Electronics 2014, 3, 282-302; doi:10.3390/electronics3020282
`
`OPEN ACCESS
`
`electronics
`
`ISSN 2079-9292
`www.mdpi.com/journal/electronics
`
`Review
`Wearable Photoplethysmographic Sensors—Past and Present
`Toshiyo Tamura 1,*, Yuka Maeda 2, Masaki Sekine 1 and Masaki Yoshida 1
`
`1 Osaka Electro-Communication University, Faculty of Biomedical Engineering /18-8, Hatsu-Cho,
`Neyagawa, Osaka 572-8530, Japan; E-Mails: m-sekine@isc.osakac.ac.jp (M.S.);
`yoshida@isc.osakac.ac.jp (M.Y.)
`2 Faculty of Engineering, Information and Systems, University of Tsukuba/1-1-1, Tennodai,
`Tsukuba, Ibaraki 305-8573, Japan; E-Mail: maeda@iit.tsukuba.ac.jp
`
`* Author to whom correspondence should be addressed; E-Mail: tamurat@isc.osakac.ac.jp;
`Tel.: +81-72-824-1131; Fax: +81-72-824-0014.
`
`Received: 25 February 2014; in revised form: 15 April 2014 / Accepted: 18 April 2014 /
`Published: 23 April 2014
`
`Abstract: Photoplethysmography (PPG) technology has been used to develop small,
`wearable, pulse rate sensors. These devices, consisting of infrared light-emitting diodes
`(LEDs) and photodetectors, offer a simple, reliable, low-cost means of monitoring the
`pulse rate noninvasively. Recent advances in optical technology have facilitated the use of
`high-intensity green LEDs for PPG, increasing the adoption of this measurement
`technique. In this review, we briefly present the history of PPG and recent developments in
`wearable pulse rate sensors with green LEDs. The application of wearable pulse rate
`monitors is discussed.
`
`Keywords: photoplethysmography; pulse rate; reflectance; transmittance; green light;
`infrared light; adaptive filter; least mean square algorithm
`
`1. Introduction
`
`It is important to monitor the perfusion of the circulation. The most important cardiopulmonary
`parameter is blood pressure, but monitoring it is complicated. A second important parameter is blood
`flow, which is related to blood pressure. We can monitor the blood perfusion in large vessels using
`ultrasound devices, but it is not practical to use these routinely. Several devices for monitoring blood
`MASIMO 2006
`Apple v. Masimo
`IPR2020-01526
`
`

`

`283
`
`Electronics 2014, 3
`
`perfusion have been developed [1], but, unfortunately, it is difficult to find a practical device.
`However, the perfusion of blood flow and blood pressure can be determined easily using a pulse
`rate monitor.
`Wearable pulse rate sensors based on photoplethysmography (PPG) have become increasingly
`popular, with more than ten companies producing these sensors commercially. The principle behind
`PPG sensors is optical detection of blood volume changes in the microvascular bed of the tissue. The
`sensor system consists of a light source and a detector, with red and infrared (IR) light-emitting diodes
`(LEDs) commonly used as the light source. The PPG sensor monitors changes in the light intensity via
`reflection from or transmission through the tissue. The changes in light intensity are associated with
`small variations in blood perfusion of the tissue and provide information on the cardiovascular system,
`in particular, the pulse rate. Due to the simplicity of this device, wearable PPG pulse rate sensors have
`been developed. This review describes the basic principles of PPG, previous and current developments
`in wearable pulse rate monitors with a light source, and the elimination of motion artifacts.
`
`2. Photoplethysmography (PPG)
`
`2.1. Principle of PPG
`
`The principle of PPG has been reviewed previously [2–4], and is explained briefly here. Light
`travelling though biological tissue can be absorbed by different substances, including pigments in the
`skin, bone, and arterial and venous blood. Most changes in blood flow occur mainly in the arteries and
`arterioles (but not in the veins). For example, arteries contain more blood volume during the systolic
`phase of the cardiac cycle than during the diastolic phase. PPG sensors optically detect changes in the
`blood flow volume (i.e., changes in the detected light intensity) in the microvascular bed of tissue via
`reflection from or transmission through the tissue.
`Figure 1 shows an example of a photoplethysmographic waveform, consisting of direct current
`(DC) and alternating current (AC) components. The DC component of the PPG waveform corresponds
`to the detected transmitted or reflected optical signal from the tissue, and depends on the structure of
`the tissue and the average blood volume of both arterial and venous blood. Note that the DC
`component changes slowly with respiration. The AC component shows changes in the blood volume
`that occurs between the systolic and diastolic phases of the cardiac cycle; the fundamental frequency of
`the AC component depends on the heart rate and is superimposed onto the DC component.
`
`Figure 1. Variation in light attenuation by tissue.
`
`
`
`
`
`

`

`Electronics 2014, 3
`
`2.2. Light Wavelength
`
`284
`
`The interaction of light with biological tissue can be quite complex and may involve scattering,
`absorption and/or reflection. Anderson and Parrish examined the optical characteristics and penetration
`depth of light in human skin [5]; within the visible region, the dominant absorption peak corresponded
`to the blue region of the spectrum, followed by the green-yellow region (between 500 and 600 nm)
`corresponding to red blood cells. The shorter wavelengths of light are strongly absorbed by melanin.
`Water absorbs light in the ultraviolet and longer IR regime; however, red and near-IR light pass easily.
`As a result, IR wavelengths have been used as a light source in PPG sensors.
`Blood absorbs more light than the surrounding tissue. Therefore, a reduction in the amount of blood
`is detected as an increase in the intensity of the detected light. The wavelength and distance between
`the light source and photodetector (PD) determine the penetration depth of the light. Green light is
`suitable for the measurement of superficial blood flow in skin. Light with wavelengths between 500
`and 600 nm (the green-yellow region of the visible spectrum) exhibits the largest modulation depth
`with pulsatile blood absorption. IR or near-IR wavelengths are better for measurement of deep-tissue
`blood flow (e.g., blood flow in muscles). Thus, IR light has been used in PPG devices for some time [6].
`However, green-wavelength PPG devices are becoming increasingly popular due to the large intensity
`variations in modulation observed during the cardiac cycle for these wavelengths [7–9].
`A green LED has much greater absorptivity for both oxyhaemoglobin and deoxyhaemoglobin
`compared to infrared light. Therefore, the change in reflected green light is greater than that in
`reflected infrared light when blood pulses through the skin, resulting in a better signal-to-noise ratio
`for the green light source.
`Several green-light-based photoplethysmographs are available commercially. For example, MIO
`Global has developed the MIO Alpha in cooperation with Philips; this measures the electrocardiogram
`(ECG) with 99% accuracy, even while cycling at speeds of up to 24 kmph [10]. For daily use, Omron
`has developed a green light pulse rate monitor (HR-500U, OMRON, Muko, Japan).
`Furthermore, the use of video cameras using the signal based on the red green blue (RGB) colour
`space has been considered, as shown in Section 3.3. The green signal was found to provide the
`strongest plethysmographic signal among camera RGB signals [11,12]. Haemoglobin absorbs green
`light better than red and green light penetrates tissue to a deeper level than blue light. Therefore, the
`green signal contains the strongest plethysmographic signal.
`
`2.3. Reflected and Transmitted Signals
`
`The wearable PPG has two modes—transmission and reflectance—as shown in Figure 2. In
`transmission mode, the light transmitted through the medium is detected by a PD opposite the LED
`source, while in reflectance mode, the PD detects light that is back-scattered or reflected from tissue,
`bone and/or blood vessels.
`
`
`
`

`

`Electronics 2014, 3
`
`
`285
`
`Figure 2. Light-emitting diode (LED) and photodetector (PD) placement for transmission-
`and reflectance-mode photoplethysmography (PPG).
`
`
`
`The transmission mode is capable of obtaining a relatively good signal, but the measurement site
`may be limited. To be effective, the sensor must be located on the body at a site where transmitted
`light can be readily detected, such as the fingertip, nasal septum, cheek, tongue, or earlobe. Sensor
`placement on the nasal septum, cheek or tongue is only effective under anesthesia. The fingertip and
`earlobe are the preferred monitoring positions; however, these sites have limited blood perfusion. In
`addition, the fingertip and earlobe are more susceptible to environmental extremes, such as low
`ambient temperatures (e.g., for military personnel or athletes in training). The greatest disadvantage is
`that the fingertip sensor interferes with daily activates.
`Reflectance mode eliminates the problems associated with sensor placement, and a variety of
`measurement sites can be used (as discussed in the following section). However, reflection-mode PPG
`is affected by motion artifacts and pressure disturbances. Any movement, such as physical activity,
`may lead to motion artifacts that corrupt the PPG signal and limit the measurement accuracy of
`physiological parameters. Pressure disturbances acting on the probe, such as the contact force between
`the PPG sensor and measurement site, can deform the arterial geometry by compression. Thus, in the
`reflected PPG signal, the AC amplitude may be influenced by the pressure exerted on the skin.
`Reflectance PPG sensors such as the MaxFast (Nellcor™, Mansfield, MA, USA) have been used
`clinically to measure continuous oxygen saturation non-invasively. Anecdotally, it gives false-positive
`readings occasionally. Further research is needed in this area.
`
`3. Factors Affecting PPG Recordings
`
`Previous research has identified several factors that affect PPG recordings, including the measurement
`site (i.e., probe attachment site), the contact force, mechanical movement artifacts, subject posture, and
`breathing, as well as ambient temperature. This chapter briefly discusses these factors.
`
`3.1. Measurement Site of Probe
`
`The location of the LED and PD is an important design issue that affects the signal quality and
`robustness against motion artifacts. Therefore, suitable measurement sites must be located to optimize
`sensor performance. PPG sensors are commonly worn on the fingers due to the high signal amplitude
`that can be achieved in comparison with other sites [13]. However, this configuration is not well suited
`to pervasive sensing, as most daily activities involve the use of the fingers.
`In recent years, different measurement sites for PPG sensors have been explored extensively,
`including the ring finger [14], wrist [15,16], brachia [9,17,18], earlobe [19–21], external ear
`cartilage [22–24], and the superior auricular region [25–27]. In addition, the esophageal region has
`been used in clinical practice [28–30]. Commercial clinical PPG sensors commonly use the finger,
`earlobe and forehead [31]. In addition, use of a glass-type wireless PPG has been examined [32].
`
`
`

`

`Electronics 2014, 3
`
`
`286
`
`The perfusion values of 52 anatomical sites in healthy subjects showed that the fingers, palm, face,
`and ears offer much higher perfusion values compared with other measurement sites [33]; the transmitted
`PPG signal amplitude from the earlobe provides the largest perfusion value. In addition, earlobe
`sensors are easy to fabricate, and have become popular as pulse rate monitors (Table 1). However, a
`spring-loaded ear-clip, although effective, can become painful over extended monitoring periods.
`There was little improvement in the wearable earlobe PPG sensor design until the development of
`micro-electromechanical system (MEMS) technology. MEMS facilitated the fabrication of a
`lightweight, comfortable, fully integrated, self-contained sensor earpiece. For example, an earring PPG
`sensor with magnetic attachment to the earlobe was developed that allowed good contact for
`monitoring during physical activity [22].
`
`Table 1. Key features of a wearable ear photoplethysmography (PPG) device.
`
`In-mount
`Features
`Year published 2009 [20]
`Auditory
`canal
`Otoplastic
`insertion
`
`Sensing site
`
`Probe
`attachment
`Wireless
`communication
`Motion
`cancellation
`
`CUHK
`2008 [26]
`Inferior
`auricle
`
`e-AR
`2007 [19]
`
`Superior auricle
`
`Imperial
`2009 [27]
`Superior
`auricle
`
`Earhook
`
`Earhook
`
`Tape
`
`MIT
`2010 [22]
`
`Earlobe
`
`Magnetic
`earring
`
`MIT
`2012 [24]
`External ear
`cartilage
`
`Earphone
`
`Pulsear
`2004 [23]
`External ear
`cartilage
`Earcup
`headphones
`
`Yes
`
`Yes
`
`No
`
`No
`
`Yes
`
`Yes
`
`None
`
`None
`
`Passive motion
`cancellation
`
`None
`
`
`cancellation
`
`Automatic noise Automatic noise
`cancellation
`
`No
`
`PCA
`
`Samsung
`2009 [21]
`
`Earlobe
`
`Spring-loaded
`clip
`
`No
`
`Automatic noise
`cancellation
`
`Earphone/earbud PPG sensors are also available and provide greater comfort for the user. In this
`design, a reflective photosensor is embedded into each earbud, as shown in Figure 3. The sensor
`earbuds are inserted into the ear and positioned against the inner side of the tragus to detect the amount
`of light reflected from the subcutaneous blood vessels in the region. The PPG sensor earbuds look and
`work like a regular pair of earphones, requiring no special training for use [24].
`
`Figure 3. Earpiece PPG sensor (with permission [24]).
`
`
`
`
`
`

`

`Electronics 2014, 3
`
`
`287
`
`A headset with an ear-clip, transmission-type PPG sensor allows continuous, real-time monitoring
`of heart rate while listening to music during daily activities. In addition, the proposed headset is
`equipped with a triaxial accelerometer, which enables the measurement of calorie consumption and
`step-counting. However, over the course of a variety of daily activities (e.g., walking, jogging, and
`sleeping), the PPG sensor signal may become contaminated with motion artifacts [20].
`The most common commercially available PPG sensor is based on finger measurement sites. Finger
`sites are easily accessed and provide good signal for PPG sensor probes. For example, a ring sensor
`can be attached to the base of the finger for monitoring beat-to-beat pulsations. Data from the ring
`sensor are sent to a computer via a radiofrequency transmitter, as shown in Figure 4. To minimize
`motion artifacts, a double ring design was developed to reduce the influence of external forces,
`acceleration and ambient light, and to hold the sensor gently and securely to the skin, so that the blood
`circulation in the finger remained unobstructed. Experiments have verified the resistance of the ring
`sensor to interfering forces and acceleration acting on the ring body. Benchmark testing with
`FDA-approved PPG and ECG sensors revealed that the ring sensor is comparable in the detection of
`beat-to-beat pulsations, despite disturbances [14,34].
`
`Figure 4. PPG ring sensor.
`
`
`
`Wristwatch-type pulse oximetry and blood pressure sensors have been developed and
`commercialized by several companies. These devices, although much easier to wear, are not usually
`used in clinical settings, due to several technical issues. However, a novel PPG array sensor module
`with a wristwatch-type design has been developed. The proposed module measures the PPG signal
`from the radial artery and the ulnar artery of the wrist, whereas previous methods obtained signals
`from the capillaries in the skin. Phototransistors and IR-emitting diodes were placed in an array format
`to improve the PPG signal sensitivity and level of accuracy. Various arrays were considered for
`optimization. A conductive fiber wristband was used to reduce external noise. In the experiments, the
`proposed module was assessed and compared with the commercially available product produced by
`BIOPAC [16].
`A reflective brachial PPG sensor has also been examined. Although the pulse amplitude is lower
`than those from the finger and earlobe, the PPG pulse waveforms from regions in the vicinity of a
`human artery could be detected and measured easily [17].
`Forehead sensors have shown greater sensitivity to pulsatile signal changes under low perfusion
`conditions, compared with other peripheral body locations [31]. The thin-skin layer of the forehead,
`coupled with a prominent bone structure, helps to direct light back to the PD. Sensor placement
`on the forehead has been shown to result in decreased motion artifacts during certain types of
`physical activity.
`
`
`
`

`

`Electronics 2014, 3
`
`
`288
`
`Glass-type PPG sensors have also been investigated. A reflectance mode-PPG sensor, equipped
`with an accelerometer for detecting kinetic signals and a wireless controller for transmitting both
`signals to the receiver installed on the frame of the glass, was used to obtain PPG data from a user
`performing alternating sitting and sit-to-stand motions—the correlation between the peak-to-peak
`intervals in the signals of a BIOPAC device and the developed device was 97.5% and 87% for sitting
`and the sit-to-stand motion, respectively, given 100% data transmission without error [32].
`
`3.2. Probe Contact Force
`
`In reflectance-type PPG, the PPG signal waveform may be affected by the contact force between
`the sensor and the measurement site. The waveform of the obtained PPG signal differs depending on
`the PPG probe contact pressure. A number of studies have suggested that the PPG waveform coincides
`with arterial stiffness and vascular reactivity. Several studies have shown that moderate pressure on the
`sensor can improve the PPG signal. Ideally, the best PPG signal can be obtained under conditions of
`transmural pressure, defined as the pressure difference between the inside and outside of the blood
`vessel (i.e., the pressure across the wall of the blood vessel). Insufficient pressure results in inadequate
`contact and consequently low AC signal amplitude. However, PPG signal recording under excessive
`pressure conditions can also lead to low AC signal amplitude, as well as distorted waveforms caused
`by the occluded artery beyond the PPG probe. Optimal contact pressure corresponds to the maximal
`pulsatile amplitude; this occurs when the transmural pressure approaches zero (i.e., under maximal
`arterial compliance).
`Despite numerous attempts, no generally accepted standards have been adopted for clinical or
`fundamental PPG measurements of contact pressure. Changes in the AC pulse amplitude, DC
`amplitude, ratio of AC/DC amplitudes and normalized pulse area of the reflected PPG signal have
`been investigated [35]. In one study, a contact force ranging from 0.2 to 1.8 N was applied to the
`finger [36]. As a result, the AC amplitude increased and then decreased with increasing contact force.
`For different arterial stiffness with gender and age, the pulse amplitude and the AC/DC ratio peaked at
`different contact forces ranging from 0.2 to 1.0 N; most subjects achieved their maximum pulse
`amplitude within 0.2–0.4 N. Note that a calibration load of 0.65 N was applied to the fingertip skin to
`provide a uniform pressure force during the experiment. A similar contact force study was conducted
`on the upper arm [37]; an average compression pressure of 30 mmHg (4 kPa) produced a PPG peak
`amplitude for IR and green light sources.
`Studies have shown that contact pressures ranging from 8 to 12 kPa (60–90 mmHg) resulted in the
`largest PPG amplitude for a reflectance sensor attached to the forehead region above the eye, although
`the signal-to-noise ratio (SNR) did not improve significantly [38]. Note that in SpO2 studies, the ratio
`between the relative change in the red and IR light intensities (R/IR) was used to indicate accurate
`oxygen saturation; R/IR values varied from 0 to 80 mmHg for the forehead of a newborn infant, with a
`portion of the PPG signal derived from tissue underneath the skull [39–41].
`Transmission-type PPG also affects the transmural pressure. Increasing the amplitude of arterial
`pulsations improves the PPG SNR. For example, when measuring the pulsatile amplitude of a
`finger-based PPG sensor, as the cuff pressure increases, the PPG amplitude increases and then
`decreases to zero when the arterial blood vessel becomes occluded. The largest PPG amplitude (i.e.,
`
`
`
`

`

`289
`
`Electronics 2014, 3
`
`approaching zero transmural pressure), in general, corresponds to the mean arterial pressure. However,
`this pressure is too high for application over any significant period. To prevent the collapse of capillary
`beds, the cuff pressure must be on the order of 10 mmHg, which is too low to obtain sufficient PPG
`amplitude. Thus, even for transmission-type photoplethysmographs, care must be taken in how the
`sensor is attached to the finger.
`In one particular study, blood pressure and finger PPG waveforms were simultaneously measured in
`healthy volunteers under different contact pressures ranging from 0 to 200 mmHg [42]. The results of
`PPG regression indicated that the amplitude ratio was highest for a contact pressure of 60 mmHg.
`The optimal contact pressure of the probe is important, particularly with regard to arterial stiffness.
`Arterial stiffness, estimated from the PPG pulse-wave second-derivative parameter, b/a, is strongly
`inconsistent when recorded at non-optimal probe contact pressures. To determine the optimal contact
`pressure, one group obtained in vivo PPG readings from conduit artery sites in five healthy subjects
`recorded with probe contact pressures ranging from 0 to 15 kPa [43]. In this study, an 880-nm
`reflectance-type sensor and force transducer derived the optimal contact pressure criterion from the
`PPG AC signal area; the b/a values showed high repeatability (coefficient of variability < 5%) for a
`contact pressure of 10 kPa (75 mmHg), indicating optimal contact [43].
`
`3.3. Video Plethysmography
`
`A video camera can be used as a PD; this is termed video plethysmography. There are two types of
`video plethysmography. One is non-contact for long distance measurements. This uses a charge
`coupled device (CCD) camera, such as a mobile phone camera or web camera, and examines the light
`reflected from skin. Plethysmographic signals can be measured remotely (>1 m) using ambient
`light and PD with either a simple consumer-level digital camera in movie mode or a cellular
`phone [11,44–46]. Although the green channel features the strongest plethysmographic signal,
`corresponding to an absorption peak by (oxy-) hemoglobin, the red and blue channels also contain
`plethysmographic information. These results suggest that ambient light PPG may be useful for medical
`purposes, such as the characterization of vascular skin lesions or remote sensing of vital signs for
`triage or sports purposes. The other is the contact type for short-distance measurements, such as
`placing the finger on the mobile phone camera to cover the entire camera view and using the phone
`camera flash LED light (white light) to illuminate the finger [47]. This type of PPG measurement
`method has been commercialized by Azumio [48].
`
`3.4. Signal Processing
`
`3.4.1. Introduction
`
`Motion artifacts mainly consist of random low-frequency interference. Therefore, most artifact
`reduction is accomplished by signal processing. The signal-processing algorithm basically assumes
`that the original PPG signal has power only at certain frequencies and that the rest is noise. Several
`signal processing techniques can be applied, including those that use referencing from an acceleration
`signal or those that minimize the motion artifact with synthetic noise generation. However, attempts to
`minimize motion artifacts reported to date do not appear to correlate well with real-world noise
`
`
`
`

`

`Electronics 2014, 3
`
`sources. Moreover, studies have shown that a high degree of randomness is not necessarily correlated
`with true motion artifacts in PPG signals. This section briefly reviews current signal processing
`methods used to reduce motion artifacts.
`
`290
`
`3.4.2. Moving Average Filter
`
`The moving average method is commonly used to reduce motion artifacts and works well for a
`limited artifact range. However, this method does not account for sudden changes. The periodic
`moving average filter (PMAF), based on the quasi-periodicity of the PPG signals, segments the PPG
`signal into periods and resamples each period. Thus, with PMAF, the motion artifacts are removed
`without degrading the signal [49].
`In-band noise occurs when the spectra of the motion artifact and that of the PPG signal overlap
`significantly. However, it is not advisable to use fixed-frequency filtering techniques to eliminate
`motion artifacts due to in-band interference and the overlapping frequency spectra in the PPG signal.
`Instead, motion artifacts can be removed using a filter bank and a matched filter consisting of several
`frequency bands [50]. In this case, the adaptive filter exhibited many variations and errors due to
`amplitude variations during convergence when measuring PPG in real-time, whereas the moving
`average filter exhibited a more stable output. Compared with traditional adaptive filtering methods,
`the ratio variation was 50% lower than that of the moving average filter, allowing more stable
`measurement of oxygen saturation, despite the patient’s movement [50].
`
`3.4.3. Fourier Analysis
`
`A Fourier series is applicable only to periodic signals, and, therefore, cannot be directly applied to a
`PPG signal, which is non-stationary and quasi-periodic. To overcome this problem, Fourier series
`analysis can be applied on a cycle-by-cycle basis. In this case, the acquired data are first filtered using
`a Savitzky-Golay (SG) smoothing filter to remove high-frequency noise. Once the noise has been
`removed, the proposed cycle-by-cycle Fourier series (CFS) analysis is carried out, and the PPG signals
`(IR and red) are reconstructed on a cycle-by-cycle basis. Experimental results have verified the
`efficacy of the proposed method [51]. Moreover, the results showed that artifacts induced by patient
`movement were attenuated by at least 35 dB, reducing the measurement error of the PPG signal
`from 37% to 3% using the technique described above.
`
`3.4.4. Adaptive Filter
`
`The basic function of a filter is to remove the unwanted signal from the signal of interest. A
`commonly used filtering technique separates the signal from noise using the peak frequencies of the
`signal and artifact. In addition, the mean square of the error signal (the difference between the desired
`response and actual response of the filter) can be minimized. However, when the peak frequency of the
`pulse rate is similar to the noise frequency, the pulse rate cannot be separated from the noise. In PPG
`measurements, active noise cancellation with acceleration-based adaptive filters has been attempted by
`several groups (Figure 5). Obtaining the best design usually requires a priori knowledge of certain
`statistical parameters (such as the mean and correlation functions) within the useful signal. With this
`
`
`
`

`

`Electronics 2014, 3
`
`information, an optimal filter can be designed that minimizes the unwanted signals according to
`statistical criteria.
`
`291
`
`Figure 5. Block diagram of an adaptive filter with an accelerometer.
`
`
`
`Adaptive noise cancellation (ANC) is an alternative technique used to estimate signals corrupted by
`additional noise or interference. The adaptive filter is inherently self-designed through the use of a
`recursive algorithm that updates the filter parameters. This approach can be used to obtain the desired
`level of noise rejection, without a priori estimates of the signal or noise. Two inputs are required for
`this method: (i) the primary input containing the corrupted signal and (ii) a reference input containing
`some potential noise correlation to the primary noise; note that an acceleration sensor can be
`used as the reference input. Figure 5 shows a motion-tolerant wearable biosensor using a MEMS
`accelerometer [52,53], which operates based on adaptive noise cancellation utilizing an acceleration
`reference. This development has motivated new and improved PPG sensor design.
`
`3.4.4.1. Least Mean Square Adaptive Algorithm
`
`The least mean square (LMS) adaptive algorithm removes motion artifact noise by estimating the
`synthetic noise reference signal and adapting the filter coefficients based on filter order. LMS
`algorithms are a class of adaptive filters used to mimic the desired filter by finding the filter
`coefficients that relate to producing the least mean square of the error signal.
`Many adaptive techniques have been applied to the reduction of motion artifacts from PPG signals,
`including the normalized least mean squares (NLMS) method, recursive least squares (RLS) filter,
`time varying step-size LMS (TVS-LMS), adaptive step-size LMS (AS-LMS), and Kalman filters. A
`LMS filter with automatic step-size control was used to mitigate the effects of motion artifacts in PPG
`recordings for long-term patient monitoring [54]. The experimental results indicated that the proposed
`variable step-size LMS filter provided better performance than the LMS filter with a fixed step-size.
`In another study, a two-dimensional (2-D) active noise cancellation algorithm was applied to
`compensate for motion-distorted signals using directional accelerometer data [55]; a NLMS adaptive
`filter (fourth-order) was used in the algorithm, resulting in a reduction in signal distortion from 52.34%
`to 3.53% over the frequency range of 1–2.5 Hz (the frequency range for daily motion, such as walking
`and jogging). In addition, this wearable health-monitoring device, equipped with a motion artifact
`
`
`
`

`

`292
`
`Electronics 2014, 3
`
`reduction algorithm, can be integrated as a terminal in a so-called ubiquitous healthcare system, to
`provide continuous health monitoring without interrupting daily life.
`In another study, a triaxial MEMS accelerometer was attached to a PPG sensor to detect wrist
`movement [56]. In this case, a fast transversal RLS algorithm was used to reduce the computational
`complexity of the adaptive filter, by providing an estimate of the linear motion-to-artifact transfer
`function. The resulting output represented an estimation of the noise, which was then minimized by the
`RLS algorithm. Experimental results showed that this device produced more reliable signals that were
`stable against motion artifact corruption under typical types of movement, such as the swinging of
`the arms [56].
`A Laguerre series was implemented to compactly represent the system dynamics for joggers using a
`few parameters, such as the heart rate [57]. This study determined that the standard artifact reduction
`scheme does not work when the physiological signal is correlated with wearer motion. Thus, adaptive
`blind-source separation techniques can be used to recover the physiological signal. However, the
`success of this method is currently limited.
`In one study, the correlations among six representative daily motions and patterns of motion
`artifacts were examined to obtain more precise analysis [58]. An artifact reduction algorithm was
`designed for the analysis using the motion data. In addition, in this study, the short operating times and
`small number of variables for the portable device required the use of an LMS adaptive filter. The
`results using this filter were compared with those using other noise reduction algorithms. For validation
`purposes, a real-time motion artifact reduction experiment was performed using a wearable device
`during the motions. The results indicated that the motion artifacts of wearable PPG devices can be
`effectively reduced using the proposed method and that such devices can potentially be used in daily
`life with any type of motion.
`The TVS-LMS algorithm offers a fast convergence rate. However, performance studies have shown
`that the AS-LMS algorithm provides not only a fast convergence rate, but also minimal mean square
`error (MSE), as indicated by the high SNR values.
`A simple, efficient approach, based on the AS-LMS adaptive filter, was applied to reduce motion
`artifacts in corrupted PPG signals [59]. In this study, a synthetic noise reference signal, representing
`motion artifact noise, was generated internally from the motion artifact-corrupted PPG signal itself,
`and therefore no additional hardware (e.g., accelerometer or source-detector pair) was used. The
`generated noise reference signal was then filtered through the AS-LMS adapt

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket