throbber
INTERNATIONAL
`STANDARD
`
`ISO
`7864
`
`Third edition
`1993-05-15
`
`Sterile hypodermic needles for single use
`
`Aiguilles hypodermiques sleriles, non reut/lisables
`
`;:: C:
`
`O= = C:
`:ii -
`
`Reference number
`ISO 7864:1993(E)
`
`Regeneron Exhibit 1042.001
`
`

`

`ISO 7864:1993(E)
`
`Contents
`
`Page
`
`=
`
`....
`-
`
`1 Scope .... .. .......... ......... ................ ................ .. ... ................................... 1
`
`2 Normative references .. .. . ..... ..... .. ... .. .... .. .. ......... ....... .. ... .. .. ................ 1
`
`3 Nomenclature
`
`........... ........... ................ ...... .............. ,........ ......... .. 1
`
`4 Clean liness
`...... .. .. .. ......... ... ..... .................................... ......... ...... .. 1
`s Limits for acid ity or alkalinity
`
`....... .. ..................... ............ .......... 1
`
`6
`
`lim its for extractab le metals
`
`............. ........................ .............. .. 1
`
`7 Size designation
`
`..... ........... ..... ...... .. ........ ............... .. .................. .. 2
`
`8 Colou r coding
`
`......... , ... ,, ....... ...... ..... ... ... ,,.... .................. ............... 2
`
`9 Needle hub
`
`. ..... .. ....... .. ....... .. .... .. ... .. ............................ . ..... .. ....... .. 2
`
`10 Sheath
`
`.......... ....... ....... .. .. ....... .... ....... ......... ... ....... ... .. ... ... ... .. ... ... .. 2
`
`11 Need le tube
`
`....... ........... ... .................................... ....... ............... 3
`
`12 Need le point
`
`...... .. ... ....... ............ ...... , ................ .. ,...... ............... . 4
`
`13 Performance
`
`...... ................... ... ....... .................. ......... ........ ... ..... 4
`
`14 Packaging
`
`............... ................ , ..... ......... .............. ..... , ...... .......... 5
`
`15 Labelling
`
`,, ... ... ...... ..... .............. , ................. ............. .. . , ..... .. .... ,.... 5
`
`Annexes
`
`A Melhod for preparation of extracls
`
`..... .. .... .. .. ..... .. .... ..... ........... 7
`
`B Example of method for determination of fragment prod uction from
`rubber closures
`.. ..... .. ...... ....... .............. , ..... , .. , ...... . , .................... 8
`
`C Symbol for " do not re-use "
`
`....... ..... .......... .. ...... .. ... .... ... .. ... ...... 10
`
`D Bibliography
`
`..... ....... ... ....... ...... .. .. ........ ............... ............... .. .. ... . 11
`
`tl ISO 1993
`All rights reserved. No part of this publication may be reproduced or utilized in any form
`or by any means, electronic or mechanical, Including photocopying and mlcrornrn, wlttiout
`permission In writing from the publisher.
`International Organization Jor Standardization
`Case Postale 56 • CH-1211 Geneve 20 • Switzerland
`Printed In Switzerland
`
`ii
`
`Regeneron Exhibit 1042.002
`
`

`

`ISO 7864:1993(E)
`
`=
`
`Foreword
`
`ISO (the lnlerna!ionat Organization for Standardization) Is a worldwide
`federation of national standards bodies (ISO member bodies). The work:
`of preparing International Standards is normally carried out through ISO
`technical committees. Each member body interested in a subject for
`which a techni cal committee has been established has the right to be
`represented on that committee. International organizations, govern(cid:173)
`mental and non-governmental, in liaison with ISO, also take part in the
`work. ISO collaborates closely with the in!erna1ional Electrotechnical
`Commission (IEC) on al l matters of electrotechn ical standardization .
`
`Draft International Standards adopted by the technical committees are
`circu lated to the member bodies for voting. Publication as an Inter(cid:173)
`national Standard requires approval by at least 75 % of the member
`bodies casting a vote.
`
`International Standa rd ISO 7864 was prepared by Technical Comm ittee
`ISO/TC 84, Medical devices
`for injections , Sub-Committee SC 1,
`Syringes, needles and intravascu/ar catheters for single use.
`
`second edition
`the
`replaces
`third edition cancels and
`This
`(ISO 7864:1988). of which it constitutes a technical revision.
`
`The major differences between this edition and 1he 1988 edition are as
`follows .
`
`a) Th is International Standard specifies the use of needle tubing com(cid:173)
`ISO 9626. As
`requirements
`for metallic materials,
`plying With
`stiffness , res istance to breakage and resistance to corrosion are
`given in ISO 9626, they have been deleted from this International
`Standard. The preparation of ISO 9626 has also allowed the intro(cid:173)
`duction ot new, smaller outside diameters of needle tubing and of
`tubing of thin- and extra-thin-walled types into this lnternaUonal
`Standard , In order to avoid inh ibiting innova1ion , this International
`Standard no longer recommends combinations of needle diameter
`and length.
`
`b) Additional information and guidance have been introduced on needle
`point geometry and frag mentation properties , and the Jimjted num(cid:173)
`ber of tests for toxicity given i_n the 1988 edition has been replaced
`by an informative annex that lists a significantly greater number of
`re levant biologi.cal tests.
`
`c} This International Standard permits the use on package labell ing of
`the ISO symbol for ''do not re-use ", but continues to require the
`written word. Manufacturers a.re encouraged to use the symbol so
`as to increase familiarity wlth it among purchasers and users.
`
`Annex A forms an integral part of this International Standard. Annexes
`B. C and D are for information only.
`
`iii
`
`Regeneron Exhibit 1042.003
`
`

`

`ISO 7864:1993(E)
`
`Introduction
`
`T.his International Standard covers sterile hypodermic needles intended
`for single use primarily In humans.
`
`- This International Standard does not give requ irements or test methods
`for freedom from biological hazard because international agreement
`upon the methodology and the pass/fail criteria is incomplete. Guidance
`is given
`in
`tests relevant to hypodermic needles
`on biological
`ISO 10993-1 , and it is suggested that manufacturers take th is guidance
`into account when evaluating products. Such an evaluation should in(cid:173)
`clude the effects of the process whereby the need les are sterilized .
`However, national regulat ions may exist in some countries, and these
`will override the guidance in ISO 10993-1.
`
`- Plasttcs materials to be used for t he construction of needles are not
`- specified as their selection will depend to some extent upon the design,
`process of manufacture and method of sterilization employed by indi(cid:173)
`vidual ma nufacturers. The materials should be compatible with injection
`fluids included in relevant pharmacopoeiae .
`
`Hypodermic needles specified in this International Standard ate in(cid:173)
`tended for use with hypodermic syringes specined in ISO 595 and
`ISO 7886-1. They will also fit syringes of types 1 and 2 specified in
`ISO 8537.
`
`In some countries, na1ional pharmacopoeiae or government regulations
`are 1egally binding and their requirements may take precedence over
`th is International Standard.
`
`Regeneron Exhibit 1042.004
`
`

`

`INTERNATIONAL STANDARD
`
`ISO 7864:1993(E)
`
`Sterile hypodermic needles for single use
`
`1 Scope
`
`This International Standard specifies requirements
`ror sterile hypodermic needles ror single use of
`nominal outside diameters 0,3 mm and 1,2 mm.
`
`It does not apply to dental needles.
`
`2 Normative references
`
`The following stahdatds contain provisions which,
`through reference In this text, constitute provisions
`of this International Standard. At the time of publi(cid:173)
`cation, the editions indicated were valid . All stan(cid:173)
`to
`dards are subject
`to
`revision, and parties
`agreements based on this International Standard
`are encouraged to investigate the possibility of ap(cid:173)
`plying the most recent editions of the standards in(cid:173)
`dicated below. Members of IEC and ISO maintain
`reg isters of currently va lid International Standards.
`
`ISO 594-1:1986 , Conical fittings with a 6 % (Luer) ta(cid:173)
`per for syringes, needles and certain other med/ca/
`equipment - Part 1: General requirements ,
`
`ISO 594-2:1991, Conical fittings with a 6 % (Luer) ta(cid:173)
`per for syringes, needles and certain other medical
`equipment - Part 2: Loci< fittings.
`
`ISO 3696:1987, Water for analylical laboratory use -
`Specification and test methods.
`
`ISO 8601:1988, Data elements and interchange for(cid:173)
`Information interchange - Representation of
`mats -
`dates and times.
`
`ISO 9626:1991, Stainless steel needle tubing for the
`manufacture of medical devices.
`
`3 Nomencjature
`
`The nomencla1ure ror components of hypodermic
`needles ror single use is shown in figure 1 together
`with the designation for length /; nomenclature for
`needle points is shown in figure 2.
`
`4 Cleanliness
`
`When inspected by normal or corrected-to-normal
`vision without magnification under .an ilfuminance
`of 300 Ix to 700 Ix, the surface of the hypodermic
`needle tube shatl appear free from particles and
`extraneous matter.
`
`When exam ined under x 2,5 magnification, the hub
`socket shall appear free from particles and extrane(cid:173)
`ous matter.
`
`5 Limits for acidity or ._,kallnlty
`
`When determined with a laboratory pH meter and
`using a general purpose electrode, the pH value of
`an extract prepared in accordance with annex A
`shall be within one unit of pH of that of the control
`fluid.
`
`ISO 6009:1992, Hypodermic needles tor single use -
`Colour coding for identification.
`
`6 Limits for extractable metals
`
`ISO 7886-1:-11, Sterile hypodermic syringes for single
`use - Part 1: Syringes for manual use.
`
`recognized microanalytical
`tested by a
`When
`method,
`for example by an atornic absorption
`
`I} To be published.
`
`1
`
`Regeneron Exhibit 1042.005
`
`

`

`ISO 7864:1993(E)
`
`method, an extract prepared in accordance with an(cid:173)
`nex A shal l, when corrected for the metals content
`of the control Ouid, contain not greater than a com(cid:173)
`bined total of 5 mg/I of lead, tin, zinc and iron. The
`cadm ium content of the extract shall , when cor(cid:173)
`rected for the cadmium content of the control fluid,
`be lower than 0,1 mg/I.
`
`7 Size designation
`
`ance with ISO 6009 applied to the unit conta iner
`and/or part of the needle assembly such as the
`needle hub or the sheath .
`
`9 Needle hub
`
`9.1 Conical fitting
`
`The size of hypodermic needle shall be designated
`by the following :
`
`=
`
`a) lhe nominal outside diameter of the needle tube ,
`expressed in mill imetres;
`
`The conica l socket or the hypoderm ic needle hub
`shall be in accordance with ISO 594-1,
`
`If the hub has a locking fitting. It shal l be In accord(cid:173)
`ance with ISO 594-2.
`
`b) the nominal length of the needle tube, expressed
`in millimetres.
`
`9.2 Colour of hub
`
`The size shall be referred to as "the designated
`metric size " and shaJI be expressed in mill imetres.
`
`The hub shall be made either of pigmented or of
`unpigmented materia l. If plgrnen1ed, the colour shall
`be in accordance with ISO 6009.
`
`EXAMPLE
`
`0,8 X 40
`
`8 Colour coding
`
`The nomina l outside diameter of hypodermic nee(cid:173)
`dles shall be identified by colour coding in accord-
`
`10 Sheath
`
`If a separate needle sheath is provided, it sha ll be
`made either of pigmen1ed or of unplgmented ma(cid:173)
`terial. If pigmented , the colour shall be in accord(cid:173)
`ance with ISO 6009.
`
`1
`
`2
`
`3
`
`4
`
`Key
`
`1
`2
`3
`4
`
`Hub
`Jointing medlum
`Needle tube
`Sheath
`
`Figure 1 - Example of typical hypodermic needle and sheath for single use
`
`2
`
`Regeneron Exhibit 1042.006
`
`

`

`-,!lt,_ ir t.il
`•J j_::-,,. =-~'J1.1111u I..J1•.lJ
`
`da
`
`d1
`
`' t-
`
`Co
`
`C2
`
`Bo
`
`82
`
`t
`
`View showing Vlll'lant when primary
`bevel Intersects needle tip
`
`·--·
`
`Neede rotated by llJ'lgl.e GO
`
`Key
`d O outside diameter of needle tube
`d 1 Inside dla.meter of needle tube
`A Point length
`B0 Pr1mn.ry bevel nominal Length Bo• (A-CO )
`8 1 Rlght !Jrllllllry bevel length
`82 Left primary bevel Length
`CO Secandclry bevel nom/no.l length
`C, Right secondary be11el length
`
`A
`
`C 2 Lett secondary bevel length
`a. Primary benl angle
`!II Secondary bevel angle
`(J Tip angle
`81 Right Sl!lconda.ry bevel rctn tlon angle
`8z Lett secondary bevel. rotation angle
`1 Camblned secondlll"y bevel angle
`
`~
`:.i.
`
`CJ>
`
`0 ... m a,
`IO :
`"' -
`
`.,,
`c
`C
`al
`
`~
`
`1
`0
`(D
`Ill ca·
`
`:::,
`
`:::,
`
`Al .. 5·
`0 -a.
`
`3 (D
`:::,
`Ill
`0 :::,
`
`Ill
`Al
`::,
`a.
`:I
`0
`3
`CD
`:::,
`n
`
`;' e-... Ill
`0 -:::,
`
`(D
`(D
`Cl.
`ii"
`"ti
`0
`....
`5'
`CiCI
`CD
`0
`:I
`Ill 5
`
`~
`(II
`i,
`~
`n
`Qi"
`C:
`(h
`~
`
`~
`
`:::a
`C'D cc
`C'D
`:J
`..,
`C'D
`0
`:J
`m
`><
`::::r
`-· C"
`;::::;:
`
`~
`0
`.i::i,.
`l'IJ
`0
`0 ......
`
`(.)
`
`

`

`ISO 7864:1993(E)
`
`11 Needle tube
`
`11 .1 General
`
`The needle shall be made of tubing in accordance
`with ISO 9626.
`
`11.2 Tolerance.son length
`
`The actual length of the need le tube (see dimension
`I in figure 1) shall equa l the nominal length within the
`= tolerances given in table 1.
`
`needle points shown are of configurations com(cid:173)
`mon ly manufactured: other configurations may be
`equally satisfactory. It may not be necessary to use
`all the dimensions when describing the point con·
`figuration .
`
`The needle point should be designed so as to mini(cid:173)
`mize coreing and fragmentation when penetrating
`vial closures . This International Standard does not
`specify requirements or test methods for these
`properties, but an example of a test method for de·
`termining the production of fragments from rubber
`closures is given in annex B.
`
`Table 1 - Tolerances on length of needle tube
`Dimensions in millimetres
`
`13 Performance
`
`Nominal length of needle
`tube
`
`Tolerance
`
`13.1 Bond between hub and needle tube
`
`< 25
`
`25 to 39
`
`40
`
`>40
`
`+1
`-2
`
`+1,5
`- 2,5
`
`Q
`-4
`
`+1,5
`-t5
`
`The union of the hub and needle tube shall not be
`broken by the minimum force given in table 2 ap(cid:173)
`plied as push or pull in 1he direction of the needle
`axis.
`
`Table 2 - Force to test bond between hub and
`needle tube
`
`11.3 Freedom from defects
`
`When examined by normal or corrected vision, the
`needle tube shall appear straight and of regular
`cross-section and wall thickness.
`
`11 .4 Lubricant
`
`If the hypoderm ic need le tube is lubricated, the
`lubricant shall not be visible, under normal or cor(cid:173)
`rected vision, as droplets of fluid on the outside or
`inside surfaces of the needle tube.
`
`NOTE 1 An acceptable lubricant, applied undiluted, is
`polydimethylsiloxane complying with a national or the
`European pharmacopoeia,. The quantily of lubricant used
`should not exceed 0,25 mg per square centimetre of the
`surface area of the needle tut)e,
`
`12 Needle point
`
`When examined under x 2,5 magnification , the nee(cid:173)
`dle point sha ll appear sharp and free from feather
`edges, burrs and hooks.
`
`NOTE 2 The needle point usually has a bevel with a
`(11 ± 2) 0
`primary bevel angle of
`in
`(as
`illustrated
`figure 2). but a ''short" bevel with other angle, e,g.
`(17 ± 2)°, may be provided.
`
`The designation of needle point dimensions and the
`nomenclature used to describe the dimensions and
`features is shown for information In figure 2. The
`
`4
`
`Nominal outside diameter
`of needle
`
`mm
`
`0,3
`0,33
`0,36
`0,4
`0,45
`0,5
`0,55
`0,6
`0,7
`0,8
`0,9
`1,1
`1,2
`
`Force
`
`min.
`
`N
`
`22
`22
`22
`22
`22
`22
`34
`34
`40
`44
`54
`69
`69
`
`13.2 Patency of lumen
`
`The patency of the lumen shall be such that either
`
`a) a stainless steel stylet of the appropri ate diam(cid:173)
`eter se lected from the diameters given in table 3
`sha ll pass through the needle; or
`
`b) the rate of flow of water through the needle Un·
`der a hydrostatic pressure not exceeding
`
`Regeneron Exhibit 1042.008
`
`

`

`ISO 7864:1993(E)
`
`1 x 105 Pa 2J shall be not less than 80 % of that of
`a needle of equ ivalent outside dfameter and
`length having a minimum inside diameter in ac(cid:173)
`cordance with ISO 9626 when tested under the
`same pressure.
`
`d) that once opened , the container cannot be easily
`resealed, and it should be obvious that the con(cid:173)
`ta iner has been opened.
`
`14.2 Secondary container
`
`Table 3 - Size of slylet to test patency of lumen
`Dimensions In milllrnetres
`
`One or more pri mary containers shall be packaged
`in a secondary container.
`
`Nominal
`outside
`diameter of
`needle
`
`Diameter of stylet
`0
`-D,01
`
`for needle
`of normal-
`walled
`tubing
`
`for needle
`of
`thin-walled
`tubing
`
`for needle
`of
`extra-thin-
`walled
`tubing
`
`The secondary container should be sufficiently ro(cid:173)
`bust to protect the contents during handlftig, transit
`and storage.
`
`One or more secondary containers may be pack(cid:173)
`aged in a storage and/or a transit container.
`
`0,3
`0,33
`0,36
`0,4
`0,45
`0,5
`0,55
`0,6
`0.7
`0,8
`0,9
`1, 1
`1,2
`
`0,11
`0, 11
`0,11
`0,15
`0,18
`0,18
`0,22.
`0,25
`0,30
`0,40
`0,48
`0,58
`0,70
`
`0,13
`0,15
`0,15
`0,19
`0,23
`0,23
`0,27
`0,29
`0,35
`0,42
`0,49
`0,60
`0,73
`
`-
`
`-
`-
`-
`-
`-
`-
`0,30
`0,37
`0,44
`0,50
`0,68
`0,83
`
`14 Packaging
`
`14.1 Primary container
`
`Each hypodermic needle shall be sealed in a pri(cid:173)
`mary container. ihe material and design of this
`container shall be such as to ensure that the colour
`coding of the contents is visible.
`
`The materials of the container should not have det(cid:173)
`rimental elfects on the contents. The materlals and
`design of this container should be such as to ensure
`
`a) the maintenance or sterility of the contents under
`dry, clean and adequately ventilated storage
`conditions;
`
`b) the minimum risk of contamination of the con(cid:173)
`tents during removal from the container;
`
`c) adequate protection of the contents during
`normal handling. transit and storage;
`
`2) 1 standard atmosphere (aim) = 101 325 Pa
`1 technical atmosphere (all = 98 066 ,5 Pa
`
`15 Labelling
`
`15.1 Primary container
`
`The primary container shall be marked with at least
`the following inrormation:
`
`a) a description of the contents, including the des-
`ignated metric size in accordance with ciause 7;
`
`b)
`
`the word "STERILE";
`
`c)
`
`the lot number, prefixed by the word " LOT";
`
`d) the name or trade-mark or trade-name or logo
`or the manufacturer or supplier.
`
`15.2 Secondary container
`
`The secondary container shall be marked with at
`least the following informat ion:
`
`a) a description of the contents, including the des(cid:173)
`ignated metric size in accordance with clause 7,
`the number, the type or angle or bevel (see
`clause 12) and, if appropriate, the words ''thtn(cid:173)
`walled" or "extra-thin-walled " or equivalent or
`an abbreviation;
`
`b) the word "STERILE" ;
`
`c)
`
`the words "FOR SINGLE USE" or equivalent (ex(cid:173)
`cepting the term "disposable '');
`
`NOTE 3 The symbol given In annex C may ad-
`ditionally be given.
`
`d) a warning to check th e integrity of each primary
`container before use;
`
`e) the lot number. prefixed by the word "LOT" :
`
`5
`
`Regeneron Exhibit 1042.009
`
`

`

`ISO 7864:1993(E)
`
`f)
`
`the date (year and month expressed as specified
`subdause 5.2.1.1 of
`ISO 8601: 1988) of
`in
`sterilization ;
`
`b) the lot number, prefixed by the word "LOT":
`
`c)
`
`the word "STERILE":
`
`NOTE 4 The date or sterilization may be incorpor(cid:173)
`ated in the first several digits of the lot number.
`
`d) the date of sterilization as specified in 15.2 f) :
`
`g) the name and address of the manufacturer or
`supplier,
`
`h) information for handling, storage and transpor·
`talion .
`
`15.3 Storage container
`
`If secondary containers are packaged in a storage
`- container, the storage container sha ll be marked
`with at least t.he following information:
`
`a) a description of the contents as specified in
`15.2 a);
`
`e) the name and address of the manufacturer or
`supplier;
`
`f)
`
`information for handling, storage and transpor(cid:173)
`tation of the contents.
`
`15.4 Transport wrapping
`
`If a storage container is not used but the secondary
`containe rs are wrapped for transportation, the in·
`formation required by 15.3 shall either be marked
`on the wrapping or shall be visible through the
`wrapping .
`
`6
`
`Regeneron Exhibit 1042.010
`
`

`

`ISO 7864:1993(E)
`
`Annex A
`(normative)
`
`Method for preparation of extracts
`
`A.1 Principle
`
`A.3 Procedure
`
`The needle, including the inside of the needle tube ,
`is immersed in water In order to extract soluble
`components.
`
`A.2 Apparatus and reagents
`
`A.2.1 Freshly distilled or deionized water, of grade
`3 in accordance with ISO 3696.
`
`Immerse 25 needles in 250 ml or water
`A.3.1
`from
`containet made
`(A.2.1)
`in
`a
`suitable
`boros ilicate glass (A.2.2) . Ensure that all surfaces of
`the needles, including the inside of the needle tube ,
`are in contact with the water. Maintain the water at
`a temperature of (37 +~) •c for (60 ± 2) min . Re(cid:173)
`move the needles and ensure that all water from the
`inside and outside su rfaces of the needles is re(cid:173)
`turned to !he container.
`
`A.2.2 Selection of laboratory borosilicate glass(cid:173)
`ware.
`
`A.3.2 Prepare the conlrol fluid by following the
`procedure given in A.3.1 but omitling the needles.
`
`7
`
`Regeneron Exhibit 1042.011
`
`

`

`ISO 7864:1993(E)
`
`Anne>< B
`(informative)
`
`Example of method for determinatioh of fragment production from rubber closures
`
`B.1 Principle
`
`=
`
`Penetration or ISO rubber closures for injection vials
`by the needle, and collection and counting or t he
`fragments generated by t he penetrations. Because
`the number of fragments generated is affected by
`variables of needle desig n, each type of needle to
`be tested is compared with the performance of a
`reference needle.
`
`B.2.8 Distilled or deionized water, of grade 3 in ac(cid:173)
`cordance with ISO 3696 .
`
`8.3 Procedure
`
`8 .3.1 Place n m I of water (B.2 .8)
`into th e vial
`(B .2.1), where n is 50 % of the nominal capacity of
`the vial .
`
`B.2 Apparatus and reagents
`
`Injection vials , in accordance Wi1h ISO 8362·1
`B.2.1
`or ISO 8362-4.
`
`B.3.2 Place a closure (B.2.2) on the via l and seal
`with an aluminium cap (B.2.3) us ing the hand(cid:173)
`operaled ca pping device (B.2.4). Inspect the vial by
`normal or corrected-to-normal vision for the pres(cid:173)
`ence or fragments , and discard any vial in which
`rubber fragments are vis ible.
`
`closures,
`B.2.2 Rubber
`ISO 8362-2, as .follows:
`
`in
`
`accordance with
`
`B.3.3 Fill the syringe (B.2.6) with water (B .2.8) . At(cid:173)
`tach a reference needle to the syringe and remove
`any water from the outside of the needle tube.
`
`Type A, nom inal size 20
`
`Hardness; 40 IRHD
`
`Base elastome r: Halogenated butyl.
`
`caps 1
`8 .2.3 Alumlnlmn
`accordance with
`in
`1SO 8362-3, having a central hole , to fit the injection
`vials (8.2.1 ).
`
`B.2.4 Hand-operated capping device, suitable for
`vials (B.2.1) and caps (B.2.3).
`
`B.2.5 Membrane filter, of pore size 5 µm, and filter
`holder.
`
`8.3.4 Place the via l vertically and , holding lhe
`syringe vertically , pierce the closure.
`
`8.3.5
`Inject 2 rnl of water into the vial , and with(cid:173)
`draw the needle from the vial Remove the needle
`from ttie syringe and replace It with a fresh refer(cid:173)
`ence need le. Remove any water from the outside of
`the needle tu be.
`
`B.3.6 Repeat B.3.4 and B.3.5 for a total of five
`penetrations, choosing a different area of the clo(cid:173)
`sure each time.
`
`8.2.6 Hypodermic ,yringe, of 10 ml nominal ca·
`pacity In accordance with ISO 7886-1 .
`
`B.3.7 Repeal 8.3.3 to B.3.6 on a total of five vials,
`i.e. 25 reference needles and 25 penetrations In all.
`
`B.2.7 Rererence needles1 of the same nominal out(cid:173)
`side diameter of the type of needle to be tested, and
`having a nom inal value of ix (see figure 2) of 13° .
`
`8 .3.8 Repeat 8 .3.3 to 8 .3 7, using 25 needles of the
`type to be tested.
`
`8
`
`Regeneron Exhibit 1042.012
`
`

`

`ISO 7864:1993(E)
`
`8.3.9 Remove the closure from the five vials from
`B.3.7 and fi lter the contents through the membrane
`fllter (B.2.5) . Ensure that no fragments remain in the
`vial.
`
`a) th.e numbe r of fragments generated by the refer(cid:173)
`ence needles, as counted in 8.3 .11;
`
`b) the number of fragments generated by the type
`of needle tested, as counted in 8.3.11 ;
`
`B.3.10 Repeat B.3.9 on the five vials from B.3.8,
`using a second membrane filter.
`
`c)
`
`the date of testing.
`
`B.3.11 Count and record the number or fragments
`on each filter visible with norma l or corrected vision
`without magnification, the distance between the eye
`and the filter being (250 ± 5) mrn.
`
`8.5 Validity
`
`B.4 Test report
`
`The test report shall contain a1
`information:
`
`least the following
`
`Consider the resu lts obtained from the test needles
`to be invalid if the results from the reference nee(cid:173)
`dles lack consistency with previous results. Deter(cid:173)
`mine the reasons for the Inconsistency.
`
`9
`
`Regeneron Exhibit 1042.013
`
`

`

`ISO 7864:1993(E)
`
`Annex C
`(informative)
`
`Symbol for "do not re-use"
`
`C.1 General
`
`=
`
`The ISO symbol to denote equi pment Intended for
`single use
`is
`ISO symbol registration number
`ISO 7000/1051 , given in ISO 7000:1989.
`
`Further information on design, dimensions and
`NOTE 5
`application of ISO symbols is given in ISO 3461 .
`
`C.2 Original design
`
`Symbol ISO 7000/1051 is shown in figure C.1 .
`
`The thickness or the lines is 2 mm . Dirnensioh a is
`the nominal size of the original design of all ISO
`symbols and is normally made equal to 50 mm. In
`many cases, including ISO 7000/1051 , the actua l di(cid:173)
`:me-nsi'on differs slightly, and the outside diameter of
`
`the circle (dimension It) of the original design Is
`1,1 6 a, i.e. 58 mm.
`
`No colour is specified in ISO 7000 or in this Inter(cid:173)
`national Standard for symbol 1051 .
`
`C.3 Reduction and enlargement of
`original design
`
`For the app lication of the symbol it may be neces(cid:173)
`sary to reduce or enlarge the srze of the orig inal to
`a suitable size at wh ich it w ill actually appear. The
`nominal dimension a should be used as a gauge .
`Practice has shown that a may be reduced to 3 mm
`without the sym bol losing its legibility . However, the
`legibility of the symbol when reduced In size shou ld
`be checked.
`
`L
`
`_J
`
`NOTE - The visual centring lines do nol form part of the syrnbol.
`
`Figure C.1 -
`
`ISO symbol for "do not re-use'', number ISO 7000/1051
`
`10
`
`Regeneron Exhibit 1042.014
`
`

`

`ISO 7864:1993(E)
`
`Annex D
`(informative)
`
`Bibliography
`
`[1] ISO 595-1:1986, Reusable all-glass or metal(cid:173)
`and-gfass syringes for medical use - Part f:
`Dimensions.
`
`for
`containers
`Injection
`[6 ] ISO 8362-1:1 989,
`in]ectables and accessories - Part 1: Injection
`vials made of glass tubing.
`
`[2] ISO 595-2: 1987, Reusable all-glass or meta/(cid:173)
`and-glass syringes for medical use - Part 2:
`Design, performance requirements and tests.
`
`[7] ISO 8362-2:1988 ,
`for
`containers
`Injection
`injec!ab/es and accessories - Part 2: Closures
`for injection vials ,
`
`the
`for
`[3] ISO 3461-1:1988, General principles
`Part 1:
`-
`creation of graphical symbols
`Graphical symbols for use on equipment.
`
`[8] ISO 8362-3:1989 ,
`Injection
`injeclables and accessories
`minium caps for injection vials.
`
`for
`containers
`Part 3: Alu-
`
`[4] ISO 3461 -2:1987, General principles for the
`creation of graphical symbols
`-
`Part 2:
`Graphical symbols for use in technical product
`documentation.
`
`[5] ISO 7000:1989, Graphical symbols for use on
`equipment -
`Index and synopsis.
`
`[9] ISO 8362-4:1 989,
`for
`containers
`Injection
`injectab/es and accessories - Part 4: Injection
`vials made of moulded glass.
`
`[ 1 OJ ISO 8537:1991, Sterile single-use syringes, with
`or without needle, for insulin .
`
`[ 11] ISO 10993-1.: 1992, Biological evaluation of
`medical devices - Part 1: Guidance on se(cid:173)
`lection of tests.
`
`11
`
`Regeneron Exhibit 1042.015
`
`

`

`ISO 7864:1993(E)
`
`UDC 615.473.2.014.45
`De$Crlplor$: medica l equipment, hypodermic needles, specifications, di mensions,
`labelling.
`
`Price based on 11 pages
`
`tes ts, designation, colour marking, packaging,
`
`Regeneron Exhibit 1042.016
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket