant tank circuit, has an intrinsic dc resistance associated
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Page 3 of 10
`
`

`

`870
`
`IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL 45, NO 7, JULY 1998
`
`TABLE II
`COIL DIMENSIONS FOR THE 5-TURN CIRCULAR SPIRAL COILS WITH FIXED
`INNER AND OUTER DIAMERERS AND DIFFERENT W TO S RATIOS
`
`with it. This dc loss resistance of the coil mainly depends on
`the nature of the conducting material used in the fabrication
`of the coil, the length of the coil and the cross-sectional area
`of the coil. When the coil is operated at a high frequency, the
`losses associated with the coil are commonly represented by
`the series resistance of the coil (
`), represented in series with
`the coil. However, in some cases, the losses associated with
`the coil,
`, are represented by a parallel loss resistance,
`.
`This ac resistance,
`of the coil, depends on several factors
`such as the frequency at which the coil is being operated, the
`skin effect, the proximity effect, and eddy current effects due
`to metallic objects around the coil.
`The quality factor (
`) of a coil is defined as how much
`energy it can store (in a coil inductor) versus how much energy
`it dissipates (by a series loss resistance,
`). The relationship
`between the coil
`, coil inductance (
`), and
`is defined as
`
`(1)
`
`The representation of
`is one way to describe
`in terms of
`the quality of the coil. Another way is to represent coil losses
`is by a parallel loss resistance (
`) that is represented as being
`in parallel with the coil. The relationship between the coil
`,
`coil inductance (
`), and the parallel loss resistance (
`) is
`defined as
`
`(2)
`
`Further, the relationship between
`is defined as
`and
`and can be further simplified as
`], for
`. Thus, in order to
`[or
`characterize the coil
`one should know coil inductance, coil
`operating frequency, and either its series loss resistance (
`)
`or its parallel loss resistance
`.
`We estimated coil inductances using appropriate equations
`described in the literature [1], [4], [6], [11]. To validate the
`calculated values, coil inductances were measured using two
`different methods. The first method involved making a parallel
`tuned tank with a PSC and a capacitor of a known value
`and connecting that parallel tank in series with a resistor. A
`sinusoidal voltage was applied to the circuit from a variable
`frequency generator and the waveforms across the tank and
`tank plus the resistor were monitored with an oscilloscope. By
`bringing both the waveforms into phase, we could determine
`
`the resonant frequency of the tank. From the value of resonant
`frequency, the inductance of the PSC was then calculated.
`The second technique involved the use of an HP 4194A
`impedance/gain-phase analyzer. The calculated values of coil
`inductances differed from the measured values by no more
`than 10%–15% for all PSC’s.
`We measured the dc resistances of all PSC’s with an
`LCR (inductance/capacitance/resistance) meter. Values of both
`the theoretically calculated and the measured dc resistances
`of PSC’s were in close agreement. We also measured loss
`resistances
`and
`of all
`these coils at different res-
`onating frequencies (the coils were resonated using different
`values of capacitance connected in parallel with the coil)
`in the frequency range of 5–40 MHz using the HP 4194A
`impedance/gain-phase analyzer. As the goal of the experiment
`was to achieve the greatest
`for a given coil, from the
`coils described herein, we characterized coils by calculating
`coil
`, using the relation between
`,
`, and
`:
`is the
`, where
`is the coil inductance,
`is the series
`is the parallel loss resistance.
`
`resonating frequency,
`loss resistance, and
`
`III. RESULTS
`The electrical properties of the PSC’s were measured in
`an unloaded, parallel tuned tank model made from a parallel
`connection of the PSC and a capacitor. Since the capacitors
`were selected such that they had negligible losses, all tank
`circuit losses were attributed to coil losses and represented
`either by
`or
`. The parallel tank model was chosen
`so that direct comparison could be made between the
`of
`the tank (
`of the coil) and the load (computed from the
`power requirement of the electrode array) that will be placed in
`parallel with the parallel tuned tank. For an efficient telemetry
`system, most of the power that is received at the receiver tank
`should be delivered to the load, which can be achieved by
`making
`of the parallel tuned receiver tank relatively high
`compared to the actual load (at resonance the receiver tank
`appears simply as a resistor). Thus, the goal of our study was
`to determine if a PSC could be fabricated that has a value of
`substantially greater than the anticipated load resistance
`and that the PSC could provide the required value of
`.
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Page 4 of 10
`
`

`

`SHAH et al : PRINTED SPIRAL COILS FOR NEUROPROSTHETIC TRANSCRANIAL TELEMETRY APPLICATIONS
`
`871
`
`The frequency dependence of series loss resistance (Rs), parallel
`Fig 3
`loss resistance (Rp), and quality factor (Q) of a 5-turn circular spiral coil
`(W = S = 313 m) at different resonating frequencies
`
`To investigate the frequency dependence of loss resistance
`of PSC’s, we measured values of
`,
`, and
`for our sets
`of 16 circular spiral PSC’s, 16 square spiral PSC’s (with four
`different values of
`and
`and four different numbers of
`turns), and 5-turn circular spiral coils with fixed inner and
`outer diameters having different
`to
`ratios. These tests
`were conducted at various resonating frequencies (by creating
`a parallel tank using each PSC and resonating it at a particular
`frequency with an appropriate value of a capacitor), ranging
`from 5–40 MHz. The results, as shown in Fig. 3, show that
`and
`have a square root dependence (
`) on
`the operating frequency. This has also been widely reported
`in the literature for other types of coils [4], [8], [25]. Since
`the relation between
`and
`can be approximated as
`, if the frequency dependence of
`is
`substituted in the relation,
`it
`is obvious that
`has the
`frequency dependence of the order of
`. The
`dependence of frequency on the
`, and
`,
`of a 5-turn
`circular spiral coil (
`313 m) is shown in Fig. 3. A
`similar dependence was found for the coils with 2–4 turns.
`The representation of the tuned tank losses (coil losses) by
`a parallel loss resistance (
`) is one way to model the tank.
`A functionally identical model of tuned tank losses (the most
`common representation) is to represent the loss resistance of
`the coil as a series loss resistance
`. A plot of values of
`of PSC’s with
`313 m and different number of
`turns (Fig. 4) illustrates the effect of resonating frequency and
`number of turns on the coil losses.
`These results are similar to the results described in the
`literature for other types of coils [4], [8]. The coil losses
`increase with an increase in the number of turns and the
`operating resonant frequency. As it is convenient to replace
`of the coil by
`, similar PSC’s were used to measure
`values of
`at different resonating frequencies. Fig. 5 shows
`a comparison between the values of
`of the PSC’s (with
`313 m) having different numbers of turns, with the
`increasing as either the number of turns or the resonating
`frequency (or both) increases. Although this effect is similar to
`
`Fig 4 Values of series loss resistance (Rs) of 2- to 5-turn circular spiral
`coils (W = S = 313 m) at different resonating frequencies
`
`Fig 5 Values of parallel loss resistance (Rp) of 2- to 5-turn circular spiral
`coils (W = S = 313 m) at different resonating frequencies
`
`(Fig. 4), the frequency dependence
`the results shown for
`than for
`. This behavior has also been
`is greater for
`described by others for conventional wire-wound coils [8].
`Since the frequency and number of turns of the coil have
`effects on coil losses that are similar to those seen in the
`conventional wire-wound coils, coil geometry is an important
`parameter in the design of the spiral configuration of the
`for 5-turn circular PSC’s with dif-
`printed coils. Values of
`ferent
`and , measured at different resonating frequencies,
`are shown in Fig. 6. As the width of the coil increases,
`of
`the coil decreases. The coil with the biggest
`achieved the
`smallest
`. This was also found to be true for any number
`of turns in the coil [2–4 in our case].
`Values of
`for 5-turn circular PSC’s with different
`and
`, measured at different resonating frequencies, are also
`shown in Fig. 7. The PSC with
`313 m achieved
`the highest
`at each resonating frequency (the PSC’s have
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Page 5 of 10
`
`

`

`872
`
`IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL 45, NO 7, JULY 1998
`
`Fig 6 Values of series loss resistance (Rs) of 5-turn circular spiral coils
`with different W and S at different resonating frequencies
`
`Fig 7 Values of parallel loss resistance (Rp) of 5-turn circular spiral coils
`with different W and S at different resonating frequencies
`
`and ). This was
`different outer diameters due to different
`also found true for all other coils with different numbers of
`turns (2–4). This is because
`is inversely proportional to
`.
`The coil with a larger
`has a smaller
`(see Fig. 6). As
`is related to
`by the relation,
`, a coil
`with a larger
`(smaller
`) has a higher
`. The coil with
`250 m,
`313 m had only a small change in
`compared to the coil with
`250 m,
`375 m. As
`was the same in these two coils, the small change in
`was
`due to the change in inductance resulting from the different
`and the different outer diameter.
`Since losses in the PSC’s are reflected in coil
`, the results
`shown in Fig. 7 can be represented in terms of
`. These results
`are shown in Fig. 8, which show values of
`for 5-turn circular
`PSC’s with different
`and
`. A PSC with
`313
`m achieved the highest
`at all resonating frequencies. This
`was because the
`associated with the coil with
`
`Fig 8 Values of quality factor (Q) of 5-turn circular spiral coils with
`different W and S at different resonating frequencies
`
`313 m was minimum. Further, as shown in Fig. 8, a change
`in width of the coil from 163 to 313 m, resulted the change
`in
`by approximately 50% at 20 MHz. Further, the coils that
`had identical
`, the coil with
`250 m and
`375
`m and the coil with
`250 m and
`313 m, had a
`small change in
`caused by a change in
`of the coil.
`The above described coils (Figs. 3–8) had a fixed inner
`diameter and variable outer diameters. In these coils, the outer
`diameter was dependent on
`,
`, and the number of turns in
`the coil. However, in the design of a cortical neuroprosthetic
`device where miniaturization is an important issue, there is a
`restricted space available for the receiving coil. This constrains
`the size of the receiving coil and thus defines the inner and
`outer diameters of the coil. In between these inner and outer
`diameters of the coil, a certain number of turns can be provided
`by three different approaches: by making
`less than
`,
`equal to , and
`greater than . In order to study the effect of
`various
`to
`ratios on coil
`, we fabricated a set of 5-turn
`circular spiral coils that had fixed inner and outer diameters of
`1.0 and 1.625 cm, respectively. Further, by keeping the sum of
`and
`approximately constant, various
`to
`ratios were
`obtained. We studied the effect of various
`to
`ratios on
`,
`, and
`of these coils at different resonating frequencies in
`the range of 5–40 MHz. The results of these coils at 21 MHz
`are shown in Fig. 9.
`The top part of Fig. 9 shows the values of
`for these coils
`for different
`decreases
`to
`ratios. As shown in Fig. 9,
`as the
`to
`ratio of the coil increases. This is due mainly
`to the larger
`associated with the coil. This is similar to
`what has been shown in Figs. 6–8. The remaining parts of
`Fig. 9 show the results in terms of
`and
`. As
`and
`are related to
`such that any decrement in
`is reflected
`as an increment in both
`and
`, the coils with
`. Finally,
`achieved greater
`and
`than coils with
`the coil that had the biggest
`1.94), in the set of
`(
`all these coils, represented the lowest
`and the highest
`of
`,
`, and
`and
`. Finally, at 21 MHz, the values of
`the circular spiral coil with
`to
`ratio of 1.94 were 0.985
`, 4.774 k , and 69.6, respectively.
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Page 6 of 10
`
`

`

`SHAH et al : PRINTED SPIRAL COILS FOR NEUROPROSTHETIC TRANSCRANIAL TELEMETRY APPLICATIONS
`
`873
`
`(a)
`
`(b)
`
`(c)
`
`Fig 9 Values of (a) Rs, (b) Rp, and (c) Q of 5-turn circular spiral coils with fixed inner and outer diameters and different W to S (W=S) ratios
`
`IV. DISCUSSION
`
`For the visual prosthesis under development at the Uni-
`versity of Utah, we needed “surface mount space” in the
`center of the PSC for mounting the electrode array and the de-
`multiplexing chip, as well as various discrete microelectronic
`components. Surrounding this required “surface mount space,”
`we placed coils with 2–5 turns of either square or circular
`spirals. Because a circle has the minimum circumference for
`a given spacing between the inner edges of the coil, a square
`spiral geometry provides more area in the center of the coil
`and more inductance compared to a circular spiral coil of the
`similar dimensions. For similar numbers of turns, frequencies,
`
`dimensions, the dependence of coil losses (and
`and
`and
`) of square spiral PSC’s were similar to those measured
`for the circular spiral PSC’s. However, the consequences of
`misalignment between transmitting and receiving coils are
`greater for square spiral coils than for circular spiral coils
`(personal observations). Thus it would be preferable to use
`circular spiral coils rather than square spiral coils for a
`neuroprosthetic transcranial telemetry system.
`
`A. Coil Electrical Properties
`The PSC’s that we have built achieved electrical properties
`similar to other coils that have been described in the literature.
`A previous study on the coil losses of resonant wire-wound
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Page 7 of 10
`
`

`

`874
`
`IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL 45, NO 7, JULY 1998
`
`of circular wire-wound coils
`coils demonstrated that the
`changes by a factor of 1.5 for every doubling of the resonating
`frequency [8]. However, this compares only the relative values
`of coil losses at different resonating frequencies. In order to
`compare the efficiencies of wire-wound coils and PSC’s, we
`have normalized the series loss resistance (ac resistance) of
`both the types of coils (measured at 30 MHz for PSC’s and
`extrapolated to 30 MHz for wire-wound coils) to their dc
`resistances. The ratio (
`to
`) for PSC’s was approxi-
`mately four times smaller than their wire-wound counterparts
`described in the literature [8]. This suggests that our coils are
`less lossy than the wire-wound coils because the increase of
`coil resistance from its dc value to ac value (at a particular
`frequency) was smaller in our coils than the wire-wound
`coils.
`The qualities of PSC’s have been the subject of one other
`study in which the printed thin film coils were fabricated on
`ceramic and glass substrates [4]. This study focused on the
`electrical properties of printed thin film coils in the frequency
`range of 100 MHz to 1 GHz for nonbiological applications.
`The coils differed from our PSC’s in terms of size, substrate
`material, substrate thickness, frequency of operation, and the
`model of the coil circuit (resonated tank in our case, self-
`resonated tank in their case). An empirical relationship was
`derived in the previous study that allows the determination of
`the series loss resistance,
`, of any coil (with any dimension)
`at any frequency [4]. However, a constant that was used in
`the empirical relationship was derived to fit the data that were
`taken for their printed thin film coils in the frequency range of
`100 MHz to 1 GHz. We calculated values of
`for our PSC’s
`based on this relationship in order to compare the results. The
`predicted frequency dependence of our PSC’s agreed with the
`measured values. However, the absolute values of
`for our
`PSC’s agreed only roughly with this empirical relationship.
`We do not have any solid reasons to explain this, however
`we can only speculate that the constant that was calculated
`in order to fit the data taken for the printed thin film coils
`operated in the frequency range of 100 MHz to 1 GHz, did
`not fit well with the values of loss resistances of our PSC’s.
`Further, for our PSC’s in the frequency range of 5 to 40 MHz,
`we could not find the effect of proximity that was described
`for the printed thin film coils by others [8]. We could not
`find any optimum
`ratio, instead we found a dominating
`effect on the coil losses for the our PSC’s: the greater
`is the width (
`) of the coil, the greater is the coil
`(see
`Fig. 9).
`
`B. Mutual Inductance Between Spiral Coils
`
`Because we have developed a PSC with a 2-D geometry
`in order to reduce the height of the coil, we are forced to
`use a spiral configuration of the coil
`to achieve multiple
`numbers of turns. Since we are proposing the use of a PSC
`for telemetry applications,
`the mutual
`inductance between
`two circular spiral PSC’s (transmitting and receiving coils)
`should be calculated. The calculation of the mutual inductance
`between two circular coils is described in the literature [18].
`Similar analysis can be done between two circular spiral coils.
`
`However, because the radius of the spiral coi

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket