throbber
JOURNAL OF THE EUROPEAN SOCIETY FOR
`
`THERAPEUTIC RADIOLOGY AND ONCOLOGY
`
`IQXDIOTHERAPY
`ONCOLOGY
`
`Page 1 of 15
`
`ELSEVIER
`
`ViewRay Exhibit 1024
`
`ViewRay Exhibit 1024
`Page 1 of 15
`
`

`

`International Standard Serial Number 0167-8140
`
`
`
`© 1999 Elsevier Science Ireland Ltd. All rights reserved.
`
`This journal and the individual contributions contained in it are protected under copyright by Elsevier Science Ireland Ltd., and the
`following terms and conditions apply to their use:
`Photocopying
`
`Single photocopies of single articles may be made for personal use as allowed by national copyright laws. Permission of the publisher and
`payment of a fee is required for all other photocopying, including multiple or systematic copying, copying for advertising or promotional
`purposes, resale, and all forms of document delivery. Special rates are available for educational institutions that wish to make photocopies
`for non—profit educational classroom use.
`
`Permissions may be sought directly from Elsevier Science Rights & Permissions Department, PO Box 800, Oxford OX5 1DX, UK; Tel.:
`(+44) 1865 843830, fax: (+44) 1865 853333, e-mail: pennissions@elsevier.co.uk. You may also conlact Rights & Permissions directly
`through Elsevier’s homc page (http://www.elsevier.nl), selecting first ‘Custunier Support'. then General Information’, then ‘Pcrmissions
`Query Form’.
`
`In the USA, users may clear permissions and make payment through the Copyright Clearance Center, Inc., 222 Rosewood Drive, Danvers,
`MA 01923, USA; Tel.: (+1) (978) 7508400, fax: (+1) (978) 7504744, and in the UK through the Copyright Licensing Agency Rapid
`Clearance Service (CLARCS), 90 Tottenham Court Road, London WlP OLP, UK; Tel.: (+44) 171 631 5555; fax: (+44) 171 631 5500.
`Other countries may have a local reprographic rights agency for payments.
`Derivative Works
`
`Subscribers may reproduce tables of contents or prepare lists of articles including abstracts for internal circulation within their institutions.
`Permission of the publisher is required for resale or distribution outside the institution.
`
`Permission of the publisher is required for all other derivative works, including compilations and translations.
`Electronic Storage or Usage
`
`Permission of the publisher is required to store or use electronically any material contained in this journal, including any article or part of an
`article.
`
`Except as outlined above. no part of this publication may be reproduced, stored in a retrieval system or transmitted in any form or by any
`means, electronic. mechanical, photocopying. recording or otherwise. without prior written permission of the publisher.
`Address permissions requests to: Elsevier Science Rights & Permissions Department, at the mail, fax and e—mail addresses noted above.
`Notice
`
`No responsibility is assumed hy the publisher for any injury andfor damage to persons or property as a matter of products liability,
`negligence or otherwise, or from any use or operation of any methods, products, instructions or ideas contained in the material herein.
`Because of rapid advances in the medical sciences, in particular, independent verification of diagnoses and drug dosages should be made.
`Although all advertising material is expected to conform to ethical (medical) standards, inclusion in this publication does not constitute a
`guarantee or endorsement of the quality or value of such product or of the claims made of it by its manufacturer.
`@ The paper used in this publication meets the requirements of ANSI/NISO 239.48—1992 (Permanence of Paper).
`Printed in the UK
`
`Orders, claims, and product enquiries: please contact the Customer Support Department at the Regional Sales Office
`nearest you:
`
`[Note (Latin America): for orders, claims and help desk information, please contact the Regional Sales Office in New York as listed above]
`
`New York
`Elsevier Science
`PO. Box 945
`New York, NY 10159-0945
`USA
`Tel : (+l)(212) 633 3730
`[toll free number for Nonh
`American customers:
`1-888-4ES-[NFO (437-4636)]
`Fax: (+1)(212) 633 3680
`c—mail: usinfo-f@elsevicr,com
`
`Amsterdam
`Elsevier Science
`PO. Box 211
`1000 AE Amsterdam
`The Netherlands
`Tel.: (+3l)(20) 4853757
`Fax: (+31)(20) 4853432
`e—mail: nlinfo-f@elsevier.nl
`
`Tokyo
`Elsevier Science
`9-15 Higashi—Azabu l—chome
`Minato—ku, Tokyo 106-0044
`Japan
`Tel.: (+81)(3) 5561 5033
`Fax: (+81)(3) 556] 5047
`e—mail: info@elsevierrco.jp
`
`Rio de Janeiro
`Singapore
`Elsevier Science
`Elsevier Science
`Run Sete de Setembro 111/16 Andar,
`No. l Tcmasek Avenue
`200507002 Ccntro,
`#17-01 Millenia Tower
`Rio de Janeiro -RJ,
`Singapore 039192
`Brazil;
`Tel.: (+65)434 3727
`Tel.: (+55) (21) 509 5340
`Fax: (+65)337 2230
`e—mail:'asiainfo©elsevier.com sg Fax: (+55) (21) 507 I991
`e—mail: elsevier@campus.com.br
`
`ViewRay Exhibit 1024
`Page 2 of 15
`
`ViewRay Exhibit 1024
`Page 2 of 15
`
`

`

`Radiotherapy and Oncology, 1999, Volume 50, Number 3, March, pp. 247—378
`
`CONTENTS
`
`Cited in: Chemical Abstracts. Excerpta Medica (EMBASE), Current Contents (Clinical Medicine; Life Sciences), Index Medicus (MEDLINE), Current
`Awareness in Biological Sciences (CABS)
`
`Review article
`Altered fractionation: limited by mucosa] reactions?
`J.H.A.M. Kaanders, A.J. van der Kogel, K. Kian Ang (The Netherlands, USA)
`
`Original articles
`Modulation of accelerated repopulation in mouse skin during daily irradiation
`K—R. Trott, A. Shirazi, F. Heasman (UK)
`
`Loco-regional recurrences after mastectomy in breast cancer: prognostic factors and implications for postoperative irradiation
`J.J. Jager, L. Volovics, L.J. Schouten, J.M.A. de Jong, P.S.G.J. Hupperets. M.F. von Meyenfeldt, B. Schutte, G.H. Blijham (The Netherlands)
`
`Acute and late morbidity of using a breast positioning ring in women with large pendulous breasts
`GE. Home], L.B. Marks, CS. Whiddon, L.R. Prosnitz (USA)
`
`Evaluation of predictive factors for looal tumour control after electron-beani—rotation irradiation of the chest wall in locally advanced breast cancer
`T. Hahr, W. Budach, F. Paulsen, C. Gromoll, G. Christ, M. Bamberg (Germany)
`
`The use of compensators to optimise the three dimensional dose distribution in radiotherapy of the intact breast
`L.J. Can-uthers, A.T. Redpath, I.H. Kunkler (UK)
`
`Clinical delivery of intensity modulated conformal radiotherapy for relapsed or second-primary head and neck cancer using a multileaf collimator
`with dynamic control
`W. Dc Neve. W. Dc Gersem, S. Dcrycke, G. De Meerleer, M. Moerman. M.—T. Bate, B. Van Duyse. L. Vakaet, Y. De Deene, B. Mersseman, C. De
`Waeter (Belgium)
`
`A conformal technique for a ring shaped conjunctive lymphoma treatment (Technical note)
`R. Arrans, S. Alonso, F. Sénchez—Doblado, J.A. Sénchez-Calzado, A. Leal. M. Perucha (Spain)
`
`A single»variable method for the derivation of collimator scatter correction factors in symmetrical and asymmetrical X-ray beams (Technical note)
`J.L.M. Venselaar, N. Beckers (The Netherlands)
`
`Commissioning of a micro multi-lcaf collimator and planning system for stereotactic radiosurgery
`V.P. Cosgrove, U. Jahn, M. Pfaender, S. Bauer, V. Budach, R.E. Wurm (Germany)
`
`Daily positioning accuracy of frameless stereotactic radiation therapy with a fusion of computed tomography and linear accelerator (focal) unit:
`evaluation of z-axis with a z—rnarker (Technical note)
`M. Ucmatsu, M. Sonderegger, A. Shioda, K. Tahara, T. Fukui, Y. Hama, T. Kojima. J.R. Wong. S. Kusano (Japan, USA)
`
`Small-field fractionated radiotherapy with or without stereotactic boost for vestibular schwannoma
`K. Kagei, H. Shirato, K. Suzuki, T. Isu, Y. Sawarnura, T. Sakamoto, S. Fukuda, T. Nishioka, S. Hashimoto, K. Miyasaka (Japan)
`
`Seminorna of the testis: is scrotal shielding necessary when radiotherapy is limited to the para—aortic nodes‘.7
`S. Bieri, M. Rouzaud, R. Miralbell (Switzerland)
`
`Characteristics and clinical application of a treatment simulator with Ct—option
`D. Verellen, V_ Vinh—Hung, P. Bijdckerke, F. Nijs, N. Linthout, A. Bcl, G. Storme (Belgium)
`
`Three-dimensional movement of a liver tumor detected by high—speed magnetic resonance imaging
`S. Shimizu. H. Shirato. B. Xo, K. Kagei, T. Nishioka, S. Hashimoto, K. Tsuchiya, H. Aoyama, K. Miyasaka (Japan)
`Errata
`ESTRO Courses
`ESTRO Meetings
`Calendar of events
`
`247
`
`261
`
`267
`
`277
`
`283
`
`291
`
`301
`
`315
`
`319
`
`325
`
`337
`
`341
`
`349
`
`355
`
`367
`
`371
`375
`375
`376
`
`
`
`_—_—_-_—__—.—_.—_.—_—_— available on the Elsevier Science website at www.clsevier.nl or www.elsevier.com or www.clscvier.co.jp
`
`This journal is part of ContentsDirect, thefree alerting service which sends tables of contents by e—mail for Elsevier
`Science books and journals. The quickest way to register for ContentsDirect is via the World Wide Web at:
`www.elsevier.nl/locate/ContentsDirect
`If you don’t have access to the WWW you can register for this service by sending an e—mail message to
`cdsubs@elsevier.co.uk specifying the title of the publication you wish to register for. The tables of contents are also
`
`
`ViewRay Exhibit 1024
`Page 3 of 15
`
`ViewRay Exhibit 1024
`Page 3 of 15
`
`

`

`
`The material may be protected by Copyright law {Title 17 u.s. Code)
`
`
`
`
`
`
`ELS EVI E R
`
`
`
`MDIOTHERAPY
`& ONCOLOGY
`
`Radiotherapy and Oncology 50 (1999) 325—336
`
`Commissioning of a micro multi-leaf collimator and planning system for
`stereotactic radiosurgery
`
`Vivian P. Cosgrove*, Ulrich Jahn, Mathias Pfaender, Susanne Bauer, Volker Budach,
`Reinhard E. Wurm
`
`Klinikfiir Strahlemherapie, Universiu'irsklinikmn Charite’, Schumann Strafle 20721, 10117 Berlin, Germany
`
`Received 15 July 1998; received in revised form 9 November 1998; accepted 28 December 1998
`
`
`Abstract
`
`Purpose: A computer controlled micro multi-leaf collimator, m3 mMLC, has been commissioned for conformal, fixed-field radiosurgery
`applications. Measurements were made to characterise the basic dosimetric properties of the m3, such as leaf transmission, leakage and beam
`penumbra. In addition, the geometric and dosimetric accuracy of the m3 was verified when used in conjunction with a BrainSCAN v3.5
`stereotactic planning system.
`Materials and methods: The m3 was detachably mounted to a Varian Clinac 2100C accelerator delivering 6 MV X—rays. Leaf transmis-
`sion, leakage, penumbra and multiple, conformal fixed field dose distributions were measured using calibrated film in solid water. Beam data
`were collected using a diamond detector in a scanning water tank and planned dose distributions were verified using LiF TLDs and film. A
`small, shaped phantom was also constructed to confirm field shaping accuracy using portal images.
`Results: Mean transmission through the closed multi—leaves was 1.9 i 0.1% and leakage between leaves was 2.8 i 0.15%. Between
`opposing leaves abutting along the central beam-axis transmission was ~ 15 i 3%, but was reduced to a mean of 4.5 i 0.6% by moving the
`abutmen position 4.5 cm off-axis. Beam penumbrae were effectively constant as a function of increasing square field size and asymmetric
`fields and was seen to vary non—linearly when shaped to diagonal, straight edges. TMR, OAR and relative output beam data measurements of
`circular m3 fields were comparable to conventional, circular stereotactic collimators. Multiple, conformal field dose distributions were
`calculated with good spatial and dosimetric accuracy, with the planned 90% isodose curves agreeing with measurements to within 1—2 mm
`and to i 3% at isocentre. Portal films agreed with planned beams eye-view field shaping to within 1 mm.
`Conclusions: The m3 micro multi—leaf collimator is a stable, high precision field-shaping device suitable for small—field, radiosurgery
`applications. Dose distributions can be accurately calculated by a planning system using only a few beam data parameters. © 1999 Elsevier
`Science Ireland Ltd. All rights reserved.
`
`Keywords: Micro multi—leaf collimator; MLC; Conformal therapy; Dosimetry; Radiosurgery planning
`
`
`1. Introduction
`
`From the earliest clinical applications of stereotactic
`radiosurgery and radiotherapy for the treatment of intra—
`cranial pathological processes it has been the primary
`goal to improve therapeutic ratio. This involves maximis—
`ing the radiation dose to damage or destroy a lesion while
`minimising the dose to nearby, uninvolved, normal tissue.
`Stereotactic
`radiosurgery using linear
`accelerators
`is
`conventionally carried out with single or multiple isocen—
`tre, non—coplanar, arcing techniques for radiation delivery.
`As an alternative, non-coplanar,istatic, conforrnally shaped
`beams can be employed [2,21,22,29,32,33,39,41,42].
`
`* Corresponding author. Joint Department of Physics, Royal Marsden
`NHS Trust, Downs Road, Sutton, Surrey SM2 SPT, UK.
`
`Conformal radiosurgery or radiotherapy involves accu-
`rately tailoring a dose of radiation to match the shape of a
`target volume in three dimensions. Conformal beam shap-
`ing has been shown to offer a number of advantages over
`conventional, arcing radiosurgery using single and multiple
`isocentres. This includes improved normal tissue sparing
`(especially for larger, irregularly shaped target volumes),
`more homogeneous dose distributions throughout the target
`volume and reduced treatment times [2,22,29,32,39]. A
`report by Nedzi et a1. [30] indicated that radiosurgery treat-
`ment complications were associated With multiple isocenr
`tre treatments, particularly for larger target volumes, due to
`tumour dose inhomogeneity.
`Multi—leaf collimators (MLC) are currently regarded as
`the state—of-the-art device for producing arbitrary, irregu—
`larly shaped radiation fields. MLC design and use are now
`
`0167-8140/99/$ - see front matter © 1999 Elsevier Science Ireland Ltd. All rights reserved.
`PH: SOl67—8140(99)00020-1
`
`ViewRay Exhibit 1024
`Page 4 of 15
`
`

`

`326
`
`V.P. Corgmve et al. / Radiotherapy and Oncology 50 (1999) 325—336
`Table 1
`Designspecifications of the m3 micro multi-leaf collimator
`
`Number of leaves
`Leaf width (at isocentre)
`
`26 Pairs
`14 X 3.0 mm
`6 X 4.5 mm
`6 X 5.5 mm
`10.2 X 10 ‘cm
`Maximum field size
`5 cm
`Maximum leaf over-travel
`31 cm
`Clearance from isocentre
`1.5 cm/s
`Maximum leaf speed
`
`Weight 30 kg
`
`leaves have this shaft
`inserted into each leaf. Adjacent
`inserted at vertical increments to permit optimum position—
`ing of each leat‘s driving motor. Single, 10 mm diameter
`motors are used to drive each micro—leaf. Leaf ends are
`
`milled to three angled straight edges, each. covering a
`third of the leaf edge length. These edges correspond to
`the divergence of the beam when the leaf is fully extended
`(—5 cm), centred (0 cm) and fully retracted (+5 cm). The
`vertical central edges allow opposing leaves to meet when
`closed. Each leaf is also shaped in cross-section to match
`the change in divergence of the beam across the field area.
`The maximum square field area that can be defined at
`isocentre is 10.2 x 10 cm2.
`Monitoring of the position of each leaf is carried out
`using two independent methods. Primary positional feed—
`back is derived from the rotation of the motor shaft to each
`
`leaf, with the number of turns of each motor being related
`to leaf displacement. Secondary feedback derives from two
`mechanical brushes physically mounted on and along the
`longitudinal plane of each leaf. Together these ensure
`precise (to 0.1 mm) leaf positioning.
`the m3 is manually
`With the gantry turned to 180°,
`transferred to the accessory mount of the linac by means
`of a specially designed trolley. The mounting and initiali—
`sation procedure
`is
`straightforward and is usually
`
`+—120m
`
`
`
`.
`
`i
`
`Iii/~30
`
`—r
`
`*-—-1.9mm
`
`'
`
`'
`
`ill
`
`-—~
`
`4—2.1‘mm
`
`(b)
`
`6cm
`
`‘
`
`|I|
`
`.‘I
`
`l
`
`,..,,,,,
`
`fl ,1
`
`(a)
`
`'
`
`Fig. 1. Schematic diagram of a multi—leaf in (a) plane View and (b) cross—
`section.
`
`ViewRay Exhibit 1024
`Page 5 of 15
`
`Well established [4,9,l3,l4,l9,20,23,25,34]. The advantage
`of using automatic field shaping devices over conventional
`lead alloy shielding blocks has also been documented
`[10,11,16,26]. More recently, MLCs have been shown to
`have the additional potential for modulating the intensity of
`radiation across a field [1,6,7,15,24,40,44].
`Accurate conformation of dose to small
`
`intracranial
`
`targets (up to 5 cm in cross—section) using conventional
`multi-leaf collimators is impeded by the relatively large
`leaf width at isocentre (usually ~ 1 cm). Linear accelera-
`tor based stereotactic radiosurgery and radiotherapy of
`these small targets have generally been carried out using
`lead alloy blocks or manually shaped miniature multi-
`leaves [2,12,27,37]. Between four and eight static beams
`are generally used, shaped with 3D ‘beams—eye view’
`image displays. The process of treatment delivery can be
`very time consuming and labour intensive in terms of block
`construction or manual field shaping, verification and treat—
`ment delivery. There has, therefore, been much interest in
`developing an automated, miniature multi-leaf collimator
`to improve dose conformation to small target volumes as
`well as take advantage of the known labour and time
`saving attributes of a computer controlled field shaping
`device.
`
`Here we report on the dosimetric measurements, initial
`acceptance testing and commissioning of the m3 computer
`controlled micro multi-leaf collimator (m3). The measure—
`ments were principally used to verify that
`the unique
`features of the m3, such as leaf design, positioning and
`monitoring,
`fulfilled the demands
`for high precision,
`stereotactic radiosurgery applications. Beam data were
`then collected and transferred to a BrainSCAN v3.5 stereo-
`tactic planning system and used to calculate conformal,
`multiple-field,
`non-coplanar
`treatment plans. Further
`measurements were carried out
`in phantom to compare
`and verify the calculated and delivered dose distributions.
`
`2. Materials and methods
`
`2.1. m3 Design features
`
`The m3 micro multi—leaf collimator investigated in this
`work (a joint development project between BrainLAB
`GmbH, Germany, and Varian Inc., USA) is based upon
`the architecture of a standard Varian MLC [3,14,20]. It
`has 52 tungsten leaves (26 pairs), which move perpendicu—
`larly to the beam central axis (i.e. unfocussed). However,
`unlike the standard Varian MLC that has a constant leaf
`width of 1 cm at isocentre, the m3 has variable leaf widths
`(see Table 1). The finer, 3 mm wide leaves, located in the
`centre of the field area, allow improved shaping around
`smaller targets, of the size treated by radiosurgery.
`Leaf design differs from the standard Varian MLC (see
`Fig.
`la,b). A more complicated ‘tongue and groove’ leaf
`cross—section was necessary to allow drive shafts to be
`
`ViewRay Exhibit 1024
`Page 5 of 15
`
`

`

`V.P. Cosgrove et al. / Radiotherapy and Oncology 50 (1999) 325—336
`
`327
`
`completed within 5—10 min. The linac used during this
`work has a standard 52 leaf Varian MLC already built
`into the gantry head. For all measurements the MLC leaves
`were fully retracted so as not to interfere with the primary
`beam.
`
`2.2. Dosimetry tools
`
`In common with other work, radiotherapy verification
`film (X-OMAT V2, Kodak, Inc, Rochester, NY, USA)
`was used for dosimetry measurements [3,13,14,19,20].
`All
`irradiated films were scanned using a computer
`controlled digital densitometer (FIPS Plus laser' Scanner,
`PTW GmbH, Freiburg, Germany). The minimum resolu—
`tion of this laser densitometer is 0.3 mm. Film optical
`density was calibrated to absolute dose using a set of
`films irradiated in a standard calibration geometry (10 X
`10 cm square field, 6 MV X—rays, films at isocentre and
`SSD = 98.5 cm). This was repeated whenever films from a
`new batch of 50 were used. All measurements were carried
`out using 6 MV X-rays.
`Measurements of beam data parameters used to calculate
`stereotactic radiosurgery dose distributions requires cham—
`bers or detectors with dimensions small enough to resolve
`steep dose gradients and avoid the problems associated
`with the lack of lateral electronic equilibrium of very
`narrow beams
`[35,38]. Therefore, beam data were
`measured in a scanning water tank using a diamond detec-
`tor (PTW Gmbl-I, Freiburg, Germany). The properties of
`the diamond detector make it ideal for high precision small
`field dosimetry [17,36].
`Similarly, measurements of absolute dose were made
`using small, 0.5 X 6 mm LiF therrnoluminescent detectors
`(TLDs). Irradiated TLDs were read out on a Rialto (NE
`Technology, UK) automatic reader. Before use, TLDs were
`individually calibrated in a 60C0 therapy beam using a
`traceably calibrated 0.3 cm3 PTW ionisation chamber.
`TLDs with a reproducibility of s3% were selected for
`the measurements.
`
`2.3. Leaf transmission, leakage and penumbra
`
`Measurements of leaf leakage and transmission were
`carried out using films positioned perpendicularly to the
`beam central axis in Solid Water (Gammex, RMT, USA)
`at a depth of 1.5 cm (dmax), SSD = 98.5 cm and with the
`primary jaws of the linac set to 10 X 10 cm. The leaves
`were fully closed so that opposing leaves abutted either
`along the central beam axis or 4.5 cm off-axis (i.e. a
`bank of leaves was set to over-travel
`the central beam
`
`axis by 4.5 cm). Films were scanned with the densitometer
`to produce profiles across the closed leaves and at the leaf
`ends. Scans were normalised to the output measured in a
`10 X 10 cm2 square field without the m3 attached to the
`linac. Ten times the monitor unit settings (500 MU) were
`used compared to open field films to increase the intensity
`and distribution statistics of the transmitted X-rays. (Film
`
`was calibrated over a large dose range (0—3 Gy) to ensure
`all optical density values would be accounted for.)
`80% to 20% beam penumbra was first measured as a
`function of square field size ranging from 2 X 2 cm2 to
`10 X 10 cm2 in steps of ~l cm. (The exact size of each
`square field was defined taking into account the combina-
`tions of the variable leaf widths.) Films were also placed at
`dmu, SSD : 98.5 cm, in solid water and profiles across the
`irradiated films were scanned both perpendicular and paral—
`lel
`to leaf motion (i.e. across leaf sides and leaf ends,
`respectively).
`Beam penumbra as a function of asymmetric leaf posi—
`tioning was investigated by defining a 2 X 10 cm rectan-
`gular field shape with the m3 leaves. (The 2 cm field width
`was parallel to the direction of motion of the leaves). Three
`films were simultaneously exposed, positioned at dmx, 5
`cm and 10 cm deep in a solid water phantom, SSD = 95
`cm. The field shape was then moved asymmetrically in 1—
`cm steps away from the beam central axis so that the final,
`fifth field shape was centred 4 cm from the central axis,
`with one set of leaves fully retracted {—5 cm), the other set
`extended by +3 cm. A total of five sets of three films were
`irradiated and beam profiles were measured on each film,
`parallel to the direction of motion of the leaves.
`The variation in the effective beam penumbra was also
`investigated as a function of straight but diagonal field
`edges. Two irregular field shapes were created by fitting
`leaves to a number of diagonal, straight field edges with
`differing angles to the horizontal (leaves were fitted so that
`the mid-point of each leaf-end intersected with the diago-
`nal field edge). Films were irradiated at dmflx, SSD = 98.5
`cm and 80, 50 and 20% isodose distributions measured. For
`all penumbra measurements the primary jaws were fixed at
`a field size of 10 X 10 cm2.
`
`2.4. Mechanical stability of the linac with the m3 attached.
`
`Measurements were made to investigate whether the 30
`kg m3 causes any additional gantry or collimator sagging
`when attached to the linac gantry head. To do this, a test
`identical in principle to the isocentre verification procedure
`described by Winston and Lutz for stereotactic radiosur-
`gery applications was used [43]. A 3 mm diameter tungsten
`ball, supported on a rigid, Perspex rod, was fixed at the
`radiation isocentre as defined by the treatment room lasers
`(this reference point is not strictly a true radiation isocen-
`tre, which can move due to mechanical instability of the
`linac. It is instead a mean point defined after taking into
`account observed field displacements as a function of
`gantry, collimator and table angle). The gantry was then
`moved in steps of 45° and at each angle three separate port
`films were exposed with the collimator turned to angles of
`0°, 90° and 270°. On development of the films the displa-
`cement of the field—edges relative to the tungsten ball was
`measured both with and without the m3 installed. Fields
`
`were either a 3 cm diameter circular field defined by the
`
`ViewRay Exhibit 1024
`Page 6 of 15
`
`ViewRay Exhibit 1024
`Page 6 of 15
`
`

`

`328
`
`V.P. Cosgrave et al. /Radiotherapy and Oncology 50 (1999) 325—336
`
`m3 (primaryjaws at 5 X 5 cm2) or a 3 X 3 cm2 square field
`defined by the primary jaws alone. The light—field cross-
`hairs were marked on all film exposures and were used
`together with the images of the field edges to determine
`a mean displacement.
`
`2.5. Stereatactic beam data measurements
`
`A commercial hardware and software system (BrainScan
`V3.5, BrainLAB GmbH, Germany) is used for conven—
`tional, stereotactic are planning of intracranial
`lesions.
`Circular collimators are utilised which are diverging, 9
`cm deep blocks constructed from lead and positioned so
`that the base of the collimator is 27 cm from the isoeentre.
`
`For dose calculation, a standard algorithm based upon
`measurements of tissue maximum ratios (TMR), off-axis
`ratios (OAR) and relative output factors is used [28,35]. A
`software module has been added to this system for non—
`coplanar,
`irregularly shaped, conformal fixed~field plan—
`ning and dose calculation with the m3. The calculation
`model also uses the above beam data parameters but
`these are instead measured with circular fields defined by
`the m3. To allow dose distributions and relative output
`factors to be calculated for irregularly shaped fields a
`Clarkson sector integration algorithm [8] has been imple—
`mented. An irregular field shape is divided into a finite
`number of sectors of differing radial extent and the beam
`data parameters relating to each sector are then used to
`calculate dose.
`
`Beam data for twenty—one circular fields defined by the
`m3 (from 10 to 50 mm diameter in 2 mm steps) were
`measured using a scanning water tank and a diamond
`detector. Percentage depth—dose measurements were first
`collected with an SSD = 98.5 cm. These were converted
`
`to tissue maximum ratios using a standard computation [5].
`Relative output factors were measured for each circular
`field at a depth of 5 cm in water, was normalised to dose
`measured in a 10 X 10 cm2 field. Off-axis ratios were
`measured in a direction parallel to the leaf motion (i.e.
`the leaf ends defined the field edges, not their sides) at
`isoeentre and SSD= 92.5 cm. All
`the beam data were
`
`measured with a fixed primary jaw setting of 5 X 5 cm2.
`Fixed jaws were used to avoid undertaking more complex
`calculations for beam output, which will vary both as a
`function of the irregular m3 field area and primary jaw
`field size.
`
`2.6. Quality assurance of the m3 and treatment planning
`system
`
`After the beam data had been transferred to the treatment
`
`planning system various quality assurance procedures Were
`carried out to verify that dose calculation and other aspects
`of the planning and treatment process are performed to the
`required accuracy. An initial test set out to verify a number
`of parameters simultaneously,
`including the accuracy of
`image localisation, three—dimensional shape reconstruction,
`
`set—up and irradiation. To do this an irregularly
`target
`shaped phantom was constructed from a hollow aluminium
`form and Optosil® P dental
`impression material. This
`material combination was used so that the phantom could
`be visualised without distortion both on CT scans and high-
`energy X-ray portal images, acquired on the linear accel»
`erator. The phantom dimensions were approximately 8 cm
`high X 4 cm wide X 1.5 cm thick.
`The phantom shape was solidly fixed to a BrainLAB
`stereotactic head frame and CT scanned with a fiducial
`
`localiser box in place. Images were acquired with scan
`steps and slice thickness of 1.5 nun. After the CT data
`was transferred to the planning system, images were loca—
`lised relative to the stereotactic fiducial co—ordinate frame
`
`and a treatment isoeentre was defined. The outer phantom
`surface was used to define a volume to which forty confor—
`mal m3 fields were shaped using various combinations of
`collimator, gantry and table angle. A 1 mm margin was
`arbitrarily added between the volume and the field edges to
`help with identifying any displacements.
`For plan verification, the head frame and attached phan—
`tom were fixed to the linac treatment table and positioned
`to the planned treatment isoeentre using standard stereo—
`taetie patient positioning procedures. The planned field
`shapes were transferred from the planning system to the
`m3 workstation via a floppy disk. Portal VisionT'“ images
`of each field were then collected for the defined collimator,
`gantry and table angle combinations. Deviations between
`these images and the planned beams—eye Views
`(BEV)
`were then measured.
`
`that the
`it is crucial
`treatment planning,
`For clinical
`planned isodose distributions accurately represent
`the
`actual delivered dose distributions. Therefore,
`relative
`dosimetry measurements were made to verify the spatial
`accuracy of the isodose calculations produced by the plan-
`ning system. Measurements were performed using film and
`a cubic, Solid Water phantom. This was constructed from
`14.8 X 14.8 cm2 sheets of Solid Water, ranging in thickness
`from 1 mm to 1 cm. The cubic shape enabled films to be
`positioned in either the axial, sagittal or coronal planes by
`simply changing the orientation of the phantom. The phan—
`tom geometry was also considered to be more challenging
`than a spherically shaped phantom in terms of dose calcu—
`lation, particularly for beams that are incident obliquely to
`the phantom surface. The phantomv'shape was also simple
`to handle, load films and position for irradiation.
`The cubic phantom was CT scanned within the fiducial
`localiser box, and the image data transferred to the treat-
`ment planning system. A treatment volume was drawn onto
`the images for use as an irradiation target. Several volumes
`were used,
`including spheres and more irregular shapes,
`transcribed from actual patient data files. These were
`placed in various positions within the phantom and six—
`field, non-coplanar plans were
`calculated. Treatment
`geometries were selected so that beams were Well spaced
`in three dimensional space,
`to minimise dose overlap
`
`ViewRay Exhibit 1024
`Page 7 of 15
`
`ViewRay Exhibit 1024
`Page 7 of 15
`
`

`

`V.P. Corgrove et a1. / Radiotherapy and Oncology 50 {1999) 325—336
`
`
`—w—
`
`329
`
`
`
`3'
`
`8 f
`
`6 e
`
`. --
`
`Malt
`2 W
`
`l
`
`n
`l\ :1
`[\fuj J VKJJ
`
`.
`
`~
`
`.
`
`”flux—‘1‘
`
`‘—
`
`4— E.
`4— C
`4—- D
`
`.
`
`
`
`32
`
`0 8
`
`DE
`
`.9m
`.2
`s
`
`C2
`
`“
`
`v—l—‘—I—*—-l—‘—l—-—l-
`
`
`
`-
`
`l ~9—l—-
`
`l—‘—4
`
`'—l—‘
`
`:
`
`-70
`
`-60
`
`-50 -40 -30 -20
`
`10
`0
`-10
`X Axis (mm)
`
`20
`
`30
`
`40
`
`50
`
`60
`
`70
`
`Fig. 2. Transmission through a single bank of leaves that abut with the opposing set 4.5 cm off—axis. Light grey line, A — 19.7 mm from central axis; grey
`line, B central axis; dark grey line. C + 25 mm from central axis; black line, D leaf ends.
`
`regions, while at the same time reflecting practical treat-
`ment arrangements.
`the treatment
`Before irradiation within the phantom,
`isocentre position was marked on each film. This was
`achieved by marking the film with a pin (distant from the
`irradiation area) using the treatment room lasers as a guide.
`The isocentre position was used as the dose normalisation
`point after the film was scanned with the densitometer and
`isodose boundaries calculated. Absolute dose calibration of
`the films proved to be unreliable (measured dose was
`always lower than expected), with variations of up to —
`6% in the measured dose at isocentre This was most notice-
`
`able whenever films were irradiated with oblique fields, or
`when fields were incident on the edge of a film, i.e. the
`field central axis was parallel
`to the film surface. This
`phenomenon will be discussed later in more detail.
`Finally,
`dose was measured
`using
`an Alderson
`RANDO® anthropomorphic head phantom and TLDs.
`This phantom was also fixed inside a stereotactic head
`frame immobilised using self—penetrating head pins.
`It
`was CT scanned in the usual way and the images trans—
`ferred to the treatment planning system. Six—field, non—
`coplanar, conformal treatment plans were then calculated
`using arbitrary target volume shapes ranging in size from
`5—20 cma. Several TLD chips were loaded inside the phan-
`tom at the isocentre position and were irradiated. Care was
`
`take, so that the planned dose distributions were homoge—
`neous over the length of the TLDs.
`
`3. Results
`
`3.]. Leaf transmission and leakage
`
`Fig. 2 plots the relative dose distributions obtained from
`scans across the closed multi—leaves. In this example, one
`bank of leaves was mov

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket