throbber
OPEN ACCESS
`
`Research Article
`
`Thomas von Woedtke1
`Axel Kramer2
`
`1 INP Greifswald e.V. (Leibniz
`Institute for Plasma Science
`and Technology), Greifswald,
`Germany
`2 Institute for Hygiene and
`Environmental Medicine,
`University Greifswald,
`Germany
`
`The limits of sterility assurance
`
`Die Grenzen der Sterilisationssicherheit
`
`Abstract
`Sterility means the absence of all viable microorganisms including vir-
`uses. At present, a sterility assurance level (SAL) of 10–6 is generally
`accepted for pharmacopoeial sterilization procedures, i.e., a probability
`of not more than one viable microorganism in an amount of one million
`sterilised items of the final product. By extrapolating the reduction rates
`following extreme artificial initial contamination, a theoretical overall
`performance of the procedure of at least 12 lg increments (overkill
`conditions) is demanded to verify an SAL of 10–6. By comparison, other
`recommendations for thermal sterilization procedures demand only
`evidence that the difference between the initial contamination and the
`number of test organisms at the end of the process amount to more
`than six orders of magnitude. However, a practical proof of the required
`level of sterility assurance of 10–6 is not possible. Moreover, the attain-
`ability of this condition is fundamentally dubious, at least in non-thermal
`procedures. Thus, the question is discussed whether the undifferentiated
`adherence to the concept of sterility assurance on the basis of a single
`SAL of 10–6 corresponds with the safety requirements in terms of patient
`or user safety, costs and energy efficiency. Therefore, in terms of prac-
`tical considerations, a concept of tiered SALs is recommended, analog-
`ous to the comparable and well-established categorization into “High-
`level disinfection”, “Intermediate-level disinfection” and “Low-level dis-
`infection”. The determination of such tiered SALs is geared both to the
`intended application of the sterilized goods, as well as to the character-
`istics of the products and the corresponding treatment options.
`In the case of aseptic preparation, filling and production procedures, a
`mean contamination probability of 10–3 is assumed. In automated pro-
`cesses, lower contamination rates can be realized. In the case of the
`production of re-usable medical devices, a reduction of at least 2 lg in-
`crements can be achieved through prior cleaning in validated cleaning
`and disinfecting devices. By chemical disinfection, a further reduction
`of ≥5 lg increments is achieved. In the case of sterilized surgical instru-
`ments, an additional concern is that they lay opened in contaminated
`air for the duration of the operation, at least in conventionally ventilated
`operating theaters. Finally, the amount of pathogens necessary to cause
`an infection must be considered. By logical consideration of all aspects,
`it seems possible to partially reduce sterility assurance levels without
`any loss of safety. Proceeding from this, we would like to make the fol-
`lowing suggestions for tiered SAL values, adjusted according to the re-
`spective sterilization task:
`• SAL 10–6 for heat-resistant pharmaceutical preparations (parenter-
`als), suggested term: “Pharmaceutical sterilization”,
`• SAL 10–4 for heat-resistant medical devices, suggested term: “High-
`level sterilization”,
`• SAL 10–3 for heat-sensitive re-usable medical devices, under the
`precondition of a validated cleaning efficacy of >4 lg increments,
`suggested term: “Low-level sterilization”.
`Keywords: sterility, sterility assurance level (SAL), draft of tiered SAL
`values
`
`GMS Krankenhaushygiene Interdisziplinär 2008, Vol. 3(3), ISSN 1863-5245
`
`1/10
`
`I
`
`Novartis Exhibit 2200.001
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`von Woedtke et al.: The limits of sterility assurance
`
`Zusammenfassung
`Sterilität bedeutet die Abwesenheit aller vermehrungsfähigen Mikroor-
`ganismen einschließlich Viren. Derzeit wird für Sterilisationsverfahren
`ein Sterilitätssicherheits-Wert (SAL-Wert) von 10–6 gefordert, d.h. in einer
`Menge von einer Million sterilisierten Gütern darf höchstens ein lebens-
`fähiger Mikroorganismus zu erwarten sein. Durch Extrapolation der
`Mikroorganismen-Reduktionsraten nach artifizieller extremer Ausgangs-
`kontamination (=106 Test-Mikrorganismen pro Prüfobjekt) wird zum
`Nachweis eines SAL-Werts von 10–6 eine theoretische Gesamtredukti-
`onsleistung des Verfahrens um mindestens 12 lg-Stufen abgeleitet
`(„Overkill“). Demgegenüber verlangen andere Empfehlungen lediglich
`den Nachweis, dass die Differenz zwischen Ausgangszahl und Zahl der
`Testorganismen nach Ende des Prozesses mehr als sechs Zehnerpoten-
`zen beträgt. Da der praktische Nachweis des geforderten Sterilisations-
`sicherheitsniveaus von 10–6 unmöglich ist und zumindest bei nichtther-
`mischen Verfahren die Erreichbarkeit dieses Zustands grundsätzlich
`zweifelhaft ist, wird die Fragestellung diskutiert, ob ein undifferenziertes
`Festhalten an dem gegenwärtigen praktizierten Konzept der Sterilisati-
`onssicherheit auf der Basis eines SAL-Wertes von 10–6 unter dem Aspekt
`der Patienten- bzw. Anwendersicherheit sowie der Kosten und des
`Energieverbrauchs den tatsächlichen Sicherheitsanforderungen ent-
`spricht.
`Unter praktischen Gesichtspunkten wäre daher ein Konzept von abge-
`stuften SAL-Werten analog der Differenzierung in „High-level“, „Interme-
`diate-level“ bzw. „Low-level disinfection“ sinnvoll, deren Festlegung sich
`sowohl an der vorgesehenen Anwendung des Sterilisierguts als auch
`an dessen Eigenschaften und den damit verbundenen Behandlungs-
`möglichkeiten orientiert.
`Bei aseptischen Zubereitungs-, Abfüllungs- und Herstellungsverfahren
`wird von einer mittleren Kontaminationswahrscheinlichkeit von 10–3
`ausgegangen, bei automatisierten Prozessen können auch geringere
`Kontaminationsraten realisiert werden. Im Fall der Aufbereitung wieder
`verwendbarer Medizinprodukte ist durch die vorausgehende Reinigung
`in Reinigungs-Desinfektions-Geräten eine Reduktion um mindestens
`2 lg-Stufen erreichbar. Durch die chemische Desinfektion wird nach der
`Reinigung eine weitere Reduktion um ≥5 lg-Stufen erreicht. Bei sterili-
`siertem chirurgischem Instrumentarium kommt hinzu, dass das Sterilgut
`im konventionell belüfteten Operationssaal für die Dauer der Operation
`geöffnet auf dem Instrumentiertisch lagert. Schließlich muss auch die
`Erregermenge berücksichtigt werden, die eine Infektion auszulösen
`vermag. Bei konsequenter Berücksichtigung aller Teilaspekte erscheint
`es möglich, die terminale Sterilisationsbehandlung ohne Sicherheitsver-
`lust deutlich zu reduzieren. Hierfür wird ein Vorschlag für an die jeweilige
`Sterilisationsaufgabe angepasste abgestufte SAL-Werte unterbreitet:
`• SAL 10–6 für thermostabile Arzneizubereitungen, vorgeschlagener
`Terminus: „Pharmazeutische Sterilisation“,
`• SAL 10–4 für thermostabile Medizinprodukte, vorgeschlagener Ter-
`minus: „High-level-Sterilisation“,
`• SAL 10–3 für thermolabile Medizinprodukte unter der Voraussetzung
`einer validierten Reinigungseffektivität >4 lg-Stufen, vorgeschlagener
`Terminus: „Low-level-Sterilisation“.
`Schlüsselwörter: Sterilität, Sterilitätssicherheits-Wert (SAL), Konzept
`abgestufter SAL-Werte
`
`GMS Krankenhaushygiene Interdisziplinär 2008, Vol. 3(3), ISSN 1863-5245
`
`2/10
`
`I
`
`Novartis Exhibit 2200.002
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`von Woedtke et al.: The limits of sterility assurance
`
`The concept of sterility assurance
`– a compromise between overkill,
`viability and safety requirements
`Sterility of a product or object means the complete ab-
`sence of viable microorganisms, including viruses, which
`could pose a risk during administration [67], [81], [61],
`[22]. A sterility assurance level (SAL) of 10–6 is currently
`required for sterilization procedures, i.e., a probability of
`not more than one viable microorganism in one million
`sterilized items of the final product [18].
`The inherent problem with these requirements is that
`evaluating the success of such sterilization by means of
`a final inspection is all but impossible, since contamina-
`tion rates on the order of an SAL of 10–6 cannot be recor-
`ded in experiments [59], [67], [46], [26].
`Thus, model situations have to be created, with the help
`of which conclusions can be drawn regarding the treat-
`ment conditions necessary to attain sterility meeting the
`SAL. Therefore, representative test organisms with a
`maximum resistance to the procedure to be examined
`are used for the purposes of auditing and qualifying
`sterilization procedures. Consequently, the inactivation
`of such highly resistant microorganisms encompasses
`all less resistant organisms, including most pathogens.
`Furthermore, the test organisms should be cultivable
`under simple and most easily reproducible conditions. In
`general, innocuous bacterial endospores fulfill these re-
`quirements [41], [50], [31], [4]. In order to verify the ef-
`ficacy of sterilization procedures, and to determine
`treatment parameters, extremely large volumes of these
`test organisms are used, typically ≥106 bacterial spores
`per test object, e.g., in the form of bio-indicators [18].
`When using the “Half-cycle method”, the action of half of
`the intended sterilization cycle – usually half of the
`treatment period – is examined. If, following this action,
`a certain number of bio-indicators contaminated with 106
`resistant bacterial spores are inactivated, it can be con-
`cluded that when applying the full cycle, an SAL of 10–6
`is guaranteed at a theoretical spore-inactivation rate of
`≥12 lg increments. This corresponds to “Overkill condi-
`tions” [37], [41], [30], [26].
`However, in a course of action of this kind, which is based
`on the complete inactivation of a limited number of test
`objects, it must be taken into account that there is a
`statistical connection between the mean number of
`contaminated test objects after treatment and the total
`number of identically treated test objects. Accordingly,
`the more test objects are included in the test, the greater
`the duration of treatment, concentration or dosage of
`antimicrobially active agents must be to completely inac-
`tivate a limited number of bio-indicators [66], [64], [87],
`[65].
`In contrast, other directions and recommendations – also
`for heat sterilization procedures – only demand evidence
`that the difference between the initial number of test
`microorganisms and the number of test organisms at the
`end of the process amount to more than six orders of
`
`magnitude, i.e., an inactivation rate ≥6 lg increments in
`order to consider the SAL of 10–6 as attained and indicate
`the product treated in this way as sterile [14], [3].
`What all methods used to prove the efficacy of steriliza-
`tion procedures have in common is that the conditions
`necessary to attain an extremely low probability of con-
`tamination of 10–6 are inferred from the treatment condi-
`tions required to reduce extremely high artificial test
`contaminations. This procedural method is based on the
`general assumption of exponential inactivation kinetics
`for microorganisms under the influence of antimicrobially
`effective parameters, from which a linear mortality curve
`results, given a semi-logarithmic diagram.
`There have been detailed trials and discussions, in par-
`ticular concerning heat inactivation kinetics. As early as
`1921, Bigelow put forward logarithmic inactivation kinet-
`ics for microorganisms under the influence of heat. The
`monograph by Konrich and Stutz [35], which was con-
`sidered to be a standard work for many years, as well as
`works by Machmerth [39], Pflug and Holcomb [48], Rus-
`sell [51], Gould [23] and Knöller [34] provide detailed
`analyses and discussions of the relevant literature, includ-
`ing experiments on this issue. The inactivation kinetics
`for microorganisms initially postulated for heat treat-
`ments, which corresponded to first-order reaction kinetics,
`was also applied in principle to the conditions of the ac-
`tion of ionizing irradiation as well as chemical agents with
`antimicrobial effect [28], [83], [63], [42], [53].
`Divergence from a strict semi-logarithmic course of the
`mortality curve is usually explained by inhomogeneity of
`the test organism populations used. The actual complexity
`of microbiological inactivation kinetics cannot be clearly
`and comprehensively described using simple mathemat-
`ical models, so that there is no uniform theory which takes
`into account all possible courses of mortality curves for
`the inactivation of microorganisms under the influence
`of noxa with antimicrobial effect [83]. Consequently, while
`in many cases the assumption of the linearity of inactiva-
`tion kinetics in the semi-logarithmic standard simplifies
`the actual circumstances, it offers the only practicable
`possibility for interpretation and utilization of data gained
`through experiments [39], [48], [37], [81], [34]. The cur-
`rent European Pharmacopoeia also puts forward the view
`that “the inactivation of microorganisms by physical or
`chemical means follows an exponential law...” [18].
`The fact that the objective of sterilization is to ensure a
`rate of microbial contamination of ≥10–6 surviving microor-
`ganisms per test object results in the necessity of extra-
`polating mortality curves from the area that can be recor-
`ded in experiments to determine SAL-compatible steriliz-
`ation parameters. As the description of the inactivation
`kinetics by means of a simple mathematical model, such
`as first-order kinetics, already shows an approximation
`within the area ascertainable in experiments, the associ-
`ated uncertainties must considerably increase in the case
`of extrapolation outside this area [83]. Nevertheless,
`procedural parameters intended to guarantee the attain-
`ment of sterility that meets the SAL are usually defined
`on the basis of extrapolation of this kind [41], [48], [4].
`
`GMS Krankenhaushygiene Interdisziplinär 2008, Vol. 3(3), ISSN 1863-5245
`
`3/10
`
`I
`
`Novartis Exhibit 2200.003
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`von Woedtke et al.: The limits of sterility assurance
`
`Thus, both the determination of procedural parameters
`for sterilization processes and the proof of a SAL ≤10–6
`as the quantitative end-point which has to be guaranteed
`by a sterilization process are not based on scientifically
`proven data, but are only rules of thumb and approximate
`values [34].
`Conclusion: Using the SAL concept as a basis for the
`evaluation of the performance of sterilization procedures
`constitutes a situation (which is probably unique) in which
`the certain attainment of a condition is required by law,
`but there is no way of proving the attainment of this
`condition in practical terms [22], [25], [24], [2]).
`
`Concept for proving sterilization
`assurance for heat-sensitive
`medical devices
`In our own tests to examine new concepts for the gentle
`sterilization of
`irritable goods, applying non-thermal
`physical and/or chemical treatment procedures and using
`Bacillus subtilis spores as test organisms, test conditions
`were selected that, after treatment, either viable test or-
`ganisms were still in evidence on the test objects or both
`sterile and unsterile test objects were present simultan-
`eously. By combining direct cell counting methods using
`classical microorganism recovery and counting techniques
`[65], [48], [1] with the “fraction negative method” [64],
`[65], [48], [62], mortality curves on the basis of experi-
`mental data were maintained in a range of about 8 orders
`of magnitude (≥106 to around 10–2 test organisms per
`test object). With reference to the known initial contam-
`ination of the test objects, exact reduction factors could
`be stated for the treatment parameters applied in each
`case, to derive from these the treatment conditions ne-
`cessary to attain sterility assurance in accordance with
`the SAL [73], [74], [75].
`Under this condition, the attainable degree of reduction
`of the test organisms can be exactly quantified on the
`one hand, and the actual inactivation kinetics can be
`depicted on the other, at least in the range that can be
`recorded using microbiological methods of proof. Addi-
`tionally, possible inhomogenities of the mortality curves
`can be taken into account for extrapolation into the SAL
`area.
`Conclusion: Exact quantitative statements as to the inac-
`tivation kinetics and, accordingly, the antimicrobial effic-
`acy can be made only if not all test objects are fully inac-
`tivated, starting from a known initial contamination follow-
`ing sub-effective treatment. Thus, for the quantitative
`characterization of the efficacy of antimicrobial proced-
`ures, the experimental conditions are to be chosen so
`that in the result of treatment, surviving test organisms
`are detectable for various individual treatments.
`Various experiments by the current authors demonstrated
`that inactivation kinetics within the experimentally access-
`ible range between 106 and 10–2 bacterial spores per test
`object never exhibited a linear course [33], [77], [76],
`
`[74], [75]. In part, there were very pronounced concave
`curves, consisting of an initial steep section which then
`levelled off. At least with the non-thermal antimicrobially
`active treatments examined, it was apparently possible
`to reduce the high starting incidence in the test objects
`to a low level using relatively short action duration, low
`substance concentrations or low irradiation. However,
`the efficacy against the residual contamination was con-
`siderably lower. Consequently, an extrapolation of the
`steep section of the respective inactivation curve into the
`SAL range of 10–6 would result in treatment durations,
`irradiation doses or substance concentrations which are
`actually much too brief to guarantee an adequate sterility
`assurance. On the other hand, an extrapolation of the
`second, flat part of the mortality curve would result in all
`cases in extreme sterilization conditions, which could not
`be applied practically.
`Analogous mortality curves can be found again and again
`primarily in older publications. Seidl et al., for example,
`reported on experiments on radio sterilization of medicinal
`products, in which it was possible to destroy 99.9% of a
`test organism population with an irradiation dose of
`0,1 Mrad (1 kGy); for the remaining 0.1%, however, a
`dosage at least five times stronger was necessary [57].
`Pronounced concave mortality curves for Bacillus subtilis
`spores dependent on varying gamma irradiation doses
`can also be found in Wallhäußer [80]. Pfeiffer also reports
`non-linear inactivation kinetics when applying ionizing ir-
`radiation, because of disproportionate survival of radi-
`ation-resistant microorganisms in the range of higher ir-
`radiation doses [45], [47]. Furthermore, various works
`on sporocidal efficacy of hydrogen peroxide show con-
`cave, flattening-out mortality curves [72], [11], [5]. Van
`Ooteghem describes non-linear survival curves for microor-
`ganisms under the influence of preservatives [71].
`Such inactivation curves, also called “tailing” curves, are
`explained predominantly (e.g., by Hermann [28] as well
`as Wickramanayake and Sproul [83]) by the existence of
`microorganism populations on the test object with incon-
`sistent resistance to the antimicrobially active treatments
`examined. Consequently, the less resistant fraction is
`killed first (steep curve section); the predominance of the
`surviving, more resistant fraction then results in a flatter
`course of the mortality curve. Spicher explains this phe-
`nomenon in detail, coming to the conclusion that the
`determination of the parameters of a sterilization proced-
`ure must be oriented to such extreme values represented
`by highly resistant test organism fractions, since that re-
`flects the actual circumstances [67].
`The range between experimentally detectable contamin-
`ation rates up to ca. 10–2 test organisms per test object
`and the SAL of 10–6, still encompasses 4 lg increments,
`which cannot be proven by experimental data. Due to the
`inhomogeneity of the inactivation kinetics already present
`in the experimentally accessible range, it is not possible
`to make any certain statement regarding the continuing
`course of such mortality curves, which may increasingly
`flatten out. Spicher already in 1993 expressly pointed
`out that, in unfavorable cases, the highly resistant test
`
`GMS Krankenhaushygiene Interdisziplinär 2008, Vol. 3(3), ISSN 1863-5245
`
`4/10
`
`I
`
`Novartis Exhibit 2200.004
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`von Woedtke et al.: The limits of sterility assurance
`
`organism fractions, and thus the flat part of the mortality
`curve, may no longer be determined using the usual mi-
`crobiological testing procedures. As a result, the custom-
`ary extrapolation of such curves into the SAL range in-
`volves considerable risks [67].
`Thus, it is clear that the general assumption of first-order
`exponential inactivation kinetics, which was used origin-
`ally to describe heat inactivation of bacteria, cannot be
`applied without restrictions to non-thermal processes.
`One approach for the interpretation of these experimental
`results is the hypothesis that there is a basic difference
`between thermal and non-thermal antimicrobial modes
`of action, from which follows the assumption that such
`non-linear mortality curves are not attributable solely to
`the inhomogeneity of the test populations. Possibly, they
`display a characteristic that can be generalized for all
`inactivation procedures based on non-thermal actions
`[77], [76]. From those treatment parameters that effect
`elimination of the conventionally high test organism
`numbers present on bioindicators for sterilization control,
`it is therefore not permissible – for sterilization proced-
`ures in which efficacy relies on irradiation and/or chem-
`ical effects – to directly derive treatment conditions which
`are intended to guarantee a reduction into the experiment-
`ally no longer detectable but for sterilization assurance
`essential range. Consequently, treatment conditions that
`ensure an SAL-compliant reduction up to a contamination
`rate of at least 10–6 cannot be clearly determined for such
`procedures. The conclusion to be drawn from this is that
`when non-thermal treatment procedures are applied, the
`condition of sterility that would conform with a contamin-
`ation probability of 10–6 cannot in fact be guaranteed.
`Thus, the SAL concept is not a procedure suitable for
`showing the efficacy of non-thermal sterilization pro-
`cesses.
`Consequently, only steam sterilization and sterilization
`using dry heat should be described as sterilization pro-
`cedures in the proper or traditional sense.
`Sterility according to a SAL of 10–6 should, logically, only
`still be required for medical devices and preparations
`that can be subjected to steam or hot air sterilization
`using the required standard and equivalent procedures.
`This is because it is possible that a homogeneous linear
`mortality curve, and thus the sufficiently certain determ-
`ination of the treatment conditions necessary to guaran-
`tee an SAL of 10–6, can be presumed only in thermal
`procedures.
`Other authors who carried out a detailed mathematical
`analysis of the intrinsic uncertainties of the exponential
`model of mortality of test organisms reach the same
`conclusions in principle (while also taking into account
`thermal inactivation kinetics). They explain that for dec-
`ades, this evidently inadequate theoretical basis has been
`adhered to without question, using the argument that,
`due to extreme safety premiums, the safety of sterilized
`products in practice is secured by assuming higher con-
`tamination rates with extremely resistant test organisms
`when examining sterilization procedures [8], [9].
`
`In order to ensure the highest possible level of safety in
`the application of non-thermal procedures, a proof of
`“Antimicrobial efficacy on the highest experimentally ac-
`cessible level” should be required.
`This proof should show that the inactivation kinetics de-
`pendent on the number of test organisms in the entire
`range ascertainable in experiments can be evidenced
`with performance data. As a rule, the performance char-
`acterization for non-thermal antimicrobial procedures
`should be carried out using test bodies contaminated
`with low levels of highly resistant test organisms. This is
`in order to reflect the fact that, apparently in contrast to
`the relatively simple option of reducing high numbers,
`the inactivation of low levels of residual contamination
`is disproportionately more difficult to achieve. Test mi-
`croorganisms with a high level of resistance to the pro-
`cedure to be examined should be used as test organisms,
`e.g., bacterial spores. It must be proven that a reduction
`of the number of test organisms by at least five lg incre-
`ments up to a contamination rate of 10–2, which can only
`just be proven in experiments, has been achieved. In
`general, only the strict concentration on data that can be
`recorded in experiments affords the possibility of being
`able to directly compare various procedures and proced-
`ural steps using inactivation kinetics and, consequently,
`to make available differing, but equivalent inactivation
`procedures for various products. The extrapolation of
`such inactivation kinetics recorded in experiments by
`merely one additional lg increment to a contamination
`level of 10–3 would guarantee a sufficient "Safety premi-
`um” for the determination of the necessary treatment
`parameters. Here, a “tiered” SAL of 10–3 could be intro-
`duced for non-thermal sterilization procedures. This value
`is also referred to repeatedly in the literature on this
`subject [22], [24], [86].
`In order to differentiate it from actual sterilization with
`an SAL of 10–6, which should be restricted to thermal
`procedures, such a gentle sterilization procedure targeted
`at a contamination probability of 10–3 could be called
`”Low-level sterilization”. The efficacy of ”Low-level steril-
`ization” concentrates primarily on the range of low levels
`of residual contamination with highly resistant microor-
`ganisms on goods to be sterilized, following effective
`preparation (aseptic processing and/or cleaning and
`subsequent disinfection), which is very important in
`practice.
`
`Do the current theoretical sterility
`assurance requirements reflect the
`actual safety requirements?
`It was ascertained that the practical proof of the required
`level of sterility assurance of 10–6
`is not possible.
`Moreover, the attainability of this condition is fundament-
`ally questionable, at least in non-thermal procedures.
`Furthermore, it is questionable whether the undifferenti-
`ated adherence to the currently practiced concept of
`
`GMS Krankenhaushygiene Interdisziplinär 2008, Vol. 3(3), ISSN 1863-5245
`
`5/10
`
`I
`
`Novartis Exhibit 2200.005
`Regeneron v. Novartis, IPR2021-00816
`
`

`

`von Woedtke et al.: The limits of sterility assurance
`
`sterility assurance on the basis of an SAL of 10–6 complies
`with the actual safety requirements in terms of patient
`or user safety [2]. The practical relevance of an SAL of
`10–6 is not conclusively proven, i.e., it is practically im-
`possible to find a difference in the rate of infection result-
`ing from the application of products that were sterilized
`with a (theoretical) SAL of 10–6 on the one hand, or on
`the other hand were treated using a procedure in which
`only a contamination likelihood of 10–3 is ensured [85].
`By extrapolating the reduction rates following extreme
`artificial initial contamination to an SAL of 10–6, a theor-
`etical overall performance of the procedure can be estab-
`lished by at least 12 lg increments (overkill conditions).
`Such an SAL, however, is neither evidence based nor can
`it be accurately ascertained in experiments, or attained
`practically.
`Furthermore, it must be taken into consideration that, by
`adhering to an SAL of 10–6 as a “gold standard” for steril-
`ization, unnecessarily high costs are incurred, the intro-
`duction of new sterilization procedures is made consider-
`ably more difficult [2], and no allowance is made for the
`requirement of sustainable development.
`Therefore, for practical reasons, a concept of tiered SALs
`would make sense, the establishment of which is oriented
`towards the intended application of the sterilized goods,
`the characteristics of such goods, and the corresponding
`treatment options. In disinfection procedures, an analog-
`ous categorization of “High-level disinfection”, “Interme-
`diate-level disinfection”, and “Low-level disinfection”, has
`been customary for some time already [69], [10], [52],
`[55], [60].
`Since the mid-1990s, there have been repeated sugges-
`tions for the definition of tiered sterility assurance values,
`such as a “Concept of tiered SALs”, the suggestion of the
`declaration of “Asepsis assurance values” [22], [86], or
`a declaration as “Aseptic: Safe for its designated use”,
`which is oriented more towards the intended application
`of the product [24], [2].
`The determination of the respective safety level should
`be oriented towards both the features and the quality of
`the products (e.g., new products being used for the first
`time and single-use products, as opposed to re-usable
`and re-processed goods etc.), as well as the intended
`application.
`In practice, in the course of modern quality-controlled
`production and preparation procedures for pharmaceut-
`icals and medical devices, only a very low level of contam-
`ination is to be expected. In the case of aseptic prepara-
`tion, filling and production procedures, a contamination
`probability of 10–3 is assumed; lower contamination rates
`can also be realized in automated processes [18], [45],
`[41], [85], [81], [17], [24], [13], [44]. For the use of
`gamma radiation for sterilization, possibilities of determ-
`ining dosage by ascertaining the number, type and radi-
`ation resistance of microbiological contaminants on the
`goods to be sterilized have been discussed and applied
`in practice for several years [45], [49], [27], [47].
`The possible inclusion of the real contamination risk for
`the products to be sterilized in the course of the quality
`
`control of production processes could become general
`practice to determine the necessary sterilization treat-
`ment conditions and the resulting level of safety. This
`would also correspond to the suggested concentration
`of the proof of efficacy of a sterilization procedure on the
`low level of residual contamination with more highly res-
`istant microorganisms.
`In addition to the degree of contamination prior to steril-
`ization, the period between sterilization and intended use
`as well as the risk of recontamination and/or proliferation
`of microorganisms during the storage period must also
`be taken into account. Thus, e.g., for aqueous parenterals,
`it is by all means reasonable to demand an SAL of 10–6.
`In the case of the production of re-usable medical devices,
`a reduction of at least 2 lg increments can be achieved
`through prior cleaning in validated cleaning and disinfect-
`ing devices (CDD). In CDDs for containers for human
`egesta, the reduction rate was 3 to >5 lg increments,
`depending on the procedure [82]. Due to the require-
`ments for chemical disinfection, a further reduction by
`≥5 lg increments is achieved through disinfection follow-
`ing cleansing in the reprocessing procedure [21]. This
`means that because of the preceding reprocessing pro-
`cedure for medical devices, at least 7 lg increments are
`usually added to the actual sterilization performance.
`Based on a consistent consideration of the reduction
`performance of all partial steps of reprocessing and
`sterilization procedures, it would be possible to markedly
`reduce the final sterilization treatment without a reduction
`in safety. The prerequisite for this is the validation of all
`partial process steps, which is required in any case.
`In the case of sterilized operating instruments, an addi-
`tional factor must be taken into account in the risk assess-
`ment. In conventionally ventilated operating theaters,
`sterile goods are removed from their sterile packaging in
`the theater and are stored opened on the instrument
`table for the duration of the operation. During this time,
`they are exposed to the risk of contamination from sedi-
`mentation of airborne or particle-borne pathogens, re-
`leased mainly by the operating team. Even in the case of
`low-turbulence displacement flow (LDF) and subject to
`the premise that the sterile goods are opened within the
`LDF and the table of instruments is also situated com-
`pletely within the LDF, according to DIN 1946-2 and -4,
`up to 10 CFU/m3 (CFU – colony forming units) may be
`contained in the airflow following the filter emissions of
`the ventilation equipment (around 10 cm apart) in an
`empty, unused, and previously cleaned and disinfected
`operating theater [15], [16]. The pathogens released by
`the operating team, which cannot be fully removed by
`the LDF, are added to this. Thus, the ov

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket