throbber
DIABETES TECHN OLOGY & THERAPEUTICS
`Volume 3, Number 4, 2001
`© Mary Ann Liebert, Inc.
`
`Dose Accuracy Testing of the Humalog® /
`Humulin® Insulin Pen Device
`
`MICH AEL J. RO E, P.E., M .B.A., DEBRA IGNAUT, C.D.E., RN., TOM MIYAKAWA, M.S.,
`and CH ERYL HU LTMAN, Ph .D.
`
`ABSTRACT
`
`The primary purpose of the study was to determine whether pen users would challenge the
`insufficient remaining dose (IRD) stop mechanism with sufficient force to affect the dose ac(cid:173)
`curacy of the final dose. The secondary purpose was to determine the participant's positive
`and negative impressions of the Humalog®/ Humulin® pen and the likelihood of using the
`new prefilled pen. Three different modifications to the prefilled pen's IRD stop feature were
`made. These three pen models then underwent environmental dose accuracy tes ting at vari(cid:173)
`ous temperatures and humidities, and user dose accuracy testing by 64 patients with diabetes.
`Evaluation also involved challenging the IRD stop at various dialing torques. Thirty pens
`from each model were tested to failure of the IRD stop. A model of the prefilled pen was se(cid:173)
`lected for commercialization that met the dose accuracy targets of ± 1 unit (U) for insulin
`doses less than 20 U and ± 5% of dose volume for doses equal to or over 20 U. The selected
`pen model was superior at the minimum (1 unit), median (30 unit) and maximum (60 unit)
`dose volumes. Also 92% (n = 59) of patients interviewed felt that the stop mechanism for the
`final dose was clear. Extensive testing in the development of a prefilled insulin delivery de(cid:173)
`vice demons trates an accurate and reliable medical device.
`
`INTRODUCTION
`
`W HEN DEVELOPING NEW DRUGS, clinical trials
`
`are cond ucted to provide information on
`safety and efficacy in order to obtain regula(cid:173)
`tory approval. In gen eral, an insulin injector
`must be able to comply with global regulatory
`stan dards as well as U.S. Food an d Drug Ad(cid:173)
`ministra tion (FDA) regulations when these de(cid:173)
`vices are manufactured and distributed in the
`United States. In addition, all types of insulin
`delivery devices developed for distribution in
`Europe must comply with the Internation al Or(cid:173)
`ganiza tion of Standards (ISO). These stan dards
`are a series of laboratory-based tests that con(cid:173)
`firm the device will deliver an appropriate dose
`
`over a variety of environm ental conditions, for
`examp le, after being dropped from on e meter
`in a variety of positions. The ISO regulations
`con tain specific requirements regarding dose
`accuracy. The testing in th is article specifically
`addresses the dose accuracy requirements
`specified in the ISO guidan ce docum ent. For
`medical devices such as the insulin injector
`pens, there are no requirements to perform
`clinical trials in order to obtain European reg(cid:173)
`ulatory approval, termed the Conformity Eu(cid:173)
`ropean (CE) mark. However, the CE mark re(cid:173)
`q uires th at the pen incorporate a n umber of
`safety fea tures to ensure accurate dosage.
`The H umalog® / H um ulin® Disposable In(cid:173)
`sulin Injection pen was developed by Eli Lilly
`
`Lilly Research Laboratories, Eli Lilly and Company, Lilly Corporate Center, Indianapolis, Indiana.
`
`FRESENIUS EXHIBIT 1074
`Page 1 of 7
`
`

`

`624
`
`I.R.D. Key
`
`hook detail on Nut
`engages with hook
`on Screw key in
`self-locking mode
`
`FIG. 1. "Hook and Key" IRD stop.
`
`and Compan y (Indianapolis, IN) and first mar(cid:173)
`keted in Europe in 1998. It is capable of ad(cid:173)
`ministering Humalog® / H umulin® subcuta(cid:173)
`neously in accurate doses from 1 to 60 units, in
`1-unit increments. At the time, it was the only
`insulin pen on the market that allowed for sin(cid:173)
`gle-unit dose adjustmen t. The injector is de(cid:173)
`signed to meet the current draft of ISO stan(cid:173)
`dards1 (ISO/DIS 11608-1, 11608-2, 11608-3 TC
`84) governing design verification of pen injec(cid:173)
`tors for medical use.
`In the pen's development, it was importan t
`to establish that all doses delivered would have
`the same degree of accuracy when used by pa(cid:173)
`tien ts. This created a challen ge for Lilly's me(cid:173)
`chanical/ design engineers. The result was a
`prefilled pen that con tains an insufficient re(cid:173)
`maining dose (IRD) stop, which p revents the
`user from setting a dose greater than that re(cid:173)
`maining in the cartridge. This is don e by means
`of a hook and key detail on the nut an d screw,
`respectively. When the nut reaches the end of
`the screw, a stop in th e nut threads (hook) hits
`a stop at the end of the screw threads (key),
`thus preventing the user from dialing a dose
`greater than the amount remaining in the car(cid:173)
`tridge (Fig. 1). When this feature is engaged,
`the screw is prevented from turning by the anti(cid:173)
`backup device (ABD) fingers, w h ich are
`molded into the body halves (Fig. 2).
`In or der for the IRD stop feature to function
`properly, the hook an d key feature must be po(cid:173)
`sitioned such that the user feels an increase in
`dialing stiffness (the stop). This increase in stiff-
`
`ROE ET AL.
`
`ness should occur as the last true remaining
`dose starts to appear in the dose window. This
`indicates to the user that the number appear(cid:173)
`ing in the window is the last remaining dose,
`and signals him or her to stop turning the dial.
`In addition to the above, the IRD stop should
`be able to withstand the maximum dialing
`torque expected to be exerted by users while
`still delivering an accurate dose; otherwise, the
`user could exert enough force to override the
`stop feature, causing an under dosing of insulin
`on the final dose.
`This document will summarize a study of
`some of the torque and dose accuracy tests in(cid:173)
`volved in the development of the H uma(cid:173)
`log® /Humulin® Disposable Insulin Injection
`pen, along with the criteria used for selection
`of the appropriate model for launch-that is,
`production and distribution to health care
`providers and patients with diabetes. The pri(cid:173)
`mary purpose of the study was to determine
`whether pen users would challenge the IRD
`stop mechanism with sufficient force to affect
`the dose accuracy of the final dose. And the sec(cid:173)
`ondary purpose was to determine th e partici(cid:173)
`pant's positive an d negative impressions of the
`Humalog® /Humulin® pen an d the likelihood
`of using the new prefilled pen.
`
`MATERIALS AN D METHODS
`
`In order to evaluate the IRD stop feature and
`the IRD stop strength, three different modifi-
`
`- -
`
`·, ··,..___ -~
`
`....=A=n=ti-=Ba=c=ku=p==;-----,,
`Device
`(ABO)
`
`FIG. 2. ABO detail.
`
`FRESENIUS EXHIBIT 1074
`Page 2 of 7
`
`

`

`INSULIN DEVICE TESTING PROCESSES
`
`cations to the feature in question were evalu-
`ated and then tested by actual patients with di-
`abetes. These pen modifications will be con-
`sidered prototypes Pen1, Pen2, and Pen3. The
`modifications made were different phasings of
`the nut feature—that is, the position of the hook
`and key feature relative to the dial interface.
`
`Pens
`The Humalog®/Humulin® Disposable In-
`sulin Injection pen contains 3.0-mL cartridges
`that hold at least 300 units of Humalog®/Hu-
`mulin® or other insulin mixtures, and the pen
`is disposed of when expended. The maximum
`dose possible with this pen is 60 Units (U), or
`0.6 mL, and the smallest dose possible is 1 U,
`or 0.01 mL.
`
`Patients
`Sixty-four patients with diabetes who were
`currently taking insulin injections participated
`in the study. They completed questionnaires
`that contained both open-ended and closed-
`ended questions on new pen models and were
`also interviewed one-on-one at the end of the
`protocol exercises. The study was conducted in
`Columbus, Ohio, on April 8, 1998. Participants
`did not know that Lilly was sponsoring the
`study or that the Humalog®/Humulin® pre-
`filled pen was a Lilly product.
`
`Insufficient remaining dose accuracy
`Pens were tested for IRD dose accuracy by
`challenging the IRD stop at various torques,
`reading the indicated dose, and then expelling
`
`625
`
`the dose into a measuring device. From the
`weight and the specific gravity of the fluid, the
`exact volume expelled could be determined.
`At least 15 pens were tested at each torque.
`The challenge torque values were selected
`based on the specification for overall strength
`of the feature and evaluation of actual user
`test data.
`The graph in Figure 3 shows the results of
`the dose accuracy testing. The horizontal axis
`contains the challenge torque applied in inch-
`ounces, while the vertical axis contains the av-
`erage dose error expressed as a percentage of
`the indicated dose.
`
`User test insufficient remaining dose accuracy
`In this test, each individual was given one
`each of the three pen models, as well as a prac-
`tice pen to familiarize the individual with the
`device. The three different pens were tested
`using a Graeco-Latin square design. This de-
`sign allowed for randomization of the pen
`type as well as randomization of the dialing
`protocol sequence used. Dose sizes of 12, 18,
`28, and 41 units were chosen in order to eval-
`uate the feature under typical user conditions.
`Dose sequences were selected so that equal
`numbers of pens would hit the IRD stop at ap-
`proximately the four chosen doses. The user
`did not know when the IRD stop would be
`reached and so would adequately challenge
`the stop. Twenty-five actual uses at each
`unique dose were measured and the accuracy
`of the last dose was compared to the previous
`24 doses. The results of this testing are in-
`cluded in Figure 4.
`
`.. e .. w
`
`Cl) ,,,
`0
`0
`'#.
`
`---o- Pen 1
`_,.__Pen2
`
`-Pen3
`
`FIG. 3. Percent dose error versus patient challenge torque.
`
`
`
`FRESENIUS EXHIBIT 1074
`Page 3 of 7
`
`

`

`626
`
`ROE ET AL.
`
`.. e .. w
`
`QI .,
`0
`0
`c
`"' QI
`:Ii:
`
`□ Pen1 ■ Pen2
`
`□ Pen3
`
`- -~ -(cid:173)
`
`Absolute Difference
`
`Actual Difference
`
`FIG . 4. Patient dose accuracy tests on three different pen models.
`
`Insufficient remaining dose stop strength
`Thir ty pens from each build were tested to
`failure of the IRD stop, with the results detailed
`in the g raph in Figure 5.
`
`Level 1 and insufficient remaining dose
`accuracy simulation
`To simula te the performance of each config(cid:173)
`uration in real-world use, the user test dose ac(cid:173)
`curacy data for Pens 1, 2, and 3 were combined
`with w h at is termed level 1 dose accuracy da ta
`for con trol pens. That is, dose events a t the IRD
`stop were combined with a group of level 1
`dose accuracy data.
`In order to determ ine h ow man y IRD doses
`to include at each dose size, the following for(cid:173)
`mula was used:
`
`30-unit dose: [(30 units/ dose) x
`(60 doses/ test)]/(300 units/pen ) =
`6 pens per test
`
`ture [cool (5°C), standard (23°C with 50% rela(cid:173)
`tive h umidity), an d ho t (40°C with 50% relative
`humidity)], 1,000 trials were run in a simula(cid:173)
`tion program and the resulting K values (tol(cid:173)
`erance limit factors) were recorded. The results
`are expressed as th e probability of obtaining a
`K value high er than the required target K value.
`The target K value varied slightly for each
`group due to the different number of IRD doses
`includ ed at each dose size.
`These results are summarized in Figure 6. In
`each instance, a probability of less than 100%
`mean s that the IRD doses have an a d verse im(cid:173)
`pact on the likelihood of the prefilled pen pass(cid:173)
`ing dose accuracy standards. It is important to
`note that the representa tions in Figure 6 should
`not be confused w ith the con fidence interval or
`p con ten t of a dose accuracy test. They repre(cid:173)
`sen t the impact that the inclusion of the IRD
`dose has on th e overall dose accuracy of a cer(cid:173)
`tain body of data, with high er percen tages be(cid:173)
`ing more desirable than lower percentages.
`
`In the example above, on e would expect six
`pens to be used in th e test. At the end of each
`pen test, the IRD feature would be engaged, so
`six IRD doses are included a t the 30-unit level.
`
`60-unit dose: [(60 units/ dose) x
`(60 doses/ test)]/(300 units/pen ) =
`12 pens per test
`
`.c
`'c,
`C
`~
`ui
`0
`g;
`
`The IRD doses were combined with actual
`level 1 test data from a control model build. For
`each dose size (30 and 60 units) and tempera-
`
`Co ntrol
`
`Pen1
`
`Pen2
`
`Pen3
`
`FIG. 5. IRD stop strength versus pen type.
`
`FRESENIUS EXHIBIT 1074
`Page 4 of 7
`
`

`

`INSULIN DEVICE TESTING PROCESSES
`
`A
`
`%
`
`30 Unit K-Value
`Probabilities
`
`60 Unit K-Value
`Probabilities
`
`Pen 3
`
`%
`
`Pen 2
`
`627
`
`B
`
`Cool
`
`Std
`
`Hot
`
`Cool
`
`Std
`
`Testing Conditions
`
`Testing Conditions
`
`FIG. 6. 30U (A) and 60U (B) K value probabilities versus temperature.
`
`Pen 1
`
`Hot
`
`RESULTS
`
`Patients
`A total of 64 patients with diabetes were in(cid:173)
`volved in evaluating the three pen models (52
`patients produced a complete set of three test re(cid:173)
`sults). There were 26 males (40%) and 38 females
`(60%) included in the study; all were currently
`using insulin injections prior to the study. The
`majority of patients were using a needle and sy(cid:173)
`ringe (88%; n = 56), while 6%, or n = 4, were us(cid:173)
`ing the insulin pump, and 3%, or n = 2, were us(cid:173)
`ing the pen and 3%, or n = 2, were using both
`the pen and syringe and needles. The patient's
`age at diagnosis was 30 or younger for 30%, n =
`19, and over age 30 for 70%, n = 45.
`
`Analysis A: binomial method
`
`The absolute values of the differences from
`nominal were compared. This is the actual dif(cid:173)
`feren ce in the dose delivered by th e pen versus
`the dose indicated by th e dial on the pen at the
`beginning of the injection . This difference was
`calcula ted a t small, medium, and large dose
`sizes. Each pair-wise comparison was made
`and tested: Pen 1 versus Pen 2, Pen 1 versus
`Pen 3, Pen 2 versus Pen 3. For example, in the
`comparison of 1 versus 2, in 32 of 58 subjects,
`Pen 1 delivered a dose closer to the indicated
`value. Pen 2 was closer to indicated value in 26
`of 58 subjects. If the pens were identical, one
`would expect on the average a 29 /29 break(cid:173)
`down . Based on th e laws of probability of the
`
`binomial distribution, the p value for a 32/26
`breakdown is 0.5114, meaning that there is
`about a 50% chance of obtaining a result of
`32/26 or more extrem e (such as 34/24 or
`26/32), given the pens are n o different. There(cid:173)
`fore, 32/26 is not strong evidence to suggest
`that they are different with respect to deliver(cid:173)
`ing doses close to indicated value.
`
`Comparisons
`
`Closer to nominal
`
`Pen 1 vs. Pen 2
`Pen 1 vs. Pen 3
`Pen 2 vs. Pen 3
`
`Pen 1, 32; Pen 2, 26
`Pen 1, 18; Pen 3, 38
`Pen 2, 15; Pen 3, 42
`
`p value
`
`0.5114
`0.0111
`0.0006
`
`Analysis A results. Pens 1 an d 2 are n ot sig(cid:173)
`nificantly different at the 95% confidence level.
`Pen 3 is significantly better in dose delivery
`than Pen 1 and Pen 2.
`
`Analysis B: nonparametric test on absolute
`difference from indicated dose
`Again, the absolute value of the differential
`was calculated for each pen, each patient. The
`mean delta between the two p ens was graph ed
`and tested against 0 (zero). If a test shows a mean
`delta significantly different from 0, the devia(cid:173)
`tions (from indicated) on one pen are deemed
`larger than deviations from the other pen. Be(cid:173)
`cause a couple of the deltas were far askew from
`the other data points, the distributions in two of
`the three cases were not norm al and thus, a non(cid:173)
`parametric signed rank test was used.
`
`FRESENIUS EXHIBIT 1074
`Page 5 of 7
`
`

`

`628
`
`Analysis B results. Mean absolute value from
`indicated dose (in units) was as follows: pen 1,
`1.082; pen 2, 0.994; and pen 3, 0.354.
`
`Signed rank test results
`
`p value
`
`Pen 1 vs. Pen 2
`Pen 1 vs. Pen 3
`Pen 2 vs. Pen 3
`
`0.316
`0.000
`0.000
`
`Pen 3 is again significantly lower in devia-
`tion from the indicated dose, and no significant
`difference was found between Pen 1 and Pen 2
`(Fig. 4).
`
`Analysis C: actual differences from indicated
`The actual differences were calculated to as-
`certain the direction of bias from indicated, if
`a bias exists (Fig. 4). A negative value repre-
`sents a dose delivered which was less than the
`indicated value. The p value is asociated with
`the mean differential compared to zero (over-
`all, no bias).
`
`Analysis C results.
`
`Mean differential
`(difference in units
`from indicated dose)
`
`p value
`(difference from 0)
`
`Pen 1
`Pen 2
`Pen 3
`
`20.725
`20.895
`20.106
`
`.001
`.000
`.045
`
`As expected, all three pen means are nega-
`tive (delivered less than indicated). While all
`three means are different from zero, the bias of
`Pens 1 and 2 is almost a full magnitude larger
`than Pen 3.
`
`Insufficient remaining dose feature strength
`The graph in Figure 5 indicates that Pen
`2 and Pen 3 demonstrated higher IRD stop
`strengths, and Figure 6 indicates that Pen 3 pro-
`vides the higher percentage and therefore bet-
`ter results across all temperature conditions for
`both the 30- and the 60-U dosage amounts. Un-
`der the indicated conditions, Pen 3 was the
`more accurate of the three pen models tested.
`
`Patient results
`Patients rated the pens qualitatively on ease
`of use, accuracy, and the adequacy of the IRD
`
`ROE ET AL.
`
`stop. The majority (84%) of the patients felt that
`the pen was very easy to use, and 92% felt that
`the stop mechanism for the final dose was clear.
`Also, 77% of patients were confident in the ac-
`curacy of the pen. The majority of patients
`(70%) reported that they would be somewhat
`or very likely to use the new prefilled pen if it
`was on the market.
`The main patient concerns about using the
`new prefilled pen included concerns about not
`being able to mix different types of insulin and
`the cost/insurance issue, since, currently in the
`United States, the cost of insulin pen injectors
`is not covered by most types of medical insur-
`ance.
`
`DISCUSSION
`
`We believe that safety standards for insulin
`delivery systems should include both envi-
`ronmental and wet testing by actual patients
`with diabetes who used the insulin pens with-
`out any formal training. The results are pre-
`sented and summarized in Figures 3–6. In
`summary, the data indicate that Pen 3 dem-
`onstrated the best performance over the
`evaluation criteria, and therefore Pen 3 was
`considered the best candidate for commercial-
`ization efforts.
`The patients’ results indicated that most par-
`ticipants (84%) thought that the Humalog®/
`Humulin® pen was very easy to use. The stop
`mechanism for the final dose was rated “very
`clear” by 92%, or n 5 59, of the respondents,
`and 70% said they would be likely to use the
`new pen if it were on the market.
`These data have been corroborated by sub-
`sequent studies on the new prefilled pen’s
`functionality and patient acceptance of its use.
`In a study conducted in Spain on 42 patients
`with either type 1 or type 2 diabetes, 77% of pa-
`tients preferred the new prefilled pen over their
`previous pen, and 82% would recommend it to
`another patient with diabetes.2 Although stud-
`ies have shown that the use of insulin pens
`causes less perceived pain and results in more
`accurate insulin dosing and fewer missed in-
`jections that the use of insulin syringes,3,4 there
`are no or few clinical trials that prove they are
`associated with improved health outcomes.
`
`
`
`FRESENIUS EXHIBIT 1074
`Page 6 of 7
`
`

`

`INSULIN DEVICE TESTING PROCESSES
`
`629
`
`CONCLUSION
`
`Additional research involving prospective
`studies is needed to determine if these new pre-
`filled pens actually improve patient compli-
`ance and glycemic control, especially with type
`2 diabetes patients who may delay starting in-
`sulin therapy. If such research were to show
`that patient’s use of insulin pens improve
`health outcomes, it may be more cost effective
`for health insurers in the United States to cover
`the cost of these new medical devices.
`
`ACKNOWLEDGMENTS
`
`Appreciation is expressed to James Ander-
`son, Jr., John Lake, and Andrew Burroughs for
`technical assistance and to David C. Klonoff,
`M.D. for editorial assistance with the manu-
`script. Eli Lilly and Company sponsored this
`work.
`
`REFERENCES
`
`1. [ISO Standards] 11608-1: Pen-injectors for medical
`use—Part 1: Pen injectors—requirements and test
`
`methods. 11608-2: Pen injectors for medical use Part 2:
`Needles—requirements and test methods 11608-3 Pen
`injectors for medical use Part 3: Finished cartridges—
`requirements and test methods. Copyright by ISO, Au-
`gust 1999.
`2. Cobo A, Ignaut D, Hultman C, Reviriego J: Patients
`and health care professionals evaluate a new pre-filled
`insulin pen device. Avances en Diabetologia 2000;
`197–202.
`3. Kadiri A, Chraibi A, Marouan F, Ababou MR, el Guer-
`mai N, Wadjinny A, Kerfati A, Douiri M, Bensouda JD,
`Belkhadir J, Arvanitis Y: Comparison of NovoPen 3
`and syringes/vials in the acceptance of insulin therapy
`in NIDDM patients with secondary failure to oral hy-
`poglycaemic agents. Diabetes Res Clin Pract 1998;41:
`15–23.
`4. Graff MR, McClanahan MA: Assessment by patients
`with diabetes mellitus of two insulin pen delivery sys-
`tems versus a vial and syringe. Clin Ther 1998;20:486–
`496.
`
`Address reprint requests to:
`Michael J. Roe, P.E., M.B.A.
`Pharmaceutical Delivery Systems
`Eli Lilly and Company
`8770 Guion Road, Suite K
`Indianapolis, IN 46268
`
`E-mail: roe michael j@lilly.com
`
`
`
`FRESENIUS EXHIBIT 1074
`Page 7 of 7
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket