throbber

`
`
`
`
`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`MEDTRONIC, INC., AND MEDTRONIC VASCULAR, INC.,
`Petitioner,
`v.
`TELEFLEX LIFE SCIENCES LIMITED,
`Patent Owner
`
`
`
`DECLARATION OF MICHAEL JONES
`SUBMITTED IN SUPPORT OF PETITIONER’S REPLIES AND
`PETITIONER’S OPPOSITION TO PATENT OWNER’S MOTION TO
`AMEND
`
`IPR2020-01341
`IPR2020-01342
`U.S. Pat. No. 8,142,413
`
`IPR2020-01343
`IPR2020-01344
`U.S. Pat. No. RE 46,116
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 1 of 76
`
`

`

`
`
`TABLE OF CONTENTS
`
`Page
`
`V.
`
`VI.
`
`VI.
`
`
`INTRODUCTION ........................................................................................... 1
`I.
`QUALIFICATIONS ........................................................................................ 1
`II.
`III. LEVEL OF ORDINARY SKILL IN THE ART .............................................. 3
`IV. MECHANISM OF BACK-UP FORCE .......................................................... 3
`A. A Catheter-in-Catheter Assembly Decreases the Distance that
`an Interventional Cardiology Device Travels in the Vasculature. ......... 7
`A Catheter-in-Catheter Assembly Increases the Moment-of-
`Inertia, Which is an Increase in the Stiffness of the Assembly. ............ 8
`ITOU IS SUFFICIENTLY SIZED TO RECEIVE A BALLOON OR
`STENT CATHETER ....................................................................................... 9
`ITOU DESCRIBES A SYSTEM IN WHICH THE SUCTION
`CATHETER REMAINS SEATED IN RESPONSE TO AN
`OPPOSING BACKWARD FORCE EXERTED BY THE
`PROTECTIVE CATHETER 5 ......................................................................16
`VII. A POSITA WOULD UNDERSTAND THAT THE ENTIRE
`INTERIOR LUMEN OF ITOU’S CATHETER IS LUBRICOUS
`AND NOT TACKY .......................................................................................18
`ITOU IN COMBINATION WITH RESSEMANN .......................................20
`A.
`Rationale for Modifying Itou with Collar 2141 ..................................20
`1.
`Use of flexibility transition members in the art ........................30
`VIII. MODIFICATION OF ITOU WITH KATAISHI ...........................................35
`IX. KONTOS .......................................................................................................41
`A. Kontos Resists Axial and Shear Forces that Would Otherwise
`Tend to Dislodge the Guide Catheter. .................................................43
`Kontos Does Not Require a Narrow Tube 16 to Protect the
`PTCA Catheter Contained Therein. ....................................................46
`After Replacing the Funnel with a Side Opening, Kontos’s
`Support Catheter 10 could be made bigger .........................................49
`D. Modifying Kontos with Ressemann’s support collar ..........................58
`E. Modifying Kontos with Kataishi’s shape ............................................66
`
`B.
`
`B.
`
`C.
`
`
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 2 of 76
`
`

`

`
`
`X. GIVEN RESSEMANN’S SUPPORT COLLAR 2141, AND THE
`SHAPE OF KATAISHI’S DESIGN, THE ULTIMATE PROXIMAL
`SIDE OPENING SHAPE IS AN OBVIOUS DESIGN CHOICE
`DETERMINED THROUGH ROUTINE EXPERIMENTATION AND
`OPTIMIZATION ...........................................................................................68
`
`
`
`
`
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 3 of 76
`
`

`

`
`
`I.
`
`Introduction
`I have been retained by Robins Kaplan LLP on behalf of Medtronic,
`1.
`
`Inc. and Medtronic Vascular, Inc. (“Medtronic”) as an independent expert.
`
`2.
`
`I am informed that Medtronic intends to use my opinions in support of
`
`petitions pending at the Patent Trial and Appeal Board (“PTAB”) concerning U.S.
`
`Patent Nos. 8,142,413, and RE46,116 (the “Teleflex Patents”): IPR2020-01341,
`
`IPR2020-01342, IPR2020-01343, IPR2020-01344.
`
`3.
`
`I have been asked to provide my opinion on the disclosures of certain
`
`patents and articles in view of the knowledge of a POSITA as of the date of
`
`invention of the Teleflex Patents.
`
`4.
`
`I am informed that the parties dispute the date of invention of the
`
`patents that are the subject of the IPRs. My opinions herein are the same whether
`
`the relevant date is May 3, 2006 or early 2005.
`
`5.
`
`I make this Declaration based on my personal education, experience,
`
`and knowledge of medical device product development since the mid-1980s. I
`
`have also reviewed the materials cited herein.
`
`II. Qualifications
`6. My curriculum vitae is attached as Appendix A.
`
`7.
`
`I received a Bachelor of Science degree in Chemical Engineering
`
`from the University of California, Berkeley in 1984. I later earned a Master of
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 4 of 76
`
`

`

`
`
`Science degree in Engineering, Mechanics and Materials, in 1995 from California
`
`State University, Long Beach.
`
`8.
`
`From the mid-1980s through the late-1990s I worked as an engineer
`
`responsible for various aspects of medical device design, development, and
`
`manufacturing for products that included catheters for treating thrombo-embolic
`
`stroke, catheters for treating subarachnoid hemorrhage, balloon catheters for
`
`treating vascular spasm (Micro Therapeutics, Inc.), an electrosurgical balloon
`
`catheter and a ureteral stent product line (Applied Medical Resources), urological
`
`balloon catheters (Baxter Healthcare), embolectomy catheters, endo tracheal and
`
`tracheostomy tubes, and angioplasty catheters (Shiley).
`
`9.
`
`Since 1997, I have worked at Design Development and Fabrication,
`
`Inc. in contract mechanical engineering, focused on product design and
`
`development. My projects included prototype designs for minimally invasive,
`
`surgical, and implantable devices, development and fabrication for surgical
`
`devices, test fixture design and fabrication, and process validation in support of
`
`PMA filings. Clients included: Edwards Lifesciences Inc., SenoRx Inc, Vascular
`
`Control Systems Inc., Micro Therapeutics and Neuroperfusion.
`
`10. Over the course of my work, I have been named as an inventor on
`
`over 100 patents.
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 5 of 76
`
`

`

`
`
`11.
`
`I am being compensated at $425 per hour for my time, and my
`
`compensation is not contingent on the results of these proceedings.
`
`III. Level of Ordinary Skill in the Art
`I understand that Medtronic proposed the following level of ordinary
`12.
`
`skill in the art for these IPRs: a person of ordinary skill in the art (“POSITA”) at
`
`the time of the alleged invention would have had: (a) a medical degree; (b)
`
`completed a coronary intervention training program; and (c) experience working as
`
`an interventional cardiologist. Alternatively, a POSITA would have: (a) an
`
`undergraduate degree in engineering, such as mechanical or biomedical
`
`engineering; and (b) three years of experience designing medical devices,
`
`including catheters or catheter-deployable devices. Extensive experience and
`
`technical training might substitute for education, and advanced degrees might
`
`substitute for experience. Additionally, a POSITA with a medical degree may have
`
`access to a POSITA with an engineering degree, and one with an engineering
`
`degree might have access to one with a medical degree.
`
`13. Based on my education and experience, I meet the definition of a
`
`POSITA. My opinions herein are provided with respect to this definition.
`
`IV. Mechanism of Back-up Force
` I have been asked to opine on the mechanism by which a
`14.
`
`catheter-in-catheter assembly provides additional back-up support for deployment
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 6 of 76
`
`

`

`
`
`of devices, such as guidewires, balloon catheters, stents, and stent catheters, to a
`
`lesion located in a coronary artery.
`
`15. By way of background, tough lesions located deep within the
`
`coronary arteries are difficult to reach. One challenge to treating such distal lesions
`
`is the problem of guide catheters dislodging from the coronary ostium when a
`
`guidewire, balloon catheter, stent or stent catheter meets a particularly tough distal
`
`lesion. This is shown schematically in the figure below. Image “a” illustrates by
`
`arrows the force pushing the guide catheter away from the ostium and coronary
`
`artery due to a guidewire engaging a tough stenosis, for example.
`
`Ex-1814 (Takahiko Suzuki, M.D.)
`http://cct.gr.jp/2003/books/bookspdf/suzuki05cii.pdf
`16. Prior to the 2005-2006 timeframe, a known solution to prevent
`
`back-out of the guide catheter was the use of a “double-catheter,” sometimes
`
`
`
`referred to as a “catheter-in-catheter assembly.” By way of the example provided in
`
`image “b” above, Dr. Takahiko Suzuki, M.D. of Toyohashi Heart Center, describes
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 7 of 76
`
`

`

`
`
`using a 120 cm 6 Fr Goodman Road-Master™ catheter in a 8 Fr guide catheter “to
`
`make a ‘double-catheter’” to achieve the “back-up force needed to cross [a] wire”
`
`in an ostial RCA or LCA lesion.” Ex-1814.
`
`17. Similarly, in a 2004 article, Takahashi quantitatively illustrates the
`
`increased back-up support of a catheter-in-catheter system. See Ex-10101. In
`
`particular, Takahashi describes achieving the backup support of an 8 Fr guide
`
`catheter with a 5 Fr-in-6 Fr system. Id. A figure of Takahashi’s catheter-in-catheter
`
`assembly deploying a balloon catheter is shown below.
`
`
`1 I cite the references I discuss in this declaration according to the way they are
`
`numbered in IPR2020-01341 (e.g., Ex-1010.). However, I understand that the
`
`same set of exhibits were filed in each of the four IPRs concerning the two Root
`
`patents, always retaining the same final two numbers. There may be exceptions to
`
`this general rule, where slightly different versions of these documents were filed in
`
`each of the IPRs, although there is much duplication of content. When I cite these,
`
`I indicate the respective IPR number.
`
`
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 8 of 76
`
`

`

`
`
`Id., Fig. 2.
`18. Suzuki and Takahashi disclose a catheter-in-catheter system that
`
`
`
`improves back-up support. Takahashi teaches that the distance between the inner
`
`diameters of the nested catheters is less than 1 French. Ex-1010 (disclosing an
`
`inner diameter of the outer 6 Fr catheter of 0.071 inches and inner diameter of the
`
`inner 5 Fr catheter of 0.059 inches—a difference of 0.012 inches (or 0.30 mm,
`
`which is less than 0.33 mm (1/3 mm = 1 Fr)). Suzuki teaches improved back-up
`
`support without necessitating a 1 French relationship between the inner diameters
`
`of the nested catheters (i.e., describing a 6Fr in 8Fr system).
`
`19. As shown in Suzuki, a catheter-in-catheter assembly can provide
`
`back-up support even if the distance between the inner diameters of the nested
`
`catheter is greater than 1 French. This is for two reasons. First, because the inner
`
`catheter extends beyond the ostium and into the coronary artery, the
`
`catheter-in-catheter assembly decreases the distance that an interventional
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 9 of 76
`
`

`

`
`
`cardiology device travels in the vasculature. Second, by making a “double
`
`catheter,” the catheter-in-catheter assembly is stiffened, which leads to additional
`
`back-up support.
`
`A. A Catheter-in-Catheter Assembly Decreases the Distance that an
`Interventional Cardiology Device Travels in the Vasculature.
` By using a catheter-in-catheter assembly, it is possible to extend the
`
`20.
`
`inner catheter (sometime referred to as the “child catheter”) beyond the ostium of
`
`the coronary artery. In so doing, the interventional cardiology device has to travel a
`
`shorter distance in the vasculature, which in turn reduces the amount of force
`
`necessary to advance the interventional cardiology device to the target location.
`
`This is because the vasculature can be tortuous and/or calcified, thereby requiring
`
`more force to advance the interventional cardiology device.
`
`21. As discussed above, catheters are generally lined with a lubricious
`
`coating, such as Teflon. See Ex-1015, 548. This lining is intended to promote
`
`trackability of an interventional cardiology device within the catheter assembly,
`
`and reduces the amount of force necessary to advance the interventional cardiology
`
`device to the target location. This, in turn, results in reducing the axial and shear
`
`forces exerted by an interventional cardiology device that would otherwise tend to
`
`dislodge from the ostium of a coronary artery the distal portion of a catheter-in-
`
`catheter assembly.
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 10 of 76
`
`

`

`
`
`B. A Catheter-in-Catheter Assembly Increases the Moment-of-
`Inertia, Which is an Increase in the Stiffness of the Assembly.
`22. The mechanisms involved in the movement of catheters while in the
`
`vasculature explains how a catheter-in-catheter assembly provides back-up
`
`support. In particular, the following mechanical properties of catheters influence
`
`back-up support: flexural rigidity, torsional rigidity, and buckling force. See
`
`Ex-1834, Figs. 7, 14, 11:45-48, 18:26-32. These mechanical properties are
`
`proportional to the moment of intertia, and the construction material. Id.
`
` The moment of inertia, 𝐼𝐼0, of the cross section of a catheter with
`
`23.
`
`respect to its neutral axis, is expressed by the following, and is proportional to
`
`flexural rigidity and buckling force. Id. at 12. The cross section of a catheter is a
`
`hollow circle with an outer diameter D and an inner diameter d. Id.
`
`𝐼𝐼0=𝜋𝜋(𝐷𝐷4−𝑑𝑑4)
`64
`
`
`
`24. The polar moment of inertia of a catheter shaft with outer diameter D
`
`and inner diameter d is expressed as follows, and is proportional to the torsional
`
`
`
`rigidity of a catheter. Id., 12-15.
`
`𝐽𝐽0𝑐𝑐𝑐𝑐𝑐𝑐ℎ𝑒𝑒𝑐𝑐𝑒𝑒𝑒𝑒=𝜋𝜋(𝐷𝐷4−𝑑𝑑4)
`32
`
`
`
`25. From these equations it is clear that when placing a catheter within a
`
`catheter, the outer diameter remains defined and the effective inner diameter
`
`decreases, which results in the following consequences:
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 11 of 76
`
`

`

`
`
`a.
`
`b.
`
`c.
`
`Flexural rigidity increases;
`
`Torsional rigidity increases; and
`
`Resistance to buckling force increases.
`
`26. The polar moment of inertia of a catheter shaft with outer diameter D
`
`and inner diameter d is expressed as follows, and is proportional to the torsional
`
`rigidity of a catheter. Id., 12-15.
`
`27. The increase in these properties is why backup support is increased
`
`when a catheter-in-catheter system (whether a full-length catheter in a full-length
`
`catheter or rapid-exchange catheter within a full-length catheter) is used instead of
`
`a single-guide catheter. It also explains why Suzuki—despite not describing a 1
`
`French relationship—provides back-up support.
`
`V.
`
`Itou is Sufficiently Sized to Receive a Balloon or Stent Catheter
`I have been asked to opine on whether suction catheter (2) of Itou is
`28.
`
`sufficiently sized to receive a balloon or stent catheter. In connection with my
`
`analysis, I have reviewed the position of Medtronic on this limited topic, as well as
`
`Patent Owner’s position as articulated in Patent Owner’s response and the
`
`declaration of Peter T. Keith. See Paper 1, 29-31; Paper , Paper 23
`
`(IPR2020-00135), 37-40; Ex-2138, ¶¶ 170-73; IPR2020-01343, Paper 21, 33-38.
`
`29. With reference to Figures 1 and 3 of Itou, Itou teaches an assembly
`
`consisting of a guidewire (6), which fits within a distal end protective catheter (5).
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 12 of 76
`
`

`

`
`
`Catheter (5) fits within suction catheter (2), and catheter (2) fits within guide
`
`catheter (1).
`
`Ex-1007, Figs. 1A-F.
`
`
`
`
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 13 of 76
`
`

`

`
`
`Id., Fig. 3 (illustrating suction catheter 2, color added).
`30. Table 1 of Itou teaches that the inner diameter of catheter 2 is 1.5 mm,
`
`which is 0.059 inches. Id., 7:55-65. In the 2005 period, there were commercially
`
`available balloon and stent catheters that could be delivered through a guide
`
`catheter that had an inner diameter (“ID”) of >0.056 inches. One example is the 5
`
`Fr Heartrail catheter that has an inner diameter of 0.059 inches. Ex-1010, 452.
`
`Such an inner diameter can “accept normal balloons or stent delivery systems less
`
`than 4.0 mm in diameter.” Id.
`
`31.
`
`I am aware that Itou teaches that catheter 5 is inserted into catheter 2,
`
`and the catheters are together advanced through guiding catheter 1. Ex-1007, 5:29-
`
`51. I am also aware that Itou teaches that suction catheter 5 should then be
`
`removed from catheter 2. Id., 7:13-19.
`
`32.
`
`I am also aware that Itou teaches that when the distal part of the distal
`
`tubular portion (24) is extended into the coronary artery, the proximal part of
`
`tubular portion (24) remains within the guide catheter. Id., 5:35-42.
`
`33.
`
`I understand Patent Owner has argued that the “effective” size of the
`
`opening of catheter 2 is smaller than Itou discloses. See Paper 23, 44-46; Ex-2138
`
`(IPR2020-00135), ¶¶ 170-73; IPR2020-01343, Paper 21, 36-37. Patent Owner
`
`argues that wire-like portion 25 to catheter 2 partially blocks the proximal opening
`
`of catheter 2. Id. Mr. Keith argues that wire 25 reduces the effective size of the
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 14 of 76
`
`

`

`
`
`lumen from 0.059 inches to 0.046 inches. Ex-2138, ¶ 174. Patent Owner argues
`
`that once suction catheter 2 was placed in a guide catheter, it would be impossible
`
`to advance a balloon or stent into catheter 2 for delivery into the coronary
`
`vasculature.
`
`34.
`
`I disagree with Mr. Keith and Patent Owner’s position.
`
`35. First, as reported in Itou’s table 1, the wire-like pushrod, 25, does
`
`have a diameter of 0.45 mm (~0.017 inches). However, the portion of wire 25 with
`
`a full, circumferential cross-sectional shape does not abut the proximal opening of
`
`tubular portion 24. Itou is explicit that, adjacent the location of tubular portion 24’s
`
`proximal opening, pushrod 25 is “crushed into a form of a flat plate.” Ex-1007,
`
`4:32-36, Figs. 3, 4. A change in diameter of pushrod 25 is not unsurprising.
`
`Tapering a shaft along its longitudinal axis in the proximal to distal direction was
`
`known. Ex-1063, [0025].
`
`36. Second, catheter (2) bends. It must, in order for its distal tubular
`
`portion to be maneuverable into the coronary vasculature. See, e.g., Ex-1007,
`
`Abstract, Fig. 6. A POSITA would understand Itou’s disclosure to teach that
`
`catheter (2) has varying rigidity along its length, such that its proximal portion is
`
`more rigid, and its distal portion is more flexible. This is not unique. The art taught
`
`that balloon angioplasty catheters should have a decreasing rigidity along their
`
`proximal to distal lengths. Ex-1063, [0025]. The same was taught for embolic
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 15 of 76
`
`

`

`
`
`protection catheters, as I discuss herein, in regard to Ressemann, as well as for
`
`aspiration catheters. Ex-1019, 2:60-65 (“The catheters are provided with varying
`
`flexibility along the length of the shaft, such that they are soft and flexible enough
`
`to be navigated through the vasculature of a patient without causing damage, but
`
`are stiff enough to sustain the axial push required to position the catheter properly
`
`and to sustain the aspiration pressures.”).
`
`37. Third, the proximal opening of catheter (2) is cut at an oblique angle,
`
`as illustrated below.
`
`Ex-1007, Figs. 3 (left), 4.
`38. The three features I discuss above — (a) the flattening of wire (25)
`
`adjacent to the proximal opening of catheter (2), (b) catheter 2’s flexibility, and (c)
`
`the angle of the proximal opening means that the proximal opening of catheter (2)
`
`
`
`would readily accept a balloon catheter or stent catheter once catheter 2 was
`
`positioned within a guide catheter as shown in Fig. 6.
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 16 of 76
`
`

`

`
`
`39.
`
`In order for wire 25 to obstruct Itou in the manner that Patent Owner
`
`and Mr. Keith allege, Itou’s catheter 2 would have to be entirely rigid, which it is
`
`not.
`
`40.
`
`In the alternative, for wire 25 to obstruct Itou in the manner suggested
`
`by Patent Owner and Mr. Keith, catheter 2 would have to be modified such that the
`
`wire 25 is welded to proximal end portion 231 without any flattening of the wire
`
`adjacent the opening. See Ex-1007, 4:32-36. Such a modification would look like
`
`the modified Itou Fig. 3, below.
`
`
`
`Id., Fig. 3 (modified).
`41. This is clearly not what Itou contemplates or teaches. Nor would a
`
`
`
`person of ordinary skill in the art understand Itou to disclose such an obstruction at
`
`the proximal opening.
`
`42. Moreover, even if Itou’s proximal opening was obstructed by wire 25
`
`so that the “effective size” of catheter 2’s opening went from 0.059 inches (1.5mm)
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 17 of 76
`
`

`

`
`
`to 0.046 inches (1.16 mm) as Patent Owner and Mr. Keith allege (it is not), such an
`
`opening is still large enough to receive a standard coronary stent.
`
`43. By the early 2000s, standard coronary stents, guidewires, balloon
`
`catheters, and stent catheters were available with an outer diameter sufficient to
`
`pass through Itou’s allegedly constricted opening of 0.046 inches. See Ex-1015,
`
`641 (“All current slotted tube designs are ‘bare mounted’ on a delivery balloon,
`
`with deflated profiles smaller than 0.040-in. (1mm)…”); Ex-1802, 104 (Genic®
`
`stent with less than 0.9 mm (0.035 inch) profile), 143 (Lunar stent with 0.0382
`
`inch profile), 269 (Spiral Force stent with 0.039 to 0.042 inch profile), 274
`
`(Tsunami stent with 0.95 mm (0.038 inch) profile); see also, Ex-1803, 4:46-56;
`
`Ex-1804, Table 2.
`
`44. Such sized interventional devices would have sufficient clearance to
`
`be received into a proximal diameter opening of 0.046 inches. Even in the case of a
`
`stent with an outer diameter (“OD”) of 0.044 inches, there would be enough
`
`clearance because the size tolerances on the diameter dimensions can be very
`
`precise (less than plus or minus 0.02 mm, which is less than 0.0008 inches) due to
`
`precision extrusions.
`
`45. Current catheter construction technology using PTFE liners with a tie
`
`layer, stainless steel braids and Pebax outer jackets, and run a tolerance of ±.001
`
`inch on the inner diameter and ±.003 inch on the outer diameter of the completed
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 18 of 76
`
`

`

`
`
`catheter body for catheters up to 4.4 mm in outer diameter. These tolerances were
`
`typical in 2005-2006.
`
`46. For the reasons stated above, it is my opinion that the lumen of Itou’s
`
`suction catheter 2 is suitable for receiving a stent catheter once catheter 2 has been
`
`advanced through a guide catheter and has been partially extended from the guide
`
`catheter’s distal end, and when the proximal opening of catheter 2’s tubular portion
`
`is still within the guide catheter.
`
`VI.
`
`Itou Describes a System in Which the Suction Catheter Remains Seated
`in Response to an Opposing Backward Force Exerted by the Protective
`Catheter 5
`Itou describe as system that includes nested devices, as illustrated in
`47.
`
`Itou’s Fig. 5, shown below.
`
`
`
`48.
`
`Itou states:
`
`[T]he suction catheter 2 is disposed in the lumen of the guiding catheter
`1; the distal end of the distal end protective catheter 5 is inserted in the
`lumen of the suction catheter 2; and the guide wire 6 is inserted in the
`lumen of the distal end protective catheter 5.
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 19 of 76
`
`

`

`
`
`Ex-1007, 5:12-17.
`Itou also teaches that protective catheter 5 has a tip that is rounded
`49.
`
`and flexible. Id., 4:52-55; see also Fig. 1E. This design explains how catheter 5
`
`serves a “protective” function. A rounded, flexible tip is designed to not damage
`
`the coronary vasculature.
`
`50.
`
`Itou additionally teaches that protective catheter (5) is longer than
`
`suction catheter 2. Id., Table 1. While catheter 2 is described as 1250 mm in length,
`
`catheter 5 is described as 1320 mm in length. This makes sense, as, in order to be
`
`“protective,” catheter 5 must extend beyond the distal end of catheter 2.
`
`51. As described above in ¶¶ 22-27, supra, having a nested configuration
`
`of a structure within a structure increases the moment of inertia of a catheter
`
`system and increases backup support. In the case of the distal end protective
`
`catheter 5 disposed within suction catheter 2, which is disposed within guiding
`
`catheter 1, when the distal end protective catheter 5 contacts the coronary
`
`vasculature when disposed beyond the distal end of the suction catheter, the
`
`increased backup support found due to the nested nature of the suction catheter 2
`
`within guiding catheter 1, would result in the suction catheter remaining seated in
`
`response to an opposing backward force exerted by the protective catheter 5 when
`
`it came into contact with the vasculature.
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 20 of 76
`
`

`

`
`
`VII. A POSITA Would Understand that the Entire Interior Lumen of Itou’s
`Catheter is Lubricous and not Tacky
`I have been asked to describe a POSITA’s understanding of the
`52.
`
`characteristics of the interior lumen of Itou’s suction catheter 2. Itou discloses that
`
`its guiding catheter 1 includes an inner layer 110 “made of a resin material having
`
`a sliding property such as a fluorocarbon resin represented by PTFE.” Ex-1007,
`
`3:30-33. Itou describes its suction catheter 2 as having a “tubular portion 24 [that]
`
`includes a tubular body portion 21 which in turn includes an inner layer 210 made
`
`of a resin material having a sliding property such as a fluorocarbon resin
`
`represented by PTFE.” Id., 3:50-54.
`
`53. Notably, having a lumen with an inner layer made of a resin having a
`
`sliding property, or what is known as a lubricious property, was common in the art
`
`at the time of the Teleflex patents. For example, it was well known to a POSITA
`
`that catheters used for the transport of other devices or thrombi should include a
`
`lubricious lining in order to reduce friction. See Ex-1015, 548 (discussing a Teflon
`
`liner in a guide catheter).
`
`54. As mentioned above, Itou discloses that suction catheter 2’s tubular
`
`body 24 comprises a body portion 21 that is lined with, e.g., PTFE. See Ex-1007,
`
`Fig. 3, below. Itou is silent on the lining of portion 23 which also comprise tubular
`
`body 24. POSITA would understand that portion 23 would also include a lining
`
`having a sliding or lubricious property.
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 21 of 76
`
`

`

`
`
`55.
`
`Itou discloses both an interventional device (e.g., protective catheter
`
`5) and foreign matter as having to travel within the entire lumen of tubular portion
`
`24. See, e.g., Ex-1007, 7:13-27, 2:39-42. For example, Itou teaches that catheter 5
`
`is withdrawn from catheter 2 after the catheter assembly is positioned within the
`
`coronary vasculature. Id., Fig. 6, 7:16-19. Additionally, the whole point of a
`
`suction catheter is to remove foreign matter from the vasculature. It was known in
`
`the art to reduce friction within a catheter to line the catheter with a material
`
`having a sliding property. See Ex-1015, 548. For this reason, a POSITA would
`
`understand that the inner lining of portion 23 of tubular portion 24 would have the
`
`same sliding properties as the inner lining of body portion 21.
`
`
`
`Id., Fig. 3.
`
`56. Moreover, the manner in which catheters are typically manufactured
`
`would result in a PTFE lining within tubular body portion 21 as well as portion 23.
`
`The manner in which catheters are typically constructed starts with the lining itself.
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 22 of 76
`
`

`

`
`
`For example, first a Teflon liner for the entire length of a catheter would be snugly
`
`fit over a support mandrel with a very thin layer of Pebax fused on the top layer of
`
`the PTFE. Then, over the Teflon, would be fitted the braiding or coiling in the
`
`region to have reinforcement, and the collar from portion 23 would also be fitted
`
`over the Teflon. Then, over the braiding/coiling and collar would be fitted the
`
`catheter’s support structure polymer, for example a Pebax layer. A heat shrinking
`
`process would be carried out to pull all the pieces together into a single cohesive
`
`structure. After heat shrinking the heat shrink tubing is peeled off. For at least these
`
`reasons a POSITA would understand the portion 23 of Itou’s suction catheter to
`
`have a sliding property of for example, PTFE, and not a characteristic that is tacky.
`
`This methodology dates back to the early 2000s.
`
`57. Regardless of the exact material, no POSITA would design a suction
`
`catheter that has a sticky or tacky interior.
`
`VIII. Itou in Combination with Ressemann
`A. Rationale for Modifying Itou with Collar 2141
`I have reviewed Petitioner’s arguments regarding reasons to combine
`58.
`
`the proximal opening of Itou’s catheter 2 with Ressemann’s collar 2141, as well as
`
`Patent Owner’s arguments to the contrary. IPR2020-01343, Paper 1, 64-67; Paper
`
`21, 42-55. I agree with Medtronic that a POSITA would have been motivated to
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 23 of 76
`
`

`

`
`
`make the discussed modification. A schematic showing the modification of Itou
`
`with Ressemann’s support collar 2141 is shown below.
`
`Ex-1007, Fig. 3 (modified with Ressemann’s support collar 2141).
`
`59.
`
`I agree that the modification to Itou with Ressemann’s collar 2141 is
`
`supported by motivations identified by Petitioner, which include:
`
`
`
`1.
`
`2.
`
`to increase the area for receiving a stent and balloon catheter;
`
`the tab portion provides a flexibility transition between the
`
`proximal end of evacuation lumen 140 and shaft 120; and
`
`3.
`
`the support collar reinforces the opening of the lumen.
`
`IPR2020-01343, Paper 1, 64-67.
`
`60. As noted by Medtronic, combining Ressemann’s collar yields at least
`
`two inclines. Notably, Ressemann’s collar has more than two inclines, as shown in
`
`the schematic below at inclines A, B, and C.
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 24 of 76
`
`

`

`
`
`
`Ex-1008, Fig. 16J (annotations showing at least three inclines on the support collar
`
`
`
`at the proximal end (A), the transition from the concave track of the tab portion
`
`and incline (B), and the incline near the distal most portion of the opening (C)).
`
`61. A POSITA would be motivated to modify Itou because the area of
`
`opening provided by the Ressemann collar creates a longer and more gradual
`
`entryway into the lumen of catheter 2. While Itou and Ressemann do not report the
`
`area of each of their angled side openings, these areas can be estimated based on
`
`the figures and dimensions reported in each patent. I compared what the area of the
`
`opening would be based on the inner diameter of Itou’s catheter 2, which is 1.5
`
`mm. Ex-1007, Table 1, 7:60. To compare to Ressemann’s support collar, I scaled
`
`Ressemann’s support collar such that it has the same inner diameter of Itou. Since
`
`Ressemann’s support collar’s inner diameter is ~0.067 inches (1.7mm) (Ex-1008,
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 25 of 76
`
`

`

`
`
`23:4), I scaled Ressemann’s collar down by 12% to achieve the same 1.5 mm inner
`
`diameter.
`
`62. Once Itou’s side opening and Ressemann’s collar are equivalently
`
`scaled, I measured the relative length of each configuration’s side opening. I then
`
`approximated the area of the inner opening by an area of an ellipse. By doing so,
`
`the opening area of Ressemann’s support collar—when scaled to the same inner
`
`diameter of Itou—is over 3 times larger than Itou’s angled opening. A schematic of
`
`my analysis is shown below.
`
`Schematic illustrating scaled comparison of Itou’s collar area as compared to
`
`
`
`Ressemann’s collar area.
`
`
`
`
`
`IPR2020-01344
`
`Medtronic Ex-1807
`Medtronic v. Teleflex
`Page 26 of 76
`
`

`

`
`
`63. Ressemann’s collar provides two benefits that Itou’s existing metal
`
`collar does not. The first, as Mr. Keith acknowledges, is that the Ressemann collar
`
`is thinner than the Itou collar. IPR2020-01343, Ex-2138, ¶ 187 (explaining the wall
`
`thickness of Itou is 0

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket