throbber
British Journal of Anaesthesia 112 (6): 1109–14 (2014)
`Advance Access publication 15 April 2014 . doi:10.1093/bja/aeu095
`
`TRANSLATIONAL RESEARCH
`
`Downloaded from https://academic.oup.com/bja/article/112/6/1109/2919827 by guest on 02 August 2021
`
`Feasibility and accuracy of nasal alar pulse oximetry
`T. E. Morey1*, M. J. Rice1, T. Vasilopoulos2, D. M. Dennis1,3 and R. J. Melker1,3
`
`1 Department of Anesthesiology, University of Florida, PO 100254, 1600 SW Archer Road, Gainesville, FL 32601-0254, USA
`2 Department of Psychiatry and Behavioral Neuroscience, University of Chicago, 5841 S. Maryland Ave., MC 3077, CNPRU, Room L-603,
`Chicago, IL 60637, USA
`3 Xhale, Inc., 3630 SW 47th Ave., Suite 100, Gainesville, FL 32608, USA
`* Corresponding author. E-mail: morey@ufl.edu
`
`Editor’s key points
`† Nasal pulse oximetry is
`an attractive alternative
`when digital perfusion
`is poor or inaccessible.
`† Use of monitoring devices
`outside of their intended
`use requires specific
`validation.
`† Results obtained from
`volunteer studies may not
`translate to the clinical
`environment.
`† The performance of
`monitoring devices is
`most crucial at the
`margins of the normal
`range.
`
`Background. The nasal ala is an attractive site for pulse oximetry because of perfusion by
`branches of the external and internal carotid arteries. We evaluated the accuracy of a novel
`pulse oximetry sensor custom designed for the nasal ala.
`Methods. After IRB approval, healthy non-smoking subjects [n¼12; aged 28 (23–41) yr; 6M/6F]
`breathed hypoxic mixtures of fresh gas by a facemask to achieve oxyhaemoglobin saturations
`of 70–100% measured by traditional co-oximetry from radial artery samples. Concurrent alar
`and finger pulse oximetry values were measured using probes designed for these sites. Data
`were analysed using the Bland–Altman method for multiple observations per subject.
`Results. Bias, precision, and accuracy root mean square error (ARMS) over a range of 70–100%
`were significantly better for the alar probe compared with a standard finger probe. The mean
`bias for the alar and finger probes was 0.73% and 1.90% (P,0.001), respectively, with
`corresponding precision values of 1.65 and 1.83 (P¼0.015) and ARMS values of 1.78% and
`2.72% (P¼0.047). The coefficients of determination were 0.96 and 0.96 for the alar and
`finger probes, respectively. The within/between-subject variation for the alar and finger
`probes were 1.14/1.57% and 1.87/1.47%, respectively. The limits of agreement were
`3.96/22.50% and 5.48/21.68% for the alar and finger probes, respectively.
`
`Conclusions. Nasal alar pulse oximetry is feasible and demonstrates accurate pulse oximetry
`values over a range of 70–100%. The alar probe demonstrated greater accuracy compared
`with a conventional finger pulse oximeter.
`
`Keywords: alar nasal cartilages; lateral nasal cartilages; nasal cartilages; oximetry; pulse
`oximetry
`
`Accepted for publication: 22 November 2013
`
`Pulse oximetry remains a vital tool in healthcare facilities and a
`standard monitor to measure the oxygenation of patients re-
`ceiving anaesthesia. Fingers are a commonly used site for
`probe placement, but may be suboptimal because diminished
`perfusion from numerous causes may result in loss of sufficient
`pulse signal to provide an accurate pulse oximetry oxygen sat-
`).1 – 3 Additionally, finger use may be limited by
`uration (SpO2
`injury, presence on the surgical field, non-invasive arterial pres-
`sure cuff interruption, arm tucking, and shivering.4 – 6 Finally,
`digital pulse oximetry produces an unacceptably high inci-
`dence of failure (79%) during emergency pre-hospital airway
`management.7 Therefore, application to sites other than
`fingers may be useful. Several alternative probe sites on the
`head have been suggested (e.g. earlobe, forehead, cheek,
`tongue) because perfusion to these structures may provide ad-
`equate pulsatile signal even in the presence of significant
`pathophysiology and because the time to detect desaturation
`
`is less when measured from a site on the head, compared with
`a finger or toe.8 – 12
`Recently, Saban and colleagues13 demonstrated that
`branches of both the facial and ophthalmic arteries perfuse the
`nasal alae, the lateral walls of the nares (Fig. 1).13 The ophthalmic
`artery, which is a branch of the internal carotid artery, has been
`shown to autoregulate during experimental hypotension to pre-
`serve blood flow.14 15 Integrating these anatomical and physio-
`logical data, we hypothesize that the nasal ala should be a
`suitable pulse oximetry location. As a first step, the primary aim
`of this investigation was to validate the performance character-
`istics ofalarpulse oximetry probes similar to the seminal pulse ox-
`imeter evaluations of Yelderman and New16 in 1983. In healthy
`subjects receiving hypoxic mixtures of gas to generate oxygen
`saturations of 70–100%, we compared novel alar pulse oximetry
`and conventional finger pulse oximetry with the reference stand-
`ard, co-oximetry measurement of radial artery blood specimens.
`
`& The Author [2014]. Published by Oxford University Press on behalf of the British Journal of Anaesthesia. All rights reserved.
`For Permissions, please email: journals.permissions@oup.com
`
`1
`
`Masimo Ex. 2008
`Apple v. Masimo
`IPR2020-01722
`
`

`

`Downloaded from https://academic.oup.com/bja/article/112/6/1109/2919827 by guest on 02 August 2021
`
`Morey et al.
`
`time-stamped data (SpO2
`, heart rate) were collected from the
`probes at a rate of 1 Hz and automatically logged onto a spread-
`sheet supported by a personal computer for later analysis.
`After instrumentation, subjects began the experimental
`protocol for desaturation similar to that previously conducted
`in this laboratory.17 In brief, subjects were in a 30–458 head-
`up position. Two arterial blood samples were obtained ,30 s
`apart while each subject breathed room air. After aspiration of
`blood (2 ml), the samples were immediately processed and ana-
`lysed to measure the SaO2. Investigators placed a tight-fitting
`full facemask to prevent entrainment of room air. Thereafter,
`hypoxia was induced by supplying a fresh gas mixture of nitro-
`gen, room air, and carbon dioxide. Each plateau level of oxy-
`haemoglobin saturation was maintained for at least 30 s until
`pulse oximeter readings stabilized. Then, two arterial blood
`samples were obtained (cid:3)30 s apart. Each stable plateau was
`therefore maintained for at least 60 s defined by SpO2 variation
`≤3%. The different plateaus were nominally at 98–100% (room
`air saturation), 93%, 90%, 87%, 85%, 82%, 80%, 77%, 75%, and
`70%. This procedure was performed twice with the second
`epoch completing any saturation values missed during the
`first run. Approximately 25 samples were obtained across this
`anticipated range of saturation (70–100%) for each subject
`that completed the protocol.
`
`Statistical analysis
`Descriptive data are reported as mean (SD). Data were analysed to
`calculate bias, precision, and the accuracy root mean square error
`(ARMS). ARMS is a composite value determined by both bias and
`precision and is used as an overall index of uncertainty of the
`SpO2
`value by regulatory authorities such as the US National Insti-
`tute of Standards in Technology with values ,2–3% considered
`(cid:3)
`
`bias2 + precision2
`to be acceptable.18 ARMS was calculated as
`.
`These values were calculated for oxyhaemoglobin saturation
`bands of 70–80%, 80–90%, and 90–100%, and 70–100% (all
`data).18 Differences in bias between the nasal alar probe and
`finger probe were assessed using a paired Student’s t-test, utiliz-
`ing the full sample of subject measurements across all saturation
`bands and within each band. A folded F-test for homogeneity of
`variances was similarly used to determine significant differences
`in precision between the probes.
`In review of the medical literature, we did not discover a
`method to statistically evaluate ARMS values. We did learn,
`however, of a potential approach for similar metrics that have
`been utilized in other fields by Nilsson and colleagues.19 We
`adapted these methods to evaluate statistical differences in
`ARMS between probes. ARMS values were calculated for each
`subject (n¼12). Specifically, subject-level bias and precision
`were first calculated using the multiple measurements obtained
`from each subject. Then, these bias and precision values were
`combined to determine ARMS for each subject. This resulted in
`two sets of ARMS values, with one set for each type of probe.
`The average ARMS values for each probe were then compared
`using a paired Student’s t-test. To the best of our knowledge,
`this is the first study to adapt this type of approach to examine
`statistical differences in ARMS.
`
`BJA
`
`Fig 1 Reproduction of the anastomotic polygonal system forming
`the arterial supply of the nose that includes the facial artery from
`the external carotid artery and the nasal branch of the ophthalmic
`artery from the internal carotid artery. Note the anastomosis of the
`external and internal carotid arterial system at the right and left
`ala. Reproduced from Saban and colleagues13 with permission.
`&(2012) American Medical Association. All rights reserved.
`
`Methods
`Subject recruitment
`This non-treatment investigation was approved by the Univer-
`sity of California at San Francisco Committee on Human Re-
`search under protocol H6301-01706-24 and was conducted
`at the HYPO2XIALAB at that institution (San Francisco, CA,
`USA). Written,
`informed consent was obtained from all
`recruited subjects, who were aged 18–50 yr. Subjects self-
`identified their own ethnic and racial categories as defined by
`the National Institutes of Health. Exclusionary criteria were
`history of tobacco smoking, hypertension, respiratory disease,
`or haemoglobin ≤10 g dl21.
`
`Protocol
`On the day of study, a catheter was inserted using aseptic tech-
`nique into a radial artery of each subject in order to aspirate
`blood for determination of arterial oxyhaemoglobin saturation
`(SaO2 ). Values for SaO2 were measured using a multi-wavelength
`oximeter (OSM3w Hemoximeter, Radiometer Medical A/S,
`Copenhagen, Denmark), which served as the reference standard
`for subsequent comparisons with SpO2
`readings. An alar pulse
`oximeter sensor
`(Assurance Biosense,
`Inc., Glastonbury,
`CT, USA) was placed on the right ala to measure SpO2
`. An add-
`itional, conventional pulse oximeter probe and monitor was
`placed on the index finger for reference purposes (N-595 Pulse
`Oximeter, Mallinckrodt Inc., St Louis, MO, USA). Date- and
`
`1110
`
`2
`
`

`

`Downloaded from https://academic.oup.com/bja/article/112/6/1109/2919827 by guest on 02 August 2021
`
`BJA
`
`to calculate the bias and limits of agreement as suggested by
`Myles and Cui.22 P,0.05 was considered significant.
`
`Results
`Patient characteristics data
`Subjects (n¼12) aged 28 (23–41) yr were divided evenly
`between males (n¼6) and females (n¼6). The ethnic and
`racial categories were as follows: six Asians, four non-Hispanic
`whites, one Hispanic white, and one African-American. Skin
`tone was variable and characterized as light (n¼3), light-
`medium (n¼2), medium (n¼3), medium-dark (n¼2), and
`dark (n¼2). Eleven subjects completed the entire protocol of
`the study with 25 or more saturation observations. One
`subject complained of anxiety during the first episode of desat-
`uration and self-terminated participation after collection of
`eight saturation observations. All subjects’ data were included
`in this analysis.
`
`Oximetry data
`A representative study (Subject 1) of the alar SpO2
`values and
`heart rate over time are shown in an illustrative example
`(Fig. 2). A number of co-oximetry oxyhaemoglobin saturations
`were episodically collected as well. Using these data from
`Figure 2 and from all other subjects (n¼12 for all subjects), ana-
`lysis was performed to determine the bias, precision, and ARMS
`for oxyhaemoglobin saturation bands of 70–80%, 80–90%,
`and 90–100% (Table 1). The bias from the alar probe was sig-
`nificantly less than that from the finger probe for all bands of
`saturation. However, across the separate bands, there were
`not consistent significant differences in precision and ARMS
`between probes. Using the method comparison technique for
`multiple measurements, no magnitude effect on bias was
`evident. Further, the Bland–Altman method comparison ana-
`lysis was also performed with all saturation results (70–100%)
`as noted in Table 2.
`Revised Bland–Altman analysis of pooled data for repeated
`measurements showing bias plots for the alar (Fig. 3) and
`finger (Fig. 4) probes are shown.21 22 Overall, less bias across
`all saturations was noted with the alar probe compared with
`the finger probe (P,0.001). Additionally, the nasal alar probe
`
`Alar pulse oximetry accuracy
`
`Additionally, method comparisons were performed using
`revised Bland–Altman analysis using SaO2 as the reference
`standard. In the original Bland–Altman technique, only one
`measurement per subject was described under constant con-
`ditions.20 Some investigators, however, measure subject vari-
`ables (e.g. SpO2
`, SaO2) repeatedly under different conditions
`such as variable inspired concentrations of oxygen. Measure-
`ment of the subject more than once and under changing
`conditions violates assumptions of independent sampling.
`Therefore, they proposed a newer method that we have used
`herein.21 In the technique described in section 3 of their publi-
`cation [a required statistical analysis by the US Food and Drug
`Administration for 510(k) medical device clearance], Bland
`and Altman provide a way to determine within-subject vari-
`ance using analysis of variance (SigmaPlot 12.2, Systat Soft-
`ware, Inc., San Jose, CA, USA), between-subject variance,
`total variance, bias, and limits of agreement for subjects
`providing multiple samples under changing physiological con-
`ditions that would normally violate assumptions of independ-
`ent sampling.21 Additionally, we pooled data for each subject
`
`Pulse (min–1)
`
`125
`
`100
`
`75
`
`50
`
`SpO2
`SaO2
`Pulse
`
`0
`
`10
`
`20
`Time (min)
`
`30
`
`40
`
`100
`
`90
`
`80
`
`70
`
`60
`
`50
`
`Oxyhaemoglobin (%)
`
`Fig 2 Representative study of oxyhaemoglobin saturation and
`heart rate over time in a human subject receiving variable concen-
`trations of inspired oxygen by a facemask. Shown is the oxyhaemo-
`globin concentration measured by alar pulse oximetry (n¼2620)
`with some points also measured concurrently by co-oximetry
`(n¼25).
`
`Table 1 Bias and ARMS for alar and finger pulse oximetry data stratified by co-oximeter oxygen saturation bands from human subjects (n¼12)
`receiving variable fresh gas flow concentrations of inspired oxygen by a facemask. Oxygen saturation band grouping was determined by
`co-oximetry measurement
`
`Saturation (%)
`
`Observations
`
`Parameter
`
`70–80
`
`80–90
`
`90–100
`
`85
`
`97
`
`93
`
`Bias
`Precision
`ARMS
`Bias
`Precision
`ARMS
`Bias
`Precision
`ARMS
`
`Alar
`
`0.82
`1.77
`1.95
`1.13
`1.49
`1.87
`0.22
`1.51
`1.53
`
`Finger
`
`2.93
`1.98
`3.53
`2.06
`1.52
`2.56
`0.82
`1.24
`1.49
`
`P-value
`
`,0.001
`0.305
`0.023
`,0.001
`0.862
`0.155
`,0.001
`0.063
`0.396
`
`1111
`
`3
`
`

`

`BJA
`
`Morey et al.
`
`Downloaded from https://academic.oup.com/bja/article/112/6/1109/2919827 by guest on 02 August 2021
`
`Upper limit=4.53
`
`Bias=2.02
`
`Lower limit=–0.50
`
`80
`
`85
`Mean [(S pO2
`+SaO2
`
`)/2] (%)
`
`90
`
`6 4
`
`2 0
`
`–2
`
`–4
`
`–6
`
`) (%)
`
`–SaO2
`
`Finger difference (SpO2
`
`Fig 4 Revised Bland–Altman analysis for multiple measurements
`of oxyhaemoglobin concentrations measured by finger pulse oxim-
`) and co-oximetry (SaO2) for human subjects (n¼12) re-
`etry (SpO2
`ceiving variable concentrations of inspired oxygen by a facemask.
`Shown are individual subjects’ pooled observations, bias (solid
`blue line), and upper and lower limits of agreement (dashed blue
`line).
`
`Bland–Altman analysis using non-pooled data is shown in
`Supplementary Figs S1–S4.)
`
`Discussion
`We designed and tested the accuracy of a novel, alar pulse ox-
`imeter sensor in healthy volunteers during both normal (room
`air) and hypoxic conditions, using a standard method for
`testing pulse oximeters and sensors. The ARMS of the alar
`data set compares favourably with the accepted instrument
`specifications of a ≤4% ARMS for functional oxygen saturation
`promulgated by the International Organization for Standard-
`ization (IOS), ≤3% for the US Food and Drug Administration,
`and ≤2% ARMS for most other regulatory agencies.18 Moreover,
`the sensor demonstrated acceptable accuracy not only for all
`aggregated data (70–100%), but also at all incremental satur-
`ation bands tested. An ARMS of 2% does not mean that all points
`are within 2% of the reference standard co-oximetry, but rather
`that most points ((cid:3)67%) are within this range. Furthermore, a
`sensor could theoretically have a relatively poor bias or low pre-
`cision, but not both, and meet the ambiguity target of ≤4%
`ARMS. Manufacturers of pulse oximeters generally claim an ac-
`curacy of 2–3% over the saturation range of 70–100%, which
`would roughly correspond to an ARMS of 2%, but the Inter-
`national Standard Organization reminds us that ‘. . . only
`about two-thirds . . .’ of measurements will fall within the
`stated twice the ARMS.1 – 3 23 Likewise, Batchelder and Raley18
`point out that the ARMS represents a group of values and does
`not allow for consideration of temporary ‘pop-ups’ or assess
`dynamic performance conditions. Notwithstanding,
`the
`present alar probe possessed acceptable characteristics as cur-
`rently assessed by various agencies. Furthermore, via several
`significance testing approaches, the nasal alar probe demon-
`strated significantly less bias, better precision, and lower ARMS
`
`Table 2 Bland–Altman analysis of oxyhaemoglobin saturation for
`alar and finger oximetry probes with comparison with co-oximetry
`saturations as the reference standard method in human subjects
`(n¼12) receiving variable concentrations of inspired oxygen by a
`facemask
`
`Parameter
`
`Alar
`
`Finger
`
`n
`Regression
`Coefficient of determination
`(r2)
`Within-subject variation
`Between-subject variation
`Total variation
`Standard deviation
`Bias
`Upper limits of agreement
`Lower limit of agreement
`Overall ARMS
`
`275
`y¼0.96 x+3.92
`0.96
`
`275
`y¼0.89 x+10.97
`0.96
`
`1.14
`1.57
`2.71
`1.65
`0.73
`3.96
`22.50
`1.78
`
`1.87
`1.47
`3.34
`1.83
`1.90
`5.48
`21.68
`2.72
`
`Upper limit=3.44
`
`Bias=0.56
`
`6 4 2 0
`
`–2
`
`–4
`
`–6
`
`80
`
`Lower limit=–2.31
`
`85
`Mean [(S pO2
`+SaO2
`
`)/2] (%)
`
`90
`
`Fig 3 Revised Bland–Altman analysis for multiple measurements
`of oxyhaemoglobin concentrations measured by alar pulse oxim-
`) and co-oximetry (SaO2) for human subjects (n¼12) re-
`etry (SpO2
`ceiving variable concentrations of inspired oxygen by a facemask.
`Shown are individual subjects’ pooled observations, bias (solid
`blue line), and upper and lower limits of agreement (dashed blue
`line).
`
`was overall more precise than the finger probe (P¼0.017). For
`both Figures 3 and 4, the outlying point originates from the
`subject that completed only eight saturation values at the
`greater end of the desaturation protocol before withdrawing
`from the investigation. After calculating ARMS for each subject
`for each probe, the average ARMS for the alar probe (1.76%)
`and finger probe (2.54%) using this approach closely
`matched the overall ARMS values that were calculated using
`each subject measurement separately (Table 2). Using this ap-
`proach, the average ARMS for the alar probe was significantly
`less than that for the finger probe (P¼0.047). (Note: The
`
`1112
`
`) (%)
`
`–Sa
`
`O2
`
`O2
`
`Alar difference (Sp
`
`4
`
`

`

`Downloaded from https://academic.oup.com/bja/article/112/6/1109/2919827 by guest on 02 August 2021
`
`BJA
`
`population was entirely healthy with a normal perfusion
`state. Specific testing is required with subjects in a low perfu-
`sion state and subjects with impaired circulation to determine
`if alar pulse oximetry is accurate in these subjects when digital
`pulse oximetry fails. Finally, the long-term effects of continu-
`ous alar oximetry over days and weeks at this anatomical loca-
`tion are not known.
`In summary, we have demonstrated acceptable accuracy of
`a novel nasal alar pulse oximetry sensor in subjects both with
`normal and low oxygen saturations and also have provided evi-
`dence that nasal alar pulse oximetry may be less biased, more
`precise, and more accurate than conventional finger pulse ox-
`imetry. Future studies should focus on the use and accuracy of
`this sensor in subjects under a variety of pathophysiological
`conditions such as trauma, cardiopulmonary bypass, and in
`subjects in critical care units.
`
`Supplementary material
`is available at British Journal of
`Supplementary material
`Anaesthesia online.
`
`Authors’ contributions
`The authors individually contributed to the study design and
`data acquisition (R.J.M., T.E.M.), data analysis (T.E.M., D.M.D.,
`T.V.), and early drafts of the manuscript (T.E.M., M.J.R., R.J.M.)
`and critical revisions (T.E.M., M.J.R., T.V., R.J.M., D.M.D.). All
`authors have approved the final version of the manuscript.
`
`Acknowledgements
`The authors appreciate the time and careful effort of the staff
`of the HYPO2XIALAB located at the Department of Anesthesi-
`ology, University of California, San Francisco, San Francisco,
`CA, USA.
`
`Declaration of interest
`R.J.M. and D.M.D. are employees of Xhale, Inc., the company
`that sponsored the study. T.E.M. is a consultant for Xhale, Inc.
`R.J.M., D.M.D., and T.E.M. also own equity in Xhale, Inc. In add-
`ition, the University of Florida owns equity in Xhale, Inc. If a
`device is sold commercially, then the authors listed above
`and the University of Florida could benefit financially.
`
`Funding
`The project was supported by Xhale, Inc., Gainesville, FL, USA.
`
`References
`1 Brimacombe J, Keller C. Successful pharyngeal pulse oximetry in
`low perfusion states. Can J Anaesth 2000; 47: 907–9
`2 Secker C, Spiers P. Accuracy of pulse oximetry in patients with low
`systemic vascular resistance. Anaesthesia 1997; 52: 127–30
`3 Wilson BJ, Cowan HJ, Lord JA, Zuege DJ, Zygun DA. The accuracy of
`pulse oximetry in emergency department patients with severe
`sepsis and septic shock: a retrospective cohort study. BMC Emerg
`Med 2010; 10: 9
`
`1113
`
`Alar pulse oximetry accuracy
`
`when compared with a finger probe when all saturations
`(70–100%) were analysed. These differences in bias were also
`evident when analyses were performed across oxyhaemoglo-
`bin saturation bands of 70–80%, 80–90%, and 90–100%.
`However, the significant differences in precision and ARMS were
`not similarly consistent across each saturation band. These
`inconsistent results could be due to the reduced sample sizes
`of the analyses using the separate bands, which would reduce
`power to detect group differences.
`We specifically selected the nasal ala as a potential new site
`for pulse oximetry because of its unique arterial perfusion
`characteristics. Recently, Saban and colleagues13 used ultrason-
`ography on 40 living subjects and anatomical dissection on 20
`cadavers to demonstrate that the nose is perfused by a four
`component, multidirectional, arterial arcade to form a polygon-
`al system of vessels that anastomose the external and internal
`carotid arteries (Fig. 1). In contrast, the arterial supply of the ear,
`pharynx/cheek, and tongue originates solely from the external
`carotid artery. During periods of severe pathophysiology (e.g.
`shock, hypothermia, stress) that potentially require vasopressor
`therapy, pulse oximeter sensors on the fingers may cease to
`function because of reduction in systemic blood flow, particular-
`ly to the periphery. During these conditions, however, cerebral
`perfusion may be maintained via the internal carotid artery
`with autoregulation of blood flow to the central nervous
`system. Although not answered in the present feasibility in-
`vestigation, we hypothesize that alar pulse oximetry will be a
`superior site for pulse oximetry during periods of low systemic
`blood flow because of preserved arterial perfusion to the
`central nervous system. Since the nasal branch of the ophthal-
`mic artery provides a sample of internal carotid artery flow
`and partially perfuses the ala, an adequate pulse oximetry
`signal could be preserved to provide accurate SpO2
`values.
`The only other region of the head with a blood supply origin-
`ating from the internal carotid artery and assessable non-
`invasively for oximetry is the forehead which is perfused by a
`branch of the internal carotid artery (supraorbital artery).
`This oximetry site has been demonstrated to have acceptable
`bias and limits of agreement for patients undergoing vascular
`surgery, is more accurate than finger SpO2
`measurement in crit-
`ically ill patients receiving high-dose vasopressor therapy, and
`monitor.8 – 12 24 – 26
`responds more rapidly than a finger SpO2
`Monitoring at this site requires reflectance oximetry, however,
`wherein precise localization of the sensor and the presence of
`a corresponding vein in close proximity may cause erroneously
`values.13 27 To prevent erroneous values, the
`decreased SpO2
`use of a circumferential headband is suggested to dampen
`venous pulsations.14 15 28 Because these data demonstrate
`that monitoring from the nasal ala is feasible, accurate SpO2
`monitoring using transmittance oximetry may be useful for
`patients where traditional finger placement fails and is an alter-
`native approach to the supraorbital artery.
`The major limitation of this study was subject selection. That
`is, only healthy adult volunteers were observed over a short
`time period. In this initial investigation, no paediatric or neo-
`natal subjects were included that would have required a
`smaller alar pulse oximetry probe. Moreover, the subject
`
`5
`
`

`

`Downloaded from https://academic.oup.com/bja/article/112/6/1109/2919827 by guest on 02 August 2021
`
`BJA
`
`4 Gehring H, Hornberger C, Matz H, Konecny E, Schmucker P. The
`effects of motion artifact and low perfusion on the performance
`of a new generation of pulse oximeters in volunteers undergoing
`hypoxemia. Respir Care 2002; 47: 48–60
`5 Barker SJ, Shah NK. Effects of motion on the performance of pulse
`oximeters in volunteers. Anesthesiology 1996; 85: 774–81
`6 Barker SJ. ‘Motion-resistant’ pulse oximetry: a comparison of new
`and old models. Anesth Analg 2002; 95: 967–72
`7 Davis DP, Aguilar S, Sonnleitner C, Cohen M, Jennings M. Latency
`and loss of pulse oximetry signal with the use of digital probes
`during prehospital rapid-sequence intubation. Prehosp Emerg
`Care 2011; 15: 18–22
`8 Yu H, Liu B. Successful use of pharyngeal pulse oximetry with the
`oropharyngeal airway in severely shocked patients. Anaesthesia
`2007; 62: 734–6
`9 Rogers SR, Gan TJ. Pulse oximeter probe sheath for buccal use.
`Anesth Analg 1997; 85: 1414
`10 Reynolds LM, Nicolson SC, Steven JM, Escobar A, McGonigle ME,
`Jobes DR. Influence of sensor site location on pulse oximetry
`kinetics in children. Anesth Analg 1993; 76: 751–4
`11 Cote´ CJ, Daniels AL, Connolly M, Szyfelbein SK, Wickens CD. Tongue
`oximetry in children with extensive thermal injury: comparison
`with peripheral oximetry. Can J Anaesth 1992; 39: 454–7
`12 Phillips JP, Kyriacou PA, Jones DP, Shelley KH, Langford RM. Pulse ox-
`imetry and photoplethysmographic waveform analysis of the
`esophagus and bowel. Curr Opin Anaesthesiol 2008; 21: 779–83
`13 Saban Y, Andretto Amodeo C, Bouaziz D, Polselli R. Nasal arterial
`vasculature: medical and surgical applications. Arch Facial Plast
`Surg 2012; 14: 429–36
`14 Kolodjaschna J, Berisha F, Lung S, Schima H, Polska E, Schmetterer L.
`Comparison of the autoregulatory mechanisms between middle
`cerebral artery and ophthalmic artery after thigh cuff deflation in
`healthy subjects. Invest Ophthalmol Vis Sci 2005; 46: 636–40
`15 Ne´meth J, Kne´zy K, Tapaszto´ B, Kova´cs R, Harka´nyi Z. Different auto-
`regulation response to dynamic exercise in ophthalmic and central
`retinal arteries: a color Doppler study in healthy subjects. Graefes
`Arch Clin Exp Ophthalmol 2002; 240: 835–40
`
`Morey et al.
`
`16 Yelderman M, New W. Evaluation of pulse oximetry. Anesthesiology
`1983; 59: 349–52
`17 Feiner JR, Severinghaus JW, Bickler PE. Dark skin decreases the
`accuracy of pulse oximeters at low oxygen saturation: the effects
`of oximeter probe type and gender. Anesth Analg 2007; 105:
`S18–23
`18 Batchelder PB, Raley DM. Maximizing the laboratory setting for
`testing devices and understanding statistical output in pulse oxim-
`etry. Anesth Analg 2007; 105: S85–94
`19 Nilsson H, Olsson H, Juslin P. The cognitive substrate of subjective
`probability. J Exp Psychol Learn Mem Cogn 2005; 31: 600–20
`20 Bland JM, Altman DG. Statistical methods for assessing agreement
`between two methods of clinical measurement. Lancet 1986; 1:
`307–10
`21 Bland JM, Altman DG. Agreement between methods of measure-
`ment with multiple observations per individual. J Biopharm Stat
`2007; 17: 571–82
`22 Myles PS, Cui J. Using the Bland–Altman method to measure agree-
`ment with repeated measures. Br J Anaesth 2007; 99: 309–11
`23 Milner QJW, Mathews GR. An assessment of the accuracy of pulse
`oximeters. Anaesthesia 2012; 67: 396–401
`24 Wax DB, Rubin P, Neustein S. A comparison of transmittance and
`reflectance pulse oximetry during vascular surgery. Anesth Analg
`2009; 109: 1847–9
`25 Nesseler N, Fre´nel J-V, Launey Y, Morcet J, Malle´dant Y, Seguin P.
`Pulse oximetry and high-dose vasopressors: a comparison
`between forehead reflectance and finger transmission sensors.
`Intensive Care Med 2012; 38: 1718–22
`26 Choi SJ, Ahn HJ, Yang MK, et al. Comparison of desaturation
`and resaturation response times between transmission and re-
`flectance pulse oximeters. Acta Anaesthesiol Scand 2010; 54:
`212–7
`27 Dassel AC, Graaff R, Sikkema M, Meijer A, Zijlstra WG, Aarnoudse JG.
`Reflectance pulse oximetry at the forehead improves by pressure
`on the probe. J Clin Monit 1995; 11: 237–44
`28 Covidien. NellcorTM Forehead SpO2 Sensor with OxiMax Technology
`Application Guide. 2011
`
`Handling editor: P. S. Myles
`
`1114
`
`6
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket