throbber
Performance Comparison of Nebulizer Designs:
`Constant-Output, Breath-Enhanced, and Dosimetric
`
`Joseph L Rau PhD RRT FAARC, Arzu Ari MSc CRT CPFT, and Ruben D Restrepo MD RRT
`
`INTRODUCTION: Design differences among pneumatically powered, small-volume nebulizers affect drug
`disposition (percentage of the dose delivered to the patient, lost to deposition in the equipment, and lost via
`exhalation to ambient air) and thus affect drug availability and efficacy. OBJECTIVE: Evaluate in vitro the
`dose disposition with 5 nebulizer models, of 3 types (constant-output, breath-enhanced, and dosimetric), using
`simulated normal, adult breathing. METHODS: We compared 5 nebulizer models: 2 constant-output (Misty-
`Neb and SideStream), 1 breath-enhanced (Pari LCD), and 2 dosimetric (Circulaire and AeroEclipse). Each
`nebulizer was filled with a 3-mL unit-dose of albuterol sulfate and powered by oxygen at 8 L/min. The
`nebulizers were connected to an induction throat, connected to a breathing simulator. We measured (1)
`inhaled drug (subdivided into mass deposited in the induction throat and mass deposited in the filter at the
`distal end of the induction throat), (2) exhaled drug (lost to the ambient air), (3) drug lost to deposition in the
`apparatus, and (4) drug left in the unit-dose bottle. The duration of nebulization (until sputter) was measured
`with a stopwatch. All drug amounts were analyzed via spectrophotometry and expressed as a percentage of
`the total dose. RESULTS: The mean ⴞ SD inhaled drug percentages were: Misty-Neb 17.2 ⴞ 0.4%, Side-
`Stream 15.8 ⴞ 2.8%, Pari LCD 15.2 ⴞ 4.2%, Circulaire 8.7 ⴞ 1.0%, and AeroEclipse 38.7 ⴞ 1.3%. The
`mean ⴞ SD percentages of drug lost to the ambient air were: Misty-Neb 26.8 ⴞ 0.7%, SideStream 17.3 ⴞ
`0.4%, Pari LCD 18.3 ⴞ 0.8%, Circulaire 12.3 ⴞ 0.8%, and AeroEclipse 6.6 ⴞ 3.3%. The mean ⴞ SD
`percentages of drug lost to deposition in the apparatus were: Misty-Neb 52.3 ⴞ 0.6%, SideStream 63.4 ⴞ
`3.0%, Pari LCD 62.5 ⴞ 4.0%, Circulaire 75.8 ⴞ 0.5%, and AeroEclipse 51.0 ⴞ 2.1%. Duration of nebuli-
`zation was shortest with the Circulaire and longest with the AeroEclipse (p < 0.05 via 1-way analysis of
`variance). CONCLUSIONS: The nebulizers we tested differ significantly in overall drug disposition. The
`dosimetric AeroEclipse provided the largest inhaled drug mass and the lowest loss to ambient air, with the
`test conditions we used. Key words: nebulizers; aerosols, drug therapy; drug administration, inhalation; respi-
`ratory drug administration. [Respir Care 2004;49(2):174–179. © 2004 Daedalus Enterprises]
`
`Introduction
`
`Gas-powered jet nebulizers are commonly used for de-
`livering medications in the clinical and home-care settings.
`
`Joseph L Rau PhD RRT FAARC, Arzu Ari MSc CRT CPFT, and Ruben
`D Restrepo MD RRT are affiliated with Cardiopulmonary Care Sciences,
`Georgia State University, Atlanta, Georgia.
`
`Arzu Ari MSc CRT CPFT presented a version of this report at the
`American Association for Respiratory Care OPEN FORUM at the 47th
`International Respiratory Congress, held December 1–4, 2001, in San
`Antonio, Texas.
`
`Correspondence: Joseph L Rau PhD RRT FAARC, Cardiopulmonary
`Care Sciences, MSC 8R0319, Georgia State University, 33 Gilmer Street
`SE, Unit 8, Atlanta GA 30303. E-mail: jrau@gsu.edu.
`
`Over the past few years nebulizer design changes have
`created nebulizer categories, termed constant-output,
`breath-enhanced, and dosimetric.1 Constant-output nebu-
`lizers are the traditional T-piece nebulizers that generate
`aerosol constantly, during the inhalation, exhalation, and
`breath-hold. With constant-output nebulizers some of the
`aerosol is lost during exhalation, which causes release of
`aerosol to the ambient air through the expiratory limb of
`the T-piece.2– 4 Constant-output nebulizers have been crit-
`icized as unreliable and inefficient, because a low percent-
`age of the dose reaches the patient.5–7 A length of large-
`bore tubing is usually attached to the expiratory side of the
`constant-output nebulizer T-piece, to reduce drug loss and
`increase the inhaled amount.8 –9
`Breath-enhanced nebulizers are designed to allow re-
`lease of more aerosol during inhalation, when ambient air
`
`174
`
`RESPIRATORY CARE • FEBRUARY 2004 VOL 49 NO 2
`
`IPR2021-00406
`United Therapeutics EX2077
`
`

`

`PERFORMANCE COMPARISON OF NEBULIZER DESIGNS
`
`is drawn through the nebulizer.9 During exhalation, gas
`flow through the nebulizer falls back to the power-gas
`flow only, exhaled gas is routed out the expiratory valve in
`the mouthpiece, and aerosol is contained in the nebulizer
`chamber. Examples of breath-enhanced nebulizers include
`Pari LC Plus and Pari LCD. Coates et al10 and Dennis11
`found better drug output with breath-enhanced nebulizers.
`Dosimetric nebulizers release aerosol only during inha-
`lation.1 The Circulaire represented an early attempt to con-
`vert a constant-output nebulizer to a dosimetric device, by
`attachment of a storage bag with a 1-way valve in the
`mouthpiece connector.12,13 A recently introduced nebu-
`lizer, the AeroEclipse, has a breath-actuated valve that
`triggers aerosol generation only during inhalation, elimi-
`nating the need for a storage bag or reservoir.
`Theoretically, both breath-enhanced and dosimetric
`nebulizers would have reduced or no aerosol loss during
`exhalation. However, although there may be reduced ex-
`halation loss of aerosol, does the emitted drug amount in
`fact increase, or is there a shift in the location of lost
`aerosol, from exhaled/ambient to device? With the Circu-
`laire does the storage of aerosol increase or decrease the
`inhaled drug? We found no studies of all 3 categories of
`nebulizer, including the Circulaire, using the same set of
`realistic breathing conditions14 and that characterized the
`total drug disposition, including emitted drug, device loss,
`and exhaled/ambient drug loss. The purpose of the present
`study was to evaluate in vitro the total drug disposition of
`constant-output, breath-enhanced, and dosimetric nebuliz-
`ers, using simulated normal adult breathing.
`
`Methods
`
`Study Design
`
`The nebulizer brands tested were AirLife Misty-Neb
`(Allegiance Healthcare, McGaw Park, Illinois), AirLife
`SideStream (Allegiance Healthcare, McGaw Park, Illinois),
`Circulaire (Westmed, Tucson, Arizona), Pari LCD (PARI
`Respiratory Equipment, Monterey, California), and Aero-
`Eclipse (Monaghan Medical, Plattsburgh, New York). The
`Misty-Neb and SideStream are traditional constant-output
`nebulizers. The Pari LCD is a breath-enhanced nebulizer.
`The Circulaire and AeroEclipse were considered dosimet-
`ric devices, based on Dennis’s definition.1 The Circulaire
`was tested with the supplied nebulizer. Figure 1 shows the
`principle of operation of each nebulizer brand tested.
`Three of each of the 5 nebulizer brands were tested,
`using a simulated normal adult breathing pattern. Each
`device nebulized a unit-dose of albuterol sulfate solution,
`2.5-mg base equivalent (Proventil, Schering, Kenilworth,
`New Jersey), with a 3 mL total fill volume. No additional
`diluent was added to any nebulizer. All the nebulizers
`were powered by 50-psi oxygen at 8 L/min.
`
`Lung Model
`
`The nebulizers were connected to a breathing simulator
`(Series 1101, Hans Rudolph, Kansas City, Missouri), which
`provides a complete breathing cycle with both inhalation and
`exhalation phases. Tidal volume was set at 600 mL, inhala-
`tion flow at 30 L/min, and respiratory rate at 12 breaths/min,
`giving a 1:3 inspiratory-expiratory ratio. Figure 2 shows the
`equipment configuration. An induction port (throat) (Thermo
`Andersen, Franklin, Massachusetts), as described in the United
`States Pharmacopeia (USP) for use with cascade impactor
`testing,15 was placed between the nebulizer outlet and the
`breathing simulator. The throat, which has a diameter of ap-
`proximately 19 mm and a right angle, was used as a simple
`geometric analogue of the upper airway, to allow inertial
`impaction of larger aerosol particles. This allowed a stan-
`dardized basis for comparison of inhaled aerosol from each
`nebulizer tested. A filter (2-way nonconductive anesthesia
`filter, Baxter Healthcare, Deerfield, Illinois) was attached to
`the distal end of the throat, between the throat and the breath-
`ing simulator. We defined the total inhaled drug mass as the
`amount in the throat plus the filter.
`In cascade impactor testing the throat is placed verti-
`cally, but in our experiments the throat was placed hori-
`zontally to prevent the inhalation filter from collecting
`drug that might condense on the throat wall and then drip
`onto the inhalation filter. The mouthpieces were removed
`from the 2 constant-output nebulizers (Misty-Neb and Side-
`Stream) and the T-piece was connected directly to the
`throat and the breathing simulator. A 15-cm length of
`large-bore corrugated tubing, as supplied by the nebulizer
`manufacturer, was attached to the exhalation outlet of the
`T-piece, and exhaled drug was collected by a filter at the
`end of the tubing (see Fig. 2).
`The breath-enhanced Pari LCD, which is a disposable unit,
`has a nonvalved opening at the top of the chamber and open
`exhalation ports in the mouthpiece (see Fig. 1). With the Pari
`LCD we placed exhalation filters at the chamber top and at the
`outlet of a T-piece, which replaced the mouthpiece (see Fig. 2).
`The Circulaire contains a 1-way valve in the nebulizer T-piece,
`which directs aerosol toward the mouth. There is also an exha-
`lation port with a size-adjustable opening between the 1-way
`inspiratory valve and the mouthpiece (see Fig. 1). An exhalation
`filter was attached to the exhalation port, with maximum open-
`ing; the T-piece assembly, without the mouthpiece, was attached
`to the throat and breathing simulator (see Fig. 2). The Aero-
`Eclipse has a 1-way exhalation flapper valve integrated into the
`mouthpiece assembly. An inhalation flow of approximately 6
`L/min causes the spring-loaded valve to engage and generate
`aerosol (see Fig. 1). The mouthpiece was replaced with a T-
`piece, and an outwardly directed 1-way valve was added to the
`exhalation outlet of the T-piece so that inhalation flow came
`from the nebulizer and allowed breath-actuated triggering of the
`nebulizer (see Fig. 2).
`
`RESPIRATORY CARE • FEBRUARY 2004 VOL 49 NO 2
`
`175
`
`

`

`PERFORMANCE COMPARISON OF NEBULIZER DESIGNS
`
`Fig. 1. Functional diagrams of the 5 nebulizer brands tested, illustrating principle of operation and patterns of gas flow during inhalation and
`exhalation.
`
`Measurement of Drug
`
`In each nebulizer trial the total aerosol drug mass was
`measured and consisted of the total inhaled drug mass,
`exhaled/ambient drug loss, and drug lost in the device.
`We also measured drug remaining in the unit-dose bot-
`tle. The total inhaled drug mass was divided into drug
`collected in the throat and drug from the collecting filter
`attached to the throat outlet. Exhaled drug was collected
`on a filter attached to the exhalation outlet of the neb-
`ulizer. The drug remaining in the nebulizer apparatus
`(including adapters, T-piece, and mouthpiece) was col-
`lected by washing, and analyzed. Each nebulizer was
`weighed empty, after filling, and at the end of nebuli-
`zation, to calculate the volume left, as described by
`Coates et al.10 Solvent was added to the calculated vol-
`ume, drug concentration was then determined by spec-
`trophotometry, and the drug mass was calculated. Each
`nebulizer was operated until the onset of sputter, with
`no tapping of the nebulizer (as is usually done when
`
`administering aerosol to a patient), and the time to sput-
`ter was recorded with a stopwatch. All drug amounts
`were analyzed via spectrophotometry (Beckman Instru-
`ments, Fullerton, California), at a wavelength of 276
`nm. The solvent was 0.1 molar normal hydrochloric
`acid (JT Baker Company, Phillipsburg, New Jersey).
`Collecting filters were washed for 1 min with gentle
`agitation. Longer washing did not yield additional drug.
`Measurements with 2 filters in series verified that no
`drug was lost through the first filter. The spectropho-
`tometer was calibrated prior to trials, using a holmium
`oxide filter (Beckman Instruments, Fullerton, Califor-
`nia) to determine wavelength accuracy, and set to zero
`using the solvent alone before each analysis. A regres-
`sion curve and prediction equation were developed from
`serial dilutions of known albuterol sulfate solution
`(Sigma, St Louis, Missouri). Concentrations of sample
`solutions, and thereby drug amounts of albuterol, were
`calculated from this known concentration/absorbance
`relationship.
`
`176
`
`RESPIRATORY CARE • FEBRUARY 2004 VOL 49 NO 2
`
`

`

`PERFORMANCE COMPARISON OF NEBULIZER DESIGNS
`
`Fig. 2. Setup for nebulizer performance testing. Right Panel: Configuration for the constant-output nebulizers (Misty-Neb and SideStream).
`Left Panels: Configurations for the Pari LCD, Circulaire, and AeroEclipse. NEB ⫽ nebulizer. Tee ⫽ T-piece.
`
`Data Analysis
`
`Means and standard deviations were calculated for each
`component of the total drug mass and for time of nebuli-
`zation. Differences between the total inhaled mass and the
`inhalation-filter mass (ie, total inhaled drug minus drug
`deposited in the throat) were compared with 1-way anal-
`ysis of variance. Differences were considered statistically
`significant when p ⬍ 0.05. Multiple follow-up compari-
`sons to identify differences among nebulizers were per-
`formed using Scheffe´’s S method.16 All statistical calcu-
`lations were performed using commercially available
`software (SYSTAT 7.0, SPSS, Chicago, Illinois).
`
`Results
`
`Table 1 shows the dose disposition results, expressed as
`percentages of total drug recovered from the throat, inha-
`lation filter, exhalation filter, nebulizer apparatus, and drug
`remaining in the unit-dose bottle. The mean ⫾ SD total
`drug mass of albuterol sulfate recovered from all sources
`and expressed as the base was 2.56 ⫾ 0.09 mg, which
`corresponds well to the 2.5-mg nominal dose of the unit-
`dose albuterol nebulizer solution we used.
`
`The percentage of total inhaled drug mass differed sig-
`nificantly (by 1-way analysis of variance) among the 5
`nebulizer brands tested (p ⫽ 0.0001). Table 1 shows which
`groups of individual brands did not significantly differ
`from each other (homogeneous subsets), based on fol-
`low-up comparisons. The total inhaled drug obtained from
`the constant-output nebulizers, Misty-Neb and SideStream,
`was similar to that from the breath-enhanced Pari LCD,
`ranging from 15% to 17%. The 2 nebulizers considered
`dosimetric (Circulaire and AeroEclipse) differed from each
`other, and the Circulaire differed from the Misty-Neb. The
`inhaled drug mass from the Circulaire was approximately
`half that of the constant-output and breath-enhanced nebu-
`lizers, whereas the inhaled mass of the AeroEclipse was
`about 2.5 times greater than the constant-output and breath-
`enhanced nebulizers.
`The inhalation filter mass differed significantly (by
`1-way analysis of variance) among the 5 nebulizer brands
`(p ⫽ 0.0001). Table 1 shows groups of brands that did not
`significantly differ from each other, based on follow-up
`comparisons. The dosimetric nebulizers (Circulaire and
`AeroEclipse) had the least exhaled drug loss: approxi-
`mately 7–12%. The duration of nebulization also differed
`significantly among the nebulizers (p ⫽ 0.0001).
`
`RESPIRATORY CARE • FEBRUARY 2004 VOL 49 NO 2
`
`177
`
`

`

`PERFORMANCE COMPARISON OF NEBULIZER DESIGNS
`
`Table 1. Aerosol Deposition and Loss, and Nebulization Time With 5 Nebulizer Brands*
`
`Total inhaled (%)
`Inhalation filter (%)
`Exhaled to ambient (%)
`Deposited in nebulizer apparatus (%)
`Remained in unit-dose bottle (%)
`Nebulization time (min)
`
`Misty-Neb
`
`17.2 ⫾ 0.4†
`14.4 ⫾ 0.5†§
`26.8 ⫾ 0.7
`52.3 ⫾ 0.6
`3.7 ⫾ 1.1
`11.9 ⫾ 0.3
`
`SideStream
`
`15.8 ⫾ 2.8†‡
`14.7 ⫾ 2.7†
`17.3 ⫾ 0.4
`63.4 ⫾ 3.0
`3.6 ⫾ 0.5
`9.5 ⫾ 0.1
`
`Pari LCD
`
`15.2 ⫾ 4.2†‡
`13.3 ⫾ 4.2†§
`18.3 ⫾ 0.8
`62.5 ⫾ 4.0
`4.1 ⫾ 0.6
`8.4 ⫾ 1.2
`
`Circulaire
`
`8.7 ⫾ 1.0‡
`7.4 ⫾ 1.0§
`12.3 ⫾ 0.8
`75.8 ⫾ 0.5
`3.0 ⫾ 0.4
`7.0 ⫾ 0.5
`
`AeroEclipse
`
`38.7 ⫾ 1.3
`34.2 ⫾ 1.3
`6.6 ⫾ 3.1
`51.0 ⫾ 2.1
`3.7 ⫾ 0.6
`14.4 ⫾ 1.1
`
`*The percent values represent percent of total dose. The inhalation filter percentage is a subset of the total inhaled percentage: specifically, the inhalation filter percentage equals the total inhaled
`minus the amount deposited in the throat. Subsets of nebulizer brands with no significant differences (p ⬍ 0.05) based on follow-up comparisons are indicated for total inhaled percentage and
`inhalation filter percentage. Nebulization time was until sputter.
`†No significant difference between Misty-Neb, SideStream, and Pari LCD
`‡No significant difference between SideStream, Pari LCD, and Circulaire
`§No significant difference between Misty-Neb, Pari LCD, and Circulaire
`
`Discussion
`
`The focus of the present study was to measure drug
`disposition with different nebulizer design categories tested
`under the same conditions. We found differences among
`the nebulizers in total inhaled mass of aerosolized bron-
`chodilator and in the location and amounts of aerosol drug
`lost to ambient air and to the nebulizer apparatus. Though
`the total inhaled drug mass was similar for the constant-
`output (Misty-Neb and SideStream) and the breath-
`enhanced (Pari LCD) nebulizers, the 2 dosimetric nebu-
`lizers differed in opposite directions. The Circulaire had
`the lowest and the AeroEclipse the highest inhaled mass of
`all the devices tested. We hypothesize that greater appa-
`ratus drug loss in the Circulaire’s bag storage system ac-
`counts for the smaller inhaled mass. With the AeroEclipse
`the apparatus drug loss was similar to the Misty-Neb, but
`inhalation-only aerosol generation shifted aerosol from ex-
`haled to inhaled. Both the Circulaire and the AeroEclipse
`lost less to the ambient air, as expected, based on their
`design and function. The Circulaire contains aerosol dur-
`ing the exhalation phase, and the AeroEclipse’s breath-
`actuation limits aerosol generation to the inhalation phase.
`With breath-actuation, exhaled/ambient loss from the Aero-
`Eclipse was half that of the Circulaire.
`We could find only 1 study, in the form of an abstract,
`by Hess et al, that tested the same nebulizers under a
`uniform set of breathing variables to allow direct compar-
`ison of nebulizer performance.17 Their study measured to-
`tal nebulizer output of albuterol and calculated fine parti-
`cle mass from particle size measurements. They did not
`measure exhaled or nebulizer apparatus drug loss. The fine
`particle mass output was greatest with the AeroEclipse and
`least with the Circulaire. The Pari LCD, SideStream, and
`Misty-Neb were intermediate between the AeroEclipse and
`Circulaire. That is the same order of output found in our
`study for total inhaled mass and for inhalation filter mass
`(total inhaled mass minus throat loss).
`
`Other studies have measured drug output from one or
`several of the nebulizers tested in our study, using various
`breathing conditions. Our measurements of inhaled drug
`mass from the constant-output SideStream agree well with
`Dennis’s in vitro research on the SideStream.11 Devadason
`et al measured inhaled drug mass from a Pari LC, using
`volunteers and filter collection at the mouth.9 They found
`19% for the total inhaled drug, which is higher than the
`15% in our study of the Pari LCD with simulated breath-
`ing. That difference may be due to design differences among
`Pari models, notably between the Pari LC Plus, a nondis-
`posable, reusable unit, and the Pari LCD, a disposable
`unit. The Pari LC Plus has a 1-way valve in the top of the
`nebulizer chamber, which allows ambient air to be en-
`trained during inhalation, with no loss of aerosol on ex-
`halation. The LCD has a simple opening with no valve in
`the top of the nebulizer chamber, and we observed visible
`loss of a small amount of aerosol through that opening
`during exhalation. The Pari LC Plus also has a 1-way
`flapper valve in the mouthpiece, whereas the Pari LCD has
`simple nonvalved openings on either side of the mouth-
`piece. Design differences may also account for the differ-
`ence in exhaled loss between our study (18%) and the
`study by Dennis, who found approximately 11% exhaled
`loss with the Pari LC Plus.11
`Inhaled drug from a breath-enhanced nebulizer also in-
`creases or decreases as a function of inhalation flow.18
`Measuring inhaled mass at a single flow with the Pari
`LCD, as we did, could be seen as limiting, but our peak
`inhalation flow corresponded to the highest uniform flow
`in a study by Knoch et al of the Pari IS-2; this was also the
`flow that gave the highest emitted drug mass in their study.18
`The effect of variable flow or different inhalation wave-
`forms (eg, uniform flow versus a sine waveform) with
`breath-enhanced nebulizers requires further investigation.
`The Circulaire represents an adaptation of a constant-
`output nebulizer to create a dosimetric device, if we accept
`Dennis’s definition of dosimetric as a nebulizer that re-
`
`178
`
`RESPIRATORY CARE • FEBRUARY 2004 VOL 49 NO 2
`
`

`

`PERFORMANCE COMPARISON OF NEBULIZER DESIGNS
`
`leases “aerosol only during the inhalation cycle” and that
`makes “all released aerosol available for patient inhala-
`tion.”1 In an vitro study of the Circulaire, using albuterol,
`Piper found an emitted drug mass of 0.32 ⫾ 0.01 mg,13
`which is approximately 13% of the nominal dose of 2.5
`mg albuterol, compared to 8.7 ⫾ 0.99% found in our
`study. That difference may be due to the fact that in Pip-
`er’s study emitted drug on inhalation was not directly mea-
`sured but rather calculated based on intermittent sampling.
`Measures of drug mass lost in the constant-output and
`breath-enhanced nebulizers have ranged between 55% and
`66% in a number of studies.5,9,11,19,20 That range agrees
`well with the 52– 63% apparatus loss we found for those
`types of nebulizers. Our measurement of drug remaining
`in the AeroEclipse (51%) was identical to that reported by
`Fink et al.21
`A limitation in the present study was the use of the USP
`throat as a simple model of the upper respiratory tract,
`rather than measuring particle size distribution and the fine
`particle fraction. The USP throat was designed to capture
`the large and high-velocity aerosol particles emitted from
`a metered-dose inhaler, when testing at a constant flow of
`approximately 30 L/min. The throat has also been adopted
`by the USP as a model throat for testing dry powder in-
`halers,15 in the testing of which the flow varies through the
`throat, with the testing conditions prescribed in the USP,
`Chapter 601, on aerosols. With a dry powder inhaler there
`is no high-velocity, large-particle fraction, as there is with
`a metered-dose inhaler.15 The measurements of throat loss
`in our study do not represent a certain particle size nor
`provide an estimate of the fine particle mass that could
`reach the lower respiratory tract. Our use of the throat
`provides a standard model for the comparative evaluation
`of nebulizer designs. Based on the mechanism of inertial
`impaction, which is a function of particle mass and veloc-
`ity, we would expect the model throat loss (the difference
`between the total inhaled mass and the inhalation filter
`mass) to be a very approximate measure of larger aerosol
`particles. Measurements of fine particle fractions have been
`reported elsewhere for the Misty-Neb,17 SideStream,11,17
`Pari LC Plus,11 Pari LCD,17 Circulaire,13,17 and Aero-
`Eclipse.17
`
`Conclusions
`
`Our results indicate that design differences among nebu-
`lizers affect drug disposition in inhaled mass, apparatus
`loss, and exhaled/ambient loss. Use of reservoir systems to
`store aerosol during the exhalation phase can cause large
`apparatus losses and thus decrease inhaled mass, whereas
`generating aerosol only during inhalation (ie, breath-actu-
`ated nebulization) increases inhaled mass and decreases
`ambient drug loss. Clinical comparisons are necessary to
`
`determine if these differences substantially affect clinical
`outcomes.
`
`REFERENCES
`
`1. Dennis JH. A review of issues relating to nebulizer standards. J
`Aerosol Med 1998;11 Suppl 1:S73–S79.
`2. Hess D, Fisher D, Williams P, Pooler S, Kacmarek RM. Medication
`nebulizer performance: effects of diluent volume, nebulizer flow,
`and nebulizer brand. Chest 1996;110(2):498–505.
`3. Hess D, Horney D, Snyder T. Medication-delivery performance of
`eight small-volume, hand-held nebulizers: effects of diluent volume,
`nebulizer flow, gas flowrate, and nebulizer model. Respir Care 1989;
`34(8):717–723.
`4. Loffert DT, Ikle D, Nelson HS. A comparison of commercial jet
`nebulizers. Chest 1994;106(6):1788–1792.
`5. Lewis RA, Fleming JS, Fractional deposition from a jet nebulizer:
`how it differs from a metered dose inhaler. Br J Dis Chest 1985;
`79(4):361–367.
`6. Kradjan WA, Lakshminarayan S. Efficiency of air compressor-drive
`nebulizers. Chest 1985;87(4):512–516.
`7. Alvine GF, Rodgers P, Fitzsimmons KM, Ahrens RC. Disposable jet
`nebulizers. How reliable are they? Chest 1992;101(2):316–319.
`8. Pisut FM. Comparison of medication delivery by T-nebulizer with
`inspiratory and expiratory reservoir. Respir Care 1989;34(11):985–
`988.
`9. Devadason SG, Everard ML, Linto JM, Le Soue¨f PN. Comparison of
`drug delivery from conventional versus “Venturi” nebulizers. Eur
`Respir J 1997;10(11):2479–2483.
`10. Coates AL, MacNeish CF, Lands LC, Meisner D, Kelemen S, Vadas
`EB. A comparison of the availability of tobramycin for inhalation
`from vented vs unvented nebulizers. Chest 1998;113(4):951–956.
`11. Dennis JH. Drug nebulizer design and performance: breath enhanced
`jet vs constant output jet vs ultrasonic. J Aerosol Med 1995;8(3):
`277–280.
`12. Mason JW, Miller WC, Small S. Comparison of aerosol delivery via
`Circulaire system vs conventional small volume nebulizer. Respir
`Care 1994;39(12):1157–1161.
`13. Piper SD. In vitro comparison of the Circulaire and AeroTee to a
`traditional nebulizer T-piece with corrugated tubing. Respir Care
`2000;45(3):313–319.
`14. Smaldone GC. Drug delivery by nebulization: “reality testing” (ed-
`itorial). J Aerosol Med 1994;7(3):213–216.
`15. U. S. Pharmacopeia and National Formulary. ⬍601⬎ Aerosols, me-
`tered-dose inhalers, and dry powder inhalers. In: USP 24-NF 19
`Supplement, Rockville MD: United States Pharmacopeial Conven-
`tion Inc, January 1, 2000:2674–2688.
`16. Kirk RE. Experimental design: procedures for the behavioral sci-
`ences. Belmont CA: Wadsworth Publishing Co; 1968:90–91.
`17. Hess D, Mitchell JP, Coppolo D, Nagel MW, Archer AD, Blacker R.
`Effect of nebulizer design on fine particle mass (abstract). Respir
`Care 1999;44(10):1289.
`18. Knoch M, Wunderlich E, Geldner S. A nebulizer system for highly
`reproducible aerosol delivery. J Aerosol Med 1994;7(3):229–237.
`19. Newman SP, Pitcairn GR, Hooper G, Knoch M. Efficient drug delivery
`to the lungs from a continuously operated open-vent nebulizer and low
`pressure compressor system. Eur Respir J 1994;7(6):1177–1181.
`20. Wildhaber JH, Dore ND, Wilson JM, Devadason SG, LeSoue¨f PN.
`Inhalation therapy in asthma: nebulizer or pressurized metered-dose
`inhaler with holding chamber? In vivo comparison of lung deposi-
`tion in children. J Pediatr 1999;135(1):28–33.
`21. Fink JB, Simon M, Uster P. Does use of a nebulizer result in an increase
`in drug concentration? (abstract) Respir Care 2001;46(10):1085.
`
`RESPIRATORY CARE • FEBRUARY 2004 VOL 49 NO 2
`
`179
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket