`Volume 14 Number 6 August 1998
`
`en a
`
`379-380 BriefReviews
`J. S. Gravenstein
`Original Articles
`381-384 INFLUENCE OF THE REFERENCE GAS OF PARAMAGNETIC
`OXYGEN ANALYZERS ON NITROGEN CONCENTRATIONS
`DURING CLOSED-CIRCUIT ANESTHESIA
`Janu FP. A. Hendrickx,MD, Andre A, J. van Zundert,MD,
`PhD, and Andre M. de Wolf,MD
`385-391 AVAILABILITY OF RECORDS INANOUTPATIENT =
`PREANESTHETIC EVALUATION CLINIC
`Gordon L, Gibby, MD, and Wilhem K. Schwab, PhD
`393-402 IN VIVOEVALUATION OF A CLOSED LOOP MONITORING
`STRATEGY FOR INDUCED PARALYSIS
`Deepak Ramakrishna, MS, Khosrow Belibehani, PhD, PE,
`Kevin Klein, MD, Jeffrey Mokhtar, MS,
`
`WolfW.vonMaltzalin,PhD,PE,RobertC.Eberhart,PhD,
`
`and Michael Dollar, MS
`403-412 REFLECTANCE PULSE OXIMETRY—PRINGIPLESAND =~
`OBSTETRIC APPLICATION IN THE ZURICH SYSTEM
`Volker Konig, Renate Huch, and Albert Huch
`413-420 ACCURACY OF VOLUME MEASUREMENTS IN MECHANIGALLY
`VENTILATED NEWBORNS:
`A COMPARATIVE STUDY OF COMMERCIAL DEVICES
`Kai Roske, Bertram Foitzik, Roland R. Wauer, and
`Gerd Schimatisch
`421-424 AN EVOLUTIONARY SOLUTION TO ANESTHESIA AUTOMATED
`RECORD KEEPING
`Alvin A. Bicker, PhD, John S, Gage, MD,and
`PaulJ. Poppers,MID
`425-431 EVALUATION OF APITOT TYPE SPIROMETER IN HELIUM/
`OXYGEN MIXTURES
`Soren Sondergaard, MD, Sigurbergur Karason, MD,
`Stefan Lundin, MD, and Ola Stenqvist, MD
`433-439 DETECTION OF LUNG INJURY WITH CONVENTIONAL AND
`NEURAL NETWORK-BASED ANALYSIS OF CONTINUOUS DATA
`Jukka Rasénen, MD, and Mauricio ¢A. Leon,MD
`Algorithm
`441-446 AREAL-TIME ALGORITHM TO IMPROVE THE RESPONSE TIME
`OF A CLINICAL MULTIGAS ANALYSER
`Lawdy Wong, MSc, Ruth Hamilton, MB, ChB,
`Eileen Palayiwa, PhD, and Clive Halin, D. Phil
`Book Review
`L. G6. HENSON AND A. C. LEE (EDS): SIMULATORS IN
`ANESTHESIOLOGY EDUCATION
`Richard Morris
`449-450 INFORMATION FORCONTRIBUTORS =
`
`~J
`
`447
`
`KLUWER ACADEMIC PUBLISHERS
`
`OFFICIAL JOURNAL OF THE EUROPEANSOCIETY
`FOR COMPUTING AND TECHNOLOGY
`IN ANAESTHESIA AND INTENSIVE CARE, AND OF
`THE SOCIETY FOR TECHNOLOGY IN ANESTHESIA
`
`(cid:44)
`
`JUL 3.0 1999
`
`
`HOY Hiikana iin,
`(cid:36)(cid:83)(cid:83)(cid:79)(cid:72)(cid:3)(cid:44)(cid:81)(cid:70)(cid:17)(cid:3)
`AppleInc.
`APL1014
`(cid:36)(cid:51)(cid:47)(cid:20)(cid:19)(cid:20)(cid:23)(cid:3)
`U.S. Patent No.8,989,830
`(cid:56)(cid:17)(cid:54)(cid:17)(cid:3)(cid:51)(cid:68)(cid:87)(cid:72)(cid:81)(cid:87)(cid:3)(cid:49)(cid:82)(cid:17)(cid:3)(cid:27)(cid:15)(cid:28)(cid:27)(cid:28)(cid:15)(cid:27)(cid:22)(cid:19)
`
`CODEN JCMOEH
`ISSN 1387-1307
`
`APPLE 1014
`
`1
`
`APPLE 1014
`
`

`

`
`
`JOURNAL OF CLINICAL MONITORING
`AND COMPUTING
`ee
`OFFICIAL JOURNAL OF THE EUROPEAN SOCIETY FOR COMPUTING AND TECHNOLOGY IN ANAESTHESIA AND INTENSIVE GARE (ESCTAIC)
`AND OF THE SOCIETY FOR TECHNOLOGY IN ANESTHESIA (STA)
`
`
`Volume 14 Number 6 August 1998
`
`THE PARENT JOURNALS
`International Journal of Clinical Monitoring and Computing,
`founded by OmarPrakash in 1984; Reed Gardner, senior
`editor; and the
`founded in 1985 by N. Ty
`Journal of Clinical Monitoring,
`Smith, Allen K. Ream, and J. S. Gravenstein.
`
`EDITORS
`Jeffrey M. Feldman, MD
`Allegheny University Hospitals, MCP
`Dept. of Anesthesiology
`3300 Henry Avenue
`Philadelphia, PA 19129, U.S.A.
`E-mail: Feldman]@auhs.cdu
`J. S. Gravenstein, MD, Coordinating Editor
`UF/College of Medicine
`Dept. of Anesthesiology
`P.O. Box 100254
`Gainesville, FL 32610, U.S.A.
`E-mail: jgravens(@anest2.anest.ufl.edu
`Ilkka Kalh, MD, PhD
`City Maternity Hospital
`Dept. of Anaesthesiology
`P.O. Box 61
`FIN-00971 Helsinki, Finland
`E-mail:
`ilkka.kalli@iki.Fi
`Publisher
`Kluwer Academic Publishers
`P.O, Box 990
`3300 AZ Dordrecht, The Netherlands
`E-mail: Loes.Wils@ wkap.nl
`
`
`EDITORIAL BOARD
`Paul G. Barash, MD
`~ Gerard Bashein, MD, PhD
`Charlotte Bell, MD
`Jan E. W. Beneken, PhD
`J. A. Blom, PhD
`J.G. Bovill, MD, PhD
`Peter Carstensen
`William P. Coleman, PhD
`Jeffrey B. Cooper, PhD
`John Doyle, MD
`Ronald Dueck, MD
`Peter C. Duke, MD
`
`ISSN 1387-1307
`All rights reserved
`
`© 1998 Kluwer Academie Publishers
`No part of the material protected by this copyright notice may be
`reproduced or utilised in any form or by any means, electronic or
`mechanical, including photocopying, recording or by any informa-
`tion storage and retrieval system, without written permission from
`the copyright owner.
`
`Printed in the Netherlands
`
`Printed on acid-free paper
`
`nif
`
`W CTA. pine
`Es T
`JU
`L 3 0 1999
`giro
`$
`Cea
`“MAUIOUN, Wi
`vane AY
`
`A. Dean Forbes
`
`Harvey L- Edmonds, MD
`John H, Eichhorn, MD
`TammyY. Euliano, MD
`M.Fisher, MD
`Wolfgang Friesdorf, MD
`Yasuhiro Fukui, PhD
`Leslie Geddes, PhD
`Nikolaus Gravenstein, MD
`A. Gerson Greenburg, MD, PhD
`G. Hedenstierna, MD
`Wolfgang Heinrichs, MD
`..C. G. Gallandat Huet, MD
`Kazuyuki Ikeda, MD
`Ville Jantti, MD
`Gavin N.C. Kenny, MD
`Samsun Lampotang, PhD
`J. W. R. McIntyre, MD
`James H. Philip, MD
`Michael L. Quinn, PhD
`Pradeep Ramaaya, MD
`Ira J. Rampil, MD
`Gunter Rau, MD
`Michael F. Roizen, MD
`J.-A. Romand, MD
`William B. Runciman, MD
`Keith Ruskin, MD
`David A. Sainsbury, MD
`H. Schillings, MD
`John W. Severinghaus, MD
`M. Michael Shabot, MD
`DeanForrest Sittig, PhD
`Bradley E. Smith, MD
`Richard Teplick, MD
`Kevin K. Tremper, MD
`Jorge Urzua, MD, PhD
`Karel H. Wesseling, PhD
`Dwayne Westenskow, PhD
`J.C. Wyatt, PhD
`
`VOLUME 14, 8 ISSUES, 1998
`Issues include:
`Brief Reviews, Editorials, Clinical Articles, Research Ar-
`ticles, Educational Articles, Book Reviews, Correspondence,
`Selections from the Anesthesia Patient Safety Foundation
`Newsletter.
`
`Kluwer Academic Publishers, Dordrecht, Boston, London
`
`Kluwer Academic Publishers Online
`Visit our website at http://www.wkap.ol
`
`The Journal of Clinical Monitoring and Computing is indexed in Index
`Medicus, Current Contents/Clinical Medicine, Excerpta Medica, Current
`Awareness in Biological Sciences, and the Biomedical Enginecring Citation
`Index, and is indexed/abstracted in Engineering Index, Science Citation
`Index, Expanded Research Alert, Medica Documentation Service.
`
`Il
`(cid:44)(cid:44)
`
`7
`
`2
`
`

`

`
`
`This material may be protected by Copyright law (Title 17 U.S. Code)
`
`403
`403
`
`
`
`REFLECTANCE PULSE OXIMETRY — PRINCIPLES AND
`OBSTETRIC APPLICATION IN THE ZURICH SYSTEM
`Volker Konig, Renate Huch, and Albert Huch
`
`Fromthe Perinatal Physiology Research Department, Department
`of Obstetrics, Zurich University Hospital, CH-8091 Zurich, Switzer-
`land.
`
`Received Nov13, 1997, and in revised form Jun 23, 1998, Accepted
`for publication Aug 4, 1998.
`
`Address correspondence to Volker Kénig, Perinatal Physiology Re-
`search Department, Department of Obstetrics, Zurich University
`Hospital, CH-8091 Zurich, Switzerland.
`E-mail: vkg@fhk.usz.ch
`
`Journal ofClinical Monitoring and Computing 14; 403-412, 1998,
`© 1998 Kluwer Academic Publishers. Printed in the Netherlands.
`
`Konig V, Huch R, Huch A. Reflectance pulse oximetry — principles
`and obstetric applicationin the zurich system.
`J Clin Monit 1998; 14: 403-412
`
`ABSTRACT. Transmission and reflectance are the two main
`modes of pulse oximetry. In obstetrics, due to the absence of a
`transilluminable fetal part for transmission oximetry, the only
`feasible option is the reflectance mode, in which sensor and
`detector are located on the same surface of the body part.
`However, none of the reflectance pulse oximeters developed
`for intrapartumuse are fully satisfactory, as indicated by the
`fact that none have entered routine use. We have designed,
`developed, constructed and tested a reflectance pulse oximeter
`with the possibility to adjust the electronic circuits and signal
`processing in orderto determine the effects of various param-
`eters on signal amplitude and wave-form and to optimize the
`sensitivity and spatial arrangementof the optical elements.
`Following an explanation of the principles of reflectance
`pulse oximetry, we report our experience with the design,
`development, construction and field-testing of an in-house
`reflectance pulse oximetry system for obstetric application.
`
`KEY WORDS. Oxygen saturation, reflectance pulse oximetry,
`intrapartumfetal monitoring.
`
`INTRODUCTION
`
`Pulse oximetry is the combination of spectrophotome-
`try and plethysmography. It permits rapid noninvasive
`measurement of arterial oxygen saturation with the
`added advantages of simple sensor application and direct
`measurement, requiring neither calibration nor pre-
`adjustment, Pulse oximeters are thus in widespread and
`fast-increasing use, ¢.g. in intensive care, anesthetics and
`neonatology [1]. All these applications employ “trans-
`mission” pulse oximetry, so called because the light
`used to determine blood oxygensaturation is “trans-
`mitted” from a light emitter on one side of the body
`part to a light receiver on the otherside; suitable sites
`are the fingers in adults or hands and feet in neonates or
`children, whichare said to be “transilluminated.”
`In obstetrics, fetal oxygen status during labor is a
`crucial parameter. However, no transilluminable fetal
`part is available. The only optionin this case is reflec-
`tance oximetry [2], using a sensor with its light emis-
`sion and detection elements on the same surface of
`the bodypart. Various types of such a reflectance pulse
`oximeter have been developed for intrapartumuse at
`various locations. However, for a wide variety of reasons,
`all are still experimental and not infull routine use [3-7].
`Basically a reflectance measurement can be achieved
`using planar sensors — which can be produced,
`for
`example, by modifying conventional transmission sen-
`sors — and a sensitive modern pulse oximeter. However,
`
`3
`
`

`

`404 Journal of Clinical Monitoring and Computing Vol 14 No6 August 1998
`
`such instruments come with a “black-box” microproces-
`sor-controlled mode of operation making constructional
`adjustments to the electronic circuits and signal process-
`ing virtually impossible. As a result, it becomes difficult
`to determine the effect of various parameters on signal
`amplitude and wave-form, optimize sensorsensitivity to
`light intensity and the arrangement of the optical ele-
`ments, and hence assess the dependenceof arterial oxygen
`saturation measurement on key physical, technical and
`above all physiological variables. This was the aim
`driving our decision to design, develop and construct
`in-house a system dedicated to obstetric applications.
`Following a brief review of the principles of pulse
`oximetry, we report our experience with the develop-
`ment of the new device,
`together with some field-
`testing results.
`
`Ep
`
`d(t)
`
`absorption coefficient of reduced hemoglobin
`(function of wavelength)
`time function of mean pulsatile change in artery
`thickness, with amplitude d = d(diastole) —
`d(systole)
`
`transmitted
`Measurement may be impaired by light
`directly from the light source to the receiver or light
`which does not pass througharterially perfused tissue.
`If this “direct light” Igi, is taken into account, Equation
`(1) changes to:
`
`I(t) = Ihissue * exp(—(SO2 - Expo +
`(1 _ $O2) -Enb) * d(t)) + lair
`
`(ia)
`
`where
`
`
`
`Tessue = Io° exp(—Erisue . s)
`
`PRINCIPLES OF PULSE OXIMETRY
`
`Sigial recording
`
`Light is absorbed on passing through matter. The de-
`gree of absorption depends onthe nature of the trans-
`illuminated material and the wavelength of the light
`employed. All optical techniques for determining arte-
`rial oxygen saturation use the marked difference in the
`absorption of red light between oxygenated and re-
`duced hemoglobin.
`The absorption of light passing through bone or
`nonpulsatile tissue is constant over time. Oxygenated
`and reduced hemoglobinin the arterial vascular bed, on
`the other hand, cause changes in absorption timed by
`the heart rate due to the pulsatile variation in artery
`thickness. The total intensity ofthe light after passing
`through tissue can be measured, for example, as the
`photocurrent I(t) of a photodiode, andis obtained from
`the Lambert-Beerabsorptionlawas:
`
`I(t) = Ip + exp(—Etisue 8) + exp(—(SO2 + Enpo +
`(1 — SOz) - Evy) - d(t))
`
`(1)
`
`where
`intensity of incident light
`Ip
`Etissue Mean absorption coefficient of tissue (function of
`wavelength)
`mean thickness of transilluminatedtissue
`5M
`SO oxygen saturation to be determined (= HbO/
`(HbO + Hb), i.e. ratio of oxygenated hemoglo-
`bin concentration to sum of oxygenated and
`reduced hemoglobin concentrations)
`Epo absorption coefficient of oxygenated hemoglobin
`(function of wavelength)
`
`Since the pulsatile component of the absorption ts at
`most a few percent,
`i.c, the exponent of the second e
`function in Equation (1)
`is very small, we can use the
`approximation:
`
`exp(x) =1+x for |x| <1
`
`to obtain the veryclose approximation:
`
`I(t) =
`.
`.
`Leissue * a _ (SOo *Enpo + (1 - $O2) *EHb * d(t)) + Tair
`
`(1b)
`
`This light intensity is measured in the photodiodes and
`can be broken downelectronically into two compo-
`nents, a time-independent signal
`
`DC = leissue + Laie
`
`with amplitude equal to the value of this signal
`
`de = DC = Trissue F lair
`
`(2)
`
`(3)
`
`and a signal which varies in time with the pulsatile
`changeinartery thickness
`
`AC = lrissue * (SO2 *Enbo + (1 ~ SOz) "ey tb) . d(t)
`
`(4)
`
`with amplitude
`
`ac = I(diastole) — I(systole) =
`Iissue * (SO2 - Ero + (1 — SO2) + Eu) «a.
`
`Theratio between the ac and de amplitudesis then
`
`4
`
`

`

`Konig et al: The Zurich Obstetric Reflectance Pulse Oximeter
`
`405
`
`r= ac/de a (Tsesne/ tissue + lair):
`(6)
`(SO- Enno + (1 — SO2) > erm) >
`this ratio r is
`In the case that “direct light” Ii, = 0,
`independent ofthe incident light intensity Ip and of the
`absorptionin the nonpulsating tissue valuelrissue?
`
`6
`
`However, despite various theoretical models [9, 10],
`the scatter coefficients of the various tissue types are
`not known with sufficient accuracy to permit exact
`calculation. Experimental calibration thus has to be
`performed bydirectly comparing the pulse oximeter
`readings witharterial blood sample values.
`
`r= ac/de =
`(SO2 + Enbo + (1 —SO2)- env) +d
`
`for lair =0.
`
`6a
`
`(s )
`
`Transmission pulse oximetry
`
`is then dependent only on the oxygen
`This ratio r
`saturation SQ> to be determined, the knownabsorption
`coefficients Epo and ey, and the meanpulsatile change
`d in the thickness of the arterial vessels in the trans-
`illuminated region.
`To eliminate this dependence on d, the measurement
`is performed at two wavelengths with maximally dif-
`fering absorptioncoefficients. On the assumption that
`the d values are the same for both wavelengths, we
`obtain a variable
`
`R
`
`ll
`
`ll
`
`Cred /Tir = (ac/de),.4/(ac/de);,
`
`(SO+ Ern0 + (1 — SO2) > Erb) sea /
`(SO- E140 + (1 — SO2) + Erb) ie
`
`(7)
`
`(7a)
`
`from which the unknown SQz is readily calculated
`without knowing theincident light intensity or tissue
`thicknesses.
`Calculation assumesthe following physical prerequi-
`sites:
`
`e No light must be measured that has not passed
`through the pulsatile vascular bed e.g. light passing
`directly fromlight source to receiver(Igir).
`e The pulsatile changes in artery thickness must be the
`same for both wavelengths,
`i.c. both wavelengths
`must transilluminate the sametissue region.
`© Valid measurement assumes that the pulsatile signal
`originates only from varying absorption byarterial
`oxygenated and reduced hemoglobin. Theresults are
`falsified by othercauses of pulsatile changesin optical
`thickness, e.g. hemoglobin derivatives, circulating
`pigments, pulsatile changes in thickness produced
`mechanically in nonarterially perfused tissue by car-
`diac action, and, above all, venous pulsation,
`e To simplify descriptionof the principle behind meas-
`urement and its limitations,
`the Lambert-Beer law
`was assumedvalid for the passage of light through
`tissue. However,as light is not only absorbedin tissue
`but also scattered, the lawis of limited applicability
`[8]. The exact absorption coefficients must be cor-
`rected by taking the scattering effect into account.
`
`The optical elements are located on opposite sides of a
`bodypart. The sensorsare applied mainlyto the fingers
`and toes. Ears and nose are used onlyrarely due to poor
`perfusion. In neonates the sensoris applied around the
`hand orfoot. This arrangement largely ensures that the
`optical paths are the same for both wavelengths. Never-
`theless, incorrect sensor attachment can give spurious
`results, e.g. if some of the transmitted light reaches the
`receiving diodes around the outside of a fingeras “direct
`light.”
`Signal magnitudes are an important determinant of
`measurement accuracy: in normal fingertips, the ratio
`of the signal due to absorption in pulsating blood (ac)
`to the signal due to absorption intotal tissue (dc), r =
`ac/de, is 0.02—0.05.
`
`Reflectance pulse oximetry
`
`In this method the light backscattered in the bodyis
`used to determine oxygen saturation. The optical ele-
`ments are thus located on the same plane on the same
`body surface. Reflection originates from nonhomo-
`geneityin the optical path, i.e. at the interfaces between
`materials with different refractive indices. This means
`that on physiological grounds,strong reflections can be
`expected on the entry of light into bone. The trans-
`illuminated tissue must also be well perfused to obtain
`as strong a signal as possible, Not all bodyparts are as
`well perfused as the fingers or hands, but an ac/dc ratio
`of (.001—0.005 can be achieved on the forehead. Perfu-
`sion is also good over the sternum. One method of
`signal enhancement is to heat the measurement site to
`induce hyperperfusion, which can safely be performed
`up to 42°C. A rubefacient, e.g. nicotinic acid (Rubri-
`ment), can also be applied to the measurement site.
`Theprincipal physical limitations are the following:
`
`© The sensor design must eliminate “direct light,”i.c.
`light passing directly from the light sources to the
`photodiodes orthatis onlyscattered in the outerpart
`of the skin.
`e The measured AC signals are some 10 times weaker
`
`5
`
`

`

`406 Journal of Clinical Monitoring and Computing Vol14 No 6 August 1998
`
`than in the transmission method. The conditions
`governing the heating of the light-emitting diodes
`(LED) limit
`the potential
`for producing stronger
`signals by increasing the incident light intensity: not
`only can high uncontrolled temperatures damage
`tissue at the measurement site, but the wavelength
`of the emitted light changes as the LEDs become
`warmer. For this reason the photodiode area must be
`as large as possible.
`As in the transmission mode, the principle of meas-
`urement is the determination of absorption, except
`that this now refers to incoming reflected light, The
`light path is less well defined than in transmission
`mode, and thus may differ between the two wave-
`lengths. The effective absorption coefficients of the
`calibration inserted in the Lambert-Beer law must be
`checked and if necessary corrected by comparison
`with photometrically measured arterial blood values.
`
`neonatal head. Experimentationled to the choice ofa
`vacuumsystem using sensors cast fromsilicone rub-
`ber, with a suction grooveforfixation, a guard ring
`against direct light, and a connector for a suction
`pump. The photoelectric componentsare identical in
`both types of sensor.
`e Some metal sensors were fitted with a resistance-wire
`heating coil of maximal output 200 mW to induce
`local hyperemia. The temperature was monitored by
`a negative temperature coefhicient (NTC)resistor in-
`corporatedin the sensor unit.
`
`Numerous types of sensor meeting the above require-
`ments were built. A sensor used for intrapartum meas-
`urements — cast from silicone rubber with suction
`channel and pump connection — is shown in cross-
`section in Figure 1. It is attached to the fetal head with a °
`vacuumof approximately 100 mbar.
`
`THE ZURICH REFLECTANCE PULSE OXIMETER
`
`Electronics
`
`Sensors
`
`In constructing planar reflection sensors, i.e. sensors in
`whichthe photoelectric emitting and receiving elements
`lie next to eachotherin virtually the same plane, special
`attention was paid to the following points (Figure 1).
`
`e For maximal independence from local tissue differ-
`ences, a radially-symmetric pattern was selected for
`the photoelements. The light source — a chip with
`two LEDs for the wavelengths red = 660 nm and
`infrared = 920 nm — was placed in the center of the
`sensor and surrounded bya radial photodiode array
`for detecting the reflected light. To obtain a good
`signal at minimal
`light
`intensity,
`the area of the
`photodiodeshad to be as large as possible. After some
`preliminary experiments,
`six BX33 photodiodes
`(Siemens) were used with a mean radius of 17 mm.
`Their connections are led outwards individually, so
`that the sensor as a whole remains operational if a
`wire breaks or an individual diode is lost. This arrange-
`ment gives an external sensor diameter of 22 mm.
`e A guard ring around the LEDs acts as a barrier to
`“direct light.” The sensor must also fit snugly to the
`skin to minimize the risk of ambient light reaching
`the photodiodes.
`fixation to the
`Unlike with transmission sensors,
`measurementsite can pose problems. Aluminum sen-
`sor units are readily fixed with double-sided adhesive
`ECG rings. However,
`these rigid sensor heads are
`unsuited to the small radii of curvature of the fetal/
`
`Initially, we decided to separate signal processing in the
`analog part before the analog/digital converter (ADC),
`including signal amplification, filtering, and separation
`into DCand ACsignals, all handled electronically ; the
`post-ADC digital part was handled by software which
`input the data, averaged and evaluated amplitudes, cal-
`culated saturation andheart rate, and produced a some-
`what complicated screen display. Now, using modern
`techniques, we have a systemin preparation in which a
`fast separate microprocessor unit handles most signal
`processing and digital filtering tasks.
`Figure 2 shows the block diagram of the current
`apparatus with the following individual units:
`
`e Time control of measurement
`Froma rectified mains power supplysignal (100 Hz),
`a phase-locked loop (PLL) — connected as a frequency
`multiplicr — generates a square-wave signal of 64
`kHz. Coupling to the line frequency eliminates data
`acquisition faults due to interferences with the line
`frequency. The 64 kHz from the PLL clock drives a
`* 7-bit counter that addresses an EPROM giving a data
`cycle of 1 kHz, 64 pulses long. The output pulses of
`the EPROM control
`the entire sequence of light
`emission, signal acquisition andsignal processing.
`e LED drive
`The oppositely poled red and infrared LEDs are
`located in the output circuit of a current-stabilized
`push-pull output stage. They are triggered bydigital
`signals from the contro] unit via analog switches at
`the 1 kHz sampling rate in the equally spaced se-
`
`6
`
`

`

`Kénig et al: The Zurich Obstetric Reflectance Pulse Oximeter
`
`407
`
`
`
`
`
`(4) Vacuum-tube
`Photodiodes
`(1)
`(s)
`Signal-cable
`Red and Infrared LED's
`(2)
`
` (G3) Light barrier
`Fig. 1. Cross-section througha silicone-rubberreflectionsensorforintrapartummeasurement o ‘oxygensaturation.
`
`7
`
`

`

`408 Journal ofClinical Monitoring and Computing Vol14 No6 August 1998
`
`
`
`
` Vacuum
`
`
`Pump
`and
`Control
`
`Pre-
`Amplifier
`
`
`
`S&H
`
` Filter
`| Dark-Curr.
`
`Subtract.
`
`
`
`
`
`Ext. Device
`
`
`Oscillator/Control Unit
`
`(HP, Nellcor)
`
`
`
`
`Fig. 2. Schematic representation of measurement electronics.
`
`quence: infrared-dark-red-dark (overall length: 1 ms).
`Light intensity is determined by the amplitude of the
`signal driving this output stage. Input signal intensity
`canbeselected in two ways:
`— Manually: The red and infrared LED intensities
`can be manually adjusted independently using two
`potentiometers (Helipot). This permits the use of
`any desired light intensity within the limits stipu-
`lated for test and research purposes.
`— Automatically: DC voltage as the input signal
`controls the LEDs so that the DC voltages at the
`computerinput for both wavelengths are 2.0 + 0.5
`V. Outside this range the control circuit changes
`the LED currents to reset the DC voltages to 2.0 V.
`This setting is used for normal clinical applications.
`To prevent skin damage from overheating even in
`the event of electronic component failure, maximal
`LEDintensityis limited by an electronic circuit.
`Input amplifier and sample-and-hold stages
`Using operational amplifiers the photocurrent sup-
`plied by the photodiodes is converted to.a voltage
`and then amplified. Six switch positions permit am-
`plifications of 50 mV/nA to 5 V/nA. Afterwards,
`three sample-and-hold (S&H)
`stages — switched
`bythe correspondingsignals from the control stage —
`resolve the signals into the three components infra-
`red, red and dark. The dark currents are then sub-
`tracted fromthe red andinfrared signals in a subtrac-
`tion stage which also eliminates small ambient light
`components that may have reached the photodiodes.
`
`e Filters
`Using low-passfilters the discrete-time signals at the
`S&H output are reconverted to continuous-time sig-
`nals and trimmed of high-frequency components
`using cighth-order Bessel
`filters with a cut-off
`frequency of 7 Hz. The DC components are then
`separated using a low-pass filter with a cut-off fre-
`quency of 0.1 Hz. The AC componentsare passed to a
`further Bessel high-pass 0.7 Hzfilter for separation of
`slow motionartifacts, and then to a 40-fold amplifi-
`cation stage. The four signals DC;,, DC,.g, ACj, and
`ACyeq are thus available at
`the output. As oxygen
`saturation is calculated from the ratios (ac/dc);, and
`(ac/de),eq, it was essential to ensure by careful com-
`ponent selection that the amplifications for the am-
`plification and separation stages were as near as possi-
`ble identical for both wavelengths.
`Patient insulation
`the LED controller output
`For patient protection,
`stage and photodiode input stage were electrically
`‘isolated from the mains using Burr & Brownisola-
`tion amplifiers. These units were powered by an
`insulated powerpack.
`Heating
`Sensorsincorporating aresistance coil were fitted with
`a precision controller stabilizing temperature within
`the range 38.0—41.0 °C to an accuracy of 0.1 °C.
`Vacuum pump
`The sensoris fixed to the skin using a small pump
`producing a maximal -300 mbar vacuum. The pump
`
`8
`
`

`

`Konig etal: The Zurich Obstetric Reflectance Pulse Oximeter
`
`409
`
`
`
`60 min
`O sec File: LAOBLY22.PK2
`Start: 23. 1.97
`B8:19°521 Tot
`RPOXLL
`RPOX
`$o2 %
`LxxxeMMM Axxx geb. 20.03.63 US Datum 23.
`
`
`
`
`
`- aoormocooococo
`
`
`
`RPOX HR bpm
`
`
`
`File:
`
`
`
`
`
`Fig. 3. Screen display of a 10-minute measurement. Fromtop to bottom: oxygen saturation, heart rate, heart rate from HP CTGmonitor,
`uterine contraction from HP CTG monitor. Underneath the DC and AC signals for infrared andred light for the last 5 seconds (note the
`different ordinate scalesfor DC and AC).
`
`is maintained electronically via a manometer at a
`preset negative pressure. In normal medical use, ad-
`equate fixation is achieved with a vacuum of about
`—100 mbar.
`
`Softivare
`
`Apart from varioustest programs, a programfor data
`acquisition, display, calculation and storage and a pro-
`gram for subsequent data postprocessing were written
`in PASCAL.
`
`e Signal acquisition, calculation and display
`For data acquisition a 486 DOS computer was
`equipped with a 12-bit analog/digital conversion
`
`(ADC) card (Metrabyte). A counter on this card
`triggers analog/digital conversion at 400 Hz, which
`starts an interrupt programin the computerfor read-
`ing the 4 measured values DCj,, DCreq and AC;,,
`ACyea into a cyclic buffer of 5-second length. From
`the signal (AC;,+AC,.4) the maxima and minimaare
`determined for each cardiac cycle, and hence the
`instantaneous heart rate. Oxygen saturations are cal-
`culated from the amplitudes ac and de of the AC and
`DCsignals.
`Saturation is calculated from the Lambert-Beer
`law using the absorption coefficients [11]
`for
`the
`nominal wavelengths of our LEDs. However,
`as
`the actual wavelengths may deviate from these nom-
`inal values and the applicability of the Lambert-Beer
`law is limited by scattering in tissue, experimental
`
`9
`
`

`

`410 Journal of Clinical Monitoring and Computing Vol 14 No6 August 1998
`
`calibration of the measured saturation valuesis essen-
`tial.
`Saturation and heart rate can be averaged over 1-9
`cardiac cycles and are displayed every second. Analog
`heart rate and uterine contraction signals from any
`CTG monitor with analog output
`(e.g. Hewlett
`Packard model 8040) can be input to the ADC every
`second and likewise displayed.
`The measurement display (Figure 3) shows, along the
`bottom, the 4 measured values DC;,,DCyq and AC},
`ACyea, over the last 5 seconds of measurement. This
`serves to monitor signal quality during acquisition.
`Poor-signal periods can be marked and excluded from
`data processing. Along the top, measured arterial
`oxygensaturation andheartrate values per second are
`displayed cyclically over a 10-minute interval, with
`the CTG heart rate and contraction input under-
`neath. Any time point can be marked for subsequent
`identification and all values and comment stored ina
`file at any time.
`e Data analysis
`The values from a stored file can be redisplayed in
`measurement mode using an evaluation program.
`Timeintervals can be marked with the arrow keys or
`mouse. Means and standard deviations — including
`the CTG data — are then calculated and displayed,
`
`
`MEASUREMENTS AND DISCUSSION
`
`Clinical application of any newinstrument or measure-
`ment system presumes:
`
`— mechanicalreliability, accuracy andcalibration,
`— feasibility in the clinical situation, including accept-
`ance by both medical personnel and patients,
`the ability not only to determine physiological pa-
`rameters not previously measured in both physiolog-
`ical and pathological situations, but also to evaluate
`the diagnostic significance of such parameters,in this
`case oxygensaturation.
`Forthe first two more technical points is to say, that
`our instrument required calibration before clinical
`use, together withfield tests of the suction device and
`long-term oximeter performance during birth.
`For
`these points controls and clinical
`trials were
`performed in our own unit and with colleagues in
`Copenhagen (DK), Graz (A), Oulu (SF) and Berlin
`(D). The majorinvestigations comprised:
`
`Calibration
`
`To calibrate a pulse oximeter, an oxygen saturation
`value must be assigned to the measured variables
`
`R = trea/Tir = (ac/de),.4/(ac/de),,
`
`cf (7)
`
`on the basis of an experimental or theoretical relation-
`ship. Initially we used the absorption coefficients of
`Zijlstra et al [1i]. The general problems ofcalibrating a
`pulse oximeter have been discussed elsewhere [12, 13].
`For fine calibration we performed the following inves-
`tigations:
`
`e Tests in the arterial oxygen saturation range 88—
`100% were conducted in 14 healthy adult voluntcers
`breathing normal air and then air with approxi- ~
`mately 80% normal oxygen content for 10 minutes
`in each case. The reflection sensor was fixed to the
`forehead or sternumwith an adhesive ring. Owing to
`the invasive nature of arterial catheterization, refer-
`ence values were provided by a MINOLTA PUL-
`SOX 8 transmission pulse oximeter attached to the
`index finger. Data analysis [13] showed a 4.5% differ-
`ence in oxygen saturation between the MINOLTA
`and the preliminary results of ourreflectance system
`based onthe absorptioncoefficents of Zijlstra et al.
`e At lowersaturation levels, measurements were per-
`formed in cyanotic children before surgery. The chil-
`dren had arterial lines, permitting direct comparison
`with arterial blood readings[14].
`@ Lowsaturations in vive can also be measured in the
`fetal lamb [15]. We used this method to compare our
`pulse oximeter readings directly with arterial values
`in the oxygensaturation range 10-80% [16].
`
`Preliminary evaluation of these data shows that our
`previous calculations of oxygen saturation have to be
`corrected in the 10-100% saturation range by a factorof
`1.045.
`
`Fixation
`
`Thefirst experiments in sensorfixation to the head and
`other parts of the human body were performed in
`adults [17] and neonates[18].
`Attachmentis simple in practice, even during birth.
`After rupture of the membranes, the sensor canbe fixed
`to the fetal head once the cervix has dilated to at least 2
`em. Initial fixation takes 30-60 seconds and allowsfull
`freedom of movement[19].
`Approximately 100 mbaris the most suitable pressure
`
`10
`
`10
`
`

`

`at which to maintain the sensor as it ensures good
`sensor-skin contact and reliable continuous fetal oxygen
`saturation values. Measurements at twodifferentsites of
`the head of the fetus have shownno significant differ-
`ences in the oxygen saturation values,
`indicating no
`effect of suction on the local blood supply. The same we
`have observedin the case of caput succedaneum[20]. °
`
`Long-term performance
`
`Vacuumfixation did not impair oxygen measurement,
`even overfixation periods of 4.3 hours [20], Suction
`marks had disappeared normally within 10 to 20 mi-
`nutes after removal of the sensor for neonates [18] and
`for fetuses.
`
`Heating
`
`in the
`AC signals at 41.0°C are twice as strong as
`unheated state [21]. However, the silicone rubber sen-
`sors required for vacuumfixation in fetal monitoring
`cannot be fitted with a

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket