`Edi1m:-.: W. Pauhl',, M. Hallc11, P.M. Rossini, J.C. Ro1hwcll
`11', J'JlJ() El:-.cvicr Scicm:c 8.V. All righh rc~crvcd.
`
`3
`
`Chapter 1
`
`The history and basic principles of magnetic nerve stimulation*
`
`Anthony T. Barker*
`
`Department of Medical Physics and Clinical Engineering, Royal Hallamshire Hospital, Sheffield SJO 2JF (UK)
`
`Introduction
`
`Magnetic nerve stimulation has rapidly become
`established as a clinical tool following its develop(cid:173)
`ment at the Royal Hallamshire Hospital and the
`University of Sheffield, and the first demonstration
`of cortical magnetic stimulation in 1985. Over 3000
`stimulators are now in use world-wide, primarily
`for diagnostic and basic research purposes, but
`with rapidly increasing interest being shown in a
`variety of therapeutic applications. This chapter
`describes the development of magnetic stimulation
`as a clinical tool and its basic principles. It also
`introduces some of the technical issues surrounding
`the design of stimulators and how they affect the
`delivered stimulus.
`
`Basic principles of magnetic nerve stimulation
`
`Ever since the work of Galvani and Volta in the
`1790s, it has been known that nerves and muscles
`can be stimulated with externally applied electrical
`
`·:.- Reproduced by permission of Arnold, London (a
`member of the Hodder Headline Group): Barker, A.T. The
`history and basic principles of magnetic nerve stimulation.
`In: Pascual-Leony, Davey, Wassermann and Rothwell
`(Eds.), Handbook of Transcranial Magnetic Stimulation.
`Arnold, London, 1999.
`* Correspondence to: Dr. A.T. Barker, Degartment of
`Medical Physics and Clinical Engineering, Royal Hallam(cid:173)
`shire Hospital, Sheffield S 10 2JF (UK).
`
`currents. Electrical stimulation, in which current is
`injected into the body through surface, needle or
`implanted electrodes, has been in widespread clin(cid:173)
`ical use for many years. In electrical stimulation
`charge is carried by electrons flowing in the wires
`to the electrodes, and is transferred to a flow of ions
`at the electrode-tissue interface. A small fraction of
`the charge on these ions is transferred onto nearby
`excitable membranes and can result in depolarisa(cid:173)
`tion.
`Electrical stimulation is very effective when
`applied
`to superficial peripheral nerves using
`surface electrodes, or to deeper structures where
`needle electrodes can be placed close to the
`required site of stimulation. However it is difficult
`to stimulate deep nerves from the surface and,
`because bone has high electrical resistance, it is
`necessary to use high voltage stimuli to penetrate
`the skull in order to achieve cortical stimulation
`(Merton and Morton 1980).
`Magnetic stimulation differs from electrical
`stimulation in that it uses a pulse of magnetic
`field to cause current to flow in the tissue. The
`mechanism of stimulation, at the cellular level, is
`the same for both techniques. In both cases charge
`flows into an excitable cell membrane, causing a
`change in transmembrane potential. This can result
`in depolarisation of the membrane and the initiation
`of an action potential, which then propagates along
`the structure by the normal nerve conduction
`mechanisms.
`
`LUMENIS EX1087
`Page 1
`
`
`
`4
`
`Magnetic stimulation is based on the scientific
`principle of electromagnetic induction, discovered
`by Michael Faraday in 1831. In one of the most
`famous of scientific experiments he wound two
`coils of wire on opposite sides of an iron ring and
`observed that, when current was turned on or off to
`one coil (called the primary), a current flowed
`briefly in the secondary coil (Fig.
`la). A few
`weeks later he showed that the iron ring, whilst
`enhancing the induction by 'guiding' the magnetic
`field from primary to secondary, was not essential,
`and he demonstrated the same effect with two
`closely positioned air-cored coils (Fig. 1 b ).
`Faraday's experiments showed that currents (and
`voltages) were only induced by a changing, or
`'time-varying' magnetic field, and not by a static
`field. In magnetic stimulation of the body, tissue
`forms the secondary circuit. The primary circuit is
`the stimulating coil, through which the stimulator
`drives current pulses, but which is not in electrical
`contact with the tissue. The magnetic field gener(cid:173)
`ated by the current flow in the coil is proportional to
`the current passing through it, and the electric field
`induced in the tissue is proportional to the rate of
`change of the magnetic field with respect to time.
`At the frequencies used in magnetic stimulation the
`magnetic field is not affected by the electrical prop(cid:173)
`erties of the body, and passes through both bone and
`soft tissue (and even clothing and air) without being
`affected by them. The magnetic field pulse induces
`
`a)
`
`b}
`
`Fig. 1. Schematic of the coil arrangements used by Michael
`Faraday (a, b) and in a modern magnetic stimulator (c).
`
`an electric field (a voltage difference between two
`points) in the tissue which causes an ionic current to
`flow (Fig. le). If the amplitude, spatial characteris(cid:173)
`tics and duration of this induced current are such
`that they cause depolarisation of a nerve membrane
`then stimulation, and the generation of an action
`potential, will occur.
`Thus the magnetic field does not itself directly
`stimulate the tissue and hence 'magnetic stimula(cid:173)
`tion'
`is a slight misnomer for the
`technique,
`although it is a convenient and well-established
`name.
`
`Development of magnetic stimulation as a
`clinical technique
`
`The first example of a physiological effect due to
`a time-varying magnetic field was reported by
`d' Arson val (1896) who observed that phosphenes
`(flicking lights 'seen' by a subject) and vertigo were
`produced when a volunteer's head was placed
`inside a coil driven at 42 Hz. This finding was
`subsequently confirmed by Silvanus P. Thompson
`(1910) and others (Fig. 2).
`The retina of the eye is the structure in man that
`is most sensitive to induced currents, and there has
`been much subsequent work on what are now
`known as magnetophosphenes. Threshold values
`for magnetophosphenes in normal subjects have
`been reported to be as low as 10 mT r.m.s. for a
`20 Hz sinusoidal magnetic field (Lovsund et al.
`1980).
`Kolin et al. (1959) demonstrated that exposed
`nerves could be stimulated by looping a frog sciatic
`nerve around the pole piece of an electromagnet
`and produced contractions in the gastrocnemius
`muscle with both 60 Hz and 1 kHz currents.
`Bickford and Freeming ( 1965) reported non(cid:173)
`invasive magnetic stimulation of human and animal
`peripheral nerves. They used a 500 Hz damped
`sinusoidal magnetic field with a peak amplitude
`of 4 Tesla decaying to zero over approximately
`40 ms, and a stimulating coil of mean diameter
`approximately 3 cm. Their choice of an oscillatory
`magnetic field waveform made it impossible to
`record nerve or muscle action potentials because
`of interference between the stimulating field and
`
`LUMENIS EX1087
`Page 2
`
`
`
`the recording equipment, and their work was not
`pursued further.
`In 1974 Barker started to investigate the use of
`short duration pulsed magnetic fields as a possible
`method of achieving velocity selective stimulation
`of human peripheral nerves. He constructed an
`electromagnet from a closed 'C' core made of
`grain-orientated mild steel of cross section 30 X
`25 mm, with a 50 mm length removed from one
`of the long sides to form an airgap. Two capacitor
`banks, each of 800 µF charged to 200 V, were
`simultaneously discharged into separate windings
`on the core, each of ten turns. The resultant current
`pulse had a peak value of 2300 A with a risetime of
`approximately 120 µs and gave a peak magnetic
`field in the core of 2 T, Clear sensation could be
`felt, and
`slight muscular contractions were
`observed, when a wrist was placed in the airgap
`(Barker 1976).
`After these early experiments Barker and collea(cid:173)
`gues decided to pursue the goal of a practical clin-
`
`Fig. 2. Silvanus P. Thompson carrying out experiments to
`induce currents in his own head.
`
`5
`
`ical magnetic nerve stimulator that used field pulses
`short enough to allow evoked nerve and muscle
`action potentials to be recorded. They changed
`their coil to an air-cored design to avoid magnetic
`saturation (most ferromagnetic materials saturate at
`fields of approximately 2 T), and to give a relatively
`small and light structure which could be moved
`readily around the body. In 1982 they reported the
`stimulation of superficial nerves at the wrist and the
`recording of supramaximal evoked potentials over
`the thenar eminence using a 500 V, 6000 µF capa(cid:173)
`citor discharge system with a risetime of 180 µs
`(Polson et al. 1982). Supramaximal responses
`were achieved with the capacitor bank discharged
`from 340 V, giving a peak coil current of 6800 A and
`peak field in the centre of the coil of 2.2 T. (Fig. 3).
`Using a new high voltage design of stimulator
`which was more efficient at transferring energy
`from the storage capacitor to the coil (transfer effi(cid:173)
`ciency having been improved from -20% to -80%),
`the Sheffield group achieved the first magnetic
`stimulation of the human motor cortex (Barker et
`al. 1985a). Placing a coil of 100 mm outside
`diameter centrally over the vertex of a normal
`subject they demonstrated clear hand movements
`and recorded evoked muscle action potentials
`from Abductor Digiti Minimi using surface elec(cid:173)
`trodes (Figs. 4 and 5).
`There was no pain or discomfort associated with
`this first cortical magnetic stimulation, a consider(cid:173)
`able contrast to the sensations felt with electrical
`stimulation using electrodes placed on the scalp.
`Public demonstrations of cortical magnetic stimula(cid:173)
`tion using the prototype stimulator on volunteers at
`the 11th International Congress of Electroencepha(cid:173)
`lography and Clinical Neurophysiology in London
`and at the Physiological Society in Oxford (Barker
`et al. 1985b,c) caused considerable interest. In
`response to a number of requests, the first stimula(cid:173)
`tors specifically for routine clinical use were
`designed and built in the department of Medical
`Physics and Clinical Engineering at the Royal
`Hallamshire Hospital, Sheffield, for five groups in
`the U.K. and the U.S.A., who wished to evaluate the
`technique (Fig. 6).
`Early clinical studies using magnetic stimulation
`were presented in 1985 and 1986 (Barker et al.
`
`LUMENIS EX1087
`Page 3
`
`
`
`6
`
`Fig. 3. Dr. M.J.R. Polson with the stimulator used to produce supramaximal peripheral responses in 1982.
`
`1985d, 1986). Interest in magnetic stimulation grew
`quickly and, to expedite its adoption as a clinical
`tool, the Sheffield group introduced a number of
`manufacturers
`to
`the
`technique during 1985.
`Commercial stimulators from at least three manu(cid:173)
`facturers are in widespread use at present. The basic
`theory, advantages and safety considerations of
`magnetic stimulation, along with the first detailed
`clinical evaluation of the technique in the study of
`multiple sclerosis and motor neurone disease, were
`described in 1987 (Barker et al. 1987).
`
`Characteristics of the output field pulses from a
`magnetic nerve stimulator
`
`A magnetic field pulse of peak amplitude in the
`range 1-4 T is usually require to stimulate nerves,
`the exact value depending on a number of para(cid:173)
`meters such as the stimulating coil geometry, the
`depth and path of the nerve, and the local anatomy.
`The
`first generation of magnetic stimulators
`produced these pulses at relatively low maximum
`repetition rates, typically one pulse every few
`seconds
`('standard
`rate'
`stimulators). Their
`magnetic field pulses were monophasic, with rise(cid:173)
`times of order 100 µs, decaying back to zero over
`about 800 µs. This was achieved by discharging a
`
`capacitor, previously charged to a high voltage, into
`a stimulating coil via an electronic switch called a
`thyristor (Fig. 7).
`The coil current reaches its peak after a time
`determined by the value of the storage capacitor
`and by the inductance of the stimulating coil. The
`current then decays back to zero, at a rate deter(cid:173)
`mined by the coil inductance and value of resistor,
`
`Smsec
`..........,..r--,t
`
`Fig. 4. Upper trace: response due to a single magnetic
`stimulus applied to the motor cortex. Lower trace: response
`due to a single magnetic stimulus applied to the ulnar nerve
`at the elbow (from Barker et al., 1985a).
`
`LUMENIS EX1087
`Page 4
`
`
`
`7
`
`Fig. 5. From left to right: Dr. R. Jalinous, Prof. I.L. Freeston and Dr. A.T. Barker with the magnetic stimulator used to achieve
`the first cortical responses in February 1985.
`
`R, The stored energy, which starts in the capacitor
`as electrical energy, is transfeITed to the coil as
`magnetic energy as the coil cuITent increases, and
`then returns to be dissipated as heat in resistor R as
`the magnetic field collapses The electric field and
`cuITent induced in the body is proportional to the
`rate of change of magnetic field (Fig. 8), which in
`turn is propo1tional to the coil cuITent, just as it was
`in the secondary coil of Faraday's classic experi(cid:173)
`ments on electromagnetic induction.
`The first stimulators were designed to produce
`the waveforms of Fig. 8 because the resultant
`signals induced in tissue were similar in time course
`to those from charge-balanced electrical stimula(cid:173)
`tors, whose effectiveness was well proven. They
`have the. technical advantage that the reverse
`voltage, which appears on the storage capacitor
`during the decay of the coil cuJTent, is limited to
`about 20% of the discharge voltage, thus length(cid:173)
`ening the life of the capacitor. Typical maximum
`values found in such stimulators are: capacitor
`discharge voltage 3 kV, stored energy 500 J, peak
`coil CUJTent 8 kA and maximum stimulusrepetition
`rate 0.3 Hz at full output. Some standard rate stimu-
`
`lators have been designed to produce an oscillatory
`output similar to that of the rapid rate devices
`described below. Standard rate stimulators continue
`to be widely used, and are ideally suited to diag(cid:173)
`nostic and research applications where responses to
`single stimuli are to be studied.
`
`Rapid rate stimulators
`Widespread interest has recently been shown in
`the use of rapid rate stimulators, capable of produ(cid:173)
`cing tens of pulses per second in bursts lasting up to
`about one minute, primarily for therapeutic appli(cid:173)
`cations. They are usually based on a variation of the
`circuit shown in Fig. 7, with the resistor omitted and
`the diode repositioned across the thyristor switch.
`This produces an oscillatory magnetic field output,
`as opposed to the monophasic one of the standard
`rate stimulator (Fig. 9).
`Oscillatory output stimulators have two main
`advantages, both relating to their energy require(cid:173)
`ments. Firstly, at the end of the stimulating pulse
`(the complete waveform shown in Fig. 9), approxi(cid:173)
`mately 40% of the original energy stored on the
`capacitor has returned to it. Thus only 60% of the
`
`LUMENIS EX1087
`Page 5
`
`
`
`8
`
`Fig. 6. The first clinical magnetic stimulator in use, Shef(cid:173)
`field, November, 1985.
`
`original stored energy needs to be supplied to the
`capacitor in order to recharge it in readiness for the
`next stimulus. A variety of configurations which
`can be used to achieve such energy saving have
`been described by Jalinous (1988).
`Secondly, oscillatory waveforms achieve stimu(cid:173)
`lation at lower levels of peak magnetic field than
`monophasic ones, and hence less energy is required
`to be stored in the stimulator. McRobbie and Foster
`( 1984) investigated the effects of different magnetic
`stimulator waveforms on the stimulation threshold
`of the median nerve in the forearm. They found
`that, using an oscillatory waveform of period 280
`µs (approximately the same as that in Fig. 9), only
`59% of the peak rate of change of magnetic field of
`a monophasic waveform was required to achieve
`the same level of stimulation. This would suggest
`that only about 35% of the stored energy of a mono(cid:173)
`phasic waveform stimulator is required for an oscil(cid:173)
`latory stimulator.
`
`Wada et al. (1996) investigated the effect of
`stimulator waveform both analytically, using the
`Frankenhaeuser-Huxley membrane model, and
`experimentally using
`the
`frog
`sciatic nerve.
`Comparing a monophasic waveform with a one(cid:173)
`cycle oscillatory waveform of period 280 µs they
`found the latter required 67% of the storage capa(cid:173)
`citor voltage (equivalent to 44% of the stored
`energy) to achieve the same response in their
`analytic model. With their sciatic nerve experiment
`they obtained similar results when using a stimu(cid:173)
`lating waveform of period 0.8 ms.
`The physical mechanism which causes nerves to
`be more sensitive to oscillatory waveforms has yet
`to be explored in detail, but may be related to
`membrane time constant (Wada et al. 1996) and
`membrane non-linearity (Reilly 1992).
`
`The effect of magnetic field risetime
`For a given stimulating coil in a fixed position
`relative to the nerve of interest, and assuming that
`the nerve membrane is regarded as a lossless capa(cid:173)
`citor, it can be shown that the charge transferred to
`that membrane is proportional to both the peak
`magnetic field and the square root of the magnetic
`energy in the coil, irrespective of the risetime of the
`magnetic field (Barker et al. 1987). In practice,
`nerve membranes do not behave like lossless capa(cid:173)
`citors. Leakage resistance within the membrane
`causes any applied charge to decay with time,
`resulting in the familiar stimulus 'strength-duration
`curve' seen with electrical stimulation. The effects
`of pulse duration are well known in the context of
`electrical
`stimulation, but apply equally
`to
`magnetic stimulation. A magnetic field pulse that
`
`S1
`
`energy
`storage
`capacitor
`
`D
`
`R
`
`Fig. 7. Simplified schematic diagram of a standard rate
`magnetic nerve stimulator.
`
`LUMENIS EX1087
`Page 6
`
`
`
`9
`
`KT : : ; ; , · · :
`
`• . ...... ; ................ : .......... j .......... ; ................ , ................ ; ................ ; ................ ; .......... ..!. ............. .
`jl r
`
`5
`
`4
`
`3
`
`2
`
`1
`
`0
`
`-1
`
`'iii
`~
`C
`::,
`~
`
`l'G ... -:e
`..!!. -~ "ti
`
`"ti
`C
`l'G
`al
`
`-2 ~~~-~~~~~~-~~~-~~~-~~-~~~-~~~'----'
`0.4
`0.6
`-0.1
`0.1
`0.2
`0.3
`0.5
`0.7
`0.8
`0.9
`1
`0
`time msec
`
`Fig. 8. The time course of the magnetic field measured at the centre of a stimulating coil driven by a standard rate magnetic
`stimulator (Magstim 200), and the resultant induced electric field waveform.
`
`rises rapidly to its peak allows less time for charge
`to leak away, and hence achieves depolarisation at
`lower peak fields and magnetic energies. This has
`practical advantages because it decreases
`the
`energy storage requirements of the stimulator, the
`
`heat dissipated in the coil and the stimulator energy
`consumption, the latter being primarily a problem
`for fast repetition rate stimulators.
`The effect of field risetime on the efficiency of
`stimulation has been experimentally quantified
`
`5
`
`4
`
`'iii 3 -·c ::, 2
`~ e 1
`l'G --:E
`~ -e 0
`
`-1
`
`al
`"ti
`"ti
`C
`l'G
`al -3
`
`-2
`
`-4
`-5 L-----..1.--'-...1-_....__,L. _ __, _____ ..._......_....i...._.___.___.___._ _______ ....._....._....___.__.
`-0.1
`0
`0.1
`0.2
`0.3
`0.4
`0.5
`0.6
`0.7
`0.8
`0.9
`1
`time msec
`
`Fig. 9. The time course of the magnetic field measured at the centre of a stimulating coil driven by a rapid rate magnetic
`stimulator (Magstim Rapid), and the resultant induced electric field waveform.
`
`LUMENIS EX1087
`Page 7
`
`
`
`10
`
`using a stimulator with an output waveform having
`the same shape as that of Fig. 8 but whose timescale
`could be varied. Shorter rise time (-60 µs) pulses
`were shown to need approximately half the stored
`energy of those with longer risetimes (- 180 µs)
`(Fig. 10).
`The use of a stimulator with two or more
`magnetic field risetimes enables membrane time(cid:173)
`constant to be measured and the first measurements
`of this type have shown values of approximately
`150 µs for both peripheral and cortical stimulation
`(Barker et al. 1991 ). Non-invasive determination of
`membrane timeconstant has yet to be systemati(cid:173)
`cally studied in man, using either magnetic or elec(cid:173)
`trical stimulation, but such measurements may
`contain useful clinical information relating to the
`condition of the nerve membrane.
`
`The mechanism of stimulation - the 'activating
`function'
`
`The site of stimulation of a nerve fibre is that
`point along its length at which sufficient current
`to cause depolarisation passes out of the axon and
`through its membrane. When using electrical stimu(cid:173)
`lation with surface electrodes, stimulation
`is
`normally assumed to occur under the cathode elec(cid:173)
`trode, and this is a reasonable approximation in the
`case of a long, relatively straight, superficial nerve.
`In the case of magnetic stimulation the situation is
`less well defined and depends on a number of
`factors.
`For both electrical and magnetic stimulation the
`'activating function', which causes transmembrane
`current to flow and hence stimulation to occur, can
`be described mathematically as 'the spatial deriva(cid:173)
`tive of the electric field along the nerve' (see for
`example Reilly 1992; Abdeen and Stuchly 1994;
`Garnham et al. 1995). This can best be visualised
`by considering a long straight nerve in a homoge(cid:173)
`neous volume. Magnetic stimulation induces an
`electric field in tissue. If this electric field is
`uniform, and parallel to the nerve, it will cause
`current to flow both inside and outside the fibre
`but not across its membrane (Fig. I la). However,
`electric current is a continuous function, in other
`words the current entering any volume has to be
`
`1.8
`
`1.6
`
`1.4
`
`1.2 -
`
`0.8
`
`0.6
`
`0.4
`
`0.2 -
`
`I Data from 10 normal subjects ± 1 S.D.
`
`>, QI
`0111
`._ C
`QI 0
`CC.
`QI Ill
`
`,:, ! e,:, o-.. 0
`
`Ill .C
`QI Ill
`> QI
`.:;~
`ca ..
`a, 0
`a::-
`
`0 W ~ ~ M 1001W1~1~1MW0
`Risetime to Speak 1,1sec
`
`Fig. 10. The relative stored energy required to achieve
`threshold cortical stimulation versus magnetic field risetime.
`(from Barker, 1991 ).
`
`the same as that leaving it. Thus if the current
`within the axon changes along its length, that
`change must pass through the membrane and can
`cause stimulation (Fig. 11 b ).
`
`a)
`
`- - - -
`- - - -
`- - - -
`- - - -
`- - - -
`
`b)
`
`- -
`-!J
`- -
`- - 7
`- - -:)
`- -
`
`4
`4
`
`Fig. 11. Diagrammatic representation of (a) the cun-ent flow
`due to a uniform electric field parallel to a nerve fibre,
`causing no transmembrane cun-ent, (b) the cun-ent flow
`due to an electric field varying along the length of a nerve
`fibre nerve, resulting in transmembrane cun-ent, (c) a
`uniform electric field and a bent nerve also resulting in a
`transmembrane cun-ent.
`
`LUMENIS EX1087
`Page 8
`
`
`
`ll
`
`V/m
`
`0 06
`
`0 04
`
`0 02
`
`0
`
`-0 02
`
`-0 04
`
`.o 06
`
`.Q 08
`
`.o 08
`
`.o 06
`
`.o 04
`
`.o 02
`
`o o 02 o 04
`
`0.06 o os o 1 m
`
`Fig. 12. The distribution of induced electric field, 20 mm below a 66.5 mm mean diameter circular coil.
`
`A change of electric field along the axon is
`required to change the current flowing within it,
`and hence the activating function is proportional
`to the rate of change of electric field, otherwise
`known as its spatial derivative, along the nerve.
`The situation becomes more complex in the case
`of a bent nerve. In this case a spatially uniform
`electric field can cause stimulation. Current flow
`will again be approximately parallel to the electric
`field. Where the axon bends across the field, the
`current will tend to continue in a straight line and
`pass out of the fibre across the membrane. Again
`the relevant parameter is the spatial derivative of
`electric field along the nerve, but in this case the
`bend in the nerve causes it to be non-zero, as shown
`in Fig. I le (Maccabee et al. 1993; Abdeen and
`Stuchly 1994).
`
`Coil geometry and the site of stimulation
`
`focused at a point. The physics of magnetic fields
`causes them to diverge after they leave their source
`and hence it is not possible to focus magnetic fields
`at a distance using only a single coil.
`The geometry of the stimulating coil is just one
`of the factors, along with others such as nerve
`geometry and local variations in tissue conduc(cid:173)
`tivity, which determines the amplitude and spatial
`distribution of the activating function. This in turn
`will determine the site of stimulation. The effect of
`coil geometry can be illustrated by considering the
`induced fields from the two most commonly used
`coil shapes, circular and figure-of-eight. In the
`following analysis it has been assumed, for simpli(cid:173)
`city, that nerve is situated within a homogeneous
`semi-infinite volume conductor, and parallel to the
`surface on which
`the coil rests. The results
`presented below can be extrapolated to coils having
`the same geometry but different dimensions by
`appropriate scaling of the spatial axes.
`
`Magnetic stimulation does not, in general, give a
`small and predicable site of stimulation._Ihe term
`'focal' stimulation is sometimes used, but this is
`misleading because it implies that the stimulus is
`
`The circular coil
`The first magnetic stimulator coils to be used
`were circular, and this is still the most widely
`used shape. Circular coils have the advantage that
`
`LUMENIS EX1087
`Page 9
`
`
`
`12
`
`m
`
`0.08
`
`0.06
`
`004
`
`002
`
`0
`
`-002
`
`-0 04
`
`-006
`
`-0 08
`
`V/m2
`2500
`
`2000
`
`1500
`
`1000
`
`500
`
`0
`
`-500
`
`-1000
`
`-1500
`
`-2000
`
`-2500
`
`-0.1
`
`-0 08
`
`-0 06
`
`-0.04
`
`-0 02
`
`o o 02 o 04 o 06 o 08
`
`o 1 m
`
`Fig. 13. The rate of change of electric field, calculated in the direction of the shown nerve, 20 mm below a 66.5 mm mean
`diameter circular coil.
`
`they are relatively easy to construct and may conve(cid:173)
`niently be positioned over many parts of the body.
`Their primary disadvantage is the relative uncer(cid:173)
`tainty as to the exact site of stimulation.
`In a semi-infinite homogeneous medium the
`induced electric field and current loops will be
`circular and concentric. Fig. 12 shows this circular
`distribution of the amplitude of induced electric
`field, calculated in a plane at a depth of 20 mm
`below a realistic model of a commercial coil
`(d//dt = 108 Ns, inside tum diameter 41.5 mm,
`outside
`tum diameter 91.5 mm, number of
`turns= 15).
`As previously discussed, the activating function
`is the rate of change of electric field along the
`nerve. Fig. 13 shows this function calculated
`along a straight nerve positioned in the direction
`shown.
`A nerve positioned as shown relative to the coil
`will experience an induced transmembrane current
`which will tend to depolarise it in region A and
`hyperpolarise it in region B. These regions can be
`thought of, by analogy with electrical stimulation,
`as a 'virtual cathode' and a 'virtual anode' respec-
`
`tively. If the nerve of Fig. 13 had a sharp bend
`somewhere between these regions, the rate of
`change of electric field along the nerve would be
`increased at the bend and the site of stimulation
`would tend to shift to it.
`It should also be borne in mind that, because of
`circular symmetry, all structures which have the
`same stimulation threshold and which lie tangential
`
`Approximate
`-site of
`stimulation
`
`--Coil
`
`Fig. 14. Schematic representation of the site of stimulation
`for three nerves lying at different angles to a circular stimu(cid:173)
`lating coil.
`
`LUMENIS EX1087
`Page 10
`
`
`
`m
`
`0.15 .-------------......i.---""-----'-----1-
`
`0.1
`
`0.05
`
`0
`
`-005
`
`-0.1
`
`13
`
`Vim
`
`160
`
`140
`
`120
`
`100
`
`20
`
`-0.1
`
`-0.05
`
`0
`
`0.05
`
`0.1
`
`o.1sm
`
`Fig. 15. The electric field amplitude calculated in a plane 20 mm below a 73 mm mean diameter figure-of-eight coil.
`
`to a loop of given radius will be equally likely to be
`stimulated. This can be seen in Fig. 13 by the
`presence of an inverted virtual anode-cathode pair
`on the other side of the coil, and schematically in
`Fig. 14 which shows the approximate site of stimu(cid:173)
`lation of three nerves lying at different angles to the
`coil. Additionally, as
`the stimulus strength is
`increased, the site of stimulation will move further
`along the nerve, away from the coil, due to the
`shape of the 'activating function' (Garnham et al.
`1995).
`A practical example of the effect of the inverted
`virtual anode-cathode pair shown in Fig. 13 can be
`observed by stimulating using a 40 mm mean
`diameter coil placed centrally over the wrist. Both
`the ulnar and median nerve will be stimulated
`simultaneously, but at different positions along
`their lengths. Uncertainty as to the site of stimula(cid:173)
`tion can be decreased by using smaller coils, but
`there are considerable
`technical problems
`in
`making coils below approximately 25 mm mean
`diameter that can withstand the energies required,
`and that do not rapidly overheat. Depth or-penetra(cid:173)
`tion and lack of discomfort, which are two of the
`
`main advantages of magnetic stimulation when
`compared to conventional electrical stimulation,
`also decrease with smaller coils.
`In order to maximise the electric field induced in
`the tissue, stimulating coils are usually placed with
`one of their flat surfaces parallel to the skin as shown
`in Fig. le. It is also possible to achieve stimulation
`by placing a circular coil with its surface perpendi(cid:173)
`cular to the skin and parallel to the nerve of interest.
`This has the advantage of improving selectivity but
`has the disadvantage of very poor coupling of the
`magnetic field to the tissue. As a result the depth of
`penetration is greatly decreased and much higher
`stimulating energies are required. For these reasons
`this orientation is rarely used.
`Finally, perhaps the most common misconcep(cid:173)
`tion about magnetic stimulation is that the site of
`stimulation is underneath the centre of a circular
`coil. Figs. 12 and 13 show that both the electric
`field, and its spatial derivative, are zero under the
`centre of the coil. Hence, unless major tissue inho(cid:173)
`mogeneities causing gross distortion of the electric
`field patterns are nearby, this is the position where
`stimulation is least likely to occur.
`
`LUMENIS EX1087
`Page 11
`
`
`
`14
`
`The jigure-o_feight coil
`
`The only other coil geometry that is widely used
`at present is the figure-of-eight configuration first
`proposed by Ueno et al. (I 988). Two circular coils
`are placed side by side, and connected such that the
`current from the stimulator in one coil rotates in the
`opposite direction to that in the other. The advan(cid:173)
`tage of this geometry is that it can, under some
`conditions, decrease the uncertainty as to the site
`of stimulation. Fig. 15 shows the distribution of
`electric field loops calculated beneath a realistic
`model of a commercial
`figure-of-eight coil
`(d//dt = 108 Als, inside tum diameter 56 mm,
`outside tum diameter 90 mm, number of turns per
`winding = 9).
`The electric field (and current) loops induced in
`the tissue can be seen to add below the position
`where the two coils approach each other, but not
`elsewhere. This results in electric fields under the
`centre of the figure-of-eight coil typically twice
`those elsewhere under the coil. Fig. 16 shows the
`rate of change of electric field in a direction parallel
`
`to the mid-line between the two coils, calculated for
`the same plane.
`In contrast to the circular geometry of Fig. 13, the
`virtual cathode and anode regions (A and B, respec(cid:173)
`tively), where the rate of change of electric field are
`a maximum and minimum, occur only on the
`midline between the two coils for a nerve in this
`direction. Thus stimulation is more likely to occur
`on the centreline of this configuration than else(cid:173)
`where. The secondary peaks (C and D) which
`occur at the sides of the figure-of-eight coil are
`typically about half the amplitude of the central
`peaks. As is the case with circular coils, the site
`of stimulation of a nerve on the midline of this
`configuration and having a sharp bend near the
`centre of the coil, will tend to be at that bend
`(Abdeen and Stuchly 1994 ). Other examples of
`electric field distributions from both circular and
`figure-of-eight coils have been presented by Jali(cid:173)
`nous (1991 ).
`Figure-of-eight coils have been used to map
`cortical representation (see for example Levy et
`al. 1991) and are effective in stimulating structures

Accessing this document will incur an additional charge of $.
After purchase, you can access this document again without charge.
Accept $ ChargeStill Working On It
This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.
Give it another minute or two to complete, and then try the refresh button.
A few More Minutes ... Still Working
It can take up to 5 minutes for us to download a document if the court servers are running slowly.
Thank you for your continued patience.

This document could not be displayed.
We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.
You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.
Set your membership
status to view this document.
With a Docket Alarm membership, you'll
get a whole lot more, including:
- Up-to-date information for this case.
- Email alerts whenever there is an update.
- Full text search for other cases.
- Get email alerts whenever a new case matches your search.

One Moment Please
The filing “” is large (MB) and is being downloaded.
Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!
If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document
We are unable to display this document, it may be under a court ordered seal.
If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.
Access Government Site