throbber
INTERNATIONAL
`STANDARD
`
`ISO
`10993-7
`
`Second edition
`2008-10-15
`
`Biological evaluation of medical
`devices-
`Part 7:
`Ethylene oxide sterilization residuals
`
`Evaluation biologique des dispositifs medicaux -
`Parlie 7: Residus de sterilisation a l'oxyde d'ethylene
`
`C0P'Y'/'¥1ht International OrQaniz.atlon for Slandardizatlon
`Provided by IHS und&r lloeose w1th 1$0
`No reproduction or networking pennitted without license from IHS
`
`NOi for Resale
`
`Reference number
`ISO 10993-72008(E)
`
`© ISO2008
`
`Novartis Exhibit 2196.001
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`ISO 10993-7:2008(E)
`
`PDF disclaimer
`
`This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but
`shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
`downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
`accepts no liability in this area.
`
`Adobe is a trademark of Adobe Systems Incorporated.
`
`Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
`parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In
`the unl kely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.
`
`
`
`
`
` COPYRIGHT PROTECTED DOCUMENT
`
` ©
`
` ISO 2008
`All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
`electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
`ISO's member body in the country of the requester.
`ISO copyright office
`Case postale 56 CH-1211 Geneva 20
`Tel. + 41 22 749 01 11
`Fax + 41 22 749 09 47
`E-mail copyright@iso.org
`Web www.iso.org
`Published in Switzerland
`
`
`Copyright International Organization for Standardization
`Provided by IHS under license with ISO
`No reproduction or networking permitted without license from IHS
`
`ii
`
`
`Not for Resale
`
`© ISO 2008 – All rights reserved
`
`Novartis Exhibit 2196.002
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`ISO 10993-7:2008(E)
`
`Contents
`
`Page
`
`Foreword............................................................................................................................................................ iv
`Introduction ....................................................................................................................................................... vi
`1
`Scope ......................................................................................................................................................1
`2
`Normative references ............................................................................................................................1
`3
`Terms and definitions ...........................................................................................................................1
`4
`Requirements .........................................................................................................................................2
`4.1
`General....................................................................................................................................................2
`4.2
`Categorization of devices .....................................................................................................................2
`4.3
`Allowable limits......................................................................................................................................3
`4.4
`Determination of EO and ECH residuals .............................................................................................5
`5
`Product release ....................................................................................................................................10
`5.1
`General..................................................................................................................................................10
`5.2
`Release of products without dissipation curve data .......................................................................10
`5.3
`Procedure for product release using residue dissipation curves ..................................................10
`Annex A (normative) Evaluation of gas chromatograms..............................................................................12
`Annex B (informative) Gas chromatographic determination for EO and ECH............................................15
`Annex C (informative) Flowchart and guidance for the application of this part of ISO 10993 series
`of standards to the determination of EO and ECH residuals in medical devices.........................19
`Annex D (informative) Factors influencing product residual........................................................................26
`Annex E (informative) Extraction conditions for determination of residual EO .........................................28
`Annex F (informative) Rationale for the provisions of this part of ISO 10993 ............................................29
`Annex G (informative) Establishment of allowable limits for EO .................................................................33
`Annex H (informative) Establishment of allowable limits for ECH...............................................................50
`Annex I (informative) Establishment of allowable limits for EG ...................................................................59
`Annex J (informative) Preparation of EO and ECH standards......................................................................63
`Annex K (informative) Ethylene oxide residue measuring methods ...........................................................67
`Bibliography ......................................................................................................................................................74
`
`
`© ISO 2008 All rights reserved
`Copyright International Organization for Standardization
`Provided by IHS under license with ISO
`No reproduction or networking permitted without license from IHS
`
`Not for Resale
`
`iii
`
`Novartis Exhibit 2196.003
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`ISO 10993-7:2008(E)
`
`Foreword
`
`ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
`(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
`technical committees. Each member body interested in a subject for which a technical committee has been
`established has the right to be represented on that committee. International organizations, governmental and
`non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
`International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
`
`International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
`
`The main task of technical committees is to prepare International Standards. Draft International Standards
`adopted by the technical committees are circulated to the member bodies for voting. Publication as an
`International Standard requires approval by at least 75 % of the member bodies casting a vote.
`
`Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
`rights. ISO shall not be held responsible for identifying any or all such patent rights.
`
`ISO 10993-7 was prepared by Technical Committee ISO/TC 194, Biological evaluation of medical devices.
`
`This second edition cancels and replaces the first edition (ISO 10993-7:1995) which has been technically
`revised.
`
`ISO 10993 consists of the following parts, under the general title Biological evaluation of medical devices:
`
` Part 1: Evaluation and testing within a risk management system
`
` Part 2: Animal welfare requirements
`
` Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
`
` Part 4: Selection of tests for interactions with blood
`
` Part 5: Tests for in vitro cytotoxicity
`
` Part 6: Tests for local effects after implantation
`
` Part 7: Ethylene oxide sterilization residuals
`
` Part 9: Framework for identification and quantification of potential degradation products
`
` Part 10: Tests for irritation and skin sensitization
`
` Part 11: Tests for systemic toxicity
`
` Part 12: Sample preparation and reference materials
`
` Part 13: Identification and quantification of degradation products from polymeric medical devices
`
` Part 14: Identification and quantification of degradation products from ceramics
`
` Part 15: Identification and quantification of degradation products from metals and alloys
`
`iv
`Copyright International Organization for Standardization
`
`Provided by IHS under license with ISO
`No reproduction or networking permitted without license from IHS
`
`Not for Resale
`
`© ISO 2008 – All rights reserved
`
`Novartis Exhibit 2196.004
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`ISO 10993-7:2008(E)
`
` Part 16: Toxicokinetic study design for degradation products and leachables
`
` Part 17: Establishment of allowable limits for leachable substances
`
` Part 18: Chemical characterization of materials
`
` Part 19: Physico-chemical, morphological and topographical characterization of materials [Technical
`Specification]
`
` Part 20: Principles and methods
`Specification]
`
`for
`
`immunotoxicology
`
`testing of medical devices
`
`[Technical
`
`© ISO 2008 All rights reserved
`Copyright International Organization for Standardization
`Provided by IHS under license with ISO
`No reproduction or networking permitted without license from IHS
`
`Not for Resale
`
`v
`
`Novartis Exhibit 2196.005
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`ISO 10993-7:2008(E)
`
`Introduction
`
`Requirements for the development, validation and routine control of an ethylene oxide sterilization process for
`medical devices are given in International Standards developed by ISO/TC 198. Certain requirements relating
`to medical devices for biological testing, selection of tests, and the allocation of devices to categories are dealt
`with in a variety of International Standards developed by ISO/TC 194. The specific requirement for ethylene
`oxide and other sterilization process residuals was referred to ISO/TC 194. Other International Standards
`delineate particular requirements for biological testing for specific products.
`
`As noted in the introduction to ISO 11135-1:2007, when determining the suitability of ethylene oxide (EO) for
`sterilization of medical devices, it is important to ensure that the levels of residual EO, ethylene chlorohydrin
`(ECH) and ethylene glycol (EG) pose a minimal risk to the patient in normal product use. Therefore, it is
`important that the use of alternative materials and sterilization processes be considered during product design
`and development. EO is known to exhibit a number of biological effects. In the development of this part of
`ISO 10993, consideration was given to these effects, which include irritation, organ damage, mutagenicity and
`carcinogenicity in humans and animals, and reproductive effects in animals. Similar consideration was given
`to the harmful effects of ECH and EG. In practice, for most devices, exposure to EO and ECH is considerably
`lower than the maximum values specified in this part of ISO 10993.
`
`Moreover, when the choice for EO sterilization has been made, irrespective of the provisions of this part of
`ISO 10993, exposure to EO residues should be minimized. Requirements herein are in addition to the
`biological evaluation and testing requirements for each individually designed medical device as indicated in
`ISO 10993-1. The biological evaluation and testing requirements, combined with the EO-sterilization process
`residue limits, form the justification that an EO-sterilized device is acceptable for use. Maximum allowable
`residues for ethylene chlorohydrin (ECH), when ECH has been found to be present in medical devices
`sterilized with EO, are also specified. Local effects (e.g., irritation) have been considered and are incorporated
`in the tolerable contact limit (TCL) as given in 4.3.5.2 and Annex G for EO, and in 4.3.5.3 and Annex H for
`ECH.
`
`
`
`vi
`Copyright International Organization for Standardization
`
`Provided by IHS under license with ISO
`No reproduction or networking permitted without license from IHS
`
`Not for Resale
`
`© ISO 2008 – All rights reserved
`
`Novartis Exhibit 2196.006
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`INTERNATIONAL STANDARD
`
`ISO 10993-7:2008(E)
`
`
`
`Biological evaluation of medical devices —
`
`Part 7:
`Ethylene oxide sterilization residuals
`
`1 Scope
`
`This part of ISO 10993 specifies allowable limits for residual ethylene oxide (EO) and ethylene chlorohydrin
`(ECH) in individual EO-sterilized medical devices, procedures for the measurement of EO and ECH, and
`methods for determining compliance so that devices may be released. Additional background, including
`guidance and a flowchart showing how this document is applied, are also included in the informative annexes.
`
`EO-sterilized devices that have no patient contact (e.g., in vitro diagnostic devices) are not covered by this
`part of ISO 10993.
`
`NOTE
`
`This part of ISO 10993 does not specify limits for ethylene glycol (EG).
`
`2 Normative references
`
`The following referenced documents are indispensable for the application of this document. For dated
`references, only the edition cited applies. For undated references, the latest edition of the referenced
`document (including any amendments) applies.
`
`ISO 10993-1:—1), Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk
`management process
`
`ISO 10993-3, Biological evaluation of medical devices — Part 3: Tests for genotoxicity, carcinogenicity and
`reproductive toxicity
`
`ISO 10993-10, Biological evaluation of medical devices — Part 10: Tests for irritation and delayed-type
`hypersensitivity
`
`ISO 10993-12, Biological evaluation of medical devices — Part 12: Sample preparation and reference
`materials
`
`ISO 10993-17:2002, Biological evaluation of medical devices — Part 17: Establishment of allowable limits for
`leachable substances
`
`3 Terms and definitions
`
`For the purposes of this document, the terms and definitions given in ISO 10993-1, ISO 10993-17 and the
`following apply.
`
`3.1
`simulated-use extraction
`extraction to demonstrate compliance with the requirements of this part of ISO 10993, by evaluating residue
`levels available to the patient or user from devices during the routine use of a device with water extraction to
`simulate product use
`
`
`1) To be published. (Revision of ISO 10993-1:2003)
`
`© ISO 2008 All rights reserved
`Copyright International Organization for Standardization
`Provided by IHS under license with ISO
`No reproduction or networking permitted without license from IHS
`
`Not for Resale
`
`1
`
`Novartis Exhibit 2196.007
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`ISO 10993-7:2008(E)
`
`3.2
`exhaustive extraction
`extraction until the amount of EO or ECH in a subsequent extraction is less than 10 % of that detected in the
`first extraction, or until there is no analytically significant increase in the cumulative residue levels detected
`
`As it is not possible to demonstrate the exhaustive nature of residual recovery, the definition of exhaustive
`NOTE
`extraction adopted is as above.
`
`4 Requirements
`
`4.1 General
`
`Information on the derivation of the limits in this part of ISO 10993 as well as other important background
`NOTE
`information and guidance relevant to the use of this document is contained in the informative annexes.
`
`This clause specifies maximum allowable residues for ethylene oxide (EO) for each individual medical device
`sterilized with EO. As noted in the introduction to ISO 11135-1:2007, when determining the suitability of EO for
`sterilization of medical devices, it is important to ensure that the levels of residual EO, ethylene chlorohydrin
`(ECH) and ethylene glycol (EG) pose a minimal risk to the patient in normal product use. Moreover, when the
`choice for EO sterilization has been made, irrespective of the provisions of this standard, exposure to EO
`residues should be minimized. Maximum allowable residues for ECH, when ECH has been found to be
`present in medical devices sterilized with EO, are also specified. Local effects (e.g., irritation) have been
`considered and are incorporated in the tolerable contact limit (TCL) as discussed in 4.3.5.2 and Annex G for
`EO, and 4.3.5.3 and Annex H for ECH. No device limits are specified for EG because a risk assessment
`(Annex I) indicates that calculated allowable levels are higher than those likely to occur in a medical device.
`However, the potential exists for acute haemodynamic and haemolytic effects to occur following rapid
`intravenous administration of hyperosmolar compounds like EG. Ethylene oxide sterilization of medical
`devices would not be expected to produce hyperosmolar solutions. Methods for the determination of EO and
`ECH are given in 4.4.
`
`The requirements in this part of ISO 10993 are in addition to the biological testing requirements set out in
`ISO 10993-1. For devices sterilized using ethylene oxide, attention shall be paid in particular to ISO 10993-3
`and ISO 10993-10. All applicable requirements of ISO 10993-1 shall take into account the EO residual level at
`the time of release for each individually designed medical device.
`
`Results of the biological assessment of the device may dictate more stringent limits than those specified
`in 4.3, which are designed to protect against systemic effects.
`
`4.2 Categorization of devices
`
`In establishing the maximum daily doses of EO and ECH that a medical device is allowed to deliver to
`patients, devices shall be categorized according to the duration of contact.
`
`Devices shall be placed into one of three exposure categories in accordance with ISO 10993-1:—, 5.3:
`
`a)
`
`limited exposure (A) – devices whose cumulative single, multiple or repeated use or contact is up to 24 h;
`
`b) prolonged exposure (B) – devices whose cumulative single, multiple, or repeated long-term use or
`contact is likely to exceed 24 h but not 30 d;
`
`c) permanent contact (C) – devices whose cumulative single, multiple or repeated long-term use or contact
`exceeds 30 d.
`
`If a material or device can be placed in more than one duration category, the more rigorous testing and/or
`evaluation considerations should apply. With multiple exposures, the decision into which category a device is
`placed should take into account the potential cumulative effect, bearing in mind the period of time over which
`these exposures occur.
`
`As it is applied in this part of ISO 10993, “multiple use” is defined to mean repeated use of the same device
`NOTE
`type, e.g. dialyser cartridges.
`
`2
`Copyright International Organization for Standardization
`
`Provided by IHS under license with ISO
`No reproduction or networking permitted without license from IHS
`
`Not for Resale
`
`© ISO 2008 – All rights reserved
`
`Novartis Exhibit 2196.008
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`ISO 10993-7:2008(E)
`
`4.3 Allowable limits
`
`4.3.1 General
`
`For each medical device, the maximum allowable doses of EO and ECH delivered to patients shall not exceed
`the values given below for the exposure category that the device has been placed into in accordance with 4.2.
`
`The limits for permanent contact and prolonged exposure devices are expressed as maximum average daily
`doses. These limits carry additional constraints for the first 24 h of the exposure period and, in the case of the
`permanent contact devices, for the first 30 days. These constraints place limitations on the amount of EO and
`ECH that can be delivered to the patient during these early time periods. If data are available, consideration
`should be given for proportioning the limits downward if multiple devices with the residue of concern are used
`at one time, or proportioning the limits upward when device use is only for a part of the exposure period of
`concern. These concomitant exposure factors (CEF) and proportional exposure factors (PEF) are given in
`ISO 10993-17. The procedure that was used to establish the allowable limits is described in Annex G for EO,
`in Annex H for ECH, and the rationale for considering the establishment of allowable limits for EG is described
`in Annex I.
`
`4.3.2 Permanent contact devices
`
`The average daily dose of EO to patient shall not exceed 0,1 mg/d. In addition, the maximum EO dose shall
`not exceed:
`
` 4 mg in the first 24 h;
`
` 60 mg in the first 30 d;
`
` 2,5 g in a lifetime.
`
`The average daily dose of ECH to patient shall not exceed 0,4 mg/d. In addition, the maximum ECH dose
`shall not exceed:
`
` 9 mg in the first 24 h;
`
` 60 mg in the first 30 d;
`
` 10 g in a lifetime.
`
`4.3.3 Prolonged exposure devices
`
`The average daily dose of EO to patient shall not exceed 2 mg/d. In addition, the maximum EO dose shall not
`exceed:
`
` 4 mg in the first 24 h;
`
` 60 mg in the first 30 d.
`
`The average daily dose of ECH to patient shall not exceed 2 mg/d. In addition, the maximum ECH dose shall
`not exceed:
`
` 9 mg in the first 24 h;
`
` 60 mg in the first 30 d.
`
`4.3.4 Limited exposure devices
`
`The average daily dose of EO to patient shall not exceed 4 mg.
`
`The average daily dose of ECH to patient shall not exceed 9 mg.
`
`© ISO 2008 All rights reserved
`Copyright International Organization for Standardization
`Provided by IHS under license with ISO
`No reproduction or networking permitted without license from IHS
`
`Not for Resale
`
`3
`
`Novartis Exhibit 2196.009
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`ISO 10993-7:2008(E)
`
`4.3.5 Tolerable contact limits for surface contacting devices and implants
`
`4.3.5.1 Overview
`
`The tolerable contact limit (TCL) is expressed in units of micrograms per square centimetre for EO and
`milligrams per square centimetre for ECH. The unit of square centimetre represents the surface area of the
`patient-device interface.
`
`NOTE
`
`The intent of this subclause is to prevent localized irritation due to EO or ECH released from the device.
`
`4.3.5.2
`
`Tolerable contact limit for EO
`
`Either the EO TCL for surface contacting devices and implants shall not exceed 10 µg/cm2 or it shall exhibit
`negligible irritation as specified in ISO 10993-10.
`
`4.3.5.3
`
`Tolerable contact limit for ECH for surface contacting devices
`
`Either the ECH TCL for surface contacting devices and implants shall not exceed 5 mg/cm2 or it shall exhibit
`negligible irritation as specified in ISO 10993-10.
`
`4.3.6 Special situations
`
`For multi-device systems the limits shall apply to each individual patient-contact device.
`
`Residue of EO in intraocular lenses shall not exceed 0,5 µg EO per lens per day, or 1,25 µg per lens. Prorate
`limits for other intraocular devices are set on the basis of the mass of the device, with the mass of an
`intraocular lens taken as 20 mg. The acceptability of ECH levels in intraocular devices made from viscoelastic
`materials that contain chlorine may need to be evaluated, as the level of ECH that results in ocular toxicity is
`about four times greater than the corresponding EO level.
`
`For blood cell separators used in patient and donor blood collection, the maximum allowable dose of EO is
`10 mg and the maximum allowable dose of ECH shall not exceed 22 mg.
`
`For blood oxygenators and blood separators, the maximum allowable dose of EO to patient is 60 mg and the
`maximum allowable dose of ECH shall not exceed 45 mg.
`
`For devices used in cardiopulmonary bypass procedures, the maximum allowable limits shall be 20 mg for EO
`and 9 mg for ECH.
`
`For extracorporeal blood purification devices, the EO and ECH limits specified shall be 4,6 mg/device, but the
`allowable EO dose for a lifetime may be exceeded.
`
`For drapes that are intended to contact only intact skin, the maximum allowable limits shall be the TCL of
`10 g/cm2 for EO and 5 mg/cm2 for ECH, or the drapes shall exhibit negligible irritation as specified in
`ISO 10993-10.
`
`The rationale for specifying EO limits for certain devices that are at variance with the general requirements
`NOTE
`appears in Annex F.
`
`A flowchart providing guidance for the application of this part of ISO 10993 to the determination of EO
`residuals in medical devices is presented in Annex C.
`
`4
`Copyright International Organization for Standardization
`
`Provided by IHS under license with ISO
`No reproduction or networking permitted without license from IHS
`
`Not for Resale
`
`© ISO 2008 – All rights reserved
`
`Novartis Exhibit 2196.0010
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`ISO 10993-7:2008(E)
`
`4.4 Determination of EO and ECH residuals
`
`4.4.1 General
`
`4.4.1.1
`
`Procedure
`
`The procedure for determining compliance with 4.3 consists of extracting the residue from samples,
`determining the amount of residue, determining the contact surface of the device, and analysing and
`interpreting the data.
`
`DANGER — Analysts and others who obtain samples should perform all work involving the use of the
`chemicals and solvents required for these methods in a fume cupboard whilst wearing appropriate
`protective clothing, and should review the Material Safety Data information for each chemical prior to
`such use. Healthcare workers using EO-sterilized medical devices shall take appropriate precautions
`to protect against exposure to residues, which may be required by local occupational health and
`safety regulations.
`
`4.4.1.2
`
`Ethylene oxide
`
`This is a flammable gas that is irritating to body surfaces and highly reactive. It is mutagenic under many
`conditions, has fetotoxic and teratogenic properties, can adversely effect testicular function and can produce
`injury to many organ systems in the body. In cancer studies in animals, inhalation exposure produced several
`types of neoplastic changes including leukaemia, brain tumours and mammary tumours while ingestion or
`subcutaneous administration produced tumours only at the site of contact. One investigator has reported
`higher cancer and mortality rates in some subpopulations of exposed workers. However, the results of several
`studies in workers have shown even weaker associations. See References [177], [178] and [181]. In 1994 the
`International Agency for Research on Cancer (IARC) reclassified EO as a human carcinogen (class 1) based
`mainly on its mechanism of action. See Reference [75].
`
`4.4.1.3
`
`Ethylene chlorohydrin
`
`This is a flammable liquid that is irritating to body surfaces, acutely toxic and readily absorbed through the skin
`in toxic amounts. It has weak mutagenic potential, has some potential to produce fetotoxic and teratogenic
`changes and can produce injury to several organ systems in the body including lungs, kidneys, central
`nervous system and cardiovascular system. It was negative in cancer bioassays in animals.
`
`4.4.2 Determination of residue
`
`A valid method of extraction and measurement shall be used to determine the amount of EO and, where
`necessary, ECH delivered to the patient.
`
`If ECH is not detected based on the results of analyses performed using the methods given in either K.4.2 or
`K.4.7, no further monitoring for ECH is required.
`
`Many gas chromatography (GC) methods that use a capillary column instead of a packed column will produce
`NOTE
`EO, ECH and EG results during a single sample run.
`
`The guiding principle in selecting appropriate extraction methods (4.4.6) for the quantitative determination of
`EO and, where necessary, ECH is the evaluation of the dose to the patient in order to show compliance with
`the requirements set out in 4.3.
`
`Where residues are shown to be within the requirements for products tested by exhaustive extraction, there is
`no need to further challenge the device by simulated-use extraction, provided all applicable limits in 4.3 are
`met. When exhaustive extraction is used, particular attention shall be paid to the limits expressed for the first
`24 h and for the first 30 days in 4.3.
`
`Many analytical methods for these EO-sterilization residuals have been described and are reviewed in the
`Bibliography. However, the enormous diversity of materials and methods of construction of sterile medical
`
`© ISO 2008 All rights reserved
`Copyright International Organization for Standardization
`Provided by IHS under license with ISO
`No reproduction or networking permitted without license from IHS
`
`Not for Resale
`
`5
`
`Novartis Exhibit 2196.0011
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`ISO 10993-7:2008(E)
`
`devices may, in certain cases, still present problems in determining residual EO and ECH levels using the
`methods given in the Bibliography. Therefore, any method that has been shown to be analytically sound (i.e.
`demonstrated accuracy, precision, linearity, sensitivity, and selectivity) may be used, provided that it has been
`validated. Annex A contains general validation requirements for gas chromatographic methods.
`
`4.4.3 Product sampling and sample “blank”
`
`4.4.3.1
`
`Product sampling
`
`Samples to be used for residual analysis shall be selected in such a manner as to be truly representative of
`the product. When selecting samples, attention shall be given to the many factors described in Annex D.
`Since many of these factors influence not only the initial levels of residuals in device components but also the
`rate of residue dissipation, they shall also be considered when test samples are drawn from a processed load
`and sent to the laboratory for analysis. Removal of the product samples from the processed load soon after a
`sterilization cycle is completed and shipment to a laboratory far from the sterilization site or storage in the
`laboratory for later analysis can jeopardize correlations of residual levels on the samples with those on the
`rest of the load. Moreover, if samples cannot be drawn from the load and handled so that the effect on
`aeration conditions for the sample will be negligible, an experiment to establish the relationship between the
`sample aeration and load aeration at various seasons of the year shall be carried out.
`
`Precautions shall be taken to minimize or control the effects of laboratory conditions on the rate of aeration for
`test samples that have been removed from a product load (see D.1.5). In addition, operator and analyst safety
`shall be ensured. Samples should remain with the product load until the day of analysis or until test samples
`are retrieved and immediately frozen. The time between removal of samples from a controlled aeration area
`and the beginning of extraction should be held to a minimum. Samples shall be sealed, shipped and stored
`frozen when analysis is delayed. Samples shall be shipped on dry ice on overnight delivery. Dry ice shall
`remain in the shipping container throughout the shipment and be present when the package is opened in the
`laboratory. Test samples may also be taken directly from the product load at the desired aeration interval and
`immediately placed into a headspace vial, which is sealed and then shipped to the laboratory for analysis. As
`an alternative, samples may be extracted and the extraction fluid shipped to the analytical laboratory for
`analysis. If the extraction fluid is water, then shipment shall be done such that the fluid is kept at ice-cold
`temperatures (< 10 °C) until arrival. Testing should be carried out to measure hydrolysis of EO to EG.
`
`Samples to be analysed shall be placed in a fume cupboard and removed from the packaging. Samples shall
`be prepared according to any applicable pre-use instructions in the product labelling. Extractions shall be
`started as soon as possible after the device has been removed from the packaging or pre-use preparations
`have been completed.
`
`4.4.3.2
`
`Sample “blank”
`
`To ensure that no other sample matrix components with the same retention time as any of the residues being
`determined are present, a “blank” sample shall be evaluated for the possible presence of such interferences
`by the extraction of a non-sterilized sample using the identical procedure being applied to the EO-sterilized
`samples. In the event of materials being extracted from such a “blank” with conflicting or overlapping retention
`times in the GC analysis, chromato

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket