throbber
International Journal of Cardiology 166 (2013) 15–29
`
`Contents lists available at SciVerse ScienceDirect
`
`International Journal of Cardiology
`
`j o u r n a l h o m e p a g e : w w w . e l s e v i e r . c o m / l o c a t e / i j c a r d
`
`Review
`How accurate is pulse rate variability as an estimate of heart rate variability?
`A review on studies comparing photoplethysmographic technology
`with an electrocardiogram
`Axel Schäfer a,b, Jan Vagedes a,c,⁎
`a Arcim institute, Im Haberschlai 7, 70794 Filderstadt, Germany
`b University of Tübingen, Institute of Medical Psychology and Behavioral Neurobiology, Gartenstr. 29, 72074 Tübingen, Germany
`c University of Tübingen, Children's Hospital, Department of Neonatology, Calwerstrasse 7, 72076 Tübingen, Germany
`
`a r t i c l e
`
`i n f o
`
`a b s t r a c t
`
`Article history:
`Received 7 September 2011
`Received in revised form 26 January 2012
`Accepted 10 March 2012
`Available online 17 July 2012
`
`Keywords:
`Heart rate variability
`Photoplethysmography
`Pulse wave analysis
`Continuous blood pressure monitoring
`Electrocardiography
`
`Background: The usefulness of heart rate variability (HRV) as a clinical research and diagnostic tool has been
`verified in numerous studies. The gold standard technique comprises analyzing time series of RR intervals
`from an electrocardiographic signal. However, some authors have used pulse cycle intervals instead of RR
`intervals, as they can be determined from a pulse wave (e.g. a photoplethysmographic) signal. This option
`is often called pulse rate variability (PRV), and utilizing it could expand the serviceability of pulse oximeters
`or simplify ambulatory monitoring of HRV.
`Methods: We review studies investigating the accuracy of PRV as an estimate of HRV, regardless of the under-
`lying technology (photoplethysmography, continuous blood pressure monitoring or Finapresi, impedance
`plethysmography).
`Results/conclusions: Results speak in favor of sufficient accuracy when subjects are at rest, although many
`studies suggest that short-term variability is somewhat overestimated by PRV, which reflects coupling effects
`between respiration and the cardiovascular system. Physical activity and some mental stressors seem to
`impair the agreement of PRV and HRV, often to an inacceptable extent. Findings regarding the position of
`the sensor or the detection algorithm are not conclusive.
`Generally, quantitative conclusions are impeded by the fact that results of different studies are mostly incom-
`mensurable due to diverse experimental settings and/or methods of analysis.
`© 2012 Published by Elsevier Ireland Ltd.
`
`1. Introduction
`
`The term ‘Heart rate variability’ (HRV) refers to the fact that the
`duration of cardiac cycles is not constant, but varies from one heart-
`beat to the next. The extent of variability is determined by digital
`processing of an electrocardiographic (ECG) signal. Because of their
`distinct profile, the R peaks of an ECG signal are suitable for automated
`detection by computer algorithms; hence the standard method to
`assess cardiac cycles is to place their limits at the R peaks. As a result,
`one obtains a time series of such consecutive RR intervals. Ectopic
`beats and arrhythmic events are usually not processed when deter-
`mining HRV; only regular beats should be considered, which is why
`one often encounters the alternative term NN intervals (‘normal to
`normal’). Analysis of HRV comprises the computing of meaningful
`parameters from RR interval time series called HRV variables. For a
`comprehensive presentation of the methodology see [1].
`
`⁎ Corresponding author at: University of Tübingen, Children's Hospital, Department
`of Neonatology, Calwerstrasse 7, 72076 Tübingen, Germany. Tel.: +49 7071 2984742.
`E-mail address: Jan.Vagedes@med.uni-tuebingen.de (J. Vagedes).
`
`0167-5273/$ – see front matter © 2012 Published by Elsevier Ireland Ltd.
`doi:10.1016/j.ijcard.2012.03.119
`
`HRV has become a very useful tool in clinical diagnostics within
`the last decades. Reduced HRV is correlated with the risk of cardiac
`events like myocardial infarction and congestive heart failure [2,3],
`and with sudden cardiac death [4,5]. From the early 1980s on, the
`frequency domain HRV variables, gained from a power spectrum of
`the RR interval series, have been found to reflect autonomic cardio-
`vascular control [6,7]. Physical fitness and social integration are both as-
`sociated with reduced cardiac risk and enhanced HRV [8,9]. Treatment
`of psychiatric patients with tricyclic antidepressants seriously dimin-
`ishes HRV, although there is ambiguity as to whether psychopharmaco-
`logical treatment increases mortality [10]. Also, depression itself seems
`to influence autonomic control and HRV [11]. All in all, HRV seems to
`be a reliable and multifunctional parameter indicating cardiovascular
`and autonomic health as well as general psychic and somatic fitness.
`Evaluation of HRV variables can be gained from sessions of various
`duration, up to Holter recordings lasting 24 h or more. However,
`short-term recordings of only a few minutes have been found to be
`comparably useful [12], and even ultra-short sequences of only 10 s
`seem to have reasonable diagnostic value [13].
`Photoplethysmography (PPG) is a technique developed in the 1930s
`for monitoring blood volume changes in the micro vascular bed of tissue
`
`APPLE 1018
`
`1
`
`

`

`16
`
`A. Schäfer, J. Vagedes / International Journal of Cardiology 166 (2013) 15–29
`
`[14]. In more recent decades, progress in semiconductor technology and
`optoelectronics, as well as advancements in digital signal processing,
`have facilitated a renaissance of PPG, which today is probably the
`most widespread method used in clinical monitoring. Its basic principle
`requires a light source to illuminate subcutaneous tissue (typically
`an LED, i.e. a light emitting diode) and a photo detector with spectral
`characteristics matching those of the light source (e.g. a photodiode
`or phototransistor). Current PPG sensors use low-cost optoelectronic
`components operating in the domain of red and/or near infrared
`wavelengths.
`There are two basic configurations used in PPG: transmission
`mode, where the perfuse tissue (like a fingertip or an earlobe) is
`placed between the source and the detector, and reflection mode,
`where the two electronic components are placed side-by-side near
`the skin, e.g. at the forehead. In both cases the detector registers
`small variations in the transmitted or reflected light, respectively,
`caused by changes in microcirculation. Major factors affecting the
`detected light intensity are blood volume, blood vessel wall move-
`ment and the orientation of erythrocytes [15].
`Changes in a PPG signal arise from variations both in the path-
`length between source and detector and in the optical density of
`the blood. The signal can be decomposed into two parts [16]. The
`small pulsatile component, or AC component, arises from arterial
`blood pulsation; hence its oscillation parallels momentary cardiac
`activity. It is superimposed on the much larger DC component,
`where DC refers to direct current, suggesting a static behavior. How-
`ever, the DC component is not entirely static, but includes variations
`slower than the heart rate due to venous volume fluctuations, vaso-
`motor activity and thermoregulation. Changes in the intrathoracic
`pressure due to respiration cause fluctuations in the venous return
`to the heart, which in turn modulates cardiac output and blood pres-
`sure. Ventilation thus induces fluctuation of both AC and DC compo-
`nents, which enables one to monitor respiratory activity by filtering
`and processing a PPG signal appropriately [17,18].
`PPG technology is a very versatile tool, and its usefulness in a wide
`range of clinical applications has been demonstrated. The most
`common one is pulse oximetry, which utilizes the difference of red
`and near infrared light absorption by oxyhemoglobin and reduced
`hemoglobin to estimate arterial blood oxygen saturation. Further
`applications include the estimation of cardiac output, the diagnosis
`of atherosclerosis, peripheral arterial occlusion and other peripheral
`vascular diseases, as well as the assessment of arterial compliance
`and aging, of endothelial and venous function, micro vascular blood
`flow and other functions [15]. Additionally, PPG offers a number of
`ways to assess cardiovascular autonomic function and skin vasomotor
`function. Due to this property, it is also a valuable instrument for
`monitoring nociception during general anesthesia [16]. For a compre-
`hensive review on PPG and its clinical applications see [15].
`Arterial blood pressure can also be estimated from a PPG, because
`the higher the pressure, the quicker the propagation of the pulse
`wave from the heart to the periphery. Therefore, the determination
`of pulse wave velocity or its reciprocal, pulse transit time (cf.
`Section 3.1 below) for each pulse cycle can serve as an approximation
`of continuous blood pressure (CBP) monitoring. However, estimators
`of systolic and diastolic blood pressure are only in moderate agree-
`ment with their directly measured counterparts, although this can
`be somewhat improved by including the amplitude of each PPG
`pulse wave in the analysis [19,20]. In order to provide a more reliable
`method for CBP monitoring, the Finapres™ (for FINger Arterial PRES-
`sure) technology was introduced in the early 1980s. It is based on the
`dynamic vascular unloading of the arterial walls of the finger using an
`inflatable finger cuff [21]. The built-in PPG sensor is used as a control
`sensor to regulate cuff pressure for optimum CBP detection, but does
`not measure blood pressure directly.
`Given that PPG is such a simple and ubiquitous technology in clin-
`ical monitoring, it is preferable to maximize its potential. As already
`
`indicated above, the signal can also be used to monitor the heart
`and respiratory rate of a patient, including instantaneous rates or
`cycle lengths. In principle, this would allow HRV variables to be deter-
`mined as well from a PPG signal. Compared to the current standard of
`an ECG-based HRV analysis, this would involve certain benefits. In
`clinical situations where a pulse oximeter (PO) device is already at
`hand by default, being able to include HRV analysis in the monitoring
`process without requiring an ECG means a significant advantage. In
`addition, during magnetic resonance imaging (MRI), for example,
`ECG electrodes or other metal-containing sensors are not permitted,
`as they interfere with strong electromagnetic fields.
`Another fact speaking in favor of PPG technology is that it is
`noninvasive, cost-effective and straightforward to use. Detecting the
`signal usually requires no more than attaching a single sensor to a
`finger or an earlobe, compared to at least three leads and Ag/AgCl
`electrodes required for an ECG. Furthermore, ECG electrodes often
`have to be applied to the chest, requiring the patients to undress—
`which can delay recordings and pose a problem for embarrassed
`patients. On the other hand, a major disadvantage of PPG technology
`is that the signal is susceptible to motion artifacts, which can impair
`the accuracy of the detected cardiac activity [15].
`
`2. Scope of this review
`
`The purpose of this review is to summarize the hitherto existing
`literature about the accuracy of estimating HRV and/or (instanta-
`neous) heart rates from a continuously recorded pulse wave signal.
`All of the considered articles contrast the results of the latter with
`the gold standard of an ECG-based method.
`We performed a search in PubMed and Embase for publications
`that matched at least one keyword in both of the following word
`groups:
`
`1) oximetr…/oxymetr… or plethysmogr… or “pulse wave” or “pulse
`applan…”
`2) electrocard… or ECG or “rate variability”
`
`Here the ellipsis (…) indicates arbitrary completion of words, and
`quotation marks require the enclosed words to appear exactly in
`the specified order. The retrieved abstracts were then inspected for
`their thematic agreement with the topic of this review. Such selected
`articles and correspondingly eligible references cited therein were all
`included.
`Most of the retrieved studies use a PPG signal to detect a pulse
`wave. However, some employ a Finapres™ system monitoring CBP
`or an impedance plethysmography system instead. The analysis
`techniques used to discern individual pulse cycles in these cases are
`similar to those utilized for PPG signals. Additionally, many of the
`findings therein complement those in PPG studies. For these reasons,
`all those references assessing the quality of results from non-PPG
`pulse waves are included in this review.
`The main focus is on the question whether the ECG-based method
`of evaluating HRV can be replaced by a technique using a pulse wave.
`From now on the latter approach will be referred to as pulse rate
`variability (PRV), as it is based on the varying length of pulse cycles,
`not cardiac cycles. Heartbeats can be more accurately determined
`from an ECG; therefore we will henceforth use the term HRV mainly
`for ECG-derived heart rate variability.
`It is clear that the methods used to determine the limits between
`adjacent pulse cycles affect the individual lengths of the latter and,
`hence, PRV results. The statistical techniques used to compare, say
`PRV and HRV, will likewise have an impact on the outcomes of an
`investigation. Section 3 outlines some important approaches used in
`analyses; additionally, Tables 1 and 2 summarize the most important
`abbreviations regarding HRV variables and HRV/PRV analysis.
`A number of researchers were merely interested in the accuracy of
`estimating the average heart rate with a PPG device for monitoring
`
`2
`
`

`

`A. Schäfer, J. Vagedes / International Journal of Cardiology 166 (2013) 15–29
`
`17
`
`Table 1
`Abbreviations of common HRV variables.
`
`Time domain variables
`
`Mean NN, mean HR
`
`SDNN
`
`RMSSD
`
`SDSD
`
`NN50/pNN50 (pNN10)
`
`Mean of all NN intervals. Mean HR is its reciprocal value,
`converted into beats per minute (bpm)
`Standard deviation of all NN intervals;
`measure of overall variability.
`Root of mean of squared subsequent differences;
`measure of short-term variability.
`Standard deviation of subsequent differences
`(almost identical to RMSSD).
`Number/percent of subsequent differences with an
`absolute value > 50 ms (or 10 ms).
`
`Frequency domain variables (spectral domains according to Task force definitions [1])
`
`ULF
`VLF
`LF
`HF
`TP
`
`Nonlinear variables
`
`SD1
`SD2
`ApEn
`SampEn
`
`Ultra low frequencies (b0.003 Hz)
`Very low frequencies (0.003–0.04 Hz)
`Low frequencies (0.04–0.15 Hz)
`High frequencies (0.15–0.40 Hz)
`Total power (sum contribution of all spectral domains)
`
`Standard deviation of short diagonal axis in Poincaré plot
`Standard deviation of long diagonal axis in Poincaré plot
`Approximate entropy
`Sample entropy
`
`purposes; hence they did not concern themselves with individual
`heart or pulse cycles or HRV variables. However, some of them use
`analytical approaches that are interesting in context, while other in-
`vestigations can be perceived as historical predecessors of the studies
`in the focus of this review. We thus summarize these results briefly in
`Section 4 and Table 3.
`The central topic, compatibility of PRV with HRV, is treated in
`Section 5 and its subsections. We should note that the various find-
`ings are almost never comparable directly, since a variety of devices,
`analytical methods, experimental conditions etc. have been used.
`Therefore, we can present and contrast all results only descriptively,
`itemized in Tables 4 and 5.
`A discussion and conclusions can be found in Sections 6 and 7.
`
`3. Analytical methods
`
`3.1. Pulse wave analysis
`
`A typical pulse wave cycle can be subdivided into two parts (cf.
`Fig. 1). The anacrotic phase is the rising part of the pulse due to systole.
`Shortly after the QRS complex appears in the ECG, the ventricular systole
`generates a pulse wave travelling distally. In the arteries and arterioles
`this leads to a rapid increase in blood pressure and blood volume, i.e. a
`steep rise in the pulse wave. The subsequent decline corresponding to
`cardiac diastole is termed the catacrotic phase and is more prolonged
`than the anacrotic phase. Often it contains a secondary peak separated
`
`Table 2
`Abbreviations of HRV and PRV phenomena used in the text.
`
`Time intervals
`RRI
`PPI
`
`Frequencies/rates
`HR
`PR
`IHR
`IPR
`
`Result of analysis
`HRV
`PRV
`
`RR interval length(s) from an ECG
`Pulse interval length(s) from a pulse wave signal
`
`Heart rate, determined by an ECG signal
`Pulse rate, determined by a pulse wave signal
`Instantaneous heart rate (reciprocal of RRI)
`Instantaneous pulse rate (reciprocal of PPI)
`
`Heart rate variability, gained by analyzing RRI values
`Pulse rate variability, gained by analyzing PPI values
`
`by the so-called dicrotic notch, an effect diminishing with ageing and
`increasing arterial stiffness [15]. This phenomenon is often attributed
`to the closure of the aortic valve [22], although recent findings suggest
`a causation by reflection of the pressure peak from small arteries in
`the trunk and lower limbs [23].
`Corresponding to each RR interval (RRI) of the ECG, which is
`considered as the “true” instantaneous heart cycle length, there is
`an interval comprising a full pulse cycle length. We will henceforth
`denote it as PPI (“pulse to pulse interval”). Its exact location depends
`on the definition of its boundaries and the computer algorithm used
`to detect them [24]. The black disks in Fig. 1 highlight three possible
`alternatives. One can determine the beginning of the anacrotic or,
`alternatively, the catacrotic phase, i.e. the pulse foot or peak (marked
`with an “f” or “p” in the figure), as the respective boundary. This leads
`to different definitions of pulse intervals, as is indicated in the figure
`by PPI (f) and PPI (p). A third option is to use the maximum 1st deriv-
`ative representing the steepest part of the upstroke as a boundary, in-
`dicated by the disk marked “d”. In the existing literature researchers
`have utilized all of these methods as well as others (cf. Table 5); com-
`parative studies on some approaches can be found in [25,26].
`Depending on the pulse wave velocity and the vascular path from
`the heart to the location of the detector, there is a delay between each
`R peak and the onset of its corresponding pulse wave. The delay is
`usually termed pulse transit time (PTT) and is negatively correlated
`with blood pressure, arterial stiffness, and age [15]. As defined
`above, it ranges from an R peak to the next pulse foot or the beginning
`of systole; however, many studies employ a different definition of a
`PTT extending to the subsequent peak. The two cases are denoted
`as PTT (f) and PTT (p) in Fig. 1.
`Deviations of the PPI from the RRI series can arise from two possi-
`ble causes: 1) an inaccurate detection of pulse cycle boundaries due
`to hardware limitations, artifacts and/or noise, or 2) a physiological
`variability in PTT.
`
`3.2. Heart and pulse rate variability analysis
`
`In order to quantify the agreement of PPI and RRI series, researchers
`have mostly used one or more of the following three strategies:
`
`a) Often one is only interested in the accuracy of the determined
`mean pulse rate (PR) compared to the mean heart rate derived
`from an ECG. This is usually sufficient when a PPG device is used
`for monitoring or telecare.
`b) The focus of this review is on the reliability of PRV as a substitute
`for HRV. Most of the studies in this domain compare HRV variables
`for both the RRI and PPI series, which we will henceforth refer to
`as HRV and PRV variables. The most common HRV variables eval-
`uated are time and frequency domain variables, which have been
`extensively standardized, but nonlinear variables are also used [1].
`Abbreviations of the most commonly used HRV variables are listed
`in Table 1.
`c) A number of authors also compare the RRI and PPI series directly,
`i.e. they compare the lengths of individual heart beats to the cor-
`responding pulse cycles—or they do the same thing with the in-
`stantaneous heart (IHR) and pulse (IPR) rates, which are simply
`the reciprocal values of the cycle lengths.
`
`Table 2 gives an overview of the respective quantities and their
`abbreviations used throughout the subsequent sections.
`
`3.3. Statistical analysis
`
`In all the studies considered in this review, quantities determined
`from an ECG and a simultaneously recorded pulse signal are com-
`pared. Data pairs are given where each pair comprises corresponding
`values measured by each method in question, respectively. In case
`a) of the preceding subsection, these values are average HR and PR
`
`3
`
`

`

`18
`
`A. Schäfer, J. Vagedes / International Journal of Cardiology 166 (2013) 15–29
`
`Table 3
`Overview of studies on monitoring heart rate with PPG technology.
`
`Study
`
`N=number of subjects×
`T=duration of records
`
`Device(s) used
`
`Position of
`sensor
`
`Experimental
`conditions
`
`Methods of pulse detection
`and analysis
`
`Results
`
`Adults
`Altemeyer et al.
`(1986) [31]
`
`Atlasz et al.
`(2006) [37]
`
`Lindberg et al.
`(1992) [34]
`Nakajima et al.
`(1996) [35]
`
`(Monitored patients
`during operation,
`ages 25–84) N = 19
`
`Biox III, and
`Nellcor N-100
`pulse oximeters
`
`Index or middle
`finger
`
`Both normal
`respiration and
`inhalation of hypoxic
`gas mixture.
`5 min supine rest,
`5 min standing
`
`Fingertip
`
`(Healthy, ages 22.2±1.9)
`
`N = 35 × T = 10 min
`(Males, ages 20–30)
`N = 11 × T = 10 min
`(Males, ages 22–34)
`N = 11 × T = 20 min
`
`Self-made
`reflection PPG
`sensor
`Self-made
`PPG sensor
`Self-made
`PPG sensor
`
`Left forearm
`
`Supine rest
`
`Earlobe
`
`Ergometer: sitting +
`Exercise (30–130 W)
`
`“Good agreement” with ECG-
`determined HR, not quantita-
`tively specified.
`
`No significant differences
`between HR from PPG and ECG.
`
`Exact agreement with HR from
`ECG, when signal clean
`PPG and ECG determined heart
`rates “agreed well”, maximum
`|ΔHR| = 10 bpm
`
`PCC of 0.95, Bland-Altman
`analysis says 95% CI: ΔHR =
`−0.1 ± 6.5 bpm
`
`Spectral method and infrared
`light both more reliable. Results
`in this case: PCC = 0.92, ΔHR =
`−0.6 ± 1.3 bpm (mean ± SD)
`
`Auto-detection method of
`monitoring devices.
`
`Peak counting? (not exactly
`specified)
`Statistics: ANOVA/t-test
`Manual counting of PPG
`peaks
`Band-pass filtering PPG sig-
`nal, then zero-crossing count.
`Take median of 10 consecu-
`tive HRs.
`Band-pass filtering PPG sig-
`nal, finding mean HR by au-
`tocorrelation analysis.
`
`a) Peak detection by
`threshold algorithm,
`reciprocal median pulse
`interval → mean HR.
`b) Looking for freq. with max.
`power in FFT spectrum.
`Analysis by PCC and
`mean ± SD of difference.
`
`Auto-detection method of
`monitoring devices. Test for
`|ΔHR| > 10 bpm
`
`Depending on condition,
`11.9–29% of HR values found
`unreliable (|ΔHR| > 10 bpm)
`
`High-pass filtering PPG sig-
`nal, then zero-crossing count.
`Evaluation of 30 s epochs and
`binary classification.
`Auto-detection method of
`monitoring devices.
`Bland-Altman analysis of
`ΔHR
`Manual elimination of
`motion artifacts. Visual
`analysis of PPG charts, man-
`ual counting of PPI. PCC of PR
`vs. HR
`Auto-detection method of
`monitoring devices.
`Bland–Altman analysis of
`ΔHR
`
`PPG recorded 1.1% (±0.7%)
`false negative and 0.9% (±0.6%)
`false positive beats, as
`compared to ECG.
`95% CI: ΔHR =−2 ±26 bpm
`
`Buttock: PCC = 0.999
`Leg: PCC = 0.995
`Back: PCC =−0.134 (sic!)
`
`95% CI: ΔHR =−0.4 ± 12 bpm
`
`Kornowski et al.
`(2003) [36]
`
`(Cardiac patients vs.
`healthy controls,
`ages 43 ± 18) N = 144
`
`Medic4All
`wristwatch
`telecare system
`
`Wrist (sensor
`worn like a
`watch)
`
`Vogel et al.
`(2007) [38]
`
`N = 1, T not specified,
`additionally 6 h of data
`from the “MIT-BIH
`normal sinus rhythm
`database”
`
`Self-developed
`IN-MONIT system,
`using both red
`and infrared LEDs
`
`Inside auditory
`canal
`
`Rest on chair,
`evaluation of up to 10
`intervals of 30 s for
`each patient,
`respectively.
`Not specified
`
`Neonates
`Barrington et al.
`(1988) [32]
`
`(Lung or cardiorespiratory
`disease) N=22×T=3 min
`
`Roche Medical
`Electronics 634
`
`“Gently
`restrained” limb
`
`Johansson et al.
`(1999) [33]
`
`N = 6 (4 preterm) ×
`T = 8 h
`
`Self-made PPG
`sensor, connected
`to HP 66S monitor
`
`Lateral side of
`left thigh
`
`Comparison of HR
`detected from PPG-
`and ECG-based
`monitors.
`Continuous
`monitoring during 8 h
`
`Kamlin et al.
`(2008) [39]
`
`N = 55 × T = 3 min
`
`Masimo Radical
`PO monitor
`
`Olsson et al.
`(2000) [41]
`
`N = 10
`(preterm) × T = 30 min
`
`Siemens photo
`diode SFH2030
`(940 nm)
`
`Right hand/wrist Experimenters
`reading HR displayed
`by PPG and ECG
`monitors from video.
`Rest in incubators
`
`Leg, buttock and
`interscapular back
`
`Singh et al.
`(2008) [40]
`
`N = 30 × T = 3 min
`
`Masimo Radical
`PO monitor
`
`Right hand
`
`Experimenters
`reading HR displayed
`by PPG and ECG
`monitors from video.
`
`values, in case b) they are HRV and PRV variables. In both approaches
`they are determined from detected cardiac cycles within the identical
`epoch of simultaneously recorded signals. In case c) the values under
`investigation consist of the time series of the detected cardiac cycle
`lengths themselves. Since each PPI belongs to an RRI, namely the
`one beginning directly before and separated only by PTT, one can
`also arrange them in pairs here. Whatever the case, one has a number
`of paired results derived from ECG and pulse signals.
`In order to assess the agreement of data pairs from two methods
`of measurement, the correct statistical approach is not obvious. Often
`researchers use Pearson's correlation coefficient (PCC); however, it
`quantifies linear correlation, not agreement. As examples to illustrate
`this, one can envisage situations where the second method either con-
`sistently overestimates values by an additive constant (location shift)
`or a multiplicative factor (scale shift) or both, which still yield a perfectly
`linear correlation. Aside from this, results of two techniques measuring
`the same data will almost always be linearly correlated. In addition,
`PCC tends to be greater, if the range of the true quantity in the sample
`is increased. Since investigators usually test a method in question
`
`with a wide range of data, a highly significant correlation is almost guar-
`anteed. The arguments speaking against PCC apply even more to Spear-
`man's rank correlation coefficient, which was used in one of the studies
`[27] summarized in this review.
`Testing the equality of the mean, e.g. by Student's t-test, is another
`insufficient way to verify agreement. It is merely able to detect
`whether the lack of accuracy of a method (i.e. its bias) is significantly
`worse than its lack of precision.
`In order to cure the shortcomings of the mentioned analyses,
`several approaches have been proposed. An example is the concor-
`dance correlation coefficient, a more suitable recast of the PCC [28].
`Most of the studies cited in this review, however, either simply
`itemize mean and standard deviation of the differences between
`ECG and pulse measures, or make use of the approach suggested
`by Bland and Altman [29]. In the latter, one plots the differences
`between the two methods versus the average as the best estimator
`of the true value. Even when one technique is considered as the
`gold standard, the average values should be plotted at the abscissa
`in order to avoid statistical artifacts [30]. A 95% confidence interval
`
`4
`
`

`

`A. Schäfer, J. Vagedes / International Journal of Cardiology 166 (2013) 15–29
`
`19
`
`500HzFinger
`200HzRightmiddlefinger
`
`500HzRightmiddlefinger
`
`Middlefinger
`
`Finapresdevice
`IHRasprovidedby
`Notspecified:used
`
`sensorinamattress
`b)Contactless
`1000Hza)Fingerclip
`
`b)Indexfinger
`1000Hza)Leftearlobe
`1000HzRightmiddlefinger
`
`400HzEarlobe
`
`shoulder,forehead
`forearm,wrist,
`
`66HzIndexfinger,medial
`
`100HzLeftearlobe
`
`400HzNotspecified
`
`Finger
`
`(reconstructedat2092Hz)
`128Hz
`
`200HzRadialartery
`1000HzRadialartery
`
`BiopacAcqKnowledgeII
`Self-madehandhelddevice
`
`1000HzNotspecified
`
`Colin-System
`
`OhmedaFinapres2300
`OhmedaFinapres2300
`
`OhmedaFinapres2300
`
`OhmedaFinapres2300
`
`→50×5min
`N=10,a)–e)5mineach
`
`InfraredPPGsensors(850nm)
`
`ADInstrumentsML305
`BIOPACtransducerTSD200
`
`1×5hofnightsleep
`15×4min
`15×6min
`
`ADInstrumentsPowerLab410
`
`44×6min
`
`aA/DconverterinaPC
`VariousPPGsensorslinkedto
`transmissionsensorPS-2105
`FeedbackInstrumentsPPG
`pulseoximetermodule
`NellcorMP506
`
`NellcorN-100pulseoximeter
`
`48×10min
`
`42×7min
`
`1724×2minepochsused
`Nightsleeprecords,
`
`40×30sanalyzed)
`(randomlyselected
`40×5min
`20×10s
`
`9×10min
`4minepochs
`subdividedinto
`Variousrecords,
`20×3×5min
`b)20×5min
`a)20×5min
`
`20HzDorsalsideofwrist
`
`30×6.0±0.8hofnightsleepDENSO“PrototypeC”
`
`Indexfinger
`
`(interpolatedto1000Hz)
`250Hz
`
`Leftmiddlefinger
`
`Exp.2)991Hz
`(downsampledto10–100Hz)
`Exp.1)1000Hz
`
`1000HzLeftfinger
`
`250HzIndexfinger
`
`TSD200transducer
`BiopacPPG100Camplifier+
`System1.5B
`2)FlexcompBiomonitoring
`
`1)BeckmanRMDynograph
`MP150+TSD123Btransducer
`BiopacPPG100Camplifier+
`sensor(860nm)
`BiopacMP30+SS4LPPG
`
`17×(4+5+4)min
`
`Exp.2)10×26min
`
`Exp.1)16×5min
`
`9×(10+20+20)min
`signals>1min)
`Notspecified(‘highquality’
`
`Positionofsensor
`
`Samplingrate
`
`Numberofrecords×durationDevice(s)used
`
`occlusion,followedbyreleaseperiod.
`Supinerest.Supradiastolicandsubsystolic
`Restinchair
`
`Treoetal.(2005)[49]
`N=40(hospitaloutpatients)
`Kristiansenetal.(2005)[48]N=20(healthy,ages19–51)
`Impedanceplethysmography(IP)studies
`
`Supinerest
`
`cardiomyopathy,ages45.1±7.3)
`N=9(patientswithdilated
`
`Suhrbieretal.(2006)[47]
`
`physicalormentaltasks
`Supineorsittingrest,various
`Supinerest
`4mincontrolledbreathing
`a)supineb)orthostatic1minspontaneous,
`
`studies,variousageranges)
`N=234(patientsfrommultiple
`N=20(healthy,ages22–75)
`and10controls,bothaged6–13)
`N=20(10childrenwithpace-maker
`
`McKinleyetal.(2003)[46]
`Dawsonetal.(1998)[44]
`
`Constantetal.(1999)[43]
`
`c)standingd)exercisee)recovery
`a)supineb)supineandcontrolledbreathing
`
`N=10(healthy,ages22.5±1.6)
`
`Carrascoetal.(1998)[42]
`Continuousbloodpressure(CBP)studies
`
`Supinerest
`
`N=15(healthy,ages23–38)
`
`Wongetal.(2010)[65]
`
`14×10min
`10×5min
`
`Supinerest
`Rest(nototherwisespecified)
`andcontrolledbreathing,respectively
`Restinchair;3minofspontaneous
`
`N=14(healthy,ages25.8±4.2)
`N=10(healthy,ages21–28)
`
`Shietal.(2008)[54]
`Selvarajetal.(2008)[53]
`
`N=44(healthy,ages16–60)
`
`Rauhetal.(2004)[27]
`
`Lyingsemi-erectinhospitalbed
`
`N=48(healthy,ages20–68)
`
`Nilssonetal.(2007)[67]
`
`Semi-recumbentrest
`
`N=42(healthy,age21±3)
`
`Luetal.(2009a)[58]
`
`10mininuprightpos.,10mininsupinepos.10×(10+10)min
`
`Nightsleep
`pulsedetectionbydevice)
`Nightsleep(otherwisenostable
`
`N=10(healthy,ages26±7.5)
`(OSA,ages54.9±16.3)
`
`Luetal.(2008)[57]
`
`Khandokeretal.(2010)[105]N=29(healthy,ages51.3±8.5)+22
`
`N=30(healthy,ages22–47)
`
`Hayanoetal.(2005)[55]
`
`Tilttable:supine–head-uptilt–supine
`
`N=17(healthy,ages28.5±2.8)
`
`Giletal.(2010)[56]
`
`Exp.2)mentalexercisetests
`
`Exp.2)N=10(healthy,ages25–50)
`
`Exp.1)restinchair
`slowwalk,bicycleexerciseinsupinepos.
`3sympatheticstimulations:orthostatictest,
`
`Exp.1)N=16(healthy,ages23–35)
`
`Giardinoetal.(2002)[52]
`
`N=9(healthymen,ages26.3±4.5)
`
`Charlotetal.(2009)[59]
`
`Restinchair
`
`N=10(healthymen,ages25–35)
`
`Changetal.(2007)[104]
`Photoplethysmographicstudies
`
`Experimentalpostureorconditions
`
`Nandtypeofsubjects
`
`Study
`
`SurveyofstudiesontheaccuracyofPRVcomparedtoHRV.Foranalysismethodsandresultscf.Table5.
`Table4
`
`5
`
`

`

`20
`
`A. Schäfer, J. Vagedes / International Journal of Cardiology 166 (2013) 15–29
`
`distributions
`PPIandRRI
`kurtosissimilarfor
`Skewnessand
`inconspicuous(t0)
`variables
`t-testsforall
`
`pacemakersubjects
`fornormaland
`supineandstanding,
`Resultsvalidfor
`respectively
`(almostall)HRVvars,
`differenceforall
`findsignificant
`(duringstanding)
`t-testsduringexercise
`
`ingrest
`HF(logarithms)dur-
`BA+forLF,BA−for
`
`studies
`datafromearlier
`secondaryanalysesof
`Resultsobtainedby
`
`inconspicuous(t0)
`variables
`t-testsforall
`
`overestimateHRV
`mostlytendsto
`available:PRV
`t-testsalso
`
`TP:BA+(c++)
`
`BA+(c++)
`
`BA+(c+)
`
`BA+(c++)BA+(c++)BA+(c++)
`
`ApEn:(c0)
`ApEn:(c0)
`
`(c++)
`
`c−
`
`BA+(c+)
`
`BA+(c++)
`
`t−
`
`(t0)
`
`TP:(c++)Δ+
`TP:c−Δ−
`TP:(c+)Δ−
`TP:(c++)Δ+
`TP:(c++)Δ++
`
`(c0)Δ0
`c−Δ−
`(c0)Δ−
`(c+)Δ+
`(c++)Δ+
`
`(c+)Δ0
`c−Δ−
`(c0)Δ−
`(c++)Δ0
`(c++)Δ+
`
`TP:BA−(c++)
`VLF:BA0(c++)
`
`BA−(c+)
`
`(c++)t−
`
`c−t−
`
`BA−(c++)BA−(c++)
`
`BA−(c+)
`
`(c++)t−
`
`c−t−
`
`(c+)t−
`(c++)t−
`
`SD1,SD2:BA+
`
`BA+
`
`BA+
`
`SD2:(c++)
`SD1:(c0)
`
`SampEn:u−
`SampEn:u−
`
`(c0)
`
`u−
`(u0)
`
`u−
`(u0)
`
`BA−
`
`BA+
`
`(c0)/c−
`
`(c0)
`(c0)
`
`(c+)
`
`(c+)
`(c++)
`(c++)
`
`(c++)
`
`BA0
`
`c−
`
`BA0
`
`c−
`
`(c0)
`
`(c0)
`
`u−
`(u0)
`
`(c0)
`
`(c++)
`
`(c+)
`(c++)
`
`(c+)
`(c++)
`
`(c++)
`(c++)
`
`(c++)
`(c++)
`
`BA−(c++)
`differs(−)
`Statistics
`
`BA−c−
`BA−c−
`BA−c−
`
`BA+(c++)
`
`BA−c−
`BA−c−
`BA−c−
`
`BA0(c++)
`BA0(c++)
`
`BA−c−
`BA−c0
`
`BA−(c+)
`BA−(c++)
`BA0(c++)
`BA+(c++)
`
`LF/HFratio
`
`[n.u./%]
`LFandHF
`
`HFpower
`
`LFpower
`
`pNN50
`
`Remarks
`
`Othervariables
`
`HRVfrequencydomainvariables
`
`BA+(c+)
`
`BA+(c++)
`
`BA++(c++)
`
`(c+)
`(c++)
`
`(c++)
`(c++)
`
`oftotalpower)
`(Diff.=1.35–3%
`
`Standing
`Supine
`
`Shi2008[54]
`
`Lu2008[57]
`
`BA0(c++)
`
`BA++(c++)
`
`BA++,(c++)
`
`Kristiansen2005[48](IP)
`Hayano2005[55]
`
`Detectionofdiastolicminima.
`manufacturer'ssoftware.
`andarrhythmiaswithdevice
`Preprocessingforartifacts
`negativeslopeincatacroticphase.
`(EMD).Detectingmaximum
`Empiricalmodedecomposition
`samplepulsecycle
`Crosscorrelationwitha
`Pulsefr

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket