AAPS PharmSciTech, Vol. 12, No. 4, December 2011 (© 2011)
`DOI: 10.1208/sl2249-0ll-9644-8
`
`Review Article
`Theme: Sterile Products: Advances and Challenges in Formulation, Manufacturing, Devices and Regulatory Aspects
`Guest Editors: Lavinia Lewis, Jim Agalloco, Bill Lambert, Russell Madsen, and Mark Staples
`
`Radiation and Ethylene Oxide Terminal Sterilization Experiences with Drug
`Eluting Stent Products
`
`Byron J. Lambert,1
`
`3 Todd A. Mendelson,1 and Michael D. Craven2
`'
`
`Received 19 May 2010; accepted 10 June 2011; published online 2 September 2011
`
`Abstract. Radiation and ethylene oxide terminal sterilization are the two most frequently used processes
`in the medical device industry to render product within the final sterile barrier package free from viable
`microorganisms. They are efficacious, safe, and efficient approaches to the manufacture of sterile
`product. Terminal sterilization is routinely applied to a wide variety of commodity healthcare products
`(drapes, gowns, etc.) and implantable medical devices (bare metal stents, heart valves, vessel closure
`devices, etc.) along with products used during implantation procedures (catheters, guidewires, etc.).
`Terminal sterilization is also routinely used for processing combination products where devices, drugs,
`and/or biologics are combined on a single product. High patient safety, robust standards, routine process
`controls, and low-cost manufacturing are appealing aspects of terminal sterilization. As the field of
`combination products continues to expand and evolve, opportunity exists to expand the application of
`terminal sterilization to new combination products. Material compatibility challenges must be overcome
`to realize these opportunities. This article introduces the reader to terminal sterilization concepts,
`technologies, and the related standards that span different industries (pharmaceutical, medical device,
`biopharmaceuticals, etc.) and provides guidance on the application of these technologies. Guidance and
`examples of the application of terminal sterilization are discussed using experiences with drug eluting
`stents and bioresorbable vascular restoration devices. The examples provide insight into selecting the
`sterilization method, developing the process around it, and finally qualifying/validating the product in
`preparation for regulatory approval and commercialization. Future activities, including new sterilization
`technologies, are briefly discussed.
`
`KEY WORDS: combination devices; drug eluting stents; ethylene oxide sterilization; material
`compatibility; radiation sterilization.
`
`INTRODUCTION
`
`Medical device, pharmaceutical, and biologic products
`provide a significant, positive impact to the quality of life
`of patients who receive
`them. Combination devices,
`which utilize technology spanning the medical device,
`pharmaceutical, and biopharmaceutical industries, have
`been growing and evolving. Combination devices are
`products comprised of two or more regulated compo(cid:173)
`nents, i.e., drug/device, biologic/device, drug/biologic, or
`drug/device/biologic, that are physically, chemically, or
`otherwise combined or mixed and produced as a single
`entity (1). More and more companies are creating novel
`
`1 R&D, Abbott Vascular, PO Box 9018 MS T561, Temecula, California
`92589, USA.
`2 R&D, Abbott Vascular, Santa Clara, California, USA.
`3 To whom correspondence should be addressed. ( e-mail: byron.
`lambert@av.abbott.com)
`
`drug delivery devices or are expanding the scope of existing
`devices with the addition of a drug or biologic compound
`(2,3). Abbott Vascular examples of combination devices
`are drug eluting stents (DES) ( 4,5) and bioresorbable
`vascular scaffolds (BVS) ( 6,7). At present, the DES
`market represents 60-70% (as high as 90% in China) of
`the $4B vascular stent industry and is growing at more
`than 7% per year worldwide (8). The use of temperature
`sensitive bioresorbable polymers for timed release of
`active agents is emerging, as are devices that utilize active
`electronics. Common to all of these medical product sectors
`with their sensitive materials as shown in Fig. 1 (9), is the need
`for safe, robust, cost-effective sterilization of product.
`In the world of medical devices, "sterilization" is defined
`as a "validated process used to render product free from
`viable microorganisms." Terminal sterilization is defined as
`the "process whereby product is sterilized within its sterile
`barrier system." (10) The terminal sterilization process is
`considered a manufacturing process step itself and usually
`takes place at, or near, the end of the manufacturing process.
`
`1530-9932/11/0400-1116/0 © 2011 American Association of Pharmaceutical Scientists
`
`1116
`
`Regeneron Exhibit 1049.001
`
`

`

`Terminal Sterilization and Drug Eluting Stent Products
`
`1117
`
`Devices
`
`Vascular
`Scaffolds
`
`Polymers
`
`Fig. 1. Sterilization: a common need across evolving combination
`product sectors with sensitive materials (9)
`
`Sterilizing product within the sterile barrier system is a very
`efficient approach for the manufacture of sterile product.
`Furthermore, terminal sterilization has exceptional process
`control and provides a high assurance of sterility (11). Note
`that sterilization or re-sterilization of products within the
`hospital setting is out of scope of this discussion.
`By definition, and in practice, terminal sterilization
`differentiates itself from aseptic processing where the final
`sterile product is realized over several manufacturing process
`steps. For aseptic processing, the products/components are
`sterilized separately and combined later in a sterile environment
`to produce the final sterile product. Great care must be taken to
`assure control over each process step to maintain sterility of the
`products/components. This involves capital expenditures and
`ongoing quality control expenses to achieve a comparatively
`lower assurance of sterility than terminal sterilization (11- 13).
`However, both sterilization approaches provide for the safe
`sterilization of the final medical product.
`Terminal sterilization is routinely applied to a wide variety
`of implantable medical devices and other medical products that
`are used during implantation procedures (14). Combination
`products with the device as the primary mode of action are
`sterilized using only terminal sterilization; there are no other
`options at this time. The practice of aseptic processing of solid
`combination devices, e.g., drug delivery devices, has only
`recently been considered (15). The application of terminal
`sterilization, apart from steam sterilization, with pharmaceut(cid:173)
`icals has been limited due to material compatibility challenges
`(16). Terminal sterilization of biologic products using radiation
`is also limited with the exception of tissue products for tissue
`banks (17). As the combination product market expands and
`evolves, so does the need to expand and evolve the application of
`terminal sterilization solutions.
`In this article, the authors will:
`
`• Introduce the basic concepts, definitions, benefits, and
`types of terminal sterilization used in the medical
`industry and provide an overview of related interna(cid:173)
`tional sterilization standards
`• Provide guidance and offer strategies for successful
`terminal sterilization process development and product
`sterilization qualifications highlighting case studies
`involving material compatibility challenges with drug
`eluting stents and vascular restoration devices
`• Outline next steps and future opportunities in develop(cid:173)
`ing effective terminal sterilization solutions for combi(cid:173)
`nation devices
`
`OVERVIEW OF TERMINAL STERILIZATION
`
`Terminal sterilization concepts, technologies, and standards
`are reviewed in this section. These perspectives provide a
`foundation for understanding the strong patient safety record
`of industrial terminal sterilization processes.
`
`Terminal Sterilization Concepts
`
`Patient Safety Issues Related to Infection
`
`Hospital acquired infections are a major societal concern.
`It is important to differentiate the sources of this problem. In
`particular, related to the topic of this article, it is important to
`ask the question if product processed by industrial terminal
`sterilization contributes to the problem. The answer appears
`to be a resounding "no."
`The Center for Disease Control (CDC) reviewed sources
`of hospital acquired infections for two sequential decades and
`found no incidents directly linked to terminally sterilized
`product (11,18,19). Why is this? The reasons become clear
`when industrial terminal sterilization processes are under(cid:173)
`stood and compared to hospital sources of infection (20) and
`other methods of manufacturing sterile product, e.g., aseptic
`processing or disinfection/liquid chemical methods.
`Exceptional process control is the primary reason for the
`strong quality record of terminal sterilization. As discussed in
`some detail below, terminal sterilization modalities provide a
`high level of process control to achieve a given sterility
`assurance level (SAL). In practice, while all parts of the
`product in the sterile barrier package confidently achieve the
`SAL, most locations of the product receive considerably
`greater assurance of sterility, often by several orders of
`magnitude (see "Sterility Assurance Level-Exponential
`Decay Curves" below).
`In contrast, aseptic processes are designed to exclude
`microbial contamination during the manufacturing process as
`opposed to killing it after the product is packaged. Process
`control over all variables that could contribute to microbial
`contamination is much more difficult to achieve than process
`control of a robust terminal sterilization process with a
`packaged product. Likewise, despite significant recent advan(cid:173)
`ces with liquid chemical sterilization processes (21), disinfec(cid:173)
`tion of geometrically complex devices followed by liquid
`chemical sterilization cannot match the process control of
`terminal sterilization. The superior patient safety results from
`terminally sterilized product explain the preference of
`regulatory bodies for terminal sterilization whenever possible
`(12) as well as their active participation in the sterilization
`standards development process.
`
`Definition of Sterility for Terminally Sterilized Products
`
`The International Organization for Standardization (ISO)
`definition of sterility is "free from viable microorganisms" (10).
`This definition implies zero microorganisms. A problem
`with this definition is the ability to test for and statistically
`verify achievement of the condition. Even with a practical
`surrogate, such as only one non-sterile unit in 1,000 or one
`million units, testing large quantities of expensive medical
`devices to this level is not practical.
`
`Regeneron Exhibit 1049.002
`
`

`

`1118
`
`Lambert, Mendelson and Craven
`
`Terminal sterilization process validation solves this prob(cid:173)
`lem. Microbial kill rates from ethylene oxide (EO) sterilization,
`radiation sterilization, and other sterilization modalities are
`exponential in nature (22). This allows the sterility of a product
`to be expressed as a probability based on the extent of exposure
`to the sterilization modality and the corresponding microbial log
`reduction. Achievement of a practical surrogate for sterility
`becomes experimentally achievable. This led the medical device
`industry and other industries facing similar challenges to
`quantify the effectiveness of a sterilization process by the
`probability of a non-sterile unit using the term SAL. The basis
`of quantification is microbial inactivation rate data, e.g., D
`values, the time or radiation dose required to achieve inactiva(cid:173)
`tion of 90% of a population of the test microorganism under
`stated conditions (10).
`
`Sterility Assurance Leve"-Exponential Decay Curves
`
`In North America, two healthcare SAL values have been
`used in practice, 10-3 or 10-6
`, the probability of one non-sterile
`unit in 1,000, or one million, units processed, respectively (23).
`Since SAL is a probability of contamination, the smaller
`number, 10-6
`, provides a greater assurance of sterility than the
`larger number, 10-3
`. An SAL of 10-3 has been permitted "if the
`patient risk is negligible, e.g., products not intended to come into
`contact with breached skin or compromised tissue or topical
`products that contact intact skin or mucous membranes."
`Examples include surgical drapes and gowns (14). Most
`combination devices are required to utilize a sterilization
`process that achieves the higher assurance of sterility, an SAL
`of 10-6 or one non-sterile unit in 1,000,000 units.
`An example of the relationship between the extent of the
`sterilization process and the resultant microbial log reduction
`is seen in Fig. 2. In this terminal sterilization example with
`radiation, as radiation dose increases, the number of surviving
`microorganisms drops essentially exponentially. The total
`dose required to get to a target SAL of 10-6 depends on the
`initial bioburden of the product. In this example, a 25-kGy
`
`dose is required to achieve the nine log reduction in
`bioburden from the initial level of 1,000 to an SAL of 10-6
`•
`For product with an initial bioburden level of 10, a seven log
`reduction in bioburden is required to achieve an SAL of 10-6
`,
`which could be achieved with a dose of less than 18 kGy.
`Importantly, the achievement of the one in a million sterility
`assurance level is the minimal requirement. Dose is not delivered
`as a mono-dose, but rather as a distribution of doses. It is the
`minimum portion of the dose distribution curve ( or below, if a
`common statistical safety factor is used) that achieves the SAL of
`10-6
`• The portion of the product that receives the top end of the
`dose distribution may receive a sterility assurance level better
`than one in 10,000,000 (SAL=l0-7 corresponding to the
`maximum dose with a dose uniformity ratio of 1.2; DUR=l.2)
`and more commonly something better than one in 100,000,000
`(SAL=l0-8 corresponding to the maximum dose with DUR=
`1.4) (24). This is an extraordinary margin of safety.
`Similar curves for EO sterilization demonstrate microbial
`log reduction as a function of time of exposure to EO gas for a
`given EO concentration, humidity level, and sterilization
`temperature (25). It is common when utilizing EO sterilization
`to use an overkill method of sterilization validation (26). This
`method essentially assumes that the initial bioburden consists of
`1,000,000 hardest to kill microorganisms. An EO cycle is
`validated to reduce this bioburden all the way to an SAL of
`10-6
`, resulting in a 12 log reduction of bioburden. In practice,
`with a reasonably well-controlled product bioburden of less than
`1,000, the assurance of sterility for all product in an EO sterile
`load is one in 1,000,000,000 (SAL=l0-9
`). Again, this is truly
`overkill and provides exceptional patient safety.
`In light of the very high levels of microbial reduction
`from these standard terminal sterilization processes, the
`strong patient safety record for industrial sterilization is
`indeed not surprising. The technologies that achieve this
`robust assurance of sterility are discussed in the next section.
`This is followed by a discussion of challenges involved in
`qualifying sensitive materials associated with combination
`devices.
`
`~
`
`...._
`
`..___
`
`1.E+03
`
`1.E+02
`
`1.E+01
`
`Ill 1.E+OO
`"' ::, ..,
`"' 1.E-01
`C > >
`:i
`V)
`'o
`..,
`.; 1.E-03
`E
`::, z 1.E-04
`
`1.E-02
`
`1.E-05
`
`1.E-06
`
`1.E-07
`
`-
`
`-
`
`Selected SAL
`
`'\ "'
`""
`"" I'..
`"'
`""'
`""' I'...
`
`High Initial Bioburden -
`-
`-
`
`Low Initial Bioburden
`
`~
`
`0
`
`5
`
`10
`
`15
`Dose(kGy)
`Fig. 2. Log reduction of microorganism as a function of radiation dose
`
`20
`
`25
`
`30
`
`Regeneron Exhibit 1049.003
`
`

`

`Terminal Sterilization and Drug Eluting Stent Products
`
`1119
`
`Terminal Sterilization Technologies: a Brief Introduction
`
`There are several technologies that can provide terminal
`sterilization. Some of these technologies are EO, radiation,
`moist heat (steam), dry heat, hydrogen peroxide, ozone,
`chlorine dioxide, supercritical carbon dioxide, and nitrogen
`dioxide. Ethylene oxide and radiation are the most commonly
`used technologies to terminally sterilize medical devices (27)
`due to their robust microbial kill, broad material compati(cid:173)
`bility, and ability to process high volumes of product at
`reasonable costs. Moist heat and dry heat are most commonly
`applied to pharmaceutical products and components (16);
`they are not applied to combination devices since the devices
`typically have polymeric components that cannot withstand
`the high temperatures.
`
`Ethylene Oxide Sterilization
`
`Fig. 3. Abbott Vascular two-pallet EO chamber
`
`Ethylene oxide sterilization accounts for approximately
`50% of the industrial terminal sterilization market (27) and is a
`conceptually simple terminal sterilization process. A fully func(cid:173)
`tional finished good device is placed into a sealed breathable
`packaging system that allows ingress and egress of EO and
`humidity but is microbially resistant. For most industrial
`applications, packaged product is palletized (approximately
`three cubic meters) in well-defined and validated configurations.
`Based on the size of the EO chamber, one to 40 pallets are
`combined to create an ethylene oxide sterile load. Product must
`be humidified to assure microbial kill; this is sometimes
`accomplished prior to placing product in the EO chamber but
`increasingly it is accomplished in the EO chamber itself through
`dynamic humidity pulsing.
`In the ethylene oxide chamber product is exposed to a
`validated combination of humidity, ethylene oxide gas,
`temperature, and time. Deep vacuum cycles are often used to
`drive humidity and ethylene oxide into palletized product.
`Following the sterilization process, EO levels are brought below
`permissible exposure limits through completion of a validated
`in-chamber vacuum purge process or a post-sterilization aera(cid:173)
`tion process. Product is released for distribution following
`review and documentation of routine monitoring parameters
`and, in many instances, biologic indicator test results (28). Total
`cycle times range from 6 hours to several days.
`Ethylene oxide is a highly reactive cyclic ether with two
`carbons and one oxygen, CH2CH20. It is a gas at room
`temperature with a boiling point of 11 °C. It is pressurized and
`stored as a liquid for use in EO processing plants. The mechanism
`of microbial kill is alkylation of the amine groups of DNA (25).
`Moisture facilitates microbial kill; as noted above, product and
`thus the microbes, must be exposed to a humid environment
`before EO exposure (25). EO kill rate is a function of temper(cid:173)
`ature and concentration of EO gas (29). Shown in Fig. 3 is a two(cid:173)
`pallet EO sterilization chamber used by Abbott Vascular.
`
`Radiation Sterilization
`
`Radiation sterilization accounts for most of the remain(cid:173)
`ing 50% of the industrial terminal sterilization market (27).
`Fully functional finished good devices are placed and sealed
`within a sterile barrier packaging system according to a
`defined product orientation. The product is loaded onto a
`
`conveyor system using a specified orientation and passed
`in front of a radiation source that emits electrons or
`photons that penetrate through the packaging and inacti(cid:173)
`vate the device's microbial load. One parameter, radiation
`dose, correlates directly with microbial kill and is easily
`measured to provide process control. The mechanism for
`microbial kill is radiation induced scission of DNA chains,
`either "direct" (i.e., direct scission of DNA chains) or
`"indirect" (i.e., scission mediated by formed radicals),
`which stops microbial reproduction (25). There are three
`radiation sterilization modalities: gamma, electron beam,
`and X-ray (24).
`
`Gamma Sterilization. Gamma sterilization uses cobalt-
`60, a radioactive element that undergoes nuclear decay
`producing useful gamma radiation. These photons have a
`very large penetration capability, easily penetrating
`through two or more pallets of product (30). Racks of
`cobalt-60 rods provide the radiation source. A conveyor
`system moves many totes of fully packaged product into
`the sterilization chamber and around the racks, often
`passing by multiple times, to sterilize the product. The
`dose is related to
`the amount of exposure time the
`product experiences, typically ranging from 4 to 8 hours.
`
`Electron Beam Sterilization. Electron beam (E-beam)
`sterilization relies on high-energy electrons to accomplish
`sterility. Electrons are commonly accelerated up to 0.2 to
`10 Me V and delivered as a continuous curtain or magnetically
`focused into a 1-5-cm-diameter beam that is magnetically
`scanned at high frequency across the product as it moves
`in front of the beam on the conveyor system. Low energy
`E-beam is used for surface sterilization of pharmaceutical
`packaging whereas high energy E-beams are used for fully
`packaged medical devices. Electrons from accelerators do
`not penetrate nearly as far as photons from gamma
`sources (30), so product is often processed in single
`product cartons or small corrugated shipper boxes. Shown
`in Fig. 4 is an illustration of a self-shielded E-beam
`accelerator and conveyor system. Products packed in
`corrugated shipper boxes are loaded on the conveyor
`system. Product is carried through the electron beam to
`achieve the desired irradiation dose, typically in a few
`
`Regeneron Exhibit 1049.004
`
`

`

`1120
`
`Product L:oa i:I ~I
`
`Fig. 4. Abbott Vascular self-shielded electron beam sterilization
`system
`
`seconds. Product is then returned sterile to an unload/
`product release station.
`
`X-ray Sterilization. X-ray sterilization is a hybrid between
`gamma sterilization and e-beam sterilization. Radiation is
`generated from high-energy electrons from accelerators, typi(cid:173)
`cally using electrons with energies of 5-7.5 MeV The X-ray
`photons behave nearly identical to photons from gamma sources
`in terms of energy deposition and high penetration capabilities.
`Utilization of X-ray sterilization is limited but increasing.
`
`Overview of Standards
`
`A great asset in the application of terminal sterilization to
`combination devices is the availability of clear requirements and
`guidance in the form of national and international consensus
`standards for major sterilization technologies (see Table I).
`These standards are developed cooperatively by regulatory
`authorities, industry users of terminal sterilization, industry
`providers of contract terminal sterilization services or equipment,
`and, as needed, academia. The standards are robust with
`sterilization validation and routine control practices that have
`been in practice for decades. The standards were born out of the
`realization that all parties benefit from having a common
`understanding of best practice. National and international stand(cid:173)
`ard requirements for these important horizontal sterilization
`technologies touch the entire medical device industry and certain
`portions of the pharmaceutical and biopharmaceutical industries.
`
`Lambert, Mendelson and Craven
`
`The standards for each sterilization technology use a
`common template (31) to establish a sterilization process that
`reliably and reproducibly provides the intended sterility assur(cid:173)
`ance level when sterilizing medical products. The concepts in the
`sterilization standards are reviewed below to give the reader a
`sense of their scope which leads to their strong safety
`record. The systematic approach of the standards ensures
`consistency across the sterilization technologies in key
`areas such as utilization of a quality management system,
`characterization and definition of the sterilizing agent,
`sterilization process and equipment, product qualification,
`validation of the process, and monitoring, control, and
`maintenance of the process. This provides for a robust,
`safe approach to sterilize medical products. Although the
`intent to understand and control the process is analogous
`to process analytical technology (PAT), the approach for
`terminal sterilization is concerned with inputs and outputs of the
`sterilization process. Aseptic processing, and PAT, on the other
`hand, involves the characterization and control of inputs,
`outputs, and interactions of multiple processes.
`A typical first step in the application of a terminal
`sterilization process is the identification of a process compatible
`with the product, which includes all product components and
`packaging. Once a suitable process is identified, it is common to
`consider "Product Definition" which includes establishment of
`product families based on product characteristics germane to the
`given sterilization process. This is typically followed by "Process
`Definition" which includes experimental establishment of the
`minimum extent of processing required to assure sterility and
`the maximum extent of processing above which product
`functionality will be compromised. The key challenge for
`combination devices that incorporate pharmaceuticals and/or
`biologics is finding a process window that fits within these
`constraints, as discussed later.
`Once the product and process are defined the process
`can be validated. Each standard provides requirements and
`best practice guidance for installation qualification and
`operational qualification of the sterilization equipment. The
`heart of performance qualification involves the definition of a
`load configuration and experimental verification that the
`proposed production process will achieve sterility (process
`stays above the minimum extent of processing) and avoid
`product functionality concerns (process stays below the
`maximum extent of processing). For radiation sterilization,
`this involves mapping the dose received in the load config(cid:173)
`uration. For ethylene oxide sterilization, this involves assuring
`EO penetration and kill within the load configuration as well
`as mapping product temperature and humidity distributions
`
`Table I. Sterilization Standards and Guidance-References
`
`Radiation sterilization
`
`Ethylene oxide
`
`Moist heat (saturated steam)
`
`Other
`
`EN/ANSI/AAMI/ISO 11137-1 Sterilization of health care products-radiation-part 1: requirements for
`development, validation, and routine control of a sterilization process for medical devices
`EN/ANSI/AAMI/ISO 11135-1 Sterilization of health care products-ethylene oxide-part 1: requirements
`for the development, validation, and routine control of a sterilization process for medical devices
`EN/ANSI/AAMI ISO 10993-7 Biological evaluation of medical devices, part 7:
`ethylene oxide sterilization residuals
`EN/ANSI/AAMI/ISO 17665-1 Sterilization of health care products-moist heat-part 1: requirements
`for the development, validation, and routine control of a sterilization process for medical devices
`AAMI ST67 Sterilization of health care products-requirements for products labeled "STERILE"
`AAMI TIR 17 Compatibility of materials subject to sterilization
`
`Regeneron Exhibit 1049.005
`
`

`

`Terminal Sterilization and Drug Eluting Stent Products
`
`1121
`
`within the product. Demonstrating the reduction or dissipa(cid:173)
`tion of sterilant residuals is also required for EO sterilization.
`Once validated, requirements of a quality management
`system are designed to ensure the process is monitored,
`appropriately controlled and maintained to provide the
`intended sterilization of medical products.
`In summary, terminal sterilization technologies provide
`efficient and robust validated processes to assure patient
`safety. However, relative to combination devices, terminal
`sterilization technologies will provide no benefit to patients if
`sensitive materials are not compatible with them. Avoiding
`this problem is the focus of the next section.
`
`APPLICATION OF TERMINAL STERILIZATION:
`GUIDANCE AND CASE STUDIES
`
`Successful application of terminal sterilization requires the
`selection of an appropriate sterilization modality, qualification
`of materials subject to the sterilization process, optimization of
`the sterilization process, demonstration of stability of the
`product over its shelf-life, and regulatory approval. Combination
`products provide an additional challenge to terminal sterilization
`since these products can incorporate technologies from other
`industries that are not typically terminally sterilized. These topics
`are addressed in this section along with case studies of drug
`eluting stents and bioresorbable vascular restoration devices.
`
`Guidance on Selecting a Sterilization Method
`
`Selecting a terminal sterilization method for a product
`depends on many factors, but two primary factors are central to
`the decision: ability to achieve the desired sterility assurance
`level and compatibility and stability of the associated materials.
`The selected sterilization method must demonstrate the
`required sterility assurance level for the packaged product, and
`the product (and package), once sterilized, must meet intended
`performance requirements, which include lifecycle/shelf-life
`requirements. Secondary factors that may also influence the
`decision include company preferences, sterilization costs, avail(cid:173)
`ability of in-house sterilization technologies, relationships with
`sterilization service providers, knowledge of use, and impact on
`predicate or similar products. In selecting a method the use of
`standards and guidance documents is recommended, especially
`if new materials or sterilization methods are being considered.
`Examples of two such documents are the Technical Information
`Report titled Compatibility of Materials Subject to Sterilization
`(16) (for healthcare manufacturers; covers six sterilization
`modalities and relates to products manufactured from polymers,
`ceramics, and metal with brief discussion of pharmaceuticals and
`biologics) and a Committee for Proprietary Medical Products
`document titled Decision Trees for the Selection of Sterilization
`Methods (32) (for development of pharmaceutics; applies to
`aqueous products, non-aqueous liquids, semi-solids, and dry
`powder products). Sterilization technology review articles
`related to implantable materials and different technologies may
`also be helpful (33- 35).
`Selecting a terminal sterilization method can begin by
`determining if the desired sterility assurance level is achievable
`for the product packaged within the sterile barrier system.
`Product designed, or assembled, such that an interior surface of
`the product is nearly closed off to the external environment
`
`would not be a likely candidate for EO sterilization due to the
`need for moisture and EO gas to reach and interact with
`microbes on that interior surface to destroy their DNA. If EO
`sterilization was desired, based on other factors influencing this
`choice, consideration could be given to how the product is
`assembled and packaged ( e.g., if a stopcock is attached to a
`syringe, can the stopcock be left in an open position to allow
`moisture and EO gas to reach the inner surface of the syringe?).
`On the other hand, if a device includes active electronics, then
`radiation sterilization is not likely to be compatible (16).
`An important and challenging next step in selecting a
`terminal sterilization process would be to understand the
`compatibility of a product subject to a particular sterilization
`process. Product design, materials, and how the product/
`materials are manufactured are key compatibility factors to
`consider (16). For example, if the design or manufacturing
`processes result in residual stresses within the material, one
`sterilization method may be more aggressive than another in
`terms of impact to physical property degradation and the
`resultant product performance. Consideration must be given to
`potential changes in physical properties, chemical properties,
`and the functional performance of the product.
`The most significant part of understanding product com(cid:173)
`patibility with a terminal sterilization modality, once initial
`literature and guidance documents have been reviewed, is
`clinically relevant evaluation of the product performance.
`Regardless of the sterilization method being considered, it is
`important that the effects of that method on the particular
`product being developed are well understood early during the
`design and development phases of the project. Emphasizing this
`point further, consider that

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket