throbber

`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`
`
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`
`
`REGENERON PHARMACEUTICALS, INC.
`Petitioner,
`v.
`
`NOVARTIS PHARMA AG,
`NOVARTIS TECHNOLOGY LLC,
`NOVARTIS PHARMACEUTICALS CORPORATION,
`Patent Owner
`
`
`Case No. IPR2021-00816
`U.S. Patent No. 9,220,631
`
`
`
`DECLARATION OF JAMES AGALLOCO
`
`
`
`
`Regeneron Exhibit 1100.001
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`TABLE OF CONTENTS
`
`Page
`INTRODUCTION ........................................................................................... 1
`I.
`QUALIFICATIONS AND COMPENSATION .............................................. 1
`II.
`III. LEGAL STANDARDS ................................................................................... 5
`IV. A PERSON OF ORDINARY SKILL IN THE ART ...................................... 7
`V. DEFINITION OF “TERMINAL STERILIZATION” IN THE '631
`PATENT .......................................................................................................... 8
`VI. BACKGROUND ............................................................................................. 9
`A.
`Sterilization of Pre-filled Syringes ........................................................ 9
`B.
`Sterility Assurance and Validation ...................................................... 12
`VII. RELEVANT PRIOR ART TO THE '631 PATENT ..................................... 18
`A.
`“Sigg” – WO 2011/006877 ................................................................. 18
`B.
`“Lam” – WO 2008/077155 ................................................................. 19
`VIII. A Person with Specialized Skills in Microbiology Would Be
`Motivated to Use Sigg’s VHP Method to Achieve an SAL of 10-6 and
`Would Have a Reasonable Expectation of Success in so Doing ................... 21
`A.
`Sigg Provides a Motivation to Use its VHP Method .......................... 21
`B.
`A Person with Specialized Skills in Microbiology Would
`Understand that Sigg’s VHP Method Could Be Used to
`Achieve an SAL of 10-6 ....................................................................... 23
`C. A Person with Specialized Skills in Microbiology Would Have
`Reasonably Expected to Succeed in Using Sigg’s VHP Method
`to Achieve an SAL of 10-6 ................................................................... 25
`
`
`
` .................................. 27
`IX. A Person with Specialized Skills in Microbiology Would Have Been
`Motivated to Use Lam’s EtO Method to Achieve a SAL of 10-6, and
`Would Have a Reasonable Expectation of Success ...................................... 29
`X. DECLARATION ........................................................................................... 35
`
`D.
`
`
`
`i
`
`Regeneron Exhibit 1100.002
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`I.
`
`INTRODUCTION
`I have been retained by Petitioner Regeneron Pharmaceuticals, Inc.
`1.
`
`("Petitioner" or "Regeneron"), as an independent expert witness in the above-
`
`captioned inter partes review ("IPR"), in which Regeneron has requested that the
`
`U.S. Patent and Trademark Office cancel as unpatentable all claims of U.S. Patent
`
`No. 9,220,631 ("the '631 patent") (Ex. 1001).
`
`2.
`
`This declaration sets forth my analyses and opinions based on my
`
`knowledge, experience, and the materials I have considered.
`
`II. QUALIFICATIONS AND COMPENSATION
`I have a M.B.A. in Pharmaceutical Studies from Fairleigh Dickinson
`3.
`
`University, a M.S.Ch.E. from the Polytechnic Institute of New York, and a
`
`B.E.Ch.E. from Pratt Institute.
`
`4.
`
`I have over fifty years of management experience in pharmaceutical
`
`manufacturing, pharmaceutical process, pharmaceutical process engineering,
`
`technical services and research and development. Since 1991, I have been the
`
`President of Agalloco & Associates, which provides to the pharmaceutical,
`
`biotechnology, and medical device industry a wide range of technical services such
`
`as process and product validation, sterilization, aseptic processing, processing,
`
`isolation technology, sterility assurance, compliance and facility design.
`
`1
`
`Regeneron Exhibit 1100.003
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`5.
`
`Prior to 1991, I was Director of Validation and Technology at Bristol-
`
`Myers Squibb; Director of Worldwide Validation and Automated Technology and
`
`Pharmaceutical Engineering; Department Manager at Bristol-Myers Squibb, and
`
`held a number of other positions, including Engineering Project Manager and
`
`Senior Production Supervisor at Pfizer Pharmaceuticals.
`
`6.
`
`As Director of Validation and Technology at Bristol-Myers Squibb, I
`
`directed validation, automation and technical documentation activities and served
`
`as an important technical resource for worldwide pharmaceutical manufacturing. I
`
`was an active participant on product introduction and facility upgrade task forces.
`
`As Director of Worldwide Validation and Automated Technology at Bristol-Myers
`
`Squibb, I was responsible for facilities in 27 countries around the world and served
`
`as a major technical resource for facility design, facility start-up, sterilization,
`
`aseptic processing, validation and automation. I participated actively on major
`
`product, process, facility and equipment projects, directed the validation and
`
`automation phases of a $25 million expansion of existing parenteral facility in New
`
`Brunswick, and provided major support to sterile bulk manufacturing. As head of
`
`the Validation department at Squibb New Brunswick, I was responsible for
`
`validation of all processes at the site including various sterilization processes. I
`
`also led the validation effort for a $60 million parenteral facility.
`
`2
`
`Regeneron Exhibit 1100.004
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`7.
`
`In previous positions, I provided validation expertise for sterile
`
`facilities, acted as a spokesperson for validation to the Food and Drug
`
`Administration (“FDA”) and other regulatory agencies, managed production
`
`operations for sterile and oral liquid and powder products, and had major areas of
`
`responsibility including cost control, cost reduction, Current Good Manufacturing
`
`Practice (“CGMP”) compliance, scheduling, equipment selection, and process
`
`trouble shooting.
`
`8.
`
`In addition to my work experience, I have many years of experience
`
`participating in professional organizations, and pharmacopoeias relating to
`
`sterilization and sterility assurance for pharmaceuticals and medical devices,
`
`including microbiology as it relates to sterilization. For example, I have been an
`
`active member of the United States Pharmacopoeia (“USP”) Microbiology Expert
`
`Committee since 2005, and the lead author on the comprehensive revision of USP
`
`<1211> Sterilization & Sterility Assurance of Compendial Items. The new USP
`
`content (Chapters <1211> Sterility Assurance and <1229> Sterilization) includes
`
`substantially expanded content addressing the full range of sterilization processes
`
`and means for the aseptic processing for drugs and medical devices. In addition, I
`
`have led and participated in the development of numerous sterilization and aseptic
`
`processing industry guidance documents as a member of the Parenteral Drug
`
`Association. I have been a member of the Parenteral Drug Association since 1980
`
`3
`
`Regeneron Exhibit 1100.005
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`and served as its President in 1988 and 1989. I am also the co-founder and a
`
`current member of the Validation Discussion Group. I have also served on several
`
`scientific and editorial advisory boards. I have also co-authored more than 60 book
`
`chapters and over 160 technical papers.
`
`9.
`
`In addition to the above, I have given numerous presentations at
`
`industry meetings and in-house at pharmaceutical companies, including many
`
`presentations over the years that relate to aseptic processing, sterilization, and
`
`process validation.
`
`10. Through my professional experience, I have gained extensive
`
`expertise in both aseptic processing and terminal sterilization. I have experience in
`
`all types of medical device and pharmaceutical sterilization, including chemical
`
`sterilization, such as by hydrogen peroxide and ethylene oxide. I have also gained
`
`knowledge of sterilization of container closure systems, including aseptic
`
`processing and terminal sterilization of pre-filled syringes.
`
`11.
`
`I am being compensated at my standard rate of $350/hour. My
`
`compensation is in no way contingent upon my opinions or the outcome of the
`
`proceeding.
`
`12.
`
`I may testify on any or all of the opinions expressed in this expert
`
`report. In addition, I will, as needed, explain principles and terminology referred to
`
`in this report as well as materials referenced herein. I provide this declaration to
`
`4
`
`Regeneron Exhibit 1100.006
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`provide specific background and explanation regarding sterilization technology, as
`
`well as explaining how certain claim limitations of the ’631 patent relating to
`
`sterilization are disclosed and/or rendered obvious by the prior art.
`
`13. My curriculum vitae is attached as Attachment A, and provides
`
`further information about my experience, expertise, publications, and
`
`presentations.
`
`III. LEGAL STANDARDS
`I understand that the invalidity analysis must be conducted from the
`14.
`
`perspective of the person of ordinary skill in the art, and must consider whether the
`
`skilled artisan would consider any differences between the prior art and what is
`
`claimed to have been obvious. To make this assessment, I have been informed that
`
`the concept of patent obviousness involves four factual inquiries: (1) the scope and
`
`content of the prior art; (2) the differences between the claimed invention and the
`
`prior art; (3) the level of ordinary skill in the art; and (4) secondary considerations
`
`of non-obviousness. I have been instructed that one must not engage in hindsight.
`
`Rather, the better approach is to consider what the person of ordinary skill in the
`
`art would have reason to pursue further, and steps that were routinely done, such as
`
`in response to known problems, steps or obstacles.
`
`15.
`
`It is my understanding that some teaching, suggestion, or motivation
`
`in the prior art that would have led one of ordinary skill to modify the prior art
`
`5
`
`Regeneron Exhibit 1100.007
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`reference or to combine prior art reference teachings to arrive at the claimed
`
`invention may support the obviousness of an invention. Other rationales that may
`
`support the obviousness of the invention include combining prior art elements
`
`according to known methods to yield predictable results, simple substitution of one
`
`known element for another to obtain predictable results, and applying a known
`
`technique to a known device ready for improvement to yield predictable results.
`
`16.
`
`It is my understanding that the motivation to combine prior art
`
`references may be implicit and may be found in the knowledge of one of ordinary
`
`skill in the art, or in the nature of the problem to be solved. Specifically, it is my
`
`understanding that an implicit motivation to combine exists not only when a
`
`suggestion may be gleaned from the prior art as a whole, but when the
`
`“improvement” is technology-independent and the combination of references
`
`results in a product or process that is more desirable, for example, because it is
`
`stronger, cheaper, cleaner, faster, lighter, smaller, more durable or more efficient.
`
`It is my further understanding that the motivation to combine references may be
`
`found in the nature of the problem to be solved where prior art references are
`
`directed to precisely the same problem.
`
`17.
`
`I also understand that prior art may be relied on for its express
`
`disclosure and teachings. I also understand that the prior art may be relied upon
`
`6
`
`Regeneron Exhibit 1100.008
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`for a teaching of features that are necessarily present in the prior art reference even
`
`if that specific feature is not expressly or explicitly disclosed.
`
`IV. A PERSON OF ORDINARY SKILL IN THE ART
`I understand that Novartis has asserted that the following definition of
`18.
`
`a person of ordinary skill in the art (POSITA) should apply in this proceeding:
`
`A POSA would have an advanced degree (i.e., an M.S., a Ph.D., or
`equivalent) in mechanical engineering, biomedical engineering, materials
`science, chemistry, chemical engineering, or a related field, and at least 2–3
`years of professional experience, including in the design of a PFS and/or the
`development of ophthalmologic drug products or drug delivery devices.
`Such a person would have been a member of a product development team
`and would have drawn upon not only his or her own skills, but also the
`specialized skills of team members in complementary fields including
`ophthalmology, microbiology and toxicology.
`
`Ex. 2203.0011-.0012 at ¶ 26. Dr. Miller further asserts that a person “expected to
`
`contribute to a larger team responsible for developing sterile pharmaceuticals,
`
`medical devices, or related dosage forms” would be “[a] microbiologist or one
`
`who is versed in the scientific principles of sterilization techniques and validation
`
`strategies.” Ex. 2203.0014 at ¶ 32 (emphasis added). In my opinion, microbiology
`
`is relevant to the '631 patent to the extent it references sterility, terminal
`
`sterilization, and sterility assurance. See also Ex. 1210.029-.030 at 29:8-30:11.
`
`Furthermore, a person with experience and well-versed in sterilization techniques
`
`7
`
`Regeneron Exhibit 1100.009
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`would be a suitable product development team member with specialized skills in
`
`microbiology insofar as that person would be knowledgeable in microbiology as it
`
`relates to sterilization processes. See also Ex. 1210.037 at 37:3-12.
`
`19. As an expert in terminal sterilization, and a member of the USP
`
`Microbiology Expert Committee, I am qualified to opine from the perspective of
`
`team member with specialized skills in microbiology as it relates to terminal
`
`sterilization. I have an in-depth knowledge of terminal sterilization concepts that
`
`implicate the subject of microbiology, such as bioburden, bioindicators, log-
`
`reduction and the calculation of sterility assurance level. Furthermore, I have been
`
`a member of numerous product development teams, specifically as one with
`
`expertise in microbiology as it relates to the sterilization of medical devices,
`
`pharmaceuticals and ophthalmic preparations, as well as one who is versed in the
`
`scientific principles of sterilization techniques and validation strategies. Thus, I am
`
`able to offer my opinions in this declaration from the perspective of a member of a
`
`product development team who has specialized skills in microbiology, and have
`
`done so in the sections below.
`
`V. DEFINITION OF “TERMINAL STERILIZATION” IN THE '631
`PATENT
`20. Although “terminal sterilization” traditionally refers to a process
`
`whereby the drug and its container are sterilized within their packaging, I
`
`understand that, as used in the '631 patent, the parties have agreed that “terminally
`
`8
`
`Regeneron Exhibit 1100.010
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`sterilized” refers to a process whereby the outside of a pre-filled syringe is
`
`sterilized, while contact between the sterilizing agent and the drug product within
`
`the syringe is minimized.
`
`VI. BACKGROUND
`Sterilization of Pre-filled Syringes
`A.
`21. Regulatory agencies such as the FDA and EMA require all ophthalmic
`
`products, such as pre-filled syringes for intravitreal injection, to be sterile. See 21
`
`C.F.R. § 200.50(c) (“[D]ispensers intended for ophthalmic use should be sterile….
`
`These articles, which are regulated as drugs if packaged with the drugs with which
`
`they are to be used, should be packaged so as to maintain sterility until the package
`
`is opened.”). Maintaining the sterility of pre-filled syringes for intravitreal
`
`injection was known to be important because of the risk of infection from a non-
`
`sterile syringe used on an exposed human eye. See, e.g., Ex. 1029.002 at 1:28-29
`
`(“[T]here is an increased risk of endophthalmitis after intraocular injection1 if the
`
`surface of the syringe used for injection is not sterilized.”).
`
`
`
`1 Intraocular injection refers to all injections into the chambers of the eye.
`
`Intravitreal injection is a subset of intraocular injection, and refers to injection into
`
`the vitreous cavity specifically. Ex. 1015.035.
`
`9
`
`Regeneron Exhibit 1100.011
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`22. The FDA, in its guidance on aseptic filling, indicates that “terminal
`
`sterilization” is required when possible. See Ex. 1036.007. “Terminal sterilization”
`
`(outside of the context of the '631 patent) traditionally refers to a process in which
`
`the drug and container closure system (which are not already sterile) are sterilized
`
`together in a single process. See id. For drugs that are not heat sensitive, this one-
`
`step terminal sterilization can be accomplished by heating (for example, in an
`
`autoclave) with steam to sterilize the drug in its primary container-closure system.
`
`See generally Ex. 1029.002 at 1:14-22.
`
`23. The '631 patent uses the term “terminal sterilization” in a different
`
`context, specifically stating that “[t]he package is exposed to the sterilising gas
`
`until the outside of the syringe is sterile,” and also that “it is a requirement that
`
`significant amounts of the sterilising gas should not enter the variable volume
`
`chamber of the syringe.” Ex. 1001.008 at 9:49, 9:55-56; 10:2-4 (emphasis added).
`
`Thus, the '631 patent describes “terminal sterilization” of only the outer surface of
`
`the syringe after the syringe has been filled with the drug product, generally under
`
`aseptic conditions.
`
`24. There are many ways to sterilize medical products, including via
`
`“steam sterilization, radiation sterilization, gas sterilization (e.g., with ethylene
`
`oxide), and chemical sterilization.” Ex. 1029.002 at 1:14-16. It was already known
`
`prior to 2012 that protein drug formulations can be degraded by high temperature
`
`10
`
`Regeneron Exhibit 1100.012
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`sterilization processes, which are therefore disfavored for use with protein
`
`therapeutics such as VEGF-antagonist solutions in pre-filled syringes. See Ex.
`
`1018.003 at [0010]-[0011], [0021]-[0022]; Ex. 1007.008-.009 at 7:29-8:1; Ex.
`
`1029.002 at 1:18-25. With respect to sterilization processes that do not use high
`
`temperatures, several such processes were known, including sterilization using
`
`ethylene oxide (“EtO”) gas, and using vaporized hydrogen peroxide (H2O2)
`
`(“VHP”). See e.g., Ex. 1046.001 (describing ethylene oxide sterilization); id. at
`
`002 (describing vaporized hydrogen peroxide (VHP) sterilization). Ethylene oxide
`
`(EtO) is a gas-based sterilization method, and hydrogen peroxide (H2O2)
`
`sterilization is a vapor-based (2-phase system of liquid suspended in a gas)
`
`sterilization method. Both EtO and VHP were used to sterilize medical devices for
`
`many years prior to 2012. See id.
`
`25. EtO sterilization has been used since “the 1950s to sterilize heat and
`
`moisture-sensitive medical devices.” Id. at .001. A POSITA would have been well
`
`aware that EtO was “[t]he most prevalent gas utilized for sterilization” such that
`
`“sterilization using other agents is based on methods used for ETO.” Ex.
`
`1016.260-.261 (describing EtO sterilization).
`
`26. Similarly, the use of hydrogen peroxide (H2O2) for decontamination
`
`has been around since at least the early 1990s. Ex. 1046.002-.003. There were two
`
`well-known methods of H2O2-based sterilization that developed prior to 2012. Id.
`
`11
`
`Regeneron Exhibit 1100.013
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`at .002-.003. The first is sterilization via the creation of a hydrogen peroxide
`
`plasma. Id at .002. The second method, vaporized hydrogen peroxide (VHP), is
`
`simpler because it does not require formation of a plasma and has a large
`
`processing capacity. Id at .002-.003. Prior to 2012, VHP systems were available
`
`for “processing packaged, heat and moisture-sensitive instruments for terminal
`
`sterilization and storage.” Id. at .002.
`
`Sterility Assurance and Validation
`B.
`27. The sterility of a medical product can be measured in terms of the
`
`sterility assurance level, or SAL, which was a well-known concept in the art long
`
`before 2012, and is a mathematical measure of the probability that a sterilized
`
`article may not in fact be sterile. Ex. 1007.008 at 7:8-9 (“‘Sterility’ as used herein
`
`is meant to refer to complete absence of microbial life as defined by a probability
`
`of nonsterility or a sterility assurance level (SAL).”). An SAL of 10-6, the
`
`preferred sterility level for health care products, refers to the probability that only 1
`
`in 1,000,000 processed units would be non-sterile. Id. at 7:10-13 (“For example,
`
`required SALs for health care products are defined to be at least 10-6, i.e. a chance
`
`of less than 1:1 million of a non-sterile product”); Ex. 1049.003.
`
`28. Determining the SAL of a sterilization process would be routine and
`
`well within the ordinary skill of a POSITA prior to 2012. Because it is not feasible
`
`to test millions of products to directly measure the rate of non-sterility, sterilization
`
`12
`
`Regeneron Exhibit 1100.014
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`process validation is used. Ex. 1049.002-.003. This type of process validation
`
`utilizes the fact that microbial kill rates from most sterilization techniques are
`
`exponential in nature. Id. at .003. When a plot of log population versus time is
`
`prepared a line can be drawn indicating the death curve. This allows the calculation
`
`of SAL level “based on the extent of exposure to the sterilization modality and the
`
`corresponding microbial log reduction,” which ultimately derive from the
`
`measured time or dose. The D-value is the time “required to achieve inactivation
`
`of 90% of a population of the test microorganism” (that is, each log reduction is a
`
`90% reduction —e.g., from 100 to 10 or from 10 to 1). Id. Performing such a
`
`calculation would be routine to a POSITA, and indeed the adoption of SAL across
`
`multiple industries is due in part to its easily adapted standardized approach.
`
`29.
`
`In the context of sterilizing a medical product, such as a prefilled
`
`syringe, the actual number of microorganisms present on the medical product prior
`
`to sterilization is referred to as the “bioburden.” The test organisms acting as a
`
`stand-in for the bioburden, which can be used to estimate the microbial log
`
`reduction of a sterilization process, are called “bioindicators.” The commonly used
`
`bioindicator has about 106 bacterial spores of a species (usually from the genus
`
`Bacillus or Geobacillus) that is particularly resistant to the type of sterilization
`
`used, and thus can represent the worst-case scenario for the bioburden. Ex. 1016.
`
`264, .267. Bioindicators are placed alongside or within the products to be
`
`13
`
`Regeneron Exhibit 1100.015
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`terminally sterilized, and after undergoing a sterilization process, the number of
`
`remaining spores on the bioindicator are counted. This final number is then
`
`expressed as a log reduction in comparison to the starting population.
`
`30. For example, if the bioindicator has 106 (that is 1,000,000) spores to
`
`start, but 103 (that is, 1,000) spores remain after the sterilization cycle, the number
`
`of spores has been reduced by 103 (that is, 1,000-fold), which is 3 logs. If 100
`
`spores (that is, one spore), remain, then there has been reduction by106 (that is,
`
`1,000,000-fold), or a 6-log reduction. If all 106 spores are killed, such that no
`
`spores remain, the estimated log reduction is greater than 6 logs, since fractions of
`
`a living spore cannot be recovered. To test for a reduction greater than 6 logs, the
`
`starting population of the bioindicator would need to be greater than 106 spores.
`
`Thus, to reach higher sterility probabilities, a POSITA would look to use certain
`
`validation methods, such as the half-cycle method, first developed for use with
`
`EtO, and commonly used for sterilization with gases or vapors. Id. at .264-.265.
`
`31. As explained above, the standard goal for sterilization of health care
`
`products, such as those intended to come into contact with breached skin or
`
`compromised tissue, is a sterility assurance level, or SAL, of 10-6. Ex. 2187.0015.
`
`To validate a sterilization process using the half-cycle method, a bioindicator with
`
`106 microorganisms would be placed in the sterilization chamber, and the time for
`
`killing all of the microorganisms would be determined. For example, if a one hour
`
`14
`
`Regeneron Exhibit 1100.016
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`sterilization cycle killed all 106 spores of the bioindicator, that one hour cycle
`
`would show a 6-log reduction. In order to validate a 12-log reduction, which
`
`would result in a SAL of 10-6, the time demonstrated to kill 106 spores would then
`
`be doubled. Thus, the initial 6-log reduction is considered a “half-cycle.”2 This is
`
`shown on the graph below, taken from Dr. Miller’s declaration.
`
`
`
`2 Dr. Miller equates the half-cycle method with the overkill method. The overkill
`
`method, which is used for sterilization methods other than those utilizing gas or
`
`vapor, refers generally to a sterilization methodology in which the destruction of a
`
`high concentration of a resistant biological indicator assures the reliable
`
`elimination of the bioburden, such that its microbial population and resistance can
`
`be ignored. The half-cycle method is one example of an overkill method.
`
`15
`
`Regeneron Exhibit 1100.017
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`Ex. 2203.0021
`
`
`
`32. Although Dr. Miller points out in his declaration that a lesser SAL, for
`
`example 10-3 to 10-5, may be acceptable to regulatory authorities if a SAL of 10-6
`
`were not possible, the goal for a POSITA and person with specialized skills in
`
`microbiology for an ophthalmic pre-filled syringe would nevertheless be a SAL of
`
`10-6. The ANSI source that Dr. Miller cites for the proposition that a SAL of 10-6
`
`was not absolutely required, explains that, prior to accepting a less stringent SAL,
`
`other validation methods beyond the traditional 6-log half-cycle approach would
`
`be attempted. See Ex. 2187.0015 (“For those products that require a 10-6 SAL and
`
`are incapable of withstanding the sterilization process chosen, alternative
`
`16
`
`Regeneron Exhibit 1100.018
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`sterilization processes and/or validation methods should be investigated before
`
`selecting an alternative SAL (e.g., 10-5, 10-4, or 10-3)”).
`
`33. An individual with specialized skills in microbiology as it relates to
`
`terminal sterilization would understand that the extent of treatment required to
`
`validate an SAL of 10-6 can be reduced if a different validation approach is used.
`
`Id. For example, if the initial bioburden can be controlled, a treatment with a
`
`sterilization process that achieves less than a 12-log reduction of the biological
`
`indicator (for example, a sterilization process with a shorter treatment time) can
`
`validate a SAL of 10-6 for the bioburden. This is referred to as the combined
`
`bioburden-bioindicator approach, and was a well-documented method in the art for
`
`achieving an SAL of 10-6 before the '631 patent. Ex. 2188.016-.018, .030-.031.
`
`
`
`
`
`
`
`Likewise, it is possible to validate a particular SAL by controlling the initial (pre-
`
`sterilization cycle) bioburden and determine the extent to which the actual
`
`bioburden is reduced, which can shorten the treatment time further. This is
`
`referred to as a bioburden-based validation method. Ex. 2187.0015 (“[I]n general, a
`
`bioburden-based validation method will give a shorter extent of treatment to
`
`achieve a particular SAL than either a biological indicator-bioburden or an
`
`‘overkill’ method.”).
`
`17
`
`Regeneron Exhibit 1100.019
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`VII. RELEVANT PRIOR ART TO THE '631 PATENT
`“Sigg” – WO 2011/006877
`A.
`34. Sigg (Ex. 1007) is a patent publication assigned to Novartis. Sigg
`
`teaches terminal sterilization that involves the “treatment of prefilled containers in
`
`secondary packaging by an application of vaporized-hydrogen peroxide, in which
`
`vapors are controllable by certain post-treatment measures, and exposure to
`
`tunable-beta radiation, in which the depth of penetration of beta rays into
`
`secondary packaging are controllable.” Ex. 1007.009 at 8:8-12. The VHP method
`
`includes “applying post-treatment measures, within a decontamination chamber”
`
`that ensure full removal of the VHP from the chamber and, as a result, protection
`
`of the sensitive biologic product and future handlers from the VHP. Id. at .011,
`
`10:5-6.
`
`35. Sigg explains that the disclosed VHP terminal sterilization methods
`
`are especially suitable for use on pre-filled syringes. Id. at .003, 2:13-15
`
`(“Prefilled syringes, although filled under aseptic conditions, are not packed into
`
`their secondary packaging in an aseptic environment and are therefore likely to be
`
`microbiologically contaminated at their outside.”); id. at .004, 3:8-11 (“Described
`
`herein is a terminal sterilization and surface decontamination treatment of prefilled
`
`containers, specifically for sterilization of prefilled containers containing sensitive
`
`solutions, such as a drug product or biological therapeutic, within secondary
`
`18
`
`Regeneron Exhibit 1100.020
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`packaging.”). Sigg notes that “[t]erminal sterilization of prefilled containers in
`
`secondary packaging is one way to provide the device to an end user with a low
`
`bio-burden and low risk of contaminants.” Id. at .003, 2:15-17.
`
`36. Relevant to the claims of the '631 patent, Sigg teaches the terminal
`
`sterilization (surface sterilization) of low volume (0.5 mL) pre-filled syringes for
`
`intravitreal injection containing an ophthalmic solution comprising a VEGF-
`
`antagonist. Sigg describes that the VHP sterilization process does not impact the
`
`sensitive biologic product contained in the pre-filled glass syringe. Id. at .010,
`
`9:16-17 (“the contents of the container are sterile and unaffected by surface
`
`decontamination methods as described herein.”). Example 1 of Sigg describes that
`
`the protein concentration of an anti-VEGF antibody in pre-filled syringes was
`
`relatively unchanged after sterilization with VHP, which indicates that little or no
`
`VHP penetrated the syringe. Id. at .021-.022, 20:18-21:3 (describing syringes
`
`comprising a “formulation as described in U.S. Patent No. 7,060,269”); Ex. 1023
`
`(U.S. Patent No. 7,060,269 describing Anti-VEGF antibodies).
`
`“Lam” – WO 2008/077155
`B.
`37. Lam (Ex. 1029) is a patent publication that is assigned to Genentech.
`
`Lam discloses the terminal sterilization of pre-filled glass syringes containing a
`
`VEGF-antagonist intended for intravitreal injection, using EtO sterilization. Ex.
`
`1029.014 at 13:14-15 (“We performed EtO sterilization runs on syringes
`
`19
`
`Regeneron Exhibit 1100.021
`Regeneron v. Novartis
`IPR2021-00816
`
`

`

`
`
`containing a ranibizumab solution….”). Lam teaches that the EtO sterilization
`
`methods taught therein are especially suitable for drug formulations containing
`
`active ingredients that may be damaged by the high temperatures, radiation, or
`
`direct interaction with chemical gases used in some sterilization processes. Id. at
`
`.002, 1:18-23.
`
`38. Lam discloses that the surface of the syringe is EtO sterilized after the
`
`syringe has already been filled with drug formulation via aseptic fill, and that
`
`contact between the sterilizing agent and the drug formulation is avoided. See id.
`
`at .002, 1:22-33 (“Consequently, pharmaceutical compositions are generally
`
`sterilized by an alternative method, e.g. by filtration…. Thus, there remains a need
`
`for efficient and cost-effective methods of surface-sterilizing objects containing
`
`ethylene-oxide-sensitive, temperature-sensitive compounds, such as biological
`
`molecules….”).
`
`39. By design, the “surface-steril

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket