throbber

`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`
`___________________________
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`__________________________
`
`MEDTRONIC, INC., AND MEDTRONIC VASCULAR, INC.
`
`Petitioners,
`
`v.
`
`TELEFLEX LIFE SCIENCES LIMITED,
`
`Patent Owner.
`_____________________________
`
`Case No.: IPR2020-01344
`U.S. Patent No. RE 46,116
`______________________________
`
`PETITIONERS’ REPLY
`
`
`
`
`

`

`TABLE OF CONTENTS
`
`4.
`
`5.
`
`2.
`3.
`
`Page
`I.
`Introduction ...................................................................................................... 1
`Kontos’s Extension Catheter Is Not a Narrow Tube. ...................................... 1
`II.
`III. The Asserted claims are Obvious. ................................................................... 3
`A.
`The Kontos-Ressemann combination teaches the single-incline and
`double-incline side openings of the ʼ116 patent. .................................. 3
`1.
`Replacing Kontos’s funnel with a side opening maximizes the
`usable area in the catheter assembly. .......................................... 4
`Petitioner’s other motivations are not based in hindsight. .......... 7
`Replacing Kontos’s funnel with a side opening would not create
`a problematic gap. ....................................................................... 9
`A POSITA had a reasonable expectation of success when
`replacing Kontos’s funnel with Ressemann’s collar. This
`combination creates a two-incline side opening. ......................11
`Even if a POSITA placed Ressemann’s collar underneath the
`pushrod, the Kontos-Ressemann combination still teaches the
`single-incline and double-incline openings. .............................13
`Claim 32: After replacing the funnel with a side opening, support
`catheter 10 would remain coaxial with the GC. ..................................16
`Claims 25-40, 42, and 44-48: Kontos-Ressemann-Takahashi
`combination teaches the not-more-than-one-French limitation. .........17
`The Kontos-Ressemann-Kataishi combination teaches a double-
`inclined side opening. ..........................................................................19
`IV. Root is Prior Art and Renders the Claims Obvious .......................................23
`V.
`SECONDARY CONSIDERATIONS DO NOT OVERCOME
`OBVIOUSNESS. ...........................................................................................25
`A.
`Teleflex is not entitled to a presumption of nexus. .............................25
`B.
`Teleflex’s secondary consideration evidence all relates to prior art
`features and functionality. ...................................................................27
`1.
`Teleflex’s alleged evidence of copying is actually copying of
`the prior art. ...............................................................................29
`2. Medtronic did not copy GuideLiner. ........................................30
`VI. CONCLUSION ..............................................................................................31
`
`B.
`
`C.
`
`D.
`
`ii
`
`

`

`TABLE OF AUTHORITIES
`
` Page(s)
`
`Cases
`Butamax Advanced Biofuels LLC v. Gevo, Inc.,
`IPR2013-00539, 2015 WL 1263027 (PTAB Mar. 3, 2015) ............................... 23
`Dr. Reddy’s Labs. S.A. v. Indivior UK Ltd.,
`IPR2019-00329, 2019 WL 2361782 (PTAB June 3, 2019) ............................... 23
`Fox Factory, Inc. v. SRAM, LLC,
`944 F.3d 1366 (Fed. Cir. 2019) .......................................................................... 26
`In re Kao,
`639 F.3d 1057 (Fed. Cir. 2011) .................................................................... 29, 30
`Iron Grip Barbell Co. v. USA Sports, Inc.,
`392 F.3d 1317 (Fed. Cir. 2004) .................................................................... 29, 30
`WBIP, LLC v. Kohler Co.,
`829 F.3d 1317 (Fed. Cir. 2016) .......................................................................... 25
`ZUP, LLC v. Nash Mfg., Inc.,
`896 F.3d 1365 (Fed. Cir. 2018) .......................................................................... 28
`
`
`
`iii
`
`

`

`I.
`
`INTRODUCTION
`Patent Owner (“PO”) does not dispute, because it cannot, that Kontos
`
`describes its “support catheter” as a “mini guide catheter.” Ex-1409, 3:40-49. Nor
`
`does PO dispute that Kontos teaches, just like the coaxial guide catheter 12 of the
`
`Teleflex patent, that support catheter 10 includes a short distal lumen (body 12)
`
`coupled to a pushrod (wire 14). Kontos also describes the method of passing the
`
`support catheter “further through the coronary ostia than can guide catheter 38” to
`
`function as a “guide extension catheter” when “extending beyond the distal end of
`
`guide catheter 38” to help prevent a PTCA balloon catheter from bending,
`
`buckling, or kinking en route to treat a lesion. Other than the various side openings,
`
`Kontos teaches each structural limitation of the Challenged Claims. But as
`
`explained herein, the use of the claimed side openings was an obvious modification
`
`from several pieces of prior art—including Ressemann, which the Board already
`
`found anticipated similar claims in other IPRs. The challenged claims in this IPR
`
`are likewise invalid as obvious.
`
`II. KONTOS’S EXTENSION CATHETER IS NOT A NARROW TUBE.
`PO argues that Kontos’s body 12 must be a “narrow tube” because it “is
`
`important for Kontos to provide these functions” of protecting a balloon catheter,
`
`crossing a lesion, or acting as a temporary stent. Patent Owner Response (POR),
`
`Paper 91, 22. PO is wrong. Ex-1806 ¶¶99-106; Ex-1807 ¶¶113-17. Kontos never
`
`1
`
`

`

`states that it is “important” for body 12 to be “narrow” or have a “snug fit.” Ex-
`
`1806 ¶¶101-02; Ex-1801, 52:5-8 (language is silent concerning “snugness”);
`
`Ex-1800, 77:6-11, 78:12-20. In fact, Kontos teaches the opposite, explaining that
`
`the size of body 12 should be suitable “for existing PTCA catheters” and that “[o]f
`
`course, other sizes may be used for other applications.” Ex-1409, 4:46-48, 4:61-
`
`5:2.
`
`PO’s argument is also contradicted by how Kontos actually prevents kinking
`
`of the PTCA catheter. Ex-1806 ¶103; Ex-1807 ¶¶14-27, 113-17. Kontos prevents
`
`kinking by reducing the distance between the distal-most portion of the catheter
`
`assembly and the occlusion. Ex-1806 ¶103; Ex-1807 ¶¶20-21, 117. “[B]ody 12
`
`functions as a guide catheter extension,” by reducing “the gap that PTCA catheter
`
`40 must negotiate without assistance …, [which] lessens considerably the tendency
`
`of the PTCA catheter 40 to bend, buckle or kink.” Ex-1409, 5:49-56. This is
`
`because, by comparison to advancement within the vasculature, significantly less
`
`force is needed to advance the interventional cardiology device (“IVCD”) within
`
`body 12 due to its lubricious inner coating. Ex-1806 ¶103; Ex-1807 ¶¶20-21, 117;
`
`Ex-1813, 52:14-53:8, 72:24-73:14. It is Kontos’s ability to shorten the distance that
`
`the IVCD traverses in the vasculature—especially in regions that are not straight or
`
`heavily calcified—and not the spatial relationship with the IVCD that reduces the
`
`likelihood of kinking. Ex-1806 ¶103; Ex-1807 ¶¶20-21, 117.
`
`2
`
`

`

`This conclusion is buttressed by the real-world experience of PO’s expert.
`
`Dr. Graham testified that he has never had an IVCD kink within a catheter
`
`assembly. Ex-1813, 52:14-53:8. In other words, the closeness of support assembly
`
`10 and the PTCA catheter 40 is irrelevant, as kinking (if it does occur) would not
`
`happen until the IVCD is in the vasculature. Ex-1806 ¶103-04.
`
`Not only is a close-fitting tube unnecessary to prevent kinking of the PTCA
`
`catheter, but, as Dr. Brecker explains, it would be detrimental. Ex-1806 ¶102. If the
`
`relationship is too “snug,” the IVCD can be damaged when loaded into the
`
`extension catheter, especially post-expansion, when the PTCA balloon is retracted
`
`proximally into the extension catheter. Id.
`
`Finally, even if Kontos taught a small body 12 with a “snug fit” to the
`
`IVCD, PO’s argument still fails because it is premised on Kontos’s invention being
`
`limited to fixed-wire catheters. Ex-1800, 76:2-8, 80:12-81:17; Ex-1801, 56:1-10,
`
`60:16-21. Kontos is not so limited, and instead provides that “the device of this
`
`present invention may be used with almost any type of catheter, including over-
`
`the-wire catheters.” Ex-1409, 9:47-50; Ex-1806 ¶¶105-06.
`
`III. THE ASSERTED CLAIMS ARE OBVIOUS.
`A. The Kontos-Ressemann combination teaches the single-incline
`and double-incline side openings of the ʼ116 patent.
`A POSITA was motivated, with a reasonable expectation of success, to
`
`replace Kontos’s proximal funnel with Ressemann’s support collar 2141
`
`3
`
`

`

`(“Ressemann’s collar”). Petition (Pet.), Paper 1, 15-32. Replacing Kontos’s
`
`proximal funnel with Ressemann’s support collar 2141 results in a proximal
`
`opening with two inclines. Pet., 51-53. PO’s entire opposition is based on the
`
`mistaken belief that after adding Ressemann’s collar, a POSITA would make no
`
`further modification to Kontos’s support assembly 10. Even if a POSITA made no
`
`further modifications, though, PO’s argument still fails.
`
`1.
`
`Replacing Kontos’s funnel with a side opening maximizes
`the usable area in the catheter assembly.
`Replacing Kontos’s proximal funnel with a side opening maximizes the
`
`usable real estate within the catheter assembly. Ex-1806 ¶115. This is particularly
`
`important given that catheter assemblies used in PCI procedures are “not much
`
`bigger than a toothpick.” Ex-2137, 199:11-200:16. PO’s expert agrees that
`
`“maximizing the usable real estate inside a guide catheter is extremely important.”
`
`Ex-2145 ¶90; Ex-1801, 102:2-8; Ex-1813, 91:18-92:5. Specifically, PO’s expert
`
`testified that it was “important” to “maximize the inner diameter of the extension
`
`catheter without having to increase the outer diameter of the guide catheter.”
`
`Ex-1801, 102:2-8; Ex-2138 ¶266. And PO’s expert agreed that the use of a
`
`proximal funnel does not maximize the useable “real estate” within the catheter
`
`assembly. Ex-1813, 92:19-24.
`
`
`
`
`
` Ex-1851, 296:23-297:1.
`
`4
`
`

`

`a. Maximizing usable real estate: reducing diameter of GC.
`Replacing Kontos’s funnel with a side opening, reduces the outer diameter
`
`of the extension catheter, which permits the use of a smaller-diameter GC.
`
`
`
`Ex-1405 ¶211. PO does not disagree, but instead argues that a POSITA would not
`
`replace the funnel with a side opening because Kontos was already sized to fit
`
`inside a 6 French GC and that a 5 French GC offered no advantage. POR, 21-22.
`
`Kontos has a 0.065-inch outer diameter at base portion 18 (Ex-1409, Fig. 3),
`
`meaning that to fit inside a 6 French GC having a 0.070-inch inner diameter,
`
`Kontos’s funnel would only have a 0.005 inch profile. Ex-1807 ¶¶120-21. This
`
`would provide no—or significantly limit the—funneling function. Id., ¶121.
`
`Further, even at 0.005 inches, the funnel would “rub” in a 6 French GC, which, as
`
`admitted by PO’s expert, would “hinder the ability to facilitate smooth passage.”
`
`Ex-1801, 115:14-117:7. Thus, replacing Kontos’s funnel with a side opening
`
`would improve the trackability within the GC. Ex-1806 ¶124. Further, a POSITA
`
`5
`
`

`

`was motivated to use a side opening because it would more easily permit
`
`compatibility with a 5 French GC, which has advantages over a 6 French GC.1 Id.,
`
`¶125.
`
`b. Maximizing usable real estate: increasing diameter of
`extension catheter.
`The corollary is also true: a POSITA would be motivated to replace
`
`Kontos’s funnel with a side opening because the inner diameter of body 12 could
`
`then be increased without similarly increasing the diameter of the GC. Ex-1405
`
`¶¶163-64.
`
`
`
`Ex-1405 ¶164. A POSITA was motivated to make this modification, as PO’s
`
`expert acknowledged, because “the larger the available area within a guiding
`
`device, the larger the variety of types, numbers and sizes of therapeutic devices
`
`
`1 For compatibility with a 5 French GC, tube 16 would be molded onto tapered
`
`wire 14. § III.C, infra.
`
`6
`
`

`

`that can be delivered.” Ex-2145 ¶140. Doing so, “ultimately improves the available
`
`treatment options for the patient.” Id.; Ex-1813, 132:22-133:10.
`
`Despite its own expert agreeing to the advantages of increasing the diameter
`
`of the extension catheter without causing a commensurate increase in the GC, PO
`
`argues that a POSITA would not apply this teaching because Kontos requires a
`
`close fitting. POR, 22. But for the reasons discussed above, Kontos does not
`
`mandate such a tight fit. § II, supra.
`
`PO also argues that it would not be possible to increase the diameter of body
`
`12, without also increasing the diameter of the GC, due to Kontos’s raised marker
`
`bands. POR, 26-27. Kontos teaches, however, that “[o]f course, numerous other
`
`methods for disposing marker band 30 at distal end 24 will be readily apparent to
`
`those skilled in the art.” Ex-1409, 4:21-24. Embedded marker bands were routine
`
`in the art. Ex-1806 ¶119; Ex-1762, 56:9-23. PO’s expert agreed. Ex-1801, 66:19-
`
`67:6, 70:5-17, 75:15-19. It was well within the skill of a POSITA to recess
`
`Kontos’s marker bands. Ex-1807 ¶¶128-36. PO cannot defeat Petitioners showing
`
`that a POSITA would replace Kontos’s funnel with Ressemann’s collar.
`
`Petitioner’s other motivations are not based in hindsight.
`2.
`A POSITA was also motivated to replace Kontos’s funnel with Ressemann’s
`
`collar, as it would increase the area of entry—by more than five-fold—into the side
`
`opening. Ex-1807 ¶141. PO’s expert agreed:
`
`7
`
`

`

`Even though [Kontos is] funneled, the opening is perpendicular …. And
`I’ve have agreed on record that an angled opening gives you a greater
`surface area than a perpendicular one …. With the -- the combination,
`using the -- the angled opening at 2141 tab from Ressemann compared
`to what Kontos has originally described … would increase the opening
`area into … Kontos. It would increase the area.
`
`Ex-1801, 120:16-121:13.
`
`Further, Ressemann teaches that “[t]he proximal end 2140a of the
`
`evacuation lumen 2140 is preferably angled to facilitate smooth passage of other
`
`therapeutic devices through the evacuation lumen 2140 of the evacuation head
`
`2132.” Ex-1408, 23:17–20; see also id., 6:52-60. Despite this clear teaching, PO
`
`argues that side openings actually cause device “catching/hang up issues.” POR,
`
`16-17. This argument, though, is contradicted by a patent (“Keith”) issued to PO’s
`
`own expert, which contains Ressemann’s verbatim disclosure of side openings
`
`“facilitate[ing] smoother passage” of IVCDs. Ex-1123, 7:54-60; Ex-1800, 149:3-
`
`10. PO cannot maintain its argument that a side opening causes device hang-up,
`
`given its own expert obtained a patent reciting the benefits of side openings
`
`allowing smoother passage of interventional cardiology devices.
`
`Side openings also allow for smoother passage of the catheter assembly as it
`
`navigates the vasculature. Pet., 22. PO argues, though, that this teaching is
`
`“unsupported by the evidence.” POR, 18. PO is wrong. Not only does Ressemann
`
`explicitly state that a side opening “allow[s] for smoother passage of the
`
`8
`
`

`

`evacuation sheath assembly 100 through a guide catheter,” but an identical
`
`teaching is also recited in Keith. Ex-1123, 7:54-60.
`
`
`
`
`
` Ex-1819 ¶112.
`
`Finally, a side opening permits smooth re-entry when the extension catheter
`
`is extended beyond the distal end of the GC. Pet., 23-24. PO’s expert previously
`
`found himself in that situation (i.e., passed an extension catheter entirely outside of
`
`the GC),2 Ex-1801, 72:10-73:18, and agreed that a proximal opening, as opposed
`
`to Kontos’s funnel, permits smoother re-entry of the extension catheter. Ex-1801,
`
`79:14-21; Ex-1800, 151:17-152:1.
`
`3.
`
`Replacing Kontos’s funnel with a side opening would not
`create a problematic gap.
`Despite the clear teachings for why a POSITA would replace Kontos’s
`
`funnel with a side opening, PO argues that doing so would actually cause
`
`problems. POR, 13-18. First, PO’s argument fails because it is based on a false
`
`premise that a POSITA would not maximize the usable real estate in the catheter
`
`
`2 PO is incorrect to argue that “Kontos teaches that generally it is not desirable to
`
`extend the proximal end of body 12 beyond the distal end of the guide catheter.”
`
`POR, 24. PO ignores Figure 7, which teaches the intentional advancement into the
`
`vasculature. Ex-1409, 6:19-28, Fig. 7.
`
`9
`
`

`

`assembly. As explained above, a POSITA would have done so. § III.A.1, supra.
`
`Even assuming a POSITA made no further modifications after replacing the funnel
`
`with a side opening, Kontos would not create a “problem gap” as alleged by PO.
`
`If Kontos was used with a 6 French GC, the gap between the support
`
`assembly’s outer wall and the inner wall of the GC would be 0.005 inches. Given
`
`that a guidewire or the distal-most wire of a fixed-wire balloon typically have a
`
`0.014 inch diameter, the PTCA catheter would not catch in the gap between tube
`
`16 and the inner wall of the GC. Ex-1806 ¶¶ 120-25.
`
`
`
`Ex-1409, Fig. 1 (modified by Petitioner).
`
`Ressemann and Keith buttress this conclusion. Ex-1806 ¶124. Both
`
`references teach side openings that are “preferably angled” to “facilitate smoother
`
`passage of other therapeutic devices,” and both teach a gap of at least 0.009 inches
`
`as shown below. Id., ¶125. This is almost double Kontos’s alleged “problematic”
`
`gap. Id.
`
`10
`
`

`

`
`
`Ex-1123, Fig. 1A (annotations added). For these additional reasons, a POSITA
`
`would replace Kontos’s funnel with Ressemann’s collar. Ex-1806 ¶125.
`
`4.
`
`A POSITA had a reasonable expectation of success when
`replacing Kontos’s funnel with Ressemann’s collar. This
`combination creates a two-incline side opening.
`PO also argues no reasonable expectation of success to integrate
`
`Ressemann’s collar with Kontos. POR, 41-42. Specifically, PO argues that the
`
`proposed modification would (i) create a catch point and (ii) not be
`
`“straightforward.” Id. Both arguments fail.
`
`In support of its argument that the proposed modification would create a
`
`catch-point, PO presents the following schematic.
`
`11
`
`

`

`
`
`Id., 41. It was standard practice in the art—including, as explicitly taught by
`
`Ressemann (Ex-1408, 7:28-39, 24:2-11)—to encase Ressemann’s collar in a
`
`polymer jacket to remove any catch points. Ex-1807 ¶¶142-44. In so doing, as
`
`shown in the below schematic, the proposed modification would eliminate any
`
`concern about creating a problematic catch-point.
`
`Id.
`
`Further, Ressemann’s collar could have been attached to Kontos’s wire 14,
`
`
`
`12
`
`

`

`as taught by Ressemann, with a thin layer of solvent casting. Ex-1807 ¶¶69-71,
`
`144-45; Ex-1408, 24:2-11. Doing so would have maintained the proximal incline
`
`of Ressemann’s collar. Ex-1807 ¶159. Thus, PO is incorrect to argue no
`
`expectation of success. Accordingly, the Kontos-Ressemann combination teaches
`
`both the single-incline and two-incline proximal openings recited in the ʼ116
`
`patent. Ex-1806 ¶133-39.
`
`
`
`Ex-1409, Fig. 1 (modified by Petitioner).
`
`5.
`
`Even if a POSITA placed Ressemann’s collar underneath
`the pushrod, the Kontos-Ressemann combination still
`teaches the single-incline and double-incline openings.
`PO states that GuideLiner version 1 has three inclines, which means that
`
`Ressemann must also have at least three inclines. Ex-1806 ¶¶86-91. Therefore,
`
`even if Ressemann’s collar was placed underneath Kontos’s pushrod—as alleged
`
`by PO—the Kontos-Ressemann combination still teaches a two-inclined opening.
`
`Ex-1806 ¶¶137-39.
`
`13
`
`

`

`PO’s expert testified that he identifies where one incline ends and another
`
`begins “just sort of by eye,” explaining that for version 1 of GuideLiner, as
`
`identified below, incline 1 has a “relatively steeper quality” to it, and incline 2 has
`
`“a shallower quality” to it, and “somewhere in between there” is incline 3.
`
`Ex-1800, 45:21-46:2. He believes that even shallow inclines are inclines so long as
`
`they can be identified relative to the catheter’s longitudinal axis. Id., 46:21-47:8.
`
`Ex-1122 (color added to arrows for visibility). Similarly, Root testified that the
`
`figure below has two inclines separated by a flat section. Ex-1762, 91:24-94:1.
`
`
`
`14
`
`

`

`Ex-1854 (bottom annotation added to show inclined regions (red) and flat region
`
`
`
`(blue)).
`
`In light of PO’s testimony, Ressemann’s collar must have more than two
`
`inclines. As shown below, Ressemann’s collar has an incline at the proximal end
`
`(“A”), an incline corresponding to the curve between 2121a and 2141b (“C”), and
`
`an incline at the curved portion of tab 2141b (“B”). Ex-1806 ¶¶136-38.
`
`
`
`15
`
`

`

`Therefore, even if a POSITA placed the tab of Ressemann’s collar under
`
`wire 14, the resulting Kontos-Ressemann combination would still have two
`
`inclines. Ex-1806 ¶¶138-39 (identifying B and C).
`
`
`
`Ex-1807 ¶158.
`
`B. Claim 32: After replacing the funnel with a side opening, support
`catheter 10 would remain coaxial with the GC.
`PO next argues that after replacing the funnel with a side opening, the lumen
`
`of tube 16 and the GC would not be coaxial. PO’s argument fails because it
`
`requires an overly-exacting definition of coaxial. Even under PO’s construction,
`
`Kontos is coaxial.
`
`As used in the Teleflex patents, “coaxially aligning” means that “the axis of
`
`the lumen of the guide extension catheter is aligned in the same direction as the
`
`axis of lumen of the guide catheter.” Ex-1806 ¶¶126-28. Under this definition,
`
`Kontos remains coaxial if its proximal funnel is replaced by a side opening. Ex-
`
`1806 ¶¶126-27. PO argues that coaxial means that the lumens must be perfectly
`
`concentric, such that any deviation off axis—as shown below—does not meet the
`
`claim limitation.
`
`16
`
`

`

`
`
`POR, 31. Even if PO is correct, Kontos is still coaxial. Ex-1806 ¶28. Specifically,
`
`Kontos teaches that “tube 16 may be molded directly onto application wire 14.”
`
`Ex-1409, 4:31-32. In such a scenario, there would be no need for receiving hole 34
`
`in base portion 18. Ex-1807 ¶¶131-34. A POSITA would be motivated to taper
`
`wire 14 for attachment to tube 16, because this would optimize Kontos by either
`
`permitting (i) use with a 5 French GC or (ii) a larger diameter for body 12.
`
`§ III.A.1, supra.
`
`C. Claims 25-40, 42, and 44-48: Kontos-Ressemann-Takahashi
`combination teaches the not-more-than-one-French limitation.
`PO does not dispute that Takahashi provides motivation to achieve a 1
`
`French differential to increase back-up support. Ex-1801, 117:14-118:5. PO argues
`
`that this motivation is inapplicable, though, because Kontos “teaches away from
`
`maximizing the inner diameter.” POR, 26. As discussed above, PO is incorrect to
`
`argue that Kontos requires a snug fitting body 12. § II, supra. Thus, a POSITA was
`
`17
`
`

`

`motivated to apply the teachings of Takahashi to increase back-up support. Ex-
`
`1806 ¶¶129-32.
`
`PO also argues that a POSITA would not have an expectation of success to
`
`modify Kontos to achieve the 1 French differential. POR, 41. PO is wrong. A
`
`POSITA knew how to (i) replace the proximal funnel with a side opening and (ii)
`
`recess Kontos’s distal marker bands. § III.A.1.b, supra. Further, Kontos explicitly
`
`teaches that “tube 16 may be molded directly only application wire 14.” Ex-1409,
`
`4:31-32. In so doing, to maximize the usable “real estate” in the catheter assembly,
`
`a POSITA would taper the pushrod for attachment onto the Kontos-Ressemann
`
`combination. Ex-1807 ¶¶143-49. These modifications allow the diameter of tube
`
`16 to be increased (Ex-1409, 4:46-48), and Kontos would achieve the not-more-
`
`than-one-French differential. Ex-1806 ¶¶129-32.
`
`Id., Fig. 1 (modified by Petitioner).
`
`18
`
`

`

`D. The Kontos-Ressemann-Kataishi combination teaches a double-
`inclined side opening.
`A POSITA was motivated to combine the shape of Kataishi’s distal opening
`
`with Ressemann’s proximal opening. Pet., 58-61. Specifically, adding Kataishi’s
`
`shape to Ressemann’s proximal opening would improve loading of interventional
`
`devices and improve crossability of the catheter. Id. PO is incorrect to argue
`
`otherwise.
`
`PO argues that a POSITA would not recognize that the benefit of Kataishi’s
`
`distal opening—“superior loading of thrombus”—applies equally to Ressemann’s
`
`proximal opening for loading interventional cardiology devices. POR, 46. PO is
`
`incorrect, as the art itself and the knowledge of a POSITA in view of that art
`
`shows the benefits of Kataishi’s inclined shape applies to loading both devices
`
`and thrombi through the lumen. PO ignores, for instance, that Ressemann itself
`
`explicitly teaches that both “[t]he proximal and distal ends 140a, 140b of the
`
`evacuation lumen 140 are preferably angled,” and that this shape on either end
`
`“allows for larger deformable particulate matter to pass through the lumen more
`
`smoothly.” Ex-1408, 6:52-60.
`
`19
`
`

`

`
`
`Id., Fig. 1A. The Keith patent—from PO’s own expert—also teaches that the
`
`“proximal and distal ends … are preferably angled … to facilitate smoother
`
`passage of other therapeutic devices.” Ex-1123, 7:54-60 (emphasis added);
`
`Ex-1800, 149:3-18; Ex-1408, Figs. 6E, 6G, 13:57-14:26. Thus, Kataishi’s teaching
`
`that the shape of its distal tip improves that catheter’s ability to suction thrombus
`
`would be understood by a POSITA to also apply to the proximal opening of a
`
`catheter such as Ressemann. Ex-1807 ¶¶150-53; Ex-1806 ¶¶ 140-43. The shape of
`
`Kataishi’s opening provides similar benefits to Ressemann’s opening, and when
`
`implemented in the Kontos-Ressemann combination, provides roughly the same
`
`opening area. Ex-1807 ¶¶158-160.
`
`20
`
`

`

`
`Id., ¶159 (illustrating the larger area of Kataishi’s shape versus Ressemann’s
`
`support collar under wire 14). And an increase in the area of opening makes it
`
`easier to introduce a stent or balloon catheter. Ex-1807 ¶¶150-53; Ex-1806 ¶¶140-
`
`
`
`43.
`
`PO also incorrectly alleges that Kataishi’s “crossability” benefits would not
`
`apply to Ressemann because it does not “see” the vasculature. POR, 46-47. As a
`
`threshold matter, Ressemann (and Keith) teach that both the proximal and distal
`
`ends of evacuation lumen 140 assist in ensuring better crossabilility through the
`
`guide catheter—e.g., “smoother passage of the evacuation sheath assembly 100
`
`through a guide catheter.” Ex-1408, 6:52-57 (emphasis added); Ex-1123, 7:54-60.
`
`A POSITA would recognize Kataishi’s distal shape would also assist devices as
`
`they move through the guide catheter. Indeed, as cited in the Petition, Sakurada
`
`21
`
`

`

`performed a “passing ability test” of Kataishi that measured “[t]he length between
`
`the ostium and the distal tip of the aspiration catheter.” Ex-1455, 300-02. Sakurada
`
`found that Kataishi’s shape had improved crossability as the catheter advanced
`
`along various loops in the vasculature. Id. A POSITA would understand these
`
`benefits apply when navigating around loops in a guide catheter. Ex-1807 ¶152.
`
`After all, the guide extension catheter that goes into the guide catheter must also
`
`come out of the guide catheter, and the proximal opening will “see” the inside of
`
`the guide catheter moving out. Because Kataishi’s shape is being used in a
`
`predictive way to achieve a known and similar benefit, a POSITA appreciates that
`
`modifying Ressemann’s proximal opening to include the shape of Kataishi’s distal
`
`tip was an obvious modification determined through routine experimentation and
`
`optimization. Ex-1807 ¶¶154-60; Ex-1405 ¶¶236-43.
`
`PO further argues that Kataishi’s benefits—that allegedly flow from both its
`
`shape and flexibility of the material—is inapplicable to Ressemann’s collar. PO is
`
`again belied by its own expert witness, who testified that once a shape is provided,
`
`a POSITA would readily use routine engineering to create side openings of several
`
`different materials, some more rigid and some more flexible. Ex-1764, 27:9-17,
`
`30:2-8, 31:16-21.
`
`22
`
`

`

`IV. ROOT IS PRIOR ART AND RENDERS THE CLAIMS OBVIOUS
`The challenged claims cannot claim priority to the May 3, 2006 ’629
`
`application because it does not provide written description support for (1) a side
`
`opening outside of the substantially rigid segment, or (2) “at least two” inclines.
`
`The Board may consider written description arguments when evaluating whether a
`
`reference qualifies as prior art. Dr. Reddy’s Labs. S.A. v. Indivior UK Ltd.,
`
`IPR2019-00329, 2019 WL 2361782, at *12 (PTAB June 3, 2019); Butamax
`
`Advanced Biofuels LLC v. Gevo, Inc., IPR2013-00539, 2015 WL 1263027, at *8-
`
`15 (PTAB Mar. 3, 2015).
`
`The challenged claims require a side opening outside of the substantially
`
`rigid segment. Ex-1401, claims 25, 52; Ex-1442 ¶¶35-40. Contrary to Teleflex’s
`
`contentions, the claims do not cover Figure 4, or any figure or embodiment
`
`described in the specification for that matter. The claims require a “segment
`
`defining a side opening” separate from a “substantially rigid segment,” taught
`
`nowhere in the specification. The specification describes only one rigid portion
`
`that may include a side opening. Id. The ’629 application only ever describes a side
`
`23
`
`

`

`opening as part of the single, rigid portion 20. Ex-1500, 9-10, 16-20, 38-43, Figs.
`
`4, 12-16; Ex-1442 ¶¶41-63.3
`
`Teleflex suggests that one “integrally formed part” may have discrete
`
`segments. POR, 53. But Teleflex’s interpretation eviscerates the plain meaning of
`
`“segment” and neglects the blatant tension between its reissue claims and the
`
`specification. The claims recite a rigid segment and, separately, a segment defining
`
`a side opening. The claims are different than the specification’s disclosure, in
`
`which the rigid segment includes the side opening. Thus, the challenged claims do
`
`not have written description support and cannot claim priority to May 3, 2006.
`
`Medtronic does not contend that the claims must recite a side opening—rather, if
`
`the claims recite a side opening, the specification requires that it be part of the rigid
`
`segment. This is not the case for the challenged claims.
`
`Thus, Root is prior art because the ’116 patent’s effective filing date is
`
`November 1, 2013—well after Root’s publication date on November 8, 2007. And
`
`PO does not dispute that Root renders the challenged claims obvious in grounds IV
`
`and V.
`
`
`3 Teleflex’s attempt to take Dr. Hillstead’s testimony out of context cannot change
`
`the patent’s disclosure. Regardless, Dr. Hillstead testified that the specification
`
`describes side opening only as part of rigid portion 20. Ex-2244, 186:1-187:19.
`
`24
`
`

`

`V.
`
`SECONDARY CONSIDERATIONS DO NOT OVERCOME
`OBVIOUSNESS.
`Teleflex bases its secondary considerations arguments on features known in
`
`the prior art. Teleflex argues that the combination of rapid exchange (“RX”), guide
`
`extension, improved backup support, and a side opening drives GuideLiner’s
`
`success. POR, 78-79. This combination is entirely taught by Itou and Ressemann,
`
`and Kontos includes every element except for a side opening. The earliest RX
`
`GEC in the record is Kontos, followed by Ressemann, and Itou. These devices are
`
`RX, configured to deliver a wide variety of IVCDs, and provide increased backup
`
`support when extended partially past the end of a GEC. Ressemann and Itou also
`
`have a side opening.
`
`A. Teleflex is not entitled to a presumption of nexus.
`Teleflex recites secondary considerations for claim 25 only, which recites a
`
`side opening adapted to receive IVCDs and a one French differential between the
`
`inner diameter of the tubular structure and lumen of the guide catheter. A
`
`rebuttable presumption of nexus exists “when the patentee shows that the asserted
`
`objective evidence is tied to a specific product and that product is the invention
`
`disclosed and claimed in the patent.” WBIP, LLC v. Kohler Co., 829 F.3d 1317,
`
`1329 (Fed. Cir. 2016). Conversely, “when the thing that is commercially successful
`
`is not coextensive with the patented invention…the patentee is not entitled to a
`
`25
`
`

`

`presumption of nexus.” Fox Factory, Inc. v. SRAM, LLC, 944 F.3d 1366, 1373
`
`(Fed. Cir. 2019).
`
`Despite Keith’s declaration, GuideLiner does not practice claim 25 of the
`
`’116 patent. See Ex-2138, App’x B.
`
` Ex-1851, 154:15-155:8
`
`. Further, Teleflex’s own expert previously
`
`
`
`
`
`
`
`testified that “[m]athmatically, those of ordinary skill in the art define one French
`
`as .0131 inches.” Ex-1825 ¶¶43-48 (testifying that

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket