RESEARCH ARTICLE
`
`www.advintellsyst.com
`
`Hardware-Mappable Cellular Neural Networks for
`Distributed Wavefront Detection in Next-Generation
`Cardiac Implants
`
`Zhuolin Yang, Lei Zhang, Kedar Aras, Igor R. Efimov, and Gina C. Adam*
`
`Artificial intelligence algorithms are being adopted to analyze medical data,
`promising faster interpretation to support doctors’ diagnostics. The next frontier
`is to bring these powerful algorithms to implantable medical devices. Herein, a
`closed-loop solution is proposed, where a cellular neural network is used to
`detect abnormal wavefronts and wavebrakes in cardiac signals recorded in
`human tissue is trained to achieve >96% accuracy, >92% precision, >99%
`specificity, and >93% sensitivity, when floating point precision weights are
`assumed. Unfortunately, the current hardware technologies for floating point
`precision are too bulky or energy intensive for compact standalone applications in
`medical implants. Emerging device technologies, such as memristors, can
`provide the compact and energy-efficient hardware fabric to support these efforts
`and can be reliably embedded with existing sensor and actuator platforms in
`implantable devices. A distributed design that considers the hardware limitations
`in terms of overhead and limited bit precision is also discussed. The proposed
`distributed solution can be easily adapted to other medical technologies that
`require compact and efficient computing, like wearable devices and lab-on-chip
`platforms.
`
`1. Introduction
`
`Machine learning (ML) algorithms are being adopted to analyze
`medical data in specialties like radiology, oncology, and cardiol-
`ogy, promising faster interpretation with accuracy close to
`doctors’ diagnostics.[1] The next
`frontier
`in computing
`
`Z. Yang, L. Zhang, G. C. Adam
`Department of Electrical and Computer Engineering
`The George Washington University
`Washington, DC 20052, USA
`E-mail: ginaadam@email.gwu.edu
`
`K. Aras, I. R. Efimov
`Department of Biomedical Engineering
`The George Washington University
`Washington, DC 20052, USA
`
`The ORCID identification number(s) for the author(s) of this article
`can be found under https://doi.org/10.1002/aisy.202200032.
`
`© 2022 The Authors. Advanced Intelligent Systems published by Wiley-
`VCH GmbH. This is an open access article under the terms of the
`Creative Commons Attribution License, which permits use, distribution
`and reproduction in any medium, provided the original work is
`properly cited.
`
`DOI: 10.1002/aisy.202200032
`
`technology is to bring these powerful
`algorithms to implantable medical devices,
`which requires automation of real-time
`life-saving therapeutic decisions without
`the physician's presence. An example is
`the need for improved medical solutions
`for life-saving cardiac defibrillation thera-
`pies, that can detect bioelectric anomalies
`(e.g., cardiac arrhythmias) and act on this
`data locally for real-time therapy delivered
`within tens of seconds or minutes since
`the onset of
`life-threatening ventricular
`fibrillation (VF). The statistics put this chal-
`lenging technological need in perspective:
`ventricular arrhythmias such as VF are
`responsible
`for over 700 000 sudden
`cardiac deaths a year in the USA and
`Europe.[2] VF is a common, life-threatening
`arrhythmia
`characterized
`by
`chaotic
`asynchronous electrical activity of
`the
`cardiac muscle, which results in death
`within 10 minutes.
`Individual differences in physiological
`mechanisms, anatomic and genetic deter-
`minants, and etiologies of various arrhythmias impact the course
`of treatment. Ablation therapy, while promising, remains a work
`in progress. Therefore, on average, defibrillation therapy deliv-
`ered by implantable cardioverter defibrillators (ICDs) remains
`the most effective treatment as antiarrhythmic drugs have lim-
`ited efficacy and can be associated with adverse side effects.
`Implants have to be biocompatible, organ conformal, and small
`enough to minimize the tissue damage and be capable of inde-
`pendent autonomous operation without external intervention.
`Low power is an essential characteristic to avoid the heat damage
`to the tissue and prolong the lifetime of the embedded battery for
`many years without recharging.[3] Currently, most volume of the
`ICD has been occupied by batteries, which has limited the
`volume reduction and the computing capacity. ICD has local
`computing based on a microprocessor to detect and differentiate
`arrhythmia to offer different
`treatments, but
`the resolution
`provided by ICD is really low typically limited to only one or a
`couple of sensors; as such, the ability to detect arrhythmia
`wavefronts is non-existent. The data can be read wirelessly by
`the physician during periodic checkups. Increasing the sensing
`resolution is desired but the local computing capacity has to also
`be increased which is difficult due to power constraints. Wireless
`data transmission for processing of data outside of the body is
`not a viable solution either, as real-time data transfer between
`
`Adv. Intell. Syst. 2022, 2200032
`
`2200032 (1 of 16)
`
`© 2022 The Authors. Advanced Intelligent Systems published by Wiley-VCH GmbH
`
`1
`
`APPLE 1065
`Apple v. AliveCor
`IPR2021-00972
`
`

`

`www.advancedsciencenews.com
`
`www.advintellsyst.com
`
`the implant and the external world requires a significant amount
`of power, even increasing the volume of the implant while intro-
`ducing delays and additional security risks. That is why these sys-
`tems have focused mostly on local real-time signal processing.
`However, due to the low resolution for sensing and therapy,
`high-energy biphasic shocks are needed to effectively terminate
`life-threatening high-frequency arrhythmias. These high-energy
`shocks can lead to myocardial damage and associated comorbid-
`ities and it is a painful and traumatic experience for patients,
`especially when delivered inappropriately when arrhythmia is
`not present due to poor sensing. On the other hand, multipulse
`therapy (MPT) utilizes well-timed trains of low-energy electric
`pulses. Experiments on animal and human heart tissue showed
`that when appropriately timed, MPT significantly decreases the
`high-energy defibrillation threshold by almost an order of mag-
`nitude. Moreover, recent first-in-human clinical trial demon-
`strated safety and efficacy of MPT in patients with atrial
`fibrillation,[4] which is not possible with current high-energy
`ICDs due to pain and discomfort caused by high-energy shocks.
`However, as it is administered also through transvenous leads,
`the issue of low resolution remains.
`To study the mechanisms of arrhythmias and develop suitable
`MPT for clinical use, high-definition electrically or optically
`mapped electrocardiograms (ECG) data must be used, which
`requires a large number of sensors to map the cardiac tissue sur-
`face. High-definition ventricular arrhythmia sensing integrated
`with electrotherapy is an emerging concept enabled by organ-
`conformal electronics platforms. Prototype organ-conformal
`electronic platforms have been developed with noncontact
`sensors and actuators and tested in vivo[5] but have limited
`resolution. Increasing the density of sensors and actuators is
`underway,[6] promising a personalized electrotherapy solution
`to terminate life-threatening tachycardias with two orders of
`magnitude less energy than a typical shock.[7] Such platforms
`could be used to predict fibrillatory wavefronts and enable their
`prevention using high-definition sensing and ultralow-energy
`electrotherapy that does not cause pain and discomfort.
`The high definition is a critical requirement as multiple rotors
`can be simultaneously present in the myocardium[8] during an
`arrhythmia event and generate the seemingly chaotic pattern
`on the electrocardiogram that is the hallmark of atrial and
`ventricular fibrillation. The ventricular fibrillation rotors can
`be identified based on individual wavefronts, and wavebreaks
`are represented by phase singularities. The wavefront is defined
`as isolines of the phase that terminate either at boundaries or at
`singular points with the phase field (phase singularities[9]).
`Although the exact data resolution needed to extract these chaotic
`wavefronts is still under investigation, we estimated that >10 000
`sensors, sampled at 500 Hz with 12-bit digitization, can pro-
`duce an accurate map for the entire human heart. Such a system
`would produce >60 MB s1 of data which must be processed in
`milliseconds, an insurmountable task for serial computation,
`especially on microprocessors of miniature implantable devices
`with limited energy resources. Real-time smart and energy-
`efficient computation is needed to process the data and trigger
`the local activation of actuators. To our knowledge, no organ
`conformal electronics platform has embedded computing
`for local data interpretation and millisecond decision-making,
`
`as needed for real-time life-saving therapy such as arrhythmia
`electrotherapy.
`In this work, we propose the use of distributed computing
`neural network algorithms which are hardware mappable, to pro-
`vide high classification sensitivity, specificity, accuracy, and pre-
`cision in determining the challenging spatiotemporal dynamics
`of cardiac electrical signals. Artificial neural networks can pro-
`cess a large amount of data in a parallel fashion and “learn”
`its patterns. As their name suggests, artificial neural networks
`are inspired by biological brain and can provide intelligent
`computing solutions. Deep learning techniques, such as
`convolutional neural networks, have been demonstrated to
`perform with >93% accuracy for the classification of ECG
`heartbeats.[10–12] These complex networks can be used for clas-
`sification of heartbeat by heartbeat of data obtained from bedside
`ECG recording equipment, but they have yet to be applied in cur-
`rent low-resolution ICDs that shock the entire heart due to
`computational complexities and limited microprocessor capabil-
`ities.[13] However,
`for new types of high-definition organ-
`conformal platforms, they are impractical to physically realize
`due to their complexity for a large number of recording channels
`and also unsuitably centralized for the spatiotemporal tracking of
`wavefronts and wavebreaks as needed for precise therapy by dis-
`tributed electric field. To our knowledge, no neural network algo-
`rithm has been proposed for the identification of wavefronts.
`This work describes a distributed computing algorithm based
`on cellular neural networks that is readily mappable to memristor-
`based hardware circuitry and could enable a closed-loop solution
`
`Figure 1. Distributed computing for electrical wavefront determination:
`Proposed technology using integrated network of sensors, computing
`chiplets distributed in a cellular neural network architecture, and actuators
`that will allow high-definition mapping, interpretation, and therapeutic
`response in a closed-loop fashion.
`
`Adv. Intell. Syst. 2022, 2200032
`
`2200032 (2 of 16)
`
`© 2022 The Authors. Advanced Intelligent Systems published by Wiley-VCH GmbH
`
`2
`
`

`

`www.advancedsciencenews.com
`
`www.advintellsyst.com
`
`2. Experimental Section
`
`2.1. Data Gathering and Preprocessing
`
`This study utilized representative data obtained from a deiden-
`tified donor human heart
`from the Washington Regional
`Transplant Community (Church Falls, VA). The study was
`approved by the Institutional Review Board at
`the George
`Washington University.
`The experimental apparatus and procedures are explained in
`detail in the study by Aras et al.[23] Briefly, the ventricular tissue
`was prepared as a wedge with average dimension of 7 cm  3.5 cm
`(Figure 2a). The tissue was then mounted in a temperature-
`controlled, pressure-controlled,
`and an oxygenated optical
`mapping setup (Figure 2a). Optical action potentials were
`mapped from 7 cm  7 cm field of view using a MiCAM05
`(SciMedia, CA) CMOS camera (100  100 pixels) and sampled
`at 1 KHz sampling rate.
`The dataset consisted of 1000 optical mapping images of the
`epicardium tissue recorded at 1 kHz sampling rate with a size of
`100  50 pixels. 800 images were used for training and 200 for
`testing. The dataset included complete recordings of several
`fibrillation events, enabling the analysis of various wavefront pat-
`terns during fibrillation as part of this work. Optical recordings
`were used because they provide higher-resolution mapping than
`flexible electrode arrays. However, these results were directly
`applicable to electrically recorded data, as shown in Figure 2b,c.[24]
`Studies into the resolution required to extract any possible chaotic
`rotors in human tissue are still under investigation and higher-
`resolution setups are being developed.
`Analysis in the phase domain was typically done for such stud-
`ies, as the wavefront propagation and the singularities could be
`easily detected in the phase domain. The time domain optical raw
`data recorded by the cameras was preprocessed to transform it
`into the phase domain with a scale between π and π through the
`Hilbert transform.[25] The Hilbert transform is an efficient signal
`analysis method for nonstationary time series, especially in deter-
`mining the instantaneous frequency of time-varying signals,
`such as ventricular arrhythmias. Detection of these subtle fre-
`quency changes and potentially recognizing the initiation and/
`or termination of VT/VF is very important in understanding
`the mechanisms of arrhythmia. Given a real-time function
`x(t), its Hilbert transform was defined as[26]
`
`(1)
`
`xðτÞ
`t τ
`
`dτ
`
`Z þ∞
`
`∞
`
`1 π
`

`
`1 π
`
`t
`
`bxðtÞ ¼ H½xðtފ ¼ xðtÞ
`
`Figure 3a shows a raw optical signal and Figure 3b shows its
`phase-domain equivalent that was further preprocessed before
`looking at
`the wavefront. More details are presented in
`prior work.[23] A wavefront was located at the edge of phase
`∅ðtÞ ¼ π (red) and phase ∅ðtÞ ¼ π (blue) on the blue side.
`The wavefronts were labeled manually because the noise and
`the undesirable artifacts of the pacing electrode might affect
`the precision of
`the labels and affect
`the training results
`afterwards. For each data sample, a corresponding phase map
`3b and its wavefront mapping 3c served as input and desired
`output, respectively, for the neural network core.
`
`that includes spatially distributed sensing, data processing, and
`any required actuation for therapy (Figure 1). The cellular neural
`network maps well to a spatiotemporally distributed architecture
`and would enable a high-speed high-data-throughput computing
`solution. Any other type of neural network, for example, a multi-
`layer perceptron or a convolutional neural network, would require
`hardware implementation in a single chip which would have to be
`connected to a multitude of sensors and actuators, with density
`limitations due to the interconnects. This proposed cellular neural
`network architecture was chosen as most suitable because it takes
`advantage of its natural tiled organization to easily map it to a dis-
`tributed network of identical computing chiplets, as shown in
`Figure 1. We consider a chiplet to be a small integrated circuit
`(IC) of submillimeter dimensions that contains a well-defined
`subset of functionality and is designed to be combined with
`other chiplets in the organ-conformal platform. Each chiplet
`implements a cell unit of the cellular neural network, processing
`only local sensor information from itself and its neighbors and
`providing an output only to its local actuator.
`The size, area, and power constraints are particularly impor-
`tant for this application. Emerging computing technologies, like
`memristor crossbars,[14] have significant potential in the More-
`than-Moore era, promising orders of magnitude better energy
`efficiency and compact implementation[15,16] of use in novel com-
`puting systems for implantable devices. A memristor commonly
`uses metal/insulator/metal sandwich structures, which include
`two layers of electrodes and an intermediate layer of memristive
`functional material, which is called the insulator.[17,18] Memristor
`devices can be fabricated as small as 2 nm, and[19] their resistance
`transition characteristics are closely associated with their electro-
`des and the switching materials. The device needs “forming” to
`create filamentary path(s) in the insulator and then reversibly set
`and reset to program the device to a desired conductance state
`between low (OFF) and high (ON) states. Thanks to its ionic
`transport, the programmed state is retained without static power
`dissipation. Memristor devices can be integrated with comple-
`mentary metal–oxide–semiconductor (CMOS) control circuitry
`as dense matrices (crossbars) of artificial synapses to implement
`vector matrix multiplication using Ohm's law,[14,20–22] which is a
`fundamental operation in neural networks. This behavior ena-
`bles a natural solution for the implementation of templates
`for the proposed cellular neural network computing,
`to be
`integrated directly with sensors and actuators. This approach allows
`for flexibility, requiring the design of only one chiplet and its tape-out
`in as many samples as needed for the size of the network at hand.
`The proposed solution can be used to develop the next-generation
`implantable devices that can provide low-energy therapy, thanks to
`high-resolution sensing, local computing, and precise actuation.
`The remainder of the article is organized as follows. Section 2
`describes the methodological details, the data obtained from
`human cardiac tissue, as well as the algorithm and the perfor-
`mance metrics used. Section 3 introduces the evaluation of
`the proposed methodology on the dataset, considering the opti-
`mization of hyperparameters such as the learning rate, weight
`initialization, binarization, as well as the impact of noise and
`quantization in the input and templates on the inference results.
`Section 4 concludes with a discussion of the algorithmic results
`and their potential mapping to a memristor-based hardware
`implementation.
`
`Adv. Intell. Syst. 2022, 2200032
`
`2200032 (3 of 16)
`
`© 2022 The Authors. Advanced Intelligent Systems published by Wiley-VCH GmbH
`
`3
`
`

`

`(b)
`
`(a)
`
`(c)
`
`Figure 2. Data gathering. a) Human left ventricular tissue wedge and experimental setup. b) Simultaneous optical and electrical cardiac mapping.
`c) Corresponding representative electrical and optical signals. Figure 2b,c are reproduced with permission.[24] Copyright 2022, American Heart Association.
`
`www.advancedsciencenews.com
`
`www.advintellsyst.com
`
`(a)
`
`(b)
`
`(c)
`
`(d)
`
`Figure 3. Data preprocessing. a) Example of raw optical phase map (100  50 pixels) recorded during VF in the human heart preparation showing the
`influence of the pacing electrode on the obtained signal. b) Example of Hilbert-transformed optical phase map. A subset (70  35 pixels) was selected to
`avoid network confusion due to pacing electrode effects. c) Example of input data used for training and its labeling. d) Example of input data used for
`testing and its labeling.
`
`Adv. Intell. Syst. 2022, 2200032
`
`2200032 (4 of 16)
`
`© 2022 The Authors. Advanced Intelligent Systems published by Wiley-VCH GmbH
`
`4
`
`

`

`www.advancedsciencenews.com
`
`www.advintellsyst.com
`
`inner states of any dynamic process.[30] Genetic algorithms have
`been shown to train the network with desirable accuracy and
`robustness, but
`the evaluation of
`the fitness functions is
`computationally very expensive.[29]
`We have defined a training algorithm based on backpropaga-
`tion and batch updates robust
`to template nonidealities.
`Following initialization, the network will calculate the corre-
`sponding error for each image in the batch. The templates A,
`B, and bias I will be updated after each batch calculation. The
`process was repeated for all images in the training dataset to min-
`imize the error between the obtained wavefront map output and
`the desired output. The network took several epochs to converge
`and several performance metrics, as shown in the next section,
`could be used to track the convergence.
`For the case of the typical adapted stochastic gradient descent
`backpropagation training algorithm, the error was calculated
`based on
`
`(4)
`
`ðdij y
`ij½kŠÞ
`
`1 2
`
`eij½kŠ ¼
`
`where y
`ij½kŠ is the output calculated by the algorithm at iteration k
`and dij is the desired cell output according to the image label. The
`templates A, B, and bias I are updated based on
`
`(5)
`
`(6)
`
`(7)
`
`(8)
`
`(9)
`
`(10)
`
`Due to the unavoidable interrupts during the hour-long
`experiments and the underlying condition of
`the available
`human heart tissue, noise was an inevitable occurrence in the
`dataset. Noise is regarded as the irregular small section of pixels
`rapidly changing in the range from π to π, as well as the value of
`pixel remaining constant throughout the measurement. The pac-
`ing electrode could also introduce significant artifacts due to its
`large size, needed to provide mechanical robustness during
`insertion into the rather stiff human cardiac muscle tissue. To
`avoid these unwanted effects, the data was cropped to 70  35
`pixels and the pixels containing the pacing electrode were
`removed, as shown in 3b vs. 3c.
`
`2.2. Cellular Neural Networks
`
`Given the tight requirements for high speed and low-power
`hardware, the cellular neural network is a promising topology
`for distributed computing, based on a fixed number of intercon-
`nected processing units called “cells.” Each unit, for example,
`unit ij at row i and column j, could be implemented by a
`computing chiplet, processing only local information from itself
`and its neighbors, with small size and energy requirements. The
`inputs uijðtÞ at time t were fed into the network and outputs yijðtÞ
`were obtained. The output of a processing cell ij was determined
`by the state of the cell xijðtÞ according to Equation (2).
`
`yijðtÞ ¼ ðjxijðtÞ þ 1j jxijðtÞ 1jÞ
`
`(2)
`
`The state of cell ij at time t was calculated using the differential
`equation (3) taking into consideration all the cells in the neigh-
`borhood of size M N. This work included only the nearest
`neighbors (neighborhood size ¼ 3  3) to keep the results map-
`pable to a potential
`compact hardware implementation.
`However, the neighborhood could include further away neigh-
`bors, for example, a neighborhood of size 7  7 included one
`central cell and 48 neighbors.
`
`amn½k þ 1Š ¼ amn½kŠ þ ηΔamn½kŠ
`
`bmn½k þ 1Š ¼ bmn½kŠ þ ηΔbmn½kŠ
`
`I½k þ 1Š ¼ I½kŠ þ ηΔI½kŠ
`
`with the updates of Δ weights
`
`eij½kŠy
`iþm2,jþn2½kŠ
`
`eij½kŠuiþm2,jþn2½kŠ
`
`N X1 ≤ i ≤ M
`
`1 M
`
`1 ≤ j ≤ N
`
`N X1 ≤ i ≤ M
`
`1 M
`
`1 ≤ j ≤ N
`
`Δamn½kŠ ¼
`
`dxijðtÞ
`dt
`
`
`
`¼ xijðtÞþ X1 ≤ i ≤ MamnymnðtÞþ X1 ≤ i ≤ M
`
`1 ≤ j ≤ N
`
`1 ≤ j ≤ N
`
`bmnumnðtÞ þ I
`
`Δbmn½kŠ ¼
`
`(3)
`
`eij½kŠ
`
`N X1 ≤ i ≤ M
`
`1 M
`
`1 ≤ j ≤ N
`
`ΔI½kŠ ¼
`
`where m and n are the row and column indices, respectively, of
`the templates A and B. η is the learning rate, typically a small
`number always >0, that defines the range of weight updates
`in each iteration. As seen in Equation (8), the update Δamn½kŠ
`for the feedback template A was calculated via the weighted
`sum of the error and the desired output for each cell. A similar
`update Δbmn½kŠ was calculated for control template B based on the
`error and the respective input. The bias I was also updated
`accordingly based on the average error of each cell to increase
`the performance of the network.
`To improve the wall-clock time, we used batch training as
`defined by
`
`In Equation (3), the inputs umn and outputs ymn of its cell and
`neighboring cells were weighted via the matrix elements amn and
`bmn of two matrices A and B of size M and N. The matrix A linked
`the outputs ymn to the state x via its elements amn, while template
`B similarly linked the inputs umn to the state x, respectively.
`These matrices were called templates and were used repeatedly
`for each cell. Training the network means determining the values
`of templates A and B and of bias I.
`Several algorithms were used for training these networks,
`including, random weights change,[27] Kalman filters,[28] genetic
`algorithms,[29] and backpropagation.[30] The random weight
`change[27] is a hardware-friendly algorithm for on-chip training
`on a wide range of tasks, but it involves large number of training
`epochs to obtain accurate templates. Kalman filters have been
`used to obtain accurate output from the inaccurate input infor-
`mation, minimizing the mean of squared error by estimating the
`
`Adv. Intell. Syst. 2022, 2200032
`
`2200032 (5 of 16)
`
`© 2022 The Authors. Advanced Intelligent Systems published by Wiley-VCH GmbH
`
`5
`
`

`

`www.advancedsciencenews.com
`
`www.advintellsyst.com
`
`Table 1. Performance matrix.
`
`(11)
`
`Desired output pixel
`
`ON
`
`OFF
`
`ON
`
`OFF
`
`True Positive (TP)
`
`False Positive (FP)
`
`False Negative (FN)
`
`True Negative (TN)
`
`(12)
`
`Obtained
`output pixel
`
`(13)
`
`accuracy ¼
`
`TP þ TN
`TP þ TN þ FP þ FN
`
`(15)
`
`Precision measures the performance of correctly identifying
`positive (wavefront) pixels. For the targeted application, it is
`highly important to have very high TP and low FP (high preci-
`sion) to avoid applying unneeded pulses.
`
`precision ¼
`
`TP
`TP þ FP
`
`(16)
`
`XB
`XB
`XB
`
`Δamn½kŠ
`
`Δbmn½kŠ
`
`1 B
`
`1 B
`
`ΔAmn½kŠ ¼
`
`ΔBmn½kŠ ¼
`
`ΔI½kŠ
`
`1 B
`
`ΔIbatch½kŠ ¼
`
`where B is the batch size for which the Δ template updates were
`averaged.
`The error could be calculated on the obtained output as proc-
`essed by the network, which took grayscale values between [1,1]
`or on a binarized version of the output which could be either
`wavefront (1) or nonwavefront (1). The obtained output could
`be binarized, either during training or after the training was com-
`plete by applying a threshold as defined by the following equation
`
`yb
`ijin ¼
`
`
`
`1if yij ≥ threshold
`1else
`
`(14)
`
`Sensitivity measured how many of the positive (wavefront) pix-
`els were identified as such.
`
`As we targeted hardware mappability, we also explored the
`impact of neighborhood size as well as limited bit precision
`weights. Limited precision templates were also considered using
`traditional routing-to-nearest method versus stochastic round-
`ing. Stochastic rounding can be particularly useful
`in deep
`network training with low bit precision arithmetic.[31,32] A real
`template value a which lies between a lower weight level (A1)
`and upper weight level (A2) was stochastically rounded up to
`A2 with probability (a–A1)/(A2
`–A1) and down to A1 with proba-
`bility (A2
`–a)/(A2
`–A1). The algorithm details are included in the
`supplemental materials.
`
`2.3. Performance Metrics
`
`To provide a comprehensive assessment of the potential perfor-
`mance of the algorithm to human tachyarrhythmia events, sev-
`eral performance metrics were used in accordance with medical
`practice for binary classification tests. The desired output was a
`binary map with pixels on the wavefront(s) labeled as “positive”
`(or “ON” or black) totaling P pixels and all other pixels, not on the
`wavefront labeled as “negative” (or “OFF” or white) totaling N
`pixels. The obtained output after the image was classified by
`the network was a similar binary map. Some of the pixels on
`the wavefront were identified correctly (true positives), totaling
`TP pixels, while others were misclassified as negative (false neg-
`atives), totaling FN pixels. Similarly, some of the pixels outside
`the wavefront were identified correctly (true negatives), totaling
`TN pixels, while others were misclassified as positive (false
`positives), totaling FP pixels. These could be arranged in a typical
`2  2 contingency table or a confusion matrix (Table 1).
`Based on this classification, four important performance
`metrics, accuracy, precision, sensitivity, and specificity, are
`defined as follows. Accuracy provides a quantitative metrics of
`the overall performance of the algorithm, showing the percent-
`age of the total number of pixels correctly identified.
`
`sensitivity ¼
`
`TP
`TP þ FN
`
`(17)
`
`Specificity measures how many of the negative (nonwave-
`front) pixels were identified.
`
`specificity ¼
`
`TN
`TN þ FP
`
`(18)
`
`The goal was to optimize the algorithm to achieve high values
`for all four performance metrics, accuracy, precision, sensitivity,
`and specificity simultaneously.
`
`3. Results
`
`3.1. Training Optimization
`
`3.1.1. Hyperparameter Optimization on Single-Image Training
`
`Single-image training was used to do a comprehensive search in
`the hyperparameter space for learning rate, initialization, and
`binarization and understand the impacts and trade-offs on per-
`formance. The optimal learning rate was identified for different
`initializations by exploring a broad range from 104 to 104 in log-
`arithmic scale. The initializations are 1) zero-template matrices;
`2) randomly generated values between 1 and 1; and 3)
`pre-defined templates for edge detection; details are described
`in Table 2.
`As shown in Figure 4, an optimal learning rate window is vis-
`ible, where all the four performance metrics are optimized.
`Outside of this learning rate window, the performance drops par-
`ticularly for precision and sensitivity. All these metrics need to be
`optimized simultaneously, with precision being the most impor-
`tant metric to avoid false positives that would inadvertently apply
`unwanted pulses to the heart tissue. The highest precision is
`92.16% on learning rate ¼ 500 with all initial templates A, B
`and bias I set to 0 shown in Figure 4d. The maximum value
`
`Adv. Intell. Syst. 2022, 2200032
`
`2200032 (6 of 16)
`
`© 2022 The Authors. Advanced Intelligent Systems published by Wiley-VCH GmbH
`
`6
`
`

`

`www.advancedsciencenews.com
`
`www.advintellsyst.com
`
`Table 2. Initialization templates used for a system with eight neighbors.
`
`Zero template
`
`Example of
`random template
`
`Edge detection
`template
`
`0
`
`0
`
`0
`
`0
`
`0
`
`0
`
`Template A
`
`Template B
`
`Bias I
`
`(a)
`
`(e)
`
`0
`
`0
`
`0
`
`0
`
`0
`
`0
`
`0
`
`(b)
`
`(f)
`
`0
`
`0
`
`0
`
`0
`
`0
`
`0
`
`0.6849
`
`0.1537
`
`0.1185
`
`0.4339
`
`0.9216
`
`0.8378
`
`0.3077
`
`0.5972
`
`0.3394
`
`0.4386
`
`0.6144
`
`0.014
`
`0.777
`
`0.189
`
`0.5028
`
`0.3372
`
`0.0429
`
`0.7459
`
`0.2083
`
`0
`
`0
`
`0
`
`1
`
`1
`
`1
`
`0
`
`0
`
`0
`
`1
`
`8
`
`1
`
`1
`
`0
`
`0
`
`0
`
`1
`
`1
`
`1
`
`(c)
`
`(g)
`
`(d)
`
`(h)
`
`Figure 4. Impact of learning rate optimization and initialization optimization on training and testing performance. a,b,c,d) Evolution of sensitivity,
`specificity, accuracy, and precision for training and e,f,g,h) for testing, respectively, using different learning rates and different initializations. The
`representative image from Figure 3b and its label were used for training. Number of neighbors ¼ 8.
`
`for specificity is 99.82% at the same learning rate. The maximum
`value for sensitivity is 98.44% on learning rate ¼ 0.05 with edge
`detection template and for accuracy is 99.38% on learning
`rate ¼ 1000 with zero templates. However, it is important to note
`that large precision is obtained for different initializations and
`over a broad range of learning rates, from 0.5 to 500. As a learn-
`ing rate that is too fast will result in large weight updates that can
`induce oscillatory behavior in the training between suboptimal
`local solutions, the best lower learning rate was investigated.
`Testing was used test the performance of these results on nine
`inputs. The averaged results are shown in Figure 4e-h. The
`highest precision is 73.08% on learning rate ¼ 0.1 with random
`template initialization, while the other performance metrics at
`the same hyperparameters are accuracy 97.85%, specificity
`99.26%, and sensitivity 42.00%.
`After evaluating the training results and testing results,
`within the optimal learning rate window, a learning rate of
`0.1 with the random initialization of templates has shown
`the highest precision especially in testing results. The results
`in Figure 5a show the convergence curves for the different per-
`formance metrics and the obtained templates. The poor sensi-
`tivity and precision are due to the fact that 41 pixels are stuck
`with in-between values, as shown in Figure 5b. A test image, as
`shown in Figure 3c, was used for validation. The challenge with
`these “gray” pixels also seems to translate to the test image, as
`
`shown in Figure 5c. This method of training where the output is
`allowed to take “grayscale” values has a difficult time differen-
`tiating true versus false positive pixels,
`leading to a large
`number of pixels in the grayscale regime and precision of only
`around 70%.
`These results prompt the need for a binarization approach by
`imposing the desired output of the network to be binary, either
`ON or OFF. The two methods, binarization during training and
`binarization after training, have been explored in Figure 6. In
`binarization during training approach, the output is forced to
`be binary after each weight update (Figure 6a). In the binariza-
`tion after training approach, the network is trained by itself with
`output in grayscale and the output is forced to be binary once the
`training is complete (Figure 6b). Figure 6c vs. d shows compara-
`tively the convergence curves for the two approaches. The
`binarization during training converges quickly but experiences
`oscillations in the metrics as the true-positive and false-negative
`pixels flip values and do not stay

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket