throbber
International Journal of Pharmaceutics 283 (2004) 1–9
`
`A review of the development of Respimat® Soft MistTM Inhaler
`R. Dalbya,1, M. Spallekb,∗
`
`, T. Voshaarc
`a Department of Pharmaceutical Sciences, School of Pharmacy, University of Maryland, 20 North Pine Street, Baltimore MD 21201-1180, USA
`b Department Drug Delivery, Boehringer Ingelheim Pharma GmbH and Co. KG, Ingelheim, Germany
`c Medical Clinic III, Bethanien Krankenhaus, Moers, Germany
`
`Received 3 January 2004; received in revised form 6 June 2004; accepted 7 June 2004
`Available online 20 August 2004
`
`Abstract
`
`Respimat® Soft MistTM Inhaler (SMI) is a new generation inhaler from Boehringer Ingelheim developed for use with respiratory
`drugs. The device functions by forcing a metered dose of drug solution through a unique and precisely engineered nozzle (the
`uniblock), producing two fine jets of liquid that converge at a pre-set angle. The collision of these two jets generates the soft
`mist. The soft mist contains a high fine particle fraction of approximately 65 to 80%. This is higher than aerosol clouds from
`conventional portable inhaler devices, such as pressurised metered dose inhalers (pMDIs) and dry powder inhalers (DPIs). In
`addition, the relatively long generation time of the aerosol cloud (approximately 1.5 s) facilitates co-ordination of inhalation and
`actuation – a major problem with pMDIs. These features, together with the slow velocity of the soft mist, result in larger amounts
`of the drug reaching the lungs and less being deposited in the oropharynx compared with either pMDIs or DPIs. Generation of
`the soft mist from Respimat® SMI is purely mechanical, so propellants are not necessary. The innovative design of Respimat®
`SMI, using water-based drug formulations, ensures patients receive consistent and reliable doses of the drug with each actuation.
`The device was initially tested in scintigraphic lung deposition studies and produced encouraging results when compared with
`the chlorofluorocarbon-based pMDI (CFC-MDI). Subsequent clinical studies have confirmed that Respimat® SMI is effective
`and safe in delivering bronchodilators to patients with asthma or chronic obstructive pulmonary disease.
`© 2004 Published by Elsevier B.V.
`
`Keywords: Respimat® Soft MistTM; Inhaler (SMI); Lung deposition; Fine particle fraction; Aerosol; Obstructive lung disease; Inhaler device
`
`∗
`
`Corresponding author. Tel.: +49 6132 77 98774;
`fax: +49 6132 77 3823.
`E-mail addresses: rdalby@rx.umaryland.edu (R. Dalby),
`michael.spallek@ing.boehringer-ingelheim.com (M. Spallek).
`1 Tel.: +1 410 706 3245; fax: +1 410 706 0346.
`
`0378-5173/$ – see front matter © 2004 Published by Elsevier B.V.
`doi:10.1016/j.ijpharm.2004.06.018
`
`1. Introduction
`
`Inhalation therapy has led to considerable improve-
`ments in the treatment of obstructive airways diseases,
`such as asthma and chronic obstructive pulmonary dis-
`ease (COPD). As the drug is delivered directly to its
`site of action, a low dose can be used to produce a
`therapeutic response and, consequently, side effects
`
`Liquidia's Exhibit 1061
`Page 1
`
`

`

`2
`
`R. Dalby et al. / International Journal of Pharmaceutics 283 (2004) 1–9
`
`are minimised. Additionally, inhaled drug delivery cir-
`cumvents the limitations imposed by first-pass hepatic
`metabolism and fast absorption results in an onset of
`action that is more rapid than that achieved by oral ad-
`ministration.
`The efficacy of an inhaled drug is largely depen-
`dent on the amount of the drug deposited in the lungs
`and its topographical anatomical distribution; this is
`influenced by various interacting factors, including the
`characteristics of the aerosol, the type of delivery de-
`vice used, the mode of inhalation and the architecture of
`the airways (Ganderton, 1997; Pavia, 1997). The char-
`acteristics of the aerosol will affect the amount of drug
`reaching the lung. The method by which fine particles
`are produced for pulmonary delivery and the size distri-
`bution of these particles significantly affects drug depo-
`sition within the airways (Newman, 1984; Pavia, 1997).
`It may be possible to deliver drugs more precisely by
`using aerosols with a defined particle size distribution;
`for example, particles with a diameter of 2–5 ␮m are
`generally deposited in the smaller bronchioles and pe-
`ripheral airways (Ariyananda et al., 1996). Larger parti-
`cles tend to be deposited in the upper airways, whereas
`those smaller than 2 ␮m are, to a large extent, breathed
`in and out of the alveoli with minimal actual deposi-
`tion (Pavia, 1997; Ariyananda et al., 1996; Matthys,
`1990); small particles that do manage to deposit in the
`alveoli may be rapidly absorbed and exert no pharma-
`codynamic effect (Pritchard, 2001). A study by Zanen
`et al. found that the optimal particle size for ␤2 ago-
`nist and anticholinergic aerosols in patients with severe
`airflow obstruction was approximately 3 ␮m (Zanen et
`al., 1996). More recently, the effects of bronchodilator
`particle size on airway drug deposition in asthmatic pa-
`tients was studied (Usmani et al., 2003). Monodisperse
`salbutamol aerosols of 1.5, 3 and 6 ␮m in size were in-
`haled and lung function changes were determined; the
`3 and 6 ␮m aerosols were significantly more effective
`bronchodilators than the 1.5 ␮m aerosol.
`Several types of portable devices are currently avail-
`able for the delivery of drugs by inhalation; these
`include the chlorofluorocarbon (CFC) and hydroflu-
`oroalkane (HFA) pressurised metered dose inhalers
`(pMDIs), and the dry powder inhalers (DPIs). The
`CFC-MDI has been the cornerstone of asthma and
`COPD maintenance therapy for many years. How-
`ever, many patients experience problems in using CFC-
`MDIs and do not obtain optimal therapeutic benefit
`
`from their medication (Giraud and Roche, 2002). The
`limitations of the pMDI, and the move to eliminate
`CFC propellants for environmental reasons, have accel-
`erated the development of alternative inhaler devices.
`Inherent in the development of these new devices has
`been a determination to improve on known device de-
`ficiencies (Steed et al., 1997).
`Respimat® Soft MistTM Inhaler (SMI) is a new
`generation inhaler that uses mechanical power from
`a spring rather than liquid-gas propellant to generate
`an aerosol cloud suitable for inhalation. This article
`reviews the development of Respimat® SMI and de-
`scribes how the latest advances in aerosol technology
`have been used in order to improve upon existing in-
`haler performance.
`
`2. Respimat® SMI
`
`Respimat® SMI is a new generation, propellant-
`free, multi-dose inhaler developed by Boehringer In-
`gelheim. The term ‘soft mist’ is used to describe both
`the mechanism of aerosol generation and the qualities
`of the aerosol cloud. Respimat® SMI does not belong
`to any of the existing categories of inhaler device and
`represents an innovative approach to patient-oriented
`inhalation therapy.
`
`2.1. Rationale for the development of Respimat®
`SMI
`
`Respimat® SMI was developed in order to overcome
`the limitations of traditional inhaler devices and to meet
`the need for a convenient propellant-free inhaler that
`could effectively deliver aerosols from solutions. Cur-
`rently, the most common inhaler devices used for bron-
`chodilator and anti-inflammatory drug administration
`are pMDIs and DPIs; both have inherent disadvantages
`relating to lung deposition and ease of use.
`The pMDI produces particles that travel very fast,
`generating a high-velocity cloud over a short period
`of time. Two consequences of this are deposition of
`the drug on the back of the throat (the oropharynx)
`and difficulties in synchronising the generation of the
`dose with inspiration. Only about 10–20% of the dose
`released from CFC-MDIs is deposited in the lungs;
`the remainder of the dose is lost through impaction in
`
`Liquidia's Exhibit 1061
`Page 2
`
`

`

`R. Dalby et al. / International Journal of Pharmaceutics 283 (2004) 1–9
`
`3
`
`the oropharynx (Newman et al., 1981). HFA-MDIs are
`similarly inefficient; only one HFA-MDI formulation
`has shown lung deposition >50% (Leach et al., 1998).
`Many patients, particularly children and the elderly, are
`unable to co-ordinate actuation of pMDIs with inhala-
`tion, which is crucial for proper lung deposition; thus
`the amount of drug reaching the lungs is both small
`and variable. Numerous other inhaler technique errors
`have been observed with pMDIs, including stopping
`inhalation when the aerosol hits the back of the throat
`(Pavia, 1997). The soft mist generated by Respimat®
`SMI travels much slower and lasts much longer than
`aerosol clouds from other devices. The relatively long
`period over which the dose from Respimat® SMI is re-
`leased facilitates co-ordination of actuation and inhala-
`tion compared with pMDIs. This should help the pa-
`tient to achieve the correct inhaler technique, which is
`important for successful long-term treatment. Further-
`more, Respimat® SMI is easy and convenient to use; it
`retains the ‘user-friendliness’ of pMDIs, but does not
`require cumbersome spacer devices to slow the aerosol
`cloud and reduce oropharyngeal deposition (Denyer et
`al., 2000).
`DPIs are breath-actuated and therefore require no
`co-ordination between device actuation and inhalation.
`However, both the aerosolisation and delivery of the
`drug to the lung are dependent on an adequate inspi-
`ratory effort from the patient. Airflow achieved early
`in the inspiratory profile deaggregates the drug from
`its carrier powder (usually lactose) and determines the
`particle size distribution of the aerosol. Because of the
`great variability in inspiratory flow, both between pa-
`tients and within an individual patient, the proportion
`of the metered dose that is inhaled varies consider-
`ably, but is typically quite low (Meakin et al., 1998).
`A large fraction of the drug often remains bound to
`the carrier and deposits in the oropharynx (Ganderton,
`1997,1999). For some DPIs, lung deposition is lower
`than that seen with pMDIs (Newman, 1999; Zainudin
`et al., 1990). Importantly, some powder formulations
`are extremely moisture-sensitive; adsorption of mois-
`ture can significantly increase powder cohesiveness,
`leading to decreased generation of fine particles during
`inhalation.
`Energy to generate the soft mist delivered by
`Respimat® SMI comes from a compressed spring in-
`side the inhaler; consequently, the particle size pro-
`duced from the device is not dependent on the patient’s
`
`inspiratory effort. Moreover, Respimat® SMI generates
`an aerosol cloud from a solution rather than a powder,
`avoiding moisture adsorption and powder agglomer-
`ation problems. These characteristics ensure that the
`dose delivered with each actuation from Respimat®
`SMI remains uniform.
`Respimat® SMI is designed to be environmentally
`friendly and to increase lung deposition and reduce
`oropharyngeal deposition of the drug compared with
`pMDIs and DPIs, without the use of spacer devices.
`One design goal was to realise clinical improvements
`and minimise side effects by lowering the nominal in-
`haled dose compared with conventional delivery sys-
`tems.
`
`2.2. The concept of a soft mist inhaler
`
`The generation of an inhalable aerosol from a drug
`solution requires the metered dose of liquid to be
`converted into appropriately sized droplets without
`the use of propellants. One technique involves the
`use of electrical energy to produce vibrations (which
`is common in ultrasonic and piezo-electric devices).
`A second approach is to use mechanical energy to
`force drug solution through a nozzle. Respimat®
`SMI derives mechanical energy from a spring that
`can be easily compressed by the patient (Zierenberg
`et al., 1996). The spring mechanism ensures that the
`aerosol is generated by a reliable and reproducible
`energy source and, consequently, dose and particle
`size distribution of the aerosol are independent of the
`variable inspiratory flow of the patient.
`The soft mist concept was initially demonstrated in
`a prototype model, which consisted of a metal pump
`body and a syringe serving as a solution reservoir. A
`lever arm was used to simultaneously compress the
`spring and withdraw a metered volume of drug solu-
`tion from the reservoir. The liquid was forced through
`a two-channel nozzle upon release of the spring, result-
`ing in aerosol generation. The droplet size distribution
`in the aerosol was demonstrated to be in the range suit-
`able for inhalation (Zierenberg, 1999; Zierenberg and
`Eicher, 2002).
`Following further development of this early pro-
`totype,
`including the introduction of the uniblock
`nozzle (see below),
`the device was successfully
`used in lung deposition studies in healthy volunteers
`
`Liquidia's Exhibit 1061
`Page 3
`
`

`

`4
`
`R. Dalby et al. / International Journal of Pharmaceutics 283 (2004) 1–9
`
`(Steed et al., 1995a,b). Additional refinement of the
`device occurred after patient focus groups evaluated
`four different design prototypes. The preferred version
`of the device was used for phase II and phase III clinical
`trials. Several additional aesthetic modifications have
`been made to the device in advance of its launch onto
`the market. These include a hinged cap, colour-coded
`to identify specific drug classes contained in the device,
`and a transparent base to allow easy identification of
`the drug product (Fig. 1). A schematic illustration of
`the device is shown in Fig. 2.
`The marketed device delivers 120 metered actua-
`tions and has a dose indicator to remind patients when
`a new prescription is needed. A locking mechanism
`automatically prevents the use of the device after all
`120 actuations have been delivered. This ensures that
`there is no detectable ‘tail-off’, commonly seen with
`pMDIs, during which reduced doses are delivered close
`to container exhaustion. Spray content uniformity of
`doses delivered via Respimat® SMI was established
`gravimetrically, using 10 devices from three batches
`(Spallek et al., 2002). The delivered volume was con-
`
`Fig. 1. The marketed version of Respimat® Soft MistTM Inhaler.
`
`Fig. 2. Schematic illustration of Respimat® Soft MistTM Inhaler, showing the key components of the device and details of the uniblock (Spallek
`et al., 2002).
`
`Liquidia's Exhibit 1061
`Page 4
`
`

`

`R. Dalby et al. / International Journal of Pharmaceutics 283 (2004) 1–9
`
`5
`
`Fig. 3. Spray content uniformity data (of an aqueous solution) over 120 actuations delivered via Respimat® Soft MistTM Inhaler (with 10 devices
`from 3 batches) (delivered volume ± S.D.) (Spallek et al., 2002).
`
`sistent throughout 120 actuations of the device and no
`‘tail-off’ effect was observed (Fig. 3).
`Respimat® SMI is similar in size to pMDIs and DPIs
`(such as the Turbohaler®).
`
`2.3. Mode of action of Respimat® SMI
`
`Medication to be delivered by Respimat® SMI is
`stored as a solution in the drug cartridge. The cartridge
`consists of an aluminium cylinder containing a double-
`walled, plastic, collapsible bag, which contracts as the
`solution is withdrawn.
`The initially sterile solution may be formulated
`with either ethanol, which acts both as a solvent
`and preservative, or water, with added preservatives
`(e.g. benzalkonium chloride). Either strategy main-
`tains the microbial stability of the solution follow-
`ing initial puncture of the cartridge prior to first
`use of the device by the patient. Tests on used car-
`tridges have shown that patient use of Respimat®
`SMI does not result in microbiological contamina-
`tion of the inhalation solution (Schmelzer and Bagel,
`2001).
`◦
`The energy from a 180
`twist of the device base com-
`presses the spring. This transfers a pre-defined metered
`volume of the inhalation solution from the drug car-
`
`tridge, through a capillary tube (via a non-return valve),
`to the pump cylinder. When the patient depresses the
`dose-release button, the energy of the spring forces the
`metered volume of drug solution into the uniblock.
`The uniblock is the key element of Respimat® SMI,
`consisting of a nozzle fed by multiple extremely fine
`filter channels. In the initial prototype of Respimat®
`SMI, the nozzle openings were tiny holes pierced into
`a stainless steel disk; however, this design was not suit-
`able for mass production (Spallek et al., 2002). The
`problem was overcome by the development of a minia-
`ture ‘sandwich’ concept, the uniblock, composed of
`a structured silicon wafer bonded to a small (2 mm
`× 2.5 mm) borosilicate glass plate (Fig. 2). Inlet, out-
`let and filter channels (which prevent the nozzle from
`becoming blocked) are etched into the silicon wafer
`using a technique derived from microchip production
`technology (Zierenberg et al., 1996; Zierenberg, 1999;
`Zierenberg and Eicher, 2002; Spallek et al., 2002). This
`allows the units to be produced on a large scale with
`high precision and accuracy. The configuration of the
`inlet and outlet channels is engineered to produce a
`high fine particle fraction (droplets <5.8 ␮m in diame-
`ter). The robustness of the uniblock (and the device as
`a whole) has been confirmed by rigorous mechanical
`testing (Spallek et al., 2002).
`
`Liquidia's Exhibit 1061
`Page 5
`
`

`

`6
`
`R. Dalby et al. / International Journal of Pharmaceutics 283 (2004) 1–9
`
`Fig. 4. Photographs, taken at intervals of 0.2 s, showing the generation of the Soft MistTM from Respimat® Soft MistTM Inhaler (Hochrainer
`and H¨olz, 2001).
`
`The metered drug solution is forced through the
`channels in the uniblock, producing two fine jets of
`liquid at the outlet; these converge at a precisely set an-
`gle (Zierenberg, 1999; Zierenberg and Eicher, 2002),
`generating a slow-moving cloud of inhalable parti-
`cles – the soft mist (Hochrainer and H¨olz, 2001). The
`soft mist emerges from the nozzle with a velocity ap-
`proximately one-tenth of the speed of release of an
`aerosol cloud from a pMDI (Newman et al., 1996).
`Hochrainer and H¨olz compared the velocity of the
`aerosol cloud and the aerosol generation time (spray
`duration) from Respimat® SMI with those delivered
`from various CFC– and HFA-MDIs (Hochrainer and
`H¨olz, 2001). They found that aerosol clouds gener-
`ated from Respimat® SMI were considerably slower-
`moving and had a longer spray duration time compared
`with either CFC- or HFA-MDIs (Fig. 4).
`A high proportion of the droplets in the aerosol cloud
`from Respimat® SMI fall into the fine particle fraction
`(droplets of <5.8 ␮m in diameter) (Wachtel and Ziegler,
`2002; Spallek et al., 2002); these droplets are likely
`to be deposited in the lungs after inhalation. Studies
`on an aqueous solution of fenoterol and an ethano-
`lic solution of flunisolide delivered from Respimat®
`SMI have shown that fine particle fractions are approx-
`imately 66% for the aqueous formulation and 81% for
`the ethanolic formulations (Zierenberg, 1999; Zieren-
`berg and Eicher, 2002). Fig. 5 shows a typical example
`of the particle size distribution of an aqueous solution
`measured in an Andersen Cascade Impactor (Spallek
`et al., 2002). For comparison, the fine particle fractions
`of the aerosols delivered by pMDIs and DPIs are typ-
`ically less than 40% (van Noord et al., 2000; Steed et
`al., 1997; Newman et al., 1998; Hill and Slater, 1998;
`Kamin et al., 2002).
`
`Because Respimat® SMI generates a slow-moving
`aerosol cloud with a high fine particle fraction, ra-
`diolabelled aerosols imaged by gamma scintigraphy
`have been able to demonstrate that less of the dose
`from Respimat® SMI is deposited in the orophar-
`ynx and more reaches the lungs than with pMDIs
`and DPIs (Newman et al., 1996, 1998; Steed et al.,
`1997; Wilding et al., 2002). Newman et al. investi-
`gated lung and oropharyngeal deposition of flunisolide
`administered to 12 healthy subjects via Respimat®
`SMI, pMDI or pMDI plus Inhacort® spacer (Newman
`et al., 1996). Mean whole lung deposition of flu-
`nisolide from Respimat® SMI (39.7%) was signifi-
`cantly higher than from pMDI (15.3%) or pMDI plus
`spacer (28.0%). Typical scans of the distribution of
`radiolabelled aerosol from each device are shown in
`Fig. 6. In a more recent study, in 12 healthy volunteers,
`
`Fig. 5. Typical aerodynamic particle size distribution for the aerosol
`generated by Respimat® Soft MistTM Inhaler, using an aqueous drug
`solution and an Andersen cascade impactor (relative humidity >90%)
`(cumulative mass fraction; % ± S.D.) (Spallek et al., 2002).
`
`Liquidia's Exhibit 1061
`Page 6
`
`

`

`R. Dalby et al. / International Journal of Pharmaceutics 283 (2004) 1–9
`
`7
`
`Fig. 6. Scintigraphic scans from one individual showing the deposition of radiolabelled aerosol in the lungs immediately after administration of
`a single dose of 250 ␮g flunisolide delivered via Respimat® Soft MistTM Inhaler, pressurised metered dose inhaler (pMDI) or pMDI plus spacer,
`on each of three study days (Newman et al., 1996).
`
`Table 1
`The distribution of a metered dose of fenoterol in the lungs and
`oropharynx of 12 healthy subjects using Respimat® Soft MistTM
`Inhaler (SMI), a pressurised metered dose inhaler (pMDI) or a pMDI
`plus spacer as the delivery device
`
`Amount of fenoterol (as % metered dose)
`
`Respimat®
`SMI
`
`pMDI
`
`pMDI plus
`spacer
`
`(Newman et al., 1998). Additionally, Respimat® SMI
`deposited more of the fenoterol dose in the central,
`intermediate, and peripheral lung regions than either
`the pMDI or the pMDI plus spacer. The mean oropha-
`ryngeal deposition of fenoterol was significantly lower
`via Respimat® SMI than via the pMDI (37.1% versus
`71.7% of metered dose, respectively).
`
`39.2 (12.7)
`Whole lung
`Central lung zone 11.0 (3.7)
`Intermediate lung
`14.1 (4.9)
`zone
`Peripheral
`zone
`Oropharynx
`Delivery device
`Exhaled air
`
`lung
`
`14.1 (4.8)
`
`37.1 (10.4)
`21.9 (6.1)
`1.9 (1.7)
`
`11.0 (4.9)
`3.1 (1.1)
`3.7 (1.8)
`
`9.9 (3.4)
`2.5 (0.9)
`3.6 (1.2)
`
`4.2 (2.1)
`
`3.8 (1.5)
`
`71.7 (7.4)
`16.7 (5.4)
`0.6 (0.4)
`
`3.6 (2.4)
`86.2 (5.2)
`0.4 (0.3)
`
`2.4. Clinical performance of Respimat® SMI
`
`Because Respimat® SMI increases the amount of
`drug deposited in the lungs, Boehringer Ingelheim ex-
`pected that inhaled drugs would maintain their efficacy
`even when smaller nominal doses than would normally
`be delivered by a pMDI were administered. This con-
`cept has been tested in a number of clinical studies.
`Respimat® SMI is being developed for use with a
`range of drugs for the treatment of asthma or COPD.
`Some of these studies have been conducted using a
`fixed combination of ipratropium bromide (IB) and
`fenoterol hydrobromide (FEN) (Berodual®); IB/FEN
`is indicated for the symptomatic treatment of airway
`narrowing in patients with asthma or COPD. Five clin-
`
`The proportions of fenoterol deposited in the delivery device and
`detected in the exhaled air are also shown. Fenoterol distribution
`was detected by gamma scintigraphy. Data are given as mean and
`S.D. (Newman et al., 1998).
`
`whole lung deposition of the bronchodilator fenoterol
`was greater when delivered by Respimat® SMI com-
`pared with pMDI or pMDI plus spacer (Table 1)
`
`Liquidia's Exhibit 1061
`Page 7
`
`

`

`8
`
`R. Dalby et al. / International Journal of Pharmaceutics 283 (2004) 1–9
`
`ical studies of IB/FEN delivered via Respimat® SMI
`have already been completed.
`In phase II, a dose-ranging and a cumulative dose-
`response study were conducted in adult asthmatics
`(Goldberg et al., 2001; Kunkel et al., 2000); the first
`of these demonstrated a dose-response relationship for
`IB/FEN delivered via Respimat® SMI, producing safe
`and efficacious bronchodilation in asthmatic patients
`(Goldberg et al., 2001). The cumulative dose study
`showed that Respimat® SMI improves the delivery of
`IB/FEN compared with pMDIs, providing an equipo-
`tent bronchodilatory effect at only one-half of the cu-
`mulative dose needed with a pMDI (Kunkel et al.,
`2000).
`Three phase III trials have been completed; two of
`these were in asthma patients–one in adults (Vincken
`et al., 2004) and one in children (von Berg et al., 2004)
`and the third was in COPD patients (Kilfeather et al.,
`2004). The results showed that Respimat® SMI enables
`a 50% reduction of the nominal daily dose of IB/FEN
`in patients with asthma or COPD while offering simi-
`lar therapeutic efficacy and safety to the corresponding
`CFC-MDI (which was used with a spacer in the paedi-
`atric study) (Vincken et al., 2004; von Berg et al., 2004;
`Kilfeather et al., 2004).
`
`3. Conclusions
`
`The performance of the inhaler device is critical for
`optimal drug delivery, lung deposition and subsequent
`clinical effects. Respimat® SMI represents an inno-
`vative approach to inhalation therapy; this propellant-
`free, multi-dose inhaler generates an inhalable aerosol
`cloud (the soft mist) with superior properties to those
`produced by existing devices, such as pMDIs and DPIs.
`The soft mist generated by Respimat® SMI is pro-
`duced by the collision of two jets of liquid forced
`through a specially designed nozzle. The aerosol cloud
`is much slower moving than those produced by pMDIs
`and DPIs and contains a much higher fine particle frac-
`tion than most of these devices. In addition, the aerosol
`generation time (cloud duration) is longer compared
`with pMDIs and DPIs. The combination of these fea-
`tures improves the efficiency of drug delivery via in-
`halation by increasing the proportion of the inhaled
`dose that is deposited in the lung. This benefit has been
`
`validated in clinical studies, in which Respimat® SMI
`has been shown to allow a reduction in the dose of
`a combination bronchodilator compared with delivery
`via CFC-MDI, while offering the same level of thera-
`peutic efficacy and safety. Therefore, the development
`of Respimat® SMI represents a significant step forward
`in pulmonary drug delivery.
`
`References
`
`Ariyananda, P.L., Agnew, J.E., Clarke, S.W., 1996. Aerosol delivery
`systems for bronchial asthma. Postgrad. Med. J. 72, 151–156.
`Denyer, J., Pavia, D., Zierenberg, B., 2000. New liquid aerosol de-
`vices for inhalation therapy. Eur. Respir. Rev. 10, 187–191.
`Ganderton, D., 1997. General factors influencing drug delivery to the
`lung. Respir. Med. 91 (Suppl. A), 13–16.
`Ganderton, D., 1999. Targeted delivery of inhaled drugs: current
`challenges and future goals. J. Aerosol Med. 12 (Suppl. 1),
`S3–S8.
`Giraud, V., Roche, N., 2002. Misuse of corticosteroid metered-dose
`inhaler is associated with decreased asthma stability. Eur. Respir.
`J. 19, 246–251.
`Goldberg, J., Freund, E., Beckers, B., Hinzmann, R., 2001. Improved
`delivery of fenoterol plus ipratropium bromide using Respimat®
`compared with a conventional metered dose inhaler. Eur. Respir.
`J. 17, 225–232.
`Hill, L.S., Slater, A.L., 1998. A comparison of the performance of
`two modern multidose dry powder asthma inhalers. Respir. Med.
`92, 105–110.
`Hochrainer, D., H¨olz, H., 2001. Comparison of the velocities of
`clouds delivered from Respimat® soft mist inhaler and metered
`dose inhalers. J. Aerosol Med. 14 (3), 386P1–386P5, abstract.
`Kamin, W.E., Genz, T., Roeder, S., Scheuch, G., Trammer, T., Juene-
`mann, R., Cloes, R.M., 2002. Mass output and particle size distri-
`bution of glucocorticosteroids emitted from different inhalation
`devices depending on various inspiratory parameters. J. Aerosol
`Med. 15, 65–73.
`Kilfeather, S.A., Ponitz, H.H., Beck, E., Schmidt, P., Lee, A.L.,
`Bowen, I.R., Hesse, C.W., 2004. Improved delivery of iprat-
`ropium bromide/fenoterol from Respimat® Soft MistTM Inhaler
`in patients with COPD. Respir. Med. 98, 387–397.
`Kunkel, G., Magnussen, H., Bergmann, K.C., Juergens, U.R., de
`Mey, C., Freund, E., Hinzmann, R., Beckers, B., 2000. Respimat®
`(a new soft mist inhaler) delivering fenoterol plus ipratropium
`bromide provides equivalent bronchodilation at half the cumula-
`tive dose compared with a conventional metered dose inhaler in
`asthmatic patients. Respiration 67, 306–314.
`Leach, C.L., Davidson, P.J., Boudreau, R.J., 1998. Improved airway
`targeting with the CFC-free HFA-beclomethasone metered-dose
`inhaler compared with CFC-beclomethasone. Eur. Respir. J. 12,
`1346–1353.
`Matthys, H., 1990. Inhalation delivery of asthma drugs. Lung 168
`(Suppl.), 645–652.
`
`Liquidia's Exhibit 1061
`Page 8
`
`

`

`R. Dalby et al. / International Journal of Pharmaceutics 283 (2004) 1–9
`
`9
`
`Meakin, B.J., Ganderton, D., Panza, I., Ventura, P., 1998. The effect
`of flow rate on drug delivery from the Pulvinal, a high-resistance
`dry powder inhaler. J. Aerosol Med. 11, 143–152.
`Newman, S.P., Pavia, D., Moren, F., Sheahan, N.F., Clarke, S.W.,
`1981. Deposition of pressurised aerosols in the human respiratory
`tract. Thorax 36, 52–55.
`Newman, S.P., 1984. Production of radioaerosols. Butterworths,
`London.
`Newman, S.P., Steed, K.P., Reader, S.J., Hooper, G., Zierenberg, B.,
`1996. Efficient delivery to the lungs of flunisolide aerosol from
`a new portable hand-held multidose nebulizer. J. Pharm. Sci. 85,
`960–964.
`Newman, S.P., Brown, J., Steed, K.P., Reader, S.J., Kladders, H.,
`1998. Lung deposition of fenoterol and flunisolide delivered us-
`ing a novel device for inhaled medicines. Chest 113, 957–963.
`Newman, S.P., 1999. Use of gamma scintigraphy to evaluate the
`performance of new inhalers. J. Aerosol Med. 12 (Suppl. 1),
`S25–S31.
`Pavia, D., 1997. Efficacy and safety of inhalation therapy in chronic
`obstructive pulmonary disease and asthma. Respirology 2 (Suppl.
`1), S5–S10.
`Pritchard, J.N., 2001. The influence of lung deposition on clinical
`response. J. Aerosol Med. 14 (Suppl. 1), S19–S26.
`Schmelzer, C., Bagel, S., 2001. Microbiological integrity of Respi-
`mat soft mist inhaler (SMI) cartridges after use in COPD patients.
`J. Aerosol Med. 14 (3), 385P1–385P3.
`Spallek, M., Hochrainer, D., Wachtel, H., 2002. Optimizing nozzles
`for soft mist inhalers. Respiratory Drug Delivery VIII 2, 375–378.
`Steed, K.P., Freund, B., Towse, L., Newman, S.P., 1995a. High lung
`deposition of fenoterol from BINEB®, a novel multiple dose
`nebuliser device. Eur. Respir. J. 8 (Suppl. 19), 204s.
`Steed, K.P., Zierenberg, B., Reader, S., Newman, S.P., 1995b. Lung
`deposition of flunisolide from BINEB, a novel multiple dose
`nebuliser, is twice that from a pressurised metered dose inhaler.
`Eur. Respir. J. 8 (Suppl. 19).
`Steed, K.P., Towse, L.J., Freund, B., Newman, S.P., 1997. Lung and
`oropharyngeal depositions of fenoterol hydrobromide delivered
`from the prototype III hand-held multidose Respimat nebuliser.
`Eur. J. Pharm. Sci. 5, 55–61.
`Usmani, O.S., Biddiscombe, M.F., Nightingale, J.A., Underwood,
`S.R., Barnes, P.J., 2003. Effects of bronchodilator particle size in
`
`asthmatic patients using monodisperse aerosols. J. Appl. Physiol.
`95, 2106–2112.
`van Noord, J.A., Smeets, J.J., Creemers, J.P.H.M., Greefhorst, L.P.,
`Dewberry, H., Cornelissen, P.J., 2000. Delivery of fenoterol
`via Respimat®, a novel soft mist inhaler. Respiration 67, 672–
`678.
`Vincken, W., Bantje, T., Middle, M.V., Gerken, F., Moonen, D.,
`2004. Long-term efficacy and safety of ipratropium bromide plus
`fenoterol via Respimat® Soft MistTM Inhaler (SMI) versus a pres-
`surised metered-dose inhaler in asthma. Clin. Drug Invest. 24,
`17–28.
`von Berg, A., Jeena, P.M., Soenamtri, P.A., Vertruyen, A., Schmidt,
`P., Gerken, F., Razzouk, H., 2004. Efficacy and safety of ipra-
`tropium bromide plus fenoterol inhaled via Respimat® Soft
`MistTM Inhaler vs. a conventional metered dose inhaler plus
`spacer in children with asthma. Pediatr. Pulmonol. 37, 264–
`272.
`Wachtel, H., Ziegler, J., 2002. Improved assessment of inhaler device
`performance using laser diffraction. Respiratory Drug Delivery
`VIII 2, 379–382.
`Wilding, I.R., Pitcairn, G.R., Reader, S., Pavia, D., Newman, S.P.,
`2002. Respimat® Soft Mist Inhaler delivers inhaled corticos-
`teroid to the lungs more efficiently than a Turbuhaler dry powder
`inhaler or a pressurized metered dose inhaler. A. A. P. S. Pharm.
`Sci. 4 (4), W4130, abstract.
`Zainudin, B.M.Z., Biddiscombe, M., Tolfree, S.E.J., Short, M., Spiro,
`S.G., 1990. Comparison of bronchodilator responses and deposi-
`tion patterns of salbutamol inhaled from a metered dose inhaler,
`as a dry powder, and as a nebulised solution. Thorax 45, 469–473.
`Zanen, P., Go, L.T., Lammers, J.W., 1996. Optimal particle size for
`beta 2 agonist and anticholinergic aerosols in patients with severe
`airflow obstruction. Thorax 51, 977–980.
`Zierenberg, B., Eicher, J., Dunne, S., Freund, B., 1996. Boehringer
`Ingelheim nebulizer BINEB® a new approach to inhalation ther-
`apy. Respir. Drug Deliv. V, 187–193.
`Zierenberg, B., 1999. Optimizing the in vitro performance of
`Respimat®. J. Aerosol Med. 12 (Suppl. 1), S19–S24.
`Zierenberg, B., Eicher, J., 2002. The Respimat, a new soft mist inhaler
`for delivering drugs to the lungs. In: Rathbone, M.J., Hadgraft, J.,
`Roberts, M.S. (Eds.), Modified-release drug delivery technology.
`Marcel Dekker, New York, pp. 925–933.
`
`Liquidia's Exhibit 1061
`Page 9
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket