throbber
ARTICLE
`
`Toxic anterior segment syndrome: Common
`causes
`
`Carolee M. Cutler Peck, MD, MPH, Jacob Brubaker, MD, Sue Clouser, RN, MSN, CRNO,
`Chris Danford, Henry E. Edelhauser, PhD, Nick Mamalis, MD
`
`PURPOSE: To identify the most common risk factors associated with toxic anterior segment
`syndrome (TASS).
`
`SETTING: Ophthalmic surgical centers in the United States, Argentina, Brazil, Italy, Mexico, Spain,
`and Romania.
`
`METHODS: A TASS questionnaire on instrument cleaning and reprocessing and extraocular and in-
`traocular products used during cataract surgery was placed on the American Society of Cataract and
`Refractive Surgery web site. A retrospective analysis of questionnaires submitted by surgical cen-
`ters reporting cases of TASS was performed between June 1, 2007, and May 31, 2009, to identify
`commonly held practices that could cause TASS. Members of the TASS Task Force made site visits
`between October 1, 2005, and May 31, 2009, and the findings were evaluated.
`
`RESULTS: Data from 77 questionnaires and 54 site visits were analyzed. The reporting centers
`performed 50 114 cataract surgeries and reported 909 cases of TASS. From January 1, 2006, to
`date, the 54 centers reported 367 cases in 143 919 procedures; 61% occurred in early 2006. Com-
`mon practices associated with TASS included inadequate flushing of phaco and irrigation/aspiration
`handpieces, use of enzymatic cleansers, detergents at the wrong concentration, ultrasonic bath,
`antibiotic agents in balanced salt solution, preserved epinephrine, inappropriate agents for skin
`prep, and powdered gloves. Reuse of single-use products and poor instrument maintenance and
`processing were other risk factors.
`
`CONCLUSIONS: The survey identified commonly held practices associated with TASS. Understand-
`ing these findings and the safe alternatives will allow surgical center personnel to change their
`practices as needed to prevent TASS.
`
`Financial Disclosure: No author has a financial or proprietary interest in any material or method
`mentioned.
`
`J Cataract Refract Surg 2010; 36:1073 1080 Q 2010 ASCRS and ESCRS
`
`Toxic anterior segment syndrome (TASS) is a sterile in-
`flammatory reaction of unknown incidence that can oc-
`cur after anterior segment surgery. It typically presents
`within 12 to 48 hours of surgery. The most common
`finding is diffuse limbus-to-limbus corneal edema
`(Figure 1) secondary to damage from a toxic insult to
`the endothelial cell layer. Widespread breakdown of
`the blood–aqueous barrier is another hallmark of this
`condition, with fibrin in the anterior chamber and hypo-
`pyon present in 75% of cases (Figure 2). Damage to the
`iris may cause the pupil to dilate or become slightly
`irregular, and glaucoma secondary to trabecular mesh-
`work damage may also occur.
`Treatment with intense topical steroidal agents will
`eventually lead to resolution of the inflammation;
`
`however, in severe cases there may be lasting sequelae,
`such as permanent corneal edema, glaucoma, and
`other effects of chronic inflammation. Various entities
`have been shown to cause TASS. These include, but
`are not limited to, endotoxin; denatured ophthalmic
`viscosurgical devices (OVDs); preservatives such as
`benzalkonium chloride, bisulfites, and metabisulfites;
`heavy-metal residue,
`fine-matter particulates, and
`substances introduced into the anterior chamber that
`are at a pH or concentration that is toxic to the sensitive
`endothelial cells. In addition, residue of materials used
`in the cleaning and sterilization of ophthalmic instru-
`ments are an increasingly important source of TASS.
`Members of industry and the American Society of
`Cataract and Refractive Surgery (ASCRS) have joined
`
`Q 2010 ASCRS and ESCRS
`
`Published by Elsevier Inc.
`
`0886 3350/$
`
`see front matter 1073
`doi:10.1016/j.jcrs.2010.01.030
`
`Novartis Exhibit 2299.001
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`1074
`
`COMMON CAUSES OF TASS
`
`to create a TASS task force. The goal of this task force is
`to educate anterior segment surgeons on the causes,
`symptoms, and treatment of TASS and to help investi-
`gate outbreaks of TASS.
`Questionnaires are available on the ASCRS web
`siteA to facilitate the reporting of TASS cases. The in-
`formation entered into the questionnaires is reviewed
`by members of the TASS Task Force with the goal of
`identifying potential causes of TASS. Recommenda-
`tions on preventative measures to stop recurrences
`are then made. This study will focus on common
`practices that may lead to TASS as well as alternative
`measures that will help protect patients having
`anterior segment surgery.
`
`MATERIALS AND METHODS
`The TASS questionnaires were posted on the ASCRS web
`site in 2007. A retrospective analysis was performed on ques
`tionnaires submitted between June 1, 2007, and May 31,
`2009. The questionnaires address instrument cleaning and
`reprocessing practices and surgical protocols as well as
`substances and techniques used for cleaning phaco and
`irrigation/aspiration (I/A) handpieces. They also address
`products used extraocularly and intraocularly, such as med
`ications, and the reuse of cannulas and other disposable
`items. Responses were analyzed, and commonly held prac
`tices that lead to cases of TASS are presented here.
`In addition, members of the TASS Task Force made site
`visits at the request of centers with TASS cases. The visits
`occurred between October 1, 2005, and May 31, 2009. During
`the visits, task force members observed preoperative, intra
`operative, and instrument processing procedures to identify
`practices that could be associated with TASS. The data
`gathered from these visits are presented.
`
`RESULTS
`During the study period, 77 TASS questionnaires were
`submitted through the ASCRS web site, 9 of which
`
`Submitted: December 15, 2009.
`Final revision submitted: January 25, 2010.
`Accepted: January 26, 2010.
`
`From the John A. Moran Eye Center (Cutler Peck, Brubaker,
`Mamalis), University of Utah, Salt Lake City, Utah; UT Southwestern
`Medical Center (Clouser), Dallas, and Alcon Laboratories, Inc.
`(Danford), Fort Worth, Texas; and Emory University (Edelhauser),
`Atlanta, Georgia, USA.
`
`Supported in part by unrestricted grants from Research to Prevent
`Blindness, Inc., New York, New York, USA, to the Department of
`Ophthalmology & Visual Sciences, University of Utah.
`
`IOL and
`the ASCRS Symposium on Cataract,
`Presented at
`Refractive Surgery, San Francisco, California, USA, April 2009.
`
`Corresponding author: Nick Mamalis, MD, John A. Moran Eye
`Center, University of Utah, 65 Mario Capecchi Drive, Salt Lake
`City, Utah 84132, USA. E-mail: nick.mamalis@hsc.utah.edu.
`
`Figure 1. Toxic anterior segment syndrome after phakic IOL surgery.
`Note the limbus to limbus corneal edema.
`
`were excluded from analysis. Of those excluded, 5
`were duplicates, 2 did not report any TASS cases,
`and 2 were reporting cases of endophthalmitis. Of
`the centers included, 62 were in the United States; 6 in-
`ternational centers responded, with 1 each from Ar-
`gentina, Brazil, Italy, Mexico, Spain, and Romania.
`Members of the TASS Task Force made 54 site visits,
`all in the United States. Overall, 909 cases of TASS
`were reported in 50 114 cataract surgeries performed
`concurrently at the reporting centers that submitted
`questionnaires. From January 1, 2006, to date, the 54
`centers visited by a TASS Task Force member reported
`367 cases of TASS in 143 919 procedures performed;
`61% of them occurred in 2006.
`Table 1 shows the cleaning procedures associated
`with TASS. Inadequate flushing of the phaco and
`I/A handpieces after surgery was the most commonly
`observed and reported behavior that can lead to TASS.
`
`Figure 2. Hypopyon is seen in the anterior chamber of this patient
`with TASS.
`
`J CATARACT REFRACT SURG VOL 36, JULY 2010
`
`Novartis Exhibit 2299.002
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`COMMON CAUSES OF TASS
`
`1075
`
`Of the 68 centers that filled out questionnaires, more
`than 60% used less than the recommended 120 cc per
`port (range 2 to 100 cc per port). Other factors were
`occluded I/A tips during surgery, a symptom of
`inadequate flushing.
`Table 2 shows the observations made during site
`visits at centers with diagnosed cases of TASS. The
`top observations were inadequate flushing of phaco
`and I/A handpieces and cannulated equipment, use
`of enzymatic cleaners and detergents, use of reusable
`cannulas, and inadequate or no manual cleaning of
`instruments. Cleaned instruments were often left on
`towels that were not lint-free, and many centers did
`not
`train personnel regarding TASS and proper
`cleaning practices.
`Medications given intracamerally or added to a bal-
`anced salt solution irrigant were another potential
`source of TASS. Table 3 shows the products associated
`
`Table 1. Cleaning procedures associated with TASS.
`
`Reported Practice
`
`Amount of flush used for handpieces
`120 cc or more
`!120 cc (range 2 100 cc)
`Unknown
`Not specified
`Occluded I/A tips
`Yes
`No
`Final rinse water
`Deionized/distilled
`Sterile/pyrogen free
`Tap
`Unknown
`Not specified
`Use of enzymatic cleaners
`Yes
`No
`Unknown
`Use of US Bath
`Yes
`No
`Frequency of cleaning US bath
`After each use
`More than once daily
`Unknown
`No cleaning protocol
`Ophthalmic instruments separated for
`cleaning
`Yes
`No
`Don’t know
`
`I/A Z irrigation/aspiration; US Z ultrasonic
`
`Centers Reporting
`Practice, n (%)
`(N Z 68)
`
`18 (26)
`42 (62)
`3 (4)
`5 (7)
`
`27 (40)
`41 (60)
`
`30 (44)
`22 (32)
`12 (18)
`2 (3)
`2 (3)
`
`36 (53)
`28 (41)
`4 (6)
`
`43 (63)
`25 (37)
`
`8 (19)
`31 (72)
`3 (7)
`1 (2)
`
`55 (81)
`9 (13)
`4 (6)
`
`Table 2. Observations during site visits at centers with diag
`nosed cases of TASS.
`
`Observation Category
`
`Inadequate flushing of phaco and
`I/A handpieces and cannulated
`equipment
`Use of enzymatic cleaners and
`detergents
`Use of reusable cannulas
`Inadequate or no manual cleaning
`of instruments
`Use of preserved epinephrine
`Reuse of single use
`devices, blades, tips,
`sleeves, etc.
`Use of tap water with no sterile
`water final rinse
`Inadequate personnel or trays to
`allow proper preparation of
`instruments
`No immediate cleaning allowing
`OVD and surgical solutions to
`dry on devices
`Use of preserved medicines in the
`eye
`Reuse of tubing for flushing, latex
`bulb for irrigation, no training,
`no term sterilization,
`instruments stored on towels
`Touching of IOL or patient contact
`areas of instruments with
`gloved hands
`Off label use of lidocaine gel
`Poor instrument maintenance,
`residual autoclave residue and
`instrument mild, rust,
`particulates, lint, etc.
`Use of powdered gloves
`Additives added to balanced salt
`solution against DFU
`Improper use of prep solutions,
`detergents, and cleaners
`Failure to follow manufacturer’s
`DFU, including
`no air flush, use of unapproved
`enzymatics
`Use of postoperative ointment in
`clear cornea cases
`Povidone iodine placed in eye at
`end of procedure
`Incorrect concentration of
`detergent and enzymatic
`cleaners used
`Lack of routine cleaning of US
`
`Frequency of Observation,
`n (%) (N Z 54)
`
`48 (89)
`
`43 (80)
`
`37 (69)
`35 (65)
`
`28 (52)
`28 (52)
`
`27 (50)
`
`23 (43)
`
`20 (37)
`
`18 (33)
`
`20 (37)
`
`18 (33)
`
`17 (31)
`15 (28)
`
`15 (28)
`15 (28)
`
`14 (26)
`
`11 (20)
`
`11 (20)
`
`9 (17)
`
`4 (7)
`
`2 (4)
`
`DFU Z directions for use; I/A Z irrigation/ aspiration; IOL Z intraocu-
`lar lens; OVD Z ophthalmic viscosurgical device; US Z ultrasound
`
`J CATARACT REFRACT SURG VOL 36, JULY 2010
`
`Novartis Exhibit 2299.003
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`1076
`
`COMMON CAUSES OF TASS
`
`Table 3. Products associated with TASS and their use.
`
`Product Used
`
`Antibiotic added to BSS
`Vancomycin
`Gentamicin
`None
`Unknown
`Intracameral antibiotic agents
`Vancomycin
`Cefuroxime
`Moxifloxacin
`Cefazolin sodium
`None
`Preserved epinephrine in BSS
`Yes
`No
`Preserved intracameral anesthetic
`agent
`Yes
`No
`
`BSS Z balanced salt solution
`
`Centers Reporting
`Use, n (%)
`
`16 (24)
`1 (1)
`50 (74)
`1 (1)
`
`7 (10)
`3 (5)
`3 (5)
`1 (1)
`54 (79)
`
`18 (36)
`32 (64)
`
`3 (6)
`44 (94)
`
`with TASS. Antibiotic agents were the most common
`additives to balanced salt solution; some centers
`added epinephrine with preservatives or stabilizers.
`Slightly more than 20% of centers used intracameral
`antibiotic agents; 17 centers (31%) used preserved
`topical lidocaine gel preoperatively.
`Poor instrument maintenance was also pervasive.
`Autoclaves were found to have autoclave residue,
`instrument milk, rust, particulates, or lint at more
`than a quarter of the sites. Other sites failed to follow
`manufacturer’s directions for use when cleaning
`instruments; this included omitting an air flush of
`cannulated equipment, although this had been recom-
`mended, or using enzymatics during processing when
`this was not approved.
`In centers in which non-ophthalmologic surgeries
`were performed, ophthalmic instruments were pro-
`cessed along with instruments from other types of
`surgeries at 9 (13%) centers responding by question-
`naire; 4 (6%) respondents did not know whether
`instruments were separated at the time of processing.
`
`DISCUSSION
`The data in this analysis were self-reported in the case
`of the questionnaires or the result of self-referrals for
`site inspection in relation to TASS cases. The data
`from the site visits were taken from written reports
`submitted at the end of the visit. Although there is mo-
`tivation on the part of surgical centers to obtain
`
`assistance when cases of TASS occur, in an effort to
`prevent future cases, not all cases of TASS may be rec-
`ognized or centers may choose not to report cases if
`only a small percentage of their cataract surgeries re-
`sult in TASS and a problem is not perceived. Because
`we do not have a tally of all TASS cases that occur in
`the United States, and because those reporting cases
`all reported various timeframes in which the reported
`cases occurred for each center, we cannot calculate the
`incidence of TASS from our data.
`The most frequently identified practice associated
`with TASS is inadequate cleaning of phaco and I/A
`handpieces after use. To be most effective, the manu-
`facturer’s directions for use should be strictly followed
`and handpieces should be wiped with a lint-free cloth,
`immediately immersed in sterile water until they are
`flushed, or both.1–3 This was not done at 89% of centers
`visited. This practice is imperative to prevent residual
`OVD and debris from drying within the handpiece
`and tips before they are flushed. This is especially
`important because dried OVD denatures after sterili-
`zation at high temperatures and if this or other
`particles or debris are irrigated into the anterior cham-
`ber, TASS may result.4 While instruments are still
`moist, they must be flushed with 120 mL of fluid per
`port, or more if
`indicated in the manufacturer’s
`directions for use. This can be performed with tap
`water as deionized water. However, the final flush
`must then consist of sterile distilled or sterile deionized
`water. If cleaned instruments are stored on towels, the
`towels should be lint free.3,5,6
`In addition to addressing the problem of inadequate
`flushing, it is important to address some of the reasons
`it occurs. Even if an adequate cleaning protocol is in
`place, if surgical centers are understaffed or there are
`too few surgical trays to allow proper cleaning of
`instruments between cases (as observed at 43% of cen-
`ters visited), time constraints will lead to insufficient
`flushing and patients will be placed at risk for TASS.
`The reuse of tubing for flushing and latex bulbs for
`irrigation is also a potential issue. These items are
`difficult to clean and hard to sterilize, which can lead
`to improper processing between cases.
`It is imperative that surgical centers ensure that
`there are adequate numbers of personnel and surgical
`trays to allow all steps of cleaning and sterilization to
`be completed between uses of ophthalmic instru-
`ments. It is also crucial that personnel are trained
`sufficiently in the causes of TASS and the proper steps
`of cleaning and sterilization.2,3,7–9
`Eighty percent of surgical centers visited and 53% of
`those responding via questionnaire were using enzy-
`matic detergents in the processing of their ophthalmic
`instruments. The benefit of using enzymatic deter-
`gents to clean ophthalmologic instruments has not
`
`J CATARACT REFRACT SURG VOL 36, JULY 2010
`
`Novartis Exhibit 2299.004
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`COMMON CAUSES OF TASS
`
`1077
`
`been established and may actually be prohibited in the
`manufacturer’s directions for use for specific products.
`Furthermore, the use of detergents mixed at the wrong
`concentration has been linked to TASS outbreaks.10
`Seven percent of centers visited were using the wrong
`concentration of detergents, putting their patients at
`risk for TASS as a result of residual detergents and in-
`complete rinsing.
`Enzymatic detergents often have the exotoxin sub-
`tilisin or a-amylase enzymes as their active ingredi-
`ents. These are only deactivated at
`temperatures
`higher than 140C, and most autoclaves do not reach
`temperatures higher than 120C to 130C. Thus, it is
`likely that residue from enzymatic detergents will
`build up on reused instruments.11 Human and rabbit
`studies evaluating the effect of enzymatic detergents
`on the anterior chamber showed dose-related corneal
`swelling; ultrastructural damage to the endothelial
`layer, leading to increased corneal permeability; and
`an increased inflammatory response in the iris.12
`The purpose behind the use of enzymatic detergents
`is to rid ophthalmic instruments of debris. If instru-
`ments are kept moist immediately after use before
`flushing and if proper flushing with an adequate
`volume is performed, there should be no adherent
`debris and thus no need for enzymatic detergents. If
`detergents are used,
`it
`is imperative that strict
`attention is paid to the dilution and expiration date.
`Furthermore, instruments processed with detergents
`must be rinsed with copious amounts of fluid accord-
`ing to the manufacturer’s directions for use. Recom-
`mended volumes should be considered a minimum
`volume, and the final rinse should be performed
`with sterile distilled or sterile deionized water.1
`Sixty-three percent of reporting facilities used an
`ultrasonic bath as part of the processing of their
`ophthalmologic instruments. This has been associated
`with the accumulation of heat-stable endotoxins
`produced by bacteria in the bath water.13 Endotoxin
`remaining on instruments after cleaning and steriliza-
`tion can induce the inflammatory reaction of TASS.
`As with enzymatic detergents, the purpose of the
`ultrasonic bath is to dislodge dried debris from instru-
`ments, particularly OVDs. Again, if instruments are
`kept moist after use and then properly flushed with
`an adequate volume of water, there should be no
`adherent debris and thus no need for an ultrasonic
`bath.
`If an ultrasonic bath is used, the manufacturer’s
`directions for use for instruments should be verified
`because some instruments should not be processed
`in this manner. In addition, the ultrasonic bath should
`be designated for medical use only and the manufac-
`turer’s directions for maintenance should be strictly
`followed. Furthermore, all gross material should be
`
`completely removed from instruments before they
`are immersed in bath water. Then, after each use,
`and if in accordance with the manufacturer’s direc-
`tions for use, the bath water should be emptied and
`the tub cleaned with an Environmental Protection
`Agency–registered facility-approved disinfectant.
`This should be followed with a rinse using volumes
`of sterile or tap water adequate to remove the cleaning
`agent completely. Then, 70% to 90% ethyl or isopropyl
`alcohol should be used to clean the tub if this is in
`accordance with the manufacturer’s directions for
`use and not associated with a risk of fire. Endotoxin
`can be removed from the walls of the bath by wiping
`the walls with ethyl or isopropyl alcohol.1,14 The tub
`should then be dried completely with a lint-free cloth
`and refilled immediately before use.
`This process of cleaning must be followed after each
`use to prevent endotoxin buildup. However, only 19%
`of centers responding to the questionnaire cleaned the
`ultrasonic bath after each use as recommended and 4%
`of centers visited had no protocol for routine cleaning
`of the ultrasonic bath.
`Of centers reporting via questionnaire, 25% added
`antibiotic agents to the balanced salt solution irrigant
`and 21% used intracameral antibiotic agents. The use
`of antibiotic agents may be associated with toxicity
`when they are included in anterior chamber irrigant
`and when injected intracamerally at
`the end of
`a case.7 If antibiotic agents are improperly mixed, the
`concentration may be too high or the pH incorrect,
`both of which can prove toxic to the anterior chamber
`tissues.
`The use of vancomycin and gentamicin sulfate in
`anterior segment surgery has been described for
`prophylaxis against endophthalmitis.15 However, the
`use of these products is associated with concerns
`over vancomycin-resistant organisms as well as
`aminoglycoside-related macular toxicity, respectively.
`Furthermore,
`the concentration used in irrigating
`solution and the time of contact with a possible
`contaminant is inadequate for their bacteriostatic or
`bacteriocidal properties to function.
`Studies to evaluate intraocular cefotaxime for endo-
`thelial toxicity have been performed. A prospective
`randomized masked study of 66 patients by
`Kramann16 found no toxicity with 0.4 mL of 0.25%
`cefotaxime in the anterior chamber. Other studies
`evaluating the use of cefuroxime17,18 found no toxicity
`and a role for the agent in endophthalmitis prevention.
`These findings were strengthened with the results in
`a prospective randomized partially masked study in
`which 1.0 mg cefuroxime in 0.1 mL normal saline
`injected into the anterior chamber after surgery de-
`creased the risk for endophthalmitis (0.05% incidence
`rate when counting all
`endophthalmitis
`cases)
`
`J CATARACT REFRACT SURG VOL 36, JULY 2010
`
`Novartis Exhibit 2299.005
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`1078
`
`COMMON CAUSES OF TASS
`
`compared with a placebo or perioperative treatment
`with a third-generation fluoroquinolone drop.19 How-
`ever, an endophthalmitis incidence of 0.056% was
`found with the perioperative use of a fourth-
`generation fluoroquinolone drop.20
`At this time, because there are effective alternatives
`and because there is a risk for TASS from toxicity due
`to the wrong concentration or pH of an improperly
`mixed antibiotic agent, use of intracameral antibiotic
`agents must be viewed with caution.7
`Of surgical centers visited, 52% were using preserved
`epinephrine in balanced salt solution and 37% were
`using other preserved medications intracamerally. In
`addition, at least 36% of questionnaire respondents
`were adding preserved epinephrine to balanced salt so-
`lution. Although bisulfites are not technically a preser-
`vative but rather an antioxidant, they can also cause
`ocular toxicity. Benzalkonium chloride (BAC), a com-
`monly used preservative, is highly toxic to ocular tis-
`sues,21 and rabbit studies22 show a dose-response
`pattern of conjunctivitis,
`flare,
`iritis, and corneal
`changes. In the rabbit study, a safety factor of 10 was
`used to set the safest intraocular concentration at
`0.001%, a level at which the preservative efficacy is
`questionable.
`An outbreak of TASS was associated with the use of
`Eye Stream rinse solution (Alcon Laboratories) pre-
`served with 0.01% BAC, in which a concentration of
`0.013% was inadvertently irrigated into the anterior
`chamber.23 Toxic anterior segment syndrome has
`also been reported after use of OVD preserved with
`BAC,24 and the use of lidocaine gel preserved with
`BAC was recently implicated in a TASS outbreak
`(unpublished data). We recommend that no products
`with preservatives or additives (stabilizers), such as
`bisulfites and metabisulfites, be used in the anterior
`chamber.7,11
`Twenty percent of sites visited were using intraocular
`ointment after clear corneal cases. This practice was
`associated with a TASS outbreak in 8 cases. Of the
`patients affected, 4 required penetrating keratoplasty
`and glaucoma with no light perception developed in
`1 patient. Examination of each of these patients showed
`that petroleum-based ophthalmic ointments had
`gained access to the anterior chamber through a clear
`corneal wound that was patched postoperatively.25
`Site visits showed that 69% of centers were using
`reusable cannulas. Cleaning cannulas is difficult to
`perform and validate, and there are no manufacturer’s
`directions for use. Furthermore, if the cannulas are
`processed with enzymatic detergents, residue will
`likely build up and then be irrigated into the anterior
`chamber. Therefore, when possible, disposable cannu-
`las should be used.7,11 In addition, it is important that
`single-use devices, such as cannulas, blades, tips, and
`
`sleeves, not be reused. Fifty-two percent of centers vis-
`ited were reusing these items. The reprocessing of
`these items is regulated and enforced by the U.S.
`Food and Drug Administration.B
`Poor autoclave maintenance, including autoclave
`residue, instrument milk, rust, particulates, and lint
`within the autoclave, was documented at 28% of cen-
`ters visited. Any such contaminants would be suspect
`for causing TASS. One documented example of poor
`maintenance leading to heavy-metal contamination
`of instruments showed carryover of sulfate, silica,
`copper, zinc, and nickel from feedwater into steam
`condensates associated with TASS outbreaks.26 This
`was found to be the result of changing from flushing
`and draining the steam generator once a week to
`once every 4 to 8 weeks when maintenance was
`outsourced.
`Surgical centers should make sure that preventive
`maintenance, cleaning, and inspection of autoclaves
`are performed as directed by the manufacturer’s direc-
`tions for use and also be documented. In addition, it
`should be verified that the sterilizer is functioning
`properly on a daily basis where possible, or at least
`weekly. Steam sterilization should be performed in
`accordance with published guidelines, and instru-
`ments should be sterilized according to directions for
`use provided by their manufacturer as well as the
`sterilizer’s manufacturer. A terminal sterilization
`should be performed at the end of each surgical
`day.3,5 Glutaraldehyde should not be used in the
`sterilization process because it is highly toxic to the
`tissues of the anterior chamber and will cause TASS
`if inadequately rinsed.27 Plasma gas sterilization has
`also been shown to lead to degradation of brass to
`copper and zinc on cannulated surgical instruments
`and should not be used because this has led to heavy
`metal–induced TASS.28
`The use of powdered gloves can also be a source of
`TASS because the powder can be toxic to the anterior
`chamber structures. Of sites visited, 28 were using
`powdered gloves during anterior chamber cases.
`Powder-free gloves can also pose a risk for develop-
`ing TASS. These gloves are made with a releasing com-
`pound that allows the glove to be easily removed from
`the mold after production. The releasing compound
`stays on the surface of the gloves and can cause ante-
`rior chamber toxicity if it comes in contact with IOLs
`or any instrument entering the anterior chamber.C
`This can occur if the IOL or other instruments entering
`the anterior chamber are touched during cataract
`surgery, a practice observed at 18 surgical centers
`visited (33%).
`Of questionnaire respondents, at least 13% indicated
`that ophthalmic instruments were processed for clean-
`ing alongside dirty surgical
`trays
`from other,
`
`J CATARACT REFRACT SURG VOL 36, JULY 2010
`
`Novartis Exhibit 2299.006
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`COMMON CAUSES OF TASS
`
`1079
`
`nonophthalmic surgeries. However, ASCRS experts
`recommend that all ophthalmic instruments be kept
`separate from other types of surgical
`instruments
`during all steps of processing to avoid contamination
`with bioburden or cleaning chemicals.1
`In conclusion, analysis of data reported in question-
`naires as well as in site visits from centers reporting
`TASS were analyzed carefully. The most common fac-
`tors associated with TASS were inadequate cleaning
`and flushing of ophthalmic instruments and handpie-
`ces. This included the use of detergents and ultra-
`sound water baths. In addition, the use of intraocular
`medications with preservatives or additives was
`associated with TASS in a significant number of cases.
`The proper cleaning and sterilization of ophthalmic in-
`struments, as well as proper attention regarding the
`use of any intraocular medications during surgery,
`can help prevent the occurrence TASS. These findings
`continue to validate the need to follow the recom-
`mendations detailed in the recommended practices
`document created by the TASS Task Force.1
`REFERENCES
`1. American Society of Cataract and Refractive Surgery and
`the American Society of Ophthalmic Registered Nurses. Recom-
`mended practices for cleaning and sterilizing intraocular surgical
`instruments. J Cataract Refract Surg 2007; 33:1095 1100.
`Available
`at:
`http://www.eyeworld.org/ewsupplementarticle.
`php?idZ200. Accessed March 30, 2010
`2. AORN. Recommended practices for cleaning and caring for sur-
`gical instrumentation and powered equipment. In: Association of
`Operating Room Nurses. Standards, Recommended Practices
`& Guidelines. Denver, CO, AORN, 2006; 555 563
`3. Association for the Advancement of Medical Instrumentation.
`Comprehensive Guide to Steam Sterilization and Sterility
`Assurance in Health Care Facilities. ANSI/AAMI ST79 2006;
`4. Kim JH.
`Intraocular inflammation of denatured viscoelastic
`substance in cases of cataract extraction and lens implantation.
`J Cataract Refract Surg 1987; 13:537 542
`5. AORN. Recommended practices for sterilization in the perioper-
`ative practice setting. In: Association of Operating Room Nurses.
`Standards, Recommended Practices & Guidelines. Denver, CO,
`AORN, 2007; 673 687. Available at: http://www.aorn.org/docs/
`assets/3E41EE62-AB20-9320-21647715CD08CB26/Draft RP
`Sterilization.pdf. Accessed March 30, 2010
`6. AORN. Recommended practices for the evaluation and selec-
`tion of products and medical devices used in perioperative
`practice setting. AORN J 1998; 67:270 272
`7. Mamalis N, Edelhauser HF, Dawson DG, Chew J, LeBoyer RM,
`Werner L. Toxic anterior segment syndrome. J Cataract Refract
`Surg 2006; 32:324 333
`8. Burlew J. Care and Handling of Ophthalmic Microsurgical
`Instruments 2nd ed.. San Francisco, CA, American Society of
`Ophthalmic Registered Nurses, 2006
`9. Aorn. Explications for perioperative nursing. In: Association of
`Operating Room Nurses. Standards, Recommended Practices
`& Guidelines. Denver, CO, AORN, 2006; 167 197
`10. Breebaart AC, Nuyts RMMA, Pels E, Edelhauser HF,
`Verbraak FD. Toxic endothelial cell destruction of the cornea
`after routine extracapsular cataract surgery. Arch Ophthalmol
`1990; 108:1121 1125
`
`11. Parikh CH, Edelhauser HF. Ocular surgical pharmacology:
`corneal endothelial safety and toxicity. Curr Opin Ophthalmol
`2003; 14:178 185
`12. Parikh CH, Sippy BD, Martin DF, Edelhauser JF. Effects of
`enzymatic sterilization detergents on the corneal endothelium.
`Arch Ophthalmol 2002; 120:165 172
`13. Kreisler KR, Martin SS, Young CW, Anderson CW, Mamalis N.
`Postoperative inflammation following cataract extraction caused
`by bacterial contamination of
`the cleaning bath detergent.
`J Cataract Refract Surg 1992; 18:106 110
`14. Franken KLMC, Hiemstra HS, van Meijgaarden KE, Subronto Y,
`den Hartigh J, Ottenhoff THM, Drijfhout JW. Purification of his-
`tagged proteins by immobilized chelate affinity chromatography:
`the benefits from the use of organic solvent. Protein Expr Purif
`2000; 18:95 99
`15. Gills JP. Filters and antibiotics in irrigating solution for cataract
`surgery [letter]. J Cataract Refract Surg 1991; 17:385
`16. Kramann C, Pitz S, Schwenn O, Haber M, Hommel G, Pfeiffer N.
`Effects of intraocular cefotaxime on the human corneal endothe-
`lium. J Cataract Refract Surg 2001; 27:250 255
`17. Montan P, Wejde G, Setterquist H, Rylander M, Zetterstro¨ m C.
`Prophylactic intracameral cefuroxime; evaluation of safety and
`kinetics in cataract surgery. J Cataract Refract Surg 2002;
`28:982 987
`18. Montan P, Wejde G, Koranyi G, Rylander M. Prophylactic in-
`tracameral cefuroxime; efficacy in preventing endophthalmitis
`after cataract surgery. J Cataract Refract Surg 2002;
`28:977 981
`19. ESCRS Endophthalmitis Study Group. Prophylaxis of postoper-
`ative endophthalmitis following cataract surgery: results of
`the ESCRS multicenter study and identification of risk factors.
`J Cataract Refract Surg 2007; 33:978 988
`20. Jensen MK, Fiscella RG, Moshifar M, Mooney B. Third- and
`fourth-generation fluoroquinolones: retrospective comparison
`of endophthalmitis after cataract surgery performed over
`10 years. J Cataract Refract Surg 2008; 34:1460 1467
`21. Swan KC. Reactivity of the ocular tissues to wetting agents.
`Am J Ophthalmol 1944; 27:1118 1122
`22. Britton B, Hervey R, Kasten K, Gregg S,

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket