`Computers and Communications (ITC-CSCC 2008)
`
`Measurement Site and Applied Pressure Consideration in Wrist
`Photoplethysmography
`Eun Geun1, Hyun Heo2, Ki Chang Nam3 and Young Huh4
`1, 2, 3, 4Korea Electrotechnology Research Institute (KERI)
`1271-19, Sa-dong, Sangnok-gu, Ansan, Gyeonggi-do, Korea
`E-mail: 1kimeg917@keri.re.kr, 2gjgus1@nate.com, 3chadol@keri.re.kr, 4yhuh@keri.re.kr
`
`Abstract: The objective of this study is to describe
`preliminary evaluation of a new optical reflectance sensor
`module with air pressure cuff. In order to improve PPG
`signals from the wrist as an example of wearable PPG, the
`optical reflectance sensor module includes two identical
`photodiodes and a pair of red and infrared LED. The sensor
`module was packaged inside of a cuff to demonstrate the
`effects of pressure applied to reflectance probe at the radial
`artery in the wrist. PPG signals with large amplitude were
`measured when the induced cuff pressure was close to
`mean blood pressure. This result will be applied to
`development of a wrist type healthcare device.
`
`1. Introduction
`It is well established that recordings of the beat-to-beat
`variations
`in heart rate can be a useful diagnostic
`technology in cardiovascular medicine [1]. Although the
`majority of applications use ECG as the sensing mechanism
`or cardiac dynamics, it has been noted that photoelectric
`plethysmography [2], also known as photoplethysmography
`(PPG), can also be employed to derive essentially the same
`information [3]. PPG is a non-invasive method of detecting
`the cardio-vascular pulse wave that propagates through the
`body. The cardiovascular wave is stimulated by each
`cardiac contraction, and detected in the periphery using a
`light source and a detector. Clinical applications of PPG
`technology include the measurement of heart rate [4, 5],
`peripheral vascular properties [6-8], blood pressure [9, 10],
`and autonomous nervous activity [11, 12].
`Photoplethysmography(PPG) is measured by analyzing
`the pulsatile components of the detected red and infrared
`plethysmograms which make use of
`transmitted or
`reflected light intensities. In reflectance , the light from the
`LEDs enters the tissue, is scattered by both the moving red
`blood cells and the non-moving tissue, and a part of this
`back scattered light is detected by the photodiode(PD)[13].
`Reflectance(back-scatter) PPG has potential advantages
`compare with transmission PPG. The probe can be placed
`at virtually any site on the skin, because the light sources
`and the detector(s) are integrated side by side in the probe.
`Therefore, the probe can even be used on locations where
`light cannot be transmitted[14].
`However, there are some problems in the reflectance
`PPG method. The major practical
`limitation
`is
`the
`comparatively low-level photoplethysmograms recorded
`from low-density vascular areas of the skin. Therefore, the
`feasibility of reflectance PPG depends on the ability to
`design an optical reflectance sensor that can reliably detect
`sufficiently strong reflectance photoplethysmograms from
`various locations on the skin[15]. In order to partially
`
`overcome these limitations Dassel et al. applied pressure on
`the probe to increase the accuracy of reflectance pulse
`oximetry at the forehead[14].
`Measuring with high-quality data on the wrist is also
`difficult and often requiring the application of pressure to
`the sensor in order to reduce the physical contact between
`the sensor and
`the wrist[16]. Furthermore, sensor
`configuration and power consumption play a important role
`in the design of a wearable bio -instrumentation device[17].
`To implement wrist wearable healthcare device, we
`have developed an optical reflectance sensor module that
`includes an array of two identical photodiodes and a pair of
`red and infrared LED. The sensor module was packaged
`inside of a cuff to demonstrate the effects of pressure
`applied to reflectance probe at the radial artery in the wrist.
`The objective of this study is to describe preliminary
`evaluation of a new optical reflectance sensor module with
`air pressure cuff, in order to improve PPG signals from the
`wrist area as an example of wearable PPG.
`
`2. Method
`2. 1 Experimental Setup
`Fig. 1 shows the general PPG block diagram. The system
`was designed to detect red and infrared (IR) signal. The
`photodiodes(PDs) detected radiation back scattered by the
`tissue from both infrared and red LED and delivered output
`current proportional to the detected radiation level. The
`analog signals from the common current to voltage
`converter were subsequently separated into IR and red
`components by
`signal
`conditioning
`circuitry(Fig.2)
`Amplified and filtered analog signal components were
`measured by a oscilloscope (TDS3014, Tektronix).
`
`Fig.1. System block diagram of PPG.
`
`1129
`
`MASIMO 2017
`Masimo v. Apple
`IPR2020-01538
`
`- 1 -
`
`
`
`centered at three points as shown in Fig.5 and the measured
`amplitudes of red and IR signal were compared. In the
`second experiment, the probe was mounted on the finger,
`the palm of the hand and palmer side of the wrist. In the
`third experiment, sensor was located at ‘2’ in Fig.5 and the
`amplitudes of PPG were compared according to pressure
`application by air cuff.
`
`Fig.2. Developed PPG circuit board.
`
`The constructed reflectance sensor consists of 2 silicon
`PD chips(active chip area : 3mm x 3mm) and a pair of red
`(660nm)and IR(940nm) LEDs. LEDs were positioned
`between identical two PDs as shown in Fig. 3,4. The
`distance of PDs was set to 6 mm. The optical sensor
`module was packaged inside of a air pressure cuff.
`
`Fig.3. Air pressure cuff and reflectance PPG probe; side
`(top) and bottom (bottom) view.
`
`Fig.4. Constructed sensor module (Air pressure cuff and
`reflectance PPG probe)
`
`2. 2 Experiment
`One volunteer was participated in preliminary measurement.
`Subject was informed about the purpose and procedure of
`the study and asked written consent. During
`the
`measurement of PPG with wrist cuff, air pressure was
`induced by gradually from 0 to 140 mmHg occluding the
`radial artery at increment of 20 mmHg. Pressure was
`monitored with a analog sphygmomanometer. In the event
`the volunteer felt uncomfortable, the measurement process
`was stopped.
`Three experiments performed to compare the PPG
`amplitudes measured by the reflectance sensor probe. In
`sensor positioning experiment, the probe was mounted on
`the palmer side of the wrist. The sensor module was
`
`1
`A
`
`2
`B
`
`C
`3
`
`Radial artery
`
`Fig.5. Sensor positioning on the wrist (top) and anatomical
`position in bone (bottom)
`
`3. Results & Discussion
`Fig.6, 7, and 8 shows the measured PPG at three point ‘1, 2,
`and 3’ in the wrist described in Fig. 5, respectively. There
`were no sigficant amplitude difference among
`three
`measurement positions except that it seems to be less
`amplitude at ‘3’. It is thought that sensor attachment on the
`wrist is affected by radius bone structure and radial artery
`location.
`The PPG signal from widely distributed portion of the
`capillaries parts such as finger or palm was relatively stable.
`But the results measured at another location of the wrist got
`a weak signal. Even thougth the PPG is easily measurable
`at finger, extensional finger probe is necessary. To
`implement a wrist type device, other possible measurement
`positions which is necessary of extentional probe option
`was exclude from the experiment.
`Fig. 9 shows the measured red LED reflectance signal
`by applying pressure on the wrist. Naturally, maximum
`amplitude changes were shown when the induced pressure
`was close to the mean arterial blood pressure of the subject.
`It is considered that air cuff module can be used for stable
`attachment between sensor and skin.
`PPG signal from radial artery might different from other
`subject because every person have different radial artery
`location and depth. Furthermore, PPG signal greatly
`distorted depending on applied pressure. Through
`experiments, it was verified that proper location of the
`sensors and pressure on the the skin surface are important
`element on stable PPG measurement.
`
`1130
`
`- 2 -
`
`
`
`4. Conclusion
`The objective of this study is to describe reflectance sensor
`module configuration with air pressure cuff, in order to
`improve PPG signals from the wrist area as an example of
`wearable PPG.
`Measuring PPG on the wrist is difficult. Because people
`have different position and depth of radial artery each other.
`The position of sensor and application of pressure to the
`sensor in order to reduce the physical contact between the
`sensor and the wrist affect on the magitude of PPG. As
`preliminary study to implement wrist wearable device, we
`have developed an sensor module that includes an optical
`reflectance sensor and air cuff. Further study is necessary to
`find optimal LED and PDs array
`to
`increase
`the
`measurement efficiency. This result will be applied to
`development of a wrist type healthcare device.
`
`Fig.6. PPG signal measured at ‘1’ in Fig. 5
`
`Fig.7. PPG signal measured at ‘2’ in Fig. 5
`
`Fig.8. PPG signal measured at ‘3’ in Fig. 5
`
`Fig.9. PPG amplitude versus pressure applied to the probe
`
`References
`[1] M. Malik, “Guidelines: Heart rate variability Standards
`of measurement, physiological
`interpretation, and
`clinical use”, Eur Heart J., vol. 17, pp354–38, 1996
`[2] A. B. Hertzman, “Photoelectric plethysmograph of the
`finger and toes in man”, Proc. Soc. Exp. Biol. Med., vol.
`37, pp. 1633-1637, 1937.
`[3] M. J Drinnan, J. Allen, and A. Murray, “Relation
`between heart rate and pulse transit time during paced
`respiration”, Physiol. Meas. vol. 22, pp425–432, 2001
`[4] K. Nakajima, T. Tamura, and H. Miike, “Monitoring of
`heart and respiratory rates by photoplethysmography
`using a digital filtering technique”, Med. Eng. Phys., vol.
`18, pp. 365-372, 1996.
`[5] A. Johansson, P. A. Oberg, and G. Sedin, “Monitoring
`of heart and respiratory rates in newborn infants using a
`new photoplethysmographic technique”, J Clin. Comp.,
`vol. 15, pp. 461- 467, 1999.
`[6] M. Nitzan, H. d. Boer, S. Turivnenko, A. Babchenko,
`and D. Sapoznikov, “Spontaneous oscillations in the
`peripheral circulation
`system, as measured by
`photoplethysmography”, Proc. SPIE, vol. 2328, pp. 188-
`195, 1994.
`[7] E. A. Lopez-Beltran, P. L. Blackshear, S. M. Finkelstein,
`and J. N. Cohn, “Non-invasive studies of peripheral
`vascular
`compliance
`using
`a
`non-occluding
`photoplethysmographic method”, Med. Biolog. Eng.
`Comp., vol. 36, pp. 748-753, 1998.
`[8] Y. Tardy, J. J. Meister, F. Perret, H. R. Brunner, and M.
`Arditi, “Noninvasive estimate of
`the mechanical
`properties of peripheral arteries from ultrasonic and
`photoplethysmographic measurements”, Clin. Phys.
`Physiol. Meas., vol. 12, pp. 39-54, 1991.
`[9] K. Meigas, R. Kattai, and J. Lass, “Continuous blood
`pressure monitoring using pulse wave delay”, presented
`at 23rd Annual International Conference of the IEEE
`Engineering in Medicine and Biology Society, Oct 25-
`28 2001, Istanbul, Turkey, 2001.
`[10] X. F. Teng and Y. T. Zhang, “Continuous and
`Noninvasive Estimation of Arterial Blood Pressure
`Using a Photoplethysmographic Approach”, presented
`at A New Beginning for Human Health: Proceedings of
`the 25th Annual
`International ESC DIVISION
`RESEARCH 2005 Department of EEE, Loughborough
`
`1131
`
`- 3 -
`
`
`
`University 42.
`[11] J. Naschitz, R. Itzhak, N. Shaviv, I. Khorshidi, S.
`Sundick, H. Isseroff, M. Fields, R. Priselac, D.
`Yeshurun, and E. Sabo, “Assessment of cardiovascular
`reactivity by fractal and recurrence quantification
`analysis of heart rate and pulse transit time”, J. Hum.
`Hypert., pp. 111-118, 2003.
`[12] M. Nitzan, A. Babchenko, B. Khanokh, and D. Landau,
`“The variability of the photoplethysmographic signal -
`a potential method for the evaluation of the autonomic
`nervous system”, Physiol. Meas., vol. 19, pp. 93-102,
`1998.
`[13] J.G.Webster, Design of Pulse Oximeters, UK, 1997.
`[14] Dassel ACM, Graaff R and Sikkema M, “Reflectance
`pulse oximetry at the forehead improves by pressure on
`the probe”, J. Clin. Monit., Vol. 11, No. 4, pp237-244,
`1995.
`
`[15] Y.Mendelson and M.J.McGinn, “Skin reflectance pulse
`oximetry : in vivo measurements from the forearm and
`calf”, J. Clin. Monit., Vol. 7, No. 1, pp. 7-12, 1991.
`[16] David Thompson and Steve Warren, “A Small, High-
`Fidelity reflectance Pulse Oximeter”, The 114th Annual
`ASEE Conference in Hawaii, AC2007-2420, 2007.
`[17] Y.Mendelson and C.Pujary, “Measurement Site and
`Photodetector Size Considerations
`in Optimizing
`Power Consumption of a Wearable Reflectance Pulse
`Oximeter”, The 25th Annual International Conference
`of the IEEE EMBS in Cancun, pp.3016-3019, 2003.
`
`
`