`SKIN BEFLEGTANGE PlllSE OXIMETBY: Ill VIVIJ
`MEASUREMENTS FROM THE FDREARM AND GALF
`
`Y. Mendelson, PhD, and M. j McGinn, MSC
`
`From the Department of Biomedical Engineering, Worcester Poly—
`technic Institute, 100 Institute Rd, Worcester, MA 01609.
`
`Received Sep 18, 1989, and in revised formjan 11, 1990. Accepted for
`publication jan 17, 1990.
`
`Address correspondence to Dr Mendelson.
`
`Mendelson Y, McGinn M]. Skin reflectance pulse oximetry: in vivo
`measurements from the forearm and calf.
`
`j Clin Monit 1991;7:7~12
`
`ABSTRACT. This study describes the results from a series of
`human experiments demonstrating the ability to measure arte—
`rial hemoglobin oxygen saturation (SaOZ) from the forearm
`and calf using a reflectance pulse oximeter sensor. A special
`optical reflectance sensor that includes a heating element was
`interfaced to a temperature controller and a commercial Data-
`scope ACCUSAT pulse oximeter that was adapted for this
`study to perform as a reflectance pulse oximeter. The reflec—
`tance pulse oximeter sensor was evaluated in a group of 10
`healthy adult volunteers during steady—state hypoxia. Hy—
`poxia was induced by gradually lowering the inspired fraction
`of oxygen in the breathing gas mixture from 100 to 12%.
`Simultaneous Sa02 measurements obtained from the forearm
`and calf with two identical reflectance pulse oximeters were
`compared with $2102 values measured by a finger sensor that
`was interfaced to a standard Datascope ACCUSAT transmit—
`tance pulse oximeter. The equations for the best-fitted linear
`regression lines between the percent reflectance, SpOz(r), and
`transmittance, Sp02(t), values in the range between 73 and
`100% were 313020“) = — 7.06 + 1.09 Sp02(t) for the forearm
`(n = 91, r = 0.95) and Sp02(r) = 7.78 + 0.93 SpOz(t) for the
`calf (n = 93, r = 0.88). The regression analysis of the forearm
`data revealed a mean : SD error of 2.47 i 1.66% (5302 =
`90—100°/o), 2.35 i 2.45% (SaOz = 80—89%), and 2.42 i
`1.20% (SaOz = 70-79%). The corresponding regression anal—
`ysis of the calf data revealed a mean t SD error of 3.36 i-
`3.06% (SaOz = 90~100°/o), 3.45 i 4.12% (82102 = 80-89%),
`and 2.97 1*: 2.75% (8302 = 70—79%). This preliminary study
`demonstrated the feasibility of measuring Sa02 from the
`forearm and calf in healthy subjects with a heated skin reflec—
`
`tance sensor and a pulse oximeter.
`
`KEY WORDS. Blood gas analyses. Monitoring: oxygen. Mea-
`surement techniques: pulse oximetry; optical plethysmog-
`
`raphy; reflectance oximetry. Equipment: pulse oximeters.
`
`Transmittance pulse oximetry has become a widely
`used technique for noninvasively monitoring changes in
`arterial hemoglobin oxygen saturation (SaOZ). The
`technique is based on the spectrophotometric analysis of
`the optical absorption properties of blood combined
`with the principle of photoplethysmography.
`In transmittance pulse oximetry, which is based on
`tissue transillumination, sensor application in adults is
`limited to several specific locations on the body, such as
`the finger tips, ear lobes, and toes. In infants, additional
`monitoring sites such as the palms and the feet have
`been used.
`
`Recently, a new reflectance pulse oximeter has been
`introduced into the market. The oximeter, which is
`manufactured by Ciba—Corning (Ciba Corning Diag-
`nostics, Medfield, MA), uses a special optical reflectance
`sensor for specific application to the forehead. Among
`the advantages of this technique, as advertised by the
`
`APPLE 1018
`
`APPLE 1018
`
`1
`
`

`

`
`
`
`14'91'3'9'9'9{9 I
`99999999999999999 I
`999999999999999999.’
`. 9%‘9’9‘9‘9’9‘9’9‘9’9’9’9’l’l’
`
`
`l9'9‘9’9’9'9’9‘9’9’9‘9’9’9’9’9’9’ ‘’ ’ ‘ ’I 999999999999999999‘9’9’9’9‘9’
`999999999999999999999999’
`9999999999999999999999999’
`.9999999999999999990999999’
`i59.9.9.9.9¢.9.9,9.9.9.9.9.9.9.9.9.9 9 9 9 9 9 9 9 9 I
`999999999]
`
`;9999999999999999999999999999]
`
`
`9‘9‘9‘9’9’9’9’9’9‘9‘9‘0"‘‘“‘~’9’9°9‘9'9’9‘9’9’9’9’9‘ ”
`
`999999999)
`“9’9‘9‘9’9’9’9’9‘9‘
`'3’3'303'3'393’ '
`“"""’°”""‘"‘I
`
`
`
`
`’9‘0'9‘9’9’9‘9‘9/
`‘9‘9’9’9’9’9‘9’9’9”
`0.9‘9'9’9’9‘9’99
`99.9.9.9’9’9’0’‘ ’
`
` woo¢999i
`55.9.9.9"?
`99999999‘
`99999999
`[4.333.334
`‘~’0W~‘o‘°‘:‘-/
`”999999:
`99999999
`
`
`
`
`
`
`
`lgggzzzzg
`3.334444
`E! H
`
`
`99999999/
`
`’9‘9’9’9‘9’9’9’9‘
`'9‘9‘9‘9’9'9’9‘9‘
`
`9 999999
`
`
`
`
`
`"2402033334
`E! E
`5:”‘.’.’.‘.‘/
`
`
`99999999
`99999999-
`"9’9’9’9‘9’9’9‘9‘
`’9‘9‘9'9’9‘9’9‘9‘
`
` 59.9.9.9.9.9.9.9‘
`"9999999
`
`
`
`
`
`
`999999999
`999999999
`’~0999999o-
`‘9‘9’9’9’9’9’9’0g
`‘999999999
`999999999-
`’999999999
`999999999
`999999999
`9999999994
` ’9999999999‘
`
`
`
`9999999999
`
`’9999999999.
`(9999999999
`9t’
`
`O O O’9’9.QO OQ6 Q 9 O 9O
`9‘9.0
`O’9.0.9 O9 OO
`
`O'999 9,99‘9O
`
`O
`0
`¢€€£€€¢€€€€§$§w'I9’9‘:‘:°:‘.’””"”I
`
`O
`9
`O
`s‘)0..a:.0
`9
`O
`
`
`99999999 9‘9‘993999'9’9‘9’9’9”!O O
`‘0
`O
`9:9.999
`O O0
`9 .9
`0 9
`0O O
`'9
`O'9
`O
`
`
`O
`.09
`O.9O.0O O
`
`‘9.IIOQ.099 9
`O OO 9O O O 9'9
`2
`9’9
`
`.0O Q
`9 O
`OO O
`9",9
`9’9
`3
`9 O
`9 ‘1
`O
`9’9‘9’929
`999'9'9
`?
`0
`O4 O
`
`:9Q 0 O
`O
`s“‘h‘99
`
`§:S5\zi:\zs---
`;9.
`4‘
`
`
`'/ I5mm
`OPTICAL SHIELD
`R 5 IR LED:
`
`PEOTODIODES
`PEOTODIC’DES
`
`OPTICALLY
`BRASS RING
`
`CLEAR 2?0X!
` TEERMOFOIL
`
`9.0333333: """""""""""" r:¢:o;9’o.o:oio. .
`HEATER
`u\\\\\\\‘
`
`\\\
`////////////////’///'/
`
`
`
`S ILICONE
`RUBBER
`
`Fig 1. (A) Frontal and (B) side Views ofthe heated skin reflec-
`tance pulse oximeter sensor. See textfor explanation. R & IR
`LEDs = red and inflated light-emitting diodes.
`
`cludes an array of six identical photodetectors arranged
`symmetrically in a hexagonal configuration surround-
`ing two pairs of red (peak emission wavelength, 660
`nm) and infrared (peak emission wavelength, 930 nm)
`light—emitting diodes (LEDs)
`[1].
`In another related
`study, we showed that by locally heating the skin under
`the sensor to a temperature above 40°C, it is possible to
`achieve a four— to fivefold increase in the magnitude of
`the pulsatile component detected from the forearm, and
`thus significantly improve the detection reliability of the
`reflectance photoplethysmograms [2]. The new optical
`reflectance sensor designed for this study combines the
`two features described above.
`
`SENSOR DESIGN
`
`The temperature-controlled optical reflectance sensor
`used in this study is shown in Figure 1. The major fea-
`ture of the optical layout design is the multiple photo-
`diode array, which is arranged concentric with the
`LEDs. This arrangement maximizes the amount of
`backscattered light that is detected by the sensor. The
`technical details related to the design and geometric
`
`
`
`0O
`O.
`
`9‘
`
`9s
`
`TEMPERATURE
`TRANSDUCER
`
`7 ’
`
`8 Journal ofClinical Monitoring Vol 7 No 1 january 1991
`
`company, are better reliability in critical care situations
`such as peripheral circulatory shutdown, less interfer—
`ence from ambient light, and better accuracy because
`measurement from the forehead is relatively unsuscep—
`tible to motion artifacts.
`
`there are no commercially available re—
`Currently,
`flectance pulse oximeters for monitoring $2102 from lo—
`cations other than the forehead. Therefore, the objective
`of this work was to investigate the feasibility of moni-
`toring SaOZ with a skin reflectance pulse oximeter from
`two alternative and convenient locations on the body:
`the ventral side of the forearm and the dorsal side of the
`
`calf. Besides extending the clinical application of pulse
`oximetry, it appears also that reflectance pulse oximetry
`from peripheral tissues may have potential advantage in
`the assessment of local blood oxygenation after skin
`transplantation and regeneration following microvascu—
`lar surgery.
`In this article, we describe preliminary in vivo evalua—
`tion of a new optical reflectance sensor for noninvasive
`monitoring of Sa02 with a modified commercial trans-
`mittance pulse oximeter. We present the experimental
`evaluation of this sensor in a group of 10 healthy adult
`volunteers and compare SaOz measured with the reflec-
`tance pulse oximeter sensor, Sp02(r), with Sa02 mea-
`sured noninvasively from the finger by a standard trans-
`mittance pulse oximeter sensor, Sp02(t).
`
`REFLECTANCE PULSE OXIMETRY
`
`The principle of reflectance, or backscatter, pulse ox-
`imetry is generally similar to that of transmittance pulse
`oximetry. Both techniques are based on the change in
`light absorption of tissue caused by the pulsating arterial
`blood during the cardiac cycle. The pulsating arterioles
`in the vascular bed, by expanding and relaxing, mod—
`ulate the amount of light absorbed by the tissue. This
`rhythmic change produces characteristic photoplethys—
`mographic waveforms, two of which are used to mea—
`sure Sa02 noninvasively.
`Recently, we showed that accurate noninvasive mea—
`surements of Sa02 from the forehead can be made with
`an unheated reflectance pulse oximeter sensor [1] The
`major practical limitation of reflectance pulse oximetry
`is the comparatively low—level photoplethysmograms
`recorded from low—density vascular areas of the skin.
`Therefore, the feasibility of reflectance pulse oximetry
`depends on the ability to design an optical reflectance
`sensor that can reliably detect sufficiently strong reflec—
`tance photoplethysmograms from various locations on
`the skin.
`
`In order to partially overcome this limitation, we
`have developed an optical reflectance sensor that in—
`
`2
`
`

`

`Mendelson and McGinn: Skin Reflectance Pulse Oximetry
`
`9
`
`configuration of the optical components were described
`recently by Mendelson et al [1].
`The heater consists of a ring—shaped (dimensions:
`30-mm outside diameter; 15-min inside diameter)
`thermofoil resistive heating element (Ocean State Ther-
`motics, Smithfield, RI). The thermofoil heater was
`mounted between the surface of the optically clear
`epoxy, which was used to seal the optical components
`of the reflectance sensor, and a thin (0.005 mm) match—
`ing brass ring, which facilitates better thermal conduc-
`tion to the skin. A miniature (dimensions: 2 X 5
`X 1 mm) solid-state temperature transducer (AD 590,
`Analog Devices, Wilmington, MA) was mounted on
`the outer surface of the brass ring with the thermally
`sensitive surface facing the skin. The entire sensor as-
`sembly was potted in room-temperature vulcanizing
`silicone rubber to minimize heat losses to the surround-
`
`ing environment. The assembled sensor weighs approx—
`imately 65 g. The sensor measures approximately 38
`mm in diameter and is 15 mm thick. The heater assem—
`
`bly was separately interfaced to a temperature controller
`that was used to vary the temperature of the skin be—
`tween 35 and 45°C in 1 : 0.1°C steps.
`
`SUBJECTS AND METHODS
`
`Data Acquisition
`
`Each of the two heated optical reflectance sensors were
`separately interfaced to a temperature controller and a
`commercially available ACCUSAT (Datascope Corp,
`Paramus, NJ) pulse oximeter [3].
`Two of the three ACCUSAT pulse oximeters were
`modified to function as reflectance pulse oximeters. The
`modification, which was described in a separate study
`[1], included the adjustment of the red and infrared LED
`intensities in the reflectance sensors so that the reflec—
`
`tance photoplethysmograms were approximately equal
`to transmittance photoplethysmograms measured by a
`standard transmittance sensor from an average size adult
`finger tip.
`The third ACCUSAT transmittance pulse oximeter
`was used as a reference to measure Sp02(t) from the
`finger tip. The specified accuracy of this transmittance
`pulse oximeter is i2.0°/o and 14.0% for S2102 values
`ranging between 70 and 100% and 60 and 70%, respec—
`tively [3} The three pulse oximeters were adapted to
`provide continuous digital readouts of the AC and DC
`components of the red and infrared photoplethysmo-
`grams.
`
`Readings from each of the three pulse oximeters were
`acquired every 2 seconds through a standard RS-232C
`
`serial port interface using an AT&T 6300 personal com—
`puter. The conversions of the reflectance red/ infrared
`(R/IR) ratios measured by the two reflectance pulse ox—
`imeters to Sp02(r) were performed by using the cali-
`bration algorithm obtained in a previous calibration
`study in which measurements were made with a similar
`nonheated sensor from the forehead [1].
`
`In Vivo Study
`
`The ability to measure 513020?) from the forearm and
`calf was investigated in vivo during progressive steady-
`state hypoxia in humans.
`Measurements were acquired from 10 healthy non—
`smoking male adult volunteers of different ages and skin
`pigmentations. The study was performed in compliance
`with the University of Massachusetts Medical Center’s
`review guidelines on human experimentation. Each
`volunteer was informed of the complete procedure as
`well as the possible risks associated with breathing hy—
`poxic gas levels. Each volunteer received monetary
`compensation for participation in this study. The sub—
`ject distribution included 1 East Indian, 3 Asians, and 2
`darkly tanned and 4 lightly tanned Caucasians. Their
`ages ranged from 22 to 37 years old (mean : SD, 27.5
`i 4.9 years). Measured blood hematocrits were in the
`range of40 to 50.5% (mean i SD, 45.7 i 3.2%).
`All instruments were allowed to warm up for at least
`30 minutes before the study. The transmittance sensor
`of the pulse oximeter was attached to the index finger.
`The reflectance sensors were attached to the ventral side
`
`of the forearm and the dorsal side of the calf by using a
`double—sided transparent adhesive ring. In cases where
`an abundance of hair prevented intimate contact be—
`tween the sensors and the skin,
`the contact was im-
`
`proved by loosely wrapping the sensor and the limb
`with an elastic strap. The temperature of each reflec—
`tance sensor was set to 40°C and remained unchanged
`throughout the entire study.
`A standard lead—I electrocardiogram and end—tidal
`carbon dioxide levels were continuously monitored by
`a Hewlett—Packard 78345A patient monitor (Hewlett—
`Packard, Andover, MA). Each subject was placed in a
`supine position. A face mask was tightly fitted over the
`subject’s nose and mouth, and the subject was instructed
`to breathe spontaneously while we administered differ-
`ent gas mixtures of nitrogen and oxygen. The inspired
`gas mixture was supplied by a modified Heidbrink anes—
`thesia machine (Ohio Medical Products, Madison, WI).
`The breathing circuit of the anesthesia machine was
`equipped with a carbon dioxide scrubber (soda lime).
`The inspired oxygen concentration was adjusted be—
`tween 12 and 100% and was monitored continuously
`
`3
`
`

`

`10 journal ofClinical Monitoring Vol 7 No 1 jannary 1991
`
`throughout the study with an IL 408 (Instrumentation
`Laboratories, Lexington, MA) oxygen monitor, which
`was inserted in the inspiratory limb of the breathing
`circuit.
`
`Steady-state hypoxia was gradually induced by low-
`ering the inspired fraction of oxygen in the breathing
`gas mixture. Initially, the inspired oxygen concentra—
`tion was changed in step decrements, each step pro—
`ducing approximately a 5% decrease in Sp02(t) as
`determined from the display of
`the ACCUSAT
`transmittance pulse oximeter. The inspired oxygen was
`maintained at each level for at least 3 minutes until the
`
`pulse oximeter readings reached a steady level (i.e.,
`Sa02 fluctuations of less than i 3%). When the inspired
`oxygen level reached 12%,
`the process was reversed.
`Thereafter, the inspired oxygen level was increased in a
`similar stepwise manner to 100%. Data were recorded
`during both desaturation and reoxygenation.
`All subjects tolerated the procedure well without ad—
`verse reactions. None of the subjects showed electrocar—
`diographic abnormalities before or after the study. Each
`subject was studied for approximately 1 hour.
`
`Data Analysis
`
`To avoid operator biases, the data from each pulse ox—
`imeter were acquired automatically by the computer
`and later subjected to the same statistical tests.
`For each step change in inspired oxygen, readings
`from the three pulse oximeters were averaged consecu—
`tively over a period of 20 seconds. Averaged readings
`from the 10 subjects were pooled and a least-squares
`linear regression analysis was performed. Student’s ttest
`determined the significance of each correlation; p <
`0.001 was considered significant.
`Although the correlation coefficient of the linear re—
`gression (1) provides a measure of association between
`the Sp02(r) and Sp02(t) measurements, it does not pro—
`vide an accurate measure of agreement between the two
`variables. Therefore, the measurement accuracy was es—
`timated on the basis of the mean and standard deviations
`
`of the difference between the readings from the trans—
`mittance and reflectance pulse oxirneters. The mean of
`the difference between the pulse oximeter measure-
`ments, which is often referred to as the bias, was used to
`assess whether there was a systematic over- or underes—
`timation of one method compared with the other. The
`standard deviation of the bias, which is often referred to
`as the precision, represents the variability or random
`error. Finally, we computed the mean errors and stan-
`dard deviations of each measurement. The mean error
`
`is defined as the absolute bias divided by the corre—
`sponding SpOg(t) values.
`
`1.6
`
`
`
`R/IRREFLECTANCERATIO
`
`H N
`
`.0P4%00
`
`U
`
`0.4
`
`U. 8
`
`1.2
`
`1.6
`
`R/ IR TRANSMITTANCE RATIO
`
`Fig 2. Comparison of red/infrared (R/IR) ratios measured by the
`modified reflectance pulse oximeter (y axis) and the standard trans-
`mittance pulse oximeter (x axis) during progressive steady-state
`hypoxia in 10 healthy subjects. The solid line represents the best-
`fitted linear regression line jbr the firearm measurements. The bro-
`ken line represents the best-fitted linear regression line for the calf
`measurements.
`
`
`
`RESULTS
`
`Normalized R/ IR ratios and Sp02(r) values measured
`by the reflectance pulse oximeters from the forearm and
`calf of the 10 subjects were compared with the nor-
`malized R/ IR ratios and Sp02(t) values measured simul-
`taneously by the transmittance pulse oximeter from the
`finger. A total of 91 and 93 pairs of data points measured
`simultaneously from the forearm and calf, respectively,
`were used in the regression analysis, which provided the
`estimated slopes and intercepts of the linear regression
`lines. Each pair of data points represents a different hy-
`poxic level.
`Regression analysis of the normalized R/ IR ratios
`measured from the reflectance pulse oximeters from the
`forearm and calf (y axis) versus the normalized R/IR
`ratios measured simultaneously by the transmittance
`pulse oximeter from the finger tip (x axis) is shown in
`Figure 2. The equations for the best—fitted linear regres--
`sion lines were y = ~ 0.05 + 1.02x (r = 0.94, SEE =
`0.08, p < 0.001) for the forearm and y = 0.04 + 0.87):
`(i = 0.88, SEE = 0.11, p < 0.001) for the calf.
`A comparison of Sp02(r) readings from the reflec-
`tance pulse oximeter (y axis) and Sp02(t) readings mea-
`
`4
`
`

`

`Mendelson and McGinn: Skin Reflectance Pulse Oximetry
`
`11
`
`
`
`100
`
`U) CD
`
`CD CD
`
`
`
`REFLECTANCESpOz(515)
`
`Y = 0.93x + 7.78
`
`FOREARM
`y = 1.09x - 7.06
`_ CAL?
`
`12
`
`CD
`
`
`DIFFERENCES
`
`
`7U
`
`80
`
`90
`
`100
`
`7U
`
`80
`
`90
`
`100
`
`TRANSMITTANCE Sp02 (as)
`
`TRANSMITTANCE Sp02 (%)
`
`Fig 3. Comparison ofpercent arterial hemoglobin oxygen satura-
`tion (SpOz) measurements obtained from the modified reflectance
`pulse oximeter (y axis) and Spoz values measured by a standard
`transmittance pulse oximeter (x axis) during progressive steady-
`state hypoxia in 10 healthy subjects. The solid line represents the
`best-fitted linear regression line for the forearm measurements. The
`broken line represents the best—fitted linear regression line for the
`calf measurements.
`
`DIFFERENCES
`
`7U
`
`80
`
`90
`
`100
`
`TRANSMITTANCE Sp02 (‘8)
`
`Fig 4. Mean differences between arterial hemoglobin oxygen sat—
`uration (SpOz) measured from the forearm by the modified reflec-
`tance pulse oximeter and the standard transmittance pulse oximeter
`measurements flom the finger tip.
`
`Fig 5. Mean differences between arterial hemoglobin oxygen sat-
`uration (SpOZ) measured from the calf by the modified reflectance
`pulse oximeter and the standard transmittance pulse oximeter mea—
`surements fiom the finger tip.
`
`Statistical Analysis of Arterial Oxygen Saturation (SaOZ) Levels
`Measured from the Forearm and Calf by the Modified Reflectance
`Pulse Oximeters
`
`Location/
`"/0 Sa02
`
`No. of
`Data Points
`
`Mean Value (SD)
`
`Difference
`
`% Error
`
`Forearm
`
`90—100
`80—89
`70—79
`Calf
`90—100
`80—89
`70—79
`
`42
`37
`12
`
`43
`33
`17
`
`1.25 (2.55)
`0.52 (2.85)
`—0.82 (1.96)
`
`2.47 (1.66)
`2.35 (2.45)
`2.42 (1.20)
`
`1.57 (4.00)
`2.22 (4.00)
`1.95 (2.42)
`
`3.36 (3.06)
`3.45 (4.12)
`2.97 (2.75)
`
`sured simultaneously from the transmittance pulse OX-
`imeter (x axis) is shown in Figure 3. The equations for
`the best-fitted linear regression lines were y = —- 7.06
`+ 1.09X (r = 0.95, SEE = 2.62, p < 0.001) for the
`forearm and y = 7.78 + 0.93X (r = 0.88, SEE = 3.73,
`p < 0.001) for the calf.
`Figures 4 and 5 show the percent differences between
`Sp02(r) and Sp02(t), that is, SpOz(r) — Sp02(t), ob-
`tained from the forearm and calf data plotted in Figure
`3, respectively. The corresponding means and standard
`deviations of the differences and errors for the forearm
`and calf measurements are summarized in the Table.
`
`5
`
`

`

`12
`
`Journal ofClinical Monitoring Vol 7 No 1 january 1991
`
`Data were summarized for three different ranges of
`Sp02(t) values between 70 and 100%.
`
`DISCUSSION
`
`Commercially available transmittance sensors can be
`used on only a limited number of peripheral locations of
`the body. Brinkman and Zijlstra [4] and Cohen and
`Wadsworth [5] showed that instead of tissue transil-
`lumination, noninvasive monitoring of SaOz can be
`performed based on skin reflectance
`spectropho—
`tometry. More recently, we described an improved
`optical reflectance sensor that was used for measuring
`Sa02 from the forehead with a modified commercial
`
`transmittance pulse oximeter [1].
`Measuring large reflectance photoplethysmograms
`from sparsely vascularized areas of the skin is challeng—
`ing. Differences in capillary densities between various
`locations on the body are known to affect the magnitude
`and quality of the reflected photoplethysmograms. For
`example, estimated average capillary density of the hu—
`man forehead is approximately 127 to 149 loops/mmz,
`whereas the capillary densities of the forearm and calf
`are approximately 35 to 51 and 41 loops/mmz, respec—
`tively [6,7]. Furthermore, the frontal bone of the fore—
`head provides a highly reflective surface that signifi—
`cantly increases the amount of light detected by the
`reflectance sensor. Therefore,
`reflected photoplethys—
`mograms recorded from the forehead are normally
`larger than those recorded from the forearm and calf.
`Local skin heating could be used as a practical method
`for improving the signal—to-noise ratio of the reflected
`photoplethysmograms from the forearm or calf areas
`and thus reduce the measurement errors in reflectance
`
`pulse oximetry.
`The approach presented in this article demonstrated
`that Sa02 can be estimated by using a heated skin
`reflectance sensor from the forearm and calf over a rela—
`
`tively wide range of Sa02 values. This technique may
`provide a clinically acceptable alternative to currently
`available transmittance pulse oximeters. In a previous
`study [2], we found that the ability to measure accurate
`SaOz values with a reflectance skin oximeter is indepen—
`dent of the exact skin temperature. We noticed, how-
`ever,
`that a minimum skin temperature of approxi—
`mately 40°C is generally sufficient to detect adequately
`stable photoplethysmograms. Furthermore, our experi-
`ence in healthy adults also has shown that at this skin
`temperature, the heated sensor can remain in the same
`location without any apparent skin damage.
`Note that despite the proven advantage of local skin
`heating to increase skin blood flow, reflected photo—
`plethysmograms recorded from the forearm and the calf
`are considerably weaker than those recorded from the
`
`the mean errors for the Sp02(r)
`forehead. Therefore,
`measurements from the forearm and calf are higher than
`the corresponding errors for similar Sp02(r) measure—
`ments made with an unheated reflectance sensor from
`
`the forehead. For comparison, relative to SaOZ mea—
`sured with a noninvasive transmittance pulse oximeter,
`the SEE for Sp02(r) measurements obtained from the
`forehead using a similar unheated optical reflectance
`sensor were 1.82% [1]. The SEE obtained in this study
`using the heated reflectance sensor were 2.62% for the
`forearm and 3.73% for the calf measurements. Despite
`those differences, it is apparent that the degree of corre—
`lation obtained in this preliminary study is encouraging
`and in selected clinical applications may be acceptable.
`We conclude that reflectance pulse oximetry from the
`forearm and calf may provide a possible alternative to
`conventional transmittance pulse oximetry and reflec—
`tance pulse oximetry from the forehead. Further stud-
`ies, however, are needed in order to compare our
`reflectance pulse oximeter against Sa02 measurements
`obtained directly from arterial blood samples. Addi-
`tional work to investigate the source of variability in
`reflectance pulse oximetry is in progress.
`
`Financial support for this study was provided in part by the
`Datascope Corporation and NIH Grant R15 GM36111—01A1.
`
`The authors would like to acknowledge the clinical assistance
`of Albert Shahnarian, PhD, Gary W. Welch MD, PhD, and
`Robert M. Giasi, MD, Department of Anesthesiology, Uni—
`versity of Massachusetts Medical Center, Worcester, MA.
`We also thank Paul A. Nigroni, Datascope Corporation,
`Paramus, NJ, and Kevin Hines, Semiconductor Division,
`Analog Devices, Wilmington, MA, for technical assistance.
`The skillful art work by Yi Wang is also greatly appreciated.
`
`REFERENCES
`
`1. Mendelson Y, Kent JC, Yocum BL, Birle MJ. Design and
`evaluation of a new reflectance pulse oximeter sensor.
`Biomed Instrum Technol 1988;22(4):167—173
`2. Mendelson Y, Ochs BD. Noninvasive pulse oximetry
`utilizing skin reflectance photoplethysmography.
`IEEE
`Trans Biomed Eng 1988;35(10):798—805
`3. Mendelson Y, Kent JC, Shahnarian A, et al. Evaluation of
`the Datascope ACCUSAT pulse oximeter
`in healthy
`adults. J Clin Monit 1988;4z59—63
`4. Brinkman R, Zijlstra WG. Determination and continuous
`registration of the percentage oxygen saturation in clinical
`conditions. NethJ Surg 1949;1:177—183
`5. Cohen A, Wadsworth N. A light emitting diode skin
`reflectance oximeter. Med Biol Eng Comput 1972;10:385—
`391
`6. Moretti G. Handbuch der Haut- und Geschlechtskran—
`
`kheiten. Berlin: Springer—Verlag, 1968:491—623
`7. Rothman S. Physiology and biochemistry of the skin.
`Chicago: University of Chicago Press, 19541685
`
`6
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket